RESMED Swift LT Nasl Pillows System User Guide

June 7, 2024
ResMed

RESMED logo Swift™ LT
NASAL PILLOWS SYSTEM
User Guide

RESMED Swift LT Nasl Pillows System -

RESMED

Swift™ LT

NASAL PILLOWS SYSTEM
Thank you for choosing the Swift LT.
Rx Only
Intended Use
The Swift LT channels airflow noninvasively to a patient from a positive airway pressure device such as a continuous positive airway pressure (CPAP) or bilevel system.
The Swift LT is:

  •  to be used by adult patients (> 66 lb/30 kg) for whom positive airway pressure has been prescribed
  • intended for single-patient re-use in the home environment and multi-patient re-use in the hospital/institutional environment.

Using the Swift LT
When using the Swift LT with ResMed flow generators that have mask setting options, if available select ‘SWIFT’; otherwise select ‘MIRAGE’ as the mask option.
For a full list of compatible devices for this mask, see the Mask/Device Compatibility List on www.resmed.com on the Products page under Service & Support. If you do not have internet access, please contact your ResMed representative
Notes:

  • The Swift LT is not compatible for use with ResMed AutoSet CS™ 2 and VPAP™ Adapt SV devices.
  • SmartStop may not operate effectively when using Swift LT with some flow generators that have this feature.
  • If you experience nasal dryness or irritation, the use of a humidifier is recommended.

Swift LT mask components

RESMED Swift LT Nasl Pillows System -Swift LT mask

Fitting your Swift LT

RESMED Swift LT Nasl Pillows System - Fitting

Using the Plastic Tube Retainer (Optional)

RESMED Swift LT Nasl Pillows System - Tube Retainer

Disassembling your Swift LT

RESMED Swift LT Nasl Pillows System - Swift LT

Cleaning your Swift LT in the home

Notes:

  •  The nasal pillows system and headgear should be cleaned by handwashing.
  • The headgear may be washed without being disassembled.

WARNING
Do not use aromatic-based solutions or scented oils (eg, eucalyptus or essential oils), bleach, alcohol, or products that smell strongly (eg, citrus) to clean any of the system’s components. Residual vapors from these solutions can be inhaled if not rinsed thoroughly. They may also damage the system, causing cracks.

CAUTION
If any visible deterioration of a system component is apparent (cracking, crazing, tears, etc.), the component should be discarded and replaced.
Daily/After Each Use
Handwash the separated mask components (excluding headgear) by gently rubbing in warm (approx. 86°F/30°C) water using mild soap for one minute. Rinse all components well with drinking quality water and allow them to air dry out of direct sunlight.
Weekly
Handwash the headgear and tube retainer in warm (approx. 86°F/30°C) water using a mild soap. Rinse well and allow them to air dry out of direct sunlight.
Reprocessing your Swift LT between Patients
Only Sleep Lab Mask (SLM) variants of the mask are intended for multi-patient re-use. When using between patients, these masks must be reprocessed according to cleaning and disinfection instructions available on ResMed.com/ downloads/masks.
Storage
Ensure that your Swift LT is thoroughly clean and dry before storing it for any length of time. Store the mask in a cool dry place out of direct sunlight.
Disposal
The Swift LT does not contain any hazardous substances and may be disposed of with your normal household refuse.

Reassembling your Swift LT

RESMED Swift LT Nasl Pillows System - Reassembling

Reassembling the headgear

RESMED Swift LT Nasl Pillows System - Reassembling lt

Troubleshooting

Pillows won’t seal properly, are uncomfortable, or cause red marks

Possible reason Solution

Pillows may have
been fitted incorrectly,| Carefully follow instructions in “Fitting your Swift LT”.
Check headgear is not over-tightened.
incorrectly adjusted, or
the headgear is too tight.| Check ResMed logo on top of pillows is facing
outwards.
Inner wall position of
one/both pillows are incorrect.| Squeeze pillow sides to release the inner wall. The
the correct position is shown in “Reassembling your
Swift LT”.
The pillows are the wrong size. Pillows may be dirty.| Talk to your clinician.
Clean pillows according to instructions in “Cleaning
your Swift LT in the Home”.

Nasal pillows system is too noisy

System is assembled
incorrectly.| Reassemble the Swift LT. Make sure the pillows are
fully inserted into the frame.
---|---

Before using your Swift LT

****WARNINGS
  •  The vent holes must be kept clear.
  •  The mask should not be used unless the CPAP system is turned on and operating properly.
  • Follow all precautions when using supplemental oxygen.
  • Oxygen flow must be turned off when the flow generator is not operating so that unused oxygen does not accumulate within the flow generator enclosure and create a risk of fire.
  • At a fixed flow rate of supplemental oxygen flow, the inhaled oxygen concentration varies, depending on the pressure settings,  patient breathing pattern, mask, point of application, and leak rate.
  • The technical specifications of the mask are provided for your clinician to check that they are compatible with the flow generator. If used outside specification or if used with incompatible devices, the seal and comfort of the mask may not be effective, optimum therapy may not be achieved, and leak, or variation in the rate of leak, may affect the flow generator function.
  •  Discontinue using the Swift LT if you have ANY adverse reaction to the use of the mask, and consult your physician or sleep therapist.
  •  As with all masks, some rebreathing may occur at low CPAP pressures.
  •  Refer to your flow generator manual for details on settings and operational information.
  • Remove all packaging before using the mask.

Technical specifications

Note: The mask system does not contain latex, PVC, or DEHP materials.
Pressure–Flow Curve
Note:
The mask contains passive venting that meets the requirements for protection against rebreathing as specified in ISO 17510-2. As a result of manufacturing variations, the vent flow rate may vary.

Dead Space
Information| Physical dead space is the empty volume of the mask to the end of the swivel. Using the large cushions this is 94 mL.
---|---
Standards| CE designation in accordance with EC directive 93/42/EEC, class IIa. ISO 14971, ISO 13485, ISO 10993-1, and ISO 17510-2.
Therapy
Pressure| 4 to 20 cm H2 O.
Resistance| Drop in pressure measured (nominal) at 50 L/min: 0.6 cm H2 O and at 100 L/min: 2.0 cm H2 O.
Sound| DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO 4871. The A-weighted sound power level of the mask is 25 dBA, with an uncertainty of 3 dBA. The A-weighted sound pressure level of the mask at a distance of 1 m is 17 dBA, with an uncertainty of 3 dBA.
Environmental
Conditions| Operating temperature: +41°F to 104°F (+5°C to +40°C)
Operating humidity: 15%–95% non-condensing
Storage and transport temperature: -4°F to 140°F (-20°C to +60°C)
Storage and transport humidity: up to 95% non-condensing.
Gross
Dimensions| 380 mm (H) x 57 mm (W) x 62 mm (D). Mask fully assembled with short tube assembly – no headgear.

Note: The manufacturer reserves the right to change these specifications without notice.

Symbols

**Caution, consult accompanying documents; LOT Lot number; REF Part number; Temperature limitation; Humidity limitation; Does not contain latex; ** Indicates a Warning or Caution and alerts you to a possible injury or explains special measures for the safe and effective use of the device.

Limited warranty

ResMed Pty Ltd (hereafter ’ResMed’) warrants that your ResMed mask system (including mask frame, cushion, headgear, and tubing) shall be free from defects in material and workmanship from the date of purchase for the period of 90 days or in the case of disposable masks the maximum use period. This warranty is only available to the initial consumer. It is not transferable. If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective product or any of its components.
This limited warranty does not cover: a) any damage caused as a result of improper use, abuse, modification, or alteration of the product; b) repairs carried out by any service organization that has not been expressly authorized by ResMed to perform such repairs; and c) any damage or contamination due to cigarette, pipe, cigar or other smoke.
Warranty is void on products sold, or resold, outside the region of original purchase.
Warranty claims on defective products must be made by the initial consumer at the point of purchase.
This warranty replaces all other expressed or implied warranties, including any implied warranty of merchantability or fitness for a particular purpose. Some regions or states do not allow limitations on how long an implied warranty lasts, so the above limitation may not apply to you.
ResMed shall not be responsible for any incidental or consequential damages claimed to have resulted from the sale, installation, or use of any ResMed product. Some regions or states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation may not apply to you.
This warranty gives you specific legal rights, and you may also have other rights which vary from region to region. For further information on your warranty rights, contact your local ResMed dealer or ResMed office.

RESMED Swift LT Nasl

608657/1 20-06
SWIFT LT
USER
AMER ML

Manufacturer:
ResMed Pty Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia
See www.resmed.com for other ResMed locations worldwide.
Swift LT Nasal Pillows System
Protected by patents: AU 785376, DE 20321468.4, DE 20321469.2, DE 20321470.6, DE
20321471.4, DE 20321472.2, EP 1314445, JP 4354664, NZ 562417, NZ 562418, NZ 562419, NZ
553756, NZ 553757, NZ 553822, NZ 553824, NZ 553825, US 6581594, US 6823865, US 6907882,
US 7159587, US 7318437, US 7341060. Other patents pending.
Protected by design registrations: AU 301943, AU 322986, AU 322987, AU 322988, AU 322989,
AU 322990, AU 322991, AU 322992, AU 322993, AU 323065, AU 323066, AU 323067, AU
323068, AU 323069, AU 323070, AU 323071, CN 200430102469.3, CN 200830141233.9,
CN 200830141234.3, CN 200830141238.1, CN 200830141239.6, CN 200830141242.8, CN
200830141243.2, CN 200830141244.7, EP 966064, JP 1243026, JP 1349550, JP 1351018, JP
1351019, JP 1351020, JP 1351021, JP 1351022, JP 1351023, JP 1351024, JP 1358420, JP
1363098, JP 1363099, JP 1363100, NZ 407161, NZ 411008, NZ 412497, NZ 412498, NZ 412499,
NZ 412500, NZ 412501, NZ 412502, NZ 412503, NZ 412504, NZ 412505, NZ 412506US D552731.
Others designs pending.
Swift, AutoSet CS, and VPAP are trademarks of ResMed Pty Ltd and are registered in U.S. Patent
and Trademark Office.
Velcro is a registered trademark of Velcro Industries B.V.
© 2020 ResMed.

Global leaders in sleep and respiratory medicine
www.resmed.com