GIMA 28243 Aspiratore Supervega Battery User Manual
- June 6, 2024
- GIMA
Table of Contents
SUPERVEGA BATTERY SUCTION ASPIRATOR –
ON TROLLE
SUPERVEGA BATTERY SUCTION UNIT is a surgical suction pump with the
following electrical characteristics: 14V 4A with AC/DC (input: 100-240V~ –
50-60Hz – 1.5A) to be used for nasal, oral, and tracheal aspirations in adults
or for body liquids in children (for example mucus, phlegm, and blood)
equipped with 5 wheels with the braking device. Thanks to these
characteristics and its high-performance levels, this device is particularly
suitable for use in hospital wards, in small surgical applications, and in
post-operative treatments. A device designed to offer ease-of-transport and
almost continuous use thanks to the adoption of an electronic system to manage
the power supply. Supplied with an audible alarm and visual tell-tale
(luminous LED) to indicate the battery status. Featuring a body in plastic
with high thermal and electrical insulation in compliance with recently
introduced European safety regulations. Supplied with no. 2 complete
sterilizable polycarbonate jugs with overflow valve. Features a suction
regulator and vacuum gauge located on the front panel.
GENERAL WARNING
Read the instruction manual carefully before use.
Only highly qualified staff use reserved.
The instrument must not be disassembled. For a technical service always
contact Gima.
Keep off the reach of children or not capable people without supervision.
Full containers must be handled with great care during transfer to the
disposal areas, following the local procedures and regulations.
IMPORTANT SAFETY RULES
-
Check the condition of the unit before each use. The surface of the unit should carefully be inspected for visual damage. Check the mains cable and do not connect to power if the damage is apparent;
-
Before connecting the appliance always check that the electric data indicated on the data label and the type of plug used, correspond to those of the mains electricity to which it’s to be connected;
-
Respect the safety regulations indicated for electrical appliances and particularly:
• Use original components and accessories provided by the manufacturer to guarantee the highest efficiency and safety of the device;
• The device can be used only with the bacteriological filter;
• Never immerse the appliance into the water;
• Position the device on stable and flat surfaces in a way that the air inlets on the back aren’t obstructed;
• To avoid incidents, do not place the aspirator on unstable surfaces, which may cause it to accidentally fall and lead to a malfunction and/or breakage. Should there be signs of damage to the plastic parts, which may expose the inner parts of the energized device, do not connect the plug to the electrical socket? Do not attempt to make the device work before it has been thoroughly checked by qualified personnel and/or the GIMA technical service department.
• Don’t use in the presence of inflammable substances such as anesthetic, oxygen, or nitrous oxide;
• Don’t touch the device with wet hands and always prevent the appliance from coming into contact with liquids;
• Don’t leave the appliance connected to the power supply socket when not in use;
• Don’t pull the power supply cable to disconnect the plug remove the plug from the mains socket correctly;
• Store and use the device in places protected against the weather and far from any sources of heat. After each use, it is recommended to store the device in its own box away from dust and sunlight.
• In general, it is inadvisable to use single or multiple adapters and/or extensions. Should their use be necessary, you must use ones that are in compliance with safety regulations, however, take care not to exceed the maximum power supply tolerated, which is indicated on the adapters and extensions.
• Prevent children from using the device without proper supervision; -
For repairs, exclusively contact technical service and request the use of original spare parts. Failure to comply with the above can jeopardize the safety of the device;
-
Use only for the purpose intended. Don’t use for anything other than the user defined by the manufacturer.
The manufacturer will not be responsible for damage due to improper use or connection to an electrical system not complying with current regulations. -
The medical device requires special precautions regarding electromagnetic compatibility and must be installed and used in accordance with the information provided with the accompanying documents: the SU-PERVEGA BATTERY SUCTION UNIT device must be installed and used away from mobile and portable RF communication devices (mobile phones, transceivers, etc.) that may interfere with the said device.
-
WARNING: Do not change this equipment without the permission of the manufacturer GIMA S.p.A. None of the electric or mechanical parts has been designed to be repaired by customers or end-users. Don’t open the device, do not mishandle the electric/mechanical parts. Always contact technical assistance
-
Using the device in environmental conditions different than those indicated in this manual may harm seriously the safety and the technical characteristics of the same.
-
The medical device is in contact with the patient by means of a disposable probe (not supplied with the de-vice). Suction tubes for insertion in the human body purchased separately from the machine should comply with ISO 10993-1 standard on material biocompatibility.
-
The product and its parts are biocompatible in accordance with the requirements of regulation EN 60601-1.
-
The operation of the device is very simple and therefore no further explanations are required other than those indicated in the following user manual.
-
The lead battery integrated into the device is not to be considered ordinary domestic waste. Such a component must be disposed of in a specific collection center in order to be recycled.
CONTRAINDICATIONS
- Before using the SUPERVEGA BATTERY SUCTION UNIT, consult the instructions for use: failure to read all the instructions in this manual can be harmful for the patient.
- The device cannot be used to drain chest fluids;
- The device must not be used for suction of explosive, corrosive, or easily flammable liquids.
- SUPERGA BATTERY SUCTION UNIT is not suitable for MRI. Do not introduce the device in MRI environments.
The manufacturer cannot be held liable for accidental or indirect damages should the device be modified, repaired without authorization or should any of its components be damaged due to accident or misuse. Any minimal modification / repair on the device voids the warranty and does not guarantee the compliance with the technicalrequirements provided by the MDD 93/42/EEC (and subsequent changes) and it’s normative.
TECHNICAL CHARACTERISTICS
Model | SUPERVEGA BATTERY SUCTION UNIT |
---|---|
Typology (MDD 93/42/EEC) | Class Ila Medical device |
UNI EN ISO 10079-1 Classification | High Vacuum / High Flow |
Power Feeding | 14V 4A with AC/DC adapter (Input: 100-240V∼ – 50/60Hz – 100VA) |
or Internally powered equipment (Pb Battery
12V 4A)
Current Consumption| 4.0A
Maximum Suction Pressure (without jar)| -80kPa (-0.80 Bar)
Minimum Suction Pressure (without jar)| Less -40kPa (-0.40 bar)
Maximum Suction Flow (without jar)| 36 l/min
Insulation Class (when used with the AC/DC adapter)| Class II
Insulation Class (when used with an internal battery)| Internally Powered
Equipment
Weight| 7.98 Kg
Size| 320 x 990 (h) x 300 mm
Battery Holding Time| 60 minutes
Battery Time Charge| 240 minutes
Dimension of the silicon aspiration tube| 0 8×14 mm
Accuracy of Vacuum Indicator| * 5%
Working Condition| Room Temperature: 5 ÷ 35°C
Room Humidity Percentage: 30 ÷ 75% RH
Atmospheric pressure: 800 ÷ 1060 hPa
Altitude: 0 ÷ 2000m s.l.m.
Conservation condition and Transport| Room Temperature: – 40°C ÷ 70°C
Room Humidity Percentage: 10 ÷ 100% RH
Atmospheric pressure: 500 ÷ 1060 hPa
The technical specifications may change without notice.
CLEANING OF THE DEVICE
Use a soft dry cloth with not – abrasive and not – solvent detergents. To
clean the device’s external parts always use a cotton cloth dampened with
detergent. Don’t use abrasive or solvent detergents. Before carrying out any
cleaning and/or maintenance operation, disconnect the appliance from the power
supply, unplug it or turn off the switch on the device.
Particular care should be taken to ensure that the internal parts of the
equipment do not get in touch with liquids. Never clean the equipment
underwater.
During all clearing operations use protective gloves and apron (if need be,
also wear a face mask and glasses) to avoid getting in contact with
contaminating substances (after each utilization cycle of the machine).
ACCESSORIES SUPPLIES
- N°2 complete aspiration jar 2000ml
- Conical fitting
- Tubes set 8 mm x 14 mm
- Hydrophobic and antibacterial filter
- AC/DC adapter
- European power supply cord for AC/DC adapter
The filter is produced with (PTFE) hydrophobic material to prevent fluids
from entering the pneumatic circuit.
It should be changed immediately if it becomes wet or if there is any sign of
contamination or discoloration.
It should also be changed if the unit is used with a patient whose risk of
contamination is unknown. Don’t use the suction unit without the protection
filter. If the suction unit is used in an emergency or in a patient where the
risk of contamination is not known the filter must be changed after each use.
The filter is not designed to be decontaminated, disassembled, and/or
sterilized. If the patient’s pathology is known and/or no risk of indirect
contamination exists, it is advisable to replace the filter after each work
shift or at any rate on a monthly basis even if the device is not used.
WARNING: Suction tubes for insertion in the human body purchased
separately from the machine should comply with ISO 10993-1 standard on
material biocompatibility.
Aspiration jar: the mechanical resistance of the component is guaranteed
up to 30 cycles of cleaning and sterilization. Beyond this limit, the
physical-chemical characteristics of the plastic material may show signs of
decay. Therefore, we recommend that you change it.
Silicone tubes: the number of cycles of sterilization and/or cleaning is
strictly linked to the employment of the said tube. Therefore, after each
cleaning cycle, it is up to the final user to verify whether the tube is
suitable for reuse. The component must be replaced if there are visible signs
of decay of the material constituting the said component.
Conical fitting: the number of cycles of sterilization and the number of
cleaning cycles is strictly linked to the employment of the said component.
Therefore, after each cleaning cycle, it is up to the final user to verify
whether the fitting is suitable for reuse. The component must be replaced if
there are visible signs of decay in the material constituting the said
component.
The service life of the device: more than 1000 hours of operation (or 3
years) in accordance with the standard conditions of testing and operation.
Shelf life: maximum 5 years from the date of manufacture.
WARNING: The medical device is provided without a specific suction probe.
If this device must be used with a specific suction probe, the end-user is
responsible for making sure it complies with the EN 10079-1 regulation.
CLEANING OF ACCESSORIES
Before using the device, the manufacturer advises you to clean and/or
sterilize the accessories.
Washing and/or cleaning the autoclavable jar as to be carried out as follows:
-
Wear protective gloves and apron (glasses and face mask if necessary) to avoid contact with contaminating substances;
-
Disconnect the tank from the device and remove the said container from the support of the device.
-
Separate all the parts of the cover (overflow device, washer).
-
Disconnect all tubes from the jar and the protection filter
-
Wash each part of the container from secretions under cold running water and then clean every single part in hot water (temperature not exceeding 60°C)
-
Once again, carefully wash every single part using, if necessary, a non-abrasive brush to remove any deposits.
Rinse with hot running water and dry all parts with a soft cloth (non- abrasive). It is possible to wash with commercial disinfectants by carefully following the instructions and dilution values supplied by the manufacturer.
After cleaning, leave the parts to dry in an open, clean environment. -
Dispose of the aspiration catheter according to that provided by local laws and regulations.
The silicone aspiration tubes and the conical fitting may be carefully washed
in hot water (temperature must not exceed 60°C). After cleaning, leave the
parts to dry in an open, clean environment.
When cleaning is complete, reassemble the container for liquid aspirations
according to the following procedure:
- Place the overflow valve into its seat in the cover (under VACUUM connector)
- Insert floating valve keeping the o-ring towards the opening of the cage
- Place the o-ring into its seat around the cover
- After completing assembling operations always make sure that covers seals perfectly to avoid vacuum leakages or liquid exit
The jar and the cover can be autoclaved by placing the parts into the
autoclave and running one sterilization stem cycle at 121°C (1 bar relative
pressure – 15 min) making sure that the jar is positioned upsidedown.
Mechanical resistance of the jar is guaranteed up to 30 cycles of
sterilization and cleaning at the indicated conditions (EN ISO 10079-1).
Beyond this limit, the physical-mechanical characteristics of the plastic may
decrease and replacement of the part is therefore recommended.
After sterilization and cooling at an environmental temperature of the parts
make sure that these are not damaged.
The aspiration tubes can be sterilized on autoclave using a sterilization
cycle at 121°C (1 bar relative pressure – 15 min).
The conical connector can be sterilized on autoclave using a sterilization
cycle at 121°C (1 bar relative pressure – 15 min).
DO NOT WASH, STERILIZE OR PUT IN AUTOCLAVE THE ANTIBACTERIAL FILTER
Disposing of disposable bags:
If the device is fitted with FLOVAC® disposable collection systems (consisting
of a rigid, reusable polycarbonate container and a single-use polyethylene
collection bag), proceed with the disposal of the bag as follows: Deactivate
the suction source and remove all tubes connected to the container, paying
particular attention to avoid accidental contamination. Attach the plugs to
the “PATIENT” and “TANDEM” connectors by inserting them firmly, paying
particular attention to avoid accidental contamination. Take the device to the
waste collection area with all the openings correctly sealed, taking into
consideration that the product may potentially be infected. Discard the
product in compliance with the regulations in force at the hospital.
PERIODICAL MAINTENANCE CHECKS
The SUPERVEGA BATTERY SUCTION UNIT suction equipment does not need maintenance
or lubrication. It is, however, necessary to inspect the unit before each use.
With regard to training, given the information contained in the user manual
and since it is easy to understand the said device, it doesn’t appear to be
necessary. Unpack the instrument and always check the integrity of plastic
parts and AC/DC switching adapter, feeding cable, they might have been damaged
during previous use. Connect the cable to the electrical network and turn the
switch on. Close the aspirator outlet with your finger and with the suction
regulator in maximum vacuum position check that the vacuum indicators reach
-80kPa (-0.80 bar) minimum (internal battery). Rotate the knob from right to
left and check the aspiration regulating control. The vacuum indicator should
go down -40kPa (-0.40 bar). Verify that loud noises are not present, these
can indicate wrong functioning. Internally, the device is protected (see
electrical specifications) by two fuses F1, F2 (T 15A L 125V)
that cannot be reached from the outside. Therefore, contact the manufacturer
to request the assistance of an authorized and qualified technician when they
need to be replaced. If it’s replaced make sure that its replacement is
always the same type and value, as indicated. The device is made up of a lead
battery that cannot be accessed from the outside. In order to replace it,
consult the technical staff authorized by the manufacturer.
USE ONLY THE RECOMMENDED BATTERIES FROM GIMA. THE USE OF OTHER BATTERIES IS
NOT RECOMMENDED AND INVOLVES THE CANCELLATION OF THE WARRANTY
In the event that the service personnel has to replace the internal battery,
pay special attention to the polarity of the same component. The + / –
polarities are indicated directly on the battery.
Fault type
| Cause|
Solution
---|---|---
1. Red light on| Battery run down| Hook up the power cord to the electricity
mains, positioning the equipment power switch on 0.
2. No light| Defective AC/DC adapter or technical internal problem| Contact
the technical service.
3. No aspiration| Jar Cap badly screwed down| Unscrewed the cap, then re-
screw it correctly
4. No aspiration| Lid seal is not in its seat| Unscrew the cap and insert
the seal properly in its seat
5. The float doesn’t close| If the cap has been washed, ensure that the
float is not partially detached| Insert the float into its place
6. The float doesn’t close| The float it’s covered by dirty material|
Unscrewed the cap, leave the and put it on autoclave
7. Low suction| The foam inside the jar| Fill the jar to 1/3 full of
ordinary water
8. No aspiration due to flow leakage of mucus| Filter blocked| Replace
filter
9. The Vacuum power on the patient side is either very low or absent|
- Vacuum regulator set to minimum
- Protection filter blocked or damaged
- Connection tubes blocked, kinked or disconnected
- Shut-off valve blocked or damaged
- Pump motor damaged
|
-
Turn the vacuum regulator clockwise and
check the value of the vacuum on the gauge -
Replace the filter
-
Replace or reconnect the tubes, check the jar connections
-
Empty the jar, or disconnect the tube from the jar and unblock the shut-off valve. The unit will only work in the upright position
-
Contact the technical service
10. Noisy| Technical internal problem| Contact the technical service
Faults 1 – 2 – 3 – 4 – 5 – 6 – 7 – 8 – 9 – 10| None of the remedies has
achieved the desired results| Contact the seller or GIMA After-sales
Assistance Service
If the overfill security system it’s activated, don’t proceed with the liquid
aspiration. If the overfill security the system doesn’t work there are two
cases:
1° case – If the overfill security system doesn’t work the aspiration will be
stopped by the bacteriological filter who avoiding the liquid penetration
inside the device.
2° case – If both the security system doesn’t work, there is the possibility
that liquid comes inside the device, in this case, return the device to GIMA
technical service.
GIMA S.p.A. will provide upon request electric diagrams, components list,
descriptions, setting instructions, and any other information that can help
the technical assistance staff for product repair.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING,
PLEASE CONTACT GIMA TECHNICAL SERVICE. GIMA DOES NOT GIVE A GUARANTEE IF THE
INSTRUMENT, AFTER THE TECHNICAL SERVICE CHECKING, APPEARS
TO TAMPER.
INSTRUCTION FOR USE
Assembly of the device SUPERVEGA BATTERY SUCTION ASPIRATOR: Take the 5 arm base and set up the 5 wheels that come with the above device. The wheels provided with the braking device must be placed one next to the order. Take the support bar that comes with the device SUPERVEGA BATTERY SUCTION ASPIRATOR and place it in the hole on the 5-arm base. From under the base, lock the two parts by means of the supplied screw. Eventually, place the device on the trolley.
- The device must be checked before each use in order to detect malfunctions and/or damage caused by transport and/or storage.
- The working position must be such as to allow one to reach the control panel and to have a good view of the empty indicator, the jar, and the antibacterial filter.
- If the device is to be transported from one place to another, to prevent the liquid collection jar from falling and consequently the liquid from spilling, removing the jar from the device is recommended.
WARNING: For proper use, place the aspirator on a flat, stable surface,
in order to have the full volume of use of the jar and better efficiency of
the overflow device. The vacuum jar, during use, must be used in vertical
mode, to prevent the action of the backflow valve. If this protection is
triggered, turn the device off and disconnect the pipe connected to the vacuum
jar (indicated with the word VACUUM) on its cover.
Operation with AC/DC power pack:
- Connect the short silicone tube with the antibacterial filter to the suction port. The other tube, connected to the filter on one end, must be connected to the spout on the vase lid with the float assembled inside (overflow device). The overflow device is triggered (the float closes off the internal lid fitting) when the maximum volume (90% of the effective vase volume) is reached, thus preventing the liquid from penetrating the inside of the machine. The device must be used on a flat, horizontal surface.
- Connect the long silicone tube to the free spout on the lid; the free end of the tube must be connected to the conical fitting for the probe coupling, to which the suction probe must then be connected.
- Connect the universal power pack to the device using the dedicated connector and insert the power cord plug into the socket. To start the treatment, press the switch into position I to turn the device on
- Set the desired vacuum level (Bar / kPa) through the vacuum regulator. Turn the knob in a clockwise direction to increase the vacuum level: these values can be read on the “vacuum gauge”.
- To suspend and/or end the treatment, press the switch again and pull the plug out of the socket
- To mitigate the formation of foam inside the vase, unscrew and remove the lid from the vase, and fill the latter with 1/3 water (to facilitate cleaning operations and speed up depressurization during operation), then screw the lid back onto the vase.
- Remove the accessories and proceed with cleaning operations.
- At the end of each use place the device back in the box, protected against dust
WARNING: The power cord plug is the element of separation from the
electrical mains, even if the device is equipped with an on/off button. Once
the device is in use, the power plug must remain accessible to allow another
method of disconnection from the electrical mains.
Filter assembling
Make sure the filter is assembled with the arrows on the side of the
patient.
WARNING: The inside of the medical device must be regularly checked for
the presence of liquids or other visible contamination (secretions). In the
presence of liquids or other visible contamination, immediately replace the
medical device due to the risk of an insufficient vacuum flow rate.
These products have been designed, tested, and manufactured exclusively for
single patient use and for a period no longer than 24 hours.
Operation with Internal Battery
- Press the switch in the position I to turn the device on (the external power pack doesn’t need to be connected)
- The fully charged battery life is about 60 minutes with continuous operation.
WARNING: Before using the device, check the battery power status. Before
each use proceed with charging the battery. To maintain the device in good
conditions, recharge the battery every 3 months (when not in use).
Recharging operations: to be able to charge the internal battery it is
necessary to connect the universal switching adapter to the electric network
for approx. 240 minutes with the main switch to position 0.
TAB. I – INDICATOR LIGHTS DURING OPERATIONS
When an external power supply (regardless of the state of the battery charger)
and when the device is working (after having turned it on), the LED stays in a
FIXED GREEN
position.
LED Signal
| Phase| Problem / Cause|
Solution
---|---|---|---
Flashing Green Led| During recharge| Battery recharge running| Wait
Steady Greed Led| During recharge| Recharging cycle complete| Remove power
supply
Steady Red Led| During battery operation| Flat battery| Start recharging cycle
WARNING: During this signal, you will hear a long, continuous beep
(duration of sound 0.8 sec/sound frequency: every 8.5 sec), which notifies the
user regarding the battery discharge.
Flashing Red Led| The device automatically turns off when the battery is flat|
Battery completely flat| When the device is restarted the LED will flash red:
begin the battery recharge cycle immediately
Steady Orange Led| During battery operation| Intermediate status| Guaranteed
battery function / Recharge when the red LED signal comes on.
NEVER USE THE DEVICE WITHOUT A JAR AND / OR PROTECTION FILTER
Using FLOVAC® disposable collection system:
Before connecting the disposable collection system, remove the blu ring fitted
on the tank holder for more comfortable insertion of the same container.
- After opening the package, fully stretch the bag and then flatten it concentrically to eliminate as much air as possible.
- Insert the bag and apply the cover to an appropriately sized reusable rigid container by pressing firmly around the entire perimeter. Make sure that the system is completely sealed.
- Close the connector marked as “TANDEM” with the lid provided.
- Connect the power source of the vacuum to the VACUUM port equipped with a specific reusable conical fitting with a “male” connection.
- Connect the patient tube to the PATIENT port of the cover
- Before use, check all closures and make sure there are no leaks, starting the aspiration source. If the bag expands to fully adhere to the walls of the rigid container and the cover bends towards the inside of the glass, the system is not leaking.
- Start the aspiration and periodically check the filling level of the container. The overflow valve will cause the interruption of aspiration if the aspirated fluids have reached the maximum filling level of the device.
- When the float valve intervenes signaling the device is too full, the suction source must be disconnected within no more than 5minutes.
RISK OF ELECTROMAGNETIC INTERFERENCE AND POSSIBLE REMEDIES
This section contains information regarding the conformity of the compliance
with the EN 60601-1-2 Standard (2015).
The SUPERVEGA BATTERY SUCTION UNIT surgical aspirator is an electro-medical
device that requires particular precautions regarding electromagnetic
compatibility and which must be installed and commissioned according to the
electromagnetic compatibility information supplied. Portable and mobile radio
communication devices (mobile phones, transceivers, etc.) may interfere with
the medical device and should not be used in close proximity with, adjacent
to, or on top of the medical device. If such use is necessary and unavoidable,
special precautions should be taken so that the electro-medical device
functions properly in its intended operating configuration (for example,
constantly and visually checking for the absence of anomalies or
malfunctions). The use of accessories, transducers, and cables differing from
those specified, with the ex-
the caption of transducers and cables sold by the appliance and system
manufacturer as spare parts can lead to an increase in emissions or in a
decrease in the immunity of the device or system. The following tables supply
information regarding the EMC (Electromagnetic Compatibility) characteristics
of the electro-medical device.
Guidance and manufacturer’s declaration — electromagnetic Emissions
The SUPERVEGA BATTERY SUCTION UNIT is intended for use in the electromagnetic environment specified below. The customers or the user of the SUPERVEGA BATTERY SUCTION UNIT should make sure that it’s used in such an environment.
Emissions test
| Compliance|
Electromagnetic environment – guidance
Irradiated / Conducted emissions CISPR11| Group 1| The SUPERVEGA BATTERY
SUCTION UNIT only used RF energy only for its internal functioning. Therefore,
its RF emissions are very low and are not cause interference in the proximity
of any Electronic appliances.
Irradiated / Conducted emissions CISPR11| Class [B]| The SUPERVEGA BATTERY
SUCTION UNIT can be used in all environments, including domestic and those
connected directly to the public mains distribution that supplies power to
environments used for domestic scopes.
Harmonic emissions EN 61000-3-2| Class [A]
Voltage fluctuations / flicker emissions EN 61000-3-3| Complies
Guidance and manufacturer’s declaration — Immunity Emissions
The SUPERVEGA BATTERY SUCTION UNIT is intended for use in the electromagnetic environment specified below. The customers or the user of the SUPERVEGA BATTERY SUCTION UNIT should make sure that it’s used in such an environment.
Immunity Test
| Level indicated by the EN 60601-1-2| Compliance Level|
Electromagnetic environment – guidance
Electrostatic discharge (ESD) EN 61000-4-2| +/-8kV on contact +/-15kV in air|
The device doesn’t change its state| Floors should be wood, concrete or
ceramic tile. If floors are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical fast transient/burst EN 61000-4-4| ± 2kV power supply lines
± 1 kV for input /output lines| The device doesn’t change its state| The main
power quality should be that of a typical commercial environment or hospital.
Surge
EN 61000-4-5| ± 1 kV differential mode +/-2 kV ordinary mode| The device
doesn’t change its state| Mains power quality should be that of a typical
commercial environment or hospital.
Loss of voltage, brief voltage interruptions and variations
EN 61000-4-11| <5% Ur (>95% dip UT) for 0,5 cycle
40% UT (60% dip UT) for 5 cycle
70% UT (30% dip UT) for 25 cycle
<5% UT (>95% dip UT) for 5 sec| – –| Mains power quality should be that of a
typical commercial environment or hospital If the user of the SUPERVEGA
BATTERY SUCTION UNIT request that the appliance operates continuously, the
use of a continuity unit is recommended.
A magnetic field with a network
frequency (50/60 Hz)
EN 61000-4-8| 30A/m| The device doesn’t change its state| The power frequency
magnetic field should be measured in the intended installation location to
make sure that it’s sufficiently low.
Note Ur is the value of the power supply voltage
Guidance and manufacturer’s declaration — Immunity Emissions
The SUPERVEGA BATTERY SUCTION UNIT is intended for use in the electromagnetic environment specified below. The customers or the user of the SUPERVEGA BATTERY SUCTION UNIT should make sure that it’s used in such an environment.
Immunity Test
| Level indicated by the EN 60601-1-2| Compliance Level|
Electromagnetic environment – guidance
Conducted Immunity
EN 61000-4-6
Radiated Immunity
EN 61000-4-3| 3Vrms 150kHz to 80Mhz (for non-life-supporting devices) 3V/m
80MHz to 2.7GHz (for non-life-supporting devices)| V1 = 3 V rms
El = 10 V / m| The portable and mobile RF communication devices, including
cables, must not be used closer to the SUPERVEGA BATTERY SUCTION UNIT device,
than the separation distance calculated by the equation applicable to the
transmitter frequency.
Recommended separation distance![GIMA 28243 Aspiratore Supervega Battery
- math](https://manuals.plus/wp-content/uploads/2021/11/GIMA-28243-Aspiratore-
Supervega-Battery-math-300x135.png) Where P is the maximum nominal output
voltage of the transmitter in Watt (W) depending on the manufacturer of the
transmitter and the recommended separation distance in meters (m). The
intensity of the field from the fixed RF transmitters, as determined by an
electro-magnetic study of the site), could be lower than the level of
conformity of each frequency interval b).
It is possible to check for interference in proximity to devices identified by the following symbol:
Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is
applied
Note 2: These guidelines may not be applicable in all situations. The
electromagnetic propagation is influenced by absorption and by reflection from
buildings, objects, and people.
a) The field intensity for fixed transmitters such as the base stations
for radiotelephones (mobile and cordless) and terrestrial mobile radio,
amateur radio devices, radio AM and FM transmitters, and TV transmitters can
not be theoretically and accurately foreseen. To establish an electro-magnetic
environment generated by fixed RF transmitters, an electro-magnetic study of
the site should be considered. If the field intensity measured in the place
where the device will be used surpasses the above-mentioned applicable level
of conformity, the normal functioning of the device should be monitored. If
abnormal performance arises, additional measures such as changing the
device’s direction or positioning may be necessary.
b) The field intensity on an interval frequency of 150 kHz to 80 MHz
should be less than 3 V/m.
The recommended separation distance between portable and mobile radio- communication devices and the monitor
The SUPERVEGA BATTERY SUCTION UNIT surgical aspirator is intended to operate in an electro-magnetic environment where RF irradiated interferences are under control. The client or operator of the SUPERVEGA BATTERY SUCTION UNIT device can help prevent electromagnetic interference by keeping a minimum distance between the portable and mobile RF communication devices (transmitters) and the SUPERVEGA BATTERY SUCTION UNIT device, as recommended below, in relation to the radio-communication maximum output power.
The maximum nominal output power of the Transmitter W
|
The separation distance from the frequency transmitter (m)
| |
0,01| 0,12| 0,12| 0,23
0,1| 0,38| 0,38| 0,73
1| 1,2| 1,2| 2,3
10| 3,8| 3,8| 7,3
100| 12| 12| 23
For transmitters with a maximum nominal output power not shown above, the
recommended separation distance in meters (m) can be calculated using the
equation applicable to the transmitter frequency, where P is the maximum
nominal output power of the transmitter in Watt (W) depending on the
transmitter’s manufacturer.
Note 1: At 80 MHz and 800 MHz the interval with the highest frequency
is applied
Note 2: These guidelines may not be applicable in all situations. The
electromagnetic propagation is influenced by the absorption and by the
reflection from buildings, objects, and people
SYMBOLS
Disposal: The product must not be disposed of along with other domestic
waste. The users must dispose of this equipment by bringing it to a specific
recycling point for electric and electronic equipment.
DISPOSAL OF WASTE BATTERIES – (Directive 2006/66/EC) This symbol on the
battery or on the packaging indicates that the battery provided with this
product shall not be treated as household waste. By ensuring these batteries
are disposed of correctly, you will help prevent potentially negative
consequences for the environment and human health which could otherwise be
caused by inappropriate waste handling of the battery. The recycling of the
materials will help to conserve natural resources. At the end of their life
hand the batteries over to the applicable collection points for the recycling
of waste batteries. For more detailed information about recycling this product
or battery, please contact your local Civic Office, your household waste
disposal service, or the shop where you purchased the product.