RESMED Mirage SoftGel Nasal CPAP Mask with Headgear User Guide
- June 6, 2024
- ResMed
Table of Contents
- RESMED Mirage SoftGel Nasal CPAP Mask with Headgear User Guide
- Mask components
- Disassembly
- Reassembly
- Cleaning your mask in the home
- Troubleshooting
- Technical specifications
- Mirage Activa™ LT NASAL MASK
- Troubleshooting
- Technical specifications
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
RESMED Mirage SoftGel Nasal CPAP Mask with Headgear User Guide
Mirage SoftGel User Guide
Fitting
Removal
Mask components
Item | Description | Part number |
---|---|---|
1 | Mask frame | 60180 (St) |
60193 (S)
2| Dial| 61289
3| Forehead support| 16393
4| Forehead support pad| 60123 (1)
60124 (10)
5| SoftGel Cushion & clip| 61631 (S)
61632 (M)
61633 (L)
61634 (LW)
6| Cushion clip| 60176 (St)
60197 (S)
7| Headgear clip| 16569 (2)
16734 (10)
8| Headgear| 16118 (S)
16117 (M)
16119 (L)
9| Ports cap| 16570 (2)
16571 (10)
10| Swivel| 16565
11| Elbow| 16387
12| Vent| –
A| Complete system| 61605 (S)
---|---|---
| | 61606 (M)
| | 61607 (L)
61608 (LW)
B| Elbow assembly| 16399
Also available
C| Acriva LT Cushion & clip| 60198 (S)
| | 60177 (M)
| | 60178 (L)
60179 (LW)
Disassembly
Reassembly
MirageTM SoftGel NASAL MASK
Thank you for choosing the Mirage SoftGel.
The Mirage SoftGel is the fi rst of ResMed’s masks offering a choice of two
high performance cushions (Mirage SoftGel and Mirage Activa LT) on the one
durable frame. For further information, refer to the Activa LT User
Information section in this User Guide.
Intended use
The Mirage SoftGel channels airfl ow noninvasively to a patient from a
positive airway pressure (PAP) device such as a continuous positive airway
pressure (CPAP) or bilevel system. The Mirage SoftGel is:
- to be used by adult patients (> 30 kg) for whom positive airway pressure has been prescribed
- intended for single-patient re-use in the home environment and multipatient re-use in the hospital/institutional environment.
Using your mask
When using your mask with ResMed CPAP or bilevel devices that have mask
setting options, refer to the Technical specifications section in this user
guide for mask selection options.
For a full list of compatible devices for this mask, see the Mask/Device
Compatibility List on www.resmed.com on the Products
page under Service & Support. If you do not have internet access, please
contact your ResMed representative.
Note: If you experience nasal dryness or irritation, use of a humidifier
is recommended.
WARNING
- The vent holes must be kept clear.
- The mask should not be used unless the CPAP or bilevel device is turned on and operating properly.
- Follow all precautions when using supplemental oxygen.
- Oxygen flow must be turned off when the CPAP or bilevel device is not operating, so that unused oxygen does not accumulate within the CPAP or bilevel device enclosure and create a risk of fire.
- At a fixed rate of supplemental oxygen fl ow, the inhaled oxygen concentration varies, depending on the pressure settings, patient breathing pattern, mask, point of application and leak rate.
- The technical specifications of the mask are provided for your clinician to check that they are compatible with the CPAP or bilevel device. If used outside specification or if used with incompatible devices, the seal and comfort of the mask may not be effective, optimum therapy may not be achieved, and leak, or variation in the rate of leak, may affect the CPAP or bilevel device function.
- Discontinue using this mask if you have ANY adverse reaction to the use of the mask, and consult your physician or sleep therapist.
- Using a mask may cause tooth, gum or jaw soreness or aggravate an existing dental condition. If symptoms occur, consult your physician or dentist.
- As with all masks, some rebreathing may occur at low CPAP pressures.
- Refer to your CPAP or bilevel device manual for details on settings and operational information.
- Remove all packaging before using the mask.
Cleaning your mask in the home
Your mask and headgear should only be handwashed by gently rubbing in warm (approximately 30°C) water using mild soap. All components should be rinsed well with drinking quality water and allowed to air dry out of direct sunlight.
Daily/After each use :
- To optimise the mask seal, facial oils should be removed from the cushion after use.
- Handwash the separated mask components (excluding headgear).
- If the vent requires cleaning use a soft bristle brush. Weekly:
- Handwash the headgear. It may be washed without being disassembled.
WARNING
Do not use aromatic-based solutions or scented oils (eg, eucalyptus or
essential oils), bleach, alcohol or products that smell strongly (eg, citrus)
to clean any of the mask components. Residual vapours from these solutions can
be inhaled if not rinsed thoroughly. They may also damage the mask, causing
cracks.
CAUTION
- If any visible deterioration of a system component is apparent (cracking, crazing, tears or cushion damage resulting in gel exposure, etc), the component should be discarded and replaced.
- Avoid connecting fl exible PVC products (eg, PVC tubing) directly to any part of the mask. Flexible PVC contains elements that can be detrimental to the materials of the mask, and may cause the components to crack or break.
Reprocessing the mask between patients
This mask should be reprocessed when used between patients. Cleaning,
disinfection and sterilization instructions are available from the ResMed
website,
www.resmed.com/masks/sterilization.
If you do not have internet access, please contact your ResMed representative.
Troubleshooting
Technical specifications
Mirage Activa™ LT NASAL MASK
Intended use
The Mirage Activa LT channels airfl ow noninvasively to a patient from a
positive airway pressure device such as a continuous positive airway pressure
(CPAP) or bilevel system.
The Mirage Activa LT is:
- to be used by adult patients (> 30 kg) for whom positive airway pressure has been prescribed
- intended for single-patient re-use in the home environment and multipatient re-use in the hospital/institutional environment.