ATMOS 41 QUICK START
Preparation
Verify that all ATMOS 41 components arrived intact. Installation will require a 13-mm (1/2-in) wrench and a secure mounting location. METER recommends a meteorological stand, pole in cement, or tripod with a 31.8- to 50.8-mm (1.25- to 2-in) diameter.
Set up and test the system (sensors and data loggers) in a lab or office. Ensure the data loggers are using up-to-date firmware and software. Verify all sensors read within expected ranges.
Before beginning installation, consider the surroundings and avoid obstructions. Many installations require the ATMOS 41 to be mounted 2 m above ground, but this can be adjusted as needed.
Please read the complete ATMOS 41 User Manual at metergroup.com/atmos41-support. All products have a 30-day satisfaction guarantee.
Connecting
P lug into Data Acquisition System Connect the stereo plug connector into any METER data logger and configure it to read the ATMOS 41 (refer to ATMOS 41 User Manual). Select a nonzero measurement interval to ensure data are being logged. To connect to a non-METER data logger, see the ATMOS 41 Integrator Guide at metergroup.com/atmos41-support.
Verify Readings
Use the scan function in the software to show a list of readings. Verify these
readings are within expected ranges.
ATTENTION
For best results, use the latest versions of METER software and firmware for
the computer or mobile device, ZENTRA products, and sensors. Please use the
software Help menu to find updates. Consult the sensor user manual for more
troubleshooting tips.
Installation
1. Mount Toward True North
The ATMOS 41 must be oriented with the engraved N on the instrument oriented
to true north (not magnetic north).
2. Level the System
Use the bubble level underneath the ATMOS 41 or a PROCHECK display to level
the weather station. The mounting pole may need to be levelled or shims may
need to be applied.
3. Secure the System
Tighten the V-bolt nuts by hand until hand-tight, and then tighten with a
wrench.
CAUTION: Do not overtighten bolt.
4. Plug Sensor In and Configure Logger
Plug stereo plug connector into the data logger. Use data logger software to
apply appropriate settings to the sensors plugged into each data logger port.
SUPPORT
Have a question or problem? Our support team can help.
We manufacture, test, calibrate, and repair every instrument in house. Our scientists and technicians use the instruments every day in our product testing lab. No matter what your question is, we have someone who can help you answer it.
NORTH AMERICA
Email:
support.environment@metergroup.com
Phone: +1.509.332.5600
EUROPE
Email: support.europe@metergroup.com
Phone: +49 89 12 66 52 0
References
- Clear Sky Calculator | Apogee Instruments Inc.
- METER Group
- ATMOS 41—How to check the precipitation reading | METER chat
- Weather station for research | ATMOS 41 All-in-one Weather Station chat
- Downloads | METER chat
- METER Environment | METER chat
- ATMOS 41 cleaning instructions | METER chat
- Terms and Conditions - METER Group
- METER Group
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>