ACCU-CHEK Blood Glucose Monitoring System User Guide

June 6, 2024
ACCU-CHEK

Blood Glucose Monitoring System
Start Here
Quick Start Guide
For in vitro diagnostic use only

For use with the MiniMed TM 770G system with Bluetooth ® wireless technology

WARNING
Keep new and used batteries away from children. See the User’s Manual for additional information.

  • This Quick Start Guide does not replace the User’s Manual for your Accu‑Chek Guide Link blood glucose meter. The User’s Manual contains important cleaning and disinfecting procedures and additional information.
  • The meter and lancing device are for single patient use. Do not share them with anyone, including other family members! Do not use on multiple patients!
  • All parts of the kit are considered biohazardous and can potentially transmit infectious diseases from bloodborne pathogens, even after you have performed cleaning and disinfecting.

NOTE

  • Always have a spare set of batteries.
  • Battery life may vary due to factors such as temperature and battery manufacturer.

Meter Overview

1. Display
2. Back Button
3. Up Arrow and Down Arrow Buttons
4. Power/Set/OK Button
5. Test Strip Slot with Light| 6. Battery Door
7. Micro USB Port*
8. Test Strip Ejector

  • USB functionality is not available for this meter.
    ---|---

Set the Meter Language

Turn the meter on by briefly pressing OK. Language appears. Press or to highlight the desired language.

Press OK to set the language and return to Main Menu.

Pair the Meter and Insulin Pump

The Accu-Chek Guide Link meter can automatically send blood glucose results to the MiniMed TM 770G system with Bluetooth ® wireless technology. This feature eliminates the need to manually enter your blood glucose result on the insulin pump. Refer to the MiniMed TM 770G System User Guide for pairing instructions.

Set up the Lancing Device

Familiarize yourself with the Accu-Chek FastClix lancing device and the lancet drum.

Get the lancet drum from the pocket in the carry case.
NOTE
Each lancet drum contains 6 sterile lancets. You can perform 6 tests before replacing them.

Remove the cap by pulling it straight off. Do not twist the cap.
WARNING You must not insert the lancet drum into the lancing device and simultaneously press the release button or hold the lancing device with the release button resting on a surface such as a tabletop. This could release a lancet and inadvertently cause injury.

Insert the lancet drum, white end first, into the lancing device. Push the drum all the way in until it clicks firmly into place.
NOTE
Do not remove the lancet drum until all 6 lancets are used since the drum cannot be reused once it has been removed from the lancing device.

Place the cap back on the lancing device. Make sure the notch on the cap lines up with the notch on the lancing device.
Adjust the lancet depth to a comfortable penetration level. Start at 2. For tougher skin, dial to a higher number.

Using the lancing device: By pressing the release button all the way down, the lancing device primes and pricks with 1 click.

Perform a Blood Glucose Test

Wash and dry your hands thoroughly before and after you perform a blood glucose test.
Check the Use By date on the test strip container. Do not use test strips past the Use By date.

Remove a test strip from the test strip container. Close the cap tightly.

Insert the metallic end of the test strip into the meter.
The meter turns on. Preparing to test appears.

Press the lancing device firmly against the side of your fingertip and press the release button all the way down to prick your finger.

Touch the yellow edge of the test strip to the blood drop. Do not put blood on top of the test strip. Remove your finger from the test strip when Analyzing appears. Failure to move your finger away from the test strip could give inaccurate test results.
The test result appears on the display in less than 4 seconds. Press to automatically send results to a paired pump. If is not pressed, there will be a delay in seeing the blood glucose result on the paired pump.

Prepare the lancing device for the next test. Slide the lever over and back once to advance to the next lancet. The window shows
the number of lancets remaining.

Roche Diabetes Care, Inc.
9115 Hague Road
Indianapolis, IN 46256
www.accu-chek.com
ACCU-CHEK, ACCU-CHEK GUIDE, and FASTCLIX are trademarks of Roche. The Bluetooth ®word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by Roche is under license.
MiniMed is a registered trademark of Medtronic, Inc.

08163022001

© 2020 Roche Diabetes Care
08163022001-0820

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals