1107-UT Differential Manometer With 2 Pressure Port
H88.0.2X.DK2RS-1.0
Operating Manual
EN
RS 1107-UT RS 1113-UT
Stock number: 205-0963 Stock number: 205-0964
rspro.com
page 2 of 24
Operating Manual RS 1107, RS 1113
Table of contents
1
About this documentation ………………………………………………4
1.1 Purpose of the document……………………………………………………….4
1.2 Legal notices ………………………………………………………………………. 4
1.3 Further information ……………………………………………………………….4
2
Safety ……………………………………………………………………………5
2.1 Explanation of safety symbols ………………………………………………..5
2.2 Foreseeable misuse ……………………………………………………………..5
2.3 Safety instructions ……………………………………………………………….. 6
2.4 Intended use………………………………………………………………………..6
3
The product at a glance ………………………………………………….7
3.1 The RS 1100 manometer series ……………………………………………..7
3.2 Display elements ………………………………………………………………….7
3.3 Operating elements ………………………………………………………………8
3.4 Connections ……………………………………………………………………….. 9
4
Operation …………………………………………………………………….10
4.1 Opening the configuration menu …………………………………………… 10
4.2 Open the adjustment menu…………………………………………………..12
5
Measurement Basics…………………………………………………….13
5.1 Special functions ……………………………………………………………….. 13
5.1.1 NVLL Tare function ………………………………………………………………. 14
5.1.2 F,NE High-resolution measurement with 0.1 Pa (RS 1107)…………14
5.1.3 F,NE High-resolution measurement with 1 Pa (RS 1113)……………15
5.1.4 AVR 0:02 / AVR 0:05 / AVR 0: 10………………………………………………..16
6
Operation and maintenance ………………………………………….17
6.1 Operating and maintenance notices………………………………………. 17
6.2 Battery………………………………………………………………………………17
rspro.com
Operating Manual RS 1107, RS 1113
page 3 of 24
6.2.1 6.2.2
7
8 8.1 8.2
9 9.1
Battery indicator…………………………………………………………………. 17 Changing battery
……………………………………………………………….. 17
Error and system messages………………………………………….18
Technical data ……………………………………………………………..20 RS 1107
…………………………………………………………………………… 20 RS 1113 …………………………………………………………………………… 22
Service ………………………………………………………………………..24 Manufacturer
…………………………………………………………………….. 24
page 4 of 24
Operating Manual RS 1107, RS 1113
1
About this documentation
1.1
Purpose of the document
_
This document is intended as a quick reference option.
1.2
Legal notices
This document is entrusted to the recipient for personal use only. Any impermissible transfer, duplication, translation into other languages or excerpts from this operating manual are prohibited.
The manufacturer assumes no liability for print errors.
1.3
Further information
_
Software version of the product:
V1.1 or later
_
For the exact product name, refer to the type plate on the rear side of the
product.
rspro.com
Operating Manual RS 1107, RS 1113
page 5 of 24
2
Safety
2.1
Explanation of safety symbols
_
DANGER
This symbol warns of imminent danger, which can result in death, severe bodily
injury, or severe property damage in case of non-observance.
_
CAUTION
This symbol warns of potential dangers or harmful situations, which can cause
damage to the device or to the environment in case of non-observance.
_
NOTE
This symbol indicates processes, which can have a direct influence on operation or can trigger an unforeseen reaction in case of non-observance.
2.2
Foreseeable misuse
_
The fault-free function and operational safety of the product can only be guaranteed if applicable safety precautions and the device-specific safety instructions for this document are observed.
If these notices are disregarded, personal injury or death, as well as property damage can occur.
DANGER
Incorrect area of application!
In order to prevent erratic behaviour of the product, personal injury and
property damage, the product must be used exclusively as described in the
chapter Description in the operating manual.
The product is not suitable for use in explosion-prone areas!
The product must not be used for diagnostic or other medical purposes on
patients!
For measurements requiring devices that are subject to authorisation or
special approvals, this product is not a substitute for such products and can
only be used as an aid in preparatory or comparison measurements!
page 6 of 24
Operating Manual RS 1107, RS 1113
_
2.3
Safety instructions
_
NOTE
This product does not belong in children’s hands!
2.4
Intended use
The device is designed as a manometer and measures even the smallest pressure differential pressures of up to ±200 hPa with a maximum resolution of up to 0.1 Pa in air or in non-corrosive/non-ionising gases between the two pressure connections.
RS 1107
RS 1113
Differential pressure
±200 hPa
±2000 hPa
Max. resolution
0.1 Pa
1 Pa
Usual applications include precise measurements of filter condition, gas flow
pressure, draught, leak integrity, dynamic pressure flow speed.
The pressure connection is made at the supplied interchangeable pressure
connection ports with suitable hoses.
The product must only be used under the conditions and for the purposes for
which it was designed.
It must be handled with care and used according to the technical data (do not
throw, strike, etc.). Suitable measures must be used to protect the pressure
connections and be protected from dirt and moisture.
rspro.com
Operating Manual RS 1107, RS 1113
3
The product at a glance
3.1
The RS 1100 manometer series
page 7 of 24
LCD Display
Front view
3.2
Display
Display elements
Top view
Battery indicator
_
Evaluation of the battery status
Unit display
_
Display of the units or Min/Max/Hold information text
Main display
_
Measurement of the current pressure or value for min/max/hold
Auxiliary display
_
Measurement of the current pressure in Min/Max/Hold mode
Bar graph
Trend display with the special function F,NE
page 8 of 24
Operating Manual RS 1107, RS 1113
3.3
Operating elements
_
On / Off button
Press briefly Long press
_
Switch on the product Activate / deactivate lighting Switch off the product Reject changes in a menu
Up / Down button
Press briefly
Display of the min/max value
Change value of the selected parameter
Long press
Reset the min/max value of the current measurement
Both simultaneously
Rotate display, overhead display
_
rspro.com
Operating Manual RS 1107, RS 1113
page 9 of 24
Function key
Press briefly
Freeze measurement (Hold)
Return to measurement display
Call up next parameter
Long press, 2s
Start menu configuration, (ONF appears in the display
Close menu, changes are saved
Long press, 4s
Depending on the selected special function: Activation of the Tare function NVLL, high-resolution measurement F,NE or rapid measurement with mean value AVR
3.4
Connections
Universal connection
Interchangeable pressure connection via G1/8″ thread.
page 10 of 24
Operating Manual RS 1107, RS 1113
4
Operation
4.1
Opening the configuration menu
1. Press the Function key for 2 seconds to open the Configuration menu.
2. (ONF appears in the display. Release the Function key.
_
Parameter Values
Meaning
_
Display unit
UN,T
Pa HPa mBaR BaR PSI
mmKG
_
Activatable special functions
(only at RS 1107 ) (only at RS 1113 )
NVLL
Tare function available
F,NE
High-resolution measurement with 0.1 Pa (RS
- or 1 Pa (RS 1113) activatable
AVR 0:02 / AVR Rapid measurement with mean value over 2 s / 5 s 0:05 / AVR 0:
10 / 10 s activatable
_
rspro.com
Operating Manual RS 1107, RS 1113
page 11 of 24
Measuring rate
RATE
_
Resolution
SLO FAST
RANG
_
AVTO K,
LO
_
Shut-off time
Selection of the measurement speed Slow Fast
Selection display resolution Automatic switchover Adjusted to the highest
value, e.g. -200.0 .. +200.0 hPa (RS 1107)
-2000 .. +2000 hPa (RS 1113) Adjusted to the lowest value, e.g. -20.00 .. +20.00 hPa (RS 1107)
-200.0 .. +200.0 hPa (RS 1113)
POFF OFF
No automatic shut-off
_
Backlight
0:15 0:30 1:00 4:00 Automatic shut-off after a selected time in hours
12:00
and minutes, during which no buttons have been
pressed
L,TE
_
OFF
Backlight deactivated
0:15 0:30 1:00 4:00 Automatic shut-off of the backlight after a selected time
in minutes and seconds, during which no
buttons have been pressed
ON
No automatic shut off of the backlight
page 12 of 24
Operating Manual RS 1107, RS 1113
Factory settings
IN,T
NO
Use current configuration
YES
Reset product to factory settings. IN,T DONE appears
in the display
4.2
Open the adjustment menu
1. The product is switched off
2. Press and hold the Down button.
3. Press the On/Off button to switch on the product.
4. Release the On/Off button after 1 second and then the Down button in order to call up the adjustment menu. The display shows the first parameter.
Parameter Values
Meaning
_
Zero point correction: offset PR.OF
0.00 -5.00 .. 5.00 -50.0 .. 50.0 Gradient correction: slope PR.SL 0.00 -5.00
.. 5.00
No offset Offset active (in hPa at RS 1107) Offset active (in hPa at RS
1113)
No slope Slope active (in %)
Formula:
Zero point correction:
Displayed value = measured value offset
Gradient correction: Display = (measured value – offset) * (1 + slope / 100)
rspro.com
Operating Manual RS 1107, RS 1113
page 13 of 24
5
Measurement Basics
_
CAUTION
Air pressure at port variant UT!
With higher pressures greater than 1 bar, the hoses must be secured to prevent
unintended loosening. Suitable hose clamps are used for this purpose.
6×1 mm PVC. Up to 5 bar rel., vacuum-suitable! 6×1 mm PE. Up to 10 bar
rel., vacuum-suitable! 6×1 mm PUR. Up to 9 bar rel., vacuum-suitable!
5.1
Special functions
_
With the special functions that can be selected via the Configuration menu, the device can be optimised for special measuring tasks. After it is switched on, the device starts up in standard measuring mode, the relevant special function is started by pressing and holding the Function key for 4 s.
page 14 of 24
Operating Manual RS 1107, RS 1113
5.1.1
NVLL Tare function
The special function FVNC NVLL has been selected in the configuration menu.
The display can be zeroed by pressing the Function key for 4 s. If the tare
function is
activated, NVLL blinks in the lower display. The tare function can be reset by
pressing the Function key again for 4 s.
NOTE
The tare function is independent of the zero point correction accessible via the settings menu.
5.1.2
F,NE High-resolution measurement with 0.1 Pa (RS 1107)
High-resolution measurement for the finest adjustment work, 4 Pascal Test (test of chimney draft with living-space-independent single combustion) and many other finelyadjusted pressure applications.
In the Configuration menu, the special function FVNC F,NE has been selected.
The high-resolution measurement can be activated by pressing and holding the
Function key for 4 s. Then the sensor is immediately zeroed and the optimised
parameters for this measurement are activated.
CAUTION
When starting the special function, make sure that there is no pressure at the
connections.
NOTE
The increased current consumption in this mode decreases battery life.
The quickly determined measurement replaces other devices, such as a U-tube
manometer. The four bars in the lower display provide additional support.
The two middle bars appear: Measurement is stable Left bars appear: the
measurement decreases Right bars appear: the measurement increases
_
By pressing and holding the Function key for 2 s., the special function can be
activated. END FVNC appears in the display.
rspro.com
Operating Manual RS 1107, RS 1113
page 15 of 24
5.1.3
F,NE High-resolution measurement with 1 Pa (RS 1113)
High-resolution measurement for the finest adjustment work and many other applications with the finest adjustment of pressure.
In the Configuration menu, the special function FVNC F,NE has been selected.
The high-resolution measurement can be activated by pressing and holding the
Function key for 4 s. Then the sensor is immediately zeroed and the optimised
parameters for this measurement are activated.
CAUTION
When starting the special function, make sure that there is no pressure at the
connections.
NOTE
The increased current consumption in this mode decreases battery life. The
quickly determined measurement replaces other devices, such as a U-tube
manometer. The four bars in the lower display provide additional support.
The two middle bars appear: Measurement is stable Left bars appear: the
measurement decreases Right bars appear: the measurement increases
_
By pressing and holding the Function key for 2 s., the special function can be
activated. END FVNC appears in the display.
page 16 of 24
Operating Manual RS 1107, RS 1113
5.1.4
AVR 0:02 / AVR 0:05 / AVR 0: 10
Fast measurement with mean value over 2 s / 5 s / 10 s
Mean value mode for measurement of heavily fluctuating pressures.
In the Configuration mode, a special function AVR 0:02, AVR 0:05 or AVR 0: 10
has been selected.
By pressing and holding the Function key for 4 s. the measurement with mean
value can be activated.
Heavily fluctuating values arise particularly with dynamic
pressure/compression measurements in chimney draft tests of forced-air burners
and, consequently, conventional electronic manometers are not adequate for
task. This special function optimises the device for this application purpose.
The different mean value times of 2, 5 or 10 seconds can be selected depending
on the requirement.
The first parameter is shown in the auxiliary display.
By pressing and holding the Function key for 2 s., the special function can be
activated. END FVNC appears in the display.
If the Tare function is activated when called up, this special function AVR
can be reset by pressing and holding the Function key for 4 s. In order to
reactivate the Tare, the special function must be switched in the
configuration menu.
rspro.com
Operating Manual RS 1107, RS 1113
page 17 of 24
6
Operation and maintenance
6.1
Operating and maintenance notices
_
NOTE
Pressure connections must be protected from soiling.
6.2
Battery
6.2.1
_
Battery indicator
If the empty frame in the battery display blinks, the batteries are depleted and must be replaced. However, the device will still operate for a certain length of time.
If the BAT display text appears in the main display, the battery voltage is no longer adequate for operation of the product. The battery is fully depleted.
6.2.2
_
Changing battery
DANGER
Danger of explosion! Using damaged or unsuitable batteries can generate heat,
which can cause the batteries to crack and possibly explode!
Only use high-quality and suitable alkaline batteries!
_
CAUTION
Damage!
If the batteries have different charge levels, leaks and thus damage to the
product can occur.
Use new, high-quality batteries! Do not use different types of batteries!
Remove depleted batteries and dispose of them at a suitable collection
point.
page 18 of 24
Operating Manual RS 1107, RS 1113
_
NOTE
This symbol indicates processes, which can have a direct influence on
operation or can trigger an unforeseen reaction in case of non-observance.
NOTE
Read the following handling instructions before replacing batteries and follow
them step by step. If disregarded, the product could be damaged or the
protection from moisture could be diminished.
1. Unscrews the Phillips screws (A)and remove the cover.
2. Carefully replace the two Mignon AA batteries (B). Ensure that the
polarity is correct! It must be possible to insert the batteries in the
correct position without using force.
3. The O-ring (C) must be undamaged, clean and positioned at the intended
depth. In order to facilitate assembly and avoid damage, a suitable grease can
be applied.
4. Fit the cover on evenly. The O-ring must remain at the intended depth!
5. Tighten the Phillips screws (A).
7
Error and system messages
Display —-
Meaning
Calculation not possible
Possible causes Remedy
Measurement data Waiting for data collection acquisition is running
rspro.com
Operating Manual RS 1107, RS 1113
page 19 of 24
No display, unclear characters or no response when buttons are pressed
Battery depleted Battery depleted Replace battery
System error
Error in the product Send in for repair
Product is defective Product is defective
BAT
Battery depleted Battery depleted Replace battery
BAT LO
Battery depleted Battery depleted Replace battery
ERR.1
Measuring range Measurement too Stay within allowable meas-
exceeded
high
urement range
Product is defective Send in for repair
ERR.2
Measuring range is Measurement too Stay within allowable meas-
undercut
low
urement range
Product is defective Send in for repair
ERR.3
Display range has been exceeded
Incorrect display unit Correct setting Incorrect resolution Deactivate function F,NE Function active
ERR.4
Display range has been undercut
Incorrect display unit Correct setting Incorrect resolution Deactivate function F,NE Function active
SYS ERR
System error
Error in the product Switch product on/off Replace batteries Send in for repair
page 20 of 24
Operating Manual RS 1107, RS 1113
8
Technical data
8.1
RS 1107
Measuring range
Accuracy Overload Pressure connection Measuring cycle Display Standard
function Activatable special functions
Adjustment
Measuring range (Hi)
Measuring range (Lo)
-200.0 .. +200.0 hPa (mbar)
-2.900 .. +2.900 PSI
-150.0 .. +150.0 mmHg (Torr)
-2000 .. +2000 Pa -20.00 .. +20.00 hPa (mbar)
-20.00 .. +20.00 mmHg (Torr)
± 0,1 % FSS typical (at nominal temperature 25 °C) ± 1 % FSS max.
Max. ± 1700 hPa
2 hose connections, interchangeable with G1/8 universal ports
FAST: approx. 25 measurements per second SLO: approx. 2.5 measurements per second
3-line segment LCD, additional symbols, illuminated (white, duration adjustable)
Min/Max/Hold
Auto-power-Off function / if activated, the product switches off automatically
NVLL: Tare function F,NE: With 0.1 Pa resolution AVR: Averaging over 2 s / 5 s / 10 s
Zero point and gradient adjustment
rspro.com
Operating Manual RS 1107, RS 1113
page 21 of 24
Housing
Break-proof ABS housing
Protection rating IP67 (pressure connections must be protected from soiling and moisture)
Dimensions
108 54 28 mm without pressure connection
LWH [mm] and weight
150 g incl. battery
Operating conditions
-20 to +50 °C; 0 to 95 % r.h. (temporarily 100 % r.h.)
Storage temperature
-20 to +70 °C
Current supply
2*AA battery (included in the scope of delivery)
Current consumption/
Battery life
approx. 1 mA (slow measurement SLO) Operating time approx. 3000 h
Battery indicator Directives and standards
4-stage battery status indicator,
Note for low battery voltage: “BAT LO”
The devices conform to the following Directives of the Council for the
harmonisation of legal regulations of the Member States:
2014/30/EU EMC Directive 2011/65/EU RoHS Applied harmonised standards: EN
61326-1:2013 Emission limits: class B
Immunity according to table 2 Additional errors: < 1 % FS EN 50581:2012
The device is intended for mobile use and/or stationary operation in the scope
of the specified operating conditions without further limitations.
page 22 of 24
Operating Manual RS 1107, RS 1113
8.2
RS 1113
Measuring range
Accuracy Overload Pressure connection Measuring cycle Display Standard
function Activatable special functions
Calibration
Measuring range (Hi)
Measuring range (Lo)
-2000 .. +2000 hPa (mbar)
-2.000 .. +2.000 bar
-29.00 .. +29.00 PSI
-1500 .. +1500 mmHg (Torr)
-200.0 .. +200.0 hPa (mbar)
-200.0 .. +200.0 mmHg (Torr)
± 0,1 % FSS typical (at nominal temperature 25 °C) ± 1 % FSS max.
Max. ± 3100 hPa
2 hose connections, interchangeable with G1/8 universal ports
FAST: approx. 25 measurements per second SLO: approx. 2.5 measurements per second
3-line segment LCD, additional symbols, illuminated (adjustable white, permanent illumination)
Min/Max/Hold
Auto-power-Off function / if activated, switches the product off automatically
NVLL: Tare function F,NE: With 1 Pa resolution AVR: Averaging over 2 s / 5 s / 10 s
Zero point and gradient adjustment
rspro.com
Operating Manual RS 1107, RS 1113
page 23 of 24
Housing
Break-proof ABS housing
Protection rating IP67 (pressure connections must be protected from soiling and moisture)
Dimensions
108 54 28 mm without pressure connection
LWH [mm] and weight
150 g incl. battery
Operating conditions
-20 to +50 °C; 0 to 95 % r.h. (temporarily 100 % r.h.)
Storage temperature
-20 to +70 °C
Current supply
Current requirement/
Battery life
2*AA battery (included in the scope of delivery) approx. 1 mA (slow measurement SLO) Operating time approx. 3000 h
Battery indicator
4-stage battery status indicator, Note for low charge level: “BAT LO”
Directives and standards
The devices conform to the following Directives of the Council for the
harmonisation of legal regulations of the Member States:
2014/30/EU EMC Directive
2011/65/EU RoHS
Applied harmonised standards:
EN 61326-1:2013 Emission limits: class B Immunity according to table 2
Additional errors: < 1 % FS
EN 50581:2012
The device is intended for mobile use and/or stationary operation in the scope
of the specified operating conditions without further limitations.
page 24 of 24
Operating Manual RS 1107, RS 1113
9
Service
9.1
Manufacturer
If you have any questions, please do not hesitate to contact us:
Contact
RS Components Limited Birchington Road Corby Northamptonshire NN17 9RS WEE/GF0002ZR
rspro.com