BRAUN IRT6520 ThermoScan Infrared Ear Thermometer Instructions
- June 6, 2024
- Braun
Table of Contents
- IRT6520 / IRT6520B
- Product description
- Body temperature
- AgeSmart color-coded display
- How does Braun ThermoScan Infrared Ear Thermometer work?
- How to use your Braun ThermoScan
- Memory mode
- Night light feature
- Changing the temperature scale
- Care and cleaning
- Replacing the batteries
- Calibration
- Errors and troubleshooting
- Model and product specifications
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
ThermoScan
Infrared Ear Thermometer
IRT6520
IRT6520B
Certain trademarks are used under license from The Procter & Gamble Company or its affiliates.
© 2021 All rights reserved.
0297
Patents
www.kaz.com/patents/braun
P/N: A002208R1
02FEB21
Braun ThermoScan Infrared Ear Thermometer
Indications for use
The Braun ThermoScan Infrared Ear Thermometer is indicated for the
intermittent measurement and monitoring of human body temperature by consumers
in a home use environment.
IRT6520 / IRT6520B
Product description
IRT6520 / IRT6520B
- Lens Filter (Box of 20; Model: LF20)
- Probe tip
- Probe (Inserted with 1pc of lens filter)
- ExacTemp light
- Lens filter ejector
- Display
- smart and Memory buttons
- Power button
- Start button
- Battery door
- Protective case
The Braun ThermoScan Infrared Ear Thermometer has been carefully developed for
accurate, safe, and fast temperature measurements in the ear.
The shape of the thermometer probe prevents it from being inserted too far
into the ear canal which can hurt the eardrum.
However, as with any thermometer, proper technique is critical to obtaining
accurate temperatures.
Therefore, read the instructions carefully and thoroughly.
The Braun ThermoScan Infrared Ear Thermometer is indicated for intermittent
measurement and monitoring of human body temperature for people of all ages.
It is intended for household use only. Use of IRT6520 along with the smart
feature is not intended as a substitute for consultation with your physician.
WARNINGS AND PRECAUTIONS
• Please consult your doctor if you see symptoms such as unexplained
irritability, vomiting, diarrhea, dehydration, changes in appetite or
activity, seizure, muscle pain, shivering, stiff neck, pain when urinating,
etc. in spite of any color of the background light or absence of fever.
• Please consult your doctor if the thermometer shows an elevated temperature
(yellow or red background lights for IRT6520 / IRT6520B).
The operating ambient temperature range for this thermometer is 10 – 40 °C (50
– 104 °F). Do not expose the thermometer to temperature extremes (below –25 °C
/ –13 °F or over 55 °C / 131 °F) or excessive humidity (> 95 % RH). This
thermometer must only be used with genuine Braun ThermoScan lens filters (LF
40).
To avoid inaccurate measurements always use this thermometer with a new, clean
lens filter attached.
If the thermometer is accidentally used without a lens filter attached, clean
the lens (see «Care and cleaning» section). Keep lens filters out of reach of
children.
This thermometer is intended for household use only. This product is not
intended to diagnose any disease but is a useful screening tool for
temperature. Use of this thermometer is not intended as a substitute for
consultation with your physician.
The smart feature is not intended for pre-term babies or small-for-
gestational-age babies. The smart feature is not intended to interpret
hypothermic temperatures. Do not allow children to take their temperatures
unattended.
Do not modify this equipment without the authorization of the manufacturer.
Parents/guardians should call the pediatrician upon noticing any unusual
sign(s) or symptom(s).
For example, a child who exhibits irritability, vomiting, diarrhea,
dehydration, seizure, changes in appetite or activity, even in the absence of
fever, or who exhibits a low temperature, may still need to receive medical
attention.
Children who are on antibiotics, analgesics, or antipyretics should not be
assessed solely on temperature readings to determine the severity of their
illness.
Temperature elevation as indicated by AgeSmart may signal a serious illness,
especially in adults who are old, frail, have a weakened immune system, or
neonates and infants. Please seek professional advice immediately when there
is a temperature elevation and if you are taking the temperature on:
- Neonates and infants under 3 months (consult your physician immediately if the temperature exceeds 37.4 °C or 99.4 °F)
- Patients over 60 years of age
- Patients having diabetes mellitus or a weakened immune system (e.g. HIV positive, cancer chemotherapy, chronic steroid treatment, splenectomy)
- Patients who are bedridden (e.g. nursing home patient, stroke, chronic illness, recovering from surgery)
- A transplant patient (e.g. liver, heart, lung, kidney) Fever may be blunted or even absent in elderly patients.
This thermometer contains small parts that can be swallowed or produce a choking hazard to children. Always keep the thermometer out of children’s reach.
Body temperature
Normal body temperature is a range. It varies by site of measurement, and it
tends to decrease with age. It also varies from person to person and
fluctuates throughout the day. Therefore, it is important to determine normal
temperature ranges. This is easily done using Braun ThermoScan. Practice
taking temperatures on yourself and healthy family members to determine the
normal temperature range.
Note: When consulting your physician, communicate that the ThermoScan
temperature is a temperature measured in the ear and if possible, note the
individual’s normal ThermoScan temperature range as an additional reference.
AgeSmart color-coded display
Clinical research shows the definition of fever changes as newborns grow into
infants, infants grow into children and children grow into adults 1.
Braun’s AgeSmart color-coded display takes the guesswork out of interpreting
temperature for the entire family.
Simply use the smart button to select the appropriate age setting, take a
temperature and the color-coded display will glow green, yellow, or red to
help you understand the temperature reading.
A green screen suggests no fever, a yellow screen suggests moderate fever, and
red signals high fever.
- Herzog L, Phillips S. Addressing Concerns About Fever. Clinical Pediatrics. 2011; 50(#5): 383-390.
How does Braun ThermoScan Infrared Ear Thermometer work?
Braun ThermoScan Infrared Ear Thermometer measures the infrared heat generated by the eardrum and surrounding tissues. To help avoid inaccurate temperature measurements, the probe tip is warmed to a temperature close to that of the human body. When the thermometer is placed in the ear, it continuously monitors the infrared energy. The measurement is finished and the result is displayed when the thermometer detects that an accurate temperature measurement has been taken.
How to use your Braun ThermoScan
-
Remove the thermometer from the protective case.
-
Push the Power button .
During an internal self-check, the display shows all segments. Then the last temperature taken will be displayed for 5 seconds.
-
The lens filter indicator will blink to signal one is needed.
To achieve accurate measurements, make sure a new, clean lens filter is in place before each measurement.
Attach a new lens filter by pushing the thermometer probe straight into the lens filter inside the box and then pulling it out.
Note: The Braun ThermoScan will not work unless a lens filter is attached.
-
Select age with the AgeSmart button as shown below.
Press the button to toggle through each age group.
NOTE: You must select age in order to take a measurement.
The thermometer is ready to take temperature when the display looks like the image below.
-
Fit the probe snuggly into the ear canal, then push and release the Start button .
-
ExacTemp light will pulse while the temperature is in progress. The light will remain solid for 3 seconds to indicate that a successful temperature reading has been achieved.
NOTE: If the probe has been properly inserted into the ear canal during the measurement, a long beep will sound to signal the completed measurement.
If the probe has NOT been constantly placed in a stable position in the ear canal, a sequence of short beeps will sound, the ExacTemp light will go out and the display will show an error message (POS = position error).
See the «Errors and troubleshooting» section for more information.
-
The confirmation beep indicates that an accurate temperature measurement has been taken. The result is shown on the display.
Age range| Green
Normal temperature| Yellow
Moderate fever| Red
High fever
---|---|---|---
0-3 Months| ≥35.8 – ≤37.4 °C
(≥96.4 – ≤99.4 °F)| | >37.4 °C
(>99.4 °F)3-36 Months| ≥35.4 – ≤37.6 °C
(≥95.7 – ≤99.6 °F)| >37.6 – ≤38.5 °C
(>99.6 – ≤101.3 °F)| >38.5 °C
(>101.3 °F)
36 Months –
adult| ≥35.4 – ≤37.7 °C
(≥95.7 – ≤99.9 °F)| >37.7 – ≤39.4 °C
(>99.9 – ≤103.0 °F)| >39.4 °C
(>103.3 °F) -
For the next measurement, press the Eject button to remove and discard the used lens filter, and put on a new, clean lens filter.
NOTE: Defaults to last age setting used if you do not change.
The Braun ThermoScan Infrared Ear Thermometer turns off automatically after 60 seconds of inactivity.
The thermometer can also be turned off by pressing the Power button .
The display will briefly flash OFF and it will go blank.
Temperature taking hints
Always replace disposable lens filters to maintain accuracy and hygiene. The
right ear measurement may differ from the measurement taken in the left ear.
Therefore, always take the temperature in the same ear.
The ear must be free from obstructions or excess earwax build-up to take an
accurate reading.
External factors may influence ear temperatures, including when an individual
has:
Factor | Yes affects |
---|---|
Poor probe placement | |
Used lens filter | |
Dirty lens |
In the cases below, wait 20 minutes prior to taking a temperature.
Factor | Yes affects |
---|---|
Extreme hot and cold room temperature | |
Hearing aid | |
Lying on pillow |
Use the untreated ear if ear drops or other ear medications have been placed in the ear canal.
Memory mode
-
This model stores the last 9 temperature measurements. To display the stored measurement, the thermometer must be turned on.
-
Press the Mem button.
The display shows the temperature, and when releasing the Mem button, the stored temperature for that memory number is displayed, together with “M”. -
Each time the Mem button is pushed, the stored reading and an M is displayed to indicate each temperature measurement (eg. M2).
Memory mode is automatically exited by not pressing the memory button for 5 seconds.
The last temperature taken is stored in its memory and will be automatically displayed for 5 seconds when it is turned on again.
Night light feature
The thermometer includes a convenient night light to illuminate the display in
a dim environment.
The light will turn on when you press any button. It will stay on until the
thermometer is inactive for up to 10 seconds even after the temperature is
taken.
Changing the temperature scale
This thermometer is shipped with the Celsius (°C) temperature scale activated. If you wish to switch to Fahrenheit (°F) and/or back from Fahrenheit to Celsius, proceed as follows:
- Make sure the thermometer is turned off.
- Press and hold down the Power button . After about 3 seconds the display will show this sequence: °C / SET / °F / SET …
- Release the Power button when the desired temperature scale is shown. There will be a short beep to confirm the new setting, then the thermometer is turned off automatically.
Care and cleaning
The probe tip is the most delicate part of the thermometer.
It must be clean and intact to ensure accurate readings.
If the thermometer is ever accidentally used without a lens filter, clean the
probe tip as follows:
Very gently wipe the surface with a cotton swab or soft cloth moistened with
alcohol. After the alcohol has completely dried out, you can put a new lens
filter on and take a temperature measurement.
If the probe tip is damaged, please contact the local service center.
Use a soft, dry cloth to clean the thermometer display and exterior. Do not
use abrasive cleaners. Never submerge this thermometer in water or any other
liquid. Store thermometer and lens filters in a dry location free from dust
and contamination and away from direct sunlight.
Replacing the batteries
The thermometer is supplied with two 1.5 V type AA (LR 06) batteries.
- Insert new batteries when the battery symbol appears on the display.
- Open the battery compartment. Remove the batteries and replace them with new batteries, making sure the poles are in the right direction.
- Snap battery door into place.
Only discard empty batteries. To protect the environment, dispose of empty batteries at your retail store or at appropriate collection sites according to national or local regulations.
Calibration
The thermometer is initially calibrated at the time of manufacture. If this
thermometer is used according to the use instructions, periodic re-adjustment
is not required. If at any time you question the accuracy of the temperature
measurement, please contact your authorized service center.
The manufacturing date is given by the LOT number located in the battery
compartment.
The first three (3) digits after LOT represent the Julian date that the
product was manufactured and the next two (2) digits represent the last two
numbers of the calendar year the product was manufactured. The last
identifiers are the letters that represent the manufacturer.
An example: LOT 11614K, this product has been manufactured on the 116th day of
the year 2014.
Errors and troubleshooting
Error Message | Situation | Solution |
---|---|---|
No lens filter is attached. | Attach a new, clean lens filter. | |
The thermometer probe was not |
positioned securely in the ear.
An accurate measurement was not possible.
POS = position error| Take care that the positioning of the
probe is correct and remains stable.
Replace lens filter and reposition.
Press the Start button to begin a new
measurement.
| Ambient temperature is not
within
the allowed operating range
(10 – 40 °C or 50 – 104 °F).| Allow the thermometer to remain for
30 minutes in a room where
the temperature is between 10
and 40 °C or 50 and 104 °F.
| Temperature taken is not within
typical human temperature range
(34 – 42.2 °C or 93.2 – 108 °F).
HI = too high
LO = too low| Make sure the probe tip and lens are
clean and a new, clean lens filter is
attached. Make sure the thermometer is properly inserted. Then, take a new
temperature.
| System error – self-check display flashes continuously and will not be
followed by
the ready beep and the ready symbol.
If the error persists,
If the error still persists,| Wait 1 minute until the thermometer
turns off automatically, then turn
on again.
… reset the thermometer by re-moving
the batteries and putting them back in.
… please contact the service center.
| The battery is low, but the thermometer will still operate correctly.|
Insert new batteries.
| The battery is too low to take correct
temperature measurement.| Insert new batteries.
Model and product specifications
Product specifications
Product | ThermoScan I RT6520 | ThermoScan I RT6520B |
---|---|---|
Temperature measurement within seconds | ■ | ■ |
Temperature range °C | 34 – 42.2 | 34 – 42.2 |
Temperature range °F | 93.2 -108.0 | 93.2 -108.0 |
AgeSmart | ■ | ■ |
ExacTemp technology | ■ | ■ |
Feedback system | ■ | ■ |
Ready beep | ■ | ■ |
Small, soft tip | ■ | ■ |
Flexible tip | ■ | ■ |
Memory storage capacity | 9 | 9 |
Auto-off after | 60 sec. | 60 sec. |
Easy-to-read LCD display | ■ | ■ |
Lens filter detector | ■ | ■ |
Lens filter ejector | ■ | ■ |
Protective case with lens filter storage | ■ | ■ |
Batteries included | ■ | ■ |
21 lens filter included | ■ | ■ |
Color | Blue | Black |
Operating ambient temperature range: | 10 – 40 °C | (50 – 104 °F) |
--- | --- | --- |
Storage temperature range: | -25 – 55 °C | (-13 – 131 °F) |
Operating and storage relative humidity: | 10 – 95 %RH (non condensing) | |
Display resolution: | 0.1 °C or °F | |
Accuracy for displayed temperature range | Maximum Laboratory Error | |
35 °C – 42 °C (95 °F – 107.6 °F): | ± 0.2 °C | (± 0.4 °F) |
outside this range: | ± 0.3 °C | (± 0.5 °F) |
Clinical repeatability: | ± 0.2 °C | (± 0.4 °F) |
Weight: | ThermoScan Infrared Ear Thermometer – 66.9g |
Protective case – 43.7g
Battery – 46.4g
Battery life:| 2 years / 350 measurements
Service life:| 5 years
Manufactured by KeyTronic Corporation (Address: Tomas Becket #2220 Parque Ind. Gema, Plant 3 Juarez, Chihuahua C.P. 32380 México) for Kaz Europe Sàrl (Address: Place Chauderon 18, CH-1003 Lausanne, Switzerland)
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>