JOYTECH DET-3015b Infrared Forehead Thermometer Owner’s Manual
- June 3, 2024
- Joytech
Table of Contents
- JOYTECH DET-3015b Infrared Forehead Thermometer
- Introduction
- Product Description
- LCD Display Introduction
- Basic Functions
- Forehead Thermometer Advantages
- Real Time Clock Setting
- Temperature Taking Hints
- Illustration For Use
- Care And Cleaning
- Battery Replacement
- Specifications
- TROUBLESHOOTING
- Calibration
- Symbol Explanation
- Service
- Warranty
- FCC Information
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
JOYTECH DET-3015b Infrared Forehead Thermometer
Introduction
Please read all instructions carefully and thoroughly before using this
product.
The DET-3015b infrared forehead thermometer is specifically designed for safe
use on the forehead. The Infrared Forehead Thermometer is a device capable of
measuring people’s body temperature by detecting the intensity of infrared
light emitted from the forehead. It converts the measured heat into a
temperature reading displayed on the LCD. When properly used, it will quickly
assess your temperature in an accurate manner.
Indications for Use:The infrared forehead thermometer is intended for the
intermittent measurement of human body temperature from the skin surface of
forehead. The device can be reused by people of all ages for home use and
clinical use.
Intended use:The infrared forehead thermometer is used to measure body
temperature by measuring forehead.
This appliance conforms to the following standards:
ASTM E1965-98 Standard Specification for Infrared Thermometers for
Intermittent Determination of Patient Temperature, ISO 80601-2-56 Medical
electrical equipment —Part 2-56:
Particular requirements for basic safety and essential performance of clinical
thermometers for body temperature measurement, IEC 60601-1-11 Medical
electrical equipment —Part 1-11: General requirements for basic safety and
essential performance –Collateral
Standard: Requirements for medical electrical equipment and medical electrical
systems used in the home healthcare environment and complies with the
requirements of IEC 60601-1-2(EMC) ,
AAMI ANSI ES60601-1(Safety) standards. And the manufacturer is ISO 13485
certified.
Warning
- There is no gender or age limitation for using infrared forehead thermometer.
- Do not touch the temperature probe with hands .
- Use of this Forehead thermometer is not intended as a substitute for consultation with your physician.
- Do not allow children to take their temperatures unsupervised, some parts are small enough to be swallowed.
- Never immerse this device in water or other liquids.
- Do not modify this equipment without authorization of manufacturer.
- Do not expose the thermometer to temperature extremes (below -25℃/-13℉ or over 55℃/131℉) nor excessive humidity (>95%RH).
- Keep the battery away from children.
- Remove battery from the device when not in operation for a long time.
- Do not put the thermometer in direct sunlight or with cotton wool, otherwise the accuracy will be affected.
- Portable and mobile RF communications can affect the devise. The device needs special precaution regarding EMC according to the EMC information provided in the accompany documents
- ME equipment should not be cleaned and wiped while in use.
- Any serious incident that has occurred in relation to the device should be reported to the manufacturer and the competent authority of the Member State in which the user and/or patient is established.
- The probe of the ME equipment shall not be serviced or maintained while in use with a patient.
- The Applied part is the shell.
Product Description
1. Probe
2. Test Button
3. START Button
4. Setting Button
5. Battery Cover
6. Display
LCD Display Introduction
Basic Functions
Real Time Clock| The real time clock will be recorded with the memory function and help you to recognize each measurement result.
→ Please see the Real time clock setting section to learn how to setup the time in the first use.
---|---
Forehead Mode| The thermometer has been designed for practical use. It’s not
meant to replace a visit to the doctor. Please also remember to compare the
measurement result to your regular body temperature.
→ Please see the To measure forehead temperature.
Object Mode
| The object mode shows the actual, unadjusted surface temperatures, which is different from the body temperature. It can help you to monitor if the object temperature is suitable for the baby or patient, for example the baby’s milk.
→ Please see the To measure object temperature.
Voice| The thermometer will broadcast the result after finishing measurement.
Memory Mode
| The machine can store up to 30 sets of measurements. Each memory also records the measurement results date/time/mode.
When the data exceeds 30 sets, the coverage starts from the earliest data.
℃/℉ Switch| Please see the Real Time Clock Setting to learn how to change
between Celsius and Fahrenheit.
Distance Detection| If the object is too far away from the probe, the
temperature will not be measured until the probe is moved within the measured
range.
Sound Switch| The thermometer can turn on or off sound.
→Please see the Illustration For Use.
Bluetooth sign| If the APP is successfully connected to the machine,the
Bluetooth sign will always be on, otherwise it will keep flashing.
Language Switch| Please see the Real Time Clock Setting to learn how to change
between Chinese and English.
Forehead Thermometer Advantages
Infrared Forehead Thermometer measures core body temperature, which is the temperature of a body’s vital organs. (See Figure 1) This thermometer is designed to measure the temperature of the skin surface over the temporal artery, a major artery of the head. The temporal artery is connected to the heart via the carotid artery, directly leading from the aorta, the main trunk of the arterial system. It offers constant blood flow. Therefore, body temperature changes are reflected sooner in the forehead than they are in other parts of the body such as oral, rectal and underarm.
Real Time Clock Setting
When using thermometer for the first time, please set the parameters of the thermometer. With the thermometer off, press and hold Setting BUTTON to enter into setting mode.
-
Set the unit
Press Button to select the unit you want.
After the unit is set,press Setting Button ,the time format figure will appear . -
Set the time format
The device can display the time in either an AM/PM (12-hour) or a 24:00(24-hour) format. Press and release Button to select the format. With the preferred time format on the display, press Setting Button, the Hour figure is flashing automatically. -
Set the hour
-
Set the minute
Press and release the Button to advance one minute until the correct minute appears.
After the minute is set, press Setting Button , the Year figure is flashing automatically. -
Set the year
Press and release the Button to advance one year until the correct year appears.
After the year is set, press Setting Button , the Month figure will appear -
Set the month
Press and release the Button to advance one month until the correct month appears.
After the month. is set, press Settin g B u tt on , the Date figure is flashing automatically. -
Set the date
Press and release the Button to advance one day until the correct month appears. After the day is set, press Setting Button ,the Language of voice figure will appear.
Set the language of voice The device can set the language of voice in either Chinese or English. Press and release Button to select the language. With the preferred language on the display, press Setting Button to exit the setting mode.
Temperature Taking Hints
To ensure that the reading always reflects the body temperature accurately, you need to take account of the following factors which may affect an accurate reading.
-
It is important to know each individual’s normal temperature when they are well. This is the only way to accurately diagnose a fever. To determine normal temperature, take multiple readings when healthy. Re-measure with a standard digital thermometer for confirmation.
-
Users must be inside for 30 minutes before taking a measurement.
Note: Users and the thermometer should be in the same ambient temperature for at least 10 minutes before taking a reading. -
Users should not drink, eat, or be physically active such as bathing, showering, shampooing and hair drying before/while taking the measurement. Remove hat and hair and wait 10 minutes before taking a reading.
-
. Oils or cosmetics on the forehead may give a lower temperature reading than the actual one. Remove dirt from the forehead before taking a measurement. Wait at least 10 minutes after washing the forehead area before taking a reading.
-
Holding a hand on the forehead for any length of time will affect the temperature reading.
-
Do not take temperature over scar tissue, open sores or abrasions.
-
Do not use the thermometer on a perspiring or sweating forehead, as this may affect the reading.
-
Don’t take a measurement while or immediately after nursing a baby.
-
Do not take temperatures with this thermometer near places that are very hot, such as fireplaces and stoves.
-
The probe window of the thermometer is the most delicate part of the device. Do not touch the probe window. The accuracy of the reading may be affected if the probe window is damaged or dirty.
-
If the thermometer is stored in a significantly different environment than testing location, place it in the testing location for approximately 30 minutes prior to use.
-
It is not intended for use in the oxygen rich environment and presence of flammable anesthetic mixture with air, oxygen or nitrous oxide.
Illustration For Use
To measure forehead temperature:
-
Press the BUTTON , The display is activated to show all After self-checking Figure 2 appears on the display screen with voice, so you can start a new measurement.
-
Aim the thermometer at the center of the forehead with the distance less than 3cm(See figure 3) and then press the Test Button.
Note1: Do not remove the thermometer from the forehead before hearing voice.
Note2:If the object to be tested is too far from the probe, a prompt will appear(See figure 4).If the thermometer remains in this state for more than 10 seconds, it will automatically return to Figure 2 state -
The thermometer will broadcast and display the result after finishing measurement. When the voice is off, the vibration is automatically turned on, and there is a vibration reminder at the end of the measurement.
-
Press the BUTTON to turn off.
How to turn on or off sound:
You can press Setting Button to turn on or off sound.
How to change the forehead mode and object mode:
You can press and hold Setting Button to switch the mode between Forehead
mode and Object mode.
To measure object temperature:
- Press the Button to turn on the thermometer, press and hold Setting Button to switch to object mode, you can take the object temperature after hearing voice.(see figure 4)
- Aim the thermometer at the center of the object you want to measure with The distance less than 3cm.
- Press the Test Button and then read the temperature on the display.
- Press the START BUTTON to turn off..
After measurement:
- Power off: Device will automatically shut off if left idle for more than 30 seconds to extend battery life.
- Clean the probe after each use to ensure an accurate reading and avoid cross contamination.
(See the section of Care and Cleaning for details.)
Back Light:
In Forehead mode:
- The display will be lighted GREEN for 3 seconds when the unit is ready for measurement and a measurement is completed with a reading less than 37.3℃(99.1℉).
- The display will be lighted YELLOW for 3 seconds when a measurement is completed with a reading less than 37.8℃(100.0℉).
- The display will be lighted RED for 3 seconds when a measurement is completed with a reading equal to or higher than 37.8℃(100.0℉).
In Object mode:
The display will only be lighted GREEN for 3 seconds when the unit is ready
for measurement and a measurement is completed.
Care And Cleaning
- The probe window must be kept clean, dry, and undamaged at all times to ensure accurate readings. The accuracy of temperature readings can be affected by damage to the probe window, or the presence of dirt, fingerprints, dust and other soiling compounds on the probe window. Degraded sensors can degrade performance or cause other problems.
- For cleaning:
- Soak a clean soft cloth in drinking water,wring it out,and then wipe the thermometer(including probe) no less than 3 times;
- Visual or use magnifying glass to observe the thermometer have no visible dirt and stains ,then use another clean soft cloth to wipe the thermometer residue water;
- Put the thermometer in the original packaging.
- For disinfection:
- Soak a clean soft cloth in drinking water,wring it out,and then wipe the thermometer(including probe) no less than 3 times;
- Visual or use magnifying glass to observe the thermometer have no visible dirt and stains ,then use another clean soft cloth to wipe the thermometer residue water;
- Using a clean soft cloth dipped in 70% medical alcohol,wipe the probe for 3 Times,each time 1 minute.
- Using a clean cotton swab dipped in 70% medical alcohol,wipe the sensor window 3 times;
- Wait at least 10 minutes to let the alcohol volatilization and put it in the original packaging.
- Do not put the thermometer into water directly.
- Store the thermometer in a dry location, free from dust and contamination and away from direct sunlight.
- Put the thermometer back to the original packaging after using.
Battery Replacement
- When flashes, it indicates that the power is low, but you can continue to measure; replace battery when ” ” remains displayed in the lower right corner of LCD display as shown in Figure 8.
- Slide battery cover down as shown in Figure 9.
- Remove battery and install 2 new AA alkaline batteries as shown in Figure 10 .
- Slide battery cover back on.
Specifications
Measuring range| Forehead mode: 34.0℃~43.0℃(93.2℉~109.4℉) Object mode:
0℃~100℃(32℉~212℉)
---|---
Measuring site| Forehead(Forehead Mode)
Reference body site| Oral (This thermometer converts the forehead temperature
to display its “oral equivalent.”)
Operation mode| Forehead mode(Adjust mode)
Laboratory accuracy
| Forehead mode:
±0.2℃ (0.4℉) during 35.5℃~42.0℃ (95.9℉~107.6℉)
at 15℃~35℃ (59.0℉~95.0℉) operating temperature range
±0.3℃ (0.5℉) for other measuring and operating temperature range
Object mode: ±4% or ±2℃(4℉) whichever is greater
Display resolution| 0.1℃ or 0.1℉
Measure time| Approximately 1 second
Operating temperature
range:
| 5℃~40℃(41℉~104℉), 15%~85%RH, non-condensing Atmospheric Pressure : 70kPa ~
106kPa
Storage and transport
temperature range
| -25℃~ 55℃ (-13℉~131℉), 15%~95%RH, non-condensing
Atmospheric Pressure : 70kPa ~ 106kPa
Clinical accuracy
| 0-1 year:
Clinical bias: -0.17℃( – 0.31℉) ; Clinical repeatability: 0.06℃( 0.11℉) ; Limits of agreement: 0.24℃( 0.43℉) 1- 5 years:
Clinical bias: -0.17℃( – 0.31℉) ; Clinical repeatability: 0.07℃( 0.13℉) ; Limits of agreement: 0.22℃( 0.40℉) over 5 years:
Clinical bias: -0.17℃( – 0.31℉) ; Clinical repeatability:
0.20℃( 0.36℉) ; Limits of agreement: 0.26℃( 0.47℉)
Shock| withstands drop of 3 feet
Dimension| 140.037.834.8mm
Weight| Approx.118grams(with batteries)
Battery| DC3V(2×AA battery)
Battery life| Approx. 3000 readings
Expected service life| Three years
Ingress protecting rating| IP22
Contraindication| No contraindication
TROUBLESHOOTING
Calibration
The thermometer is initially calibrated at the time of manufacture. If the thermometer is used according to the use instruction, periodic readjustment is not required. However, We recommends checking calibration every two years or whenever clinical accuracy of the thermometer is in question. Please send the complete device to the dealers or manufacturer.
The above recommendations do not supersede the legal requirements. The user must always comply with legal requirements for the control of the measurement, functionality, and accuracy of the device which are required by the scope of relevant laws, directives or ordinances where the device is used.
A clinical summary and procedures for checking calibration are available upon request.(Turn on the thermometer and press the button long time until entering into calibrate mode, software version will be displayed.)
Symbol Explanation
Service
The thermometer has a limited one year warranty. Do not attempt to disassemble
or repair the thermometer by yourself. Should service be required during or
after the warranty period you must contact the manufacturer. Repackage the
thermometer carefully in its original packaging or securely pack to avoid
damage during shipping. Include the original sales slip indicating the date of
purchase, a note describing the problem, and your return address. Send the
thermometer prepaid and insured.
The lay operator or lay responsible organization should contact the
manufacturer or the manufacturer’s representative:
- for assistance, if needed, in setting up, using or maintaining the thermometer; or
- To report unexpected operation or events.
Warranty
Thermometer is warranted by manufacture to be free from defects in material and workmanship under normal use and service for a period of one year from the date of delivery to the first user who purchases the instrument. This warranty does not cover batteries, damage to the probe window, or damage to the instrument caused by misuse, negligence or accident, and extends to only to the first purchaser of the product.
FCC Information
Caution: Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user authority to operate the equipment.
Note :
This equipment has been tested and found to comply with the limits for a Class
B digital device, pursuant to Part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses, and can radiate
radio frequency energy. If this equipment does cause harmful interference to
radio or television reception, which can be determined by turning the
equipment off and on, the user is encouraged to try and correct the
interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the distance between the equipment and the receiver.
- Connect the equipment to an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including interference
that may cause undesired operation.
The device has been evaluated to meet general RF exposure requirement. The
device can be used in portable exposure condition without restriction.
Electromagnetic Compatibility Information
The device satisfies the EMC requirements of the international standard IEC 60601-1-2. The requirements are satisfied under the conditions described in the table below. The device is an electrical medical product and is subject to special precautionary measures with regard to EMC which must be published in the instructions for use. Portable and mobile HF communications equipment can affect the device. Use of the unit in conjunction with non-approved accessories can affect the device negatively and alter the electromagnetic compatibility. The device should not be used directly adjacent to or between other electrical equipment.
Electromagnetic Compatibility Information
NOTE 1At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people. a The ISM(industrial, scientific and medical) bands between 0,15
MHz and 80 MHz are 6,765 MHzto6,795 MHz;13,553 MHz to 13,567 MHZ;26,957 MHz to
27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15
MHz and 80 MHz are 1,8 MHz to 2,0 MHz,3,5 MHz to 4,0 MHz,5,3 MHz to 5,4 MHz,7
MHz to 7,3 MHz,10,1 MHz to 10,15 MHz,14 MHz to 14,2 MHz,18,07 MHz to 18,17
MHz,21,0MHz to 21,4MHz,24,89 MHz to 24,99 MHz,28,0 MHz to 29,7 MHz and 50,0
MHz to 54,0 MHZ.
The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2,7 GHz are intended to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. For this reason, an additional factor of 10/3 has been incorporated into the formulae used in calculating the recommended separation distance for transmitters in these frequency ranges.
Field strengths from fixed transmitters, such as base stations for
radio(cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the device is used exceeds
the applicable RF compliance level above, the device should be observed to
verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as
re-orienting or relocating the device.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
WARNINGS!
-
This device should not be used in the vicinity or on the top of other electronic equipment such as cell phone, transceiver or radio control products. If you have to do so, the device should be observed to verify normal operation.
-
The use of accessories and power cord other than those specified, with the exception of cables sold by the manufacturer of the equipment or system as replacement parts for internal components, may result in increased emissions or decreased immunity of the equipment or system.
-
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation.
-
Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
-
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the device,including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
-
Portable and mobile RF communications can affect the device.
The device needs special pre-cautions regarding EMC according to the EMC information provided in the accompany documents. -
Do not use the devices in the MR environment.
-
The Operator should not use the system and should inform the customer service, if the ESSENTIAL PERFORMANCE is lost or degraded due to EM DISTURBANCES.
-
PRECAUTION: The performance of the device may be degraded should one or more of the following occur:
-
Operation outside the manufacturer’s stated temperature and humidity range.
-
Storage outside the manufacturer’s stated temperature and humidity range.
-
Mechanical shock (for example, drop test) or degraded sensor.
-
Patient temperature is below ambient temperature.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>