RESMED S9 Series Escape Auto Positive AirWay Pressure Device
Welcome
Thank you for choosing the S9 Series or H5i. Before operating these
devices, please read the entire Welcome and Information Guides.
S9 at a glance
The S9 system comprises the following elements:
- S9 device SlimLine air tubing 90W power supply unit S9 Travel bag SD card
- S9 SD card protective folder. Optional components include:
- H5i heated humidifier Standard air tubing 30W power supply unit (does not support H5i)
- Power Station II battery pack DC/DC Converter 24V/90W.
Traveling tips
When traveling with your S9 only:
- Ensure you pack the Slim Line or Standard air tubing as the Climate Line or Climate Line MAX heated air tubing is not designed to connect directly to the S9 device.
- Please ensure you purchase and travel with the approved power cord for the region where you will be using the S9 device.
H5i at a glance
The H5i system comprises the following elements:
- H5i heated humidifier H5i standard water tub. Optional components include:
- ClimateLine heated air tubing ClimateLineMAX heated air tubing H5i cleanable water tub.
Traveling tips
- When moving or traveling with your H5i:
- Ensure that the water tub is empty.
- Disconnect the H5i from the S9 by using the release button.
Setup
- Align your H5i with your S9 and push them together until they click into place.
- Connect the DC plug of the power supply unit to the rear of the S9.
- Connect the power cord to the power supply unit.
- Plug the other end of the power cord into the power outlet.
- Connect one end of the air tubing firmly onto the air outlet.
- Connect the assembled mask system to the free end of the air tubing.
Notes:
- Place the power supply unit away from the H5i to allow for adequate ventilation.
- For more information on assembling your mask see your mask user guide.
- Recommended masks are available on www.resmed.com on the Products page under Service & Support.
Filling the water tub
- Slide the latch and lift open the flip lid.
- Remove the water tub.
- Fill the water tub (through the center hole) with distilled or deionized water up to the maximum water level mark (380 mL).
- Return the water tub to the H5i.
- Close the flip lid ensuring that it clicks into place.
S9 Essentials
S9 Essentials is designed to make device interaction and menu navigation
easier for you. If enabled by the clinician, S9 Essentials disables the Info
and Setup functionality so that you can simply start and stop therapy and
adjust ramp, humidification and Climate Control.
Adjusting humidity level
Ranging from OFF to 6, you can adjust the humidity level at any time to find
the setting that is most comfortable for you. To adjust your level of
humidity:
Warming the H5i
The warm-up feature is used to pre-heat the water before starting treatment.
To start warming:
- Navigate to the humidity level icon.
- Push and hold icon for three seconds. The Warming Up status bar is displayed.
- You can stop warm-up at any time by pressing icon for at least three seconds. You can start treatment at any time by pressing
Note: In order to assist the heater plate in cooling, your S9 device will continue to blow air for up to an hour after treatment has stopped. However, you can unplug the device from the power outlet at any time and allow the heater plate to cool without air flow.
Adjusting ramp time
Designed to make the beginning of treatment more comfortable, ramp time is the
period during which the pressure increases from a low start pressure to the
treatment pressure. To adjust your ramp time:
Using mask-fit
Available in S9 AutoSet and S9 Elite, mask-fit helps you fit your mask properly. This feature delivers treatment pressure for a three-minute period, prior to starting treatment, during which you can check and adjust your mask- fit to minimize leaks.
-
Fit your mask as described in the mask user guide.
-
From your HOME screen, push and hold icon for three seconds.
One of the following MASK FIT screens is displayed: -
If necessary, adjust your mask, mask cushion and headgear until you have a Good mask-fit. After three minutes, the pressure reverts to the set pressure and treatment will begin. You can end mask-fit at any time by pressing .
**Selecting the mask type***
To select your type of mask:
Note: Mask Fit is only available in S9 AutoSet and S9 Elite.
- Selecting the mask type is not possible if S9 Essentials has been enabled by your clinician.
Getting started
- 1. Make sure the power is connected.
- Adjust the ramp time or humidification level if required.
- Fit your mask as described in your mask user guide.
- To start therapy, simply breathe into the mask and/or press
- Lie down and arrange the air tubing so that it is free to move if you turn in your sleep.
- To stop treatment at any time, remove your mask and/or press
Notes:
- SmartStart is only available in S9 AutoSet and S9 Elite.
- If your clinician has enabled SmartStart your device will start automatically when you breathe into the mask and stop automatically when you remove your mask.
- Once therapy has started a treatment screen is displayed.
- If power is interrupted during treatment, the device automatically restarts therapy when power is restored.
Cleaning and maintenance
You should regularly carry out cleaning and maintenance as described in this section.
- Slide the latch.
- Lift open the flip lid.
- Remove the water tub.
- Discard any excess water from the water tub.
- Unclip all four side latches.
- Pull apart the tub lid and base.
Daily cleaning
-
Remove the air tubing by pulling off the finger grips on the cuff. Hang it in a clean, dry place until next use.
-
Wash the disassembled tub lid and base in warm water using a mild detergent.
-
Rinse thoroughly in clean water and allow them to dry away from direct sunlight.
Notes:- Do not hang the air tubing in direct sunlight as it may harden over time and eventually crack.
- Do not wash the air tubing in a washing machine or dishwasher.
Weekly
- Remove the air tubing from the S9 device and the mask.
- Wash the air tubing in warm water using mild detergent.
- Rinse thoroughly, hang, and allow to dry.
- Before next use, reconnect the air tubing to the air outlet and mask.
Monthly
- Wipe the exterior of the S9 and H5i with a damp cloth and mild detergent.
- Check the air filter for holes and blockage by dirt or dust. Replace the air filter if necessary.
- Peel the flip lid seal from the flip lid and wash it in warm water using a mild detergent.
Maintenance checklist
- Inspect the water tub and flip lid seal for wear and deterioration.
- Replace the water tub if any component is leaking or has become cracked, cloudy or pitted.
- Replace the flip lid seal if cracked or torn.
- Clean white powder deposits in the water tub by using a solution of one part household vinegar to 10
parts water.
Reassembling and filling the water tub
- Place the tub lid back onto the base.
- Clip all four side latches.
- Fill the water tub with distilled or deionized water up to the maximum water level mark.
- Return the water tub to the H5i.
- Close the flip lid ensuring that it clicks into place.
Replacing the air filter
Replace the air filter every six months (or more often if necessary).
-
Remove the air filter cover from the back of the S9 device.
-
Remove and discard the old air filter.
-
Insert a new ResMed air filter ensuring that it is sitting flat in the air filter cover.
-
Replace the air filter cover.
Notes:
- Ensure the air filter and air filter cover are fitted at all times.
- Do not wash the air filter. The air filter is not washable or reusable.
SD card
An SD card has been supplied to gather therapy data from your S9 device and
provide settings updates from your clinician. When instructed to do so,
disconnect the S9 device from the power outlet, remove your SD card, insert it
in the protective folder and send it to your clinician.
Removing the card
- Push in the SD card to release it.
- Remove the card.
- Insert the card into the protective folder.
- Send the protective folder back to your clinician as instructed.
For more information on removing and inserting your card refer to your S9 SD Card Protective Folder provided with your device. Please retain the S9 SD Card Protective Folder for future use.
Notes:
- For Troubleshooting information refer to the Information Guide.
- For more product information see www.resmed.com.
Manufacturer:
ResMed Pty Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153
Australia. See www.resmed.com for other ResMed
locations worldwide.
For patent information, see www.resmed.com/ip.
S9, Auto Set, S9 Auto Set, S9 Escape, S9 Elite, S9 Escape Auto, H5i, Slim Line
and Climate Line are trademarks of ResMed Pty Ltd and S9, Auto Set, Escape,
Slim Line and Climate Line are registered in U.S. Patent and Trademark Office.
© 2017 ResMed.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>