Bovie Smoke Shark III Owner’s Manual
- June 6, 2024
- Bovie
Table of Contents
**Bovie Smoke Shark III Owner’s Manual
**
DIAGRAM
-
Figure 1
-
120 Power Cord
-
220 Power Cord
-
Automatic Activation Device
GLOSSARY
Name
|
Description
---|---
AMP
| Ampere, unit of electric current
Automatic Activation Device
| Device used to remotely operate the Bovie ® Smoke Shark Surgical Smoke Evacuator to control the suction on/standby modes in conjunction with ESU activation, saving filter life.
CISPR
| International Special Committee on Radio Interference
EMC
| Electromagnetic Compatibility
ESD
| Electrostatic Discharge
Filter
| Completely enclosed device where surgical smoke is processed through
four (4) stages of filtration.
Filter Life Indicator
| A visual status indication of the life of filter in use.
Grounded Electrical Outlet
| An electrical outlet, which in addition to the current-carrying contacts, has a third contact that serves for connection to a grounding conductor. Devices and equipment that are to benefit from this safety feature must have an appropriate three-prong plug that is inserted into this outlet. There are other possible arrangements for such outlets, including the use of lateral grounding contacts that make contact with metallic strips on the side of the plug. Also called by various other names, including grounding outlet, ground outlet, grounded receptacle, grounding receptacle, grounded socket, safety outlet, or three-prong outlet.
IEC
| International Electrotechnical Commission
LED
| Light-emitting diode (LED) is a two-lead semiconductor light source that emits light when activated.
Power Cord
| A cable used to connect the Surgical Smoke Evacuator to a grounded electrical outlet.
Suction On/Standby Button
| Button to shift between two modes of suction control: on and standby.
Bovie ® Smoke Shark Surgical Smoke Evacuator.
| Device with one or more filters designed to evacuate surgical smoke and aerosol from the operative site, filter out the contaminants, and return filtered air to the operating room.
VAC
| Volts Alternating Current
DESCRIPTION / INTRODUCTION
Introduction
The Bovie® Smoke Shark Surgical Smoke Evacuator is intended to evacuate and
filter surgical smoke and aerosols created by the interface of surgical tools
with tissue (examples: lasers, electrosurgery systems, desiccators and
ultrasonic devices).
The Bovie® Smoke Shark Surgical Smoke Evacuator has been designed to provide
appropriate suctioning, using a motor on /off button, to manage surgical smoke
instruments. The ultra-quiet motor is used to draw the surgical smoke from the
surgical site through the vacuum tubing and into the filter where the surgical
smoke is processed through four (4) stages of filtration. A single, completely
enclosed, disposable filter is used to simplify installation and removal
during filter changes to protect healthcare personnel from potential
contamination during filter changes.
- The first stage filtration utilizes a pre-filter to trap and remove gross particulate and casual fluid.
- The second stage filtration is an ULPA grade (Ultra Low Penetration Air) filter with a high-tech patented design that captures particulates and micro-organisms from 0.1 to 0.2 microns at 99.999% efficiency.
- The third stage filtration uses the highest grade virgin activated carbon. The activated carbon is known to remove toxic organic gases and may provide optimal odor removal.
- The fourth stage filtration is a woven fiberglass filtration media used to reduce the amount of activated carbon fines from migrating out of the filters.
Inspection
The Bovie® Smoke Shark Surgical Smoke Evacuator was thoroughly tested and
inspected prior to shipment. Please inspect the Bovie® Smoke Shark Surgical
Smoke Evacuator before using to ensure all items have been received and that
no damage occurred during transit. If items are missing or damage is evident,
please contact Customer/Technical Services.
Items Included:
- ms Included:
- Operator’s Manual
- Filter
- Power Cord
- Automatic Activation Device
- Wall Mounting Kit
Operational Information
- The operational information contained in this section is intended for customer review of technical specifications. The information pertains to the use of the products both domestically and internationally:
- Both the 100/120 VAC, 50/60 Hz and 220/240 VAC, 50/60 Hz Bovie® Smoke Shark Surgical Smoke Evacuators comply with IEC60601-1 electrical specifications.
- Type of protection against electrical shock (ANSI/AAMI ES 60601-1, Clause 6.1): Class I
- Degree of protection against electric shock (ANSI/AAMI ES 60601-1, Clause 6.2): Type CF Applied Part
- Degree of protection against ingress of water (ANSI/AAMI ES 60601-1, Clause 6.3): IPX0
- Method of sterilization or disinfection recommended (ANSI/AAMI ES 60601-1, Clause 6.4): Unplug Bovie® Smoke Shark Surgical Smoke Evacuator. Wipe Bovie® Smoke Shark Surgical Smoke Evacuator with a damp cloth containing mild disinfectant solution or soapy water. Wipe dry with a clean cloth. Do not steam sterilize.
- Degree of safety of application in the presence of flammable anesthetic mixture with air or with oxygen or nitrous oxide (ANSI/AAMI ES 60601-1, Clause 6.5): Not Suitable
- Mode of operation (ANSI/AAMI ES 60601-1, Clause 6.6): Continuous
- The fuses are to be serviced by an authorized technician. Please call Technical Services. 100/120 VAC, 50/60 Hz use 4 AMP 250 Volt Fuse (Slo-Blo) 220/240 VAC, 50/60 Hz use 2 AMP 250 Volt Fuse (Slo-Blo)
- The Bovie® Smoke Shark Surgical Smoke Evacuator requires special precautions regarding Electromagnetic Compatibility (EMC) and needs to be installed according to EMC information found in this manual.
- To isolate the Bovie® Smoke Shark Surgical Smoke Evacuator from supply mains, unplug the Power Cord from the power receptacle on the Bovie® Smoke Shark Surgical Smoke Evacuator or receptacle in the wall. Position the equipment to allow for ease of unplugging the Power Cord.
- The Bovie® Smoke Shark Surgical Smoke Evacuator and all filters are not intended for contact with patients.
- Type CF applied part – Pen-type (tubing) device for attachment to surgical cutting tool; no electrical connection to unit.
Cautions and Warnings
All Cautions and Warnings should be read and understood before using the
Bovie® Smoke Shark Surgical Smoke Evacuator.
Attention: Consult Instructions Before Use.
WARNINGS
The warranty on the Bovie® Smoke Shark Surgical Smoke Evacuator is void if any of the following warnings are disregarded.
- Surgical Smoke Evacuator.
- Confirm operational set-up of the Bovie® Smoke Shark Surgical Smoke Evacuator prior to a surgical procedure.
- Disconnect the Bovie® Smoke Shark Surgical Smoke Evacuator from the grounded electrical outlet prior to inspecting Bovie® Smoke Shark Surgical Smoke Evacuator components.
- The Bovie® Smoke Shark Surgical Smoke Evacuator is only intended and suitable for the applications mentioned in the operating instructions.
- The Bovie® Smoke Shark Surgical Smoke Evacuator produces a strong vacuum. Position the inlet end of the wand or tubing to prevent patient injury and to prevent suction of surgical materials and specimens.
- To prevent patient injury, the tubing or wand should not come into direct contact with tissue.
- The filter and single-use accessories are completely disposable. Please dispose according to your local codes or regulations and facility policy.
- Route Power Cord to prevent a tripping hazard or crimping of cords, which could cause unreliable operation or electric shock.
- Position Automatic Activation Device and any other attached accessories to prevent a tripping hazardv or crimping of cords, which could cause unreliable operation.
- Do not operate the Bovie® Smoke Shark Surgical Smoke Evacuator in the presence of flammable or explosive gases.
- The Bovie® Smoke Shark Surgical Smoke Evacuator is intended for use by healthcare professionals only.
- The Bovie® Smoke Shark Surgical Smoke Evacuator may cause radio interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures such as re-orienting or relocating the Bovie® Smoke Shark Surgical Smoke Evacuator or shielding the location.
- The use of accessories other than those specified by the manufacturer as replacement parts for internal components may result in increased emissions or decreased immunity of the Bovie® Smoke Shark Surgical Smoke Evacuator
- Refer routine servicing to Symmetry Surgical. See Customer/Technical Services contact information.
- Changes or modifications not expressly approved by the manufacturer could void the warranty
- To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
CAUTIONS
-
United States Federal Law restricts the Bovie® Smoke Shark Surgical Smoke Evacuator to sale by or on the order of a physician.
-
Using any other filter or accessory not specified by the manufacturer may cause damage and/or cause the Bovie® Smoke Shark Surgical Smoke Evacuator to be inoperable, voiding the warranty.
-
Care must be exercised in the installation of tubing and adapters. Failure to follow the procedures outlined in this manual may result in overheating of the motor and may void the warranty.
-
The Bovie® Smoke Shark Surgical Smoke Evacuator is not intended for evacuation of fluid. Failure to avoid fluid aspiration could cause filter blockage, electrical damage, decreased efficiency, and possible internal contamination.
-
The filter should be changed according to the Filter Life Indicator. The filter should not be used for more than the filter life specified. Failure to change the filter may result in decreased efficiency and possible internal contamination.
-
Care must be exercised in the installation and removal of filter and tubing. Filter and tubing are used to capture potentially hazardous materials. Failure to follow the procedures outlined in this manual may result in possible internal contamination.
-
Do not block the tubing during operation. An occlusion or significant restriction may cause the motor to overheat and the Bovie® Smoke Shark Surgical Smoke Evacuator to stop working.
-
Installation of the Bovie® Smoke Shark Surgical Smoke Evacuator must be performed such that the intake and exhaust vents located on the rear of the system are not obstructed. Failure to properly install the Bovie® Smoke Shark Surgical Smoke Evacuator may cause reduced performance, damage, and/or cause the Bovie® Smoke Shark Surgical Smoke Evacuator to be inoperable, voiding the warranty.
-
The ambient temperature during operation must be kept between 50°F to 104°F (10°C to 40°C).
-
The relative humidity during operation must be kept between 10% to 75%.
-
An atmospheric pressure range during operation must be kept between 700 hPa to 1,060 hPa.
-
Storage environmental ambient temperature should be kept between 14°F to 140°F (-10°C to 60°C).
-
Storage environmental relative humidity should be kept between 10% to 75%.
-
There are no user serviceable components in the Bovie® Smoke Shark Surgical Smoke Evacuator.
Refer to Symmetry Surgical Customer/Technical Services for routine servicing. -
Use only with the Power Cord provided and always plug into a grounded electrical outlet.
Symbol| Description / Meaning| Symbol| Description /
Meaning
---|---|---|---
| DANGER HIGH VOLTAGE CAUTION – ELECTRICAL SHOCK HAZARD. DO NOT REMOVE COVER.
REFER SERVICING TO QUALIFIED
SERVICE PERSONNEL.
|
|
MARKING OF EEE (ELECTRICAL AND ELECTRONIC EQUIPMENT). FOLLOW LOCAL REQUIREMENTS FOR PROPER DISPOSAL.
| DANGER
CAUTION – RISK OF EXPLOSION IF USED IN THE PRESENCE OF FLAMMABLE ANESTHETICS.
|
| CAUTION: FEDERAL LAW (USA) RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER
OF A PHYSICIAN
|
WARNING
| | CONSULT INSTRUCTIONS/ DIRECTIONS FOR
USE
|
CAUTION
| |
MEDICAL DEVICE
|
TYPE CF APPLIED PART
|
| NOT MADE WITH NATURAL RUBBER LATEX
|
ALTERNATING CURRENT
|
|
CE MARK
|
PROTECTIVE EARTH, (GROUND)
|
| AUTHORIZED REPRESENTATIVE IN THE EUROPEAN COMMUNITY FOR CE MARKED DEVICES.
| DENOTES THE DATE THE EQUIPMENT WAS MANUFACTURED| |
SERIAL NUMBER
| DENOTES THE MANUFACTURER OF THE DEVICE| | DO NOT USE IF PACKAGE IS DAMAGED
|
NON-IONIZING RADIATION
| | SERVICE INDICATOR
CONTACT CUSTOMER/TECHNICAL SERVICES
Control Panel (see Figure 2)
The Control Panel contains the following LED indicators: Motor On, Standby,
Filter Life, and Service.
Read all instructions before operating the Bovie® Smoke Shark Surgical Smoke
Evacuator or installing accessories. Failure to do so may result in damage to
the Bovie® Smoke Shark Surgical Smoke Evacuator and/or personal injury.
POWER OFF / ON
To power up the Bovie® Smoke Shark Surgical Smoke Evacuator, connect the
supplied Power Cord to a grounded electrical outlet and the power receptacle
at the bottom of the Bovie® Smoke Shark Surgical Smoke Evacuator. Flip the
Power Off/On Button to “ON” to apply power to the unit. Once power has been
applied, the Standby Indicator LED illuminates the color amber. Turn the
Bovie® Smoke Shark Surgical Smoke
Evacuator main power off by either flipping the Power Off/On Button to “OFF”
or by unplugging the Power Cord from either the power receptacle on the Bovie®
Smoke Shark Surgical Smoke Evacuator or the grounded electrical outlet.
SUCTION ON / STANDBY BUTTON
Press the Suction On/Standby Button to shift between two (2) modes: On or
Standby. The Suction On
LED illuminates the color green indicating active suction and the Standby LED
illuminates the color amber when in standby.
Filter LIFE INDICATOR
The Filter Life Indicator on the Control Panel provides a visual indication of
the status of the filter life of the filter in use.
Install an unused filter into the Bovie® Smoke Shark Surgical Smoke Evacuator
as per the installation instructions (section 2.3). When the Motor On LED is
illuminated, the Filter Life Indicator will light up the uppermost green LED
indicating 100% Filter Life. The indicator will progress through subsequent
colors, from green LEDs to amber LED as time elapses, and finally to red to
indicate the filter has expired and requires replacement. When the maximum
filter life is expended: The unit will cease operation in one (1) hour or when
the unit is manually powered off, whichever occurs first. The smoke evacuator
will not resume functionality until a new filter is installed. The filter has
a 6 hour lifetime which is automatically tracked by the Bovie® Smoke Shark
Surgical Smoke Evacuator.
Set-Up and Operation
-
Mount the Bovie® Smoke Shark Surgical Smoke Evacuator
using the wall mounting kit (SE03-M). Secure the wall
mounting kit to the wall by drilling two 3/8” diameter holes in wall, installing the provided hardware into the holes, and secure mounting plate to the provided hardware. The Bovie® Smoke Shark Surgical Smoke Evacuator can be secured to the wall mounting kit (SE03-M) by sliding the two screws protruding from the rear of the Bovie® Smoke Shark Surgical Smoke Evacuator (Figure 2.) into the grooves found on the mounting kit (SE03-M). -
Attach the Power Cord to the power receptacle on the bottom of the Bovie® Smoke Shark Surgical Smoke Evacuator and into a grounded electrical outlet. Route Power Cord to prevent a tripping hazard or crimping of cords, which could cause unreliable operation or electric shock.
-
Install the filter (see filter installation instructions, section 2.3).
-
The Automatic Activation Device may also be installed by plugging it into the Automatic Activation Device port located on the front of the Bovie® Smoke Shark Surgical Smoke Evacuator. See section 2.4 for instructions.
-
Ensure that any smoke evacuation capture accessory is fully installed in the filter port.
-
Route any other attached accessories to prevent a tripping hazard or crimping of cords, which could cause unreliable operation.
-
Activate the Bovie® Smoke Shark Surgical Smoke Evacuator by one of the following:
- Pressing the Suction On/Standby Button on the Control Panel of the smoke evacuator.
- Utilizing the automatic activation accessory provided with the unit. This device will synchronize the activation of the smoke evacuation unit with the activation of the desiccator device.
-
Deactivate the Bovie® Smoke Shark Surgical Smoke Evacuator by one of the following:
- Pressing the Suction On/Standby Button on the Control Panel of the smoke evacuator.
- Releasing the pencil button on the desiccator device if an automatic activation device is being used.
-
Replace the filter when the Filter Life Indicator flashes the color red (0% life remaining). Failure to change the filter will affect the performance of the Bovie® Smoke Shark Surgical Smoke Evacuator.
Filter Instructions
Bovie® P/N:
| SF06
---|---
Configuration:|
Wall Mounted
Bovie® SF06 Filter:
| 4-Stage Filtration In One Casing, (Pre-Filter, ULPA, Carbon, Post-
Filter)
Filter(s):|
ULPA
Particle Size, µm:
| 0.1 to 0.2 Microns At 99.999% Efficiency
Filter Life:|
6 hours, Automatic Factory Set Filter Sensor
Filter Life Indicator:
|
Time Replacement
Filter Installation Instructions:
Note: Before installing or removing any filter, be sure that the Bovie®
Smoke Shark Surgical Smoke Evacuator is placed in the Standby mode by pressing
the Suction On/Standby Button.
- Remove the filter from the shipping box and discard any protective wrapping.
- Inspect the filter for damage, which may have occurred during shipping and storage. Do not install any filter with visible signs of structural damage.
- Insert the filter into the filter chamber and ensure the filter is installed completely against the bottom of the filter chamber.
Filter Removal Instructions:
- After the filter life has expired, turn the Bovie® Smoke Shark Surgical Smoke Evacuator to the Standby mode by pressing the Suction On/Standby Button.
- Remove all accessories attached to the filter.
- Remove the Filter from the Bovie® Smoke Shark Surgical Smoke Evacuator. Dispose according to your local codes or regulations and facility policy (Section 3.2).
- Clean the Bovie® Smoke Shark Surgical Smoke Evacuator with a mild disinfectant prior to re-use and follow the indicated instructions for maintenance and installation of a new filter (Section 3.2).
Automatic Activation Device Instructions
**Performance References***
Performance
Model Name/ Number
| Bovie® Smoke Shark III / SE03-120 & SE03-220
Maximum Flow Setting (LPM-U.S.)
Standard Hose I.D.
| 9.5 mm, 3/8”
SharkSkin Adapter|
45 LPM
Dimensions (H x W x D )
| inches| 10 x 7 x 7.5
Dimensions (H x W x D )| centimeters|
25.4 x 17.8 x 19.1
Weight
| lbs| 7.0
Weight
| kg|
3.2
Noise Level, dBA| maximum|
55 dBA (120 V), 62 dBA (220 V)
Voltage Available
| 100/120 VAC, 220/240 VAC
Frequency, auto sensed|
50/60 Hz
Remote Control Activation:
| Yes (optional)
Safety Features:|
UL Classified, CE Marked, Fuse Protection
Display:
| LED, Filter Status, Suction On/Off, Service Required
*For reference purposes only
Electromagnetic Compatibility Information per IEC 60601-1-2
Table 1
Table 1
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The Bovie ® Smoke Shark Surgical Smoke Evacuator is intended for use in the electromagnetic environment specified below. The customer or user of the Bovie ® Smoke Shark Surgical Smoke Evacuator should ensure that it is used in such an environment.
Emissions Test
|
Compliance
|
Electromagnetic Environment – Guidance
RF Emissions
CISPR 11
|
Group 1
| The Bovie ® Smoke Shark Surgical Smoke Evacuator uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF Emissions
CISPR 11
| Class A| The Bovie ® Smoke Shark Surgical Smoke Evacuator is suitable for use in all establishments, other than domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
|
Class A
| Not applicable.
Voltage Fluctuations/ Flicker Emissions
IEC 61000-3-3
|
Class A
| Not applicable.
Table 2
Table 2
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Bovie ® Smoke Shark Surgical Smoke Evacuator is intended for use in
the electromagnetic environment specified below. The customer or user of the
Bovie ® Smoke Shark Surgical Smoke Evacuator should ensure that it is
used in such an environment.
Immunity Test
| IEC 60601
Test Level
| Compliance Level| ****
Electromagnetic Environment – Guidance
Electromagnetic discharge (ESD)
IEC 61000-4-2
| + 6 kV contact
+ 8 kV air
| + 6 kV contact
+ 8 kV air
| Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
| + 2 kV for power supply lines
+ 1 kV for input/ output lines
| + 2 kV for power supply lines
+ 1 kV for input/ output lines
| Mains power quality should be that of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
| + 1 kV differential mode
+ 2 kV common mode
| + 1 kV differential mode
+ 2 kV common mode
| Mains power quality should be that of a typical commercial or hospital
environment.
Voltage dips, short interruptions, and voltage variations on power supply
input lines.
IEC 61000-4-11
| <5 % UT
(>95 % dip in UT) for 0.5 cycle
40 % UT
(60 % dip in UT) for 5 cycles
70 % UT
(30 % dip in UT) for 25 cycles
| <5 % UT
(>95 % dip in UT) for 0.5 cycle
40 % UT
(60 % dip in UT) for 5 cycles
70 % UT
(30 % dip in UT) for 25 cycles
| Mains power quality should be that of a typical commercial or hospital environment.
If the user of the Bovie ® Smoke Shark Surgical Smoke Evacuator requires continued operation during power main interruptions,
it is recommended the Bovie ® Smoke Shark Surgical Smoke Evacuator be powered from an uniterruptible power supply or a battery.
<5 % UT
(>95 % dip in UT) for 5 sec
| <5 % UT
(>95 % dip in UT) for 5 sec
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
| 3 A/m| 3 A/m| Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Table 3
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The Bovie ® Smoke Shark Surgical Smoke Evacuator is intended for use in the electromagnetic environment specified below. The customer or user of the Bovie ® Smoke Shark Surgical Smoke Evacuator should ensure that it is used in such an environment.
Immunity Test
| IEC 60601
Test Level
| Compliance Level| ****
Electromagnetic Environment – Guidance
Portable and mobile RF communications equipment should be used no closer to any part of the Bovie ® Smoke Shark Surgical Smoke Evacuator including cables, than the recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Radiated RF| 3 V/m| | d = 1.7 √P 80 MHz to 800 MHz
IEC 61000-4-3| 80 MHz to 2.5 GHz| 3 V/m| d = 2.3 √P 800 MHz to 2.5 GHz
| | 3 Vrms| d = [3.5/V1} √P
Conducted RF| ****
150 kHz to 80 MHz
| | Where P is the maximum output power rating of the transmitter in Watts (W) according to the transmitter manufacturer and D is the
recommended separation distance in meters (M).
IEC 61000-4-6
| Field strength from fixed RF transmitters, as determined by an
electromagnetic site survey, should be less than the compliance level in each
frequency range.
| Interference may occur in the vicinity of equipment marked with the
following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by
absorption and reflection from structure, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Bovie ® Smoke Shark Surgical Smoke Evacuator is used exceeds the applicable RF compliance level above, the Bovie ® Smoke Shark Surgical Smoke Evacuator should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Bovie ® Smoke Shark Surgical Smoke Evacuator.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Table 4
Recommended Separation Distance Between Portable and Mobile RF Communications Equipment and
the Bovie ® Smoke Shark Surgical Smoke Evacuator @ 3 Vrms
The Bovie ® Smoke Shark Surgical Smoke Evacuator is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the Bovie ® Smoke Shark Surgical Smoke Evacuator can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Bovie ® Smoke Shark Surgical Smoke Evacuator as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter (W)
| Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = [ 3.5 ] √P
v 1
| 80 kHz to 800 MHz
d = [ 3.5] √P
E 1
| 800 kHz to 2.5 GHz
d = [ 3.5 ] √P
E 1
0.01| 0.12| 0.12| 0.23
0.1| 0.34| 0.34| 0.74
1| 1.7| 1.7| 2.3
10| 3.7| 3.7| 7.4
100| 11.7| 11.7| 23.3
For transmitters rated at a maximum output power not listed above, the
recommended separation Distance
(D) in Meters (M) can be estimated using the equation applicable to the frequency of the transmitter, where
(P) is the maximum output rating of the transmitter in Watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by
absorption and reflection from structure, objects, and people.
MAINTENANC
General Maintenance Information
It is recommended that periodic inspection and performance testing be completed by a qualified facility biomedical technician to ensure continued safe and effective operation.
Cleaning
- Unplug the Bovie® Smoke Shark Surgical Smoke Evacuator.
- Wipe Bovie® Smoke Shark Surgical Smoke Evacuator with a damp cloth containing mild disinfectant solution or soapy water.
- Wipe Bovie® Smoke Shark Surgical Smoke Evacuator with a dry clean cloth.
- Do not steam sterilize.
Periodic Inspection
The Bovie® Smoke Shark Surgical Smoke Evacuator should be visually inspected
at least once per year.
The inspection should include checks for:
- Damage to the Power Cord or Power Inlet Module.
- Obvious external or internal damage to the Bovie® Smoke Shark Surgical Smoke Evacuator.
FUSES
Two 4 AMP fuses for 100/120 Bovie® Smoke Shark Surgical Smoke Evacuators or
two 2 AMP fuses for 220/240 Bovie® Smoke Shark Surgical Smoke Evacuators are
located within the housing of the system.
The fuses electrically protect both the Bovie® Smoke Shark Surgical Smoke
Evacuator and the operator from damage or injury. If the Bovie® Smoke Shark
Surgical Smoke Evacuator is overheated or if there is an electrical surge,
fuses will break and the Bovie® Smoke Shark Surgical Smoke Evacuator will not
operate. No parts are user serviceable. When the Service LED Light Indicator
illuminates, please contact Customer/Technical Services.
Troubleshooting
PROBLEM
|
POTENTIAL CAUSE
|
CORRECTIVE ACTION
---|---|---
1. Bovie ® Smoke Shark Surgical Smoke Evacuator is on but suction is
minimal or none.| 1. Filter is not seated completely.| 1. Re-install the
filter, press firmly into place.
2. Filter is clogged.| 2. Replace the filter with the manufacturer’s filter.
(SF06)
3. Vacuum tubing is clogged.| 3. Replace vacuum tubing with manufacturer’s
products.
4. Motor is obstructed.| 4. Call Symmetry Surgical Customer Service.
2. Bovie ® Smoke Shark Surgical Smoke Evacuator does not function
even though Suction On/Standby Button is depressed.| 1. Not plugged into an
electrical outlet.| 1. Check the grounded electrical outlet connection and
the power receptacle located at the bottom of the Bovie ® Smoke Shark
Surgical Smoke Evacuator.
2. Fuses are blown.| 2. Call Symmetry Surgical Customer Service.
3. Electronic Bovie ® Smoke Shark Surgical
Smoke Evacuator failure.
| 3. Call Symmetry Surgical Customer Service.
4. Filter Life has expired or invalid filter installed.| 4. Replace the
filter.
Report complaints involving this device to bovie.complaint@symmetrysurgical.com or to your local distributor.
Product Return
For the quickest response to your service needs, please follow these
procedures:
Step 1: Write down Bovie® Smoke Shark Surgical Smoke Evacuator model name
and serial number.
Step 2: Call Customer/Technical Services: 800.251.3000 (United States) or
+1 615.964.5532 (International) and describe the problem.
Step 3: If the problem cannot be resolved over the phone and the Bovie®
Smoke Shark Surgical Smoke Evacuator must be returned for repair, a Return
Material Authorization (RMA) number must be obtained from Customer Service
before returning the Bovie® Smoke Shark Surgical Smoke Evacuator.
Step 4: Use the original packaging material to return your Bovie® Smoke
Shark Surgical Smoke Evacuator whenever possible. If you do not have the
original packaging material, ask Customer Service for advice on how to pack
for return shipment.
Step 5: Freight for all returned products should be prepaid by the
facility returning the product.
A Ship To address will be supplied by Customer Service.
Ordering Information
To reorder, obtain replacement parts, or to return Bovie® Smoke Shark Surgical
Smoke Evacuator, call Customer Service or contact your authorized
Distributor/Sales Representative.
Available accessories:
- Replacement filters (SF06)
- Replacement Automatic Activation Device (SERS2)
- Tubing (Various)
- Reducer Fittings (Various)
- Electrosurgical Pencil Adapter (SS95)
TERMS & WARRANTY
Terms & Warranty
SPECIFICATIONS
Specifications are subject to change without notice.
SHIPMENT OF ORDER
Symmetry Surgical, Inc. will try to accommodate individual customer requests
for shipping method. Symmetry Surgical, Inc. reserves the right to decide
shipping method on prepaid orders. Care is exercised in the checking and
packaging of all merchandise to avoid error, but should discrepancies arise,
claims should be made within 24 hours after delivery. Symmetry Surgical,
Inc.’s responsibility ceases with the safe delivery to the carrier at our
dock. If the merchandise is damaged in transit, a claim must be made to the
carrier involved. Symmetry Surgical, Inc. will assist customers in pursuing
these claims.
RETURN OF MATERIAL
All returned merchandise must have a pre-authorized Return Material
Authorization (RMA) from Symmetry Surgical, Inc. and be marked with that RMA
number prior to returning. Transportation costs must be prepaid by the shipper
and all risk of loss and damage of goods are the responsibility of the
shipper. Unauthorized returns will be refused. Include a copy of the packing
papers and/or invoice with the return. Exchange will be of an equivalent
dollar value of returned merchandise less a restocking and handling fee of
new, unused, unopened equipment or disposables.
EXCEPTIONS
- Defective merchandise may be returned for replacement only. Please contact Symmetry Surgical Customer Service before shipping back merchandise.
- Incorrectly shipped merchandise is exempt from restocking fees. Please contact Symmetry Surgical Customer Service before shipping back merchandise.
WARRANTY
Symmetry Surgical, Inc. warrants that the Bovie® Smoke Shark Surgical Smoke Evacuator manufactured by Symmetry Surgical, Inc. shall be free from defects in material and workmanship. Products are warranted only to the extent that Symmetry Surgical, Inc. will replace without charge any Bovie® Smoke Shark Surgical Smoke Evacuator proven to have defects within two (2) years of the date of delivery of Surgical Smoke Evacuator and provided Symmetry Surgical, Inc. has been given the opportunity to inspect the Surgical Smoke Evacuator alleged to be defective and the installation or use thereof. No warranty is included for incidental or consequential damages of any nature arising from any defect. The warranty above is the only warranty made by Symmetry Surgical, Inc. and is expressly in lieu of all other warranties, expressed or implied, including, without limitation, the warranties of merchantability and fitness for a particular purpose. All warranties implied by any course of dealing or usage between parties are expressly excluded.
Limitation of Liability for Damages. Symmetry Surgical, Inc. is not liable for
indirect, special, incidental or consequential damages of any kind, arising
from any theory, including breach of warranty, breach of contract, negligence,
tort or any other legal theory. Such damages include but are not limited to,
loss of profits or revenues, loss of use of the Product or any associated
equipment, cost of capital, cost of any substitute Product, facilities or
services, down-time costs, lost reputation, or claims of customers for such
damages. Notwithstanding anything to the contrary herein, the total limit of
Symmetry Surgical, Inc.’s liability under any theory for one or more claims
shall not exceed, in the aggregate, the amount paid to Symmetry Surgical, Inc.
for the Product that is the subject of such claim. Purchase and use of the
Product associated with this IFU constitutes acceptance of this provision and
all others in the IFU.
Entire Agreement and Amendment. All exhibits and schedules attached hereto and
referenced herein are made a part of this Agreement. This Agreement
constitutes the entire agreement between the parties concerning the subject
matter of this Agreement and supersedes all prior negotiations and agreements
between the parties concerning the subject matter of this Agreement. The terms
of any purchase order, invoice, or similar document used to implement this
Agreement shall be subject to and shall not modify this Agreement. This
Agreement may only be amended by a written agreement signed by both parties.
Warranties. All warranties provided by Symmetry Surgical, Inc. with respect to
any Product sold are as described in the package inserts or instructions for
use available with the Product or on the
www.symmetrysurgical.com or
www.boviemedical.com websites. OTHER THAN AS
SET FORTH IN THE PREVIOUS SENTENCE, THERE IS NO IMPLIED OR EXPRESS WARRANTY ON
THE PRODUCTS, INCLUDING THOSE OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE. Using any filter not distributed by Symmetry Surgical, Inc. may cause
damage to the Surgical Smoke Evacuator and will void all warranties.
Governing Law, Venue and Jurisdiction. Any dispute arising from or related in
any way to this IFU, the product(s) associated with it or described herein
shall be governed by and construed in all respects in accordance with the laws
of the state of Tennessee in the United States of America. All claims or
differences or other disputes, which arise out of or in connection with this
IFU, the product(s) related to it or otherwise shall be settled, litigated, or
resolved by the state or federal courts in Antioch, Tennessee (or those
serving Antioch, Tennessee), and the parties hereby submit to the exclusive
jurisdiction of such courts over any dispute. All proceedings shall be held in
English. Each party will bear its own costs in such proceedings.
For a period of two (2) years following the date of delivery, Symmetry
Surgical, Inc. warrants the SmokeShark Surgical Smoke Evacuator against any
defects in material or workmanship. Symmetry Surgical, Inc. will repair or
replace (at Symmetry Surgical’s option) the same without charge, provided that
routine maintenance as specified in this manual has been performed using
replacement parts approved by Symmetry Surgical, Inc.
This warranty is void if the product is used in a manner or for purposes other
than intended.
Bovie® products are exclusively distributed by:
Symmetry Surgical, Inc.
3034 Owens Drive
Antioch, TN 37013, USA
USA & Canada 800.251.3000
Fax 800.342.3272
The revision level of this manual is specified by the highest revision letter
found on either the inside front cover or enclosed errata pages (if any).
Manual Number 905214 Rev A 07/2020
MDSS GmbH Schiffgraben 41 D-30175 Hannover Germany
CONMED Corporation
525 French Road
Utica, NY 13502-5994 USA
E351320
Medical – General
Medical Equipment as to Electrical Shock, Fire and Mechanical Hazards only in
Accordance with
ANSI/AAMI ES60601-1 (2005) and AMD 1 (2012), CAN/CSA-C22.2 NO. 60601-1
(2014)
References