S-BOL PA01 Handheld Infrared Thermometer User Manual
- June 6, 2024
- S-BOL
Table of Contents
Infrared Thermometer
PA01
User Manual
Infrared Thermometer
-
Product Name
Infrared Thermometer -
Model Name
PA01 -
Software Release Version
Version 1 -
Product Composition
The infrared thermometer is mainly composed of an infrared temperature sensor, a probe cover, a display unit, a power supply circuit, and an easurement circuit. -
Application
The body temperature of the subject is displayed by measuring the heat radiation from the forehead. -
Basic Working Principle
The infrared thermometer measures the infrared radiation exchange between the forehead of the human body to be measured and the appropriate correction value and outputs the actual measurement data of the temperature of the human forehead. Finally, the actual measurement data of the measured object is displayed by measuring the heat radiation of the human forehead. -
Product Features
- High-precision infrared sensors, stable and reliable performance;
- Measured values automatically saved;
- Large-size LCD screen, high brightness backlight, clear and soft display;
- Automatic shutdown saves energy.
- Clean and hygienic: non-contact forehead temperature measurement, measuring distance 5cm-15cm, no contact with human skin to prevent cross-infection.
- Simple and practical: One-button measurement, one-button memory check, easy operation.
- Multiple sets of memory: 32 sets of measurement data can be stored for easy analysis and comparison.
Technical Indicators
Body temperature testing range | 32.0℃~42.4℃ |
---|---|
Measuring Accuracy | Temperature mode: ±0.3 ℃. |
Measuring distance | 1 cm~15 cm |
Operating environment temperature | Temperature range:10 ~ 40 ℃, humidity ≤85% |
Transportation and storage temperature | Temperature range: -20 ℃ ~ +55 ℃, |
humidity ≤93%
Power| DC9V
Auto power off| 90 seconds
Battery Alert| Battery Low Alert
Backlight| High Brightness Backlight
Size| L 100xW 46xH 160mm
Weight| 123.1g(without battery)
Usage and Operation
Advice before Use
Infrared thermometers have high requirements on the surrounding environment,
and correct use is the key to accurate testing.
- This product is an infrared forehead thermometer that professionally measures the temperature of the human body. According to the difference in human skin, the measured temperature will be different.
- When the room temperature changes greatly, the infrared frontal thermometer should be placed in the room for 20 minutes before use.
- When the person to be measured comes from a place with a large difference in temperature from the measurement environment, they should stay in the test environment for at least 5 minutes, and wait for the temperature to be consistent with the environment before measuring.
- The environment around the person to be tested must be stable. Do not test in places with large airflows such as fans and air outlets of air conditioners.
- Do not use the instrument outdoors or in a place with strong sunlight. If stored or used outside the temperature and humidity specifications specified above, the system may not meet the stated performance specifications.
- Do not hold the front end of the thermometer during measurement.
- The measurement site must not be covered by hair. If there is sweat, please wipe it with a dry towel and measure it later. Otherwise, it will affect the measurement result.
- Cold measurement of the forehead of a fever patient, sweating, and other cooling measures will make the measurement result low, and measurement should be avoided in this case.
- Comparison of different temperature measurement methods. The body temperature measured by different temperature measurement methods will be different. The specific temperature differences are shown in the table below for reference.
Measurement Area | Normal Temperature |
---|---|
Anal temperature | 36.6℃-38℃ |
Oral temperature | 35.5℃-37.5℃ |
Underarm temperature | 34.7℃-37.3℃ |
Cochlea temperature | 35.8℃-38℃ |
Forehead temperature | 35.8℃-37.8℃ |
Quick Installation
- Install the battery
- The first time you use it, you need to put in a new battery, and then you can take a temperature measurement.
- The front of the thermometer is aimed at the forehead, and the distance is 1 cm -15 cm. At the same time, make sure that there is no hair, sweat, cosmetics, or hat cover, then press the measurement switch to start the measurement.
- When the ambient temperature changes affect the forehead temperature or forehead sweating, please measure after measuring the earlobe. Also make sure that there is no hair, sweat, makeup, or hat covering.
Note: The measured temperature value will vary depending on the person’s skin
color, thickness, and body part, and when the ambient temperature changes
greatly, which is normal. The reason is that the more exposed human body parts
are, the more affected by the ambient temperature. Battery Replacement
The instrument uses a DC9V battery. Generally, it can be used continuously for
more than 10000 times. When a battery symbol appears on the top right of the
screen and flashes, it means that the battery is insufficient and the battery
needs to be replaced as soon as possible.
- Open the battery cover and replace the battery. Pay attention to the correct position of the positive and negative electrodes. The product is designed with a battery anti-reverse device. Even if the battery is installed in the wrong polarity, it will not cause damage to the product, but it will not work.
- Do not use rechargeable batteries. Use disposable batteries. Use alkaline batteries whenever it is possible.
- When the instrument is not working, the static current is extremely small. But when it is not used for a long time, it is recommended to remove the battery to prevent battery leakage from damaging the product.
Precautions
- When opening the battery cover to replace the battery, pay special attention to the polarity of the battery. Misplacement may cause damage to the product.
- When not in use for a long time, please remove the battery to prevent the thermometer from being damaged due to battery leakage.
- Do not use if the battery is leaking or moldy.
- Do not place the battery near a fire source or throw it into a fire to prevent the battery from exploding.
- Do not store the battery in a high temperature or high humidity environment.
- To avoid a short circuit, do not put the battery in the same pocket with metal objects such as coins or keys, or other containers that may shorten the battery life.
- Do not dispose of used batteries arbitrarily, they should be classified and disposed of according to local regulations.
- Check the battery for power before using the thermometer every time.
Operating Instructions
Description of the thermometer
1. Infrared sensor
2. Measurement port
3. LCD liquid crystal display
4. “+” button
5. “-” button
6. “M” button| 7. Backlight button
8. Sound button
9. Measurement button
10. Battery cover
11. Handle
---|---
Operation Steps
- How to measure temperature. Press and hold the measurement button (9) on the front of the thermometer to continuously measure temperature. If you only need one measurement, please release the button after the measurement is shown on the display.
- Measurement counter in the lower-left corner of the screen. The count value is cycled from 1 to 32. The counter will increase by 1 each time a valid measurement is performed and then return to 1 after it reaches 32. If the measurement result is “L0” or “HI”, it is an invalid measurement, invalid measurement results are not saved to the internal memory and will not add to the count.
- View stored measurements. The thermometer will automatically save up to 32 valid measurements. To view stored measurements, press the “M” button, then press “+” or “-“. Press “-” to display the latest measured temperature, and press “+” to display the first measurement. Press “+” or “-” again to cycle through the measurements.
Backlight and sound
(1) Short press the backlight button to turn the backlight on and off.
(2) Short press the sound button to turn the sound on and off.|
Symbols
Symbols | Description |
---|---|
BF-type equipment | |
Please refer to the attachments | |
Keep away from rain | |
Please refer to the instructions |
Warnings and Instructions
Read this manual carefully before use.
The ambient temperature of this product is 10 ℃ ~ 40 ℃, the optimal
temperature is 25 ℃.
Don’t use this product in an environment above 40 ° C or below 10 ° C.
Don’t place this product too close to a charged object to avoid electric
shock.
Don’t use this product in an environment with a relative humidity greater than
85%.
Don’t place this product too close to the electromagnetic range (such as
radio, mobile phone, etc.).
Don’t expose this product to the sun or near the stove. Do not touch the
water.
Don’t drop this product, do not use it if it is damaged.
Hair on the forehead, sweat, hat, or scarf will affect the accuracy of the
measurement data.
Please make sure that the measurement distance is not more than 15cm.
When forehead sweating or other reasons cause the forehead temperature to not
normally respond to body temperature, please measure from the earlobe.
When cleaning is required, wipe the surface of the product lightly with
alcohol.
In the event of a problem with the product, please contact your dealer and do
not attempt to repair it yourself.
Make sure to remove the hair and sweat from your forehead before measuring.
The use of this product is not a substitute for a doctor’s diagnosis.
The protective glass outside the LCD frame is very important, and it is also a
fragile part of the thermometer. Please use it with care. Do not charge non-
rechargeable batteries and do not throw them into the fire.
Do not expose the product to direct sunlight or water.
Common Faults
If you encounter the following problems during use, please refer to the
following guides to
help solve the problem. If the problem is still not resolved, please call our
after-sales service:
(1) HI is shown on the display.
When the measured temperature exceeds 43 degrees, the thermometer will display
HI, first check whether the measured part of the human body has been in
contact with an external heat source. If not please call after-sales service.
(2) L0 shown on the display.
When the measured temperature is lower than 32 degrees, the instrument will
display L0. Follow the prompts in the form below to find the cause. If it is
confirmed that the reason is not in the form, you need to call our after-sales
service. The main reasons why the screen displays “L0”
Possible reason | Suggestion |
---|---|
Hair or sweat | Ensure there are no obstructions or sweat on the forehead. |
Cold air blows on the forehead | Ensure stable air during the measurement |
Shortly after cold compress | Measure 10 minutes after cold compress |
The measuring distance is too far | The best measurement distance is 5cm -15cm. |
Do not place the thermometer against the skin
Maintenance
During the use of the product, please follow the prompts when you find the
following conditions.
External dirt : Wipe the dirt with a clean soft cloth, or wipe with a
cotton swab moistened with medical alcohol. Wiping with medical alcohol can
also have sterilization and disinfection.
Take care not to add too much water or alcohol to prevent damage to the inside
of the product.
Internal dirt: The lens of the internal probe is an important device. Do
not touch or press with your fingers or other objects, otherwise it will
affect the preparation of the measured value.
When the lens surface is found to be dirty, wipe the lens surface with a
cotton swab moistened with 95% or more absolute alcohol.
Tip : Do not use 75% disinfectant alcohol to wipe the lens (there will be
traces of residual water). Do not use other chemical wipes to wipe the lens
(which will cause damage to the lens).
Storage : Keep in a dry and dark place out of direct sunlight.
It is recommended to practice several times to become familiar with the
measurement method and try not to change the product factory settings. The
measurement results can’t replace physician diagnosis.
During use, if any problem, please contact the seller or manufacturer
Please dispose of the waste and residues at the end of the service life of the
product according to local laws and regulations.
If there is a product failure, please do not disassemble and repair it by
yourself, please contact the after-sales service phone. After the product
reaches the end of its life, please send it to a suitable place to protect the
environment. (Note: The disposal of wastes and residues should comply with the
corresponding national laws and regulations.
Warranty
This limited warranty will last for 1 year from the date of purchase. The
battery is not covered under warranty. The device should have a hardware
fault. Warranty will be void, even if a repair has been attempted by any
unauthorized service center. Physical, Liquid damage & Customer abuse will not
be covered under warranty. Valid proof of purchase is mandatory to claim a
warranty. (Customer Purchase Invoice is necessary).
Transportation and Storage
It should be stored in a clean, cool, dry, and ventilated place. It should not
be stored in the open air. It should not be stored in the same place as
odorous and corrosive items.Storage conditions: -10 ℃ ~ 55 ℃, 30% RH ~ 93% RH,
normal pressure.
EMC Instructions
This product complies with the EMC (Electromagnetic Compatibility) standard
YY0505-2012 required for the safe use of medical electrical equipment. The EMC
standard is a standard developed for the safe use of medical electrical
equipment. The standard stipulates that the interference of electromagnetic
waves generated by the equipment to other equipment, and the interference of
electromagnetic waves emitted by other equipment (mobile phones, etc.)
must be controlled within a certain range. YY0505-2012 stipulates detailed
information about the EMC environment required to provide users with safe
equipment operation. The following is a description of EMC-related technical
descriptions. When this product works in the electromagnetic environment
locked in this EMC technical data, the basic performance described in its
scope of application will not be affected.
Definition of EMC
- EMC (electromagnetic compatibility) refers to the ability to meet the following two requirements.
- Do not emit electromagnetic interference noise that is not allowed to other electronic devices nearby. (radiation)
- The device can perform its function normally in an electromagnetic environment in which other electronic devices emit noise and other interference. (Immunity)
- EMC (electromagnetic compatibility) related technical descriptions;
- Note : The basic performance is: “During the EMC test, the thermometer display normally displays, continuous temperature measurement, and trouble-free operation.
- Warning: Portable and mobile RF communication equipment may affect the normal operation of this instrument.
- Warning : It is the responsibility of the user to ensure the electromagnetic environment of the equipment so that the instrument can work normally.
- Warning : Do not use this equipment near strong radiation sources, otherwise it may interfere with the normal operation of this instrument.
- Warning : In addition to the transducers, accessories, and cables sold by the manufacturer of this instrument or system as spare parts for internal components, the use of unspecified accessories, transducers, and cables may result in increased emissions or immunity of the equipment or system Degree reduction.
- Warning : This instrument should not be used close to or stacked with other equipment. If it must be used close to or stacked, it should be observed to verify that it can operate normally in its used configuration.
- It is recommended to use the DC 9V 6F 22 battery provided by the manufacturer. List of accessories for this equipment: DC9V 6F22 battery, instruction manual, warranty card, certificate of conformity.
See the following table for device battery compatibility information
Guidance and manufacturer’s declaration-electromagnetic emissions
This infrared thermometer is intended for use in the electromagnetic
environment specified below, and the purchaser or user should guarantee its
use in this electromagnetic environment
Emission Test| Compliance Electromagnetic
environment Guide
---|---
Radiofrequencyemission GB 4824| Group 1| This infrared thermometer uses
radiofrequency energy only for its internal functions. As a result, its RF
emissions are very low and the possibility of interference with nearby
electronic equipment is extremely low.
Radiofrequencyemission GB 4824| Class B| This product is suitable for use in
all facilities.
Harmonic radiation
GB17625.1| Not Applicable
Voltage fluctuation/ flicker emission
GB17625.2| Not Applicable
Guidance and manufacturer’s declaration-electromagnetic immunity
This infrared thermometer is intended for use in the electromagnetic
environment specified below, and the purchaser or user should guarantee its
use in this electromagnetic environment.
Immunity test| ICE 60601
Test level| Coincidence level| Electromagnetic Environment-Guide
---|---|---|---
Electrostatic discharge GB/T17626.2| ±6kV contact discharge ± 8kV air
discharge| ±6kV contact discharge
± 8kV air discharge| The floor is recommended to bewooden, concrete or tiled
floor. When the floor is coated with syntheticmaterial, the relativehumidity
is
recommended to beat least 30%.
Electrical fast transient burst GB/T17626.4| ± 2kV power cord
± 1kV input / output
line| Not Applicable| Not Applicable
Surge GB/T17626.5| ± 1kV wire-to-wire
± 2kV line to ground| Not Applicable| Not Applicable
Voltage sags on power input linesshort-term
nterruptions andvoltage changesGB/T17626.11| <5% UT for 0.5 weeks
(On UT,> 95% dips)
40% UT for 5 weeks
(60% sag on UT)
70% UT for 25 weeks
(30% sag on UT)
<5% UT for 5s
(At UT,> 95% sag)| Not Applicable| Not Applicable
Power frequency magnetic field ( 50/60Hz )
magnetic field GB/T17626.8| 3A/m| 3A/m,50Hz| If the instrument does not work
properly, it is necessary to keep the instrument away from the power
frequency magnetic field source.
Note: UT is the AC network before the test voltage is applied.| | |
Guidance and manufacturer’s declaration-electromagnetic immunity
This infrared thermometer is intended for use in the electromagnetic
environment-specific below, and the purchaser or user should guarantee its use
in this electromagnetic environment.
Immunity test| ICE 60601
Test level| Coincidencelevel| Electromagnetic Environment-Guide
---|---|---|---
Radio
frequency
conduction
GB/T17626.6| 3V(Valid
value)
150KHz~
80M Hz| Not
applicable3V/m| Portable and mobile RF communications
equipment should not be used closer to any
part of this product, including cables, than
the recommended isolation distance.
This distance is calculated by the formula of
the transmitter frequency response.
Recommended separation distance
d=1.2
Radio
frequency
radiation
GB/T 17626.3| 3V/m
80MHz~
2.5G Hz| | 80MHz~800MHz d=2.3 p
800MHz~2.5GHz
P is the maximum rated output power of the transmitter specified by the
transmitter manufacturer, in watts (W), d is the
recommended separation distance, in meters (m). The electric field strength of
a fixed RF transmitter is determined by a site
survey a) of the electromagnetic field and should be lower than the compliance
level in each frequency band range. Interference may occur near the equipment
marked with the following symbol.
1. At the frequency points of 80MHz and 800MHz, the formula of the higher
frequency bands is adopted.
2. These guidelines may not be suitable for all situations. Electromagnetic
propagation is affected by absorption and reflection from buildings, objects,
and the human body.
3. Stationary transmitters, such as base stations for wireless (cellular /
fly-free) telephones and ground mobile radios, amateur radios, FM and FM radio
broadcasts, and television broadcasts, the theoretical field strengths cannot
be accurately predicted in theory. To assess the electromagnetic environment
of fixed RF transmitters, surveys of electromagnetic sites should be
considered. If the measured field strength of the electronic thermometer is
higher than the applicable RF compliance level above, the electronic
thermometer should be observed to verify that it can operate normally. If
abnormal performance is observed, additional measures may be necessary, such
as reorienting or relocating the electronic thermometer.
4. In the frequency range of 150KHz ~ 80MHz, the magnetic field strength
should be lower than 3V / m.
The recommended separation distance between portable and mobile RF
communication equipment and this infrared thermometer
This infrared thermometer is expected to be used in an electromagnetic
environment where radio frequency radiation disturbance is controlled.
Depending on the maximum rated output power of the communication equipment,
the purchaser or user can maintain the portable and mobile RF communication
equipment (transmitter) and the infrared thermometer as recommended below. The
minimum distance between them to prevent electromagnetic interference.
Transmitter rating Output power (W)| Distance between different transmitter
frequencies (m)
---|---
150KHz -80MHz
d=1.2 p
| 80MHz -800MHz
d=1.2 p
| 800MHz -2.5GHz
d=2.3 p
0.01| Not applicable| 0.12| 0.23
0.1| Not applicable| 0.38| 0.73
1| Not applicable| 1.2| 2.3
10| Not applicable| 3.8| 7.3
100| Not applicable| 12| 23
For the maximum rated output power of the transmitter not listed in the table above, the recommended isolation distance d is in meters (m), which can be determined by the formula in the corresponding transmitter frequency column, where p is the emission provided by the transmitter manufacturer Maximum rated output power of the unit, in watts (W). At the frequency points of 80MHz and 800MHz, the formula of a higher frequency range is adopted. These guidelines may not be suitable for all situations. Electromagnetic propagation is affected by absorption and reflection from buildings, objects, and the human body.
Validity
Three years from shipment.
Manufacturer
Shenzhen Haoyuan Electronic Co., Ltd
Registration Certification No.:20202070421
Production License No.:20203597
Read User Manual Online (PDF format)