yuwell 140603 Sphygmomanometer Aneroid User Manual
- June 6, 2024
- yuwell
Table of Contents
yuwell 140603 Sphygmomanometer Aneroid User Manual
Summary
► The product is used for measuring blood pressure.
Main performance indexes
Main performance indexes :
► Testing range: 0-40kPa(0-300mmHg)
► Minimum scale value: 0.5kPa(2mmHg)
► Tolerance: ± 0.5 kPa( ± 3.75mmHg)
Structure
► Measure in a quiet place.
► Measure at the same period every day, because blood pressure of body is
variable all the time.
► Keep clam and never Move or talk when measuring.
► Wear loose clothes.(Figure l)
► Examine the instrument before using. The needle should point to the site of
“O”. Press the ball, the needle reacts nimbly and returns to “O” quickly.
► Put an arm on the table, and keep it comfortable.
► Adjust the arm to the same height with your heart. Use a soft pad if still
can’t reach the same height. (Figure 2)
► Flat the arm-ribbon smoothly, and make the middle of the air-sack to the humours artery. The bottom of the arm-ribbon is far from elbow joint 2-3 centimeters. For accuracy, the arm can’t be bound firmly. ( Figure 3)
► Tuck the breastpiece of a stethoscope into the humours artery of elbow. (Figure 4)
► Pump into the arm-ribbon slowly and listen. When the sounds of pulse
disappear. Pump again until the pressure 2.Skpa-4kpa (20mmHg- 30mmHg).
► After stopping pumping, you can let out the air at the speed of
0.3kPa-0.6kPa (2mmHg – 4mmHg) per second. The needle will drop gradually.
► When you hear the first sound of pulse at the course of drop, the readings
of the needle are called contraction pressure.
► When the high sounds of pulse beat increase turning to noise, the tune
suddenly become muffled and disappear. The readings of the needle are called
expansion pressure.
► Measure blood pressure for 2 or 3 times to get a average value.
► There is a metal hook in the back of shell. You can place the hook on the
arm-ribbon for easily reading. ( Figure 5)
► The nylon button of the arm-ribbon is designed for convenient operation. (Figure 6)
Notice
►The instrument should be bound with arm ribbon after the measurement.
► Do not press the latex hall until hearing the sounds of pulse.
► The measured arm should be at the same height with heart or the result will
be inaccurate.
► Never make the inner membrane exceed spring limitation. When use, the
readings of needle can’t exceed 400kPa(300mmHg).
Maintenance
► The instrument should be keep stably and securely, protect from being topple or impacted.( Figure 7)
► Never fold the rubber hose too long, or fold the arm strap into small pieces, or put it near the fire. Otherwise the rubber will be ageing or cracking earlier than normal, thus shortening useful life.(Figure 8)
► The rubber ball and valve should be stored in the right space inner the box,
so as to protect the glass face from being crashed.
► If the meter is not exactly: O pointed, never attempt to dismantle it by
yourself, it should
► only be repaired and maintained in our technical service departments by
designated maintenance men. Aneroid Sphygmomanometer should be cleaned with
soft fabric neutral cleanser before storage, and kept in a dry place.
Double scales (aneroid) sphygmomanometer conversion table
The minimum value of tube is O.SkPa,
O.SkPa = 3. 75mmHg
JIANGSU YUYUE MEDICAL EQUIPMENT & SUPPLY CO., LTD. Sales Center Address: Huan
yuan East Road No.1 , Xu Zhuang Software Park, Nanjing, Jiangsu Province, P.R.
China, 210000
Manufacturing Address: Yun Yang Industrial Park, Danang, Jiangsu
Province, P.R. China, 212300
http://www.yuwell.com
331600-0A
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>