KERN ORA 3AA-AB Bee Analoges Refractometer Instruction Manual

June 6, 2024
KERN

KERN LOGO

KERN ORA 3AA-AB Bee Analoges Refractometer

KERN ORA 3AA-AB Bee Analoges
Refractometer PRODUCT

Technical data

KERN ORA 3AA-AB Bee Analoges Refractometer FIG-0

Example scale ORA 3AA/ABKERN ORA 3AA-AB Bee Analoges Refractometer
FIG-1

Example scale ORA 4AA/ABKERN ORA 3AA-AB Bee Analoges Refractometer
FIG-2

Description

KERN ORA 3AA-AB Bee Analoges Refractometer FIG-3

  1. Prism cover
  2. Prism surface
  3. Adjustment screw
  4. Eyepiece with rubber eyeshade
  5. Diopter adjustment ring
  6. Optical tubes with rubber grip
  7. Storage box
  8. Pipette
  9. Refractometer
  10. Calibration liquid (Distilled water)
  11. Adjustment tool
  12.  Cleaning cloth

General information

Intended use
The refractometer is a measuring instrument for determining the refractive index of transparent substances in the liquid or the solid state. It is used to observe the behaviour of light as it passes from a prism with known properties to the substance being tested. Use of the refractometer for other purposes is contrary to its intended use and may be hazardous. The manufacturer shall not be liable for any damages caused by improper use.

 Warranty

The warranty shall be void in the event of:

  • Failure to observe the instructions in the operating manual
  • Use for purposes other than those described
  • Modifications or opening the device housing
  • Mechanical damage and/or damage resulting from media, liquids, natural wear and tear

Basic safety information

Follow the instructions in the operating manual

  • Carefully read through the operating manual even if you have prior experience with KERN refractometers.
  • Every language version includes a non-authoritative translation. The original German document is the definitive version.

Warning

  • Do not let acids come into contact with skin or eyes. If acid comes into contact with skin, flush with copious amounts of water. Shower if larger areas of skin are affected.
  • If acid comes into contact with eyes, keep the eyelid open and flush the eye with running lukewarm water from the outer corner to the inner corner. Flush eyes for at least 15 minutes. Then consult a doctor or ophthalmologist immediately.
  • Thoroughly clean the refractometer after each use.
  • The refractometer must not be exposed to extreme temperatures, high mechanical stresses, strong direct sunlight or high humidity.
  • This refractometer is not a toy. Keep out of reach of children.
  • Make sure that you will not be hit by anything else while you are using the refractometer, as this could cause serious eye injuries
  • The rubber eyeshade may cause irritation when in prolonged contact with the skin. If this happens, consult your doctor.
  • Do not touch the lenses with your fingers.

Supplied items

After unpacking and before using the device for the first time, check that all listed parts have been supplied. Replace damaged or faulty parts immediately and do not put them into operation.

  • Refractometer
  • Storage box
  • Pipette
  • Adjustment tool
  • Cleaning cloth
  • Calibration liquid (Distilled water)

Before the first use

Remove the protective film (if present) from the prism surface [2] and check that the rubber eye-cup [4] is fitted correctly.

Use/measurement

The refractometer can be used to quickly and accurately determine the refractive index of transparent substances, liquid or solid ones. To ensure correct measure-ment, the measuring device should be adjusted before measurements are carried out. Please make sure your hands are dry before handling the measuring device.

Zero point calibration

  • Hold up the refractometer against a sufficiently bright light source and look through the eyepiece [5], holding the rubber eyeshade [4] closely against your eye/glasses.
  • Rotate the eyepiece [5] to adjust it for your eyesight, until you can see the scale sharply.
  • Open the prism cover [1].
  • Thoroughly clean the prism [2] and the underside of the prism cover [1] using a soft cloth or soft paper (with alcohol if necessary) and wipe dry.
  • Now apply a few drops of the calibration liquid [10] onto the prism surface [2].
  • Close the prism cover [1]. The amount of fluid should be sufficient to moisten most of the prism surface. There should not be any air bubbles between the measuring prism [2] and prism cover [1].
  • Wait about 30 seconds to allow the temperatures of the fluid and prism to equalise.
  • Look through the eyepiece [4] while pointing the refractometer‘s prism surface [2] at a bright light source.
  • Through the eyepiece [4], you will see a bright and a blue field. The boundary line between them shows the measured value on the scale which is also visible through the eyepiece [4].
  • Use the supplied adjustment tool [11] to turn the adjustment screw [3] behind the prism surface [2] (under the rubber cap), and adjust the scale so that the measuring device is optimally set up by moving the boundary line upwards or downwards. Calibration value ORA 3AA/3AB: 0 % (Brix) / 1.000 (SG wort) Calibration value ORA 4AA/4AB: 0 °P (Plato)
  • Repeat step 4 (cleaning).

Important!
The ambient/room temperature and the sample temperature influence the refractometer measuring result. The scales of the refractometer models, which have “AB” in its name, are designed for an ambient temperature of +20 °C! If the measurements are carried out at a temperature other than +20 °C, the results must be correspondingly corrected. A correction table can be found in the annex, Point 14. Refractometer models, which have “AA” in its name, are equipped with an automatic temperature compensation (ATC). Between +10°C and +30°C measurement differences due to temperature changes are automatically compensated.

Cleaning and maintenance

Clean the refractometer using a soft, lint-free cloth moistened with either water or, if necessary, alcohol. Do not use any aggressive or abrasive cleaning agents. Never immerse the device in water or hold it under running water. Never handle the device with wet or damp hands. Never touch the measuring prism [2] with hard tools made from plastic, wood, rubber, metal, glass etc. Hard objects can quickly damage the relatively soft prism glass, resulting in measurement errors. The refractometer is maintenance-free. Cleaning should be carried out immediately before and after each use of the refrac-tometer to maximise its life and optimise measurement results.

Storage

Store the refractometer in a dry, non-corrosive environment, preferably between 10 °C and 30 °C.

Service

After reading this operating manual, if you have any questions about setting up or using the refractometer, or if any unexpected problem occurs, please contact your dealer. The device housing may only be opened by trained service technicians authorised by KERN.

Disposal

The packaging consists of environmentally friendly materials which can be disposed of via local recycling facilities.
The device and storage box should be disposed of by the operator in accordance with applicable national or regional regulations at the place of use.

Additional information

The product may differ slightly from the illustrations. Avoid exposing the refracto-meter to direct sunlight. Never bring the refractometer into contact with solvents.

Brix to refractive index (nD) conversion table

Data from „ICUMSA“ International Commission for Uniform Methods of Sugar Analysis, at 20 °C and 589 nm wavelength.

KERN ORA 3AA-AB Bee Analoges Refractometer FIG-4

Annex

Table 1: International Temperature Correction Table for °Brix (% sugar gradient) Correct the result by the following values (refractometer must be correctly adjusted at 20

KERN ORA 3AA-AB Bee Analoges Refractometer FIG-5