ResMed AirFit N30i and AirFit P30i Masks User Guide
- June 6, 2024
- ResMed
Table of Contents
- Part List
- Intended use
- Clinical benefits
- Intended patient population/medical conditions
- Before using your mask
- Fitting your mask
- Adjusting your mask
- Sizing the mask frame
- Disassembling your mask for cleaning
- Cleaning your mask
- Reassembling your mask
- Technical specifications
- Storage **
- Disposal
- Reprocessing the mask between patients
- Symbols
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
ResMed AirFit N30i and AirFit P30i Masks
Part List
- A Elbow
- B Elbow ring
- C Headgear
- D Frame
- E P30i pillows cushion
- F N30i nasal cushion
- Vent
- QuietAir vent
- Standard vent
- Sleeve
- Side buttons
Note: These masks may not be available in all countries.
Intended use
The AirFit N30i and AirFit P30i masks are is intended to be used by patients weighing more than 30 kg who have been prescribed noninvasive positive airway pressure (PAP) therapy such as CPAP or bi-level therapy. The masks are intended for single patient re-use in the home and multi-patient re-use in the hospital/institutional environment. Please read the entire guide before use.
Clinical benefits
The clinical benefit of vented masks is the provision of effective therapy delivery from a therapy device to the patient.
Intended patient population/medical conditions
Obstructive pulmonary diseases (eg, Chronic Obstructive Pulmonary Disease), restrictive pulmonary diseases (eg, diseases of the lung parenchyma, diseases of the chest wall, neuromuscular diseases), central respiratory regulation diseases, obstructive sleep apnoea (OSA) and obesity hypoventilation syndrome (OHS).
Before using your mask
Remove all packaging and inspect each mask component for visible
deterioration.
Use the fitting template to assist in selecting the right size cushion and
frame.
Fitting your mask
-
Pull the frame and headgear over your head until the elbow is positioned on top and the headgear is behind your head.
-
Place the cushion under your nose. For the AirFit P30i mask, align the pillows into your nostrils.
-
Tighten or loosen the headgear until the cushion fits comfortably under your nose.
-
Squeeze the side buttons on the elbow and detach from the frame. Connect the air tubing from your device to the elbow, then reattach to the frame.
Adjusting your mask
- When using the mask, air will flow out of the vents in the cushion and elbow. If air leaks from the side or top of the cushion, adjust the mask fitting to improve the seal.
- You can position the top of the frame forward or backwards over your head until the mask feels comfortable.
Sizing the mask frame
If the mask frame falls back over your head or sits too close to your ears, try a small frame size. If the mask falls forward on your head or sits too close to your eyes, try a standard frame size.
Disassembling your mask for cleaning
If your mask is connected to a device, disconnect the device air tubing from the mask elbow.
-
Pull the cushion from the frame.
-
Undo the fastening tabs on the headgear strap and pull it from the frame.
-
Squeeze the side buttons on the elbow and detach from the frame.
Note: The QuietAir or standard vent can be removed to allow inspection of the vent holes.
Cleaning your mask
After each use: Cushion
Weekly: Headgear, frame and elbow
- Soak the components in warm water with a mild liquid detergent. Ensure that there are no air bubbles while soaking.
- Hand wash the components with a soft bristle brush. Pay particular attention to the vent holes.
- Thoroughly rinse the components under running water.
- For the frame sleeves, squeeze with a clean towel to remove excess water. Shake the QuietAir or standard vent to remove any excess water. Leave the components to air dry out of direct sunlight.
If the mask components are not visibly clean, repeat the cleaning steps
Reassembling your mask
- Connect the elbow into the top of the frame until it clicks.
- Insert the headgear strap into the frame and fold the fastening tabs backwards.
- Insert the ends of the frame into the cushion, ensuring that the arrows on the cushion and frame are aligned.
Note: If the elbow ring detaches, re-insert into the top of the frame.
Removing/inserting the QuietAir/standard vent
-
To remove the vent, gently pull the cushion away and slide the vent out.
-
To insert the vent, align the tab (a) on the cushion with the notch (b) on the vent, and then gently pull the cushion over the vent.
WARNING
- The mask is not suitable for patients requiring life support ventilation.
- To avoid affecting the safety and quality of therapy, the mask vents must be kept clear to protect against rebreathing. Regularly inspect the vents to ensure they are kept clean, undamaged and clear of blockages.
- The mask must be used under qualified supervision for patients who are unable to remove the mask by themselves. The mask may not be suitable for those prone to aspiration.
- Regularly clean your mask to maintain the quality of your mask and to prevent the growth of germs that can adversely affect your health.
- Visual criteria for product inspection: If there is any visible deterioration of a mask component (cracking, crazing, tears etc.), the component should be discarded and replaced.
- Only use compatible CPAP or bi-level therapy devices. The technical specifications of the mask are provided for healthcare professionals to determine compatible devices. Use with incompatible devices may affect the performance or safety of the mask.
- Discontinue using the mask if you have ANY adverse reaction and consult your physician or sleep therapist.
- The mask is not intended to be used simultaneously with nebulizer medications that are in the air path of the mask/tube.
- Follow all precautions when using supplemental oxygen.
- Oxygen flow must be turned off when the CPAP or bilevel device is not operating, so that unused oxygen does not accumulate within the device enclosure and create a risk of fire.
- Oxygen supports combustion. Oxygen must not be used while smoking or in the presence of an open flame. Only use oxygen in well ventilated rooms.
- At a fixed rate of supplemental oxygen flow, the inhaled oxygen concentration varies, depending on the pressure settings, patient breathing pattern, mask, point of application and leak rate. This warning applies to most types of CPAP or bilevel devices.
CAUTION
- Remove all packaging before using the mask.
- Using a mask may cause tooth, gum or jaw soreness or aggravate an existing dental condition. If symptoms occur, consult your physician or dentist.
- When fitting the mask, do not overtighten the headgear as this may lead to skin redness or sores around the mask cushion.
- The mask should not be used unless the device is turned on. Once the mask is fitted, ensure the device is blowing air to reduce risk of rebreathing exhaled air.
- As with all masks, some rebreathing may occur at low pressures.
- Always follow cleaning instructions and only use a mild liquid detergent. Some cleaning products may damage the mask, its parts and their function, or leave harmful residual vapours. Do not use a dish washer or washing machine to clean the mask.
- Do not iron the headgear as the material is heat sensitive and will be damaged.
Note: For any serious incidents that occur in relation to this product, these should be reported to ResMed and the competent authority in your country
Technical specifications
Compatible devices
For a full list of compatible devices for this mask, see the Mask/Device
Compatibility List at
www.ResMed.com/downloads/masks.
SmartStop may not operate effectively when using this mask with some CPAP or
bi-level devices.
Pressure-flow curve:
Mask setting options:
For AirSense, AirCurve or S9| Pillows| Pillows
Therapy Pressure:| 4 to 25 cm H2O| 4 to 25 cm H2O
Resistance: Drop in pressure measured (nominal)
at 50 L/min| 0.5 cm H2O| 0.4 cm H2O
at 100 L/min| 1.7 cm H2O| 1.5 cm H2O
Sound: Declared dual-number noise emission values in accordance with ISO4871:1996 and ISO3744:2010. A-weighted sound power level and A-weighted sound pressure level at a distance of 1 m, with uncertainty of 3 dBA, are shown
Sound power level (Quiet Air vent) | 21 dBA | 21 dBA |
---|---|---|
Sound pressure level (Quiet Air vent) | 14 dBA | 14 dBA |
Sound power level (Standard vent) | 25 dBA | 24 dBA |
Sound pressure level (Standard vent) | 18 dBA | 17 dBA |
Environmental conditions:
Operating temperature: 5°C to 40°C Operating humidity: 15% to 95% non-
condensing Storage and transport temperature: -20°C to +60°C Storage and
transport humidity: up to 95% non-condensing
Service lfe: The service life of the mask system is dependent on the
intensity of usage, maintenance, and environmental conditions to which the
mask is used or stored. As this mask system and its components are modular in
nature, it is recommended that the user maintain and inspect it on a regular
basis, and replace the mask system or any components if deemed necessary or
according to the ‘Visual criteria for product inspection’ in the ‘Warning’
section of this guide.
Note: The manufacturer reserves the right to change these specifications
without notice
Storage **
**
Ensure that the mask is thoroughly clean and dry before storing it for any length of time. Store the mask in a dry place out of direct sunlight.
Disposal
This mask and packaging do not contain any hazardous substances and may be disposed of with your normal household refuse.
Reprocessing the mask between patients
This mask should be reprocessed when used between patients. Cleaning, disinfection and sterilisation instructions are available on ResMed.com/downloads/masks. If you do not have internet access, please contact your ResMed representative.
Symbols
The following symbols may appear on your product or packaging:
| Nasal cradle mask
---|---
| Quiet Air vent
| Cushion size – small
| Cushion size – large
| Cushion size – small wide
| Small frame
| Medical device
| Caution, consult accompanying documents. See symbols glossary at
www.ResMed.com/symbols
| Nasal pillows mask
| Device setting – Pillows
| Cushion size – medium
| Cushion size – wide
| Standard frame
| Importer
LATEX ?
| Not made with natural rubber latex
ResMed acknowledges all consumer rights granted under the EU Directive 1999/44/EC and the respective national laws within the EU for products sold within the European Union.Consumer Warranty
ResMed Pty Ltd
1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia
See ResMed.com for other ResMed locations worldwide. AirFit, AirSense, AirCurve and S9 are trademarks and/or registered trademarks of the ResMed family of companies. For patent and other intellectual property information, see www.ResMed.com/ip.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>