AND TM2440-EX Ambulatory Blood Pressure Monitor Instruction Manual

June 6, 2024
AND

AND TM2440-EX Ambulatory Blood Pressure Monitor

Thank you for purchasing the Ambulatory Blood Pressure Monitor TM2440-EX. The Instruction Manual that accompanies the product has been corrected and had content added. Please read it in conjunction with the details shown below.

Automatic Blood Pressure Measurement (A-BPM)

The following changes have been made.

Wrong Correct
The initial pressurization value is set to 180 mmHg at the factory. The

initial pressurization value for automatic blood pressure measurement is AUTO.

In the case of AUTO, the appropriate pressurization value is set automatically.

Initial Pressurization Value

The following changes have been made.

Wrong Correct
The initial pressurization value is set to 180 mmHg at the The initial

pressurization value setting is AUTO
factory.| (appropriate pressurization value set automatically) when
 | the      mark is displayed and 180 mmHg when the
 | mark is hidden.

A-BPM Preset Programs

The following is added after “A-BPM can use only while the automated measurement can be performed. ”
Note: The Start time and Operation time are initially set calculating from the time at which the mark was displayed. When the next automatic measurement is made, this needs to be set again.

A-BPM Items and Parameters

The following changes have been made.

Wrong Correct

3: Example for automated measurement.

Start time :          Stores a time. (0 to 23 hour) Operation time : Set to “OFF”

Response :

A-BPM starts the blood pressure measurement at the preset Start time and continues until

A-BPM is suspended.

If the EVENT switch is pressing and holding until the mark is shown, A-BPM starts at the preset Start time.

| #3: If the Start time for automatic measurement is specified and the Operation time is set to “OFF” , long pressing the EVENT switch will start automatic blood pressure measurement from the Start time set and measurement will continue until the   EVENT   switch is long pressed. If the EVENT switch is long pressed again, automatic measurement will continue.

Note

If the Operation time is set, automatic blood pressure measurement will be performed from when the EVENT switch is initially long pressed until the Operation time has passed even if the EVENT switch is long pressed midway through automatic measurement.

4: Example for automated measurement.

Start time :  Set to “OFF” Operation time : Stores time to be continued. (1 to 27 hours)

Response :

A-BPM starts blood pressure measurement and stops after the Operation time.

If the mark is hidden by pressing and holding the EVENT switch during the Operation time , A-BPM stops.

If the mark is shown by pressing and holding the EVENT switch again, A-BPM is performed for the Operation time.

| #4: If the Start time for automatic measurement is set to “OFF” and the Operation time is set, long pressing the EVENT switch will implement automatic blood pressure measurement after initial automatic measurement until the set Operation time has passed. If the EVENT switch is long pressed midway through automatic blood pressure measurement, automatic measurement will stop. If the EVENT switch is long pressed again, automatic measurement will be performed until the set Operation time has passed.

A-BPM Program Examples

The following notes were added to “Example 2 Automatic measurement”
Addition: NB. Automatic blood pressure measurement will continue even if turned “OFF” and “ON” again during automatic measurement.

The following notes were added to “Example 3 Automatic measurement”
Addition: NB. Automatic blood pressure measurement will not be performed once the Operation time has been exceeded even if turned “OFF” and “ON” again during automatic measurement.