FOREWORD

The Non-Contact Infrared Body Thermometer operating Instructions intend to provide the necessary information for proper operation of the thermometer model. Only body mode was reviewed and certified by notified body. General knowledge of Infrared thermometer and an understanding of the features and functions of the thermometer model are prerequisites for proper use. The Non- Contact Infrared Body Thermometer is a medical device, and can be used repeatedly for the product’s recommended lifetime of 5 years. Please read the manual first before using it, if not fully understand the usages, please stop using the thermometer.

Do not operate the thermometer without completely reading and understanding these instructions.

Notice
Purchase or possession of this device does not carry any express or implied license to use with replacement parts which would, alone or in combination with this device, fall within the scope of one of the relating patents.

For further information contact:
HoMedics Group Ltd, Somerhill Business Park, Five Oak Green Road,
Tonbridge, Kent TN11 0GP.
Customer Service: service@homedics.co.uk

0598

HOMEDICS Non-Contact Infrared Body Thermometer--
Thermometer
Company Name: Wellkang Ltd
Company Address: 16 Castle St, Dover, CT16 1PW, UK
Tel: +44 (20)30869438, 32876300
Fax: +44(20)76811874
Web: www.CEmark.com, www.CE-marking.com, www.CE-marking.eu,
Email: AuthRep@CE-marking.eu

SAFETY INFORMATION

This device may only be used for the purposes described in these instructions. The manufacturer cannot be held liable for damage caused by incorrect application. The Non-Contact Infrared Body Thermometer is designed to minimize the possibility of hazards from errors in the software program by following sound and light engineering design processes, Risk Analysis and Software Validation.

  WARNING
Warnings are identified by the WARNING symbol shown above.

  • The Non-Contact Infrared Body Thermometer is to be operated by consumers in the home setting and primary care setting as screening tool. This manual, accessories, Directions for Use, all precautionary information, and specifications should be read before use.
  • This product is designed to measure human body temperature on the forehead. Do not use it for any other purpose.
  • This product is intended for use in the home setting and primary care setting as screening tool.
  • Do not use the thermometer if it malfunctions or has been damaged in any matter.
  • When the ambient temperature of the thermometer changes too much, such as moving the Thermometer from one place of lower temperature to another place of higher temperature, allow the thermometer to remain in a room for 30 minutes where the temperature is between 15°C to 40°C (59°F- 104°F).
  • Remove primary batteries if equipment is not likely to be used for long time.
  • This product is not waterproof, do not immerse in water or other liquid; If cleaning and disinfecting, please follow the “Care and Storage” section requirements.
  • Do not touch the sensor of infrared detection with your fingers.
  • Avoid using the thermometer to take a body temperature reading while using a cold compress or other measure to reduce fever.
  • If measuring human forehead temperature, please select “body” mode; for measuring other objects, liquids, food and other temperature please select “surface” mode.
  • This product must be operated in a stable environment. If the ambient environment is unstable, please check whether there is fog on the sensor and remove this before using in accordance with the “Care and Storage” section.
  • Do not near strong electrostatic field or strong magnetic fields, thus avoiding the impact on the accuracy of the measurement data.
  • Do not mix the old and new batteries to avoid damage to the product.
  • It may affect the accuracy of measurements when the forehead is covered by hair, perspiration, cap or scarf.
  • The measuring result of this product is only for your reference. If you have any doubt, please measure the temperature in other methods.

The device should be kept out of the reach of children/pets. When not in use, store the device in a dry room and protect it against extreme moisture, heat, lint, dust and
direct sunlight. Never place any heavy objects on the thermometer body.
Do not throw batteries into fire.
Only use recommended batteries. Do not use rechargeable batteries.
This thermometer does not replace a hospital diagnosis.
Do not drop, disassemble or modify the device.
Do not use this device if you think it is damaged or notice anything unusual.
This device comprises sensitive components and must be treated with caution. Observe the storage and operating conditions described in the ‘Technical
Specifications’ section.
Not servicing/maintenance while the thermometer is in use.
When using, shall not touch battery and the patient simultaneously.
Do not use the device if it is damaged/ degraded/loosened in any way. The continuous use of a damaged unit may cause injury, improper results, or serious danger.
Based on the current science and technology, other potential allergic reactions are unknown.
This equipment needs to be installed and put into service in accordance with the information provided in the ACCOMPANYING DOCUMENTS.

OVERVIEW

Intended Use
The HoMedics Non-Contact Infrared Body Thermometers are designed to be used for intermittent measurement and monitoring of human body temperature by consumers in
the home setting and primary care setting as screening tool.
Description of Non-Contact Infrared Body Thermometer

  • Device principle and introduction
  • The HoMedics Non-Contact Infrared Body Thermometer is a hand-held, reusable, battery operated device, which can measure human body temperature on forehead, the skin temperature on one’s forehead.
  • The operation principle is based on Infrared Sensor technology. The IR sensor can output different signal when measuring different object temperature or in different ambient temperature, and the ASIC can turn the signal from IR Sensor to a digital value and display it on the LCD.
  • Description on Controls, Indicators, and Symbols
    HOMEDICS Non-Contact Infrared Body Thermomet- Infrared
Thermometer
1. IR sensor 9. Data indicator
2. Liquid crystal display (LCD) 10. Indicator of measurement results
3. MODE button 11. Low Battery indicator
4. MEMO button 12. Voice on/off indicator
5. ON/scan button 13. Memory number
6. Battery cover 14. Fahrenheit
7. Surface mode 15. Celcius
8. Body mode 16. Memory indicator

THERMOMETER APPLICATIONS

Thermometer
Model Number

|

Thermometer Style

|

Adult

| Pediatric
---|---|---|---

Ear

| Forehead| Ear|

Forehead

HTD8816C TE-350-EU

|

Non-Contact Infrared Body Thermometers

| | √| |

EQUIPMENT SYMBOLS

| Warning| | Restriction of Hazardous Substances
---|---|---|---
| Non-sterile packaging| | Batch code
| Refer to operating instructions| | Compliance with WEEE Standard
| Operating Temperature| | DO NOT THROW AWAY Intended for multiple use
| Operating Humidity| | Operating atmospheric pressure
| This device compiles with Part 15 of FCC (Federal Communications Commission) Rules.|

0598

| Indicates this device is in compliance with MDD 93142/EEC. 0598 is the Notified Body Number
| Manufacturer| HOMEDICS Non-Contact Infrared Body Thermometer--
Thermometer| Authorized Representative in the European community
| Recyclable| | Serial number
| IP22: The first number 2: Protected against solid foreign objects of ® 12.5 mm and greater. The second number: Protected against vertically falling water drops when enclosure

TECHNICAL SPECIFICATIONS

Measurement unit °C/°F
Operating mode Adjusted mode (Body mode)

Direct mode (surface mode)
Reference Body Site| Axillary
Rated output range| Body mode: 34.0 – 43.0°C / 93.2 – 109.4°F
Surface mode: 0 – 100.0°C / 32 – 212°F
Output range| Body mode: 34.0 – 43.0°C / 93.2 – 109.4°F
Surface mode: 0 – 100.0°C / 32 – 212°F
Laboratory Accuracy| Body mode:
34.0 – 34.9°C: ± 0.3°C / 93.2 – 94.8°F: ± 0.5°F
35.0 – 42.0°C: ± 0.2°C / 95.0 – 107.6°F: ± 0.4°F
42.1 – 43.0°C: ± 0.3°C / 107.8 – 109.4°F: ± 0.5°F
Surface mode: ± 2°C / ± 3.6°F
Display Resolution| 0.1°C/0.1°F
Three-colour Backlight
(Colour Alarm)| 35.5 – 37.3°C / 95.9 – 99.1°F: Green (Normal Temperature)
37.4 – 38.0°C (Alarm point) / 99.3 – 100.4°F: Yellow (Slight Fever)
38.1 – 43.0°C / 100.6 – 109.4°F: Red (High Fever)
Note: 1. Surface mode is always with Green backlight
2. In body mode 34.0-35.4°C is with green backlight
Auto Power Off Time| ≤18s
Measuring Time| ≤ 2s
Measuring Distance| 1 – 5cm (0.4 – 2 in)
Memories| 50
Power Supply Requirements

Batteries| 2 x 1.5V (AAA) Alkaline battery DEC Type LR03)
Adaptable Range| 2.6V – 3.6V
Environmental
Operating Condition| Operating Temperature:
15 – 40°C (59 – 104°F)| Relative Humidity:
5 85%| Atmospheric pressure:
70-106Kpa
Transport and Storage Condition| Storage Temperature:
-20 – 55°C (-4 – 131°F)| Relative Humidity:
5 93%| Atmospheric pressure:
70-106Kpa
Dimension and Weighting

Weight (without batteries)| 80g
Size| L:122mm X W:96mm X H:42mm
Compliance
Item| Compliant with
Equipment classification| Safety Standards:

EN 60601-1: 200641:2013, EN 60601-1-2: 2015

Type of protection| Internally powered equipment (on battery power)
Degree of protection| Non-Applied part
Front panel and case labelling| EN 15015223-1:2016
Temperature| EN IS080601-2-56:2017
Home healthcare environment| EN 60601-1-11:2015

Calculated values of the indicators according to ISO 80601-2-56

Indicators Calculated value
Clinical bias, Acb -0.027
Standard deviation, of 0.14
Limits of agreement, LA 0.26
Clinical repeatability, or 0.07

Note: the above value is calculated from clinical data of HTD881 BA.

Safety classification of ME EQUIPMENT

Protection against electric shock Internally powered ME equipment
Applied part Non-Applied part
Protection against harmful ingress of water or particulate matter IP22
Mode of operation Continuous operation

Note: Not intended to be sterilized. Not for use in an OXYGEN RICH ENVIRONMENT.

OPERATION

Battery installation
Caution: The Non-Contact Infrared Body Thermometer does not operate with dead batteries or external power sources. Install new batteries.

  1. Slide the battery door downward, toward the bottom of the Non-Contact Infrared Body Thermometer, and remove the battery access door;
  2. Insert two pieces AAA size batteries according to the “+” and “-”;
  3. Close the battery cover.

How to operate
Before applying the Thermometer, be sure to read and understand all warnings listed of the instructions before use.

  • The thermometer is aligned with the middle of the forehead to measure body temperature between the eyebrows above and keep the vertical distance, press the
    On/scan button, temperature will display immediately, see figure 2.

HOMEDICS Non-Contact Infrared Body Thermomet-
Measuring

Figure 2: Measuring position and distance

  • When the ambient temperature of the thermometer changes too much, such as moving the Thermometer from one place of lower temperature to another place of higher temperature, allow the thermometer to remain in a room for 30 minutes where the temperature is between 15°C to 40°C.
  • The ambient temperature around the test person should be stable, should keep away from the larger flow fan, air-conditioning vents and so on.
  • When people moving from one place of lower temperature to another place of higher temperature, should at least remain in the test environment more than 5 minutes, to be consistent with the ambient temperature after the re-measurement.
  • Wait at least 1 second for the next measurement. If the continuous measurement of five times, it is recommended to wait at least 30 seconds and then continue measurement.
  • You cannot use the thermometer in place where the sun is strong.
  • If for some reason the forehead temperature measurement is low, try measurement behind the ears. See figure 3.

HOMEDICS Non-Contact Infrared Body Thermometer--o
measurement

Figure 3: Align behind the ears to measurement

GENERAL SET UP AND USE

Start measuring

  1. Turn on the thermometer by pressing the On/scan button. The thermometer will perform self-test with all segments displayed for 2 seconds.
  2. Align the thermometer with the forehead, keeping the recommended distance, and then press the ON/scan button to start the measurement. Data reading will be displayed.

Note:

  1. After measurement you will hear a “beep beep beep” three times which means that the measurement is complete. The temperature is displayed on the LCD. The LCD backlight colour will change red, green or yellow according to the measurement. Green means Normal Temperature (35.5-37.3°C / 95.9-99.1°F) ready for next measurement. Yellow means Slight Fever Warning (37.4-38.0°C / 99.3-100.4°F) Please pay attention to body temperature. Red means High Fever (38.1-43.0°C / 100.6-109.4°F) Please take action to cool down or go to the doctor.
  2. To ensure the accuracy of the measurement, wait at least 30 seconds after 5 consecutive measurements.

Mode conversion

  • When the device is running, pressing the MODE button to cycle conversion between “body” mode and “surface” mode.
  • “Body” mode is used for measuring human body temperature, the “surface “ mode is used to measure the surface temperature. (The factory default is “body” mode)

Recalling and Erasing Memory Data
The last temperature taken before the thermometer powers off is stored in memory, up to 50.

  1. In the boot or shutdown state, short press the MEMO button to view the history of measured values.
  2. An empty memory cell shows “—°C” or “—°F”.
  3. Temperature readings can be stored in memory. Up to 50 temperature readings can be stored into the memory cells and automatically overwrite historical data.
  4. In boot mode, press the MEMO button until the LCD display “CLR”, which means that all stored data is cleared completely.

Parameter settings
As below step:

  1. Unit Set-F1
    Under the boot mode, long press the “MODE” button to enter F1, press the “MODE” or “MEMO”button to switch Celsius and Fahrenheit temperature units, press the On/ Scan button to confirm the unit settings (factory default is Celsius).

  2. Fever alert set-F2
    Under F1 state, press ON/scan button to enter the F2, press the “MEMO” button to decrease 0.1°C, press the “MODE” button plus 0.1°C, long press to accelerate the
    speed of temperature regulation, and finally press the ON/scan button to save. (The factory default is 38.1°C)

  3. Prompt sound settings-F3
    Under F2 state, short press ON/scan button to enter F3, press MODE button or MEMO button to set voice switch, and press the ON/scan button to confirm the
    settings. (The factory default is the voice Prompt to open).
    ♦ Prompt sound settings ON/OFF function
    Under the boot mode. Short press on/off button, at the same time press the “MODE” button to take on or take off the Prompt sound function.
    ♦ Restore to factory setting function
    Under the boot mode, short press on/off button, at the same time press the MEMO button until LCD display “rst”. Two seconds later, former F1-F3 parameter back to
    factory setting.

TROUBLESHOOTING

Message

| Situation|

Solution

---|---|---
| Temperature taken in not within Typical human temperature range.
(34.0°C – 43.0°C or 93.2°F – 109.4°F).| Make sure the forehead thermometer is for forehead measurement, not other human body site.
| Measured over the distance: 1-5cm (0.4-2in)| Optimum measurement distance is 1 cm
Incorrect test position.| See figure 2 Measuring position and distance.
Subjects forehead hair, Antipyretic stickers, head with sweat, etc.| Subjects sit quietly 5-10 minutes before the test.
Some people’s body temperature is lower than the general population.| The main concern fever temperature
| Operating temperature exceeds the range of specified temperature.| Move to a room within the operating range wait 30 minutes before taking temperature.
| The screen flicker, automatic turn off.| Replace battery or the product has been damaged, needs repairs.
| Battery capacity is too low. Taking Temperature is not allowed.| Install a new battery
| Ambient temperature changes too fast| Wait until the ambient temperature is stable.
| 1. Power is off.
2. Improper battery installation.
3. The battery is exhausted.
4. Display remains blank.| 1. Press ON button again.
2. Check the battery polarity.
3. Replace with a new battery.
4. Contact the retailer or service center.

REPLACING THE BATTERY

  1. Open and release battery cover following indicator on the surface of battery cover. Before changing the battery be sure the thermometer is already turned off.
  2. Remove the battery and replace with 2 new type AAA batteries, Make sure to align properly as indicated inside the battery cover.
  3. Slide the battery cover back in until it snaps in place. Do not dispose of used batteries in household waste. Take them to special local collection sites.
  4. If the thermometer appears frozen after changing the batteries, remove batteries, wait 30 seconds and replace batteries again.

WARNING
Do not recharge, disassemble or dispose of in fire.

  1. The typical service life of the new and unused batteries is 2000 measurements for the operation time is 18s.
  2. Only use the recommended batteries, do not recharge non-rechargeable batteries and do not burn them.
  3. Remove the batteries if the thermometer is not to be used for a long period.

CLEANING, CARE AND STORAGE

  • The lens is very delicate.
  • It is very important to protect the lens from dirt and damage.
  • Use a clean, soft cloth to clean the surface of the device and LCD. Do not use solvents or immerse the device into water or other liquids.
  • Always keep the thermometer within the storage temperature range (- 20°C to 55°C or – 4°F to 131°F) and humidity range (≤93% non-condensing).
  • It is recommended to store the thermometer in a dry location free from dust. Do not expose the thermometer to direct sunlight, high temperature/ humidity or any extreme environment, otherwise, the function will be reduced.
  • When the ambient temperature of the thermometer changes too much, such as moving the thermometer from one place of lower temperature to another place of higher temperature, allow the thermometer to remain in a room for 30 minutes where the temperature is between 15°C to 40°C.

DISPOSAL

  • Used batteries should not be disposed of in the household rubbish. Used Batteries should be deposited at a collection point.
  • At the end of its life, the appliance should not be disposed of in household rubbish.
  • Enquire about the options for environment-friendly and appropriate disposal. Take local regulations into account.

WARRANTY

Our company warrants Non-Contact Infrared Body Thermometer at the time of its original purchase and for the subsequence time period of 2 years.
The warranty does not cover the followings:

  • The device series number label is torn off or cannot be recognized.
  • Damage to the device resulting from misconnection with other devices.
  • Damage to the device resulting from accidents.
  • Changes performed by users without the prior written authorization of the company.
  • Batteries and packaging are not covered under warranty

When asked to provide warranty service, you must have a purchase date and purchase stamp dealers (including dealers name and address) of the warranty card. Be sure to ask the dealer to purchase this product signature on the warranty card. When asked to provide warranty service, please put the product to get our distribution points for repair. Products outside the warranty expires, will be charged accordingly.

Note:

  1. If you have any problems with this device, such as setting up, maintaining or using, please contact with SERVICE PERSONNEL of FKA Brands Ltd. Don’t open or repair the device by yourself.
  2. Please report to FKA Brands Ltd if any unexpected operation or events occur.
  3. Calibration needed to ensure proper function for every two years or after device impact.
  4. The patient is an intended operator. The patient can measure and change battery. Under normal circumstances and maintain the device and its accessories according to the user manual.

EMC DECLARATION

  1.  This equipment needs to be installed and put into service in accordance with the information provided in the ACCOMPANYING DOCUMENTS; This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided, and this unit can be affected by portable and mobile RF communications equipment.
    • Caution: Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in incorrect operation of the unit.
  2. *Caution: This unit has been thoroughly tested and inspected to assure proper performance and operation!
    • Caution: this machine should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, this machine should be observed to verify normal operation in the configuration in which it will be used.

Guidance and manufacturer’s declaration-electromagnetic emissions


The Non-Contact Infrared Body Thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the Thermometer should assure that it is used in such an environment.

Emissions test

| Compliance|

Electromagnetic environment guidance

RF emissions CISPR 11|

Group 1

| The Non-Contact Infrared Body Thermometer uses RF energy only
for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11|

Class B

| The Non-Contact Infrared Body Thermometer is suitable for use in all establishments other that domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions

IEC 61000-3-2

| Not applicable
Voltage fluctuations

Flicker emissions

IEC 61000-3-3

| Not applicable

Guidance and manufacturer’s declaration-electromagnetic immunity


The Thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the Thermometer should assure that it is used in such an environment.

Immunity test

| IEC 60601 test level| Compliance level|

Electromagnetic environment-guidance

Electrostatic discharge (ESD) IEC 61000-4-2| ± 6kV contact
+15kV air| ±6kV contact
±15kV air| Floor should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst IEC 61000-4-4| ±2 kV for power supply lines
+1 kV air for input/output lines| Not applicable| Mains power quality should be that of a typical commercial or hospital environment.
Surge IEC

61000-4-5

| +/- lkV line(s) to tine(s)
=/- 2kV line(s) to earth| Not applicable| Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11| <5% UT (95% dip in UT for 0.5 cycle 40% UT (60% dip in UT for 5 cycles 70% UT (30% dip in UT for 25 cycles 5% UT (>95% dip in UT for 5 sec| Not application| Mains power quality should be that of a typical commercial or hospital environment. If the use of the Thermometer requires continued operation during power mains interruptions, it is recommended that the Thermometer be powered from an uninterruptible power supply or a battery.
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8| 30 Alm| 30 Alm| Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration-electromagnetic immunity

The Thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the Thermometer should assure that it is used in such an environment.
Immunity test| IEC 60601 test level| Compliance level| Electromagnetic environment-guidance
Conducted RF
IEC 61000-4-6Radiated RF
IEC 61000-4-3| 3Vrms 150kHz to 80 MHz
10 Vim 80MHz to 2.5 GHz| Not applicable

10 V/m

| Portable and mobile RF communications equipment should be used no closer to any part of the Thermometer, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance.
d=1.2
d=1.2 80 MHz to 800 MHZ
(d=1.2 800 MHz to 2,5 Ghz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,” a. should be less than the compliance level in each frequency range: b. Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

a.   Field strengths from fixed transmitters, such as base stations from radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast can not be predicted theoretic call with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Thermometer is used exceeds the applicable RF compliance level above, the Thermometer should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Thermometer.
b.  Over the frequency range 150kHz to 80MHz, field strengths should be less than 10 Vim.

Recommended separation distances between portable and mobile RF communications equipment and the Thermometer

The Thermometer is intended for use in an electromagnetic environment in which radiated RF distances are controlled. The customer or the user of the Thermometer can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RE communications equipment (transmitters) and the Thermometer as recommended below according to the maximum output power of the communications equipment.

Rated maximum output power of
transmitter / W

|

Separation distance according to frequency of transmitter /m

150 kHz to 80 MHZ
d=1.2

|

80 MHz to 800 MHZ
d=1.2

|

800 MHz to 2.5 GHz
d=2.3

0.01

| 0.12| 0.12| 0.23

0.1

| 0.38| 0.38|

0.73

1

| 1.2| 1.2|

2.3

10

| 3.8| 3.8|

7.3

100

| 12| 12|

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

BODY TEMPERATURE

  • Body temperature varies from person to person and fluctuates during the course of the day. For this reason, it is suggested to know one’s normal, healthy forehead temperature to correctly determine the temperature.
  • Body temperature runs approximately from 35.5°C to 37.8°C (95.9°F-100°F). To determine if one has a fever, compare the temperature detected with the person’s normal temperature. A rise over the reference body temperature of 1°C (1°F) or more is generally indication of fever.
  • Different measurement sites (rectal, axillary, oral, frontal, auricular) will give different readings. Therefore it is wrong to compare the measurement taken from different sites.
  • Here below are typical temperatures for adults, based on different measurement sites:

Rectal: 36.6°C to 38°C /97.9°F-99.1°F
Axillary: 34.7°C to 37.3°C /94.5°F-99.1°F

ASTM laboratory accuracy requirements in the display range of 37°C (98°F to 102°F) for IR thermometers is ±0.2.(±0.4°F), whereas for mercury-in-glass and electronic thermometers, the requirement per ASTM Standards E667-86 and E1112-86 is ±0.1.(±0.2°F).

Caution: This infrared thermometer meets requirements established in ASTM Standard (E1965-98) Except of clause 5.2.2. It displays the subject’s temperature over a range of 34.0°C ~43.0°C. Full responsibility for the conformance of this product to the standard is assumed by (Hetaida Technology Co., Ltd. Add: 4F, Baishida High-Tech Park, Xiandong Industrial Area, Dalingshan Town, Dongguan City, Guangdong, China).

2 YEAR GUARANTEE

FKA Brands Ltd guarantees this product from defect in material and workmanship for a period of 2 years from the date of purchase, except as noted below. This FKA Brands Ltd product guarantee does not cover damage caused by misuse or abuse; accident; the attachment of any unauthorised accessory; alteration to the product; or any other conditions whatsoever that are beyond the control of FKA Brands Ltd. This guarantee is effective only if the product is purchased and operated in the UK / EU. A product that requires modification or adaptation to enable it to operate in any country other than the country for which it was designed, manufactured, approved and /or authorised, or repair of products damaged by these modifications is not covered under this guarantee. FKA Brands Ltd shall not be responsible for any type of incidental, consequential or special damages. To obtain guarantee service on your product, return the product post-paid to your local service centre along with your dated sales receipt (as proof of purchase). Upon receipt, FKA Brands Ltd will repair or replace, as appropriate, your product and return it to you, post- paid. Guarantee is solely through HoMedics Service Centre. Service of this product by anyone other than the HoMedics Service Centre voids the guarantee. This guarantee does not affect your statutory rights. For your local HoMedics Service Centre, go to www.homedics.co.uk/servicecentres

HOMEDICS Non-Contact Infrared Body Thermometer--
Thermometer Wellkang Ltd
16 Castle St, Dover, CT16 1PW, UK

Manufactured by
Hetaida Technology Co., Ltd. 4F, BaiShiDa High-Tech Park,
XiangDong Industrial Area,  DaLingShan Town, DongGuan City,

Guangdong, China.
Distributed by
FKA Brands Ltd
Somerhill Business Park