ResMed AirCurve 10 Vauto User Guide
- June 6, 2024
- ResMed
Table of Contents
- Indications for use
- About your device
- About the control panel
- Setup
- Starting therapy
- Stopping therapy
- Power save mode
- My Options
- Mask Fit
- More options
- Caring for your device
- Cleaning
- Reassembling
- Therapy data
- Wireless
- SD card
- Traveling
- Traveling by plane
- Troubleshooting
- Device messages
- General warnings and cautions
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
AirCurve 10 Vauto
User guide
Welcome
The AirCurve™ 10 VAuto, AirCurve 10 S, and AirCurve 10 ST are bilevel positive
airway pressure devices.
WARNING
Read this entire guide before using the device.
CAUTION
In the US, Federal law restricts this device to sale by or on the order of a
physician.
Indications for use
AirCurve 10 VAuto
The AirCurve 10 VAuto device is indicated for the treatment of obstructive
sleep apnea (OSA) in patients weighing more than 66 lb (30 kg). It is intended
for home and hospital use. The humidifier is intended for single patient use
in the home environment and re-use in a hospital/institutional environment.
AirCurve 10 S
The AirCurve 10 S device is indicated for the treatment of obstructive sleep
apnea (OSA) in patients weighing more than 66 lb (30 kg). It is intended for
home and hospital use. The humidifier is intended for single patient use in
the home environment and re-use in a hospital/institutional environment.
AirCurve 10 ST
The AirCurve 10 ST device is indicated for the treatment of obstructive sleep
apnea (OSA) in patients weighing more than 66 lb (30 kg). It is intended for
home and hospital use. The humidifier is intended for single patient use in
the home environment and re-use in a hospital/institutional environment.
Contraindications
Positive airway pressure therapy may be contraindicated in some patients with
the following existing conditions:
- severe bullous lung disease
- pneumothorax or pneumomediastinum
- pathologically low blood pressure, particularly if associated with intravascular volume deal
- dehydration
- cerebrospinal fluid leak, recent cranial surgery, or trauma.
Adverse effects
You should report unusual chest pain, severe headache, or increased
breathlessness to your prescribing physician. An acute upper respiratory tract
infection may require temporary discontinuation of treatment.
The following side effects may arise during the course of therapy with the device:
- drying of the nose, mouth, or throat
- nosebleed
- bloating
- ear or sinus discomfort
- eye irritation
- skin rashes.
At a glance
The AirCurve 10 includes the following:
- Device with HumidAir ™ integrated humidifier
- Water tub
- Air tubing
- Power supply unit
- Travel bag
- SD card (not available in all devices).
Contact your care provider for a range of accessories available for use with the device including:
- Air tubing (heated and non-heated): ClimateLineAir ™, SlimLine ™, ClimateLineAir Oxy, Standard
- . Water tub: Standard water tub, cleanable water tub (can be disinfected)
- Side cover for use without the humidifier
- Filter: Hypoallergenic filter, standard filter
- Air10 ™ DC/DC converter
- SD card reader
- Air10 oximeter adapter
- Air10 USB adapter
- Power Station II
- Air10 tubing elbow.
About your device
1 Air outlet
2 Air filter cover
3 Power inlet
4 Serial number and device number| 5 Water tub
6 Screen
7 Adapter cover
8 SD card cover
---|---
About the control panel
| Start/Stop button| Press to start/stop therapy.
Press and hold for three seconds to enter power save
mode.
---|---|---
| Dial| Turn to navigate the menu and press to select an option.
Turn to adjust a selected option and press to save your
change.
| Home button| Press to return to the Home screen.
Different icons may be displayed on the screen at different times including:
| Ramp Time
---|---
| Humidity
| Humidifier warming
| Humidifier cooling
| Wireless signal strength (green)
| Wireless transfer not enabled (gray)
| No wireless connection
| Airplane Mode
Setup
CAUTION
Do not overfill the water tub as water may enter the device and air tubing.
- Place the device on a stable level surface.
- Plug the power connector into the rear of the device. Connect one end of the power cord into the power supply unit and the other end into the power outlet.
- Connect the air tubing firmly to the air outlet located on the rear of the device.
- Open the water tub and fill it with distilled water up to the maximum water level mark. Do not fill the water tub with hot water.
- Close the water tub and insert it into the side of the device.
- Connect the free end of the air tubing firmly onto the assembled mask. See the mask user guide for detailed information. Recommended masks are available on www.resmed.com.
Starting therapy
- Fit your mask.
- Press Start/Stop or breathe normally if SmartStart ™ is enabled. You will know that therapy is on when the Sleep Report screen is displayed. The ressure bar shows the inspiratory and expiratory pressures in green. The green bar will expand and contract as you breathe in and out. The screen will go black automatically after a short period of time. You can press Home or the dial to turn it back on. If power is interrupted during therapy, the device will automatically restart therapy when power is restored. The AirCurve 10 device has a light sensor that adjusts the screen brightness based on the light in the room.
Stopping therapy
- Remove your mask.
- Press Start/Stop or if SmartStart is enabled, therapy will stop automatically after a few seconds. The Sleep Report now gives you a summary of your therapy session.
Usage hours–Indicates the number of hours of therapy you received the last
session.
Mask Seal–Indicates how well your mask sealed:
Good mask seal.
Needs adjusting, see Mask Fit.
Humidifier–Indicates if your humidifier is working properly:
Humidifier working.
Humidifiers might be faulty, contact your care provider.
If set by your care provider, you will also see:
Events per hour–Indicates the number of apneas and hypopneas experienced per
hour.
More Info–Turn the dial to scroll down to view more detailed usage data.
Power save mode
Your AirCurve 10 device records your therapy data. In order to allow it to
transmit the data to your care provider, you should not unplug the device.
however, you can put it into power save mode to save electricity.
To enter power save mode:
- Press and hold Start/Stop for three seconds. The screen goes black. To exit power save mode:
- Press Start/Stop once. The Home screen is displayed.
My Options
Your AirCurve 10 device has been set up for your needs by your care provider, but you may find you want to make small adjustments to make your therapy more comfortable. Highlight My Options and press the dial to see your current settings. From here, you can personalize your options.
Ramp Time
Designed to make the beginning of therapy more comfortable, Ramp Time is the
period during which the pressure increases from a low start pressure of the
prescribed treatment pressure. You can set your Ramp Time to Off or between 5
to 45 minutes.
To adjust Ramp Time:
- In My Options, turn the dial to highlight Ramp Time and then press the dial.
- Turn the dial to adjust the ramp time to your preferred setting and press the dial to save the change.
Humidity Level
The humidifier moistens the air and is designed to make therapy more
comfortable. If you are getting a dry nose or mouth, turn up the humidity. If
ou re getting any moisture in your mask, turn down the humidity. You can set
the Humidity Level to Off or between 1 and 8, where 1 is the lowest humidity
setting and 8 is the highest humidity setting.
To adjust the Humidity Level:
- In My Options, turn the dial to highlight Humidity Level and then press the dial.
- Turn the dial to adjust the humidity level and press the dial to save the change. If you continue to get a dry nose or mouth, or moisture in your mask, consider using ClimateLineAir heated air tubing. ClimateLineAir together with Climate Control delivers more comfortable therapy.
Mask Fit
Mask Fit is designed to help you assess and identify possible air leaks around
your mask.
To check Mask Fit:
- Fit the mask as described in the mask user guide.
- In My Options, turn the dial to highlight Run Mask Fit and then press the dial. The device starts blowing air.
- Adjust the mask, mask cushion, and headgear until you get a good result. To stop Mask Fit, press the dial or Start/Stop. If you are unable to get a good mask seal, talk to your care provider.
More options
There are some more options on your device which you can personalize.
Leak Alert
When Leak Alert is enabled, the device beeps if the mask leaks too much air or
if you remove the mask during therapy.
SmartStart
When SmartStart is enabled, therapy starts automatically when you breathe into
your mask. When you remove your mask, it stops automatically after a few
seconds.
*When enabled by your care provider.
Caring for your device
It is important that you regularly clean your AirCurve 10 device to make sure
you receive optimal therapy. The following sections will help you with
disassembling, cleaning, checking, and reassembling your device.
WARNING
Regularly clean your tubing assembly, water tub, and mask to receive optimal
therapy and to prevent the growth of germs that can adversely affect your
health.
Disassembling
- Hold the water tub at the top and bottom, press it gently and pull it away from the device.
- Open the water tub and discard any remaining water.
- Hold the cuff of the air tubing and gently pull it away from the device.
- Hold both the cuff of the air tubing and the swivel of the mask, then gently pull apart.
Cleaning
You should clean the device weekly as described. Refer to the mask user guide for detailed instructions on cleaning your mask.
- Wash the water tub and air tubing in warm water using only mild detergent. Do not wash in a dishwasher or washing machine.
- Rinse the water tub and air tubing thoroughly and allow to dry out of direct sunlight and/or heat.
- Wipe the exterior of the device with a dry cloth.
C hecking
You should regularly check the water tub, air tubing and air filter for any
damage.
-
Check the water tub:
• Replace it if it is leaking or has become cracked, cloudy or pitted.
• Replace it if the seal is cracked or torn.
• Remove any white powder deposits using a solution of one part household vinegar to 10 parts water. -
Check the air tubing and replace it if there are any holes, tears or cracks.
-
Check the air filter and replace it at least every six months. Replace it more often if there are any holes or blockages by dirt or dust.
To replace the air filter:
- Open the air filter cover and remove the old air filter. The air filter is not washable or reusable.
- Place a new air filter onto the air filter cover and then close it. Make sure the air filter is fitted at all times to prevent water and dust from entering the device
Reassembling
When the water tub and air tubing are dry, you can reassemble the parts.
- Connect the air tubing firmly to the air outlet located on the rear of the device.
- Open the water tub and fill it with distilled room temperature water up to the maximum water level mark.
- Close the water tub and insert it into the side of the device.
- Connect the free end of the air tubing firmly onto the assembled mask.
Therapy data
Your AirCurve 10 device records your therapy data for you and your care provider so they can view and make changes to your therapy if required. The data is recorded and then transferred to your care provider wirelessly, if the wireless network is available, or via an SD card.
Wireless
Your AirCurve 10 device is equipped with cellular communication. This allows
your therapy data to be wirelessly transmitted to your care provider to
enhance the quality of your treatment. It also allows therapy settings to be
updated in a more timely manner or your device software to be upgraded. The
wireless signal strength icon displayed at the top right of your screen
indicates the signal strength.
The data is usually transmitted after therapy has topped. In order to make
sure that your data is transferred, leave your device connected to the power
outlet at all times and make sure that it is not in airplane mode.
Notes:
- Therapy data might not be transmitted if you use it outside of the country or region of purchase.
- Devices with cellular communication might not be available in all regions. Please be aware that within the wireless network, the availability and quality of the network may be affected by terrain, buildings, and weather. Wireless communication depends on network availability. Coverage is not available everywhere and varies by service.
SD card
An alternative way for your therapy data to be transferred to your care
provider is via the SD card. Your care provider may ask you to send the SD
card y mail or to bring it in. When instructed by your care provider, remove
the SD card. Do not remove the SD card from the device when the SD ight is
flashing, because data is being written to the card.
To remove the SD card:
- Open the SD card cover.
- Push in the SD card to release it. Remove the SD card from the device. Place the SD card in the protective folder and send it back to your care provider. For more information on the SD card refer to the SD card protective folder provided with your device.
Note: The SD card should not be used for any other purpose.
Traveling
You can take your AirCurve 10 device with you wherever you go. Just keep the following in mind:
- Use the travel bag provided to prevent damage to the device.
- Empty the water tub and pack it separately in the travel bag.
- Make sure you have the appropriate power cord for the region you are traveling to. For information on purchasing, contact your care provider.
- If you are using an external battery, you should turn off the humidifier in order to maximize the life of your battery. Do this by turning the Humidity Level to Off.
Traveling by plane
Your AirCurve 10 device may be taken on board as carry-on luggage. Medical
devices do not count toward your carry-on luggage limit.
You can use your AirCurve 10 device on a plane as it meets the Federal
Aviation Administration (FAA) requirements. Air travel compliance letters can
be downloaded and printed from www.resmed.com.
When using the device on a plane:
- Make sure the water tub is completely empty and inserted into your device. The device will not work without the water tub inserted.
- Turn on Airplane Mode.
To turn on Airplane Mode:
- In My Options, turn the dial to highlight Airplane Mode and then press the dial.
- Turn the dial to select On and then press the dial to save the change. The Airplane Mode icon is displayed at the top right of the screen.
CAUTION
Do not use the device with water in the water tub on a plane due to the risk
of inhalation of water during turbulence.
Troubleshooting
If you have any problems, have a look at the following troubleshooting topics. If you are not able to fix the problem, contact your care provider or ResMed. Do not try to open the device.
General troubleshooting
Problem/possible cause | Solution |
---|---|
Air is leaking from around my mask Mask may be fitted incorrectly. | Make sure |
your mask is fitted correctly. See your mask user guide for fitting
instructions or use the Mask Fit function to check your mask fit and seal.
I am getting a dry or blocked nose Humidity level may be set too low.| Adjust
the Humidity Level.If you have Climate LineAir heated air tubing, see the
Climate LineAir user guide.
I am getting droplets of water on my nose, in the mask and air tubing
The humidity level may be set too high.| Adjust the Humidity Level.
If you have Climate LineAir heated air tubing, see the Climate LineAir user
guide.
My mouth is very dry and uncomfortable Air may be escaping through your
mouth.| Increase the Humidity Level.
You may need a chin strap to keep your mouth closed or a full face mask.
Air pressure in my mask seems too high (it feels like I am getting too much
air)
Ramp may be turned off.| Wait for air pressure to build up or turn Ramp Time
off.
My screen is black
Backlight on the screen may have turned off. It turns off automatically after
a short period of time.
Power may not be connected.| Press Home or the dial to turn it back on.
Connect the power supply and make sure the plug is fully inserted.
I have stopped therapy, but the device is still blowing air
Device is cooling down.| Device blows a small amount of air in order to avoid
condensation in the air tubing. It will stop automatically after 30 minutes.
My water tub is leaking
Water tub may not be assembled correctly.| Check for damage and reassemble the
water tub correctly.
Water tub may be damaged or cracked.| Contact your care provider for a
replacement.
My therapy data has not been sent to my care provide wireless coverage may be
poor.| Make sure that the device is placed where there is coverage (ie, on
your bedside table, not in a drawer or on the floor).
The Wireless signal strength icon indicates good coverage when all bars are
displayed, and poor coverage when fewer bars are displayed.
The No wireless connection icon is displayed on the top right of the screen.
no wireless network available.| Make sure that the device is placed where
there is coverage (ie, on your bedside table, not in a drawer or on the
floor).
If instructed to do so, send the SD card to your care provider. The SD card
also contains your therapy data.
Device may be in Airplane Mode.
Data transfer is not enabled for your device.| Turn off Airplane Mode, see
Traveling by plane.
Talk to your care provider about your settings.
My screen and buttons are flashing
The software upgrade is in progress.| Software upgrade takes approximately 10
minutes to complete.
Device messages
Device message/possible cause | Solution |
---|
High leak detected, check your water tub, tub seal or side cover
Water tub may not be inserted properly.
Water tub seal may not be inserted properly.| Make sure the water tub is
correctly inserted.
Open the water tub and make sure that the seal is correctly
inserted.
High leak detected, connect your tubing
Air tubing may not be connected properly.
Mask may be fitted incorrectly.| Make sure the air tubing is firmly connected
at both ends.
Make sure your mask is fitted correctly. See your mask user guide for fitting
instructions or use the Mask Fit function to check your mask fit and seal.
Tubing blocked, check your tubing Air tubing may be blocked.| Check the air
tubing and remove any blockages. Press the dial to clear the message and then
press Start/Stop to restart the device.
SD card error, remove your card and press Start to begin therapy
SD card may not be inserted correctly.| Remove and reinsert the SD card.
Read-only card, please remove, unlock and re-insert SD card
SD card switch may be in the lock (read-only) position.| Move the switch on
the SD Card from the lock position to the unlock position and then re-insert
it.
All other error messages, for example, System fault, refer to user guide,
Error 0XX
An unrecoverable error has occurred on the device.| Contact your care
provider. Do not open the device.
Reassembling parts
Some parts of your device are designed to easily come off in order to avoid
damage to the parts or the device. You can easily reassemble them as described
below.
To insert the water tub seal:
- Place the seal into the lid.
- Press down along all edges of the seal until it is firmly in place.
- Insert one side of the lid into the pivot hole of the base.
- Slide the other side down the ridge until it clicks into place.
General warnings and cautions
WARNING
- Make sure that you arrange the air tubing so that it will not twist around the head or neck.
- Make sure the power cord and plug are in good condition and the equipment is not damaged.
- Keep the power cord away from hot surfaces.
- If you notice any unexplained changes in the performance of the device, if it is making unusual sounds, if the device or the power supply is dropped or mishandled, or if the enclosure is broken, discontinue use and contact your care provider or your ResMed Service Center.
- Do not open or modify the device. There are no user-serviceable parts inside. Repairs and servicing should only be performed by an authorized ResMed service agent.
- Beware of electrocution. Do not immerse the device, power supply, or power cord in water. if liquids are spilled into or onto the device, unplug the device and let the parts dry. Always unplug the device before cleaning and make sure that all parts are dry before plugging it back in.
- Supplemental oxygen must not be used while smoking or in the presence of an open flame.
- Always make sure that the device is turned on and airflow generated before the oxygen supply is turned on. Always turn the oxygen supply off before the device is turned off so that unused oxygen does not accumulate within the device enclosure and creates a risk of fire.
- Do not perform any maintenance tasks while the device is in operation.
- The device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the device should be observed o verify normal operation in the configuration in which it will be used.
- The use of accessories other than those specified for the device is not recommended. They may result in increased emissions or decreased immunity of the device
- Regularly check the bacterial/viral filter for signs of moisture or other contaminants, particularly during nebulization or humidification. Failure too so could result in increased breathing system resistance.
CAUTION
- Use only ResMed parts and accessories with the device. Non-ResMed parts may reduce the effectiveness of the treatment and/or damage the device.
- Use only vented masks recommended by ResMed or by the prescribing doctor with this device. Fitting the mask without the device blowing air can result in rebreathing of exhaled air. Make sure that the mask vent holes are kept clear and unblocked to maintain the flow of the fresh air into the mask.
- Be careful not to place the device where it can be bumped or where someone is likely to trip over the power cor
- Blocking the air tubing and/or air inlet of the device while in operation could lead to overheating of the device.
- Keep the area around the device dry, clean, and clear of anything (eg, clothes or bedding) that could block the air inlet or cover the power supply nit.
- Do not place the device on its side as water might get into the device.
- Incorrect system setup may result in incorrect mask pressure reading. Ensure the system is correctly set up.
- Do not use bleach, chlorine, alcohol, or aromatic-based solutions, moisturizing or antibacterial soaps or scented oils to clean the device, the water tub or air tubing. These solutions may cause damage or affect the humidifier performance and reduce the life of the products. Exposure to smoke, including cigarette, cigar, or pipe smoke, as well as ozone or other gases, may damage the device. Damage caused by any of the foregoing will not be covered by ResMed’s limited warranty.
- If you use the humidifier, always place the device on a level surface lower than your head to prevent the mask and air tubing from filling with water.
- Leave the water tub to cool for ten minutes before handling to allow the water to cool and to make sure that the water tub is not too hot to touch.
- Make sure that the water tub is empty before transporting the device.
Technical specification
90W power supply unit
AC input range:100–240V, 50–60Hz 1.0–1.5A, Class II 115V, 400Hz 1.5A, Class II
(nominal for aircraft use)
DC output:24V 3.75A
Typical power consumption:53W (57VA)
Peak power consumption:104W (108VA)
Environmental conditions Operating temperature:+41°F to +95°F (+5°C to +35°C)
Note: The airflow for breathing produced by this therapy device can be
higher than the temperature of the room. Under extreme ambient emperature
conditions (104°F/40°C) the device remains safe.
Operating humidity: 10 to 95% relative humidity, non-condensing
Operating altitude: Sea level to 8,500′ (2,591 m); air pressure range 1013 hPa
to 738 hPa
Storage and transport temperature: -4°F to +140°F (-20°C to +60°C)
Storage and transport humidity: 5 to 95% relative humidity, non-condensing
Electromagnetic compatibility
The AirCurve 10 complies with all applicable electromagnetic compatibility
requirements (EMC) according to IEC 60601-12:2014, for residential,
commercial, and light industry environments. It is recommended that mobile
communication devices are kept at least 1 m away from the device. Information
regarding the electromagnetic emissions and immunity of this ResMed device can
be found on
www.resmed.com/downloads/devices
Classification: IEC 60601-1:2005+/A1:2012
Class II (double insulation), Type BF, Ingress protection IP22.
Sensors Pressure sensor: Internally located at device outlet, analog gauge
pressure type, 0 to 40 cm H2O (0 to 40 hPa)
Flow sensor: Internally located at device inlet, digital mass flow type, -70
to +180 L/min to +180 L/min
Maximum single fault steady pressure
The device will shut down in the presence of a single fault if the steady-
state pressure exceeds: 30 cm H2O (30 hPa) for more than 6 sec or 40 cm H2O
(40 hPa) for more than 1 sec.
Sound
Pressure level measured according to ISO 80601-2-70:2015 (CPAP mode):
SlimLine: 25 dBA with uncertainty of 2 dBA
Standard: 25 dBA with uncertainty of 2 dBA
SlimLine or Standard and humidification: 27 dBA with uncertainty of 2 dBA
Power level measured according to ISO 80601-2-70:2015 (CPAP mode):
SlimLine: 33 dBA with uncertainty of 2 dBA
Standard: 33 dBA with uncertainty of 2 dBA
SlimLine or Standard and humidification: 35 dBA with uncertainty of 2 dBA
Declared dual-number noise emission values in accordance with ISO 4871:1996.
Physical – device and water tub
Dimensions (H x W x D): 4.57″ x 10.04″ x 5.91″ (116 mm x 255 mm x 150 mm)
Air outlet (complies with ISO 5356-1:2004): 22 mm
Weight (device and standard water tub): 44 oz (1248 g)
Weight (device and cleanable water tub): 44 oz (1248 g)
Housing construction: Flame retardant engineering thermoplastic
Water capacity: To maximum fill line 380 mL
Standard water tub – material: Injection molded plastic, stainless steel, and
silicone seal
Cleanable water tub – material: Injection molded plastic, stainless steel, and
silicone seal
Temperature
Maximum heater plate: 154°F (68°C
Cut-out: 165°F (74°C)
Maximum gas temperature: ≤ 106°F (≤ 41°C)
Air filter
Standard: Material: Polyester nonwoven fiber Average arrestance: >75% for
~7-micron dust
Hypoallergenic: Material: Acrylic and polypropylene fibers in a polypropylene
carrier Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron
dust
Aircraft use
ResMed confirms that the device meets the Federal Aviation Administration
(FAA) requirements (RTCA/DO-160, section 21, category M) for all phases of air
travel.
Wireless module
Bands may not be available in all regions.
The technology used: CDMA (the USA only) 4G 3G 2G
Frequencies: CDMA (800/1900 MHz) 4G (700/850/1700/1900 MHz) 3G
(850/900/1700/1900/2100 MHz) 2G (850/900/1800/1900 MHz)
Max RF power output: CDMA 24.5 dBm 4G 23.0 dBm 3G 24.0 dBm 2G 33.0 dBm
FCC ID: 2ACHL-AIR104G, 2ACHL-AIR103G IC: 9103A-AIR104G, 9103A-AIR103G The AirCurve 10 device complies with FCC Rules and Industry Canada rules. The AirCurve 10 device should be used at a minimum distance of 0.8″ (2 cm) from the body during operation. Additional information regarding the FCC Rules and IC compliance for this device can be found on www.resmed.com/downloads/devices
Operating pressure range S, ST, T: 3 to 25 cm H2O (3 to 25 hPa)
VAuto: 4 to 25 cm H2O (4 to 25 hPa)
CPAP: 4 to 20 cm H2O (4 to 20 hPa)
Supplemental oxygen Maximum flow:
For VAuto device: 4 L/min (CPAP, S, VAuto)
For S device: 4L/min (CPAP, S)
For ST device: 15 L/min (CPAP, S, ST, T)
Pneumatic flow path
- Flow sensor
- Blower
- Pressure se4. Mask
- Mask
- Air tubing
- Water tub
- . Device
- Inlet filter
Design life
Device, power supply unit: 5 years
Cleanable water tub: 2.5 years Standard water tub, air tubing: 6 months
General
The patient is an intended operator.
Humidifier performance
Mask Pressure
cm H2O (hPa)| RH output % at
63°F (17°C) ambienttemperature| RH output % at
72°F (22°C) ambient
temperature| Nominal system output AH1, BTPS2
---|---|---|---
| Setting 4| Setting 8| Setting 4| Setting 8
3
4
10
20
25
|
85
85
85
85
85
|
100
100
100
90
90
|
6
6
6
6
6
|
10
10
10
10
10
Air tubing
Air tubing|
Material
|
Length
| Inner diameter
---|---|---|---
ClimateLineAir
ClimateLineAir Oxy
SlimLine
Standard| Flexible plastic and electrical components
Flexible plastic and electrical components
Flexible plastic
Flexible plastic|
6’6″ (2 m)
6’4″ (1.9 m)
6′ (1.8 m)
6’6″ (2 m)
| 0.6″ (15 mm)
0.75″ (19 mm)
0.6″ (15 mm)
0.75″ (19 mm)
Heated air tubing temperature cut-out: ≤ 106°F (≤ 41°C)
Notes:
- The manufacturer reserves the right to change these specifications without notice.
- The electrical connector end of the heated air tubing is only compatible with the air outlet at the device end and should not be fitted to the mask.
- Do not use electrically conductive or antistatic air tubing.
- The temperature and relative humidity settings displayed are not measured values.
Air tubing resistance to flow and compliance information
Refer to the Air tubing compliance guide in ResMed.com.
Displayed values
Value | Range | Display resolution |
---|
Pressure sensor at air outlet Mask pressure
Flow derived values:
Leak
Tidal volume
Respiratory rate
Minute ventilation
Ti
I:E ratio| 3–25 cm H2O (3–25 hPa)
0–120 L/min
0–4000 mL
0–50 bpm
0–30 L/min
0.1–4.0 sec
1:100–2:1| 0.1 cm H2O (0.1 hPa)
1 L/min
1 mL
1 bpm
0.1 L/min
0.1 sec
0.1
Value
Pressure measurement1:
Mask pressure2
Flow and flow derived values1:
Flow
Leak2
Tidal volume2,3
Respiratory rate2,3
Minute ventilation2,3| Accuracy1
±[0.5 cm H2O (0.5 hPa) + 4% of measured value] ±6 L/min or 10% of reading,
whichever is greater, at 0 to 150 L/min positive flow
±12 L/min or 20% of reading, whichever is greater, 0 to 60 L/min
± 20%
± 1.0 bpm
± 20%
1 Results are expressed as STPD (Standard Temperature and Pressure, Dry).
2 Accuracy may be reduced by the presence of leaks, supplemental oxygen, tidal
volumes <100 mL or minute ventilation <3 L/min.
3 Measurement accuracy verified as per ISO 10651-1:2004 for Home Care
Ventilatory Support Devices (Figure 101 and Table 101) using nominal ResMed
mask vent flows.
Measurement system uncertainties
In accordance with ISO 80601-2-70:2015, the measurement uncertainty of the
manufacturer’s test equipment is:
For measures of flow
For measures of volume (< 100 mL)
For measures of volume (≥ 100 mL)
For measures of static pressure
For measures of dynamic pressure
For measures of time| ± 1.5 L/min or ± 2.7% of reading (whichever is greater)
± 5 mL or 6% of reading (whichever is greater)
± 20 mL or 3% of reading (whichever is greater)
± 0.15 cm H2O (hPa)
± 0.27 cm H2O (hPa)
± 10 ms
---|---
Note: ISO 80601-2-70:2015 stated accuracies and test results provided in this
manual for these items already include the relevant measurement uncertainty
from the table above.
Pressure accuracy
Maximum static pressure variation at 10 cm H2O (10 hPa) according to ISO 80601-2-70:2015
Without humidification
With humidification
| Standard air tubing
± 0.5 cm H2O (± 0.5 hPa)
± 0.5 cm H2O (± 0.5 hPa)
| SlimLine air tubing
± 0.5 cm H2O (± 0.5 hPa)
± 0.5 cm H2O (± 0.5 hPa)
|
Maximum dynamic pressure variation according to ISO 80601-2-70:2015
Device without humidification and Standard air tubing / Device with
humidification and Standard air tubing
Pressure [cm H2O (hPa)]
4
8
12
16
20
25
| 10 BPM
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.3 / 0.3| 15 BPM
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.4| 20 BPM
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.7 / 0.7
Device without humidification and SlimLine air tubing / Device with
humidification and SlimLine air tubing
Pressure [cm H2O (hPa)] 4
8
12
16
20
25| 10 BPM
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.4 / 0.3| 15 BPM
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.6 / 0.5| 20 BPM
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
Pressure accuracy – bilevel
Maximum dynamic pressure varies according to ISO 80601-2-70:2015.
Device without humidification and Standard air tubing / Device with
humidification and Standard air tubing
Breath rate | Inspiratory pressure (cm H2O [hPa]) (Means, Standard Deviations) |
---|---|
6 | 10 |
10 BPM
15 BPM
20 BPM| -0.09, 0.01 /
-0.22, 0.01
0.02, 0.08 /
-0.22, 0.01
0.17, 0.01 /
-0.23, 0.01| -0.01, 0.07 /
-0.22, 0.01
0.12, 0.01 /
-0.22, 0.01
0.21, 0.01 /
-0.28, 0.01| 0.07, 0.05 /
-0.24, 0.01
0.15, 0.01 /
-0.26, 0.01
0.25, 0.01 /
-0.34, 0.01| -0.03, 0.09 /
-0.29, 0.03
0.15, 0.01 /
-0.31, 0.02
0.21, 0.17 /
-0.38, 0.02| 0.12, 0.01 /
-0.26, 0.02
0.16, 0.12 /
-0.30, 0.02
0.32, 0.02 /
-0.40, 0.03
Breath rate| Expiratory pressure (cm HO [hPa]) (Means, Standard Deviations)
2| 6| 12| 17| 21
10 BPM
15 BPM
20 BPM| -0.14, 0.01 /
-0.27, 0.01
-0.16, 0.01 /
-0.25, 0.01
-0.27, 0.01 /
-0.37, 0.01| -0.16, 0.01 /
-0.29, 0.02
-0.20, 0.01 /
-0.33, 0.02
-0.26, 0.02 /
-0.34, 0.01| -0.11, 0.10 /
-0.34, 0.02
-0.20, 0.05 /
-0.35, 0.01
-0.25, 0.01 /
-0.38, 0.01| -0.16, 0.05 /
-0.33, 0.01
-0.21, 0.05 /
-0.38, 0.02
-0.29, 0.01 /
-0.43, 0.02| -0.17, 0.05 /
-0.33, 0.02
-0.23, 0.08 /
-0.38, 0.02
-0.31, 0.01 /
-0.45, 0.03
Device without humidification and SlimLine air tubing / Device with humidification and SlimLine air tubing
Breath rate | Inspiratory pressure (cm H2O [hPa]) (Means, Standard Deviations) |
---|---|
6 | 10 |
10 BPM
15 BPM
20 BPM| -0.26, 0.01 /
-0.52, 0.01
-0.26, 0.01 /
-0.51, 0.01
-0.25, 0.02 /
-0.52, 0.01| -0.25, 0.02 /
-0.53, 0.02
-0.25, 0.01 /
-0.54, 0.01
-0.29, 0.02 /
-0.58, 0.01| -0.24, 0.02 /
-0.53, 0.01
-0.26, 0.01 /
-0.56, 0.01
-0.34, 0.02 /
-0.62, 0.01| -0.25, 0.02 /
-0.54, 0.02
-0.31, 0.03 /
-0.58, 0.02
-0.36, 0.02 /
-0.67, 0.02| -0.20, 0.02
-0.51, 0.02
-0.30, 0.05 /
-0.60, 0.03
-0.36, 0.03 /
-0.69, 0.02
Breath
rate| Expiratory pressure (cm H2O [hPa]) (Means, Standard Deviations)
2| 6| 12| 17| 21
10 BPM
15 BPM
20 BPM| -0.28, 0.01 /
-0.43, 0.01
-0.24, 0.02 /
-0.37, 0.01
0.05, 0.21 /
-0.38, 0.01| -0.30, 0.03 /
-0.50, 0.01
-0.29, 0.02 /
-0.47, 0.01
-0.31, 0.02 /
-0.50, 0.02| -0.30, 0.01 /
-0.54, 0.01
-0.35, 0.01 /
-0.55, 0.01
-0.37, 0.02 /
-0.57, 0.02| -0.33, 0.01 /
-0.58, 0.01
-0.38, 0.01 /
-0.62, 0.02
-0.43, 0.02 /
-0.65, 0.02| -0.34, 0.01 /
-0.60, 0.02
-0.42, 0.02 /
-0.66, 0.01
-0.48, 0.02 /
-0.68, 0.02
Note: The table above is based on data that covers between 60.1 and 88.8% of
the inspiratory phase and 66.1 and 93.4% of the expiratory phase durations.
These data time slots start immediately after the initial transient
overshoot/undershoot periods and end at the point that flow diminishes to an
equivalent absolute value of its starting point, towards the end of the breath
phases (this corresponds to the % ranges of values given immediately above).
Flow (maximum) at set pressures
The following are measured accordingly to ISO 80601-2-70:2015 at the end of
the specified air tubing:
Pressure cm H2O (hPa)| AirCurve 10 and
Standard L/min| AirCurve 10,
humidification and
Standard L/min| AirCurve 10 and
SlimLine L/min| AirCurve 10,
humidification and
ClimateLineAir
L/min
---|---|---|---|---
4
8
12
16
20
25| 180
168
157
144
131
120| 143
135
136
134
123
115| 162
151
140
128
117
96| 151
142
135
121
109
84
Symbols
The following symbols may appear on the product or packaging.
Read instructions before use. Indicates a warning or caution. Follow
instructions before use. Manufacturer. European Authorized Representative.
Batch code. Catalog number. Serial number.
DN Device number. On / Off. Device weight. Protected against finger-sized
objects and against dripping water when tilted up to 15 degrees from
specified orientation. Direct current. Type BF applied part. Class II
equipment. Humidity limitation. Temperature limitation. Non-ionizing
radiation. China pollution control logo 1. China pollution control logo 2.
Prescription-only (In the US, Federal law restricts these devices to sale by
or on the order of a physician). Maximum water level. Use distilled water
only. Operating altitude. Atmospheric pressure limitation. Complies with RTCA
DO-160 section 21, category M.
Environmental information
This device should be disposed of separately, not as unsorted municipal waste.
To dispose of your device, you should use appropriate collection, reuse and
recycling systems available in your region. The use of these collection, reuse
and recycling systems is designed to reduce pressure on natural resources and
prevent hazardous substances from damaging the environment.
If you need information on these disposal systems, please contact your local
waste administration. The crossed-bin symbol invites you to use these disposal
systems. If you require information on the collection and disposal of your
ResMed device please contact your ResMed office, local distributor or go to
ResMed.com/environment.
Servicing
The AirCurve 10 device is intended to provide safe and reliable operation when
operated in accordance with the instructions provided by ResMed. ResMed
recommends that the AirCurve 10 device be inspected and serviced by an
authorized ResMed Service Centre if there is any sign of wear or concern with
device function. Otherwise, service and inspection of the products generally
should not be required during their design life.
Limited warranty
ResMed Pty Ltd (hereafter ‘ResMed’) warrants that your ResMed product shall be
free from defects
in material and workmanship from the date of purchase for the period specified
below.
Product | Warranty period |
---|
• Mask systems (including mask frame, cushion, headgear and tubing)—excluding
single-use devices
• Accessories—excluding single-use devices
• Flex-type finger pulse sensors
• Humidifier water tubs| 90 days
• Batteries for use in ResMed internal and external battery systems| 6 months
• Clip-type finger pulse sensors
• CPAP and bilevel device data modules
• Oximeters and CPAP and bilevel device oximeter adapters
• Humidifiers and humidifier cleanable water tubs
• Titration control devices| 1 year
• CPAP, bilevel and ventilation devices (including external power supply
units)
• Battery accessories
• Portable diagnostic/screening devices| 2 years
This warranty is only available to the initial consumer. It is not
transferable.
During the warranty period, if the product fails under conditions of normal
use, ResMed will repair or replace, at its option, the defective product or
any of its components.
This limited warranty does not cover: a) any damage caused as a result of
improper use, abuse, modification or alteration of the product; b) repairs
carried out by any service organization that has not been expressly authorized
by ResMed to perform such repairs; c) any damage or contamination due to
cigarette, pipe, cigar or other smoke and d) any damage caused by exposure to
ozone, activated oxygen or other gases.
Warranty is void on products sold, or resold, outside the region of original
purchase.
Warranty claims on defective products must be made by the initial consumer at
the point of purchase.
This warranty replaces all other expressed or implied warranties, including
any implied warranty of merchantability or fitness for a particular purpose.
Some regions or states do not allow limitations on how long an implied
warranty lasts, so the above limitation may not apply to you.
ResMed shall not be responsible for any incidental or consequential damages
claimed to have resulted from the sale, installation or use of any ResMed
product. Some regions or states do not allow the exclusion or limitation of
incidental or consequential damages, so the above limitation may not apply to
you.
This warranty gives you specific legal rights, and you may also have other
rights which vary from region to region. For further information on your
warranty rights, contact your local ResMed dealer or ResMed office.
Visit ResMed.com for the latest information on ResMed’s
Limited Warranty.
Further information
If you have any questions or require additional information on how to use the
device, contact your care provider.
ResMed Pty Ltd
1 Elizabeth Macarthur Drive
Bella Vista NSW 2153 Australia
See ResMed.com for other ResMed locations worldwide. AirCurve, Air10,
ClimateLine, HumidAir and SlimLine are trademarks and/or registered trademarks
of the ResMed family of companies. For patent and other intellectual property
information, see ResMed.com/ip. SD Logo is a trademark of SD-3C, LLC. © 2021
ResMed. 378962/2 2021-04
References
- Sleep apnea and COPD - learn about symptoms and treatment | ResMed
- Environmental policy - ResMed
- Legal disclaimers and intellectual property - ResMed
- Sleep apnea and COPD - learn about symptoms and treatment | ResMed
- Sleep apnea and COPD - learn about symptoms and treatment | ResMed
- resmed.com/downloads/devices
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>