SEALEY MM104 Professional Auto-Ranging Digital User Manual
- June 6, 2024
- SEALEY
Table of Contents
instructions for Professional auto-ranging digital multimeter
model no: mm104
Thank you for purchasing a Sealey product. manufactured to a high standard, this product will if used according to these instructions, and properly maintained, give you years of trouble-free performance.
IMPORTANT : PLEASE READ THESE INSTRUCTIONS CAREFULLY. NOTE THE SAFE
OPERATIONAL REQUIREMENTS, WARNINGS & CAUTIONS. USE
THE PRODUCT IS CORRECTLY AND WITH CARE FOR THE PURPOSE FOR WHICH IT IS
INTENDED. FAILURE TO DO SO MAY CAUSE DAMAGE AND/OR PERSONAL INJURY AND WILL
INVALIDATE THE WARRANTY. KEEP THESE INSTRUCTIONS SAFE FOR FUTURE USE.
| refer to instructions
---|---
| electrical shock hazard
SAFTY
Personal PreCautions
When using this meter, please observe all normal safety rules concerning:
Protection against the dangers of electric current. Protection of the meter
against misuse. full compliance with safety standards can only be guaranteed
if used with the test leads supplied. if necessary, they must be replaced with
genuine Sealey leads with the same electronic ratings. failure to do so will
invalidate the warranty.
Do not use leads if damaged or if the wire is bared in any way.
General safety instructions
familiarise yourself with the applications, limitations, and hazards of the
meter. if in any doubt consult a qualified electrician. 9 When the meter is
linked to a measurement circuit, do not touch unused meter terminals. When
the scale of the value to be measured is unknown set the selector to the
highest range available. 9 Before rotating the range selector to change
functions, disconnect test leads from the circuit under test. Warning! never
perform resistance measurements on live circuits. 9 Always be careful when
working with voltages above 60Vdc or 30Vac rms. Keep your fingers behind the
probe guards while measuring. 9 When not in use, store the meter carefully in
a safe, dry, childproof location. storage temperature range -10°c to 50°c 8 do
not apply a voltage or current to the meter that exceeds the specified
maximum. 9 the user shall ensure that test probes are correctly selected in
order to prevent danger. Probes shall be selected to ensure that adequate
barriers guard against inadvertent hand contact with live conductors under
test and that probes have minimally exposed probe tips. Where there is a risk
of the probe tip short-circuiting with other live conductors under test, it is
recommended that the exposed tip length shall not exceed 4mm.
Per ieC1010 overvoltage installation Category
Overvoltage category i
Equipment of overvoltage category i is equipment for connection to circuits in
which measures are taken to limit the transient overvoltages to an
appropriately low level.
Note – examples include protected electronic circuits.
Overvoltage category ii
Equipment of overvoltage category II is energy-consuming equipment to be
supplied from the fixed installation. note – examples include household,
office, and laboratory appliances.
Overvoltage category iii
Equipment of overvoltage category III is equipment in fixed installations.
Note – examples include switches in the fixed installation and some
equipment for industrial use with a permanent connection to fixed
installation.
Overvoltage category iV
Equipment of overvoltage category iV is for use at the origin of the
installation.
Note – examples include electricity meters and primary overcurrent
protection equipment
INTRODUCTION
Lightweight, compact digital multimeter with double molded plastic housing and a 4000 counts lcd display. drop-proof up to 2mtr. iP65 is rated as “dust tight” and protected against low-pressure water jets (all directions). features display backlight and forward-facing work light. conforms to en 61010-1 and en 61010-2-033ncAtiii 600V safety requirements for electrical equipment for measurement, control, and laboratory use. measures: Ac and dc Voltage, Ac and dc current, resistance, frequency, duty cycle, diode test, continuity test, and capacitance.
SPECIFICATION
model no ………………………………………………………….. mm104
FUNCTION
|
RANGE
|
ACCURACY
---|---|---
DC Voltage
|
400mV
|
±(1.0%reading+5digits)
4V
|
t (1.0%reading+3digits)
40V
400V
600V
AC Voltage
| |
50Hz to 60Hz
4V
|
I (1.2%reading+3digits)
40V
400V
600V
AI AC voltage ranges are specified from 5% of range to 100% of range
DC current
|
400pA
| ± (1.0%reading+3digits)
4000pA
40mA
400mA
4A
| ± (1.29fseading+3digits)
10A
(20A: 30 sec max with reduced accuracy)
AC Current
| |
50Hz to 60Hz
400pA
|
t (1.2%seading+3digits)
4000pA
40mA
400mA
4A
|
I (1.8%reading+5dIgits)
10A
(20A: 30 sec max with reduced accuracy)
Resistance
|
4000
|
t (1.5%reading+5digits)
41<0
40K0
4001<0
4M0
40M0
|
t (2.5%reading+20digits)
Capacitance
|
40nF
|
± (4.5%reading+10digits)
400nF
4pF
|
± (3.0%reading+5digits)
40pF
400pF
4000pF
|
I (5%reading+5digits)
Frequency
(electrical)
|
10KHz
|
3 (1.5%)
Duty Cycle
|
99.9%
|
3 (1.2%)
Note: Accuracy is stated at 18°c to 28°c and less than 75% RH
Note: Accuracy specifications consist of two elements:
a) (% reading) – this is the accuracy of the measurement circuit.
b) (+ digits) – this is the accuracy of the analog to digital converter.
Enclosure…………………………………………………..double molded
Shock (drop test)………………………………………. 6.5 feet (2 meters)
Transistor test……………………………………………..no
Diode test………………………………………………….Yes
Continuity Check………………………………………… An audible signal will sound if the
resistance is less than 50Ω (approx.), test current <0.35mA
Temperature sensor…………………………………….requires type K thermocouple
Input Impedance………………………………………….>10MΩ VDC & >10MΩ VAC
Ac response……………………………………………..terms
ACV Bandwidth……………………………………….50/60HZ(All WAVe) 45Hz to 1000Hz(sine WAVE)
Crest Factor…………………………………………….≤3 at full scale up to 300V, decreasing linearly
to ≤1.5 at 600V
Digits x Height………………………………………….4x14mm
Over range indication…………………………………“ol” is displayed
Auto Power-off…………………………………………15 minutes (approximately) with disabling the
feature
Polarity…………………………………………………….Automatic (no indication for positive); minus (-)
sign for negative
Measurement rate……………………………………3 times per second, nominal
low Battery indication……………………………….“ is displayed if battery voltage drops
below operating voltage
Battery……………………………………………………..1.5V AAA x2 battery supplied
Fuses……………………………………………………….mA, µA ranges; 0.5A 600V ceramic fast blow A range;
10A/600V ceramic fast blow
Operating temperature………………………………41°f to 104°f (5°c to 40°c)
Storage temperature…………………………………-4°f to 140°f (-20°c to 60°c)
Operating Humidity……………………………………max 80% up to 87°f (31°c) decreasing linearly
to 50% at 104°f (40°c)
Storage Humidity………………………………………<80%
Operating Altitude……………………………………..7000ft. (2000meters) maximum.
Weigh………………………………………………………260g
Size………………………………………………………….117x 62 x 45mm
Conformity…………………………………………………en 61010-1, en61010-2-033
1. Battery cover
2. 4,000 counts lcd negativity display
3. rAnGe button
4. mAX/min button
5. mode button| 6. Hold and flashlight button
7. function switch
8. com input jack
9. 10A input jacks
10. Positive, mA, µA input jack
---|---
| Auto power off
---|---
| Continuity
| Diode test
| Battery status
n| nano (10 -9) (capacitance)
µ| micro (10-6) (amps, cap)
m| milli (10-3) (volts, amps)
A| Amps
k| kilo (103) (ohms)
F| Farads (capacitance)
M| mega (106) (ohms)
Ω| Ohms
Hz| Hertz (frequency)
V| Volts
%| Percent (duty ratio)
AC| Alternating current
AUTO| Autoranging
DC| Direct current
HOLD| Display hold
MAX| Maximum
MIN| Minimum
OPeration
Warning! risk of electrocution. High voltage circuits, both Ac and dc are
dangerous and should be measured with great care. Always turn the function
switch to the off position when the meter is not in use.
Warning! if “ol” appears in the display during a measurement, the value
exceeds the range that has been selected. change to a higher range.
DC voltage measurements
Do not measure dc/Ac voltages if a motor on the circuit is being switched
on or off. large voltage surges may occur that can damage the meter
Set the function switch to the Vdc position.
Insert the black test lead into the negative com jack.
Insert the red test lead into the positive V jack.
Touch the black test probe tip to the negative side of the circuit
Touch the red test probe tip to the positive side of the circuit.
read the voltage in the display
AC voltage (frequency, duty cycle) measurements
Warning! risk of electrocution. the probe tips may not be long enough to
contact the live parts inside some 230V outlets for appliances because the
contacts are recessed deep in the outlets. As a result, the reading may show 0
volts when the outlet actually has a voltage on it. make sure the probe tips
are touching the metal contacts inside the outlet before assuming that no
voltage is present.
set the function switch to the VAc/Hz/% position.
Insert the black test lead plug into the negative com jack.
Insert red test lead plugs into the positive V jack.
Touch the black test probe tip to the neutral side of the circuit.
Touch the red test probe tip to the “hot” side of the circuit.
Read the voltage in the display.
Press the HZ/% button to indicate “Hz”.
Read the frequency in the display.
Press the Hz/% button again to indicate “%”.
Read the % of duty cycle in the display.
DC Current measurements
Warning! do not make 20A current measurements for longer than 30 seconds.
exceeding 30 seconds may cause damage to the meter and/or the test leads.
Insert the black test lead plug into the negative com jack.
For current measurements up to 6000µA dc, set the function switch to the µA
position and insert the red test lead plug into the µA/mA jack.
For current measurements up to 600mA dc, set the function switch to the mA
position and insert the red test lead plug into the µA/mA jack.
For current measurements up to 20A dc set the function switch to the 10A/HZ/%
position and insert the red test lead plug into the 10A jack.
Press the mode button to indicate “dc” on the display.
Remove power from the circuit under test, then open up the circuit at the
point where you wish to measure current.
Touch the black test probe tip to the negative side of the circuit.
Touch the red test probe tip to the positive side of the circuit.
Apply power to the circuit.
Read the current in the display.
Resistance measurements
Warning! to avoid electric shock, disconnect power to the unit under test
and discharge all capacitors before taking any
Resistance measurements. remove the batteries and unplug cables.
Set the function switch to the Ω CAP position.
insert the black test lead plug into the negative com jack.
Insert the red test lead plug into the positive Ω jack.
Press the MODE button to indicate “Ω” on the display.
Touch the test probe tips across the circuit or part under test. it is best to
disconnect one side of the part under test so the rest of the circuit will not
interfere with the resistance reading.
Read the resistance in the display.
Continuity Check
Warning! to avoid electrical shock, never measure continuity on circuits
or wires that have a voltage on them.
set the function switch to the Ω CAP position.
Insert the black lead plug into the negative com jack.
Insert the red test lead plug into the positive Ω jack.
Press the MODE button to indicate” “ and “Ω” on the display
Touch the test probe tips to the circuit or wire you wish to check.
If the resistance is less than approximately 50Ω, the audible signal will
sound. If the circuit is open, the display will indicate “OL”.
Capacitance measurements
Warning! to avoid electric shock, disconnect power to the unit under test
and discharge all capacitors before taking any capacitance measurements.
remove the batteries and unplug the cables.
Set the rotary function switch to the Ω CAP position.
Insert the black test lead plug into the negative com jack.
Insert the red test lead plug into the positive V jack.
Press the mode button to indicate “nF” on the display.
Touch the test leads to the capacitor to be tested.
The test may take up to 3 minutes or more for large capacitors to charge. Wait
until the readings settle before ending the test.
read the capacitance value in the display.
Diode Test
Set the function switch to green Ω CAP position.
Insert the black test banana plug into the negative com jack and the red test
lead banana plug into the positive V jack.
Press the mode button to indicate and on the display.
Touch the test probes to the diode under test. forward voltage will typically
indicate 0.400 to 0.700V. reverse voltage will indicate
“OL”. shorted devices will be indicated near of and an open device will
indicate “ol” in both polarities.
Autoranging/manual range selection
When the meter is first turned on, it automatically goes into Autoranging.
This automatically selects the best range for the measurements being made and
is generally the best mode for most measurements. for measurement situations
requiring that a range be manually selected, perform the following:
Press the rAnGe key. the “Auto” display indicator will turn off.
Press the rAnGe key to step through the available ranges until you select the
range you want.
To exit the manual ranging mode and return to Autoranging, press and hold the
rAnGe key for 2 seconds.
Note: manual ranging does not apply for the capacitance and frequency
functions.
MAX/MIN
Note: When using the mAX/min function in Autoranging mode, the meter will
“lock” into the range that is displayed on the lcd when MAX/MIN is activated.
if an mAX/min reading exceeds that range, an “ol” will be displayed. select
the desired range Before entering MAX/MIN mode.
Press the mAX/min key to activate the mAX/min recording mode. the display icon
“mAX” will appear. the meter will display and hold the maximum reading and
will update only when a new “max” occurs.
Press the mAX/min key again and the display icon “min” will appear. the meter
will display and hold the minimum reading and will update only when a new
“min” occurs.
To exit MAX/MIN mode press and hold the mAX/min key for 2 seconds.
Display Backlight
Positive display and Backlight on when the unit is turned on.
Mode
Press mode key the selection of double measured functions which are present at display is possible. in particular, this key is active in V· ·CAP·Ω position to select among resistance test, diode test, continuity test, and capacitance test, and in a current position to select between Ac or dc current measurements.
Hold/Flash Light
The hold function freezes the reading in the display. Press the Hold key
momentarily to activate or exit the Hold function.
Press the Hold key for >1 second to turn on or off the flAsH light function.
the flashlight and will automatically power off after 5 minutes of inactivity.
Auto Power-off
The auto-off feature will turn the meter off after 15 minutes. to disable
the auto power-off feature, hold down the mode button and turn the meter is
on.
Low Battery Indicator
The Icon will appear in the lower-left corner of the display when the
battery voltage becomes low. replace the battery when this appears.
MAINTENANCE
Warning! to avoid electric shock, disconnect the test leads from any
source of voltage before removing the back cover or the battery or fuse
covers.
Warning! to avoid electric shock, do not operate the meter until the
battery and fuse covers are in place and fastened securely.
This multimeter is designed to provide years of dependable service if the
following care instructions are performed:
Keep the meter dry. if it gets wet, wipe it off.
Use and store the meter in normal temperatures. temperature extremes can
shorten the life of the electronic parts and distort or melt plastic parts.
Handle the meter gently and carefully. dropping it can damage the electronic
parts or the case.
Keep the meter clean. Wipe the case occasionally with a damp cloth. do not use
chemicals, cleaning solvents, or detergents.
Use only fresh batteries of the recommended size and type. remove old or weak
batteries so they do not leak and damage the unit.
If the meter is to be stored for a long period of time, the batteries should
be removed to prevent damage to the unit.
Battery installation / removal
Warning! to avoid electric shock, disconnect the test leads from any
source of voltage before removing the battery cover
Turn the power off, and disconnect the test leads from the meter.
Open the rear battery cover by removing the screw using a Phillips head
screwdriver.
Insert the battery into the battery holder, observing the correct polarity.
Put the battery cover back in place. secure with the screw.
Warning! to avoid electric shock do not operate the meter until the
battery cover is in place and fastened securely.
Note: if the meter does not work, check the fuses and batteries to make
sure that they are still good and that they are properly inserted.
Replacing the Fuses
Disconnect the test leads from the meter.
Remove the battery cover.
Gently remove the old fuse and install the new fuse into the holder.
Always use a fuse of the proper size and value (0.5A/600V fast blow for the
600mA range, 10A/600V fast blow for the 20A range.
Replace and secure the rear cover, battery, and battery cover.
Warning! to avoid electric shock do not operate the meter until the fuse
cover is in place and fastened securely.
Parts support is available for this product. to obtain a parts listing and/or
diagram,
please log on to www.sealey.co.uk, email
sales@sealey.co.uk or telephone 01284 757500.
Environmental Protection
Recycle unwanted materials instead of disposing of them as waste. All tools,
accessories, and packaging should be sorted, taken to a recycling center, and
disposed of in a manner that is compatible with the environment. When the
product becomes completely unserviceable and requires disposal, drain off any
fluids (if applicable) into approved containers and dispose of the product and
the fluids according to local regulations.
WEEE Regulations
Dispose of this product at the end of its working life in compliance with the
EU directive on Waste electrical and electronic equipment (WEEE). When the
product is no longer required, it must be disposed of in an environmentally
protective way. contact your local solid waste authority for recycling
information.
Battery Removal
See section 6.2
Under the Waste Batteries and Accumulators regulations 2009, Jack Sealey ltd
is required to inform potential purchasers of products containing batteries
(as defined within these regulations), that they are registered with Valpak’s
registered compliance scheme. Jack Sealey Ltd’s Batteries Producer
Registration Number (BPRN) is BPRN00705.
NOTE: It is our policy to continually improve products and as such we
reserve the right to alter data, specifications, and component parts without
prior notice.
IMPORTANT: no liability is accepted for incorrect use of this product.
WARRANTY: Guarantee is 12 months from purchase date, proof of which will
be required for any claim.
Sole UK distributor, Sealey Group.
Kempson Way, Suffolk Business Park,
Bury st. Edmunds, Suffolk.
iP32 7Ar
01284 757500
01284 703534
www.sealey.co.uk