McKESSON MCK895WRFT Wireless Refrigerator/Freezer Thermometer User Guide

June 6, 2024
McKESSON

McKESSON LOGO NIST TRACEABLE

Wireless
Refrigerator/FreezerThermometer
TRIPLE DISPLAY SINGLE PROBE LARGE DIGIT

For use with MFR # MCK895WRFT.

WIRELESS MIN-MAX ALARM THERMOMETER FEATURES

  • 3 channels temperature measurement
    Internal – Device’s ambient temperature
    External – Probe with bottle
    Wireless – Bluetooth transmitter

  • Triple display for current, minimum, and maximum temperature

  • Programmable Low and High alarm limit

  • Ambient/External
    temperature display

  • Wireless temperature display

  • ºC / ºF exchange

  • Low battery indicator

  • Foldaway stand or mounting keyhole

  • FCC, CE, IC, and RoHS complaint

SPECIFICATIONS

Measuring range| 0-50°C / 32-122°F (thermometer internal sensor) -50°C-70°C / -58-158°F (External probe sensor) -40°C-50°C / -40-122°F (External wireless sensor)
---|---
Display Accuracy| ±0.5°C / ±0.9°F in -20 -50°C / -4-122°F, ±1.0°C / ±1.8°F rest range
Display Resolution| 0.1°
Display Reading Update| 10 seconds thermometer internal sensor; 30 seconds external wireless sensor
Wireless Sensor Distance| 50m (open field), 10m (inside the appliance)
Battery| 1.5 volt, type AAA x 4 pieces
Display Size| (w) 79mm x (H) 66mm
Thermometer Receiver
Battery Life| – 15,000 hours in probe mode – 4,000 hours in wireless mode
Wireless Transmitter
Battery Life| – 6,000 hours
Accessories| 1 each, external probe sensor, 1m cord 4 each, AAA batteries _ 1 each, external wireless sensor

ERROR SYMBOLS

SYMBOL DESCRIPTION ACTION
r Transmitter low battery voltage Replace battery
A  **** Receiver low battery voltage Replace battery
LL.L 1. Sensor open circuit

2. The reading exceeds the lower range
of the thermometer (-50°C/-58°F)| 1. Return the thermometer for evaluation/repair
2. Adjust for reading to be in the range of the lower limit.
HH.H| 1. Sensor open circuit
2. The reading exceeds the upper range of the thermometer (70°C/158°F)| 1. Return the thermometer for evaluation/repair
2. Adjust for reading to be in the range of the upper limit
Syn| Code rematch| No action required
FAI| Connection failure| 1. Review product is the invalid distance range
2. Power device on/off

NOTE

  1. If the Thermometer/Receiver is not in use for a long period of time, remove the batteries from the battery compartment to avoid battery leakage.
  2. Do not operate the Thermometer/Receiver in an environmental temperature lower than 0º C/32 º F or higher than 50º  C/122º  F, otherwise incorrect readings or damage to the Thermometer/Receiver may result. The Thermometer/Receiver should never be placed in an appliance for temperature monitoring, only the Probe or Wireless Transmitter.

INSTALLATION WIRELESS TRANSMITTER

  1. Install battery with correct polarity positioning. Diagram of polarity printed on an end cap.
  2. Press [ ] on the Wireless Transmitter to power on; the Blue LED light above the power button will flash 5 times.
  3. Put the Wireless Transmitter into the area of temperature monitoring. Allow the unit to reach temperature equilibrium of the area to be monitored before setting low and high alarm limits (approximately 1 hour).
  4. Continue to OPERATION for the next step.
    If you require to turn off your Wireless Transmitter press and hold [ ] for 3 seconds; the Blue LED light above the power button will flash 2 times.

INSTALLATION PROBE

  1.  Remove protective plastic end from probe sensor connection end and plug into a port on side of the unit.
  2. Turn the Thermometer/Receiver over and remove the battery cover. Install batteries with correct polarity positioning in Thermometer/Receiver. Diagram of polarity printed inside the battery compartment. Replace the battery cover.
  3. Put the Probe into the area of temperature monitoring. Allow the unit to reach temperature equilibrium of the area to be monitored before setting low and high alarm limits (approximately 1 hour).
  4. Continue to OPERATION for the next step.

OPERATION

  1. Turn on the Thermometer/Receiver using the On/Off switch on the back of the unit.
  2. Select Celsius or Fahrenheit using the ºC / ºF button on the back of the unit.
  3. Ensure that the Probe/Wireless Transmitter is in the area of temperature monitoring and has equilibrated prior to set low and high alarm.

LOW AND HIGH ALARM LIMIT SETTINGS

  1. Press [MODE] to enable alarm limit setting.
    Press [LO ↑] to set a low alarm limit. Holding down the button will increase the value automatically. If you missed the desired temperature, continue holding the button while temperatures cycle again. You can also press the button to increase by a single degree
    slowly.
    Press [HI ↑] to set a high alarm limit.
    Holding down the button will increase the value automatically. If you missed the desired temperature, continue holding the button while temperatures cycle again. You can also press the button to increase by a single degree slowly.
    **If you select LO/HI parameters for a Probe/Wireless Transmitter that exceeds your parameters, the thermometer will alarm immediately.

  2. Press [MODE] to return to NORMAL <Current/Min/Max> display mode.

  3. When the current reading is lower or higher than the alarm limits, the alarm will sound and a red LED light will flash at the bottom right of the unit.

  4. To stop the alarm sound, press any button once. The LO or HI icon will still be flashing until the current reading is within the alarm limit. Press [LO ↑] and [HI ↑] simultaneously to turn off the LED.

  5. Press [ ] to enable alarm function.

RESET MINIMUM (MIN) AND MAXIMUM (MAX) READING MEMORY

  1. In MIN-MAX display mode, press [Min-Reset] to reset minimum reading memory to a current temperature reading.
  2. Press [Max-Reset] to reset maximum reading memory to a current temperature reading.
  3. Reset the memory once after viewing/documenting new Minimum or Maximum readings to begin a new data cycle.

AMBIENT/EXTERNAL WIRELESS SENSOR TEMPERATURE DISPLAY
Press [AMB/EXT/WIRELESS] to select AMB, EXT, or Wireless [ ] temperature display.
*When powering up, displays Ambient.

CODE REMATCH

  1. Wireless Transmitter has been synchronized with the Thermometer/Receiver at ex-factory setting and serialized to match. If the rematch is needed, turn off the  Thermometer/Receiver, press and hold [MODE] and turn on the thermometer after 5 seconds. Rematch mode will start and the “Syn” icon will appear in the display.
  2. Turn off the Wireless Transmitter press and hold [ ] for 3 seconds; the Blue LED light above the power button will flash 2 times. Now turn on the Wireless Transmitter by pressing and holding [ ] for 5 seconds to start the rematch.
  3. The units will restore the new code automatically and return to NORMAL <Current/Min/Max> display mode.

LIMITED WARRANTY

1 year from date of distributor purchase due to manufacturer defect of product functionality.
General Questions? Call 1-800-777-4908
Technical Support? Call 1-973-300-9100
Manufactured by THERMCO Products, Inc.
10 Millpond Drive, Unit 10 Lafayette, NJ 07848
Distributed by McKesson Medical-Surgical Inc
Richmond, VA 23233
PVN B0220 Made in China

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals