GIMA Colour Screen Pocket Doppler User Manual
- June 6, 2024
- GIMA
Table of Contents
Colour Screen Pocket Doppler
PROFESSIONAL MEDICAL PRODUCTS
Gima S.p.A. Via Marconi, 1 – 20060 Gessate (MI) Italy gima@gimaitaly.com – export@gimaitaly.com www.gimaitaly.com
DOPPLER PORTATILE CON SCHERMO A COLORI COLOUR SCREEN POCKET DOPPLER ÉCRAN PORTATIF COULEUR DOPPLER PANTALLA DE COLOR PORTÁTIL DOPPLER MONITOR A CORES DETECTOR DOPPLER DE BOLSO FARBBILDSCHIRM-TASCHENDOPPLER
Manuale d’uso – User manual – Manuel de l’utilisateur Guía de uso – Guia para utilização Gebrauchs- und instandhaltungsanleitung
ATTENZIONE: Gli operatori devono leggere e capire completamente questo manuale
prima di utilizzare il prodotto. ATTENTION: The operators must carefully read
and completely understand the present manual before using the product. AVIS:
Les opérateurs doivent lire et bien comprendre ce manuel avant d’utiliser le
produit. ATENCIÓN: Los operadores tienen que leer y entender completamente
este manual antes de utilizar el producto. ATENÇÃO: Os operadores devem ler e
entender completamente este manual antes de usar o produto. ACHTUNG: Diese
Anleitung muss vor dem Einsatz des Produkts aufmerksam gelesen und vollständig
verstanden werden.
29480 / Sonoline C
CONTEC MEDICAL SYSTEMS CO., LTD No.112 Qinhuang West Street, Economic &
Technical Development Zone, Qinhuangdao, Hebei Province, PEOPLE’S REPUBLIC OF
CHINA Made in China
Shanghai International Holding Corp. GmbH (Europe) Eiffestrasse 80, 20537
Hamburg, Germany
0123
M29480-M-Rev.1-10.19
ENGLISH
22
Attention
This user manual is written and compiled in accordance with the council
directive MDD93/42/EEC for medical devices and harmonized standards. In case
of modifications and software upgrades, the information contained in this
document is subject to change without notice. The manufacturer makes no
warranty of any kind with regard to this material, including, but not limited
to the implied warranties of merchantability and fitness for a particular
purpose. The manufacturer assumes no responsibility for any errors that may
appear in this document, or for incidental or consequential damage in
connection with the furnishing, performance or use of this material. No part
of this document may be photocopied, reproduced or translated to another
language without prior written consent of the manufacturer. The information
contained in this document is subject to change without notice. Responsibility
of the Manufacturer The manufacturer only considers itself responsible for any
effects on safety, reliability and performance of the equipment if: Assembly
operations, repairs are carried out by persons authorized by the manufacturer,
and the device is used in accordance with the instructions for use.
WARNING: This device is not intended for treatment. The intended use is for
detecting Fetal Heart Rate. If the fetal heart rate (FHR) result is
distrustful, please use other methods such as stethoscope to verify
immediately. Warranty The unit can not be repaired by users themselves. All
services must be done by the engineers approved by manufacturer. We warrant
that each product we sell you is free from defects in labor and materials and
shall conform to its product specifications as defined in the user
documentation. If the product doesn’t function as warranted during the
warranty period, we will repair or replace it without charge. Misuse, improper
maintenance may void the warranty. Using This Label Guide This guide is
designed to give key concepts on safety precautions.
WARNING: A WARNING label advises against certain actions or situations that
could result in personal injury or death.
CAUTION: A CAUTION label advises against actions or situations that could
damage equipment, produce inaccurate data, or invalidate a procedure. Note: A
NOTE provides useful information regarding a function or procedure.
23
ENGLISH
Chapter 1 SAFETY GUIDANCE
This unit is internally powered equipment and the degree of shock protection
is type CF applied part . Type CF protection means that these patient
connections will comply with permitted leakage currents, dielectric strengths
of IEC 60601-1. WARNING and CAUTION messages must be observed. To avoid the
possibility of injury, observe the following precautions during the operation
of the device.
WARNING: This device is not explosion-proof and can not be used in the
presence of flammable anaesthetics. WARNING: Do not throw batteries in fire as
this may cause them to explode. WARNING: Do not attempt to recharge normal
dry-cell batteries, they may leak, and may cause a fire or even explode.
WARNING: Don’t touch signal input or output connector and the patient
simultaneously. WARNING: Accessory equipment connected to the analog and
digital interfaces must be certified according to the respective IEC standards
(e.g. IEC 950 for data processing equipment and IEC60601-1 for medical
equipment). Furthermore all configurations shall comply with the valid version
of the system standard IEC60601-1-1. Everybody who connects additional
equipment to the signal input connector or signal output connector configures
a medical system, and is therefore responsible that the system complies with
the requirements of the valid version of the system standard IEC 60601-1-1. If
in doubt, consult our technical service department or your local distributor.
WARNING: Pocket Fetal Doppler is a tool to aid the healthcare professional and
should not be used in place of normal fetal monitoring. WARNING: Replacing
battery shall only be done outside the patient environment (1.5m away from the
patient). CAUTION: The device must be serviced only by authorized and
qualified personnel. CAUTION: The main unit is designed for continuous
operation and is `ordinary’. Do not immerse in any liquid (i.e. not drip or
splash-proof). CAUTION: Keep the device clean. Avoid vibration. CAUTION: Do
not use high temperature sterilizing process and E-beam or gamma radiation
sterilization. CAUTION: Electromagnetic Interference-Ensure that the
environment in which the device is operated is not subject to any sources of
strong
ENGLISH
24
electromagnetic interference, such as radio transmitters, mobile telephones, etc. Keep them far away. CAUTION: The user must check that the equipment does not have visible evidence of damage that may affect patient safety or monitoring capability before use. The recommended inspection interval is once per month or less. If damage is evident, replacement is recommended before use. CAUTION: The following safety checks should be performed once every two years or as specified in the institution’s test and inspection protocol by a qualified person who has adequate training, knowledge, and practical experience to perform these tests. · Inspect the equipment for mechanical and functional damage. · Inspect the safety relevant labels for legibility. · Verify that the device functions properly as described in the instructions for use. · Test the patient leakage current according to IEC 60601-1: Limit: 10 uA (CF). The leakage current should never exceed the limit. The data should be recorded in an equipment log. If the device is not functioning properly or fails any of the above tests, the device has to be repaired. CAUTION: The battery must be properly disposed: The battery must be properly disposed according to local regulation CAUTION: The device shall only be used if the battery cover is closed. Battery must be stored in cool and dry piace. If use rechargeable battery, to insure capability and life, please fully charge batteries before first use, normally, batteries must be continuously charged over 14 hours or charged according to the guidance displayed on the battery. CAUTION: Please don’t set anode and cathode of the battery wrongly. CAUTION: The valid period of this product is five years. After the service life, please return the products to the manufacture or disposeal the products according to local regulations. When cleaning the machine: CAUTION: Don’t use strong solvent, for example, acetone. CAUTION: Never use an abrasive such as steel wool or metal polish. CAUTION: Do not allow any liquid to enter the product, and do not immerse any parts of the device into any liquids. CAUTION: Avoid pouring liquids on the device while cleaning. CAUTION: Don’t remain any cleaning solution on the surface of the device.
25
ENGLISH
When disinfecting the machine: WARNING: Never try to sterilize the probe or equipment by low temperature steam or other methods.
Refer to accompanying documents.
Chapter 2 INTRODUCTION 2.1 Overview
Pocket Fetal Doppler is a hand-held obstetrical unit, which can be used in
hospital, clinic and home for daily self-check by pregnant woman. The device
uses color LCD of high resolution to display the fetal heartbeat waveform, and
figure out the FHR to help the doctor diagnose in time. It contains components
of ultrasonic signal transmitter and receiver, analog signals processing unit,
FHR calculating unit, LCD display control unit etc. It has 3 work modes: real-
time FHR display mode, averaged FHR display mode, and manual mode. It also has
audio output, and can be connected with earphone or recorder with audio input.
2.2 Features
· Beautiful shape, portable, easy operation. · The probe has bending structure
which is easy to operate and can increase
the ease of the pregnant women, embodies the human care design. · Fetal heart
rate values, bar graph and heartbeat waveform color screen
display. · Alarming in red when the fetal heart rate range is out of the
norma! range. · Battery status indicator. · The probe can be changeable. ·
Probe inspection. · Built-in speaker. · Output for headphone. · Auto shut off.
· Two pieces of standard 1.5V alkaline battery available which can work no
less
than 8 hours.
ENGLISH
26
Chapter 3 OUTLOOK AND CONFIGURATION
LCD
Mode Button Power Button
Start/Stop Button Probe
Fig. 3-1 Front Panel
Fig. 3-2 Rear Panel
Speaker Battery Compartment
27
ENGLISH
Headphone
Fig. 3-3 Top Panel
Headphone Socket
3.1 Display
The LCD display is as follows: Probe Nominal Frequency
Work Mode
Battery Status Indicator
FBR Value
FHR Bar Graph
Fig. 3-4 LCD Display
FHR Heartbeat Waveform
3.2 Push Button
There are three push button (Power, Mode, and Start/Stop) and a volume control
button on Pocket Fetal Doppler. The primary functions are as follows:
3.2.1 Power Button
Function: Power on/off. Power on: Push the button once. Power off: Push down the button and hold 3 seconds to power off.
ENGLISH
28
3.2.2 Mode Button
Function: Mode selection, press once to enter next working mode under working
status. For the Fetal Doppler has memory function, when turning on the
machine, it will enter the mode selected before last power off automatically
after self testing.
3.2.3 Start/Stop Button
Function: Start/Stop control. Under model 3, press this button the fetal heart
rate counting starts, press this button again the counting stops.
3.2.4 Volume Control Indicator
Volume adjusting direction indicator. From left to right means that the sound
level is from high to low.
3.3 Headphone Socket
Headphone Socket: a socket for audio output, and can be connected with
earphone or recorder with audio input to record.
The socket, terminal post, or switch that connected with the headphones.
Accessory equipment connected to the analog and digital interfaces must be
certified according M the respective IEC standards (e.g. IEC 950 for data
processing equipment and IEC 60601-1 for medical equipment). Furthermore all
configurations shall comply with the valid version of the system standard
IEC60601-1-1. Everybody who connects additional equipment to the signal input
connector or signal output connector configures a medical system, and is
therefore responsible that the system complies with the requirements of the
valid version of the system standard IEC60601-1-1. If in doubt, consult our
technical service department or your local distributor.
29
Headphone Socket Signal Interface:
ENGLISH
1 2
3
4 Fig. 3-5 Headphone Socket for Audio Output
Headphone socket showed as Fig.3-5, the definition of pins showed as below:
Pin Definition
1
Signal
2
Signal
3
Signal
4
Signal
Chapter 4 GENERAL OPERATION 4.1 FHR Inspection
1. Power on by pressing the Power button. The LCD display is as Fig.3-4. 2.
Find the position of fetus: At first, please feel the position of the fetus by
hand. Find out the best direction for inspecting the fetal heart. Apply a
liberal amount of gel to the faceplate of probe; place the faceplate of probe
at the best position for detecting fetal heart. Adjust the probe to obtain an
optimum audio signal ideally by angling the probe around. Adjust the volume
according to requirements. 3. FHR Calculation: LCD displays fetal heart rate
values, bar graph and fetal heartbeat waveform. 4. Turn off the machine: Keep
pressing the power button 3 seconds to turn off.
CAUTION 1. Put the probe on the best detecting position to get better
detecting effect. 2. Don’t put the probe on the position where have strong
Placental Blood Sound (PBS) or strong Umbilical Sound (UMS).
ENGLISH
30
3. If pregnant woman adopts horizontal position and the fetus position is normal, put the probe on the position of lower navel midline to get the clearest FHR sound. 4. Do not measure FHR unless audible fetal sound has been heard. 5. Please reduce the time of ultrasonic radiation as possible as you can.
4.2 Mode Selection 4.2.1 Real-time FHR Display Mode (Mode 1)
The moment when the fetal heart rate signals are detected, the fetal heart
rate bar graph on LCD indicates the strength of the fetal heart rate signals,
and meanwhile shows the fetal heart rate values and fetal heartbeat waveform.
4.2.2 Averaged FHR Display Mode (Mode 2) This model is able to acquire more
stable fetal heart rate, displaying on LCD the latest acquisition of eight
points fetal heart rate on average. When the fetal heart rate is shown, the
fetal heart rate bar graph on LCD indicates the strength of the fetal heart
rate signals, the shown fetal heart rate values and heartbeat waveform changes
slowly.
4.2.3 Manual Mode (Mode 3) Press the start / stop button starts counting,
fetal heart rate reads as ” – – -“, the moment when the fetal heart rate
signals are detected, the fetal heart rate bar graph indicates the fetal heart
rate strength. Once again press the start / stop button to stop counting, the
equipment will automatically calculate the average fetal heart rate acquired
from the beginning to the end, and also the result will be displayed.
Numerical fetal heart rate will always remain until a repeated measurements or
patterns of change.
4.3 Probe Operation 4.3.1 Inspecting Probe
When the probe falls away from the the device, LCD screen displays the “- – -”
and displays “Probe fall !”.The probe frequency data disappeared. At this
moment the probe needs to be reconnected. After connected well, LCD screen
will clear away the “Probe fall !”and display the probe frequency data.
31
ENGLISH
4.3.2 Replacing Probe
There has been a probe connected to device while packaged by the manufacturer.
If users need to replace it with another probe, power off the device at first,
then take out the probe from the parking of device. And then pull out the plug
of the probe from its socket. Then connect the plug of the probe which needs
to be displaced with the socket. Note: Place the temporarily unused probe
carefully and avoid falling off, stress, etc. When the device is not used for
a long time, users are recommended to connect the plug of one probe to device
socket and put the probe in the parking. Then pack the device with the probe
in the wrapping box.
4.3.3 Taking out Probe and Placing Probe 1. Taking Out The Probe Hold the main unit with one hand, and hold the handle of the probe with another hand to take out the probe. (See Fig.4-1).
Fig. 4-1 Taking out Probe
2. Placing Probe It is opposite to take out probe. Hold the main unit with
one hand, and hold the top of the probe with another hand, then push the probe
into the probe holder.
ENGLISH
32
4.4 FHR Alarm
The normal fetal heart rate range is 120 BPM-160 BPM, LCD displays the fetal
heart rate numerical values as green; when the fetal heart rate is too fast or
too slow, beyond the normal range, the fetal heart rate numerical values red
alarm to remind pregnant women to go to hospital for further checks to ensure
the fetal safety.
4.5 Battery Status Indicator
When it works normally, the LCD screen displays the status of the battery as
follows:
Battery power is full
Battery power is not full
Battery power is about to run out, it needs replacing batteries.
When this machine detected the battery power is not able to maintain the
normal working of the system, LCD indicates “Low Power!”, and meanwhile the
battery power state indicative marks
is flashing, later the system will
automatically shut down.
33
ENGLISH
4.6 Replacing Battery
1. The rear panel is upturned. First open the battery compartment, then take
out the battery from the battery compartment (see Fig.4-2).
Fig. 4-2 Replacing Battery
2. Put two AA size batteries into the battery compartment (as for the
direction of battery, please refer to the instruction inside the battery
compartment), at last close the battery compartment.
CAUTION: The battery must be taken out from the battery compartment if the
device will not be used for a long time.
ENGLISH
34
Chapter 5 KEY OF SYMBOLS
Symbol Description
Symbol Description
Type CF applied part
Product code
Follow instructions for use Headphone socket Volume adjust
Lot number Serial number Manufacturer
WEEE disposal
Authorized representative in the European community
Medical Device complies with Directive 93/42/EEC
Caution: read instructions (warnings) carefully
Date of manufacture Keep in a cool, dry place Keep away from sunlight Direct current
Chapter 6 PRODUCT SPECIFICATION
Product Name: Pocket Fetal Doppler Safety: Complies with: IEC 60601-1:2005
Classification: Anti-electroshock Type: Internally powered equipment. Anti-
electroshock Degree: Type CF applied part Harmful Liquid Proof Degree: Main
unit: Degrees of protection provided by enclosure: IPXO. Probe: Prevent from
water splashing, degree of protection: IPX4. Degree of Safety in Presence of
Flamrnable Gases: Equipment not suitable for use in presence of flammable
gases. Working System: Continuous running equipment.
35
ENGLISH
EMC: Group I Class B. Suitable Using Range: Suitable for use after the 12th week of pregnancy. Physical Characteristic Size: 135mm (Length) X 92mm (Width) X 29 mm (Height) Weight: About 245g (including batteries) Environment Working: Temperature: +5°C~+40°C Humidity: <80% Atmospheric Pressure: 70 kPa~106 kPa Transport and Storage: Temperature: -10°C~+55°C Humidity: 93% Atmospheric Pressure: 50 kPa~106 kPa Display: 1.77″262K TFT display FHR Performance FHR Measuring Range: 50-240BPM (BPM: beat per minute) Resolution: 1BPM Accuracy: +-2BPM Power Consumption: <1 W Auto Shut-OFF: After 1 minute no signal, power off automatically. Battery Type Recommended: Two pieces of 1.5 V DC battery (SIZE AA LR6). Probe: Nominal Frequency: 2.0MHz Working Frequency: 2.0 MHz±10% P-: < 1 MPa Iob: < 20 mW/cm2 Ispta: < 100 mW/cm2 Ultrasonic Output Power: P < 20 mW Working Mode: Continuous wave doppler Effective Radiating Area of Transducer: < 157mm2 Note: In all working application modes, mechanical index: MI<1, thermal index: TI<1.
ENGLISH
36
Chapter 7 MAINTENANCE 7.1 Maintenance
The probe acoustic surface is frangible and must be handled with care. Gel
must be wiped from the probe after use. These precautions will prolong the
life of the unit. The user must check that the equipment does not have visible
evidence of damage that may affect patient safety or Pocket Fetal Doppler
capability before me. The recommended inspection interval is once per month or
less. If damage is evident, replacement is recommended before use. The
equipment should undergo periodic safety testing to insure proper patient
isolation from leakage currents. This should include leakage current
measurement. The recommended testing interval is once every two years or as
specified in the institution’s test and inspection protocol. The accuracy of
FHR is controlled by the equipment and can not be adjusted by user. If the FHR
result is distrustful, please use other method such as stethoscope to verify
immediately or contact local distributor or manufacturer to get help.
7.2 Cleaning
Before cleaning, switch off and take out the batteries. Keep the outside
surface of the device clean and free of dust and dirt, clean exterior surface
(display screen included) of the chassis with a dry, soft cloth. If necessary,
clean the chassis with a soft cloth soaked in a solution of soap, or water and
wipe dry with a clean cloth immediately. Wipe the probe with soft cloth to
remove any remaining ultrasound coupling gel. Clean with soap and water only.
CAUTION: Don’t use strong solvent, for example, acetone. CAUTION: Never use an
abrasive such an steel wool or metal polish. CAUTION: Do not allow any liquid
to enter the product, and do not immerse any parts of the device into any
liquids. CAUTION: Avoid pouring liquids on the device while cleaning. CAUTION:
Don’t remain any cleaning solution on the surface of the device
Note: Wipe the surface of probe with 70% ethanol, self-air dry, or clean with
a clean, dry cloth.
37
ENGLISH
7.3 Disinfecting and Sterilization
Clean the equipment case, probe, etc. as above, and then wipe the probe with
an alcohol impregnated wipe (70% ethanol). Wipe the probe with a clean, dry
cloth to remove any remaining moisture. NOTE: 1. The recommended periods of
cleaning, sterilization and disinfecting is once per month. 2. After cleaning,
sterilization and disinfecting, users must inspect whether have any obvious
damage which may affect the patient safety and instrument performance.
WARNING: Never try to sterilize the probe or equipment by low temperature.
Chapter 8 SOLUTIONS FOR POSSIBLE PROBLEMS
If it appears following problems when you use the device, please solve them as
below:
Problems Weak sound
Possible reasons
· Volume is too low · Power is low · Did not daub the gel
Solutions
· Adjust the volume louder · Change the battery · Daub the gel
Noise
· Probe is too near from the main unt
· Disturbance from the outside signal
· Power is low
· Make the distance between the probe and the main unit a little further
· Keep far away from the outside signal
· Change the battery
Low sensitivity
· Position of the probe is not correct
· Did not daub the gel
· Adjust the position of the probe
· Daub the gel
ENGLISH
38
Appendix 1
Essentiality of Fetal Domestic Monitor Modem medicine think that: FHR is an
important gist to identify fetal health, by recording FHR changes can observe
fetal hypoxia, fetal distress and the umbilical cord around the neck, and
other symptoms. Fetal domestic monitor test FHR rate changes by listening to
fetal heart sound mainly. Fetal domestic monitor is a powerful guarantee to
improve generational safety. Fetal heart rate changes most obviously in the
following three periods: · Within 30 minutes after pregnant women get up ·
Within 60 minutes after pregnant women finish lunch · Within 30 minutes before
pregnant women go to bed For the above three periods, because of the change of
the body status of pregnant women, the activity of food digesting needs the
body to provide more oxygen, relatively, the oxygen for fetus become less. It
is easy to arose symptoms such as fetus anoxia. Testing the FHR at this time
can display the healthy status for the fetus best. The above three periods can
only be tested at home by pregnant women themselves, so FHR domestic monitor
is very important. We advise the pregnant women to measure every day
respectively at early, middle and late time, every time measuring the fetal
heart rate and listening to the fetal heart rate for about one minute, and
recording the measurement results for the medical reference when go to the
hospital. Generally, the medicine deems the normal fetal heart rate as:
120BPM~160BPM; slightly too fast: 161BPM~180BPM; heavily too fast: above
181BPM; slightly too slow: 119BPM~100BPM; heavily too slow: below 99BPM. This
device can hear the fetal heart sound for fetus above twelve weeks, and check
the LCD display. Finding too fast or too slow should go to hospital for
further checks to ensure safety.
39
Appendix 2
Overall Sensitivity
Diameter of Target Reflector (mm)
Distance Reflection
(d)
Loss
(mm)
A(d)
Two-way Attenuation
B= Ba +Bw
Ba
(T:mm Ba:dB)
Bw B (dB) dB
Vs r.m.s mV
( ) Vn
r.m.s
C=20 log10
Vs (r.m.s) Vn (r.m.s)
Overall Sensitivity
mV
(S=A(d)+B+C)
dB
dB
1.58
50
45.7 T 20 4.8 4.0 – 0 57.6 186 94
5.93
A=45.7dB@ 2MHz
Ba 40 9.6 8.0 –
75
45.7 T 20 4.8 3.4 – 0 56.4 175 90
5.78
109.2 107.8
Ba 40 9.6 6.8 –
100
45.7 T 20 4.8 3.4 – 0 56.4 174 89
5.82
107.9
Ba 40 9.6 6.8 –
200
45.7 T 20 4.8 – – 0 49.6 173 90
5.68
100.9
Ba 40 9.6 – –
2.38
50
43.2 T 20 4.8 3.4 2.2 0 60.8 178 89
6.02
A=43.2dB@ 2MHz
Ba 40 9.6 6.8 4.4
75
43.2 T 20 4.8 3.4 1 0 58.4 170 90
5.52
110.0 107.1
Ba 40 9.6 6.8 2
100
43.2 T 20 4.8 3.4 – 0 56.4 165 85
5.76
105.3
Ba 40 9.6 6.8 –
200
43.2 T 20 4.8 1 – 0 51.6 160 85
5.49
100.2
Ba 40 9.6 2 –
ENGLISH
Doppler Frequency
333
Velocity
12.5
(Hz)
of Target
(cm/s)
ENGLISH
40
Disposal: The product must not be disposed of along with other domestic waste. The users must dispose of this equipment by bringing it to a specific recycling point for electric and electronic equipment. For further information on recycling points contact the local authorities, the local recycling center or the shop where the product was purchased. If the equipment is not disposed of correctly, fines or penalties may be applied in accordance with the national legislation and regulations.
GIMA WARRANTY CONDITIONS
Congratulations for purchasing a GIMA product. This product meets high
qualitative standards both as regards the material and the production. The
warranty is valid for 12 months from the date of supply of GIMA. During the
period of validity of the warranty, GIMA will repair and/or replace free of
charge all the defected parts due to production reasons. Labor costs and
personnel traveling expenses and packaging not included. All components
subject to wear are not included in the warranty. The repair or replacement
performed during the warranty period shall not extend the warranty. The
warranty is void in the following cases: repairs performed by unauthorized
personnel or with non-original spare parts, defects caused by negligence or
incorrect use. GIMA cannot be held responsible for malfunctioning on
electronic devices or software due to outside agents such as: voltage changes,
electro-magnetic fields, radio interferences, etc. The warranty is void if the
above regulations are not observed and if the serial code (if available) has
been removed, cancelled or changed.The defected products must be returned only
to the dealer the product was purchased from. Products sent to GIMA will be
rejected.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>