TE PARI TP V20 Revolution Smart Dosing Gun User Manual
- June 6, 2024
- TE PARI
Table of Contents
- Included with your Revolution Dosing Gun
- Included with your EasyDose Gun
- Overview of the control panel
- How to change the barrel
- To Calibrate the Gun if giving incorrect dose
- EasyDose Gun Set-up and Operation
- Revolution Gun Scale Connection Set-up
- Care, Maintenance and Storage
- Charging the batteries
- Te Pari Products Warranty
- Warranty Claim Form Online
- Health and Safety Warnings
- Dosing Gun Cautions
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
REVOLUTION DOSING GUN
Operational Manual for EasyDose and Revolution Models
Te Pari Dosing Guns
Congratulations on your purchase of a Te Pari Dosing Gun. This digital, self-
powered dosing gun allows you to deliver fully calibrated dosages with both
accuracy and speed.
Difference between the EasyDose and Revolution models
There are two Te Pari Dosing Gun Models – the EasyDose model and the
Revolution Smart model. Both the EasyDose and the Revolution allow for dosages
to be set manually using the digital keypad. The Revolution can also connect
to the Te Pari Scale System via Wi-Fi to automatically calculate and set the
dose rate based on the weight of the animal. Items and features only available
with the Revolution will be specified throughout this manual.
Included with your Revolution Dosing Gun
|
---|---
A. Feed tube and two anti-kink coils
B. Intellicharger – auto-sensing battery charger
C. 12-volt power supply – for the battery charger
D. Mains power supply – for the battery charger| E. Spares kit with 1x piston
O-ring, inlet, and outlet valve assemblies, and valve retainer C-clip for
outlet valve.
F. Two high-quality 18650 lithium-ion batteries
G. Blow moulded case for secure storage
H. T Bar back liner & oral nozzle
---|---
Included with your EasyDose Gun
A. Feed tube and two anti-kink coils
B. Intellicharger – auto-sensing battery charger
C. 12-volt power supply – for the battery charger| D. Mains power supply – for
the battery charger
E. Spares kit with 1x piston O-ring, inlet and outlet valve assemblies, and
valve retainer C-clip for outlet valve.
F. Two high-quality 18650 lithium-ion batteries
---|---
What is not pictured here is stored underneath the box layer
Overview of the control panel
Main Screen
The current dose is prominently displayed at the top of the main screen. The
top-left corner shows a Wi-Fi icon when it is active, and the top-right
corner shows the battery level.
| Power button
Click once to switch on.
Press and hold to switch off
---|---
| Prime button
Click to enter priming mode
| Menu button
Click through the options to access options for dose speed, refill speed, Wi-
Fi (Smart only), sound and tick confirmations and the barrel change sub-menu.
| Statistics button
Shows how many doses have been administered and how muchproduct has been used.
| Up and Down buttons
Used to increase or decrease values and change a selection when displayed.
Most setting changes are made using these buttons.
| Power button
Click once to switch on. The Te Pari logo, followed by the gun model and 2
brief safety warnings will be displayed before the main screen appears. Press
and hold momentarily to switch the gun off.
---|---
| | |
| Prime button
Click the prime button once. The word “Prime” will be displayed to show that
the gun is in Prime mode.
---|---
| | Once the gun is in prime mode, pull and hold the trigger. The gun will
dispense a full barrel load and then return to the retracted position and
stop. Release, then pull and hold the trigger again to dispense another full
barrel. Continue until fully primed.
mPrime allows for incremental adjustments to remove the last of the air in the
barrel.
Holding prime button for 10seconds will lock/unlock the dosing gun.
The trigger may be released to stop priming at any time during the cycle.
| Up and Down buttons
These buttons are used to increase or decrease values or to change a selection
when displayed. Most setting changes are made using these buttons.
---|---
| | | Changing the dose amount
Clicking the up button once increases the dose by 1 ml (70 ml barrel) or 0.5
ml (14 ml barrel). Pressing and holding it scrolls the amount up quickly. The
down button works in the same way
| | | Changing a selection
Pressing either the up or down arrow will change the selection for the current
item. The selected
option is shown with a box around it
| Menu button
Each time the menu button is clicked, it cycles through the menu for access to
the following items: dosing speed; Refill speed, Wi-Fi (G20 only), Sound, Tick
Timer, and Barrel Change Sequence
---|---
| Refill Speed
With the screen displaying the refill speed icon (arrows pointing right),
press the up arrow to increase the refill speed or the down arrow to decrease
the speed. The arrows will change to show the selected speed. This example
shows maximum speed.
Slower speeds are for more viscous liquids and faster speeds for less viscous
liquids.
| Dosing speed
With the screen displaying the dosing speed icon (arrows pointing left), press
the up arrow to increase the dosing speed or the down arrow to decrease
speed. The arrows will change to show the selected speed.
Slower speeds are best for more viscous liquids and faster speeds for less
viscous liquids
| Wi-Fi (Smart model only)
With the screen displaying the Wi-Fi icon, press the up or down arrow and then
press the menu button to enable or disable Wi-Fi. When active the Wi-Fi icon
will appear in the top left of the main screen. The number of filled bars show
the signal strength. If no connection is available no bars will be displayed.
| Sound
The gun can issue an audible double beep to confirm dose delivery, and a
single beep to notify of a dose amount received via Wi-Fi. With the screen
displaying the sound icon, press the up or down arrow to turn the sound on or
off.
| Tick Timer
The Tick Timer shows a tick mark on the screen when a dose has been delivered.
With the screen displaying the Tick Timer icon, select the length of time, in
seconds, how long the tick should stay on screen. “0” is off and no tick will
be displayed. This avoids any accidental dosing immediately after applying
product to an animal.
How to change the barrel
| Barrel Change Sequence
The barrel change sequence is the last item accessed from the menu button. You
use it to change the barrel from the standard 70ml barrel to the smaller 6ml
one. When you select the tick and press the menu button on this item you enter
the barrel change sub menu escribed below. Pressing and holding the Menu
Button at any time returns you to the main screen.
---|---
| 1. Remove the barrel The arrow flashes on the screen. Remove the barrel
collar and remove the barrel. Press the menu button for the next step| | 4.
Select the barrel size you are fitting Use the arrow keys to select the size
of the barrel being installed, then press the menu button to proceed to the
next item. To avoid damage, it is important to ensure that the correct setting
is selected for the barrel and piston size installed
| 2. Remove the piston The arrows flash on
the screen. Remove the piston by holding down the trigger of the gun and
gently pulling it out. Press the menu button for the next step.| | 5. Fit the
barrel Fit the new barrel to the gun with the collar. Press the menu button to
complete the sequence.
| 3. Load the piston for the smaller barrel The arrows flash on the screen.
Load the piston by holding down the trigger
of the gun and introduce the piston to the drive thread. The piston stops in
the “Home” position and the screen automatically advances to the next step.| |
6. Barrel change complete The screen displays a gun icon with a tick to show
the process is compl ete. The gun saves the changes and automatically returns
to the main screen. Note: When using the 14 ml barrel, dose increments change
to 0.5 ml
Barrel size, dose amount and delivery specification
70 ml Barrel:
Doses of 5 to 300 ml. Dose-volume increments of 1 ml. Volumes from 71 to 300
ml delivered in up to five sequential shots.
14 ml Barrel:
Doses of 1 to 28 ml. Dose-volume increments of 0.5 ml. Volumes from 14.5 to 28
ml delivered in two sequential shots.
6 ml Barrel:
Doses of 1ml to 12ml. Dose volume increments of 0.1ml. Volumes from 6ml to
12ml delivered in two sequential shots.
S equential shots:
These are delivered automatically and immediately after the first shot with
just the one trigger click until complete dose has been administered.
| Statistics button
The statistics button shows the volume of dose administered, the total number
of doses administered and the total number of doses administered. Each press
scrolls through the various items.
---|---
| Volume of administered dose
Shows the total volume of dose administered.
| Number of doses
Shows the total number of doses administered.
| Clear stored statistics
With the screen displaying the clear icon, press the up or down arrow to
select the tick or
the cross. With the tick highlighted, press the menu button. The display will
return to the
main menu, and the number of doses and volume statistics will be reset to 0.
To Calibrate the Gun if giving incorrect dose
| 1. Press the stats and the power buttons
simultaneously| | 3. Using up and down arrows, choose percentage(%) or the ml
as required.
---|---|---|---
| 4. Press the menu button
again and use the up and
down to set the % amount
or the ml. To save this,
press menu again.
| 2. A screen with “Info, Calibration, Features”
will appear. Scroll to “Calibration” and press menu.| | 5. Power cycle the
gun and resume dosing.
Preparing the gun for use
- onnect the feed tube from the gun to the fluid container. See instructions below• Set the gun to prime.
- Press and hold the trigger until fluid fills the barrel.
- Point the tip of the gun vertically to remove the last of the air from the barrel.
- Press the prime button to return to the main screen.
- The gun is now ready for use.
Warning: Exercise caution when dispensing fluids.
Fitting the feed tube
A. Unclip the retainer and fit the feed tube with anti-kink coil through it.
B. Press the tube and coil snugly into the retainer recess, with the
retainer’s flared end pointing backwards.
C. Refit the clip with the hose and coil fitting over the attachment.
D. Ensure the valve housing knurled ring is snug and has not loosened while
fitting the tube and coil.
EasyDose Gun Set-up and Operation
Dose on weight
| Set Up
1. Dose units ml/kg cc/lb Power gun on
2. Press menu button and select weight with up/down keys
---|---
| 3. Select dose units with up/down keys, press menu to continue
| 4. Set mL/kg or cc/ib using up/down keys
5. Save settings using the menu button
| Operation
1. Adjust weight using up/down keys to have dosage adjusted.
Warning: Exercise caution when dispensing fluids.
Dose on volume
Setup & Operation
| 1. Press menu button to switch mode, select volume using p/down keys
2. Save by pressing menu button seven times through options
---|---
| 3. Adjust volume with up/down arrows
Revolution Gun Scale Connection Set-up
Connecting to Te Pari T30 Scale System
- Power gun on > press menu button four times till ‘edit Wi-Fi’ screen appears.
- Using arrow keys to select ‘edit Wi-Fi’ and press the menu button again.
- Use arrow keys to scroll through Wi-Fi options & select ‘Te Pari T30’
- Press menu button three times more until settings are saved.
- Power on T30 Scale and the gun will automatically connect with the scale.
Set Te Pari Scale Indicator use with Te Pari Dosing Gun
-
When you turn on the Te Pari scale indicator, you will see a welcome screen and then the file list. Press “Set Up Weight’ to navigate to Main Menu
-
Scroll to DOSING using Arrows Press Record/Select
-
Dose 1 set to On using > (press once to turn on and press again to turn off )
-
Name – Key in the drench Name using keypad
-
Type – Select calculated does or fixed dose using (press once to select calculated dose and again to select fixed dose) Rate – Set using numerals and > (xxml/xxxxkg = ml Per Kg)
-
Using arrow keys scroll to save and exit Press record/select This will return you to your main screen and your scale is set up to communicate with your Te Pari Dosing Gun
Dosing Gun Set Up -
Turn gun on
-
Press menu Button three times and you will find Wifi screen
-
Make sure WiFi is turned on using Up key
Gun will automatically return to the dosing screen after a few seconds and if you are connected it will show in the top left hand of the screen.
Connecting to a Tru-test XR5000 Scale & Te Pari Dosing Gun App
-
Power Dosing Gun & Tru-Test Scales on
-
Open the Te Pari Dosing Gun App on your device
-
On the Dosing Gun, press Menu button four times to get Wi-Fi settings, use arrow keys to select ‘Edit Wi-Fi’ tick and press the menu button again to enter.
-
Use the directional arrows on the keypad to change Wi-Fi Mode to ‘Te Pari App’
-
Press Menu button and change DG SSID to ‘TePari’.
-
Press Menu button and change WiFi Security to ‘X’.
-
Press Menu button once more and the gun will prompt you to power off and back on to save the changes.
-
Open ‘Wi-Fi settings’ on device, then select Te Pari from the list. You may be prompted saying this device does not have internet connection.
-
On Tru-Test home screen press ‘Settings’, then ‘Next Page’, then ‘Connections’
Connecting to Feedlot Software
Please refer to software providers website or contact a Te Pari rep. -
On phone, open Bluetooth and press ‘Pair/Add New Device’ and select ‘ID5000’ or ‘XR5000’ (This can sometimes take a few moments to appear)
-
You will then be prompted for a password, this is either ‘0000’ or ‘1234’
-
You Wi-Fi and Bluetooth connections have now been established with each device, now open the Dosing Gun app on your device.
-
In the Dosing Gun app, start a new session, enter the session details and press ‘Save & Start’
-
Press ‘Connect Device’ and reconnect to each device (Wi-Fi = Te Pari Dosing Gun & Bluetooth = XR5000) Once correctly connected, each device will show as ‘Connected’
-
On the scale, start a new weigh session and test the dosing gun by putting weight on scales. Check that the Dosing Gun is calibrating to the correct amount of drench to weight. You are now ready to start your weigh session
Connecting a Gallagher TW series scale to a Te Pari Dosing Gun.
-
Power gun on > press menu button four times till ‘edit Wi-Fi’ screen appears.
-
Using arrow keys to select ‘edit Wi-Fi’ and press the menu button again.
-
Use arrow keys to scroll through Wi-Fi options & select ‘Gallagher TW’
-
Press menu button and adjust/choose the ‘DG SSID’.
SSID is the Wi-Fi name which will be displayed on the connecting device (Gallagher Scale) -
Press menu button again to save then power Dosing Gun off & on again
-
Take Gallagher TW Scale and power on
-
Press ‘settings’ then ‘equipment connections’ and select ‘Data Device 1’
-
Select the data device type, press ‘Input & Output’ This means data can be transferred back & forth from Gun & scale to record drench against the EID tags.
-
Now press Wi-Fi, and select the SSID of your dosing gun once search is complete (normally 5-10 seconds)
-
No passkey is required, press connect to continue.
-
Now set Data Output settings. Then press continue.
a. Set Type = Data Logger
b. Select Values = Only Weight
c. Send = On weight lock
d. Resend on data change = No -
Select Data Input settings, then press save
a. Type = Data Input Standard
b. Input Trait = Data Input -
The gun is now connected and you can begin your session. This process of connecting the devices is only required once. Next time the gun will auto connect to the scale.
Care, Maintenance and Storage
To ensure the continued high performance of your gun, always clean and
lubricate the gun immediately after use. Start by reading the cleaning
instructions provided by the fluid manufacturer. Remove the inlet tube from
the fluid container and empty the fluid from the gun and inlet tube. Place the
end of the inlet tube in a mixture of warm water and a little dish washing
detergent, then flush about 1 litre through the gun until it is clean. Flush
again with clean water. Inspect the inlet valve and outlet valve and ensure
they are clean. Remove the barrel and wipe any residue from the barrel with a
soft tissue. Important – Lubricate the valves, piston O-ring, and barrel with
a small amount of vegetable oil after cleaning and before storing the gun.
Periodically remove the valves and wash in warm water and dish washing
detergent. To access the inlet valve, unscrew the front of the inlet valve
housing. To access the outlet valve, remove the nozzle and cap then use a
small screwdriver to release the C-shaped valve retainer, being careful not to
damage it. Be sure to refit it after cleaning as the outlet valve will not
function without it. Wipe the exterior of the gun with a damp cloth and dry
off. Do not submerge the gun in water when cleaning. Store the gun in a clean,
dry place, out of direct sunlight. Recommended storage temperature is from
-10° to 35° C (14° to 95° F). Store above 0° C for optimum battery
performance. If batteries are in a low charge state, fully charge them before
storage.
Remove the batteries before storing the gun.
A. Barrel collar
B. Piston O-ring
C. Nozzle holds the outlet valve and C-clip retainer in the barrel end. See
ex-ploded view.
D. Hose entry valve housing. See exploded view.
The main barrel can be easily separated from the body to aid in cleaning and
maintenance by undoing the barrel collar. Do not over tighten the barrel
collar and ensure all O-rings and felt washers are fitting correctly when
refitting.
Exploded views of inlet and outlet valve assemblies
Note: The gun will not function correctly or at all if the valves are in
the wrong way around.
Inlet assembly – entry hose end
A. Inlet valve connector
B. Inlet valve and spring. The inlet valve is the larger of the two valves.
(70 ml barrel only)
C. Inlet valve housing
Outlet assembly – nozzle end
A. Barrel exit
B. Outlet valve and spring. The outlet valve is the smaller of the two valves.
(70 ml barrel only)
C. Valve retaining C- clip
D. Nozzle with screw collar and O-ring
Charging the batteries
The Te Pari Revolution Dosing Gun is supplied with an intelligent charger that can recharge both lithium-ion batteries at once.
- Remove the battery clip from the base of the gun handle by squeezing in both sides of the clip and gently pulling the clip out.
- Fit each of the batteries into the charger by pulling down the spring loaded retainer, slotting the battery in, and then releasing the retainer. The negative terminals go on to the spring loaded end, with the positive terminals going nearest the lights.
- Connect the charger to a standard mains supply using the cable provided and switch on. The charger’s lights will run up and down while the battery type, state, and charge required is determined. A 12-volt cigarette lighter cable is also provided to power the charger in a vehicle. The charger’s lights will flash slowly in orange showing the charge progress. When all three LED lights are lit a steady green the batteries are fully charged and ready for use.
- Before reloading the batteries, ensure that the contacts inside the gun’s battery clip are clean and dry.
- Squeezing the sides of the battery clip, push it back into the gun handle. The clip only fits into the gun in one direction; with the curved end facing forwards.
- Ensure that the polarity symbols match (+ to + and – to -) as marked on the clip and the batteries.
A. Gun handle bottom
B. Battery clip. Squeeze the sides of the tabs
on the clip to remove or refit into the handle. Note the white terminal
markings.
C. The two lithium-ion 18650 batteries. On this type, the black band is on the
positive end, but this may vary from other makes, so be sure
to check before fitting.
D. The curved end of the clip goes to the curved end of the handle when
refitting the battery clip.
E. Orange flashing lights showing that the batteries are charging and their
state of charge.
F. Three green steady lights mean the batteries are fully charged and ready
for use.
Note: Please use only the supplied batteries.|
---|---
Te Pari Products Warranty
The company warrants that the product you have purchased is free from defects
in materials or workmanship under normal use during the limited warranty
period. The warranty period starts on the date of purchase from the company or
authorised gun agent. Please keep the dated sales receipt for proof of
purchase as you may be required to provide proof of purchase as a condition of
receiving warranty service. The gun has a 1-year limited warranty. The
rechargeable batteries have a 1-year limited warranty. The battery warranty
does not cover changes in battery charge capacity and is limited to defects in
materials or workmanship resulting in battery failure. During the limited
warranty period the company will repair or replace the defective component or
product. All components removed under this warranty become the property of the
company. The replacement component will take on either the remaining warranty
period of the removed component or a ninety (90) day limited warranty for the
spare part, whichever is longer. The replacement component may be manufactured
from new or used materials equivalent to new in performance and reliability.
The company
reserves the right to improve or modify products from time to time without
assuming obligation to modify products previously manufactured.
The company is not responsible for damage that occurs as a result of (1) failure to follow the instructions that came with the product; (2) accident, misuse, abuse or other external causes; (3) voltage surge or lightning strike; (4) modification, tampering or service by an unauthorised service provider. Normal wearing parts requiring replacement or adjustment are not covered by this warranty. These parts include but are not limited to: O-rings, springs, bearings, motor brushes, and tubing. The company is not liable for any damages or personal injury caused by the product or the failure of the product to perform, including any damage or loss of data on the product or computer. The maximum liability of the company under this limited warranty is limited to the lesser of the cost of repair or the replacement of the product. In the event of the product requiring servicing under the terms of the warranty, contact your authorised dealer. The product is to be returned to the service centre at the expense of the purchaser. These conditions constitute the complete and exclusive warranty and supersede all other warranties, representations or conditions relating to this product.
Warranty Claim Form Online
To complete a warranty claim form, visit www.tepari.com/owners-
hub/warranty
and submit a claim here.
Health and Safety Warnings
- Because the toxicity and precautions required for the handling of agricultural health fluids varies, follow the manufacturer’s recommendations in regard to protective clothing, eye wear and handling precautions for the fluid being administered.
- Always be cautious when dispensing fluids.
- Always handle agricultural fluids with care.
Dosing Gun Cautions
Gun
- Operating temperature for the gun is between -10° to 40° C (14° to 104° F).
- The gun is environmentally sealed and meets IP53 standards.
- Keep the gun as dry as possible, drying it off if it becomes wet.
- Do not submerge the gun.
- Do not tamper with or remove the case screws as this will void the warranty.
Batteries
- The gun is designed to use two 18650 lithium-ion rechargeable batteries. Only use approved batteries.
- Dispose of used batteries according to the manufacturer’s instructions.
- Do not heat above 50° C (122° F).
- Do not dispose of in fire or crush, puncture or directly short circuit.
- Keep batteries clean and dry.
FCC Notice
This device complies with part 15 of the FCC rules for Class A digital
devices. It can generate, use and radiate radio frequency energy, and if not
used in accordance with the instructions may cause harmful interference to
radio communications. In a residential area it may cause harmful interference,
which will be required to be corrected by the user at their own expense.
Cautions
| Read all safety warnings and notices and instructions before use.
---|---
| Do not dispose of electric tools with household waste material.
| Contains lithium-ion batteries.
Thank you for purchasing a Te Pari Revolution Dosing Gun. If any issues arise
with its use, please do not hesitate to contact our Support Team by phone or
online, we are here to help.
– The Te Pari Team
NEW ZEALAND
Te Pari Products LTD
Postal PO Box 25, Oamaru 9444
Physical 9 Endeavour Crescent Oamaru, 9400 New Zealand
Phone +64 3 433 0077
Fax +64 3 434 8859
Sales E-mail sales.nz@tepari.com
Accounts E-mail accounts@tepari.com
Freephone 0800 837 274
Website tepari.com| AUSTRALIA
Te Pari Products Pty Ltd
Physical 44 Ricky Way Epping 3076, VIC
Phone 1800 650 682
Fax 1800 669 205
Sales E-mail sales.au@tepari.com| USA
Te Pari Products Inc
Physical 14250 Judicial Rd Burnsville, MN 55306
Phone 800 315 0433
Sales E-mail sales.usa@tepari.com
---|---|---
REVOLUTION DOSING GUN MANUAL
FREEPHONE: 0800 837 274
TEPARI.COM