beurer BM 51 EasyClip Blood Pressure Monitor Instructions

June 6, 2024
Beurer

beurer BM 51 EasyClip Blood Pressure Monitor

beurer-BM-51-EasyClip-Blood-Pressure-Monitor-pro

Dear customer,
Thank you for choosing a product from our range. Our name stands for high- quality, thoroughly tested products for applications
in the areas of heat, weight, blood pressure, body temperature, pulse, gentle therapy, massage, beauty, baby and air.
With kind regards,
Your Beurer team

Included in delivery

Check that the exterior of the cardboard delivery packaging is intact and make sure that all contents are present. Before use, ensure that there is no visible damage to the device or accessories and that all packaging material has been removed. If you have any doubts, do not use the device and contact your retailer or the specified Customer Service address.

  • upper arm blood pressure monitor
  • 1x easyClip upper arm cuff (22–42 cm)
  • 1x instructions for use
  • 1x storage bag
  • 4 x 1.5V AA LR6 batteries

Signs and symbols

The following symbols are used on the device, in these instructions for use, on the packaging and on the type plate for the device:

beurer-BM-51-EasyClip-Blood-Pressure-Monitor-2

Intended use

The BM 51 easyClip upper arm blood pressure monitor is intended for home use and serves as a non-invasive way
of measuring and monitoring arterial blood pressure values and pulse rates in adults with an upper arm circumference
of 22–42 cm. As a result, you can quickly and easily measure your blood pressure and pulse and view various average values from previous measurements. The recorded measurements are classified and evaluated graphically. A symbol in the display also warns of possible existing cardiac arrhythmia.

Warnings and safety notes

Notes on use

  •  In order to ensure comparable values, always measure your blood pressure at the same time of day.

  • Do not take a measurement within 30 minutes after eating, drinking, smoking or exercising.

  • Before the initial blood pressure measurement, always make sure to rest for about 5 minutes!

  •  Furthermore, if you want to take several measurements in succession, make sure to always wait at least 1 minute between individual measurements.

  •  Repeat the measurement if you are unsure of the measured value.

  • The measurements taken by you are for your information only – they are not a substitute for a medical examination! Discuss the measured values with your doctor and never base any medical decisions on them (e.g. medicines and their dosages).

  • Using the blood pressure monitor outside your home environment or whilst on the move (e.g. whilst travelling
    in a car, ambulance or helicopter, or whilst undertaking physical activity such as playing sport) can influence the measurement accuracy and cause incorrect measurements.

  • Do not use the blood pressure monitor on newborns or patients with pre-eclampsia. We recommend consulting a doctor before using the blood pressure monitor during pregnancy.

  • This device is not intended for use by people (including children) with restricted physical, sensory or mental skills or a lack of experience and/or a lack of knowledge, unless they are supervised by a person who has responsibility for their safety or they receive instructions from this person on how to use the device. Supervise children around the device to ensure they do not play with it.

  • Cardiovascular diseases may lead to incorrect measure-ments or have a detrimental effect on measurement accuracy. The same also applies to very low blood pressure, diabetes, circulatory disorders and arrhythmias as well as chills or shaking.

  • The blood pressure monitor must not be used in connection with a high-frequency surgical unit.

  • The easyClip cuff included in delivery is only intended for use in combination with the BM 51 easyClip blood pressure monitor from Beurer. Do not use the cuff for other blood pressure monitors.

  • Only use the device on people that have the upper arm circumference specified for the device.

  • Please note that during inflation, the functions of the limb in question may be impaired.

  • During the blood pressure measurement, the blood circulation must not be stopped for an unnecessarily long time. If the device malfunctions remove the cuff from the arm.

  •  Avoid any mechanical restriction, compression or bending of the cuff line.

  • Do not allow sustained pressure in the cuff or frequent measurements. The resulting restriction of the blood flow may cause injury.

  •  Make sure that the cuff is not placed on an arm in which the arteries or veins are undergoing medical treatment, e. g. intravascular access or intravascular therapy, or an arteriovenous (AV) shunt.

  •  Do not use the cuff on people who have undergone a mastectomy.

Notes on storage
  •  The blood pressure monitor is made from precision and electronic components. The accuracy of the measured values and service life of the device depend on its careful handling:
    • Protect the device from impacts, humidity, dirt, marked temperature fluctuations and direct sunlight.
    • Do not drop the device.
    • Do not use the device in the vicinity of strong electromagnetic fields and keep it away from radio systems or mobile telephones.
    •  Only use the cuff included with the delivery or original replacement cuffs. Otherwise incorrect measured values will be recorded.
  • We recommend that the batteries be removed if the device is not to be used for a prolonged period of time.

Notes on handling batteries

  •  If your skin or eyes come into contact with battery fluid, rinse the affected areas with water and seek medical assistance.
  • Choking hazard! Small children may swallow and choke on batteries. Store batteries out of the reach of small children.
  •  Observe the plus (+) and minus (-) polarity signs.
  • If a battery has leaked, put on protective gloves and clean the battery compartment with a dry cloth.
  •  Protect batteries from excessive heat.
  •  Risk of explosion! Do not throw batteries into a fire.
  • Do not charge or short-circuit batteries.
  • If the device is not to be used for a relatively long period, take the batteries out of the battery compartment.
  • Use identical or equivalent battery types only.
  • Always replace all batteries at the same time.
  • Do not use rechargeable batteries.
  • Do not disassemble, split or crush the batteries.

Notes on electromagnetic compatibility

  • The device is suitable for use in all environments listed in these instructions for use, including domestic environments.
  • The use of the device may be limited in the presence of electromagnetic disturbances. This could result in issues such as error messages or the failure of the display/device.
  • Avoid using this device directly next to other devices or stacked on top of other devices, as this could lead to faulty operation. If, however, it is necessary to use the device in the manner stated, this device as well as the other devices must be monitored to ensure they are working properly.
  • The use of accessories other than those specified or provided by the manufacturer of this device can lead to an increase in electromagnetic emissions or a decrease in the device’s electromagnetic immunity; this can result in faulty operation.
  • Failure to comply with the above can impair the performance of the device.

Device description

  1. Bracket
  2.  Cuff line
  3. Cuff
  4. Cuff connector
  5.  Display
  6.  Connection for mains part (right side)
  7.  -/+ function buttons
  8. START/STOP button
  9. SET setting button
  10.  M memory button
  11.  Risk indicator
  12. Connection for cuff connector (left-hand side)

Information on the display:

  1. Date/time
  2. Systolic pressure
  3.  Diastolic pressure
  4. Calculated pulse rate
  5. Pulse symbol
  6. Battery replacement symbol
  7. Number of memory spaces / memory display for average value ( ),morning ( ), evening ( )
  8. Release air ( )
  9. Cardiac arrhythmia
  10. User memory
  11. Risk indicator

Initial use

Inserting the batteries

  • Remove the battery compartment lid on the rear of the device.
  • Insert four 1.5 V AA (alkaline type LR6) batteries. Make sure that the batteries are inserted the correct way round in accordance with the markings. Do not use rechargeable batteries.
  • Carefully close the battery compartment lid again.
  • All display elements are briefly displayed, flashes in the display. Now set the date and time as described below.

If the battery replacement symbol is permanently displayed, you can no longer perform any measurements and must replace all batteries. Once the batteries have been removed from the device, the date and time must be set again. Any saved measured values are retained.

Setting the hour format, date and time
It is essential that you set the date and time. Otherwise, you will not be able to save your measured values correctly with a date and time and access them again later.
There are two different ways to access the menu from which you can adjust the settings:

  •  Before initial use and after each time you replace the battery: When inserting batteries into the device, you will be taken to the relevant menu automatically.
  • If the batteries have already been inserted: Press and hold the SET setting button on the device for approx. 5 seconds with the device switched off.

In this menu you can adjust the following settings in succession:
Hour format            Date                    Time

  • Use the -/+ function buttons to select the relevant hour format and confirm using the SET setting button.
  • Using the -/+ function buttons, select the relevant year and confirm using the SET setting button.
  • Use the -/+ function buttons to select the relevant month and confirm using the SET setting button
  • Use the -/+ function buttons to select the relevant day and confirm using the SET setting button
  • Using the -/+ function buttons, select the relevant hour and confirm using the SET setting button.
  • Using the -/+ function buttons, select the relevant minute and confirm using the SET setting button.

Operation with the mains part
You can also operate this device with a mains part (not included in delivery). The mains part can be obtained from specialist retailers or from the service address using order number 071.95. However, before connecting the device with the mains part, please ensure that you have removed the batteries from the device. During mains operation, there must not be any batteries in the battery compartment, as this could damage the device.

  • To avoid any potential damage, the device may only be operated with a mains part that meets the specifications described in chapter “Technical specifications”.
  •  Furthermore, the mains part must only be connected to the mains voltage that is specified on the type plate.
  • Insert the mains part into the connection provided for this purpose on the right-hand side of the blood pressure monitor.
  • Then insert the mains plug of the mains part into the mains socket.
  • After using the blood pressure monitor, unplug the mains part from the mains socket first and then disconnect it from the blood pressure monitor. As soon as you unplug the mains part, the blood pressure monitor loses the date and time setting but the saved measured values are retained.

Usage

Please ensure the unit is at room temperature before measuring. The measurement can be performed on the left or right arm. Circulation in the arm must not be restricted by tight clothing or similar.

Attaching the cuff

  1. 1. Place the cuff onto the bare upper arm. Press the bracket together using your fingers, in such a way that the area where the upper arm should be placed opens.
  2.  The cuff must be placed on the upper arm so that the bottom edge is positioned 2-3 cm above the elbow and over the artery. The line should point to the centre of the palm.
  3.  Close the hook-and-loop fastener. In doing so, the cuff should be fastened so that two fingers fit under the cuff.
  4.  Now fold the protruding part down. Secure the protruding part using the additional hook-and-loop fastener on the cuff.
  5. Now insert the cuff line into the connection for the cuff connector.

Ensure that your arm is not pressing on the line.
Blood pressure may vary between the right and left arm, which may mean that the measured blood pressure values are different. Always perform the measurement on the same arm. If the values between the two arms are significantly different, please consult your doctor to determine which arm should be used for the measurement.
Important: The device may only be operated with the original cuff. The cuff is suitable for an arm circumference of 22 to 42 cm.

Adopting the correct posture

  •  Before the initial blood pressure measurement, always make sure to rest for about 5 minutes! Otherwise deviations can occur.
  • You can take the measurement while sitting or lying. Always make sure that the cuff is at heart level.
  •  To carry out a blood pressure measurement, make sure you are sitting comfortably, with your arms and back leaning on something. Do not cross your legs. Place your feet flat on the ground.
  • To avoid falsifying the measurement, it is important to remain still during the measurement and not to speak.
  •  Wait at least 1 minute before taking another measurement!

Selecting memory
Press the SET setting button. Select the relevant user memory ( or 2 ) by pressing the -/+ function buttons.
You have 2 memories, each with 100 memory spaces, to enable the separate storage of measurements for 2 different people. Confirm your selection with the START/STOP button
or wait for 3 seconds. Your selection is then automatically stored.
Performing the blood pressure measurement
As described above, attach the cuff and adopt the posture in which you want to perform the measurement.

  • Press the START/STOP button to start the blood pressure monitor. All display values light up briefly. After the full-screen display, the last stored measurement ap-pears. If there is no measurement in the memory, the unit displays the value .
    The blood pressure monitor will begin measurement automatically after 5 seconds.
    The cuff automatically inflates. The measurement itself is taken during the inflation phase. As soon as a pulse is found, the pulse symbol will be displayed.
    You can cancel the measurement at any time by pressing the START/STOP button .

  • Systolic pressure, diastolic pressure and pulse measurements are displayed.

  • appears if the measurement has not been performed properly. Refer to chapter “10. What if there are problems?” in these instructions for use and repeat the measurement.

  • Press the START/STOP button to switch off the blood pressure monitor. The measurement is then stored in the selected user memory.
    If you forget to turn off the device, it will switch off automatically after approx. 1 minute. Even in this event, the value is stored in the selected or most recently used user memory.

  •  Wait at least 1 minute before taking another measurement!

Evaluating results
Cardiac arrhythmia:
This device can identify potential disruption of the heart rhythm when measuring and, if necessary, indicates this after the measurement with the symbol .
This can be an indicator for arrhythmia. Arrhythmia is a condition in which the heart rhythm is abnormal because of flaws in the bioelectrical system that regulates the heartbeat. The symptoms (skipped or premature heart beats, pulse being slow or too fast) can be caused by factors such as heart disease, age, physical make-up, excess stimulants, stress or lack of sleep. Arrhythmia can only be determined through an examination by your doctor.
If the symbol is shown on the display after the measure-ment has been taken, repeat the measurement. Please ensure that you rest for 5 minutes beforehand and do not speak or move during the measurement. If the symbol frequently, appears please consult your doctor. Self-diagnosis and treatment based on the measurements can be dangerous. Always follow your doctor’s instructions.
Risk indicator:
The measurements can be classified and evaluated in accordance with the following table. However, these standard values serve only as a general guideline, as the individual blood pressure varies in different people and different age groups etc. It is important to consult your doctor regularly for advice. Your doctor will tell you your individual values for normal blood pressure as well as the value above which your blood pressure is classified as dangerous.The bar chart on the display and the scale on the device show which category the recorded blood pressure values fall into. If the values of systole and diastole fall into two different categories (e.g. systole in the High normal category and diastole in the Normal category), the graphical classification on the device always shows the higher category; for the example given this would be High normal.

Blood pressure value category| Systole

(in mmHg)

| Diastole

(in mmHg)

| Action
---|---|---|---
Level 3: severe hypertension|

red

|

≥ 180

|

≥ 110

| Seek medical attention
Level 2: moderate hypertension|

orange

|

160 – 179

|

100 – 109

| Seek medical attention
Level 1: mild hypertension|

yellow

|

140 – 159

|

90 – 99

| Regular monitoring by doctor

High normal

|

green

|

130 – 139

|

85 – 89

| Regular monitoring by doctor
Normal| green| 120 – 129| 80 – 84| Self-monito- ring
Optimal| green| < 120| < 80| Self-monito- ring

Saving, displaying and deleting measured values

The results of every successful measurement are stored together with the date and time. If there are more than 100 measurements, the oldest measurements are lost.

User memory

  • Press the SET setting button. Select the relevant user memory ( or 2 ) by pressing the -/+ function buttons. Confirm your selection using the START/STOP button
  • Press the M memory button.lights up on the display.

The average value of all saved measured values in this user memory is displayed.

Average values

  • Press the M memory button. lights up on the display.
    The average value of the morning measurements for the last 7 days is displayed (morning: 5 a.m. – 9 a.m.).

  • Press the M memory button.
    lights up on the display.

The average value of the evening measurements for the last 7 days is displayed (evening: 6 p.m. – 8 p.m.).

  • Press the M memory button.
    lights up on the display.
    The average value of all saved measured values in this user memory is displayed.

  • Press the M memory button.
    lights up on the display.
    The average value of the morning measurements for the last 7 days is displayed (morning: 5 a.m. – 9 a.m.).

  • Press the M memory button.
    lights up on the display.
    The average value of the evening measurements for the last 7 days is displayed (evening: 6 p.m. – 8 p.m.).

Individual measured values
To delete the memory of the selected user memory, press the M memory button in switched-off state. The average value of all measurements appears on the display; in parallel, lights up.

  • Press and hold down both function buttons -/+ simultaneously for 5 seconds. appears on the display. All the values in the selected user memory have now been deleted.

  • When the M memory button is pressed again, the last individual measured values in each case are displayed with the date and time (for example measurement 03).
    When the memory button M is pressed again, you can view your individual measured values.
    To switch the device off again, press the START/STOP button . If you forget to switch off the unit, it will switch off automatically after 1 minute. You can exit the menu at any time by pressing the START/STOP button .

  • To delete the memory of the selected user memory, press the M memory button in switched-off state. The average value of all measurements appears on the display; in parallel, lights up.
    Press and hold down both function buttons -/+ simultaneously for 5 seconds.
    appears on the display. All the values in the selected user memory have now been deleted.

Cleaning and maintenance

  •  Clean the device and cuff carefully using only a slightly damp cloth.
  • Do not use any cleaning agents or solvents.
  • Under no circumstances hold the device and cuff under water, as this can cause liquid to enter and damage the device and cuff.
  • If you store the device and cuff, do not place heavy objects on the device and cuff. Remove the batteries. The cuff line should not be bent sharply.

Accessories and replacement parts

Accessories and replacement parts are available from the corresponding service address (according to the service address list). Please state the corresponding order number.

Designation Item number and/or order number
easyClip universal cuff (22–42 cm) 164.183
Mains part (EU) 071.95
Mains part (UK) 072.05

What if there are problems

Error message Possible cause Solution
Unable to record a pulse. Please wait one minute and repeat the measure-

ment.

Ensure that you do not speak or move during the measurement.

| The cuff was not attached correctly.| Please observe the infor- mation in chapter 7 under the heading “Attaching the cuff”.
| An error occurred during measure- ment.| Please wait one minute and repeat the measure- ment.

Ensure that you do not speak or move during the measurement.

| An error occurred while pumping up the cuff.| Please take another measurement to check whether the cuff can be correctly inflated.

Make sure that neither your arm nor other heavy objects are pressing on the line, and that the line is not bent.

Error message Possible cause Solution
**** System error Please contact customer service.
**** The batteries are almost empty. Insert new batteries into the device.

Disposal

  • Do not repair or adjust the device yourself. Proper operation can no longer be guaranteed in this case.
  • Do not open the device. Failure to comply will invalidate the warranty.
  • Repairs must only be carried out by Customer Services or authorised retailers. Before making a claim, please check the batteries first and replace them if necessary.
  • For environmental reasons, do not dispose of the device in the household waste at the end of its useful life. Dispose of the device at a suitable local collection or recycling point in your country.
  • Dispose of the device in accordance with EC Directive – WEEE (Waste Electrical and Electronic Equipment). If you have any questions, please contact the local authorities responsible for waste disposal.

Disposing of the batteries

  • Empty, completely flat batteries must be disposed of through specially designated collection boxes, recycling points or electronics retailers. You are legally required to dispose of the batteries.
  • The codes below are printed on batteries containing harmful substances:
    Pb = Battery contains lead, Cd = Battery contains cadmium, Hg = Battery contains mercury.

Technical specifications

Device

Model no.                                     BM 51
Type                                             BPM51
Measurement method                  Oscillometric, non-invasive blood pressure measurement on the upper arm
Measurement range                      Cuff pressure 0–300 mmHg, systolic 60–255 mmHg, diastolic 40–200 mmHg, pulse 40 –199 beats/minute
Display accuracy                           Systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement uncertainty              Max. permissible standard deviation
according to clinical testing:          Systolic 8 mmHg / diastolic 8 mmHg

Memory                                         2 x 100 memory spaces
Dimensions                                   L 138 mm x W 103 mm x H 44 mm
Weight                                           Approximately 571 g (without batteries, with cuff)
Cuff size                                        22 to 42 cm
Permissible operating                    + 10 °C to + 40 °C, ≤ 85 % relative
conditions                                      humidity (non-condensing), 700 hPa–1060 hPa environmental
pressure Permissible storage        – 20 °C to + 50 °C, ≤ 85 % relative
conditions                                     humidity (non-condensing), 700 hPa–1060 hPa environmental pressure

Power supply                                 4 x 1.5 V AA batteries
Battery life                                     For approx. 500 measurements, depending on levels of blood pressure and inflation pressure
Classification                                 Internal supply, IP22, no AP or APG, continuous operation, application part type BF
The serial number is located on the device or in the battery
compartment.
Technical information is subject to change without notification to allow for updates.

  • This device complies with the European standard EN 60601-1-2 (in compliance with CISPR 11, IEC 61000-4-2, IEC 61000-4-3, and IEC 61000-4-8) and is subject to special precautionary measures with regard to electromagnetic compatibility. Please note that portable and mobile HF communication systems may interfere with this device.

  • This device complies with the EU Medical Devices Directive 93/42/EEC, the German Medical Devices
    Act (Medizinproduktgesetz) and the standards
    EN 1060-1 (Non- invasive sphygmomanometers –

  • The accuracy of this blood pressure monitor has been carefully checked and developed with regard to a long useful life. If the device is used for commercial medical purposes, it must be regularly tested for accuracy by appropriate means. Precise instructions for checking accuracy may be requested from the service address.
    Part 1: General requirements), EN1060-3 (Non-invasive sphygmomanometers – Part 3: Supplementary requirements for electro-mechanical blood pressure measuring systems) and IEC IEC 80601-2-30 (Medical electrical equipment – Part 2 – 30: Particular requirements for the basic safety
    and essential performance of automated non-invasive sphygmomanometers).

Model no.                                               LXCP12-006060BEH
Input                                                      100 – 240 V, 50 – 60 Hz, 0.5 A max
Output                                                   6 V DC, 600 mA, in conjunction with Beurer blood pressure monitors only
Manufacturer                                          Shenzhen Iongxc power supply co., ltd.
Protection                                               The device is double protected and has a primary-side cutout switch that disconnects the device from the mains in case of malfunction. Ensure that you have removed the batteries from the battery compartment before you use the mains part. The polarity of the DC voltage connection Insulated/protection class 2
Housing and
protective covers   The housing of the mains part protects users from touching live parts or parts that could be live (for example with finger, needle, checking hook).
Do not touch the patient and the output connector of the AC/DC mains part at the same time.

Warranty/service

Beurer GmbH, Söflinger Straße 218, 89077 Ulm, Germany (hereinafter referred to as “Beurer”) provides a warranty for this product, subject to the requirements below and to the extent described as follows.

The warranty conditions below shall not affect the seller’s statutory warranty obligations which ensue from the sales agreement with the buyer.The warranty shall apply without prejudice to any mandatory statutory provisions on liability. Beurer guarantees the perfect functionality and completeness of this product. The worldwide warranty period is 5 years, commencing from the purchase of the new, unused product from the seller. The warranty only applies to products purchased by the buyer as a consumer and used exclusively for personal purposes in the context of domestic use.

German law shall apply. During the warranty period, should this product prove to be incomplete or defective in functionality in accordance with the following provisions, Beurer shall carry out a repair or a replacement delivery free of charge, in accordance with these warranty conditions.
If the buyer wishes to make a warranty claim, they should approach their local retailer in the first instance: see the at- tached “International Service” list of service addresses.The buyer will then receive further information about the pro- cessing of the warranty claim, e.g. where they can send the product and what documentation is required.

A warranty claim shall only be considered if the buyer can pro- vide Beurer, or an authorised Beurer partner, with

  • a copy of the invoice/purchase receipt, and
  • the original

The following are explicitly excluded from this warranty:

  • deterioration due to normal use or consumption of the product;
  • accessories supplied with this product which are worn out or used up through proper use (e.g. batteries, rechargeable batteries, cuffs, seals, electrodes, light sources, attachments and nebuliser accessories);
  • products that are used, cleaned, stored or maintained improperly and/or contrary to the provisions of the instructions for use, as well as products that have been opened, repaired or modified by the buyer or by a service centre not authorised by Beurer;
  • damage that arises during transport between manufacturer and customer, or between service centre and customer;
  • products purchased as seconds or as used goods;
  • consequential damage arising from a fault in this product (however, in this case, claims may exist arising from product liability or other compulsory statutory liability provisions).

Repairs or an exchange in full do not extend the warranty period under any circumstances.