ANDOR iXon Ultra and Life 888 Imaging Camera User Manual
- June 6, 2024
- ANDOR
Table of Contents
iXon Ultra & Life 888 Quick
Start Guide
Prepare your Facility
| Handle the camera and accessories with care.
---|---
| Observe anti-static precautions when installing the camera.
| Ensure 100 mm clearance around camera and power supply vents (Do not block
vents
when operating on benchtop).
| Operation of the camera close to intense pulsed sources (e.g. plasma and
X-ray
sources) may compromise performance.
| Keep dry, <70% humidity non-condensing.
| Use correct power supply and cables.
| Store between -25°C and 50°C.
Operate between 0°C and 30°C.
| Refer to the cooling guidelines in Section 6 and the camera user manual for
further
information.
Refer to the user manual for further information
Unpack the Camera
Camera
iXon Ultra or Life 888 EMCCD Camera (model as ordered)
Essentials
Cables
Install Software (Solis or SDK)
You must have administrator access on your PC to perform this
installation.
These instructions cover the installation of both Andor Solis and Andor SDK if
ordered. A copy of the software is provided within the camera box.
Additionally, the latest software files are available to download on request,
please sign in or create an account at andor.oxinst.com/downloads
Windows: Switch on the PC, run the setup exe from the supplied copy or the
downloaded file: e.g. AndorSDK2Setup
- Confirm the version of the software (minimum is v2.1x).
- Follow the on-screen prompts.
- Select the installation directory when prompted.
- If prompted for camera type select iXon.
- On the final window click on “Install”. Continue to follow the on-screen prompts.
- To complete the installation, when prompted select “Yes, restart the computer now” and click on the finish button.
Linux:
For detailed instructions on how to install the SDK drivers on Linux please
refer to the Andor SDK2 Manual, available at
andor.oxinst.com/downloads.
Install the Camera Control (PCle) Card & Driver
Ensure your PC is powered off.
Ensure anti-static precautions and use the anti-static strap provided to prevent potential damage. Keep all screws safe.
- Unplug all cables from the rear of the computer. Open the PC enclosure to gain access to the expansion slots.
- Locate a suitable PCIe slot: use an x4 PCIe slot only. If you are unsure which slot is correct, please consult the PC user manual.
- Remove the filler bracket corresponding to the slot you intend to use.
- Remove the USB 3.0 card from its protective ESD packaging and insert it fully into the expansion slot.
- Ensure the card’s mounting bracket is flush with any other mounting or filler brackets to either side of it, then secure the controller card in place.
- Replace the computer cover and secure it with mounting screws if applicable.
- Reconnect any accessories you were using previously.
Install the USB 3.0 PCIe Card Driver manually (not required for Windows
10)
USB 3.0 PCIe card drivers must be installed manually for PCs running versions
prior to Windows 10.
See Startech’s website: www.startech.com/downloads enter product code
PEXUSB3S25 and hit Search. If a later driver is available, download and
install.
Disable Power Management for the “Renesas USB 3.0 Host Controller”:
- Navigate to the ‘Renesas USB 3.0 Host Controller’ properties menu in Device Manager (Control Panel>Device Manager> Universal Serial Bus Controllers). Select the Advanced tab.
- Check the Disable USB 3.0 power management functions checkbox.
Cooling Guidelines
Note: Always ensure that the temperature of the liquid coolant circulated through the camera head is above the dew point of the camera ambient. Use of coolant at or below the dew point will result in permanent damage to the camera head, due to the formation of condensation on internal components.
- The iXon Ultra and Life 888 overclock the CCD-201 sensor which generates more heat from the sensor.
- These models retain the cooling technology and the cooling performance of other iXons at lower readout speeds (10 MHz and below ). At higher readout speeds it is necessary to set a higher cooling temperature: -55°C is acceptable in most modes – refer to the user manual for model details.
- At higher readout speeds (and shorter exposures) dark current is less important, so this higher CCD operating temperature is expected to be an acceptable compromise.
If an unusual crop mode or operating regime is required please find the camera’s temperature settling point before starting the measurement.
- Set max cooling -110°C.
- Run the camera in the intended mode for 20 minutes.
- Note cooling temperature achieved.
- Set a temperature 2-5°C warmer as the set point.
- Run required measurement.
If the temperature control circuit unlocks during a measurement it will impact measurement accuracy, in particular EM gain calibration.
Connect the Camera and Switch it on
Ensure that the Camera ON/OFF switch is in the OFF position AND your PC is switched OFF.
- Attach the camera to the lens or optical system using the camera C-mount interface as required.
- Insert the 12 V DC power cable into the power connector on the bottom plate of the camera, ensure the orientation is correct. NEVER forcibly insert the connector.
- Connect the supplied USB cable between the USB connection on the camera and the corresponding slot on the PC.
- If required, connect the multi I/O cable to the External I/O 15 way D type connector.
- Switch the camera ON using the ON/OFF switch. You should hear an audible confirmation (camera start-up tone).
- Turn ON the PC.
- Start your software.
- View and adjust the acquisition settings as required.
- Start acquiring data.
iXon Ultra & Life 888 Connections
- Ultra models only
Note: Only use the USB 3.0 cable supplied by Andor to ensure performance.
Troubleshoot
The camera start-up tone does not sound on start-up
- The camera start-up tone should be audible momentarily when the camera is switched to ON.
- If this does not occur, ensure that power is connected to the camera and the ON/OFF switch is set to ON.
Support Device not recognized when plugged into PC
- Choose another USB port.
Temperature Trip Alarm sounds (continuous tone)
Should the buzzer sound ensure the following:
Air Cooling
- Check that the air vents on the camera are not blocked and there is sufficient clearance (100 mm) around the camera.
- The ambient air temperature is not above 30ºC.
- The fan has not been deactivated (or the speed set too low) in software.
Please see the Hardware Guide for further information on Troubleshooting.