Yuwell 8F-5A Oxygen Concentrator User Manual
- January 18, 2024
- yuwell
Table of Contents
****
Yuwell 8F-5A Oxygen Concentrator User Manual
Do not operate this unit without first reading and understanding this manual.
WARNING: INDICATION FOR USE
For avoiding power outage or oxygen concentrator operating failure, exigent
oxygen user and serious patient must equip with reserve oxygen supply (for
example: oxygen cylinder, oxygen bag). This device is to be used as an oxygen
supplement. It is not used as life supporting or life sustaining.
WARNING: CONTRAINDICATIONS
The oxygen concentrator is not intended for life supporting or life sustaining
applications, nor does it provide any patient monitoring capabilities.
Symbol | Description |
---|
WARNING
| Describes a hazard or unsafe practice that if not avoided can result in
severe bodily injury, death or property damage.
CAUTION| Describes a hazard or unsafe practice that if not avoided can
result in minor bodily injury or property damage.
BEFORE INSTALLATION
To reduce the risk of electrical shock, DO NOT remove the cabinet. Refer service to qualified service personnel. Before operating the device, read and understand this manual. The concentrator should always be kept upright to prevent cabinet damage while being transported. If the electrical source power voltage is unstable that beyond normal power rang, please add the voltage stabilizer. Please use eligible, safe power set and junction box. Non- professional people DO NOT open oxygen concentrator’s cabine.
LOCATING
- You may select a room in your house where using your oxygen concentrator would be most convenient. Your concentrator can easily move from one room to other room by the wheels.
- Be certain to place the device so all sides are at least 10 cm (4 inches) away from walls, draperies, furniture, or other obstruction. DO NOT place the unit in a confined area.
- Oxygen concentrator must avoid heat source, fire source, wetness, exorbitant or over-low extra ventricular condition.
- Sundries and vessel do not be placed on top of the concentrator.
SAFETY NOTES
NEVER block the air openings of the unit or place it on a soft surface, such
as a bed or couch, where the air opening may be blocked. Keep the openings
free from lint, hair and the like.
USING
- The use of oxygen therapy requires that special care be taken to reduce the risk of fire. Users MUST NOT SMOKE while using this device. Keep all matches, lighted cigarettes or other sources of ignition out of the room in which this product is located.
- NO SMOKING signs should be prominently displayed. Textiles and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air. Failure to observe this warning can result in severe fire, property damage and cause physical injury or death.
- For optimum performance, do not open or turn off the concentrator frequently, reset after 3~5 minutes. Shorter periods of operation may reduce maximum product life.
- A spontaneous and violent ignition may occur if oil, grease or greasy substances come in contact with oxygen under pressure. These substances MUST be kept away from the oxygen concentrator, tubing and connections, and all other oxygen device.
- DO NOT use any lubricants unless recommended by manufacturer.
MAINTENANCE
- The oxygen concentrator was specifically designed to minimize routine preventive maintenance at intervals of once per year.
- Only professionals of the healthcare field or persons fully conversant with this process such as authorized or factory trained personnel should perform preventive maintenance or performance adjustments on the oxygen concentrator.
RADIO FREQUENCY INTERFERENCE
This equipment generates, uses and can radiate radio frequency energy and, if
not installed according with the instructions, may cause harmful interference
to other devices in the vicinity. However, there is no guarantee that
interference will not occur in a particular installation. If this equipment
does cause harmful interference to other devices, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
- Reorient or relocate the receiving device.
- Increase the separation distance between the equipment.
- Connect the equipment into an outlet on a circuit different from that which the other device(s) are connected.
- Consult the manufacturer or service technician for help.
TO REDUCE THE RISK OF BURNS, ELECTROCUTION, FIRE OR INJURY TO PERSONS
- Avoid using while bathing. If continuous usage is required by the physician’s prescription:
- The concentrator must be located in another room at least 2.5m from the bath.
- DO NOT come in contact with the concentrator while wet.
- DO NOT place or store product where it can drop into water or other liquid.
- DO NOT reach for product that has fallen into water. UNPLUG IMMEDIATELY.
- A product should NEVER be left unattended when plugged in.
- This device is to be used only in accordance with the prescription of a physician and this User’s Manual.
- If at any time the patient or attendant concludes that the patient is receiving an insufficient amount of oxygen, the supplier and/or physician should be contacted immediately.
- No adjustments should be made to the flowrate unless prescribed by a physician. Close supervi sion is necessary when this product is used near children or physically-challenged individuals
- Use this product for only intended use as described in this manual.
- DO NOT use parts, accessories or adapters other than those authorized by manufacturer.
- DO NOT connect the concentrator in parallel or series with other oxygen concentrators or oxygen therapy devices.
- Use of some administration accessories or certain humidifiers not specified for use with oxygen concentrator may impair the performance.
- In certain circumstances oxygen therapy can be hazardous. Manufacturer recommends that you seek medical advice before using this product.
- Avoid creation of any spark near medical oxygen equipment. This includes sparks from static electricity created by any type of friction.
- The shelf life of the device is 5 years after starting operating.
SUMMARY
- The oxygen concentrator is intended for individual use in the home. It is an electronically operated device that separates oxygen from room air.
- It provides high concentration of oxygen directly to you through a nasal cannula.
- Clinical studies have documented that oxygen concentrators are therapeutically equivalent to other types of oxygen delivery systems.
- This user’s manual will tell you about your concentrator and will serve as a reference as you use your concentrator.
CHARACTERS
- Complete plastic out shell, safe and reliable.
- Timing accumulates function, show total operating hours through the display screen.
- Timing turns off function using convenience.
- Compressor pressure relief valve help the device safer.
- Power interruption alarm function.
- Device failure alarm function (including pressure/cycle failure, compressor failure, low oxygen concentration).
- Compressor with over heat protect function to gain in safety of the compressor and the concentrator.
- Remote control function.
SPECIFICATIONS
-
Maximum Recommend Flow Rate: 5L/min
-
Flow Range: 0.5~5L/min
-
Change in maximum recommended flow when back pressure of 7kPa is applied: 0.5L/min
-
Oxygen Concentration: 95.5%~87%
-
Output Pressure: 40~70kPa
-
Oxygen Concentration: %
-
Pressure Relief Mechanism Operational at: 250kPa±25kPa (36.25psi±3.63psi)
-
Sound Level: Average (Front Side): 48dB(A), Average(Whole Unit): 52dB(A).
Power Supply:- AC100V±10% □AC120V±10% □AC220V±10% □AC230V±10%
- 50Hz±2% □60Hz±2%
-
Input Power: 400VA
-
Net Weight: 15.5kg
-
Dimension: W39cm(15.4 〃)×D24.5cm(9.7〃)×H50cm(19.7〃) Oxygen Concentration: %
-
Altitude: Up to 1828 meters (6000 ft) above sea level without degradation of concentration levels. From 1828 meters (6000 ft) to 4000 meters (13129 ft) below 90% efficiency.
-
Safety System:
- Over Current or Connection Loosen: Unit Shut Down
- Compressor Over Hot: Unit Shut Down
- Pressure, Cycle Failure: Alarming and Shut Down
- Compressor Failure: Alarming and Shut Down
- Low Oxygen Concentration
-
Minimum Operating Time: 30minutes
-
Electric classification: class II equipment, type BF applied part
-
Work system: work continuously.
-
Normal Operating Condition:
- Temperature range: 10℃~40℃ (50℉~104℉)
- Relative humidity: 30%~75%
- Atmosphere pressure: 860hPa~1060hPa (12.47psi~15.37psi)
-
Oxygen Output Temperature: ≤ 46℃
-
Cannula length DO NOT more than 15.2m (50ft) and no twist.
-
Storage and Transportation Condition:
- Temperature range: -20℃~+60℃ (-4℉~+140℉)
- Relative humidity: 10%~93% Non-condensing.
- Atmosphere pressure: 700hPa~1060hPa (10.15psi~15.37psi)
-
When storage/transportation condition is lower than 5℃, please put the device in normal operating condition more than 4 hours before operating.
-
The device should be stored with no strong sunlight, no corrosive gas and well ventilated indoor area. The device must be transported and used in the vertical position only.
UNPACKING Ⅰ. FEATURE VIEW
Unless use the oxygen concentrator, retain containers and packing materials
for storage until use of the concentrator is required.
- Check for any obvious damage to the carton or its contents. If damage is evident, please notify the carrier or local dealer.
- Remove all loose packing from the carton.
- Carefully remove all the components from the carton.
STORAGE
- Examine exterior of the oxygen concentrator for nicks, dents, scratches or other damages.
- Inspect all components.
- Store the repackaged oxygen concentrator in a dry area.
- DO NOT place other objects on top of the repackaged concentrator
OPERATING & INSTALLATION
FEATURE VIEW
PREPARE WORK
- Unscrew the bottle from the humidifier clockwise. Fill the bottle with pure water (or distilled water) to the level between MAXIMUM and MINIMUM.
- DO NOT fill humidifier bottle over MAXIMUM level. (Figure 5)
- Screw securely the bottle back counterclockwise. (Recommend to use Yuwell Humidifier p/n: 20006388)
- Connect the power supply.
CAUTION
- If the concentrator has a damaged cord or plug, if it is not working properly, if it has been dropped or damaged, or dropped into liquid, call Qualified Service Personnel for examination and repair.
- Keep the cord away from HEATED or HOT surfaces.
- Do not move or relocate concentrator by pulling on the cord.
- Do not use extension cords with this unit.
CAUTION:
NOTE: Concentrator may be used during the initial start warm-up time (approximately 30 min) while waiting for the O2 purity to reach maximum.
OXYGEN ABSORBING OPERATION
FLOWRATE (Figure 6 )
Turn the flowrate knob to the setting prescribed by your physician or
therapist.
NOTE: To properly read the flowrate, locate the prescribed flowrate line
on the flowmeter. Next, turn the flow knob until the ball rises to the line.
Now, center the ball on the L/min line prescribed
CAUTION
If the flowrate on the flowmeter ever falls below 0.5L/min, check tubing or
accessories for blocked or kinked tubing or a defective humidifier bottle.
Meanwhile, humidifier bottle will have air bubble around the sieve core. Then,
oxygen comes from the oxygen outlet. Connect oxygen cannula to oxygen outlet,
the other end match the patient.
Note: The time of absorbing oxygen and the flowrate range according to
the doctor’s advices.
ALARM SIGNAL
Concentrator has following failure alarm function:
- Pressure and cycle failure
- Compressor failure
- Low oxygen concentration
Note
All the alarms of the device are low priority.
Initial startup of the concentrator
When the unit is turned on, the green light will come on After 5 minutes, the
oxygen sensor will be operating normally and will control the indicator lights
depending on oxygen concentration values. The explanations of the indicator
light functions are as follows.
Explanation of the indicator light
Alarm signal register
-
concentration is greater than 82%. Green light illuminates, and panel 2 shows “----H elapsed time H”. Normal Operation.
-
concentration is greater than 50% and less than 82%. Green and 2 yellow light illuminate, and panel shows “----H elapsed time H”. Call Supplier Immediately. You may continue to use the concentrator unless instructed otherwise by your supplier. Be certain that BACKUP OXYGEN is nearby.
NOTE: Concentrator will reach to most stable state after warm-up (approximately 30 minutes). -
concentration is less than 50%. Red light illuminates, continuous 2 audible alarm sounds, panel shows word “LO” and the unit shutdown. Switch immediately to backup oxygen supply. Call supplier immediately.
-
Low/High pressure, cycle failure alarm Red light illuminates, continuous audible alarm sounds, and panel shows word“ E1” or“ E2” . Total unit shutdown. Switch immediately to backup oxygen supply. Call supplier immediately.
-
Compressor failure alarm Red light illuminates, continuous audible alarm sounds, and panel shows word“ E3” or“ E4” . Total unit shutdown. Switch immediately to backup oxygen supply. Call supplier immediately.
TURNNING OFF
User can press the timing button to setting the operating time in 0~2 Hours.
When it start working, the display shows “----H elapsed time H ”, it means
that timing function is closed. It will work continually until power supply
cuts off.
Press the “+” button once, operation time increases 1 minute, holding the button more than 1.5 seconds it will increase continually. Press the “-” button once, operation time decrease 1 minute, holding the button more than 1.5 seconds it will decrease continually. The device will auto turn off and the display will show “00:00H” when finishing timing. Reset timing function. During using the device, user can press down the standby button, mounted in control panel, to stop/start oxygen supply. Take off oxygen cannula from oxygen outlet first, turn off power switch, and then cut off the power source.
SYMBOLS
Remote Control
Use infrared remote controller to operate the device. (Figure 7)
Note
During remote control operating, the remote controller must be toward the
receiver which is located in the control panel.
MAINTENANCE
NOTE
Power should be disconnected before beginning preventive maintenance on the
concentrator. The concentrator is specifically designed to minimize routine
preventive maintenance at intervals of once per year. In places with high dust
or soot levels, maintenance may need to be performed more often. The following
must be performed at a minimum of one year in service to assure years of
additional reliability.
CLEAN CABINET
- CUT OFF the power supply first to avoid electrical shock.
- DO NOT remove device cabinet.Clean the cabinet with a mild household cleaner and non-abrasive cloth or sponge at least one time a month.
- Do Not drop any liquid into the seam of the device.
CLEAN OR REPLACE FILTER
Please clean or replace the filters in time, it’s very important to protect
compressor and to extend the device life.
Disassembly Filter
Remove the filter cover and take out the filter.
Clean Filter
- Clean the filter with a soft cleaner or wash in warm soapy water and rinse thoroughly.2) DRY the filter thoroughly before reinstallation.
- The filter must clean or replace.DO NOT operate the concentrator without installing filter, while filter is wet. These actions could permanently
- Remove the humidifier tube and diffuser then clean them.
- DRY the filter thoroughly before reinstallation damage the concentrator
TROUBLESHOOTING GUIDE
Symptom | Probable cause | Solution |
---|
Power Interruption Alarm:
Press the power on switch on the control panel, buzzer alarming, display not lit, device not work.
| 1) Power cord and wall outlet have poor contact.| 1) Re-insert the power
cord into wall outlet securely.
2) Wall outlet no power output.| 2) Change to another wall outlet.
- Low power of wall outlet.
| 3) Change to another wall outlet. Do not use any extension cord.
4) Breaker reset button raised.| 4) Press down the breaker reset button.
5) If the device is still not working, please contact the supplier.
Concentrator is working, the operating sound is normal, the flowmeter knob can be adjusted but no oxygen output or weak output.
| 1) Leakage between humidifier bottle and cap.| 1) Re-insert the power cord
into wall outlet securely.
2) The safety valve of humidifier is opened.| 2) Change to another wall
outlet.
The connection between oxygen output and humidifier is leaking.| 3) Re-install
the humidifier tubing.
4) The accessory (nasal cannual, mask, humidifier, tubing…., etc.) is
leaking.| 4) Replace the leaked accessory.
5) If the phenomenon is still appeared, please contact the supplier.
Symptom| Probable cause| Solution
---|---|---
Concentrator is working, but green light, yellow light illuminate at the same time.
| 1) 50% ≤ Oxygen Concentration < 82%.|
-
Clean or replace the filter.
-
Oxygen flow rate over the maximum recommend flow rate: 5L/min| 2) Adjust the flow rate only under the advice of a physician.
-
If the phenomenon is still appeared, the device can be used but please contact the supplier.
Concentrator is not working, the red light illuminate, buzzer alarm, the display show “LO”.
| 1) Oxygen Concentration
< 50 %
| 1) Clean or replace the filter.
- Oxygen flow rate over the maximum recommend flow rate: 5L/min
| 2) Turn off and re-turn on the device. Adjust the flow rate only under the
advice of a physician.
3) If the phenomenon is still appeared, the device can be used but please
contact the supplier.
Concentrator is not working, the red light illuminate, buzzer alarm, the
display show “E1”.| 1) System cycle pressure is too low.|
- Clean or replace the filter.
) If the phenomenon is still appeared, stop using the device, please contact
the supplier immediately.
Concentrator is not working, the red light illuminate, buzzer alarm, the
display show “E2”.|
- System cycle pressure is too high.
|
- Stop using the device, please contact the supplier immediately.
Concentrator is not working, the red light illuminate, buzzer alarm, the display show “E3”.|
- Compressor circuit is opened.
|
- Stop using the device, please contact the supplier immediately.
Concentrator is not working, the red light illuminate, buzzer alarm, the display show “E4”.|
- Compressor circuit is closed.
|
- Stop using the device, please contact the supplier immediately.
Symptom | Probable cause | Solution |
---|
The nasal cannula is fogged with vapor or full of liquid drops. (Remove the vapor or liquid drops method: Turn on the device and connect the tubing to the oxygen outlet. Use your finger to block and release the end of the tubing several times to remove the moisture and liquid.)
| 1) Poor ventilation around the around the device, so operating temperature
is too high.| 1) Make sure the device is at least 10 cm away from the walls,
other jam or heater.
2) The water which added to humidifier is too hot.| 2) Add cold water to
humidifier.
- Add too many water into humidifier.
| 3) Water added should between the maximum and minimum levels in humidifier.
- The device stops suddenly during operating.
| 4) Turn off the device immediately. Restart the device to remove the
moisture and liquid.
5) Accessory is connected to the device. The tubing is kinked and the device
stops suddenly.|
-
Smooth the nasal cannula and tubing.
-
Cooling fan which mounted inside the device stop running or rotate slowly, make the operating temperature is too hot.| 6) Replace the cooling fan. Note: Cut off the power first. And perform by qualified personnel only.
-
If the phenomenon is still appeared, the device can be used but please contact the supplier.
Remote control malfunction.
| 1) Do not use the remote controller in the receiving area.| 1) Operate the remote controller in the front side of the device.
- Low battery.
| 2) Replace the battery in the remote controller.
GAS PASS OPERATION SKETCH MAP
ELECTRICAL RATIONALE
PACKING LIST
- Oxygen Concentrator: 1 unit
- User Manual: 1 piece
- Filter: 1 piece
DEVICE DISPOSAL
Contact the local city or town offices for proper disposal instructions for
oxygen concentrator
Download PDF:Yuwell 8F-5A Oxygen Concentrator User Manual
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>