LINAK CAL40 Control Box for Patient Lifts Instructions

June 6, 2024
LINAK

Use 4 pcs. M10 8.8 screws, in each end, for mounting to the application. The screw depth must be min. 20 mm and max. 30 mm in aluminum profile.

Screw torque: 35 Nm.

LC3 2-Stage

15 mm distance to threadLINAK CAL40 - Dimensions 1 3 mm distance to thread

Screw depth min. 20 mm

Mounting holes, top                       Mounting holes, bottom

LINAK CAL40 - Mounting holes 1 LINAK CAL40 - Mounting holes 2

Drawing no.: 1002W9008

Notice that the cable plug must be mounted correctly. The cable slot must fit into the socket.

Remember to secure the cable mounted in the top of the column to the application, so that it cannot be pulled out of the column.
We recommend to use LINAK Cable:

  • Lock kit for minifit cable: 1002W8136-A.
  • Lock kit for hand control cable through: 1002W8137-A

Use only the screws included in the kit. Screw torque: 2 Nm

The cables coming out of the side of the column should follow below guidelines. The internal radius should not be more than 3 times the outer dimension ­ OD – of the cable. For instance if the outer cable dimension is Ø7, the internal radius of the maximum cable bending is 21 mm.

  • LC3 3-stage – protective grounding cable
    LC3 has potential equalization between top and bottom plate but the middle profile is not grounded.

  • It is recommended to use screws with thread-lock adhesive

  • Screws of high quality steel 8.8 or 10.9 must be used to secure safe mounting of the LC3 to the application.

Cable connections overview
3-stage is used as examples but variants are also applicable for 2-stage. This overview shows all possible cable connections, but please notice that some variants are upon request.

Variant| Top plate
Drawing no.:1002W9005| Side entry
Drawing no.:1002W9005| Model view
---|---|---|---
T000

Connections

| LINAK CAL40 - T000 - 15: Motor| LINAK CAL40 - T000 -
2| LINAK CAL40 - T000 - 3
T001

Connections

| LINAK CAL40 - T001 - 1 5: Motor| LINAK CAL40 - T001 -
2| LINAK CAL40 - T001 - 3

Including protective grounding cable

T100

Connections

| LINAK CAL40 - T100 - 1

5: Motor
6: Minifit through

| LINAK CAL40 - T100 - 2

3: Minifit through

| LINAK CAL40 - T100 - 3
T200

Connections

| LINAK CAL40 - T200 - 1

5: Motor
6: HB through

| LINAK CAL40 - T200 - 2

3: HB through

| LINAK CAL40 - T200 - 3
T030

Connections

| LINAK CAL40 - T030 - 1

5: Motor
6: Mains through 1

Female connection top entry, male connection side entry

| LINAK CAL40 - T030 - 2

3: Mains through 1

Female connection top entry, male connection side entry

| LINAK CAL40 - T030 - 3
T040

Connections

| LINAK CAL40 - T040 - 1

5: Motor
6: Mains through 2

Male connection top entry, female connection side entry

| LINAK CAL40 - T040 - 2

3: Mains through 2

Male connection top entry, female connection side entry

| LINAK CAL40 - T040 - 3
T101

Connections

| LINAK CAL40 - T101 - 1

5: Motor
6: Minifit through

| LINAK CAL40 - T101 - 2

3: Minifit through

| Including protective grounding cable
T201

Connections

| LINAK CAL40 - T201 - 1

5: Motor
6: HB through

| LINAK CAL40 - T201 - 2

3: HB through

| LINAK CAL40 - T201 - 3Including protective grounding cable
T031

Connections

| LINAK CAL40 - T031 - 1

5: Motor
6: Mains through 1

Female connection top entry, male connection side entry

| LINAK CAL40 - T031 - 2

3: Mains through 1

Female connection top entry, male connection side entry

| LINAK CAL40 - T031 - 3Including protective grounding cable
T041

Connections

| LINAK CAL40 - T041 - 1

5: Motor
6: Mains through 2

Male connection top entry, female connection side entry

| LINAK CAL40 - T041 - 2

3: Mains through 2

Male connection top entry, female connection side entry

| LINAK CAL40 - T041 - 3Including protective grounding cable
T110

Connections

| LINAK CAL40 - T110 - 1

5: Motor
6: Minifit through (top)
7: Minifit through (top)

| LINAK CAL40 - T110 - 2

3: Minifit through (side)
4: Minifit through (side)

| LINAK CAL40 - T110 - 3
T120

Connections

| LINAK CAL40 - T120 - 1

5: Motor
6: HB through
7: Minifit through

| LINAK CAL40 - T120 - 2

3: HB through
4: Minifit through

| LINAK CAL40 - T120 - 3
T220

Connections

| LINAK CAL40 - T220 - 1

5: Motor
6: HB through (top) 7: HB through (top)

| LINAK CAL40 - T220 - 2

3: HB through (side)
4: HB through (side)

| LINAK CAL40 - T220 - 3
T130

Connections

| LINAK CAL40 - T130 - 1

5: Motor
6: Mains through 1

Female connection top entry, male connection side entry
7: Minifit through

| LINAK CAL40 - T130 - 2

3: Mains through 1

Female connection top entry, male connection side entry
4: Minifit through

| LINAK CAL40 - T130 - 3
T140

Connections

| LINAK CAL40 - T140 - 1

5: Motor
6: Mains through 2

Male connection top entry, female connection side entry

7: Minifit through

| LINAK CAL40 - T140 - 2

3: Mains through 2

Male connection top entry, female connection side entry

4: Minifit through

| LINAK CAL40 - T140 - 3
T230

Connections

| LINAK CAL40 - T230 - 1

5: Motor
6: Mains through 1

Female connection top entry, male connection side entry

7: HB through

| LINAK CAL40 - T230 - 2

3: Mains through 1

Female connection top entry, male connection side entry

4: HB through

| LINAK CAL40 - T230 - 3
T240

Connections

| LINAK CAL40 - T240 - 1

5: Motor
6: Mains through 2

Male connection top entry, female connection side entry

7: HB through

| LINAK CAL40 - T240 - 2

3: Mains through 2

Male connection top entry, female connection side entry

4: HB through

| LINAK CAL40 - T240 - 3
S300

Connections

| LINAK CAL40 - S300 - 1| LINAK CAL40 - S300 -
23: Motor supply| LINAK CAL40 - S300 -
3
S400

Connections

| LINAK CAL40 - S400 - 1| LINAK CAL40 - S400 -
23: Motor supply| LINAK CAL40 - S400 -
3
S500

Connections

| LINAK CAL40 - S500 - 1| LINAK CAL40 - S500 -
2

3: Motor supply

| LINAK CAL40 - S500 - 3
S301

Connections

| LINAK CAL40 - S301 - 1| LINAK CAL40 - S301 -
23: Motor supply| LINAK CAL40 - S301 -
3Including protective grounding cable
S401

Connections

| LINAK CAL40 - S401 - 1| LINAK CAL40 - S401 -
23: Motor supply| LINAK CAL40 - S401 -
3Including protective grounding cable
S501

Connections

| LINAK CAL40 - S501 - 1| LINAK CAL40 - S501 -
2

3: Motor supply

| LINAK CAL40 - S501 - 3Including protective grounding cable
S310

Connections

| LINAK CAL40 - S310 - 1 6: Minifit through| LINAK CAL40 - S310 -
23: Motor supply
4: Minifit through| LINAK CAL40 - S310 - 3
S320

Connections

| LINAK CAL40 - S320 - 1 6: HB through| LINAK CAL40 - S320 -
2

3: Motor supply
4: HB through

| LINAK CAL40 - S320 - 3
S410

Connections

| LINAK CAL40 - S410 - 1 6: Minifit through| LINAK CAL40 - S410 -
2

3: Motor supply
4: Minifit through

| LINAK CAL40 - S410 - 3
S420

Connections

| LINAK CAL40 - S420 - 1 6: HB through| LINAK CAL40 - S420 -
2

3: Motor supply
4: HB through

| LINAK CAL40 - S420 - 3
S510

Connections

| LINAK CAL40 - S510 - 1 6: Minifit through| LINAK CAL40 - S510 -
2

3: Motor supply
4: Minifit through

| LINAK CAL40 - S510 - 3
S520

Connections

| LINAK CAL40 - S520 - 1

6: HB through

| LINAK CAL40 - S520 - 2

3: Motor supply
4: HB through

| LINAK CAL40 - S520 - 3
S330

Connections

| LINAK CAL40 - S330 - 1

6: Mains through 1

Female connection top entry, male connection side entry

| LINAK CAL40 - S330 - 2

3: Motor supply
4: Mains through 1

Female connection top entry, male connection side entry

| LINAK CAL40 - S330 - 3
S340

Connections

| LINAK CAL40 - S340 - 1

6: Mains through 2

Male connection top entry, female connection side entry

| LINAK CAL40 - S340 - 2

3: Motor supply
4: Mains through 2

Male connection top entry, female connection side entry

| LINAK CAL40 - S340 - 3