ResMed N20 Airfit Nasal Mask
Intended use
The AirFit N20 Classic channels airflow non-invasively to a patient from a
positive airway pressure (PAP) device such as a continuous positive airway
pressure (CPAP) or bilevel device.
The AirFit N20 Classic is:
- to be used by patients weighing more than 30 kg for whom positive airway pressure has been prescribed
- intended for single-patient re-use in the home environment and multi-patient re-use in the hospital/institutional environment.
GENERAL WARNINGS
- The vent holes must be kept clear.
- The mask should only be used with CPAP or bilevel devices recommended by a physician or respiratory therapist.
- The mask should not be used unless the device is turned on. Once the mask is fitted, ensure the device is blowing air. Explanation: CPAP and bilevel devices are intended to be used with special masks (or connectors) which have vent holes to allow a continuous flow of air out of the mask. When the device is turned on and functioning properly, new air from the device flushes the exhaled air out through the mask vent holes. However, when the device is not operating, insufficient fresh air will be provided through the mask, and the exhaled air may be rebreathed. Rebreathing of exhaled air for longer than several minutes can, in some circumstances, lead to suffocation. This applies to most models of CPAP or bilevel devices.
- Follow all precautions when using supplemental oxygen.
- Oxygen flow must be turned off when the CPAP or bilevel device is not operating, so that unused oxygen does not accumulate within the device enclosure and create a risk of fire.
- Oxygen supports combustion. Oxygen must not be used while smoking or in the presence of an open flame. Only use oxygen in well ventilated rooms.
- At a fixed rate of supplemental oxygen flow, the inhaled oxygen concentration varies, depending on the pressure settings, patient breathing pattern, mask, point of application and leak rate. This warning applies to most types of CPAP or bilevel devices.
- The technical specifications of the mask are provided for your clinician to check that they are compatible with the CPAP or bilevel device. If used outside specification or if used with incompatible devices, the seal and comfort of the mask may not be effective, optimum therapy may not be achieved, and leak, or variation in the rate of leak, may affect the CPAP or bilevel device function.
- Discontinue using this mask if you have ANY adverse reaction to the use of the mask, and consult your physician or sleep therapist.
- Using a mask may cause tooth, gum or jaw soreness or aggravate an existing dental condition. If symptoms occur, consult your physician or dentist.
- As with all masks, some rebreathing may occur at low CPAP pressures.
- Refer to your CPAP or bilevel device manual for details on settings and operational information.
- Remove all packaging before using the mask.
- The mask must be used under qualified supervision for users who are unable to remove the mask by themselves. The mask may not be suitable for those predisposed to aspiration.
- Avoid connecting flexible PVC products (eg, PVC tubing) directly to any part of the mask. Flexible PVC contains elements that can be damaging to the materials of the mask, and may cause the components to crack or break.
- The N20 line of nasal CPAP masks are not intended to be used simultaneously with nebuliser medications that are in the air path of the mask/tube.
- The mask contains an exhaust port safety feature to enable normal breathing. The mask should not be worn if this safety feature is damaged or missing.
Using your mask
When using your mask with ResMed CPAP or bilevel devices that have mask setting options, refer to the Technical specifications section in this user guide for the correct setting. For a full list of compatible devices for this mask, see the Mask/Device Compatibility List on www.resmed.com/downloads/masks. If you do not have internet access, please contact your ResMed representative.
Fitting
- With both lower headgear straps released, hold the mask against your face and pull the headgear over your head.
- Bringing the lower straps below your ears, loop the headgear into the lower hooks on the mask frame.
- Undo the Velcro® tabs on the upper headgear straps. Adjust the straps evenly until the mask sits comfortably just over the nose. Reattach the Velcro tabs.
- Undo the Velcro tabs on the lower headgear straps. Adjust the straps evenly until the mask is stable. Reattach the Velcro tabs.
- Connect the device air tubing to the elbow. Attach the elbow to the mask by pressing the side buttons and pushing the elbow into the mask, ensuring it clicks in on both sides.
- Your mask and headgear should be positioned as shown in the illustration
Adjustment
If necessary, slightly adjust the position of the mask for the most comfortable fit. Ensure that the cushion is not creased and the headgear is not twisted.
1. Turn on your device so that it is blowing air.
Adjustment tips
To resolve any leaks at the upper part of the mask, adjust the upper headgear
straps. For the lower part, adjust the lower headgear straps. Adjust only
enough for a comfortable seal. Ensure the headgear straps are not
overtightened or too loose.
Removal
- Unloop the lower headgear straps from the lower hooks of the frame.
- With both lower straps released, pull the headgear back over your head.
Disassembly
If the mask is connected to your device, disconnect the device air tubing from the elbow.
- Pull the straps out of the frame.
- Remove the elbow from the mask by pressing the side button and pulling the elbow away.
- Pull the cushion away from the frame.
Reassembly
-
Push the cushion into the frame.
Note: Ensure the cushion mask size indicator is facing downwards. -
Attach the top headgear straps to the mask frame by looping onto each of the upper headgear hooks.
-
Attach the elbow by pressing the side buttons and pushing into the mask.
Cleaning your mask at home
It is important to follow the steps below to get the best performance out of your mask.
WARNING
- As part of good hygiene, always follow cleaning instructions and use a mild liquid detergent. Some cleaning products may damage the mask, its parts and their function, or leave harmful residual vapours that could be inhaled if not rinsed thoroughly.
- Regularly clean your mask and its components to maintain the quality of your mask and to prevent the growth of germs that can adversely affect your health.
CAUTION
If any visible deterioration of a system component is apparent (cracking,
crazing, tears etc), the component should be discarded and replaced.
Daily/After each use
Before cleaning, take the mask components apart according to the disassembly
instructions.
- Soak and agitate the cushion and elbow in warm water (approximately 30°C) using a mild liquid detergent for up to 10 minutes.
- Clean the components with a soft bristle brush, paying particular attention to the small vent holes on the elbow.
- Rinse the components under warm flowing drinking quality water.
- Inspect each component to ensure that they are visually clean and free of detergent residue. If required, repeat washing.
- Allow the components to dry out of direct sunlight before assembling.
Weekly
- Soak and agitate the separated headgear and frame in warm water (approximately 30°C) using a mild liquid detergent for up to 10 minutes.
- Rinse the components under warm flowing drinking quality water until they are free of detergent residue.
- Squeeze the headgear to remove excess water.
- Allow the components to dry out of direct sunlight before assembling.
Reprocessing the mask between patients
Reprocess this mask when using between patients. Cleaning, disinfection and sterilisation instructions are available on www.resmed.com/downloads/masks. If you do not have internet access, please contact your ResMed representative.
Troubleshooting
Problem/possible cause | Solution |
---|---|
Mask is too noisy. | |
Vent is dirty. | Clean the vent according to the instructions. |
Problem/possible cause | Solution |
--- | --- |
Mask is incorrectly assembled. | Disassemble the mask, then reassemble |
according to the instructions.
Mask is leaking.| Refit or reposition your mask.
Readjust the headgear straps to bring the mask closer to the face and improve your seal.
Check that your mask is assembled correctly.
Mask leaks around face.|
Mask may have been fitted incorrectly.| Pull the mask away from your face and
reposition, ensuring that the lower corners of the mask fully enclose your
nose. If necessary, readjust the headgear straps to keep the mask in the right
place.
Cushion may be dirty.| Clean the cushion according to the instructions.
Cushion membrane is creased or folded.| Remove and refit your mask according
to the fitting instructions. Ensure that you position the cushion correctly
over your nose before pulling the headgear over your head. Do not slide the
mask down your face during fitting as this may lead to folding or curling of
the cushion.
Mask is not tightened correctly.| Ensure that the headgear straps are not
overtightened or too loose. You may find that this mask does not need to be
tightened as much as other nasal masks.
Check that the headgear is not twisted.
Mask may be the wrong size.| Talk to your clinician.
Mask does not fit properly.|
Mask is not positioned correctly.| Check that the mask size indicator on the
bottom of the cushion is facing downward.
Mask is incorrectly assembled.| Disassemble the mask, then reassemble
according to the instructions.
Mask may be the wrong size.| Talk to your clinician.
Technical specifications
Pressure-flow
curve
| The mask contains passive venting to protect against rebreathing.
As a result of manufacturing variations, the vent flow rate may vary.
---|---
| Pressure
(cm H2O)
| Flow (L/min)
4| 20
7| 27
11| 35
14| 40
17| 44
20| 49
24| 54
27| 58
| | 30| 61
Dead space
information
| Physical dead space is the empty volume of the mask to the end of
the swivel. Using the large cushions it is 55.8 mL.
Therapy pressure| 4 to 30 cm H2O| |
Resistance| Drop in pressure measured (nominal) at 50 L/min: 0.4 cm H2O
at 100 L/min: 1.5 cm H2O
| |
Environmental conditions| Operating temperature: 5°C to 40°C
Operating humidity: 15% to 95% non-condensing Storage and transport temperature: -20°C to +60°C
Storage and transport humidity: up to 95% non-condensing
Sound| DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance
with ISO 4871. The A-weighted sound power level of the mask is 26 dBA, with uncertainty of 3 dBA. The A-weighted sound pressure level of the mask at a distance of 1 m is 18 dBA, with uncertainty of
3 dBA.
Gross dimensions| 141 mm (H) x 73 mm (W) x 90 mm (D)
Mask fully assembled with elbow assembly (no headgear)
service life| The service life of the AirFit N20 Classic mask system is dependent on the intensity of usage, maintenance, and environmental conditions to which the mask is used or stored. As this mask system and its components are modular in nature, it is recommended that the user maintain and inspect it on a regular basis, and replace the mask system or any components if deemed necessary or according to the ‘visual criteria for product inspection’ in the ‘Cleaning your mask at home’ section of this guide. Refer to the ‘Mask components’ section of this guide for information of how to order replacement
parts.
---|---
Mask setting options| For AirSense, AirCurve and S9: Select ‘Pillows’.
For other devices: Select ‘SWIFT’ (if available), otherwise select ‘MIRAGE’ as the mask option.
Notes
- This product is not made with PVC or phthalates such as DEHP, DBP or BBP.
- This product is not made with natural rubber latex.
- The manufacturer reserves the right to change these specifications without notice.
Storage
Ensure that the mask is thoroughly clean and dry before storing it for any
length of time. Store the mask in a dry place out of direct sunlight.
Disposal
This mask does not contain any hazardous substances and may be disposed of
with your normal household refuse.
Symbols
The following symbols may appear on your product or packaging:
- Catalog number
- Batch code
- Humidity limitation
- Temperature limitation
- Fragile, handle with care
- Not made with natural rubber latex
- Manufacturer
- European Authorised Representative
- Keep away from rain
- This way up
- Classic nasal mask
- Device setting – Pillows
- Size – small
- Size – medium
- Size – large
Indicates a Warning or Caution and alerts you to a possible injury or explains special measures for the safe and effective use of the device Caution, consult accompanying documents
Consumer Warranty
ResMed acknowledges all consumer rights granted under the EU Directive 1999/44/EC and the respective national laws within the EU for products sold within the European Union.
ResMed Pty Ltd
1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia See ResMed.com for
other ResMed locations worldwide. AirFit is a trademark and/or registered
trademark of the ResMed family of companies. For patent and other intellectual
property information, see ResMed.com/ip. Velcro is a trademark of Velcro
Industries B.V. © 2017 ResMed. 638123/2 2017-11