RKI INSTRUMENTS 71-0595 Molecular Property Spectrometer Combustible Gas Detector User Manual
- June 3, 2024
- RKI INSTRUMENTS
Table of Contents
- RKI INSTRUMENTS 71-0595 Molecular Property Spectrometer Combustible Gas
- Product Warranty
- Overview
- Specifications
- Applications
- Description
- MPS Detector
- Installation
- Wiring the Combustible Gas Detector to a Controller
- Start Up
- Maintenance
- Troubleshooting
- Replacing the MPS LEL Detector
- Replacing the Amplifier
- Calibration Frequency
- Setting the Response Reading
- Parts List
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
RKI INSTRUMENTS 71-0595 Molecular Property Spectrometer Combustible Gas
Detector
WARNING
- Read and understand this instruction manual before operating transmitter. Improper use of the transmitter could result in bodily harm or death.
- Periodic calibration and maintenance of the transmitter is essential for proper operation and correct readings. Please calibrate and maintain this transmitter regularly! Frequency of calibration depends upon the type of use you have and the sensor types. For most applications, typical calibration frequencies are between 6 and 12 months but can be more often or less often based on your usage.
Product Warranty
RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge. This warranty does not apply to those items which by their nature are subject to deterioration or consumption in normal service, and which must be cleaned, repaired, or replaced on a routine basis. Examples of such items are:
- Absorbent cartridges
- Pump diaphragms and valves
- Fuses
- Batteries
- Filter elements
Warranty is voided by abuse including mechanical damage, alteration, rough
handling, or repair procedures not in accordance with the operator’s manual.
This warranty indicates the full extent of our liability, and we are not
responsible for removal or replacement costs, local repair costs,
transportation costs, or contingent expenses incurred without our prior
approval.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND
REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER OBLIGATIONS OR
LIABILITIES ON THE PART OF RKI INSTRUMENTS, INC. INCLUDING BUT NOT LIMITED TO,
THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO
EVENT SHALL RKI INSTRUMENTS, INC. BE LIABLE FOR INDIRECT, INCIDENTAL, OR
CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS
PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
This warranty covers instruments and parts sold to users by authorized
distributors, dealers, and representatives as appointed by RKI Instruments,
Inc.
We do not assume indemnification for any accident or damage caused by the
operation of this gas monitor, and our warranty is limited to the replacement
of parts or our complete goods
Overview
This manual describes the 65-2380 MPS combustible gas detector. This manual also describes how to install, start up, maintain, and calibrate the detector when used with a gas monitoring controller. A parts list at the end of this manual lists replacement parts and accessories for the combustible gas detector.
Specifications
Table 1 specifies the detection ranges and accuracy for each target gas the MPS sensor can detect.
Target Gas | Formula | Detection Range | Accuracy at 50% LEL |
---|---|---|---|
Butane | C4H10 | 0 – 100% LEL | ± 5% LEL |
Ethane | C2H6 | 0 – 100% LEL | ± 5% LEL |
Hydrogen | H2 | 0 – 100% LEL | ± 5% LEL |
Isobutane | HC(CH3)3 | 0 – 100% LEL | ± 5% LEL |
Isobutylene | C4H8 | 0 – 100% LEL | ± 5% LEL |
Isopropanol | C3H8O | 0 – 100% LEL | ± 10% LEL |
Methane | CH4 | 0 – 100% LEL | ± 3% LEL |
Methyl Ethyl Ketone | C4H8O | 0 – 100% LEL | ± 5% LEL |
Octane | C8H18 | 0 – 100% LEL | ± 5% LEL |
Pentane | C5H12 | 0 – 100% LEL | ± 5% LEL |
Propane | C3H8 | 0 – 100% LEL | ± 6% LEL |
Propylene | C3H6 | 0 – 100% LEL | ± 5% LEL |
Toluene | C7H8 | 0 – 100% LEL | ± 12% LEL |
Xylene | C8H10 | 0 – 100% LEL | ± 12% LEL |
NOTE: The MPS sensor will respond to other combustible gases, but the accuracy has only been determined for these 14 gases.
Area Classification | Explosion-proof for Class I, Groups B, C, and D |
---|---|
Operating Temperature Range | -20ºC to +50ºC (-4ºF to +122ºF) |
Humidity Range | 0 – 100% RH (non-condensing) |
Pressure Range | 80 – 120 kPa |
Sampling Method | Diffusion |
Response Time | T90 in 25 seconds |
Dead Bands | • Methane (CH4): 3% LEL |
• All other gases: 5% LEL
WARNING: When using the 65-2380, you must follow the instructions and warnings in this manual to assure proper and safe operation of the 65-2380 and to minimize the risk of personal injury. Be sure to maintain and periodically calibrate the 65- 2380 as described in this manual
Applications
The MPS combustible detector is most effective in the following applications:
- High concentrations of hydrogen sulfide (H2S) or silicones present where sensor poisoning is a common problem
- High-humidity environments where catalytic or IR sensors may not hold up
- Environments more likely to have mixtures of hydrocarbons present
Description
The 65-2380 combustible gas transmitter utilizes a Molecular Property
Spectrometer (MPS) sensor. The MPS sensor is immune to poisons, can be used in
high humidity, and does not require calibration as often as catalytic
combustible sensors. It also responds accurately to 14 different combustible
gases.
This section describes the components of the combustible gas transmitter. The
transmitter is a 4 – 20 mA type detector head. It consists of the combustible
gas detector, amplifier, and junction box.
MPS Detector
The MPS LEL detector is made up of the MPS combustible gas LEL detector housed and encapsulated in a pipe nipple. The pipe nipple has 3/4 NPT threads on each end and a 1-1/4 inch hex that allows removal or installation of the detector with a wrench. A porous flame arrestor coated with a hydrophobic film that repels liquids is on one end of the detector and allows sample gas to enter the detector. Four color coded leads, red, white, green, and black, extend from the other end of the detector. Use the leads when connecting the detector to the terminal block in the junction box.
Junction Box
Use the junction box to install the transmitter at a mounting site that is remote from the controller. The junction box protects the amplifier and wiring connections made to the amplifier. Use the top 3/4 NPT conduit hub to connect wiring from the amplifier to the controller. Use the cover on the front of the junction box to access the interior of the junction box. The detector and amplifier are factory installed in the junction box. Three spacers installed on the back of the junction box control the distance of the junction box from a mounting surface and ensure that there is enough room to install a calibration cup on the detector during calibration.
Amplifier
The amplifier converts the electrical output from the detector to a 4 – 20 mA signal that corresponds to the detection range and transmits the signal to a gas monitoring controller. A foam gasket that orients the amplifier and keeps it from rotating is installed on the bottom of the amplifier. A label on the amplifier indicates the detector drive current. The amplifier includes the detector terminal strip, controller terminal strip, span pot, zero pot, and test points (see Figure 2).
Controller Terminal Strip
The controller terminal strip is a three position plug-in style terminal strip located near the top of the amplifier. Use the controller terminal strip to wire the combustible gas transmitter to a controller.
Detector Terminal Strip
The detector terminal strip is a four position plug-in style terminal strip located below the controller terminal strip. Use the detector terminal strip to connect the MPS LEL detector to the amplifier.
NOTE : The MPS LEL detector is factory-wired to the detector terminal strip. See “Wiring the Combustible Gas Detector to a Controller” on page 11 for all wiring procedures related to the transmitter.
Zero Pot
The zero pot is located in the upper left corner of the amplifier (see Figure 2). Use a small flat blade screwdriver to turn the zero pot’s adjustment screw and adjust the amplifier’s zero (fresh air) output during the start-up and calibration procedures. Turn the adjustment screw clockwise to increase the zero output and counterclockwise to decrease the zero output.
Span Pot
The span pot is located to the right of the zero pot (see Figure 2). Use a
small flat blade screwdriver to turn the span pot’s adjustment screw and
adjust the amplifier’s gas response output during the calibration procedure.
Turn the adjustment screw clockwise to increase the gas response and
counterclockwise to decrease the gas response.
CAUTION : The amplifier includes two additional potentiometers. They are
factory-set. Do not adjust them.
Test Points
The test points are on the left side of the amplifier (see Figure 2). The test points produce a 100 mV to 500 mV output that corresponds to the transmitter’s 4 to 20 mA output. Use the test points and a voltmeter to measure the amplifier’s output during the start-up and calibration procedures. The black test point in the lower left corner is the negative (-) test point and the red test point below the zero pot is the positive (+) test point.
Installation
This section describes procedures to mount the combustible gas transmitter in the monitoring environment and wire the transmitter to a controller.
Mounting the Combustible Gas Transmitter
- Select a mounting site that is representative of the monitoring environment. Consider the following when you select the mounting site.
- Select a site where the transmitter is not likely to be bumped or disturbed. Make sure there is sufficient room to perform start-up, maintenance, and calibration procedures.
- Select a site where the target gas is likely to be found first. For lighter gases, mount the detector near the ceiling; for heavier gases, mount the detector near the floor
- At the monitoring site you select, hang or mount the junction box with the detector facing down (see Figure 3).
Wiring the Combustible Gas Detector to a Controller
WARNING : Always verify that the power to the controller is off before you make wiring connections.
-
Turn off the controller.
-
Turn off or unplug power to the controller.
-
Remove the junction box cover.
-
The detector leads are factory wired. Verify that the detector leads are wired to the amplifier’s detector terminal strip as shown in Figure 11.
-
To gain access to a plug-in terminal strip for wiring, pull it out of its socket by grasping the terminal strip and pulling. The detector strip is keyed so that the controller and detector terminal strips cannot be reversed inadvertently.
-
Guide a three-conductor, shielded cable, or three wires in conduit through the top conduit hub of the junction box.
-
Pull out the controller terminal strip, and connect the three wires to the terminal strip as follows (see Figure 11).
-
Connect the positive wire to the PWR/SIG “+” terminal.
-
Connect the signal wire to the PWR/SIG “S” terminal.
-
Connect the negative wire to the PWR/SIG “-” terminal.
-
CAUTION: If shielded cable is used, leave the cable shield’s drain wire disconnected and insulated at the transmitter. You will connect the opposite end of the cable’s drain wire to the controller’s chassis (earth) ground.
-
Reinstall the controller terminal strip into its socket.
-
Secure the junction box cover to the junction box.
-
Route the cable or wires leading from the combustible gas transmitter through one of the conduit hubs at the controller housing.
CAUTION: Do not route power and transmitter wiring through the same controller conduit hub. The power cable may disrupt the transmission of the transmitter signal to the controller. -
Connect the wires to the applicable detector/transmitter terminal strip at the controller as shown in Figure 11.
-
If shielded cable is used, connect the cable’s drain wire to an available chassis (earth) ground at the controller. RKI controllers typically have a ground stud that can be used to ground the cable’s drain wire
Start Up
This section describes procedures to start up the combustible gas transmitter and place the transmitter into normal operation.
NOTE: The detector must be started up in fresh air.
Introducing Incoming Power
- Complete the installation procedures described earlier in this manual.
- Verify that the power wiring to the controller is correct and secure. Refer to the controller operator’s manual.
- Turn on power to the controller.
- Turn on the controller.
- Verify that the controller is on and operating properly. Refer to the controller operator’s manual.
CAUTION: Allow the detector to warm up for 5 minutes before you continue with the next section, “Setting the Zero Reading”.
Setting the Zero Reading
WARNING: Do not remove the junction box cover while the circuits are energized unless the area is determined to be non-hazardous. Keep the junction box cover tightly closed during operation.
NOTE : If you can verify that the combustible gas transmitter is in a
fresh air environment, you do not need to apply zero air to the detector
before adjusting the zero reading.
The procedure below describes applying zero emission air, usually called zero
air, using a calibration kit that includes a calibration cup, calibration gas,
sample tubing, and a fixed flow regulator with an on/off knob. RKI
Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed flow
regulator.
- Allow the combustible detector to warm up for 2 minutes before setting the zero reading.
- Unscrew and remove the junction box cover from the junction box.
- Set a voltmeter to measure in the millivolt (mV) range.
- Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead into the red (+) test point; plug the negative lead into the black (-) test point.
- Screw the calibration cup onto the bottom of the MPS LEL detector.
- Screw the regulator into the zero air calibration cylinder.
- Use the calibration kit sample tubing to connect the regulator to the calibration cup.
- Turn the regulator’s on/off knob counterclockwise to open it. Gas will begin to flow.
- Allow the gas to flow for one minute.
- Verify a voltmeter reading of 100 mV (±2 mV).
- If necessary, use a small flat-blade screwdriver to adjust the zero pot until the voltmeter reading is 100 mV (±2 mV).
- Turn the regulator’s on/off knob clockwise to close it.
- Unscrew the calibration cup from the detector.
- Unscrew the regulator from the zero air calibration cylinder. For convenience, leave the sample tubing connected to the regulator and the calibration cup.
- Store the components of the calibration kit in a safe and convenient place.
- Remove the voltmeter leads from the test points.
- Secure the junction box cover to the junction box.
Maintenance
This section describes maintenance procedures. It includes preventive maintenance, troubleshooting, and component replacement procedures.
-
Preventive Maintenance
This section describes a preventive maintenance schedule to ensure the optimum performance of the combustible gas transmitter. It includes daily and biannual procedures. -
Daily
Verify a display reading of 0% LEL at the controller. Investigate significant changes in the display reading. -
Biannually
Calibrate the combustible gas transmitter as described in “Calibration” on page 18 of this manual.
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended action for problems you may encounter with the combustible gas transmitter.
NOTE: This troubleshooting guide describes transmitter problems only. See the controller operator’s manual for problems you may encounter with the controller.
Condition| Symptom(s)| Probable Causes| Recommended
Action
---|---|---|---
Fail Condition| • Controller indicates a fail condition.| • The
transmitter wiring is disconnected or misconnected.
• The transmitter’s zero reading is low enough to cause a fail condition.
• The detector was exposed to gas during its 2 minute warmup period.
• The transmitter is malfunctioning.
| 1. Verify that the transmitter wiring is correct and secure.
2. Calibrate the transmitter.
3. Restart the controller and ensure that the detector’s environment is free of combustible gas during the warmup period.*
4. If the fail condition continues, replace the detector.
5. If the fail condition continues, contact RKI for further instruction.
Slow or No Response/Dif- ficult or Unable to Cali-
brate
| • Unable to accurately set the zero or response reading during calibration.| • The calibration cylinder is low, out- dated, or defective.
• The calibration gas flow rate is too low.
| 1. Verify that the calibration cylinder contains an adequate supply of a fresh test sample.
2. Verify that the regulator used for calibration is a 0.5 LPM regulator.
3. Verify that the calibration gas concentration is appropriate for the transmitter. The concentration should be in the detection range, preferably about half of the detection range.
4. If the calibration/response difficulties continue, replace the detector.
5. If the calibration/response difficulties continue, contact RKI for further instruction.
| • Transmitter requires frequent calibration.| • The calibration gas is
not an appropriate concentration.
| NOTE: Under “normal” circumstances, the transmitter requires
calibration once every 6 months.| • The transmitter is malfunctioning.
| NOTE: Some applications may require a more frequent calibration
schedule.|
If a fresh air environment cannot be guaranteed, apply zero air during the warmup period.
Replacing the MPS LEL Detector
-
Turn off the controller.
-
Turn off or unplug power to the controller.
-
Remove the junction box cover.
-
Remove the detector terminal strip from its socket.
-
Disconnect the detector leads from the detector terminal strip. Note the position of the colorcoded leads as you remove them.
-
Unscrew the detector from the junction box.
-
Guide the detector leads of the replacement detector through the bottom conduit hub of the junction box, then screw the mounting threads of the detector into the conduit hub. If necessary for environmental conditions, apply thread sealant or Teflon tape to the hub and/or detector threads to seal them.
-
Connect the detector leads to the detector terminal strip as shown in Table 4 below and Figure 11 on page 12 of this manual
Detector Lead| Detector Terminal Strip
---|---
Red| LEL “R”
White| LEL “W”
Green| LEL “G”
Black| LEL “B” -
Reinstall the detector terminal strip into its socket.
-
Reinstall the junction box cover.
-
Turn on or plug in power to the controller.
-
Turn on the controller and place it into normal operation.
CAUTION: Allow the replacement detector to warm up for 5 minutes before you continue with the next step. -
Calibrate the transmitter as described in “Calibration” on page 18 of this manual.
Replacing the Amplifier
- Turn off the controller.
- Turn off or unplug power to the controller.
- Remove the junction box cover.
- Unplug the detector terminal strip and controller terminal strip from their sockets. You may leave the wires connected to the terminal strips.
- Unscrew and remove the screw with the flat and lock washers that secures the amplifier to the junction box.
- Remove the old amplifier.
- Place the new amplifier in the same position as the old amplifier. A foam gasket that orients the amplifier and keeps it from rotating is installed on the bottom of the amplifier. Make sure the amplifier is seated flat in the junction box.
- Install the new amplifier into the junction box with the screw, lock washer, and flat washer you removed in Step 5 above.
- Install the detector and controller terminals strips into their sockets on the new amplifier as shown in Figure 11 on page 12 of this manual. If controller leads or detector leads were removed during this procedure, refer to Table 5 and Table 6 below.
Amplifier Controller Terminal Strip| Controller Transmitter Terminal Strip (typical)
---|---
PWR/SIG “-”| – (DC -)
PWR/SIG “S”| S (4 – 20 mA In)
PWR/SIG “+”| + 24V
Amplifier Detector Terminal Strip| Detector Lead
---|---
DETECTOR “R”| RED
DETECTOR “W”| WHT
DETECTOR “G”| GREEN
DETECTOR “B”| BLK
NOTE: When a transmitter is first powered up with a new amplifier, the initial output may be either high or below zero depending on the setting of the zero pot. Be sure to make arrangements so that this does not cause unwanted alarms.
- Turn on power to the controller.
- Turn on the controller and place it into normal operation.
- Allow the transmitter to warm-up for 5 minutes.
- Calibrate the transmitter as described in “Calibration” on page 18 of this manual
Calibration Frequency
Although there is no particular calibration frequency that is correct for all
applications, a calibration frequency of every 6 months is adequate for most
MPS combustible gas transmitter applications. Unless experience in a
particular application dictates otherwise, RKI Instruments, Inc. recommends a
calibration frequency of every 6 months.
If an application is not very demanding, for example detection in a clean,
temperature controlled environment where calibration adjustments are minimal
at calibration, then a calibration frequency of every 9 to 12 months is
adequate.
If an application is very demanding, for example if the environment is not
well controlled, then more frequent calibration than every 6 months may be
necessary.
Calibration
This section describes how to calibrate the combustible gas transmitter. It
includes procedures to prepare for calibration, set the zero reading, set the
response reading, and return to normal operation.
WARNING : The controller is not an active gas monitoring device during
the calibration procedure.
NOTE: The following procedure assumes the use of a calibration kit which
includes a calibration gas cylinder, a 0.5 LPM fixed flow regulator with an
on/off knob, a calibration cup for the detector, and a short piece of sample
tubing to connect the regulator to the calibration cup.
NOTE : During calibration, the detector must be calibrated with a cylinder of methane in a background of real air containing argon and CO2. Only use the RKI-provided calibration cylinder 81-9070RK-03.
Materials
The following items are required for detector calibration:
- a CH4 calibration cylinder (in a background of real air which contains argon and CO2)
- a 0.5 LPM fixed flow regulator with an on/off knob
- a calibration cup for the detector
- a short piece of sample tubing to connect the regulator to the calibration cup
- a voltmeter
Preparing for Calibration
-
Unscrew and remove the junction box cover.
-
Set a voltmeter to measure in the millivolt (mV) range.
-
Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead into the red (+) test point; plug the negative lead into the black (-) test point.
-
Use the following formula to determine the correct test points output for the calibrating sample.
Output (mV) = (calibrating sample/fullscale) X 400 + 100
For example, with a calibrating sample of 50% LEL and a fullscale setting of 100% LEL, the correct output is 300 mV. 300(mV) = (50/100) X 400 +100 -
Screw the calibration cup onto the detector housing.
-
Use the sample tubing to connect the regulator to the calibration cup.
-
Place the controller into its calibration mode or disable external alarms
NOTE : Calibrating the combustible gas transmitter may cause alarms. Be sure to put the controller into its calibration mode or disable external alarms before continuing.
Setting the Zero Reading
CAUTION : If you can verify that the combustible gas transmitter is in a fresh air environment, you do not need to apply zero air to the detector before adjusting the zero reading.
- Allow the combustible detector to warm up for 2 minutes before setting the zero reading.
- Screw the regulator into the zero air calibration cylinder.
- Use the calibration kit sample tubing to connect the regulator to the calibration cup.
- Turn the regulator’s on/off knob counterclockwise to open the regulator.
- Allow the gas to flow for one minute and verify a reading of 100 mV (±2 mV). If necessary, use the zero pot on the amplifier to adjust the reading to 100 mV (±2 mV).
- Turn the regulator’s on/off knob clockwise to close the regulator.
- Unscrew the regulator from the zero air calibration cylinder.
- Leave the sample tubing connected to the regulator and the calibration cup.
Setting the Response Reading
- Ensure that only an MPS-specific methane calibration cylinder is in use.
- Screw the regulator into the methane in real air cylinder.
- Turn the regulator’s on/off knob counterclockwise to open the regulator.
- Allow the calibration gas to flow for one minute and verify that the reading matches the response reading (±2 mV) you determined earlier. If necessary, use the span pot on the amplifier to adjust the reading to match the correct response reading.
- Turn the regulator’s on/off knob clockwise to close the regulator.
- Unscrew the regulator from the calibration cylinder.
Returning to Normal Operation
-
Remove the voltmeter leads from the amplifier test points.
-
Unscrew the calibration cup from the detector.
NOTE : For convenience, leave the regulator and calibration cup connected by the sample tubing. -
Secure the junction box cover to the junction box.
-
When the controller display reading falls below the alarm setpoints, return the controller to normal operation.
-
Verify that the controller display reading decreases and stabilizes at 0% LEL.
-
Store the components of the calibration kit in a safe and convenient place.
Parts List
Table 7 lists replacement parts and accessories for the combustible gas transmitter.
Part Number | Description |
---|---|
06-1248RK | Sample tubing (3/16 in. x 5/16 in.; specify length when ordering) |
18-0400RK-01 | Junction box with spacers |
57-1053RK | Amplifier with gasket (specify detector part number when ordering) |
61-0230 | MPS LEL detector |
65-2380 | MPS transmitter (includes detector, junction box, and amplifier) |
71-0595 | 65-2380 Combustible Gas Detector Operator’s Manual (this document) |
81-1051RK | Regulator with gauge and knob, 0.5 LPM, for 103 liter calibration |
cylinders (cylinders with internal threads)
81-1103RK| Calibration cup
81-9070RK-03| Calibration cylinder, 50% LEL methane in real air, 103 liter
81-F076RK| Cal kit: 103 liter cylinder of 50% LEL CH4 in real air, 0.5 LPM
fixed flow regulator with gauge and knob, calibration cup, and 3 feet of
calibration tubing
82-0006RK| Pot adjust screwdriver, for calibration
References
- RKI Gas Detectors - Portable Gas Monitors - LEL, PID, H2S, CO, O2 Sensors
- RKI Gas Detectors - Portable Gas Monitors - LEL, PID, H2S, CO, O2 Sensors
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>