MyBones Marodyne LiV Device User Guide

June 5, 2024
MyBones

MyBones Marodyne LiV Device

MyBones Marodyne LiV Device
PRODUCT

Before using your Marodyne LiV device, please read and familiarise yourself with the ‘Instructions for Use’ guide packaged with your device

  1. Set Up

    • Place your device on a dry, hard and level floor (e.g. a tile/stone floor or a thin carpet)
    • Ensure the device is positioned where you intend to use it. Once the device is powered on, it must not be moved
  2. Power Assembly

    • Connect the power cord into the switching adaptor
    • Plug the power cord into the back of the device
    • Connect the AC power plug into a mains power socket. A green light will display on the switching adaptor to indicate the device is receiving power
    • Switch the device on using the rocker switch located at the back of the device. A blue light will signal it is successfully switched on
  3. Getting StartedMyBones Marodyne LiV Device FIG 1

    • When first switched on, the device will undergo a series of light cycles as it calibrates
    • Once the calibration is complete, the word ‘Go’ will display on the device and a beep indicates it’s ready for use. A single green light at the centre of the feet indicates the device is in standby mode, and you can begin your session
  4. LiV SessionMyBones Marodyne LiV Device FIG 2

    • Do not wear shoes while using the device. Socks are permitted

    • Step onto the device and place your feet about a shoulder width apart. It is recommended to stand on the device so that your big toes are in line with the bottom of the Marodyne LiV logo
      Correct foot placement

    • Once you step onto the device it will start to vibrate gently

    • When you are standing correctly on the device you will feel the gentle vibrations in your jaw

    • A timer on the device’s display will count down from 10 minutes to indicate how much time is left of your session

    • If you step off the device the timer will automatically pause. You have 5 minutes to step back onto the device before the timer resets and the device re-enters standby mode

    • An orange light on the display indicates you need to correct your balance. An alarm may sound, but there is no need to step off the device. If you do not correct your balance, an ‘Eror’ message will appear on the display. To restart the device, turn the rocker switch at the back of the device off and on again

    • The device will emit a short beep when your 10-minute session is complete

  5. Finish

    • Once your 10-minute session is complete, step off the device
    • The device will then automatically enter a standby mode

Do not attempt to move the device whilst it is in standby mode. Please ensure that your Marodyne LiV device is turned off at the rocker switch first
Always ensure you turn your Marodyne LiV device off using the rocker switch first before switching it off at the wall
When shipping or transporting the device in its case, always ensure the protective foam band is fastened securely around the side of the device, tucked just below the platform rim
if you have moved the device whilst in standby mode, please switch it off on the rocker switch, and then switch if off at the wall. Leave the device for 15 minutes before turning it back on. SETsquared Bristol, Approach Road, Temple Meads, Bristol, BS1 6QH info@mybones.health | 0117 253 0400 V2 – 08/04/2022

Quick Troubleshooting & FAQs

  1. Should I wear shoes?
    You should not wear shoes while using the device

  2. How do I know I am on the device correctly?
    A single green light in the centre of the display will indicate you are positioned correctly. You should be able to feel the vibrations all the way up to your jaw

  3. My device has stopped mid-session, what should I do?
    Try shifting your weight from side to side or rocking gently back and forth onto the balls of your feet to restart the vibrations. If you remain completely still for too long, the sensors inside the Marodyne LiV device may interpret that you have stepped off and will pause the session to conserve energy. Alternatively, you may need to stand slightly further forward on the device to ensure your weight is correctly positioned over the sensors. It is important that your feet are positioned either side of the display

  4. What do the yellow lights on the display indicate?
    The yellow lights indicate that you may be spreading your weight unevenly. You can use the lights to show you how to adjust. See the figure below for an explanation of what each light indicates:

  5. Why do the vibrations feel weaker when I adjust my stance on the device?
    The device may reduce the intensity of the vibrations it emits if it detects your balance is off-centre. To receive the optimum level of vibration, ensure your balance is centred with your feet placed evenly apart

  6. Why is my device making a rumbling noise?
    Incorrect posture or balance may cause the device to emit a mechanical rumbling noise. To stop the noise, adjust your posture and balance slightly

  7. How should I store my device?
    The Marodyne LiV device should be stored flat with its feet facing down

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>