JOYTECH DET-1015b Infrared Ear Thermometer Owner’s Manual
- June 6, 2024
- Joytech
Table of Contents
- Introduction
- Product Description
- LCD Display Introduction
- Ear Thermometer Advantages
- Ear Thermometer Advantages
- Body Temperature
- Real-Time Clock Setting
- Real-Time Clock Setting
- Illustration For Use
- Temperature Taking Hints
- Memory Mode
- Care And Cleaning
- Battery Replacement
- Specifications
- Troubleshooting
- Calibration
- Service
- Warranty
- Electromagnetic Compatibility Information
- Documents / Resources
Owner’s Manual
Infrared Ear Thermometer
Model DET-1015b ** Document No.: JDET-1504-005
Article-No.: 003
Version: Z
Date of Issue: 2020.12**
Introduction
The DET-1015b thermometer is specifically designed for safe use in the
eardrum. The Infrared Ear Thermometer is a device capable of measuring
people’s body temperature by detecting the intensity of infrared light emitted
from the ear canal of humans. It converts the measured heat into a temperature
reading displayed on the LCD.
When properly used, it will quickly assess your temperature in an accurate
manner. The infrared thermometer is intended for the intermittent measurement
of human body temperature by people of all ages.
Please read all instructions carefully and thoroughly before using this
product.
Warning:
This thermometer is intended for home use only.
- It’s not meant to replace a visit to the doctor. Please consult with your doctor if you have health concerns.
- There is no gender or age limitation for using an infrared ear thermometer.
- To avoid inaccurate readings, always attach a new, clean probe cover for each temperature reading.
- Do not expose the thermometer to temperature extremes (below -25℃/-13℉ or over 55℃/131℉) nor excessive humidity (>95%RH).
- This thermometer must only be used with this brand thermometer probe cover.
- Keep the battery away from children.
- Remove battery from the device when not in operation for a long time.
Product Description
- Probe Cover
- Probe
- Test Button
- Button
- Battery Cover
- Setting Button
LCD Display Introduction
Basic Functions
Real Time Clock| The real-time clock will be recorded with the memory function
and to help you to recognize each measurement result. → Please see the Real-
time clock setting section to learn how to set up the time in the first use.
---|---
Ear Temperature Mode| The thermometer has been designed for practical use.
It’s not meant to replace a visit to the doctor. Please also remember to
compare the measurement result to your regular body tempera → Please see the
Illustration For Use section to learn how to measure the body temperature.
Probe cover detection and eject function| The thermometer has been designed
for use with or without probe cover. Users can press the eject button to
remove the probe cover.
Object Temperature Mode| The object model shows the actual, unadjusted surface
temperatures, which is different from the body temperature. It c help you to
monitor if the object’s temperature is suitable for the or patient, for
example, the baby’s milk. → Please see the Illustration For Use section to
learn how to measure the object temperature.
Beep Alarm
| If the thermometer detects a body temperature ≥ 37.8℃(100.0℉) under ear
mode, there will be a long beep sound followed by th short beep sound to warn
the user.
Memory Mode| There are each 10 sets of memories for ear and object
measurements. Each memory also records the measurement date/time/mode icon.
℃/ ℉ Switch| Please see the Selecting A Temperature Scale section to learn
how to change between Celsius and Fahrenheit.
Voice Switch| You can use the button to choose whether to turn on or off the
sound.
Bluetooth sign| If the APP is successfully connected to the machine, the
Bluetooth sign will always be on, otherwise, it will keep flashing.
Ear Thermometer Advantages
Infrared Ear Thermometer measures core body temperature, which is the temperature of a body’s vital organs. (See Figure 1) Ear temperatures accurately reflect core body temperatures since eardrum is in the cranial cavity and is not affected by the environment. The eardrum shares blood supply with the hypothalamus (part of the brain.) Therefore, body temperature changes are reflected sooner in the ear than they are in other parts of the body.
Ear Thermometer Advantages
Clinical studies have shown that the ear is an excellent site for temperature
measurement because temperatures taken in the ear , reflect the body s core
temperature. Body temperature is regulated by the hypothalamus, which shares
the same blood supply as the tympanic membrane. Changes in core body
temperature are usually seen sooner at the tympanic membrane than at other
sites, such as the rectum, mouth or under the arm.
Advantages of taking temperatures at the ear versus other sites:
- Axillary temperature readings only reflect skin temperature which may not indicate the internal body temperature.
- Rectal temperatures often lag significantly behind internal body temperature changes, especially at times of rapidly changing temperatures.
- Oral temperatures are often influenced by eating, drinking, thermometer placement, breathing through the mouth, or the inability of the person to close their mouth completely.
Body Temperature
The temperature of a healthy person is affected by various factors: the
person’s individual metabolism, their age(body temperature is higher in babies
and toddlers and decreases with age. Greater temperature fluctuations occur
faster and more often in children, e.g. due to growth spurts), their clothing,
the ambient temperature, the time of day(body temperature is lower in the
morning and increases throughout the day towards evening), the preceding
physical and, to a lesser extent, mental activity.
It is suggested that users develop the habit of daily measurement, and use
this product to establish a private normal temperature range for themselves
and their families as a reference for measurement.
Real-Time Clock Setting
When using a thermometer for the first time, please set the parameters of the thermometer.
| ①Set the unit Press Button to select the unit you want.
After the unit is set, press the Setting Button, the time format figure will
appear .
---|---
② Set the time format
The device can display the time in either an AM/PM (12-hour) or a 24:00
(24-hour) format. Press and release button to select the format. With the
preferred time format on the display, press the Setting Button, the Hour
figure is flashing automatically.
| ③ Set the hour
Press and release the button to advance one hour until the correct hour
appears. After the hour is set, press the Setting Button, the Minute figure is
flashing automatically.
---|---
Real-Time Clock Setting
| ④ Set the minute
Press and release the button to advance one minute until the correct minute
appears.
After the minute is set, press the Setting Button, the Year figure is flashing
automatically.
---|---
⑤ Set the year
Press and release the button to advance one year until the correct year
appears.
After the year is set, press the Setting Button, the Month figure will appear.
⑥ Set the month
Press and release the button to advance one month until the correct month
appears.
After the month is set, press the Setting Button, the Date figure is flashing
automatically.
⑦ Set the date
Press and release the button to advance one day until the correct month
appears.
After the day is set, press the Setting Button to exit the setting mode.
Illustration For Use
To measure ear temperature:
-
is in place before each measurement. Place a new probe cover on the probe firmly; making sure the probe and the inner portion of the probe cover fit snugly.
-
Press the To achieve accurate readings, make sure a new, clean probe cover Button, The display is activated to show all segments.
After self-checking Figure 2 appears on the display screen with beeps, so you can start a new measurement. If error messages appear, it means the thermometer is not ready for measurement. Remarks:
a. Probe cover sign stands for taking the ear temperature with probe cover now and vice versa.
b. To avoid the risk of cross-contamination, it is recommended that you always measure the ear temperature with a probe cover. -
Insert the probe along the ear canal as deep as possible into the ear as depicted in Figure 3. Then press the Test BUTTON once to start the measurement. A long beep will sound approximately 1 second after the Test BUTTON is activated. This signals the end of the process as results are displayed on the LCD screen.
-
The thermometer is ready for testing again once an earmark displays on the LCD. A waiting period of 20 seconds between testing is recommended to avoid excessive cooling of the skin.
-
The thermometer will shut off automatically after 60 seconds of inactivity. To prolong battery life, press Button to turn the unit off.
To measure object temperature:
- First, eject the probe cover.
- Aim the thermometer at the center of the object you want to measure with a distance of 1 to 2 cm.
- Press the button to turn on the thermometer, you can take the object’s temperature after hearing two beep sounds.(see figure 4 Remarks: This mode shows the actual, unadjusted surface temperatures, which is different from the body temperature.
How to change the ear mode and object mode:
You can press and hold Setting Button to switch the mode between ear mode and
object mode.
After measurement:
- Power off: Device will automatically shut off if left idle for more than 1 minute to extend battery life.
- Clean the probe after each use to ensure an accurate reading and avoid cross-contamination.
(See the section of Care and Cleaning for details.)
BackLight:
In-Ear mode:
- The display will be lighted GREEN for 3 seconds with a happy face when the unit is ready for measurement and measurement is completed with reading less then 37.3 (99.1 ).
- The display will be lighted YELLOW for 3 seconds with a happy face when a measurement is completed with a reading less than 37.8℃(100.0 ).
- The display will be lighted RED for 3 seconds with a bad face when a measurement is completed with a reading equal to or higher than 37.8℃(100.0 ).
In Object mode: The display will only be lighted GREEN for 3 seconds when the unit is ready for measurement and measurement is completed.
Bluetooth requirements
The thermometer requires a Bluetooth device with:
-
Bluetooth 4.0 or later
-
Android 6.0 or later
-
IOS 8.0 or later
And works with: -
iPhone , iPod, iPad
-
Android Phones and Tablets
Using for the first time
- Download the “JoyHeath” App from the Website or APP Store (Such as Apple Store).
- Open the App on your phone or tablet. If requested, you should enable Bluetooth on your device. You can enable Bluetooth under the Settings menu on your smartphone or tablet.
- Create a new user login, or log in with your existing user name and password.
- Selection device “Thermometer”.
Match your thermometer with a Smart Device
-
If this is your first time using it, bind first. Open the “SETTING” menu, choose“Bind and unbind device” and select the appropriate model.
The date and time on your thermometer will automatically be updated when it’s connected to your phone. -
Confirm that your thermometer is connected successfully. When your thermometer is connected successfully to your smartphone, the “ ****” symbol stops flashing and keeps showing.
Transfer your readings
-
As soon as your measurement is finished, open the app on your smartphone to transfer the readings.
Note : On the matched smartphone, Bluetooth must be enabled. -
You can view your temperature readings in the app.
Temperature Taking Hints
-
The right ear reading may differ from the reading taken at the left ear. Therefore, always take the temperature in the same ear.
-
The ear must be free from obstructions or excess earwax buildup to take an accurate reading.
-
External factors may influence ear temperatures such as:
lying on one ear or the other their ears covered
exposed to very hot or very cold temperatures
recently swimming or bathing
In these cases, remove the individual from the situation and wait 20 minutes prior to taking a temperature. -
For persons wearing hearing aids or earplugs, remove the device and wait 20 minutes prior to taking a temperature.
-
When using the thermometer on infants under age 1, pull the ear up making sure the sensor faces the eardrum. (See Figure 5)
-
When using the thermometer on individuals over the age of 1, pull the ear back making sure the sensor faces the eardrum. (See Figure 6)
-
Do not touch the probe window. A cotton swab and rubbing alcohol may be used to clean the sensor window.
-
If the thermometer is stored in a significantly different environment than the testing location, place it in the testing location for approximately 30 minutes prior to use.
Memory Mode
-
The Memory Mode can be accessed ether in-ear mode or object mode:
When the thermometer has been turned on and followed by Figure 2/4 or finished testing, press and hold the button for three seconds. The letter MEM will appear at the bottom of the display. (See Figure 5) -
The thermometer will automatically memorize the last 30 temperature readings.
Each time the Button is pressed, the screen displays past readings that correspond with a number 1-30. The number 1 reflects the most recent reading, while the number 30 reveals the oldest reading stored in memory.(See Figure 6) Each memory also records the measurement date/time/mode icon. -
The user can press the Test Button to take new measurements. In the memory mode, mark or mark will not change.
Care And Cleaning
-
Probe covers are intended for single use only! Absent, damaged, perforated, or soiled probe covers can lead to inaccurate temperature readings.
-
The probe window must be kept clean, dry, and undamaged at all times to ensure accurate readings. The accuracy of temperature readings can be affected by damage to the probe window, or the presence of dirt and ear wax on the probe window.
-
Fingerprints, earwax, dust, and other soiling compounds reduce the transparency of the window and result in lower temperature readings. If the thermometer is accidentally placed in the ear without a probe cover installed, the window must be cleaned immediately.
-
The probe window is the most delicate part of the thermometer. To safely clean the window, gently wipe its surface with a cotton swab slightly moistened with isopropyl alcohol and immediately wipe dry with a clean cotton swab. After cleaning, allow at least
5 minutes drying time before taking temperatures.
Note: Do not use any chemical other than isopropyl alcohol to clean the probe window. -
Use a soft, dry cloth to clean the thermometer display and exterior.
-
Do not put the thermometer into the water directly.
-
Store the thermometer and probe cover in a dry location, free from dust and contamination and away from direct sunlight.
-
Periodic cleaning and disinfection of the device following use t prevent patient cross-infection. -Use a soft cloth slightly moistened with a 75% isopropyl alco solution to disinfect the thermometer and probe. Do not use abrasive cleaners.
-
Ensure that children do not use the instrument unsupervised; some parts are small enough to be swallowed.
-
Do not remove or modify the equipment without permission.
-
Strong electromagnetic fields may interfere with the proper operation of the thermometer. The device needs special
pre-cautions regarding EMC according to the EMC informatio -
Put the thermometer back to the original packaging after usin
-
It is not intended for use in the oxygen-rich environment and presence of flammable anesthetic mixture with air, oxygen or nitrous oxide.
Battery Replacement
- Replace the battery when it ” appears in the bottom ” of the LCD display. (See Figure 10)
- Slide battery covers down as shown in Figure 11.
- Remove the battery and install 2 new AAA alkaline batteries as shown in Figure 12.
- Slide battery covers back on.
Specifications
Measuring range
| Ear mode: 32.0℃~43.0℃(89.6℉~109.4℉) Object mode: 0℃~100℃(32℉~212℉)
---|---
Measuring site| Ear canal(Ear Mode)
Reference body site| Oral (This thermometer converts the ear temperature to
display its “oral equivalent.”)
Operation mode| Ear mode(Adjust mode)
Laboratory accuracy
| Ear mode:
±0.2℃ (0.4℉) during 35.5℃~42.0℃ (95.9℉~107.6℉)
at 15℃~35℃ (59.0℉~95.0℉) operating temperature range
±0.3℃ (0.5℉) for other measuring and operating temperature range
Object mode: ±4% or ±2℃(4℉) whichever is greater
Display resolution| 0.1℃ or 0.1℉
Measure time| Approximately one second
Operating temperature range:| Ear/Object mode:
10℃~40℃(50℉~104℉), 15%~85%RH, non-condensing Atmospheric Pressure : 700hPa ~ 1060hPa
Storage and transport temperature range| -25℃~ 55℃ (-13℉~131℉), 15%~95%RH, non-condensing Atmospheric Pressure : 700hPa ~ 1060hPa
Clinical accuracy
| Clinical bias:0.08℃( 0.14℉) Clinical repeatability:0.13℃( 0.23℉) Limits
of agreement:0.73℃( 1.31℉)
Shock| withstands drop of 3 feet
Dimension| 5714638mm
Weight| Approx.83grams(with batteries)
Battery| DC3V(2×AA battery)
Battery life| Approx. 1 year/6000 readings
Expected service life| Three years
Ingress protecting rating| IP22
Troubleshooting
Error message | Problem | Solution |
---|---|---|
Measurement before thermometer is ready | Take a measurement until JorIF |
appears on the display.
| The ambient temperature is not within the range between 10°C and 40°C
(50°F-104°F).| Place the thermometer in a room for at least 30 minutes at room
temperature between 10°C and 40°C
(50°F —104°F )
| The thermometer is placed incorrectly or unsteady.| Read Illustration For
Use thoroughly and U
take a new temperature measurement.
| The thermometer showing a rapid ambient temperature change.| Allow the
thermometer to rest in a room for at least 30 minutes at
room temperature: between 10°C and 40°C (50°F —104T )
| The thermometer is not functioning properly.| Unload the battery, wait for 1
minute and repower it. If the message reappears, contact the retailer for
service.
Troubleshooting
Error message | Problem | Solution |
---|---|---|
In object temperature mode, measurement with probe cover. | Remove probe |
cover in object temperature mode.
| In Ear mode: Temperature taken is higher than 43.0 ℃ (109.4℉). In Object
mode: Temperature taken is higher than 100 ℃ (212℉).|
Read Temperature Taking Hints Thoroughly, then check the integrity of the probe cover and take a new temperature measurement.
| In Ear mode: Temperature taken is lower than 32.0 ℃ (89.6℉). In Object mode:
Temperature taken is lower than 0 ℃ (32℉).| Read Temperature Taking Hints
thoroughly , then make sure the probe cover and lens filter are clean, then
take a new temperature measurement.
| The thermometer could not work due to low battery.| Replace two new alkaline
batteries size AAA.
Calibration
The thermometer is initially calibrated at the time of manufacture.
If the thermometer is used according to the use instruction, periodic
readjustment is not required. However, We recommend checking calibration every
two years or whenever the clinical accuracy of the thermometer is in question.
Please send the complete device to the dealers or manufacturer.
The above recommendations do not supersede the legal requirements.
The user must always comply with legal requirements for the control of the
measurement, functionality, and accuracy of the device which are required by
the scope of relevant laws, directives or ordinances where the device is used.
A clinical summary and procedures for checking calibration are available upon
request.(Turn on the thermometer and press the power button a long time until
entering into calibrate mode, the software version will be displayed.)
| Type B F applied part
---|---
Direct Current
Disposal of this product and used batteries should be carried out in
accordance with the national regulations for the disposal of electronic
products
Consult Accompanying Documents
Batch Code
Storage and Transportation Temperature Limit:
-13℉℃~131℉(-25℃~55℃)
Service
The thermometer has a limited one-year warranty. Do not attempt to disassemble
or repair the thermometer by yourself. Should service be required during or
after the warranty period you must contact the manufacturer. Repackage the
thermometer carefully in its original packaging or securely pack to avoid
damage during shipping. Include the original sales slip indicating the date of
purchase, a note describing the problem, and your return address. Send the
thermometer prepaid and insured. The lay operator or lay responsible
organization should contact the manufacturer or the manufacturer’s
representative: – for assistance, if needed, in setting up, using or
maintaining the thermometer; or
– to report unexpected operations or events.
Manufacturer:
JOYTECH HEALTHCARE Co., Ltd.
No.365, Wuzhou Road, Yuhang Economic Development Zone, Hangzhou city,
311100 Zhejiang,China
Telephone: +86-571-81957767
Fax: +86-571-81957750
The product is in compliance with the requirements of MDD 93/42/EEC, “0197” is
the identification number of notify body;
European Authorized Representative: Shanghai International Holding Corp. GmbH (Europe) Eiffestrasse 80, 20537 Hamburg, Germany
Warranty
This appliance conforms to the following standards:
EN 12470-5 Clinical thermometers —Part 5:Performance of infrared ear
thermometers(with maximum device),
ISO 80601-2-56 Medical electrical equipment —Part 2-56: Particular
requirements for basic safety and essential performance of clinical
thermometers for body temperature measurement,
IEC 60601-1-11 Medical electrical equipment —Part 1-11: General requirements
for basic safety and essential performance –Collateral Standard: Requirements
for medical electrical equipment and medical electrical systems used in the
home healthcare environment and complies with the requirements of IEC
60601-1-2(EMC) , IEC 60601-1(Safety) standards. And the manufacturer is ISO
13485 certified.
The thermometer is warranted by manufacturer to be free from defects in
material and workmanship under normal use and service for a period of one year
from the date of delivery to the first user who purchases the instrument. This
warranty does not cover batteries, damage to the probe window, or damage to
the instrument caused by misuse, negligence or accident, and extends to only
to the first purchaser of the product. Additionally, this warranty becomes
void if the thermometer is operated with anything other than this rand
thermometer probe covers.
FCC Information
Caution: Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user authority to operate the equipment.
- Note :
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try and correct the interference by one or more of the following measures:
– Reorient or relocate the receiving antenna.
– Increase the distance between the equipment and the receiver.
– Connect the equipment to an outlet on a circuit different from that to which the receiver is connected.
– Consult the dealer or an experienced radio/TV technician for help This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that may cause undesired operation. The device has been evaluated to meet general RF exposure
requirements. The device can be used in portable exposure conditions without restriction.
Electromagnetic Compatibility Information
The device satisfies the EMC requirements of the international standard IEC
60601-1-2.The requirements are satisfied under the conditions described in the
table below.The device is an electrical medical product and is subject to
special precautionary measures with regard to EMC which must be published in
the instructions for use.Portable and mobile HF communications equipment can
affect the device.Use of the unit in conjunction with non-approved accessories
can affect the device negatively and alter the
electromagnetic compatibility.The device should not be used directly adjacent
to or between other electrical equipment.
Taele 3
Guidance and manufacturer’s declaration — electromagnetic emission
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should
assure that it is used in such an environment.
Emissions test| Compliance| Electromagnetic environment –
guidance
RF emissions
CISPR 11| Group 1| The device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11| Class B| The device is suitable for use in all
establishments, including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2| Not applicable
Voltage fluctuations I
flicker emissions
IEC 61000-3-3| Not applicable
Guidance and manufacturer’s declaration — electromagnetic immunity
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should
assure that it is used in such an environment.
Immunity test| IEC 60601
test level| Compliance level| Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2| ± 8 kV contact
t2 kV, 14 kV, t8 kV, t15 kV
air| ± 8 kV contact
+.2 kV, 14 kV, t8 kV, t15 kV
air| Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrostatic
transient / burst
IEC 61000-4-4| ± 2 kV for power
supply lines
100 kHz repetition frequency
1 1 kV for input/output
lines| N/A| N/A
Surge
IEC 61000-4-5| t 0.5 kV, t 1 kV differential
mode
line-line| N/A| N/A
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11| 0 % UT
(100 % dip in UT )
for 0.5 cycle at r, 45°, 90°, 135°,180°, 225°, 270°, and 315°
0 % UT
(100 % dip in UT ) for 1 cycle at 0°
70 %UT
(30 % dip in UT )
for 25/30 cycles at 0°
0 % UT
(100 % dip in UT )
for 250/300 cycle at 0°| N/A| N/A
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8| 30 Aim, 50/60Hz| 30 Aim, 50/60Hz| Power frequency magnetic
fields should be at levels characteristic of a typical location in a typical
commercial or hospital environment.
NOTE : UT is the a. c. mains voltage prior to application of the test
level.
Guidance and manufacturer’s declaration — electromagnetic immunity
The device is intended device should assure eat| for use in the
electromagnetic environment, it is used in such an environment.|
specified below. The customer or the user of the
Immunity test| IEC 60601 test
level| Compliance
level| Electromagnetic environment – guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IF,C 61000-4-3| 3 Vrms
150 kHz to 80 MHz
6 Vrms 150 kHz to
80 MHz outside
ISM bandsa
10 V/m
80 MHz to 2.7 GHz| N/A
10 V/m| Portable and mobile RF communications
equipment should be used no closer to any part of the device, including
cables, than the
recommended separation distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the
recommended separation distance in metres(m). Field strengths from fixed RF
transmitters, as determined by an electromagnetic site survey,’ should be less
than the compliance level in each frequency range b
Interference may occur in the vicinity of equipment marked with the following
symbol:
Table 3 continued
NOTE!At 80 MHz and 800 MHz, the higher frequency range applies.NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
a The ISM(industrial, scientific and medical) bands between 0,15 MHz and 80
MHz are 6,765 MHzto6,795 MHz;13,553 MHz to 13,567 MHZ;26,957 MHz to 27,283
MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15 MHz and
80 MHz are 1,8 MHz to 2,0 MHz,3,5 MHz to 4,0 MHz,5,3 MHz to 5,4 MHz,7 MHz to
7,3 MHz,10,1 MHz to 10,15 MHz,14 MHz to 14,2 MHz,18,07 MHz to 18,17
MHz,21,0MHz to 21,4MHz,24,89 MHz to 24,99 MHz,28,0 MHz to 29,7 MHz and 50,0
MHz to 54,0 MHZ.
b The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz
and in the frequency range 80 MHz to 2,7 GHz are intended to decrease the
likelihood that mobile/portable communications equipment could cause
interference if it is inadvertently brought into patient areas. For this
reason, an additional factor of 10/3 has been incorporated into the formulae
used in calculating the recommended separation distance for transmitters in
these frequency ranges.
c Field strengths from fixed transmitters, such as base stations for
radio(cellular/ cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the device is used exceeds
the applicable RF compliance level above, the device should be observed to
verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the device.
d Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3 V/m.
Recommended separation distances between
portable and mobile RF communications equipment and the device
The device is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the
device can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment
(transmitters) and the device as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum
output of
transmitter
W| Separation distance according to frequency of the transmitter
m
| |
0.01| 0.12| 0.04| 0.07
0.1| 0.37| 0.12| 0.23
1| 1.| 0.35| 0.7
10| 4.| 1.| 2.
100| 12.| 4.| 7.0
For transmitters rated at a maximum output power not listed above the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
Recommended separation distances between RF wireless communications
equipment
The device is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the
device can help prevent electromagnetic interference by maintaining a minimum
distance between RF wireless communications equipment and the device as
recommended below, according to the maximum output power of the communications
equipment.
Frequency
MHz| Maximum
Power
W| Distance| IEC 60601
Test Level| Compliance
Level| Electromagnetic
Environment – Guidance
8
35| 2.| 0.3| 27| 27| RF wireless communications
equipment should be used no
closer to any part of the device,
including cables, than the
recommended separation distance calculated from the equation applicable to the
frequency of the transmitter. Recommended separation distace
Where P is the maximum output power rating of the ransmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in meters (m). Field strengths from fixed RF transmitter, as
determined by an electromagnetic site survey, should be less than the
compliance level in each frequency range. Interference may occur in the
vicinity of equipment marked with the following symbol:
450| | 0.3| 28| 28
710| 2| | 9| 9
745
780
810| CNI| | -)8| 28
870
930
1720| C`l| | 28| 28
1845
1970
2450| 2| 0.3| 28| 28
5240| 0.2| 0.3| 9| 9
5500
5785
Note 1: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
WARNINGS!
- This device should not be used in the vicinity or on top of other electronic equipment such as cell phones, transceivers or radio control products. If you have to do so, the device should be observed to verify normal operation.
- The use of accessories and power cord other than those specified, with the exception of cables sold by the manufacturer of the equipment or system as replacement parts for internal components, may result in increased emissions or decreased immunity of the equipment or system.
Document No.: JDET-1504-005
Article No.: 003
Version: Z
Date of Issue: 2020.12
Documents / Resources
|
JOYTECH DET-1015b Infrared Ear
Thermometer
[pdf] Owner's Manual
0013, 2AQVU0013, DET-1015b Infrared Ear Thermometer, DET-1015b, Infrared Ear
Thermometer
---|---