GIMA M24118EN Blood Glucose Test Strips Owner’s Manual
- June 7, 2024
- GIMA
Table of Contents
- PRINCIPLE AND INTENDED USE
- COMPOSITION
- STORAGE AND HANDLING
- PRECAUTIONS
- MATERIALS PROVIDED
- MATERIALS REQUIRED BUT NOT PROVIDED
- INSTRUCTIONS FOR USE
- RANGE OF EXPECTED VALUES
- CHECKING THE SYSTEM
- LIMITATIONS
- PERFORMANCE CHARACTERISTICS
- REFERENCES
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Blood Glucose Test Strips
Package Insert
REF 24118
REF 24119
REF 24120
PRINCIPLE AND INTENDED USE
The GIMA Blood Glucose Test Strips are thin strips with a chemical reagent
system. They work with the GIMA Blood Glucose Meter to measure the glucose
concentration in whole blood. Blood is applied to the end tip of the test
strip. The blood is then automatically absorbed into the reaction cell where
the reaction takes place. A transient electrical current is formed during the
reaction which is detected by the meter. The blood glucose concentration is
then calculated based on the electrical current. The result is then shown on
the meter display. The meters are calibrated to display plasma equivalent
results per the recommendation of the International Federation of Clinical
Chemistry and Laboratory Medicine (IFCC).
For in vitro diagnostic use. Test strips are to be used only outside the body
for testing purposes. For self-testing and professional use. GIMA Test Strips
are used by people with diabetes at home and by healthcare professionals for
the quantitative measurement of glucose in capillary whole blood from the
finger, forearm, and palm.
It is also used as an aid in monitoring the effectiveness of diabetes control
programs. Professionals may also test neonatal, arterial and venous blood
samples.
COMPOSITION
Each test strip has reactive and non-reactive chemicals. These chemicals are FAD-dependent Glucose Dehydrogenase < 25 IU, Mediator < 300 g, Buffer, and Non-reactive Ingredient. Each test strip vial contains a drying agent.
STORAGE AND HANDLING
-
Test strips should be stored in their protective vial.
The vial’s cap must be tightly closed to keep the test strips in good working condition. -
Store test strips in a cool, dry place at room temperature, 235 °C (3695 °F).
Store them away from heat and direct sunlight. -
Do not freeze or refrigerate.
-
To ensure accurate results, use the test strips at room temperature.
-
Keep the text side up and blank side down when you insert the strip contact bars into the strip port.
-
Do not store or use the test strips in a humid place such as a bathroom.
-
Do not store the meter, the test strips, or the control solution near bleaches or cleaners that contain bleaches.
-
Do not transfer the test strips to a new vial or any other container.
-
Repeated insertion and removal of a test strip into the meter strip port may result in reading errors.
-
Replace the vial cap immediately after removing a test strip.
-
Use the test strip immediately after removing it from the vial.
-
Do not use your test strips past the unopened expiration date printed on the vial.
Using test strips past the expiration date may produce incorrect test results.
Note: All expiration dates are printed in Year-Month-Date format. -
A new vial of test strips may be used for 6 months after first being opened. Write the opened vial expiration date on the vial label after opening.
PRECAUTIONS
-
For in vitro diagnostic use.
The test strips are to be used only outside the body for testing purposes. -
Do not use test strips after the expiration date that is shown on the vial. Expired test strips may give incorrect blood glucose readings.
-
Do not use test strips that are torn, bent, or damaged in any way.
Do not reuse test strips. -
The sample must only be applied to the tip of the test strip.
Do not apply blood or control solution to the top of the test strip as this may result in an inaccurate reading. -
Discard the vial and any unused strips 6 months after you first open it.
Constant exposure to air may destroy chemicals in the test strip. This damage can cause incorrect readings. -
Keep the test strip vial away from children and animals.
-
Consult your healthcare professional before making any changes in your treatment plan.
-
Dispose of blood samples and materials carefully. Treat all blood samples as if they are infectious materials. Follow proper precautions when disposing of materials.
MATERIALS PROVIDED
- Test Strips
- Package Insert
MATERIALS REQUIRED BUT NOT PROVIDED
- Meter
- Sterile Lancets
- Lancing Device
- Control Solution
INSTRUCTIONS FOR USE
See your User’s Manual for complete instructions for blood sample collection before use.
- Open the cap of the test strip vial only to remove a test strip for testing. Replace the cap immediately to protect the remaining test strips from moisture in the air.
- Run the blood glucose test following the instructions contained in your User’s Manual.
- The blood glucose test result will be shown on the meter display window.
This result should fall within the target range. Your healthcare professional should recommend your target range.
If your blood glucose test results are higher or lower, ask your healthcare professional what to do. Always consult your healthcare professional before making any changes to your treatment plan.
IMPORTANT: GIMA Blood Glucose Monitoring System allows alternative site
testing for forearm and palm testing in addition to fingertip testing.
There are important differences between forearm, palm, and fingertip samples
that you should know.
Important information about forearm and palm glucose testing:
- When blood levels are changing rapidly such as after a meal, insulin dose, or exercise, blood from the fingertips may show these changes more rapidly than blood from other areas.
- Fingertips should be used if testing is within 2 hours of a meal, insulin dose, or exercise and any time you feel glucose levels are changing rapidly.
- You should test with the fingertips anytime there is a concern for hypoglycemia or you suffer from hypoglycemia unawareness.
RANGE OF EXPECTED VALUES
Blood glucose monitoring requires the help of a healthcare professional.
Together you can set your own range of expected blood glucose values, arrange
your testing times, and discuss the meaning of your blood glucose results.
Expected blood glucose levels for people without diabetes1:
Time | Range, mg/dL | Range, mmol/L |
---|---|---|
Fasting and Before Meals | 70 – 100 | 3.9 – 5.6 |
2 Hours After Meal | Less than 140 | Less than 7.8 |
CHECKING THE SYSTEM
Your blood glucose meter must be handled carefully. See your User’s Manual for
detailed instructions for meter care. The quality control test should be used
to check that the meter and test strips are working together properly. Follow
the test procedure in your User’s Manual to run a quality control test. Three
ranges CTRL 0, CTRL 1, and CTRL 2 are shown on the test strip vial label.
Control Solution 1 is sufficient for most all self-testing needs. If you think
your meter or strips may not be working correctly, you may also want to do a
Control Solution 0 or Control Solution 2 test. Contact your dealer for
information on purchasing control solutions.
For confirmation of results, Control Solution 0 tests should fall within the
CTRL 0 range, Control Solution 1 tests should fall within the CTRL 1 range and
Control Solution 2 tests should fall within the CTRL 2 range. When testing
with Control Solution 1, make sure you are matching the results to the CTRL 1
range on the vial label.
CAUTION: If your quality control test result falls outside the control
range shown on the test strip vial, DO NOT use the system to test your blood,
as the system may not be working properly. If you cannot correct the problem,
contact your distributor for help.
LIMITATIONS
- The GIMA meters, GIMA test strips, and other components of the GIMA Blood Glucose Monitoring System have been designed, tested, and proven to work together effectively to provide accurate blood glucose measurements. Do not use components from other brands.
- The GIMA Test Strips are for testing fresh capillary, venous, arterial, or neonatal whole blood. Do not use serum or plasma samples.
- The venous, arterial, and neonatal blood must be obtained and tested by healthcare professionals.
- Blood glucose measurement with venous or arterial whole blood samples must be performed within 15 minutes of sample collection.
- Anticoagulant preservatives such as heparin or EDTA are recommended for best results in using venous or arterial whole blood samples. Use of anticoagulants such as iodoacetate, sodium citrate, or those containing fluoride is not advised.
- Very high (above 70%) and very low (below 10%) hematocrit levels can cause false results. Talk to your healthcare professional to find out your hematocrit level.
- Abnormally high levels of Vitamin C or other reducing substances will produce false high blood glucose measurements.
- Do not use during or soon after xylose absorption testing. Xylose in the blood will cause interference.
- The system is tested to accurately read the measurement of glucose in whole blood within the range of 0.6 – 33.3 mmol/L (10 to 600 mg/dL).
- Fats(Triglycerides up to 3,000 mg/dL or Cholesterol up to 500 mg/dL) have no major effect on blood glucose test results.
- The GIMA Blood Glucose Monitoring System shows to work properly in studies at altitudes up to 10,000 ft (3,048 meters).
- Severely ill persons should not run the glucose test with the GIMA Blood Glucose Monitoring System.
- Blood samples from patients in shock, or with severe dehydration, or from patients in a hyperosmolar state (with or without ketosis) have not been tested and are not recommended for testing with GIMA Glucose Monitoring System.
PERFORMANCE CHARACTERISTICS
The GIMA Blood Glucose Monitoring System complies with the requirements of
ISO15197:2013 (In vitro diagnostic test systems – requirements for blood
glucose monitoring systems for self-testing in managing diabetes mellitus).
The GIMA meter is calibrated by using YSI (Model 2300 STAT PLUS) Glucose
Analyzer reference instrument, which is traceable to the NIST reference
standard.
Repeatability Precision
Ten replicate assays were each run on ten GIMA Blood Glucose Meters.
Heparinized venous blood samples at five concentration levels were used in the
testing. The results provided the following estimates for reproducibility,
precision.
MEAN| 2.37 mmol/L
(42.6 mg/dL)| 4.57 mmol/L
(82.2 mg/dL)| 7.41 mmol/L
(133.3mg/dL)| 11.39 mol/L
(205.1mg/dL)|
18.59 mmol/L
(334.6 mg/dL)
---|---|---|---|---|---
Standard Deviation
mmol/L (mg/dL) or
Coefficient of Variation| 0.069 mmol/L
(1.24 mg/dL)| 0.127 mmol/L
(2.28 mg/dL)| 2.5 %| 2.5 %|
2.6
Intermediate Precision
Ten replicate assays drawn from three strip lots were run on ten GIMA Blood
Glucose Meters.
These tests were run each day for a total of ten days. Control solutions at
three concentration levels were used in the testing. The results provided the
following precision estimates.
#
| MEAN|
Standard Deviation mmol/L (mg/dL) or
Coefficient of Variation (CV)
---|---|---
Strip Lot 1| 2.00 mmol/L (36.0 mg/dL)| 0.051 mmol/L (0.91 mg/dL)
7.12 mmol/L (115.3 mq/dL)| 2.3 %
17.83 mmol/L (321.0 mq/dL)| 2.1 %
Strip Lot 2| 1.99 mmol/L (35.8 mg/dL)| 0.061 mmol/L (1.09 mq/dL)
7.10 mmol/L (115.3 mq/dL)| 2.3 %
17.83 mmol/L (320.9 mq/dL)| 2.0 %
Strip Lot 3| 2.00 mmol/L (36.0 mg/dL)| 0.058 mmol/L (1.05 mq/dL)
7.08 mmol/L (114.7 mq/dL)| 2.7 %
17.77 mmol/L (319.9 mq/dL)| 1.6 %
System Accuracy
A trained technician tested the capillary blood using the GIMA Blood
Glucose Meter (y).
The blood samples were from more than 100 participants.
Capillary blood samples were taken from the fingertip, palm, and forearm.
Fingertip samples from the same subjects were also analyzed with YSI Model
2300 STAT PLUS Glucose Analyzer (x). The results were compared.
Linear Regression Results: GIMA (y) vs. YSI Reference (x)
Sample Site| Slope| Intercept| R| N
Fingertip| 0.9892| -3.2734| 0.9964| 798
Palm| 0.9750| -2.5412| 0.9929| 666
Forearm| 0.9821| 0.4242| 0.9913| 666
Fingertip samples were used for YSI reference measurement.
The sample range was 1.1 to 29.6 mmol/L (19.6 to 534 mg/dL) for GIMA Blood
Glucose Meter testing with blood sampled from fingertip sites. The sample
range was 3.2 to 21.3 mmol/L (57 to 384 mg/dL) for GIMA Blood Glucose Meter
testing with blood sampled from palm and forearm sites.
Fingertip Site: System Accurac Results for Glucose Concentration ≥5.55 mmol/L (100 mg/dL)
Within ± 5 %| Within ± 10 %| Within ± 15 %
335/582 (57.6 %)| 530/582 (91.1 %)| 582/582 (100.0 %)
Fingertip Site: System Accurac Results for Glucose Concentration < 5.55 mmol/L (100 mq/dL)
Within ± 0.28 mmol/L
(±5 mq/dL)| Within ± 0.56 mmoUL
(±10 mg/dL)| Within ± 0.83 mmol/L
(±15 mg/dL)
156/216 (72.2 %)| 214/216 (99.1 %)| 216/216 (100.0 %)
Palm Site: System Accuracy Results for Glucose Concentration .?. 5.55 mmol/L (100 mg/dL)
Within ± 5 %| Within ± 10 %| Within ± 15 %
246/516 (47.7 %)| 447/516 (86.6 %)| 514/516 (99.6 0/0)
Palm Site: System Accuracy Results for Glucose Concentration < 5.55 mmol/L (100 mq/dL)
Within ± 0.28 mmol/L
(±5 mg/dL)| Within ± 0.56 mmoUL
(±10 mg/dL)| Within ± 0.83 mmol/L
(±15 mq/dL)
95/150 (63.3 %)| 146/150 (97.3 %)| 150/150 (100.0 %)
Forearm Site: System Accuracy Results for Glucose Concentration ≥5.55 mmol/L (100 mg/dL)
Within ± 5 %| Within ± 10 %| Within ± 15 %
297/516 (57.6 %)| 469/516 (90.9 %)| 513/516 (99.4 %)
Forearm Site: System Accurac| Results for Glucose Concentration < 5.55 mmoUL
(100 mq/dL)
Within ± 0.28 mmoVL (±5 mq/dL)| Within ± 0.56 mmol/L (±10 mq/dL)| Within ±
0.83 mmol/L (±15 mg/dL)
99/150 (66.0 %)| 141/150 (94.0 %)| 150/150 (100.0 %)
System Accuracy Results for Glucose Concentration between 1.1 mmoUL (19.6
mg/dL)
and 29.7 mmoUL (534 mq/dL)
Within ± 15 % or ± 0.83 mmol/L ( ± 15 m| /dL )
Fingertip Site| Palm Site| Forearm Site
798/798 (100.0 %)| 664/666 (99.7 %)| 663/666 (99.5 %)
Within ± 12 % or ± 0.56 mmol/L ( ± 10 mq/dL )
Fingertip Site
780/798 (97.7 %)
System accuracy according to ISO 15197:2013, >99% of measured glucose values
fall within the minimum acceptable performance criteria.
Consumer Study
A consumer study was performed by testing one test strip lot. Participants
used the GIMA Blood Glucose Monitoring System. This study showed that the
patient can run the test properly.
Linear Regression Results: GIMA (y) vs. YSI Reference (x)
Sample Site| Slope| Intercept| R| N
Fingertip| 0.9865| 0.7919| 0.9904| 100
Palm| 0.9817| 0.2111| 0.9874| 100
Forearm| 1.0152| -2.3603| 0.9888| 100
The study evaluating glucose values from fingertip capillary blood samples
obtained by 100 laypersons showed the following results:
100 % within ± 0.83 mmol/L (± 15 mg/dL) of the medical laboratory values at
glucose concentrations below 5.55 mmol/L (100 mg/dL), and 100 % within ± 15%
of the medical laboratory values at glucose concentrations at or above 5.55
mmol/L (100 mg/dL).
Venous Study
The venous blood glucose measurements from more than 103 participants were
taken by a trained technician using the GIMA Blood Glucose Meter (y). The
venous blood sample from the same subjects was also analyzed with YSI Model
2300 STAT PLUS Glucose Analyzer (x).
The results were compared in the table below.
Linear Regression Results: GIMA (y) vs. YSI Reference (x)
Blood Sample| Slope| Intercept
| R| N
---|---|---|---|---
Venous| 1.0124| 4.5900| 0.9959| 762
Neonatal Study
The neonatal blood glucose measurements from 59 participants were taken by
a trained technician using the GIMA Blood Glucose Meter (y). The neonatal
blood samples from the same subjects were also analyzed with YSI Model 2300
STAT PLUS Glucose Analyzer (x). The results were compared in the table below.
Linear Regression Results: GIMA (y) vs. YSI Reference (x)
Blood Sample | Slope | Intercept | R | N |
---|---|---|---|---|
Neonatal | 1.0054 | 2.5101 | 0.9780 | 354 |
For complete instructions, please refer to the GIMA Blood Glucose Monitoring System User’s manual. For additional questions or issues with this product, please contact your dealer for help.
REFERENCES
- ADA Clinical Practice Recommendations, 2018.
INDEX OF SYMBOLS
| Consult instructions for use| | Use by| | Control Range
---|---|---|---|---|---
| In vitro diagnostic medical device| | Lot Number| | Catalog number
| Temperature limit| | Manufacturer| | Contains sufficient for
| Authorized Representative in the European Community| | Do not reuse| | Model Number
Manufacturer
ACON Laboratories, Inc.
5850 Oberlin Drive, #340 San Diego, CA 92121, USA
www.aconlabs.com
MDSS GmbH
Schiffgraben 41 30175 Hannover,
Alemania
Number: 1151151801
Effective Date: 2020-xx-xx