HANNA Instruments HI782 Marine Nitrate High Range
Tips for an Accurate Measurement
- Ensure the sample does not contain any debris.
- Whenever the cuvette is placed into the checker, it must be dry outside and free of fingerprints, oil and dirt.
- Wipe the cuvette thoroughly with HI731318 microfiber cleaning cloth or a lint-free cloth prior to insertion.
- Shaking the cuvette can generate bubbles, causing higher readings. To obtain accurate measurements, remove such bubbles by swirling or by gently tapping the cuvette.
- Do not let the reacted sample stand too long after reagent has been added, as accuracy will be affected.
- Discard the sample immediately after the reading has been taken or the glass might become permanently stained.
Battery Replacement
To save the battery, the checker shuts down after 10 minutes of non-use.
A fresh battery lasts for a minimum of 5000 measurements. When the battery is
drained, the instrument displays “bAd” then ”bAt”, and turns off.
To replace the battery, follow the next steps:
1. Press and hold the ON/OFF button to turn the checker off.
2. Turn the instrument upside down and use a screwdriver to unfasten the
screw and remove the battery cover.
3. Remove the old battery, replace it with a new 1.5V AAA battery, inserting
the negative end first.
4. Replace the battery cover, fasten and tighten the screw.
Accessories
Certification
All Hanna Instruments conform to the CE European Directives. Disposal of
Electrical & Electronic Equipment. The product should not be treated as
household waste. Instead, hand it over to the appropriate collection point for
the recycling of electrical and electronic equipment which will conserve
natural resources.
Disposal of waste batteries. This product contains batteries, do not
dispose of them with other household waste. Hand them over to the appropriate
collection point for the recycling
Ensuring proper product and battery disposal prevents potential negative
consequences for the environment and human health. For more information,
contact your city, your local household waste disposal service, or the place
of purchase.
Recommendations for Users
Before using this product, make sure it is entirely suitable for your
specific application and for the environment in which it is used. Any
variation introduced by the user to the supplied equipment may degrade the
checker’s performance. For your and the checker’s safety do not use or store
it in hazardous environments.
Warranty
HI782 Checker®HC is warranted for a period of one year against defects in
workmanship and materials when used for its intended purpose and maintained
according to instructions. This warranty is limited to repair or replacement
free of charge. Damage due to accidents, misuse, tampering or lack of
prescribed maintenance is not covered. If service is required, contact your
local Hanna Instruments Office. If under warranty, report the model number,
date of purchase, serial number and the nature of the problem. If the repair
is not covered by the warranty, you will be notified of the charges incurred.
If the checker is to be returned to Hanna Instruments, first obtain a Returned
Goods Authorization (RGA) number from the Technical Service department and
then send it with shipping costs prepaid. When shipping any product, make sure
it is properly packaged for complete protection.
Hanna Instruments reserves the right to modify the design, construction, or
appearance of its products without advance notice.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments Inc.,
Woonsocket, Rhode Island, 02895, USA.
Dear Customer,
Thank you for choosing a Hanna Instruments product. Please read this
instruction manual carefully before using the Checker®HC handheld colorimeter.
For more information about Hanna Instruments and our products, visit
www.hannainst.com or e-mail us at sales@hannainst.com.
For technical support, contact your local Hanna Instruments Office or e-mail
us at tech@hannainst.com.
Preliminary Examination
Remove the Checker®HC handheld colorimeter and accessories from the packing
material and examine it carefully. If you require any further information,
please contact Hanna Instruments technical support team.
Each HI782 is delivered in a case with custom insert and is supplied with:
- Sample cuvette and cap (2 pcs.)
- Marine Nitrate High Range reagent starter kit (reagents for 10 tests)
- 3 mL Pasteur pipette (1 pc.)
- 1.5V AAA Alkaline battery (1 pc.)
- Instruction manual
Note: Save all packing material until you are sure that the Checker®HC handheld colorimeter works correctly. Any damaged or defective item must be returned in its original packing material with the supplied accessories.
General Description & Intended Use
HI782 Marine Nitrate High Range handheld checker is designed to measure
high levels of nitrate in aquariums and marine biology applications.
HI782 features a single-button operation system and is easy to use. The
large LCD is easy to read and the auto shut-off feature assures the battery
will not be drained.
Specifications
Functional Description & LCD Display
Measurement Procedure
- Press the ON/OFF button to turn the checker on. All segments will be displayed for a few seconds, followed by “Add”, “C.1” with “Press” blinking.
- Fill the cuvette with 10 mL of unreacted sample and replace the cap. Insert the cuvette into the checker and close the cap.
- Press the ON/OFF button. When the display shows “Add”, “C.2” with “Press” blinking, the checker is zeroed.
- Remove the cuvette, unscrew the cap and add the content of one packet of HI782-0 Marine Nitrate High Range reagent. Replace the cap and shake vigorously for 2 minutes.
- Insert the cuvette into the checker and close the cap. Press and hold the ON/OFF button. The display will show the countdown prior to the measurement. Alternatively, wait 7 minutes and press the button.
- When the timer ends the checker will perform the reading. The instrument displays the nitrate concentration in ppm. The checker automatically turns off 10 minutes after reading.
Errors & Warnings
The checker shows clear warning messages when erroneous conditions appear and when measured values are outside the expected range. The information below provides an explanation of the errors and warnings, and the recommended action to be taken.
Light High:
There is an excess amount of ambient light reaching the detector. Please check
the preparation of the zero cuvette.
Light Low:
There is not enough light to perform a measurement. Please check the
preparation of the zero cuvette.
Inverted Cuvettes:
The sample and the zero cuvettes are inverted. Swap the cuvettes and repeat
the measurement.
Under Range:
A blinking “0.0” indicates that the sample absorbs less light than the zero
reference. Check the measurement procedure and make sure to use the same
cuvette for reference (zero) and measurement.
Over Range:
The maximum concentration value displayed blinking indicates the measured
value is outside the limits of the method. Verify that the sample does not
contain any debris. Dilute the sample and repeat the measurement.
Battery Low:
The battery level is too low for the checker to function properly. Replace
the battery with a new one.
Drained Battery:
The battery is drained and must be replaced. Replace the battery with a new
one and restart the checker.