AND TM-2440 Ambulatory Monitoring System User Guide

June 6, 2024
AND

AND-TM-2440-Ambulatory-Monitoring-System-LOGO

AND TM-2440 Ambulatory Monitoring System

AND-TM-2440-Ambulatory-Monitoring-System-IMAGE

 Software installation

Insert the A&D Medical USB stick into your PC to start the installation.

  •  In your File Explorer, open your removable disc or USB drive. Find and open the Setup.exe application.

AND-TM-2440-Ambulatory-Monitoring-System-FIG 2

  •  Click on Install.

  • Follow the guided instructions. This may take a few minutes.

  • To open the application, navigate to your desktop and double click the A&D Client application. If it is not on your desktop, search for A&D Client in recently installed programs.

Inserting the batteries

  1. Open the battery cover.
  2. Refer to thev symbol inside the battery compartment. Insert two  new AA batteries into the proper “+” and “-” direction.
  3. . Close the battery cover.

Connecting the monitor

  1. Connect the micro USB cable between the monitor and computer.
    AND-TM-2440-Ambulatory-Monitoring-System-FIG 7

  2. The buzzer will sound and OFF appears on the OLED display.  The monitor enters its standby mode.

Battery Tips:

  1.  New AA batteries will last approximately 200 measurements. This varies depending on the quality of batteries used and operating temperatures.
  2.  We recommend using high-quality AA batteries.
  3.  If the AA batteries die, there is an internal coin cell battery that will retain measurement results and clock settings. This is automatically recharged when new AA batteries are inserted.

Programming the monitor:

  1. Open the A&D software.

  2.  Click on the Recorder icon.

  3. Select settings for a new ABPM study. Once complete, select Save. The device is now programmed.
    AND-TM-2440-Ambulatory-Monitoring-System-FIG 11

  4. Disconnect the micro USB cable from the device.

  5.  Turn the ABPM function ON. For the TM-2440 only, press and hold Event for at least three seconds. The icon should appear on the display. For the TM-2441 only, slide the Auto Switch to ON, then lock it.

AND-TM-2440-Ambulatory-Monitoring-System-FIG 12

Setting up the patient

  1.  Wrap the cuff around the patients’ upper arm (refer to below cuff guide).
    Part Number| Cuff| Size
    ---|---|---
    TM-CF202A| Small Cuff| 5.9-8.7” (15-22 cm)
    TM-CF302A| Medium Cuff| 7.8-12.2” (20-31 cm)
    TM-CF402A| Large Cuff| 11.0-15.0” (28-38 cm)
    TM-CF502A| Extra Large Cuff| 14.2-19.7” (36-50 cm)

  2. Position tube around the back of the neck and attach monitor to the belt loop. Clip the tube to the undershirt. If user does not have undershirt, tape to skin. Wrap the belt loop around waist. Fasten the air plug from the air hose into the air socket. Secure monitor into the carrying case.
    AND-TM-2440-Ambulatory-Monitoring-System-FIG 14

  3.  Press the START/STOP button on the top of the device to start the first measurement. The programmed intervals will automatically begin.

Downloading the data

  1.  Connect the micro USB cable between the monitor and computer.  Please ensure you are using the cable that came with your monitor.
    AND-TM-2440-Ambulatory-Monitoring-System-FIG 15

  2.  Open the A&D Client application from your computer.”

  3.  Click the Download icon. The download will begin.

  4. Enter the patient information. Ensure to enter accurate sleep information. The reporting is based on this sleep input

Reviewing data and printing report

  1.  The catalyst for all measurements appears.
    AND-TM-2440-Ambulatory-Monitoring-System-FIG 19

  2. To generate report, click the Layout icon on the top menu.

  3.  Customize your report by selecting and deselecting items in the report layout. Click Apply and Save to set layout for future report generation.

  4.  Click on the Print icon to print results.

  5. To save as .PDF, choose Views. Then, click on the Export icon. Choose the file location and file name. Click Save As. It will automatically generate a .CSV and .PDF file and save it in the selected file location.

AND-TM-2440-Ambulatory-Monitoring-System-FIG 22

Patient Tips:

DO

  1. When inflation begins, relax arm, stay still and quiet.
  2.  Ensure the tube is not compressed or kinked.
  3.  Track your bedtime and awake time then report to the clinician.

DO NOT:

  1. Operate heavy machinery. Vibrations may disrupt measurements.
  2.  Remove batteries during session.
  3.  Drop or shock the device.
  4.  Expose the device to rain or water.
  5.  Arrange the cuff hose around your neck during sleep.