Solinst 4001 Readout Unit User Guide
- June 6, 2024
- Solinst
Table of Contents
4001 Readout Unit
SRU (Solinst Readout Unit) Quick Start Guide
More Info | Instructions | Get Quote
Connect SRU to Datalogger
Model 4001
USB-C Connection for PC Communication LCD Screen
Menu Push-Buttons
Rugged Rubber
Outer Cover
Serial Number
Internal Barometric
Pressure Sensor Interface
Connect SRU to Datalogger
The SRU cable connects to a Levelogger’s Direct Read
Cable, LevelVent Wellhead or AquaVent Wellhead Connector
Cable.
An L5 Threaded or Slip Fit Adaptor can also be used to
connect the SRU cable to a Levelogger.
SRU On/Off and Navigation
Press and hold any button to turn the SRU on. The SRU has an auto-off after 2 minutes of inactivity.
Use the up/down arrows to scroll through menu items.
Use the left/right arrows to select the menu actions shown at the bottom
left/right of the LCD (e.g. Next, Back, etc.)
SRU Menu Options
There are three main menu items for the SRU: Take a Reading: allows you to display an instant reading from the connected datalogger, turn on/off barometric
Transferring Data to a PC
Download the latest version of Levelogger Software from:
https://www.solinst.com/downloads/
Connect the SRU to the PC using a USB Cable. The SRU will automatically turn on and the LCD screen will show a message indicating the SRU is connected to the PC.
compensation, and start and save a real-time logging session. Download/View Files: allows you to download and save the data from the connected datalogger’s current internal
logging session, and view all data files saved in the SRU. Status/Setup: allows you to view information about the SRU and connected datalogger, including battery level, memory, serial number, logging status and firmware version. You can also adjust the contrast of the LCD screen.
- Start the Levelogger Software and open the SRU tab.
- Select the appropriate Com Port for the connected SRU from the centre drop-down menu.
- Click the ‘Retrieve SRU Settings’ icon. A window will open prompting you to select the location to save the data files.
- Select the file(s) from the list that you want to download and click the ‘Download Data’ icon.
- Data files can be viewed in the Data Control tab by selecting “Finish and Open all”, or select “Finish” if you do not need to view the data right away – data can be viewed later by going to the Data Control tab and selecting the ‘Open Files’ icon.
- All data files saved in the SRU can be erased by clicking the ‘Erase Files’ icon.
(Page 1 of 2) High Quality Groundwater and Surface Water Monitoring Instrumentation
Idle 99%
Take a Reading
Download/View Files
Idle 99%
Level (Comp): 10.045m Temperature: 28.234ºC
Idle 99%
Turn Compensation:
On
Idle 99%
Sample Rate: 12 mins
Log 56 99%
Level (Comp): 10.045m Temperature: 28.234ºC
Log 57 99%
Barometrically Compensated
Status/Setup
Conductivity: 125.4µS
Baro Reading: 98.89kPa
Compensation: On
Conductivity: 125.4µS
Data Saved
Next
Back Next
Back
Next
Off
Back Start
Idle 99%
Sample Rate: 12 mins Compensation: Off
Back Start
Stop Save
Log 56 99%
Level (Uncomp): 10.045m Temperature: 28.234ºC Conductivity: 125.4µS
Stop Save
Home
Log 57 99%
Uncompensated Data Saved
Home
Idle 99% Idle 99%
Downloading 99% Download 99%
Take a Reading
Download/View Files Status/Setup
Next
Download Logger
View Files
Back Next
Downloading Logs: Cancel
55%
Logger Data
Saved to file
1081225-01.10.21-12.30.xle Home
Idle
Download Logger View Files
99%
3/11 Files 99%
1081225-01.10.21-12.30 D 1081225-01.15.21-15.56 C 1081225-01.16.21-22.50 U
Idle Status: indicates whether the SRU is idle, logging, or downloading/ viewing data logs.
Battery icon: represents the battery life remaining for the SRU.
Idle 99%
Take a Reading
Download/View Files Status/Setup
Next
Back Next
Idle 99%
SRU Status
Logger Status
LCD Contrast
Back Next
Home Next 3
Idle 99%
SRU Serial: 123456 SRU Memory: 50% Firmware: V1.000 Back
D = Downloaded Data File
C = Compensated Real-Time Data File
U = Uncompensated Real-Time Data File
Note: Real-time logging, saving, or downloading data will not stop/interrupt the datalogger from logging internally. The datalogger memory will not be erased.
Idle 99%
SRU Status
Logger Status
LCD Contrast
Idle 99%
Logger Serial: 1081234 Logger Memory: 50% Logger Battery: 85%
Idle 99%
Log Mode: Schedule Logging Rate: 15 secs Logger Status: Started
Idle 99%
Location: Wellhead 234 Name: Zanzibar 1 Firmware: V1.004
Back Next
Idle 99%
SRU Status
Logger Status
LCD Contrast
Back Next
Back Next
Idle 99%
Set LCD
Contrast: 55% Back Home
Back Next
Back Home
Solinst Canada Ltd. 35 Todd Road, Georgetown, Ontario Canada L7G 4R8 www.solinst.com E-mail: instruments@solinst.com Tel: +1 905-873-2255; 800-661-2023 Fax: +1 905-873-1992 September 15, 2021
(#115926) (Page 2 of 2)
References
- Solinst Water Level Meters, Water Level Indicators, Boreholes and Standpipes
- Member Modal For CF Members
- Solinst Groundwater and Surface Water Monitoring Instrumentation, Water Level Meters, Interface Meters, Dataloggers
- Solinst Levelogger 5: Reliable Water Level Data Loggers
- SRU Solinst Readout Unit Instructions
- Solinst Readout Unit: SRU: Instant Access to Water Level Readings
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>