ResMed Power Station II External Lithium-Ion Battery User Guide
- June 6, 2024
- ResMed
Table of Contents
- ResMed Power Station II External Lithium-Ion Battery
- Power Station II
- Intended use
- Compatible devices and accessories
- Control panel
- Setting up
- Charging the battery
- Using two batteries
- Indicators
- Using humidifiers
- Cleaning and maintenance
- Storing
- Servicing
- Traveling
- Troubleshooting
- Technical specifications
- Symbols
- Disposal
- General warnings and cautions
- Limited warranty
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
ResMed Power Station II External Lithium-Ion Battery
Power Station II
Battery pack
Illustrations only, parts may differ / Illustrations uniquement,
a DC plug / Fiche CC / Enchufe de CC / Plugue CC
DC output voltage & DC cables
NOTE
-
Not all devices are available in all regions. VPAP III ST-A with QuickNav and VS III are
not available in the U.S. -
To recharge the ResMed Power Station II using an Air 10 power supply unit (supplied with Air 10 devices) or the Air 10 DC converter, an Air 10 power supply unit adaptor is required (part no. 37342).
-
For information on battery run, times refer to the Battery/Compatibility List on www.resmed.com.
Intended use
The ResMed Power Station II (RPS II) is an external lithium-ion battery that
provides power when mains power is unavailable.
Please read the entire guide before using the RPS II. Refer to the device user
guides for intended patients, uses, an environment where used, and
contraindications associated with CPAP, bi-level, and ventilation therapy.
RPS II at a glance
Refer to illustration A.
The RPS II system comprises the following components:
-
Battery
-
Power supply unit adapter
-
Carry bag
Available separately: -
90W AC power supply unit or DC converter
-
AC or DC power cord
-
DC output cable
-
Air 10 Power supply unit adaptor
Note: To recharge the RPS II using an Air 10 power supply unit (supplied with Air 10 devices) or the Air 10 DC converter, an Air 10 power supply unit adaptor is required (part no. 37342).
Optional accessory: -
Battery coupler (packed with Velcro™ strap)
Compatible devices and accessories
Refer to the following list on www.resmed.com/downloads/devices:
- Battery/Device Compatibility List for all compatible devices for RPS II
- Power Supply Unit/Device Compatibility List for a list of compatible power supplies for RPS II
- Ventilation Accessories for a list of compatible accessories
If you do not have internet access, please contact your ResMed representative.
For more information on setting up your devices, see your device user guide.
Control panel
Refer to illustration B.
The control panel of your battery includes the following:
- Audible alert mute button
- Charge level check button
- DC on/off switch
- DC input/output port
- Audible alert mute indicator
- Battery level indicators
- Charging indicator
- DC switch/discharging indicator
- Voltage output selector switch (rear panel)
Setting up
CAUTION
- Ensure the RPS II and the device is turned off before setting up.
- When the RPS II is plugged into the mains power supply, ensure all cables are properly connected.
Charging the battery
Refer to illustration C.
- Attach the power supply unit adapter to the DC plug and connect into the battery DC input/output port.
- Connect the AC or DC power cord to the power supply unit.
- Plug the other end of the AC or DC power cord into the mains power outlet.
Notes:
- Charging takes less than 4 hours from 0% battery level to greater than 95%.
- To discontinue charging, unplug the power cord from the mains power outlet.
Providing power to the device
Refer to illustration D.
- Select the correct output voltage for your device using the voltage output selector from the RPS II rear panel.
- Connect the appropriate DC output cable into the battery.
- Connect the other end of the DC output cable into the device.
- Set the DC on/off switch to on.
Providing backup power to the device (for systems with both AC and DC inputs)
Refer to illustration E.
- Select the correct output voltage for your device using the voltage output selector from the RPS II rear panel.
- Connect the appropriate DC output cable into the battery.
- Connect the other end of the DC output cable into the device.
- Connect the AC power cord to the rear of the device.
- Plug the other end of the AC power cord into the mains power outlet.
- Set the DC on/off switch to on.
Note: When powering the device regularly check the battery charge level.
Using two batteries
Refer to illustration F.
- Place one battery on top of another. If required, use the Velcro strap to hold the batteries securely.
- Plug the battery coupler cables into each of the battery DC input/output ports. The battery coupler cables are marked with ‘1’ (primary) and ‘2’ (reserve) labels.
- When charging two batteries, attach the power supply unit adapter to the DC plug and connect the adapter to the battery coupler. When powering a device, connect the DC output cable of the device to the end of the battery coupler. The primary battery supplies power to the device and the reserve battery takes over when the primary battery is disconnected or discharged.
Indicators
Refer to illustration B.
The RPS II provides LEDs to indicate its current operating state.
When setting the DC on/off switch, the charge level indicator may fluctuate between different levels.
To check the battery charge level, press the charge level check button (B-2)
on the control panel. The number of green LEDs indicates the approximate
charge level.
To mute the alert, press the audible alert mute button (B-1). To permanently
mute, press and hold the audible alert mute button (B-1) for 5 seconds. To
deactivate, press the audible alert mute button (B-1) when the mute is still
active.
Using humidifiers
With the exception of AirSense 10, AirCurve 10, Lumis, S9, and H5i™, when a
device integrated with a humidifier is not connected to the mains power
supply, the humidifier will only operate in passive unheated mode despite the
warm-up message appearing on the device. When the device is connected to the
main power supply, it will return to active heated mode.
AirSense 10, AirCurve 10, Lumis, S9, and H5i devices operate normally inactive
heated mode when connected to the battery or to the mains power supply.
Cleaning and maintenance
WARNING
Do not immerse the RPS II in water, and do not use liquids to clean any part of the product.
- Disconnect the battery from the power supply unit and the device. Remove all cables.
- Wipe the exterior of the battery with a clean cloth.
Storing
CAUTION
Batteries must be charged to 100% indicated by four green LEDs and switched
off before storing. You must recharge the batteries to 100% again after six
months of storage. When not in use, all Lithium-Ion batteries self-discharge
over time. If not periodically recharged (ie, every six months), the RPS II
will eventually self-discharge to the point that it can no longer be
recharged. If this occurs the RPS II is no longer usable and is unrecoverable.
Store the RPS II in a cool, dry place.
Note: A fully charged battery when left switched on, self-discharges to 0% of full capacity within four weeks of storage. When switched off, the battery self-discharges to 0% of full capacity within six months of storage.
Servicing
The RPS II is intended to provide safe and reliable operation when used and
maintained in accordance with the instructions provided by ResMed. There is no
servicing necessary during the lifetime of the battery.
The service life of the RPS II exceeds 500 recharge cycles. After 500 recharge
cycles, the battery will last about 60% of the time from its original
condition. A full charge in an older battery will not last as long as in a new
battery. ResMed recommends that the RPS II be tested for battery duration
regularly. As with all electrical equipment, if any irregularity becomes
apparent, you should exercise caution and contact an authorized ResMed service
representative.
Traveling
Consult with your carrier if you intend to take the battery on board with your device.
Troubleshooting
If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the battery.
Problem/Possible cause Solution
The device is not running
-
Power connections are disrupted.
Check all cables and connect them as described in Setting up. -
RPS II is discharged.
Connect the device to the mains power supply and recharge the RPS II. -
RPS II is off.
Set the DC on/off switch to on. -
Incorrect output voltage selected.
Select the correct output voltage for your device using the output voltage selector switch from the RPS II rear panel.
Note: For a list of output voltages and DC cables for your device, see the attached information leaflet, and your device user guide, or see the Battery/Device Compatibility List on www.resmed.com.
RPS II sounds an audible alert continuously and amber LED is flashing
- The battery charge level is less than 5%.
Press the audible alert mute button to mute the alert. Recharge the battery as soon as possible.
RPS II sounds an audible alert for 10 seconds and a green LED is flashing
- The battery charge level is below 10%.
Recharge the battery as soon as possible.
Charging stops before completion
- When charging the battery inside the RPS II carry bag, the ambient temperature is warmer than 95oF (35oC).
Remove the battery from the RPS II carry bag or charge the battery where the ambient temperature is cooler than 95oF (35oC).
The battery turns off and stops powering a device
- When the battery is in the RPS II carry bag, the ambient temperature is warmer than 95oF (35oC).
Remove the battery from the RPS II carry bag or power the device where the ambient temperature is cooler than 95oF (35oC).
The battery charge level indicator is inaccurate
- Ambient temperature is at extreme condition [eg, 23oF (–5oC), 104oF (+40oC)].
Recharge the battery as soon as possible to ensure adequate capacity.
Technical specifications
-
Battery technology Lithium-ion
-
Capacity < 100 Wh (97 Wh)
-
UN classification UN3480 (lithium-ion batteries)
-
AC power supply Input range 100–240V, 50–60Hz, 1.0-1.5A
-
Nominal for aircraft use 110V, 400Hz
-
DC converter Input range 12V / 24V, 13A / 6.5A
-
Battery output voltage (24V / 26V) ± 0.5V, 90W continuous
-
Power supply unit output current (rated) 3.75 A
-
Battery output current (rated) 3.75 A / 3.46 A
-
Standby current < 100 μA
-
Protection Overcharge, over-discharge, excess current, short circuit, high temperature
-
Dimensions (L x W x H) 9.1” x 5.0” x 1.0” (230 mm x 126 mm x 26 mm)
-
Battery weight 2.0 lb (0.9 kg)
-
System weight 5.1 lb (2.3 kg)
-
Minimum life cycle ≥ 500 cycles at 73°F (23°C) to 60% capacity
-
Operating conditions
-
Charging
-
Discharging
41°F to +104°F (5°C to 40°C); 5–85% maximum humidity
23°F to +104°F (-5°C to 40°C); 5–85% maximum humidity -
Transport/storage conditions -4°F to +113°F (-20°C to +45°C); 5–85% maximum humidity Operating/storage air pressure 600 hPa to 1100 hPa
-
Aircraft use Product meets the Federal Aviation Administration (FAA) requirements (RTCA/D0-160, section 21, category M) for all phases of air travel.
-
Recharge time < 4 hrs at full level
-
Electromagnetic compatibility
The product complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-
1-2, for residential, commercial, and light industry environments. Information regarding the electromagnetic emissions and immunity of these ResMed devices can be found at www.resmed.com/downloads/devices. -
IEC 60601-1 classification Class II (double insulation) and/or Internally Powered
Equipment, IP21 (IP20 while charging), Continuous Operation (from mains), Limited Operation (from the battery), Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide. -
Battery run time > 8 hrs at average device settings1 For more information, see the Battery/Device Compatibility List on www.resmed.com/downloads/devices.
Using 15 cm H2O (IPAP), 5 cm H2O (EPAP), and 15 BPM (Respiratory Rate). Not applicable to devices when using heated humidification and heated tubing.
Note: The manufacturer reserves the right to change these specifications without notice.
Symbols
See symbols glossary at www.resmed.com/symbols.
Disposal
Disposal of the battery and the power supply unit should be carried out in accordance with applicable national laws and regulations. If you need information on these disposal systems, please contact your local waste administration. If you require information on the collection and disposal of your ResMed device please contact your ResMed office, local distributor or go to www.resmed.com/environment
General warnings and cautions
WARNINGS
- Lithium-ion batteries have built-in safety protection circuits, but can still be dangerous if they are not used correctly. Damaged batteries can become inoperative or catch fire.
- Due to risk of fire or electric shock:
- do not place the RPS II near open fire or heaters.
- do not expose the RPS II to direct sunlight or heat (for example behind a car window).
- do not expose the RPS II to water, rain or high moisture levels.
- do not short-circuit the RPS II.
- do not use a damaged RPS II.
- do not open the RPS II, AC charger or DC converter.
- Make sure to revert to AC power when the remaining capacity of RPS II and/or device’s internal battery is low.
- Make sure to periodically recharge the battery due to the effects of self-discharging.
- As the battery ages, the available capacity decreases. When the remaining battery capacity is getting low, do not rely on the RPS II as the primary supply.
- Ensure the internal battery of any connected device is kept charged up to provide backup power in the event of loss of power from RPS II.
- Only use a supplied ResMed 90W AC power supply unit or DC converter for charging the RPS II.
- Explosion hazard—do not use in the vicinity of flammable anesthetics.
- The RPS II system is not intended for use by persons (including children) with reduced physical, sensory or mental capabilities, or lack of experience and knowledge unless they have been given supervision or instruction concerning use of the RPS II system by person responsible for their safety.
- Children should be supervised to ensure that they do not play with the RPS II system.
- Care should be taken to keep the power supply unit adapter dry. The RPS II system, while connected to a device and discharging, is rated to IP21 (Drip-proof) according to IEC60529. When charging, the RPS II system is rated to IP20 (No protection) due to the power supply unit adapter which is rated to IP20. The AC power supply, RPS II battery and DC converter are rated to IP21.
- Only use ResMed power supplies that are compatible with the RPS II for charging. For a list of compatible power supplies for RPS II, see the Power Supply Unit/Device Compatibility List on www.resmed.com.
CAUTIONS
- Avoid hard, physical impact on the battery.
- Before using it for the first time, ensure that the RPS II and its components are in good condition and operational. If there are any defects, the system should not be used.
- The RPS II is to be used only in accordance with the intended use stated in this guide. Damage to equipment or injury can result from modifications to the equipment or operation.
- Always fully charge the RPS II before use, or before relying on as backup power.
- Monitor the charge level of the RPS II. When the charge level is low, ensure that continuity of power can be maintained.
- When in use, always plug the cable from the RPS II into the device. Keep the RPS II power switch turned on to ensure it can provide backup power.
- When transporting, turn off the RPS II, unplug all cables, and pack in the carry bag.
- Medical electrical equipment requires special precautions regarding EMC and needs to be installed and operated according to the information provided in this user guide. Portable and mobile communications equipment can affect medical electrical equipment. If EMC interference is observed, for example, static on radios, move the RPS II away from other equipment.
Notes:
- The above are general warnings and cautions. Specific warnings cautions and notes appear with the relevant instructions in the guide.
- For any serious incidents that occur in relation to this product, these should be reported to ResMed and the competent authority in your country.
Limited warranty
ResMed Pty Ltd (hereafter ’ResMed’) warrants that your ResMed product shall be free from defects in material and workmanship from the date of purchase for the period specified below.
Product Warranty period
- Mask systems (including mask frame, cushion, headgear, and tubing)—excluding single-use devices
- Accessories—excluding single-use devices
- Flex-type finger pulse sensors
- Humidifier water tubs 90 days
- Batteries for use in ResMed internal and external battery systems 6 months
- Clip-type finger pulse sensors
- CPAP and bilevel device data modules
- Oximeters and CPAP and bilevel device oximeter adapters
- Humidifiers and humidifier cleanable water tubs
- Titration control devices 1 year
- CPAP, bilevel, and ventilation devices (including external power supply units)
- Battery accessories
- Portable diagnostic/screening devices 2 years
This warranty is only available to the initial consumer. It is not transferable. If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective product or any of its components.
This limited warranty does not cover: a) any damage caused as a result of
improper use, abuse, modification or alteration of the product; b) repairs
carried out by any service organization that has not been expressly authorized
by ResMed to perform such
repairs; c) any damage or contamination due to cigarette, pipe, cigar or other
smoke; and d) any damage caused by water being spilled on or into an
electronic device. Warranty is void on products sold, or resold, outside the
region of original purchase.
Warranty claims on defective products must be made by the initial consumer at
the point of purchase.
This warranty replaces all other expressed or implied warranties, including any implied warranty of merchantability or fitness for a particular purpose. Some regions or states do not allow limitations on how long an implied warranty lasts, so the above limitation may not apply to you.
ResMed shall not be responsible for any incidental or consequential damages
claimed to have resulted from the sale, installation, or use of any ResMed
product. Some regions or states do not allow the exclusion or limitation of
incidental or consequential
damages, so the above limitation may not apply to you.
This warranty gives you specific legal rights, and you may also have other
rights which vary from region to region. For further information on your
warranty rights, contact your local ResMed dealer or ResMed office.
References
- Sleep apnea and COPD - learn about symptoms and treatment | ResMed
- Legal disclaimers and intellectual property - ResMed
- Sleep apnea and COPD - learn about symptoms and treatment | ResMed
- Sleep apnea and COPD - learn about symptoms and treatment | ResMed
- resmed.com/downloads/devices
- Environmental policy - ResMed
- resmed.com/symbols
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>