BRESSER 5760100 Science TRM 301 Microscope Instruction Manual
- June 6, 2024
- BRESSER
Table of Contents
- BRESSER 5760100 Science TRM 301 Microscope
- Product Usage Instructions
- GENERAL WARNINGS
- SCOPE OF DELIVERY
- CHOICE OF LOCATION & SETUP
- PRODUCT DESIGN SCIENCE TRM 301
- MICROSCOPE SETTINGS
- ADJUSTING THE LIGHTING
- ADJUSTING THE OBSERVATION TUBE
- FOCUSING THE SPECIMEN
- CENTRING THE CONDENSER
- APERTURE DIAPHRAGM (CONDENSER DIAPHRAGM)
- USE OF OIL IMMERSION
- LAMP REPLACEMENT
- CLEANING/MAINTENANCE/TRANSPORT
- TECHNICAL DATA
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
BRESSER 5760100 Science TRM 301 Microscope
Product Specifications
- Model: SCIENCE TRM-301
- Type: Trinocular transmitted light microscope
- Art. No.: 5760100
- Power Supply: 12V, 4A
- Input Voltage: 110-240V
- Light Source: 12V/30W Halogen Lamp
Product Usage Instructions
Location Selection & Assembly
Choose a suitable location for the microscope and assemble the components
accordingly.
Product Design – SCIENCE TRM 301
Refer to the user manual for a detailed view of the components and design
of the microscope.
Microscope Settings
- Establish Power Connection: Follow the instructions to connect the power supply.
- Adjust Illumination: Set the lighting according to your requirements.
- Adjust Observation Tube: Set the interpupillary distance and diopter adjustment.
Centering the Condenser
Follow the steps provided in the manual to center the Abbe condenser
properly.
Aperture Diaphragm Adjustment
Adjust the aperture diaphragm to set it between 70-80% for optimal viewing.
Lamp Replacement
- Turn off the main switch and unplug the microscope from the power source.
- Remove the old halogen lamp and replace it with a new one as instructed.
Cleaning/Maintenance/Transportation
Be cautious while cleaning as alcohol is flammable. Follow the maintenance
instructions provided in the manual.
Technical Data
Refer to the technical specifications section for detailed information on
the product.
Warranty
For warranty information, visit the provided link:
www.bresser.de/warranty_terms
Frequently Asked Questions (FAQ)
-
Q: How do I adjust the illumination on the microscope?
A: Refer to section “Microscope Settings” in the user manual for instructions on adjusting the illumination. -
Q: What should I do if the lamp needs replacement?
A: Follow the steps outlined in the “Lamp Replacement” section of the user manual to replace the lamp safely.
Visit our website via the following QR Code or web link to find further information on this product or the available translations of these instructions.
WARRANTY
IMPRINT
Bresser GmbH, Gutenbergstr. 2, 46414 Rhede, Germany
www.bresser.de
In the event of any warranty claims or service enquiries, please refer to the
“Warranty” and “Service” sections of these operating instructions. We ask for
your understanding that unsolicited returns cannot be processed.
Errors and technical changes reserved.
© 2024 Bresser GmbH. All rights reserved.
Reproduction of these operating instructions – in whole or in part – in any
form (e.g. photocopying, printing, etc.) as well as use and distribution by
means of electronic systems (e.g. image file, website, etc.) requires the
prior written consent of the manufacturer and is otherwise prohibited.
The designations and brand names of the respective companies used in these
operating instructions are protected by commercial, trademark and/or patent
law in Germany, the European Union and/or other countries.
VALIDITY NOTE
These operating instructions are valid for products with the following
article numbers: 5760100
Manual version: 0324
Manual name: Manual_5760100_Science-TRM-301_de-en_BRESSER_v032024a
If you have any queries, please quote this information.
ABOUT THESE OPERATING INSTRUCTIONS
NOTE
These operating instructions are to be regarded as part of the device.
Read the safety instructions and the operating instructions carefully before
using this device. Keep these instruction manual in a safe place for future
reference. If the device is sold or passed on, the operating instructions must
be handed over to the new owner/user of the product.
GENERAL WARNINGS
DANGER OF AN ELECTRIC SHOCK!
This device contains electronic parts that are powered by a power source
(mains connection). Never leave children unattended when handling the device!
Only use the device as described in the instructions, otherwise there is a
RISK of an ELECTRIC SHOCK!
- Disconnect the power supply when you are not using the device, during long breaks, and before all maintenance and cleaning work (disconnect the mains plug).
- Place the device so that it can be disconnected from the power supply at any time. The power outlet should always be near the device and should be easily accessible, as the plug of the power cord serves as a part to disconnect the device from the mains supply.
- To disconnect the device from the mains, always pull the mains plug and never the cable!
- Check the device, cables and connections for damage before use.
- Do not use the device if is damaged or contains damaged live parts! Damaged parts must be replaced immediately by an authorized service center.
- Only operate the device in a completely dry environment and do not touch the device with wet or damp body parts.
DANGER OF SUFFOCATION!
Improper use can lead to a RISK OF SUFFOCATION, especially for children! It is
therefore imperative that you observe the following safety information.
- Keep packaging materials (plastic bags, rubber bands, etc.) away from children!
- This product contains small parts that can be swallowed by children!
- If small parts are swallowed, seek medical advice immediately!
FIRE/EXPLOSION HAZARD!
Improper use of this product poses a RISK OF FIRE/EXPLOSION! Always observe
the following safety instructions to avoid fires and/or explosions.
- Do not expose the device to high temperatures. Only use the supplied mains adapter (12V, 4A). Do not short-circuit the device and batteries or throw them into a fire! Excessive heat and improper handling can cause short circuits, fires and even explosions!
- Never cover the ventilation slots or cooling fins of the device during operation or until it has cooled down completely!
RISK OF INJURY!
This device contains components/accessories that can cause minor or serious
injuries if used incorrectly. It is therefore essential that you observe the
following safety information in order to prevent body injuries.
- This device often requires the use of sharp-edged and pointed tools. For this reason, keep this device together with all accessories and tools in a place inaccessible to children!
- Keep chemicals and liquids out of reach of children! Don’t drink! Clean hands thoroughly under running water after use. In case of accidental contact with eyes or mouth, rinse with water. In case of complaints, consult a doctor immediately and show the substances.
THIS DEVICE GENERATES HEAT!
This device contains a halogen light source (12V/30W). Halogen lamps emit a
considerable amount of infrared light as heat. As a result, the surface of the
lamp housing becomes very hot. Leave sufficient space around the microscope
and DO NOT place it near flammable materials. Important: The heat generated
can also cause the specimens to warm up! Adjust the brightness to your
requirements: Set the Koehler illumination on the microscope by illuminating
only the part of the sample that you want to inspect. Switch off the light
when you are not using the microscope.
DANGER OF PROPERTY DAMAGE!
Improper handling can cause DAMAGE to the device and/or the accessories!
Therefore, use the device only in accordance with the following safety
information.
- Do not disassemble the device! In case of a defect, contact your dealer. The dealer will contact the service center and send the device for repair if necessary.
- Do not expose the device to high temperatures and protect it from water and high humidity!
- Do not subject the device to shocks!
- Only use accessories and spare parts that comply with the technical specifications!
SCOPE OF DELIVERY
- A: Transmitted light microscope
- B: 2 pcs. Eyepieces (WF 10x)
- C: 4 pcs. Objectives (4x, 10x, 40x, 100x oil / pre-assembled)
- D: Coaxial cross table with vernier adjustment (pre-assembled)
- E: Power supply unit (12V, 4A)
- F: Blue colour filter disc
- G: Immersion oil
- H: Dust cover
CHOICE OF LOCATION & SETUP
Before setting up your microscope, choose a suitable location.
Place the microscope on a surface that meets the following criteria:
- Flat/level
- Vibration-free/stable
- Hard, non-combustible surface
- Chemically and mechanically resistant
Set up the microscope so that …
- … the ventilation slots on the back are not blocked.
- … there is a distance of at least 10 cm around the device or to other appliances or the wall.
- … you can disconnect it from the power supply at any time.
ATTENTION! Always use the supplied power adapter (12V, 4A) for the halogen illumination (12V/30W) and dimmer. Using an unsuitable power supply unit can cause damage to the microscope!
PRODUCT DESIGN SCIENCE TRM 301
FRONT/SIDE VIEW
Fig. 1A: Components on TRM 301, front/side view.
EQUIPMENT
-
Eyepiece
-
Objective nosepiece
-
Stage (cross table) with Vernier scales
-
Microscope slide with sample
-
Blue filter (included in the scope of delivery)
-
6A. Field diaphragm with filter holder
6B. Field diaphragm ring -
Microscope base with lamp housing
-
Limit knob (up-stop)
-
Right coarse focusing knob
-
Right fine focusing knob
-
X-axis knob
-
Y-axis knob
-
Clamp holder
-
Objective
-
Microscope body
-
Microscope head holding screw
-
Trinocular tube
-
Connection of an external camera possible while using a suitable adapter (not included in the scope of delivery)
REAR/SIDE VIEW Power supply: power adapter 12V, 4A
Input voltage: 110-240V
Lighting: 12V/30W Halogen lamp
Fig. 1B: Components on TRM 301, rear/side view.
EQUIPMENT -
Ventilation slots
-
Mains connection
-
ON/OFF switch (main switch)
-
Dimmer
-
Left fine focusing knob
-
Left coarse focusing knob
-
Tension adjustment ring
-
Condenser up-down knob
-
27A. Condenser
27B. Aperture diaphragm lever
27C. Auxiliary condenser lens -
Condenser centering screws
-
Condenser holding screw
MICROSCOPE SETTINGS
A ESTABLISHING A MAINS CONNECTION
ATTENTION! Before setting up the microscope, check that the ON/OFF switch
(main switch) (1) is in the O (OFF) position.
To start up the microscope, first insert the hollow plug (2) on the supplied
power supply unit (12V, 4A) (3) into the mains connection (4) on the back of
the microscope. Then plug the mains plug (5) on the power supply unit into a
suitable socket (6) (max. 230V).
To switch on the microscope, move the ON/OFF switch (main switch) (1) to the
‘I’ position.
PLACING THE SPECIMEN
The clamp holder (1) of the cross table is designed to hold a microscope
slide (2). The Y-axis knob (3) (for movement in the longitudinal direction)
and the X-axis knob (4) (for movement in the transverse direction) are
coaxial. The stage (cross table) can therefore be moved according to your
requirements. Swivel the 4x objective (5) into the working position by turning
the nosepiece (6).
NOTE: The objectives engage with an audible “click” when they are
positioned in the beam path.
ADJUSTING THE LIGHTING
The lamp housing (1) is located in the microscope base.
The light is supplied by a halogen bulb (12V/30W).
You can adjust the brightness of the light source using the stepless dimmer
(2).
WARNING: Do not leave the dimmer at maximum brightness for long periods
of time. This shortens the lifespan of the lamp.
ADJUSTING THE OBSERVATION TUBE
Adjusting the interpupillary distance (A) (distance between the
eyepieces)
Set the correct interpupillary distance for binocular observation by
swivelling the eyepiece tubes (1) on the microscope head so that the right and
left fields of view overlap completely or combine to form a circular image.
Setting the visual acuity correction ( diopter compensation)
Turn the 40x objective to the working position. First, look only through the
right eyepiece with your right eye, and then use the coarse and fine focusing
knobs to focus the image. Next, look only through the left eyepiece with your
left eye, and then use the diopter adjustment ring (2) to focus the image.
FOCUSING THE SPECIMEN
Next, bring the specimen into focus and adjust the image sharpness
(focus).
The microscope has a coaxial coarse and fine focusing mechanism. Turn the
coarse focusing knob (1) to raise or lower the cross table. To make precise
adjustments to the image sharpness, use the fine focusing knob (2). The
tension adjustment ring (3) next to the coarse focusing knob (1) is used to
adjust the ease of movement of the coarse focusing knob so that the object
does not slide down by itself. The limit knob (up-stop) (4) prevents
unintentional contact between the specimen and objective when focussing. If
you lock it at a specified height (by turning it upwards), you will not be
able to use the focusing knob drive to bring the objective closer to the
specimen, which is thus protected against damage.
NOTE: The supplied blue filter (5) can be inserted into the filter holder
of the field diaphragm. The blue filter makes the light colour more neutral.
- The condenser has an auxiliary lens (6). This lens is used to adjust the condenser aperture to the objective aperture. To observe with the 4x objective, swivel the auxiliary condenser lens into the beam path. This is used to illuminate the field of view. When observing with higher magnification objectives, swivel the auxiliary condenser lens out of the beam path.
- Always start the observation with the lowest magnification: This makes it easier to center and adjust the specimen and also protects the microscope from damage.
WARNING: The halogen lamp generates a significant amount of heat and will cause the specimens to heat up! Adjust the brightness to your requirements: Set the Koehler illumination on the microscope by illuminating only the part of the sample that you want to inspect. Switch off the light when you are not using the microscope.
TRINOCULAR TUBE FOR THE PHOTOGRAPHY
The trinocular tube (1) offers the option of connecting a MicroCam or C-Mount
camera. To connect them, a C-Mount MicroCam adapter is needed (item number:
5942101). An optional adapter (item number: 5942100) for SLR or system cameras
is also available. In addition, a camera-specific T2 ring is required here.
The adapter holding screw (2) fixes the appropriate adapter/dust cap.
KOEHLER ILLUMINATION PRINCIPLE
This special technology ensures uniformly bright, high-contrast and
reflection-free illumination of the sample. In addition to making full use of
the microscope optics, it also protects sensitive samples from excessive light
irradiation. The illumination according to Koehler is therefore the standard
for scientific and photomicrography applications, and can be adjusted with the
built-in field diaphragm together with the height-adjustable and centerable
Abbe condenser (numerical aperture 1.25; with iris diaphragm).
CENTRING THE CONDENSER
Height-adjustable and centerable Abbe condenser
The condenser (1) can be moved up or down by turning the condenser up-down
knob (2). Use the condenser centring screws (3) to center the condenser. After
loosening the condenser holding screw (4), the condenser can be easily removed
from its holder. The field diaphragm and filter holder (5) has space for a
filter disc (a blue filter is included in the scope of delivery).
The following section describes step-by-step how to set the Koehler illumination using the 10x objective as an example:
- Use the condenser up-down knob (2) to bring the condenser (1) to the highest position directly under the stage.
- Switch on the light with the ON/OFF switch (6) and focus a high-contrast sample using the 10x objective. Make sure that the light is set correctly using the dimmer (7).
- NOTE: When centring with the 10x objective, swivel the auxiliary condenser lens (8) out of the beam path, otherwise you will not be able to center the condenser.
- Turn the field diaphragm ring (9) in the direction of the arrow so that the field diaphragm (5) is completely closed. Look through the eyepieces. Adjust the condenser up-down knob (2) until the blurred image of the field diaphragm is in focus.
- If the condenser is not centered, you will see an image of the field diaphragm similar to (10A).
- To center the aperture in the field of view, use the two condenser centring screws (3) on each side of the condenser holder.
- Once the image is in focus and centered in the field of view (10B), open the field diaphragm until its edge has just disappeared from the field of view and the field of view is completely illuminated (10C-10D).
If needed, you may be able to re-center the condenser a little.
You should also set the aperture diaphragm, which is explained in the next
section.
Fig. 8: Centering the Abbe condenser.
NOTE: Bright-field microscopy is the most common of all light microscopy
methods, as it allows high-contrast or coloured specimens (e.g. blood smears)
to be viewed quickly and easily. In order to fully utilise the optical
capabilities of the microscope (especially the objective), the condenser,
field diaphragm and aperture diaphragm should be configured according to the
Koehler illumination principle.
- The field diaphragm limits the diameter of the light beam entering the objective. This prevents stray light and increases the image contrast. To support the performance of the objective, the diameter of the field diaphragm must be adjusted so that its image and the field of view are the same size.
APERTURE DIAPHRAGM (CONDENSER DIAPHRAGM)
The aperture diaphragm (condenser diaphragm) determines the numerical aperture of the illumination system as well as the resolution of the microscope. It is used to set the contrast, image resolution and depth of field. Closing the aperture diaphragm increases the depth of field. The aperture diaphragm (condenser diaphragm) (1) can be opened or closed using the aperture diaphragm lever (2).
- Remove one eyepiece and look into the open eyepiece tube.
- Adjust the aperture diaphragm image (3) by adjusting the aperture diaphragm lever (2) to approx. 70-80% of the diameter of the objective pupil (4).
- Then insert the eyepiece back into the eyepiece socket.
NOTE: Each time the objective is changed, the object field size and the
objective aperture and possibly the centring will change slightly, meaning
that the light field and aperture diaphragm settings should be readjusted for
optimum results.
WARNING: The brightness can only be adjusted via the dimmer, and not via
the aperture diaphragm.
USE OF OIL IMMERSION
- Focus the object with the 40x objective (1).
- Move the Y-axis knob (2) to expose the condenser lens (3). Turn the 40x objective (1) to the side.
- Apply a drop of immersion oil (4) to the lens of the condenser (3).
- Move the slide (5) back into the beam path using the Y-axis knob (2).
- Put another drop of immersion oil (4) on the cover glass of the slide (5) and screw in the 100x objective (6), immersing the front lens in the oil.
- Slowly refocus the image using the fine focusing knob (7).
NOTE: Oil immersion can only be performed with a suitable 100x objective. The 100x oil immersion objective is included in the scope of delivery.
- We recommend immersion oil with a refractive index (RI) of nD=1.515 in accordance with ISO 8036.
NOTE: As air bubbles in the oil impair the image quality, make sure that the oil drop is free of bubbles. To remove bubbles, defocus the oil immersion objective repeatedly and refocus it, or switch to another objective and allow the visible bubbles to burst before immersing the 100x objective again.
- Clean the 100x objective by dabbing off the oil with a soft baby paper towel (lotion-free). If you use the immersion objective regularly, the lens should only be cleaned with a damp cloth once a week. Use a solution of 70% ethanol or isopropanol (to prepare such a solution, use p.A or pharmacy alcohol and double-distilled water).
WARNING: If the immersion oil comes into contact with skin, rinse it off thoroughly with soap and water. In the event of eye contact, rinse the eyes under running water for at least 15 minutes.
LAMP REPLACEMENT
-
Switch off the ON/OFF switch (main switch) (1), disconnect the mains plug from the mains socket, and then disconnect the hollow plug (2) from the mains connection (3) on the microscope. Remove the blue filter from the field diaphragm.
-
Carefully tilt the microscope and loosen the fastening screw (4) on the bottom flap (5), which carries the lamp holder and is located in the middle of the underside of the microscope. Open the bottom flap (5).
-
Pull the old halogen bulb (6) (12V/ 30W) out of the lamp holder (7). WARNING: Exercise caution, as the lamp may still be hot!
-
Plug a suitable replacement halogen bulb (6) (12V/ 30W) into the lamp holder (7).
NOTE: Take care not to touch the lamp directly with your fingers (to prevent soiling of the lamp, which reduces its lifespan and intensity). For example, grip the lamp using the lamp’s protective cover or a handkerchief. -
Close the bottom flap (5) on the underside again with the fastening screw (4).
-
After you have fitted the replacement lamp, reconnect the power supply and press the ON/OFF switch (main switch) (1). Then swivel an objective into the beam path, adjust the condenser to the correct height using the condenser up-down knob (8), and illuminate the field of view. If the light spot shifts away from the center of the field of view, loosen the screw (9) slightly and move the lamp holder (7). Guide the light spot into the center, then immediately tighten the screw (9) again.
CLEANING/MAINTENANCE/TRANSPORT
During long breaks, and before all maintenance and cleaning work, disconnect
the power supply by unplugging the mains plug.
Remove all interchangeable optical elements (e.g. objectives, eyepieces, etc.)
before cleaning.
The nosepiece, the coarse/fine focusing mechanism and the condenser are
precisely designed and calibrated to work together. Please avoid dismantling
the components. This impairs the performance of the microscope.
CLEANING
Lens cleaning:
Wipe the lenses with a lens cleaning cloth or a soft cloth moistened with a
small amount of alcohol or alcohol/ether mixture. Clean the immersion oil from
the 100x immersion objective and the condenser lens after use.
Cleaning of microscope and optical elements:
Only clean the outside of the microscope with a dry cloth.
First, blow loose dust off the lens surfaces.
Use high quality lens cleaning towels or soft cloth and moisten it with a
little bit of pure alcohol (available in a drug store). Gently wipe the lens
surface in a circular motion from the center to the edge of the lens, applying
a small amount of pressure.
WARNING: Do not use organic solvents (e.g. alcohol, ether, acetone,
xylene or other thinners) to clean painted or plastic parts!
- Alcohol is highly flammable.
- To avoid damaging the electronics, do not use any cleaning fluid.
Cleaning the 100x oil objective/condenser lens
The immersion oil should be removed from the objective/condenser lens at the
end of each working day.
Clean the 100x objective (oil) and the condenser lens by dabbing off the oil
with a soft baby paper towel (lotion-free). If you use the 100x objective with
oil immersion on a regular basis, it is sufficient to clean it once a week
with a solution of 70 % ethanol or isopropanol (to prepare such a solution,
use p.A or standard pharmacy alcohol and double-distilled water).
Dust protection:
When not in use, cover the microscope with the dust cover and place it in a
dry and mould-free place. We recommend storing all objectives and eyepieces in
a closed container with desiccant.
Transport:
Remove all loose components and the specimen from the cross table. Only use
the original packaging for transportation.
TECHNICAL DATA
-
Transmitted light microscope
Microscope head: Trinocular view, tilted by 30°, 360° rotatable -
2x eyepieces: both wide-field eyepieces 10x (23 mm plug-in diameter) Field number (FN): 20 mm, (in 30° inclined eyepiece tube)
-
4x objectives: 4x, 10x, 40x, 100x oil spring-loaded (objective type: planachromatic)
-
Condenser: Abbe condenser, height-adjustable and centerable (numerical aperture 1.25 with iris diaphragm)
-
Stage: The movement range of the cross table is 50 mm longitudinal and 75 mm transverse; coaxial cross table with vernier scales
-
Fine focusing knob: Smallest scale division: 0.002 mm
-
Adjustment range for interpupillary distance: 53 mm – 75 mm
-
Light source: Halogen lamp (12V/30W) with dimmer
-
Replacement halogen lamp (12V/30W)
DISPOSAL
Do not dispose of electrical appliances with household waste! In accordance
with Directive 2002/96/EC of the European Parliament on waste electrical and
electronic equipment and its transposition into German law, used electrical
appliances must be collected separately and recycled in an environmentally
friendly manner.
Please observe the current legal regulations when disposing of the microscope
and all accessories! Information on proper disposal can be obtained from
municipal waste disposal service providers or the Environmental Agency.
WARRANTY
The regular guarantee period is 2 years and begins on the day of purchase. To
benefit from an extended voluntary guarantee period as stated on the gift box,
registration on our website is required.
You can consult the full guarantee terms as well as information on extending
the guarantee period and details of our services at
www.bresser.de/warranty_terms.
EC DECLARATION OF CONFORMITY
Bresser GmbH has prepared a “CE Declaration of Conformity” in accordance with
the applicable guidelines and corresponding standards. The full text of the CE
declaration of conformity is available at the following website:
www.bresser.de/download/5760100/CE/5760100_CE.pdf
SERVICE
Please contact the service centre first for any questions regarding the
product or claims, preferably by e-mail.
E-Mail: service@bresseruk.com
*Telephone:* +44 1342 837 098
BRESSER UK Ltd.
Suite 3G, Eden House
Enterprise Way
Edenbridge, Kent TN8 6HF
Great Britain
Number charged at local rates in the UK (the amount you will be charged per phone call will depend on the tariff of your phone provider); calls from abroad will involve higher costs.
Bresser GmbH Gutenbergstraße 2 46414 Rhede · Germany