IMPORTANT – BEFORE FIELD DEPLOYMENT: Completely configure new X2 systems with sensors and a web connec-tion in a nearby work area. Operate the system for several hours and ensure correct sensor readings. Use this test run to become familiar with the features and functions.

Overview

The X2 with integral Wi-Fi includes an embedded Wi-Fi hotspot for smartphone, tablet, or PC connection. Three sensor ports provide industry standard protocols includ-ing SDI-12, RS-232, and RS-485. The center port offers direct communication (serial to PC) and power input.
Data is accessed and stored on the WQData LIVE web datacenter. An easy-to-use dashboard and built-in sensor library automatically facilitate setup and configuration.

What’s Included?

(1) X2 data logger
(1) X2 grounding kit
(3) Sensor port plugs, spare orings
(1) Power port plug, spare oring
(1) Oring grease
(1) Screwdriver with magnet tip
(1) Quick start guide

Wi-Fi Specifications

  • Note: The X2 embedded web does not support 5GHz net-works. Connect the X2 Wi-Fi to a 2.4GHz network.
  • Wi-Fi Communications: 802.11b/g/n (Direct to X2 or Con-nect X2 to an existing network)
  • Wi-Fi Range: 250 ft. maximum1
  • 1Range varies based on many factors including obstructions, other wireless signals in the area, elevation change, and more.
  • X2 Embedded Wi-Fi user guide link: nexsens.com/x2eweb
  1. To get started:

    • Go to WQDataLIVE.com
    • Create a new account or sign into an existing account.
    • Choose or create the project that will contain the data logger by selecting the Projects link from the bottom right footer of the page.
    • Open the project and go to the ADMIN tab located at the top of the project dashboard and click Settings.
  2. From there, choose the Project/Site pull down menu and select the site for the new data logger.

    • If a site has not been created, select New Site.
  3. Enter the claim code listed below into the space provided under Assigned Devices.

  4. Click Add Device.

    • The new device should be visible in the Assigned Devices list.
  5. Remove the center power plug and required outer port plugs (1 for each sensor to connect).

  6. If ports are labelled, match and connect the sensors with the proper outer ports (8 pin).

    • If the X2 and sensors were integrated at NexSens, the sensors must be connected to the same port for proper communication.
    • If connecting a new sensor, use the included USB cable and the CONNECT software to ensure the sensor script is enabled. The article below describes this process.
    • nexsens.com/conncss
  7. Connect power (5 to 24VDC) to the center port (6 pin).

    • Place the X2 Wi-Fi in an area with adequate Wi-Fi connection to the configured network.
  8. Using a PC or mobile device, connect to the X2’s Wi-Fi network.

    • If the X2 network is not visible, disconnect and re-apply power to the logger and refresh the PC or mobile device’s network list.
  9. Navigate to 192.168.1.1 with a web browser to connect with the X2 embedded web.

    • Access the Main Menu (Three horizontal bars at the top left of the interface)
    • Go to Settings | Network
  10. Toggle the Wi-Fi mode to Network and in the spaces provided enter:

    • Local network SSID (case sensitive)
    • Network security settings
    • Network password
      NEXSENS X2 Environmental Data Logger fig
  11. Click Save Settings.

    • Reset the Wi-Fi using the magnet swipe.
      NEXSENS X2 Environmental Data Logger fig1
  12. Disconnect from the logger network and allow up to 15 minutes for sensor detection and data acquisition.

    • At this time the updated X2 Wi-Fi profile issent to WQData LIVE.
    • At the next 10-minute transmit interval,refresh WQData LIVE and confirm all sensorparameters are shown and valid sensorreadings appear.

LED Light Indicators

LED Light Indicators

LED

Color

|

1 Second Interval

|

3 Second Interval

|

5 Second Interval

| 3 Blinks Per

5 Seconds

|

Solid

---|---|---|---|---|---

Green

| Primary power heartbeat| Secondary power heartbeat| Backup power heartbeat|

N/A

|

N/A

Blue

|

N/A

|

N/A

|

Weak/No cell signal

| Strong radio signal| WQData LIVE setup successful1
Yellow| N/A| N/A| Data acquisition in progress| N/A| Sensor calibration in progress

Red

| Significant system error2|

N/A

|

N/A

|

N/A

|

N/A

  1. WQData LIVE setup is automatically done after sensor detection.
  2. System errors can result from internal sensors detecting high current, humidity, or pressure. These errors can be indicative of a leak or other major problem with the system that requires attention.

For additional information, please reference the X2 Resource Library on the NexSens Knowledge Base. nexsens.com/x2kb

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals