RKI INSTRUMENTS 61-1004RKSS Direct Connect Infrared Explosion Proof Gas Sensor Owner’s Manual
- June 6, 2024
- RKI INSTRUMENTS
Table of Contents
- RKI INSTRUMENTS 61-1004RKSS Direct Connect Infrared Explosion Proof Gas
- Product Warranty
- Overview
- Specifications
- Description
- Installation
- Wiring the CO2 Detector to a Controller
- Start-Up
- Maintenance
- Troubleshooting
- Replacing the CO2 Detector
- Calibration
- Parts List
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
RKI INSTRUMENTS 61-1004RKSS Direct Connect Infrared Explosion Proof Gas
Sensor
WARNING
Read and understand this instruction manual before operating the detector. Improper use of the detector could result in bodily harm or death. Periodic calibration and maintenance of the detector is essential for proper operation and correct readings. Please calibrate and maintain this detector regularly! The frequency of calibration depends upon the type of use you have and the sensor types. Typical calibration frequencies for most applications are between 6 and 12 months but can be required more often or less often based on your usage.
Product Warranty
RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge. This warranty does not apply to those items which by their nature are subject to deterioration or consumption in normal service, and which must be cleaned, repaired, or replaced on a routine basis. Examples of such items are:
- Absorbent cartridges
- Pump diaphragms and valves
- Fuses
- Batteries
- Filter elements
Warranty is voided by abuse including mechanical damage, alteration, rough handling, or repair procedures not in accordance with the operator’s manual. This warranty indicates the full extent of our liability, and we are not responsible for removal or replacement costs, local repair costs, transportation costs, or contingent expenses incurred without our prior approval. THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI INSTRUMENTS, INC. INCLUDING BUT NOT LIMITED TO, THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS, INC. BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY. This warranty covers instruments and parts sold to users by authorized distributors, dealers, and representatives as appointed by RKI Instruments, Inc. We do not assume indemnification for any accident or damage caused by the operation of this gas monitor, and our warranty is limited to the replacement of parts or our complete goods.
Overview
This manual describes the 61-1004RKSS CO2 (carbon dioxide) detector. This manual also describes how to install, start up, maintain, and calibrate the detector when used with a gas monitoring controller. A parts list at the end of this manual lists replacement parts and accessories for the CO2 detector. The 61-1004RKSS CO2 detector includes the 61-0191RK detector and a junction box.
Specifications
Table 1 lists specifications for the CO2 detector.
Detection Range| With Junction Box
61-1004RKSS-01: 0 – 9,000 ppm CO2
61-1004RKSS-02: 0 – 5,000 ppm CO2
61-1004RKSS-03: 0 – 5% Volume CO2
61-1004RKSS-04: 0 – 20% Volume CO2
61-1004RKSS-05: 0 – 50% Volume CO2
61-1004RKSS-10: 0 – 100% Volume CO2
---|---
Area Classification| Explosionproof for Class I, Groups B, C, and D
Sampling Method| Diffusion
Response Time| 90% in 45 seconds
Accuracy| ± 5% of reading or ± 2% of full scale (whichever is greater)
WARNING:
When using the 61-1004RKSS, you must follow the instructions and warnings in
this manual to assure proper and safe operation of the 61-1004RKSS and to
minimize the risk of personal injury. Be sure to maintain and periodically
calibrate the 61-1004RKSS as described in this manual.
Description
The detector is an infrared type of detector. This section describes the components of the 61-1004RKSS detector. The 61-1004RKSS includes the 61-0191RK detector and a junction box. A four point terminal strip is provided inside the junction box for detector connections. Figure 1 below shows the components of the 61-1004RKSS.
61-0191RK Infrared CO2 Detector
The infrared CO2 detector is made up of a miniature infrared CO2 detector
housed and encapsulated in a pipe nipple. The pipe nipple has 3/4 NPT threads
on each end and a 1-1/4 inch hex that allows removal or installation of the
detector with a wrench. A porous flame arrestor coated with a hydrophobic film
that repels liquids is on one end of the detector and allows sample gas to
enter the detector. Four color-coded leads, red, white, green, and black,
extend from the other end of the detector. The leads allow you to connect the
detector to the terminal block in the junction box. To distinguish the
different ranges of CO2 detectors from one another, a short length of shrink
tubing is applied to the wiring where it comes out of the nipple. The
following table indicates the color of the shrink tubing and the color of the
wire to which it is applied.
Detector | Color Designations |
---|---|
0 – 5,000 ppm | black shrink tubing on white wire |
0 – 9,000 ppm | black shrink tubing on green wire |
0 – 5% volume | green shrink tubing on white wire |
0 – 20% volume | black shrink tubing on red wire |
0 – 50% volume | green shrink tubing on red wire |
0 – 100% volume | red shrink tubing on green wire |
The output of the CO2 detector mimics the output of RKI’s standard catalytic LEL detectors. For this reason, the CO2 detector’s wire colors are the same as RKI’s catalytic LEL detectors, black, white, green, and red. When the CO2 detector is wired to an RKI controller, it is wired the same way an LEL detector is wired, to the LEL detector terminals. See “Wiring the CO2 Detector to a Controller” on page 9 for wiring connections.
Junction Box
The stainless steel, corrosion-resistant junction box allows you to install
the detector at a mounting site that is remote from a controller, and it
protects the detector wiring connections. Two conduit hubs allow you to mount
the detector to the junction box and connect the wiring from the detector to a
controller. An O-ring seals the interface between the junction box cover and
the junction box base. The bottom conduit hub allows you to screw the detector
into the hub. The terminal block within the junction box facilitates the
wiring process. A cover on the front of the junction box allows access to the
interior of the junction box. A locking set screw on the junction box cover
allows you to secure the junction box cover and prevent it from being removed.
Installation
This section describes procedures to mount the CO2 detector in the monitoring environment and wire it to a controller.
Mounting the CO2 Detector
- Select a mounting site that is representative of the monitoring environment. Consider the following when you select the mounting site.
- Select a site where the detector is not likely to be bumped or disturbed. Make sure there is sufficient room to perform start-up, maintenance, and calibration procedures.
- Select a site where the target gas is likely to be found first.
- Select a site that minimizes the possibility of someone breathing on the detector. The exhaled CO2 may cause an alarm.
- At the mounting site you select, hang or mount the junction box with the detector facing down (see Figure 2).
Wiring the CO2 Detector to a Controller
WARNING:
Always verify that the power to the controller is off before you make wiring
connections.
As mentioned in “Description” on page 6, the IR CO2 detector wires to an RKI controller the same way as an RKI LEL detector, to the LEL detector terminals. See Figure 3 below for the wiring connections.
-
Turn off the controller.
-
Turn off or unplug the power to the controller.
-
Remove the cover from the junction box.
WARNING : To maintain the explosion-proof classification of the CO2 detector/junction box combination, a conduit seal must be used within 18 inches of the junction box conduit hub used for wiring to the controller. Adhere to your local electrical code when installing the conduit seal. -
Guide a four-conductor, shielded cable or four wires in conduit through the top conduit hub of the junction box. Use appropriate conduit fittings and construction techniques for the environmental rating of the junction box. The junction box is rated NEMA 4X.
-
Connect the wires to the terminals opposite the detector leads.
CAUTION : If using shielded cable. Leave the shield drain wire insulated and disconnected at the 61-1004RKSS. You will connect the opposite end of the cable’s drain wire at the controller. -
Secure the junction box cover to the junction box.
-
Route the cable or wires in conduit leading from the detector through one of the conduit hubs at the controller. Use appropriate conduit fittings and construction techniques for the environmental rating of the controller. RKI controllers are typically rated NEMA 4X, are not explosion-proof, and must be installed in a non-hazardous location.
CAUTION : Do not route power and detector wiring through the same conduit hub. The power cable may disrupt the transmission of the detector signal to the controller. -
Connect the wires to the applicable controller terminal strip. See the controller operator’s manual and the controller’s detector head specification sheet for the 61-1004RKSS detector.
-
If using a shielded cable, connect the cable’s drain wire to the available chassis ground at the controller. RKI controllers typically have a ground stud that is a convenient grounding location.
Start-Up
This section describes procedures to start up the CO2 detector and place the detector into normal operation.
Introducing Incoming Power
- Complete the installation procedures described earlier in this manual.
- Verify that the power wiring to the controller is correct and secure. Refer to the controller operator’s manual.
- Turn on or plug-in power to the controller, then turn on the controller.
- Verify that the controller is on and operating properly. Refer to the controller operator’s manual.
CAUTION : Allow the detector to warm up for 5 minutes before you continue with the next section, “Setting the Zero Signal”.
Setting the Zero Signal
Since there is a background of CO2 in air of typically 300 – 600 ppm (0.03 –
0.06% volume), it is necessary to use a calibration kit with a CO2 free gas to
set the zero signal of a low range CO2 detector. Fresh air can be used to zero
the detector if a 0-20% volume, 0-50% volume, or 0-100% volume detector is
being used. The procedure below describes applying 100% nitrogen to the
detector using a calibration kit that includes a calibration cup, a
calibration cylinder of 100% nitrogen, sample tubing, and a 0.5 LPM (liters
per minute) fixed flow regulator with an on/off knob.
- Screw the calibration cup onto the bottom of the CO2 detector.
- Screw the regulator into the 100% nitrogen calibration cylinder.
- Use the sample tubing to connect the regulator to the calibration cup.
- Turn the regulator’s on/off knob counterclockwise to open it. Gas will begin to flow.
- Allow the gas to flow for one minute.
- Verify a reading of 0% CO2 or 0 ppm CO2 depending on the detector you are calibrating at the controller. If the display reading is 0, the CO2 detector is in normal operation and start up is complete. Proceed to step 7 to disassemble the calibration kit. If the display reading is not 0, turn the regulator’s on/off knob clockwise to close it, then proceed to “Setting the Zero Reading” on page 14 to set the zero reading.
- Turn the regulator’s on/off knob clockwise to close it.
- Unscrew the cylinder from the regulator and remove the sample tubing from the regulator.
- Remove the calibration cup from the detector. You may leave the sample tubing connected to the calibration cup for convenience.
- Store the components of the calibration kit in a safe and convenient place.
Maintenance
This section describes maintenance procedures. It includes preventive maintenance, troubleshooting, and component replacement procedures.
Preventive Maintenance
This section describes a preventive maintenance schedule to ensure the optimum
performance of the CO2 detector. It includes daily and biannual procedures.
Daily
Verify a display reading at the controller of the background concentration of
CO2. Typical background concentrations of CO2 vary from about 300 to 600 ppm
(0.03 to 0.06% volume) depending on location. The 0-5,000 ppm, 0-9,000 ppm,
and 0-5% volume detectors will display a reading in a fresh air environment
because they have low ranges. The 0-20% volume, 0-50% volume, and 0-100%
volume detectors will not display a reading in fresh air because their ranges
are too large. Investigate significant changes in the display reading.
Biannually
Calibrate the CO2 detector every six months as described in “Calibration” on
page 14.
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended action for problems you may encounter with the CO2 detector. NOTE : This troubleshooting guide describes detector problems only. See the controller operator’s manual for problems you may encounter with the controller.
Condition | Symptom(s) | Probable Causes | Recommended Action |
---|---|---|---|
Fail Condition | • Controller indicates a fail condition. | • The |
detector wiring is disconnected or misconnected.
• The detector zero signal is low enough to cause a fail condition.
• The detector is malfunctioning.
| 1. Verify that the detector wiring is correct and secure.
2. Calibrate the detector.
3. If the fail condition continues, replace the detector.
4. If the fail condition continues, contact RKI for further instruction.
Condition | Symptom(s) | Probable Causes | Recommended Action |
---|---|---|---|
Slow or No Response/ Difficult or Unable to Calibrate | • Unable to |
accurately set the zero or response reading during calibration.
• Detector requires frequent calibration.
| • The calibration cylinder is low, out- dated, or defective.
• The regulator flow rate is not 0.5 LPM.
• The detector is malfunctioning.
| 1. Verify that the calibration cylinder contains an adequate supply of a fresh test sample.
2. Confirm that the regulator being used is a 0.5 LPM regulator.
3. If the calibration/response difficulties continue, replace the detector.
| Note: Under “normal” circumstances, the detector requires calibration once every six months.
Some applications may require a more frequent calibration schedule.
| | 4. If the calibration/response difficulties continue, contact RKI for further instruction.
Replacing the CO2 Detector
- Turn off the controller.
- Turn off power to the controller.
- Remove the junction box cover.
- Disconnect the detector leads from the terminal block in the junction box. Note the position of the color-coded leads as you remove them.
- Unscrew the detector from the junction box conduit hub.
- Guide the detector leads of the replacement detector through the junction box conduit hub, then screw the mounting threads of the detector into the hub. If necessary for environmental conditions, apply thread sealant or Teflon tape to the hub and/or detector threads to seal them.
- Connect the detector leads to the terminal block the same way the old detector was wired (see Figure 3 on page 10). See the controller operator’s manual and the controller’s detector head specification sheet for the 61-1004RKSS detector to verify the connections to the controller are correct.
- Reinstall the junction box cover.
- Turn on power to the controller.
- Turn on the controller.
- Calibrate the replacement detector as described in“Calibration” on page 14.
Calibration Frequency
Although there is no particular calibration frequency that is correct for
all applications, a calibration frequency of every 6 months is adequate for
most infrared CO2 detector applications. Unless experienced in a particular
application dictates otherwise, RKI Instruments, Inc. recommends a calibration
frequency of every 6 months for the infrared CO2 detector. If an application
is not very demanding, for example, detection in a clean, temperature-
controlled environment where CO2 is not normally present, and calibration
adjustments are minimal at calibration, then a calibration frequency of every
9 to 12 months is adequate. If an application is very demanding, for example,
if the environment is not well controlled, then more frequent calibration than
every 6 months may be necessary.
Calibration
This section describes how to calibrate the CO2 detector. It includes procedures to prepare for calibration, set the zero reading, set the response reading, and return to normal operation.
WARNING :
The controller is not an active gas monitoring device during the calibration
procedure. The following procedure assumes the use of a calibration kit which
includes a 100% nitrogen
calibration gas cylinder for setting the zero reading, a CO2 calibration gas
cylinder for setting the response reading, a 0.5 LPM fixed flow regulator with
an on/off knob, a calibration cup for the detector, and a short piece of
sample tubing to connect the regulator to the calibration cup.
Preparing For Calibration
- Screw the calibration cup onto the detector.
- Use the sample tubing to connect the regulator to the calibration cup.
- Place the controller into its calibration program or disable external alarms.
NOTE : Calibrating the CO2 detector may cause alarms. Be sure to put the controller into its calibration program or disable external alarms before continuing.
Setting the Zero Reading
Since there is a background of CO2 in air of typically 300 – 600 ppm (0.03 –
0.06% volume), it is necessary to use a calibration kit with a CO2 free gas to
set the zero signal of a low range CO2 detector. Fresh air can be used to zero
the detector if a 0-20% volume, 0-50% volume, or 0-100% volume detector is
being used.
- Follow the directions in the controller’s operator’s manual for setting the zero reading (sometimes called the fresh air reading).
- When the directions call for exposing the detector to zeroing gas, screw the regulator into the cylinder, turn the on/off knob counterclockwise to open the regulator, and allow the gas to flow to the detector for 1 minute.
- Turn the regulator’s on/off knob clockwise to close the regulator.
- Set the zero reading according to the controller’s operator’s manual.
- Unscrew the 100% nitrogen cylinder from the regulator.
Setting the Response Reading
- Follow the directions in the controller’s operator’s manual for setting the response reading (span).
- When the directions call for exposing the detector to calibration gas, screw the CO2 cylinder onto the regulator, turn the on/off knob counterclockwise to open the regulator, and allow the gas to flow to the detector for 1 minute.
- Turn the regulator’s on/off knob clockwise to close the regulator.
- Set the response reading according to the controller’s operator’s manual.
- Unscrew the CO2 cylinder from the regulator.
Returning to Normal Operation
-
Remove the sample tubing from the regulator.
-
Remove the calibration cup from the detector. Leave the sample tubing connected to the calibration cup for convenience.
-
Allow the reading at the controller to decrease below the alarm points before returning the controller to normal operation or enabling external alarms.
NOTE : If you do not allow the gas reading to decrease below the alarm points, then unwanted alarms may occur. -
Verify that the controller display reading decreases and stabilizes at a typical background CO2 level. The 0-5,000 ppm, 0-9,000 ppm, and 0-5% volume detectors will display a reading in a fresh air environment because they have low ranges. The 0-20% volume, 0-50% volume, and 0-100%volume detectors will not display a reading in fresh air because their ranges are too large.
-
Store the components of the calibration kit in a safe and convenient place.
Parts List
lists replacement parts and accessories for the 61-1004RKSS CO2 detector.
Part Number | Description |
---|---|
07-7151RK | O-ring |
10-5153RK | Lid locking set screw |
18-0416RK-11 | Junction box with cover, stainless steel |
61-0191RK-01 | Infrared CO2 detector, 0 – 9,000 ppm |
61-0191RK-02 | Infrared CO2 detector, 0 – 5,000 ppm |
61-0191RK-03 | Infrared CO2 detector, 0 – 5% volume |
61-0191RK-04 | Infrared CO2 detector, 0 – 20% volume |
61-0191RK-05 | Infrared CO2 detector, 0 – 50% volume |
61-0191RK-10 | Infrared CO2 detector, 0 – 100% volume |
71-0231RK | 61-1004RKSS CO 2 Detector Operator’s Manual (this document) |
81-0070RK-01 | Steel calibration cylinder, 2,000 ppm CO2 in nitrogen, 34 liter |
steel
81-0071RK-01| Steel calibration cylinder, 5,000 ppm CO2 in nitrogen, 34 liter
steel
81-0071RK-03| Steel calibration cylinder, 5,000 ppm CO2 in nitrogen, 103 liter
81-0072RK-01| Steel calibration cylinder, 2.5% CO2 in nitrogen, 34 liter steel
81-0073RK-01| Steel calibration cylinder, 15% CO2 in nitrogen, 34 liter steel
81-0078RK-01| 100% nitrogen calibration cylinder, 34 liter steel
81-1050RK| Regulator, with gauge and knob, 0.5 liter/minute continuous flow,
for 17 liter and 34 liter steel calibration cylinders (cylinders with external
threads)
81-1051RK| Regulator, with gauge and knob, 0.5 liter/minute continuous flow,
for 34 liter alu- minum/58 liter/103 liter calibration cylinders (cylinders
with internal threads)
81-1103RK| Calibration cup
81-F403RK-LV| Calibration kit, for 3/4 NPT detector, 2.5% CO2, 34 liter
81-F404RK-LV| Calibration kit, for 3/4 NPT detector, 2,000 ppm CO2, 34 liter
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>