ACCU-CHEK

ACCU-CHEK IUT526 Pump Base

ACCU-CHEK-IUT526-Pump-Base-PRODUCT

Instructions for use

Intended Use
The Accu‑Chek Solo pump base is part of the micropump. It contains the mechanical parts as well as the electronics to control and monitor the operation of the pump. The Accu‑Chek Solo pump base is intended for continuous insulin infusion in the treatment of diabetes mellitus requiring insulin.

Package Contents and Storage Conditions

See outer packaging. You can find the symbol explanations in the User’s Manual of the Accu -Chek Solo micropump system.

Component Overview

ACCU-CHEK-IUT526-Pump-Base-FIG1

  • Reservoir cap (blue)
  • Pairing code
  • Pump shield
  • Piston rod opening

WARNING
Risk of suffocation
This product contains small parts that may be swallowed. Keep small parts away from small children and vulnerable persons.

Before You Get Started

Read these instructions for use and the instructions in the User’s Manual of the Accu‑Chek Solo micropump system. Take particular note of all warnings, information and notes required for the safe and proper use of the micropump system. This product is meant to be used by 1 person only and may be reused. You may use the pump base for up to 6 months. Replace the pump base at the latest after this period. If you use the pump base past this period, this may lead to inaccurate delivery of insulin.

Putting the Pump Base Into Operation

To put the pump base into operation, you need an Accu‑Chek Solo diabetes manager, an Accu‑Chek Solo reservoir filled with insulin and an Accu‑Chek Solo cannula assembly & pump holder.  To replace a component of the micropump system, always use the replace menu on the diabetes manager. To view an animated video of the handling steps, tap Help on the diabetes manager.

ACCU-CHEK-IUT526-Pump-Base-FIG2

  1. Tap the reservoir level on the diabetes manager status screen. Tap Pump base on the Replace system components screen. The reservoir is automatically selected as well. Tap Replace and follow the instructions on the screen.
  2. Remove the blue reservoir cap (A) from the pump base.
  3. Align the piston rod of the filled reservoir such that you can insert it into the piston rod opening (D) of the pump base. Push the filled reservoir onto the pump base until both parts are tightly connected.
  4. When both components are connected to each other correctly and the battery is activated, the micropump issues the Start sound. If this sound is not issued, check whether the battery is activated and repeat step 3.ACCU-CHEK-IUT526-Pump-Base-FIG4
  5. Make sure that there is no gap between the reservoir and thepump base. The pump base and the reservoir are only connected correctly when the Start sound is issued and when the pump base and the reservoir are seamlessly connected.
  6. Use and to set the number of insulin units with which you filled the reservoir. The set fill amount will be saved as the default setting for when the reservoir is filled the next time. Tap Save.
  7. Hold the diabetes manager close to the micropump to establish the radio connection. Tap Next. Wait a moment. Tap  Scan pairing codePoint the camera of the diabetes manager at the pairing code (B) on the pump base. Hold the diabetes manager in such a way that the pairing code can be read in full. A sound is issued if the pairing code was detected. If no sound is issued, the pairing code was not detected. Tap  to return to the Detect micropump screen. Tap Enter pumpkey. You find the pump key in the envelope in the bottom drawer of the micropump system packaging or on the inside of the micropump packaging lid. If there are several micropumps within range, tap the pump serial number of your micropump on the Select pump serial number screen. You find the pump serial number on the label of the pump shield (C) and on the packaging next to the symbol.
  8. The micropump is now ready to fill the reservoir needle.ACCU-CHEK-IUT526-Pump-Base-FIG4
  9. Hold the micropump in a position so that the reservoir needle is above the grey cannula head of the pump holder.
  10. Place the reservoir needle of the micropump on the grey cannula head of the pump holder.
  11. Turn the micropump carefully in place. You may apply light pressure on the pump shield to make sure the reservoir needle stays inside the cannula head of the pump holder.
  12. After one eighth turn (approx. 45 degrees), the micropump clicks into the front and rear hooks on the pump holder.
  13. Examine the front and back hooks of the pump holder to make sure the micropump is properly clicked in place.

Cleaning and Disinfecting

For more details, see chapter Care and Maintenance in the User’s Manual of the micropump system.
Disposal
Dispose of used products according to local regulations. For more information about correct disposal, contact your local council or authority.
Distributed in the United Kingdom by:
Roche Diabetes Care Limited Charles Avenue, Burgess Hill West Sussex, RH15 9RY, United Kingdom Accu‑Chek Pump Careline 1): UK Freephone number: 0800 731 22 91 ROI Freephone number: 1 800 88 23 51 1) calls may be recorded for training purposes Some mobile operators may charge for calls to these numbers. burgesshill.insulinpumps@roche.com www.Accu -chek.co.uk, www.accu