owon OWP-H Series High Power DC Supply User Guide
- June 5, 2024
- owon
Table of Contents
OWP-H Series High Power DC Supply
User Guide
OWP-H Series High Power DC Supply
For product support, visit:www.owon.com.hk/download
Mar. 2022 edition V1.0.0
Copyright © LILLIPUT Company. All rights reserved.
LILLIPUT’s products are under the protection of the patent rights, including
ones that have already obtained the patent rights and those which are applying
for. The information in this manual will replace all that in the materials
published originally.
The information in this manual was correct at the time of printing. However,
LILLIPUT will continue to improve products and reserves the right to change
specifications at any time without notice.
is the registered trademark of the LILLIPUT Company.
Fujian LILLIPUT Optoelectronics Technology Co., Ltd.
No. 19, Heming Road
Lantian Industrial Zone, Zhangzhou 363005 P.R. China
Tel: +86-596-2130430
Web: www.owon.com| Fax: +86-596-2109272
E-mail: info@owon.com.cn
---|---
General Warranty
We warrant that the product will be free from defects in materials and
workmanship for a period of 2 years (1 year for accessories) from the date of
purchase of the product by the original purchaser from our company. This
warranty only applies to the original purchaser and is not transferable to a
third party.
If the product proves defective during the warranty period, we will either
repair the defective product without charge for parts and labor or will
provide a replacement in exchange for the defective product. Parts, modules,
and replacement products used by our company for warranty work may be new or
reconditioned like new. All replaced parts, modules, and products become the
property of our company.
In order to obtain service under this warranty, the customer must notify our
company of the defect before the expiration of the warranty period. The
customer shall be responsible for packaging and shipping the defective product
to the designated service center, a copy of the customer’s proof of purchase
is also required.
This warranty shall not apply to any defect, failure, or damage caused by
improper use or improper or inadequate maintenance and care. We shall not be
obligated to furnish service under this warranty a) to repair damage resulting
from attempts by personnel other than our company representatives to install,
repair or service the product; b) to repair damage resulting from improper use
or connection to incompatible equipment; c) to repair any damage or
malfunction caused by the use of not our supplies; or d) to service a product
that has been modified or integrated with other products when the effect of
such modification or integration increases the time or difficulty of servicing
the product.
Please contact the nearest Sales and Service Offices for services.
Excepting the after-sales services provided in this summary or the
applicable warranty statements, we will not offer any guarantee for
maintenance definitely declared or hinted at, including but not limited to
the implied guarantee for marketability and special-purpose acceptability. We
should not take any responsibility for any indirect, special or consequent
damages.
Back Panel
- DC output terminal: RED “+”, BLACK “-“
- Remote voltage compensation
- Duct outlet (No obstructions within 10 cm)
- RS485 interface(Female)
- CAN interface
- Dry contact/Analog interface
- AC Input
Interface
Interface
| PIN|
Function
---|---|---
Digital IO| 1| Normally open contact (Output of the dry contact)
2| Common contact (Output of the dry contact)
3| Normally closed contact (Output of the dry contact)
4| NC
5| The input of the dry contact
6
RS485| 1| 485-A
2| 485-B
3~9| NC
- Digital IO interface: PIN 1 to 3 is a dry contact output interface with complementary functions of normally open and normally closed. PIN2 is the common end of dry contact. Dry contact output capacity: 1A/30VDC or 0.15A/220VAC; PIN 5 to 6 are dry contact input interfaces, which can be set for external control of the output, external fault feedback, or external control of the buzzer;
- Analog interface: Analog interface is optional, and interface signals can be customized, two analog interface definitions as shown in the table above
- RS485 interface: Serial communication interface(Female), software using standard Modbus-RTU protocol;
- CAN interface: CAN1 and CAN2 are two internal parallel CAN bus interfaces, which facilitate the serial or parallel connections between devices. CAN communication also be used for communication between external devices;
Note: Analog interface is an optional interface (customizable), up to a maximum of four analogs, two analog input, and two analog output. Select 1-2 analog, interface see figure above; select 3-4 analog, the interface is RJ45-CAN1, 1-8 pin is defined as the positive and negative of analog input 1, the positive and negative of analog input 2, the positive and negative of analog output 1, the positive and negative of analog output 2. If you need an analog function, please inform us of the specific requirements in advance。
Voltage compensation
Figure 3: Schematic diagram of voltage compensation wiring
To use the remote voltage compensation function, use twisted-pair cables with high insulation. Positive and negative cables can not be connected inversely, as shown in the figure above. When not in use, the compensation terminals (SENSE) PIN1 and PIN2, PIN3, and PIN4 need to be shorted with short cables.
Parallel connection
The device identifies and controls the parallel output through CAN
communication. The diagram above shows the parallel connection.
Note:120 ohm is the CAN bus terminal resistor.
Front panel
Display area
The Home displays real-time operating status information of the device, including:
- Display area: Current real-time output information;
- Setting area: setting of voltage, current, and power reference values, And voltage/current priority Settings;
- Status area: buzzer, lock key state, date and time information, working time, and dry contact and application mode state (gray);
- Mode area: Control mode and output mode;
- Series/parallel status area: when multiple machines are used in series/parallel, each device will display the master/slave machine number and CAN data receiving and receiving the status of the machine (gray);
Note:
- Display elements of the status area can be hidden. When an application mode is enabled, the status of the application mode will be displayed, and when the dry contact is used, the corresponding status icon will be displayed.
- Output mode is divided into common mode and application mode. 1. Common mode: CV(Constant voltage), CV(Constant current), CP(Constant power), or CV/CC/CP (Output is not open); 2. Application mode: such as CV Steps(Constant voltage steps), CC Steps(Constant current steps), and Hybrid steps in step mode.
Home
Three Homes, including:
-
Home: displays the most comprehensive real-time working status information,
detailed in the section of “Display Area”; -
Auxiliary Home: maximizes the real-time output information;
-
Waveform Home: displays output real-time information in an intuitive waveform
manner.
Note
- The Home is the only interface for setting voltage, current, and power reference values.
- Press “ENTER” to set the sampling rate of the waveform displayed on the waveform Home. Whether the voltage, current, and power waveform are displayed can be controlled by pressing the “VOLT”, “CURR” or “POWER” key.
Operating area
Key
|
Description
---|---
VOLT| Voltage reference set
CURR| Current reference set
VOLT Double
Click
| Voltage priority switching
CURR Double Click| Current priority switching
VOLT+CURR| Power reference set
LOCK| Lock/Unlock
ON/OFF| Output ON/OFF
←↑| Left/Upshift
→↓| Right/Downshift
DEL| Delete
EXIT| Returns the previous level or exit setting
SAVE| Save current settings
RECALL| Recall the saved settings
L/R| local/remote control mode
CONF| Function Menu
Key| Description
---|---
0-9| Number set
.| DOT
ENTER| To Menu/ Confirm input
Switch between Home and Auxiliary Home
Knob| Description
Press| Menu
Confirm Input
Home:
1, Press once, Voltage set
2, Press twice, Current set
3, Press 3 times, Power set
Clockwise rotation| Increase value Upshift
Anti-Clock rotation| Reduce value Downshift
Table 2: Key description
The operation area includes the setting area, function area, digital area, and knob. See “Appendix 1” in 《OWP_H Series Use Manual》 for key details.
Basic operation
- Voltage reference setting: Press the “VOLT” key or press the “Knob” key to edit the voltage reference value, enter a valid value, and press the “ENTER” or “knob” key to confirm;
- Current reference setting: Press the “CURR” key or press the “Knob” key twice to edit the current reference value, enter a valid value, and press the “ENTER” or “knob” key to confirm;
- Power reference setting: Press the “VOLT” and “CURR” key simultaneously or press the “Knob” key three times to edit the power reference value, enter a valid value, and press the “ENTER” or “knob” key to confirm;
- Open and close output: Press the “ON/OFF” key to open the output (The “ON/OFF” key is light), and press the “ON/OFF” key to close the output (The “ON/OFF” key is OFF);
- Voltage/current priority switching: Close the output and double-click the “VOLT” or “CURR” key to switch the priority under the Home/auxiliary Home (Switch time is 1 second).
- SAVE Setting: 1. Under the Home, function setting UI or protection setting UI, if the Settings are valid, press the “SAVE” key to save the common mode data; On the application mode setting UI, if the Settings are valid, press “SAVE” key to save the application mode data;
- Recall setting: 1. On the Home/auxiliary Home, press the “RECALL” key to bring up the recall function option. Press “←↑the ” Or “→↓” key to select the data type and press the “ENTER” key to go to the recall UI for the data type. Press the”←↑” or “→↓” key to select the pre-called data and press the “ENTER” key to confirm the callback data. 2. On the application mode setting UI, press the “RECALL” key to bring up the relevant recall UI, press the”←↑” or “→↓” key to select the pre-called data, and press the “ENTER” key to confirm the callback data;
- Local/Remote mode switching: Press “L/R” to switch the local/remote mode temporarily in the Home/auxiliary Home (for the temporary test, the model is not saved); Buzzer control: Under the system setting UI, press “←↑” or “→↓” key to select Buzzer, and press “ENTER” key to enter buzzer control option. Select the corresponding level, and press the “ENTER” key to confirm;
Note:
- Common mode data includes voltage, current, and power reference values as well as parameters of function settings and protection settings in user Settings.
- When the key triggers the setting of the reference value, the preset area will be displayed above the corresponding operated element in the setting area. Enter a preset value through the number or knob key; When the knob triggers the setting of the reference value, the corresponding bit of the element to be operated in the setting area will flash. Through ” ← ↑ the ” Or ” → ↓ ” key to select the operation position, and then enter the preset value through the number or knob key.
- For local/remote mode Settings, see “LCD Menu – > User Settings – > Function setting ” section in 《OWP_H Series Use Manual 》 for details.
Appendix
Accessory
Certificate×1
Quick guide×1
1.5m input power line×1
6PIN terminal block×1
Key description
Area
| Abbreviation|
Description
---|---|---
Setting| VOLT| Voltage reference setting
CURR| Current reference setting
VOLT+CURR| Power reference setting
LOCK| Lock/unlock key
ON/OFF| Open/close output
Function| ←↑| Move the cursor one bit to the left (numeric Settings) Move up
one line
→↓| Move the cursor one bit to the right (numeric Settings) Move down one line
DEL| Deletes the value of the current bit
EXIT| Return to the previous level or exit the setting
SAVE| Save the normal data (In normal mode) Save App data (In data mode)
RECALL| Recall saved data on Home
L/R| Switch local/remote mode
CONF| Enter the function UI
Figure| 0~9| Enter figure
.| Enter decimal point “.”
ENTER| Enter the menu
Input confirm
Switch the home and the Auxiliary home
Knob| Clockwise| Increment the input value (numeric Settings) Move Up N line
Anticlockwise| Decrease the input value (numeric Settings) Move Down the N
line
Press| Enter the menu
Input confirm
Under the home:
1. Press to set voltage reference
2. Press twice to set the current reference
3. Press three times to set the power reference
4. In the reference setting state, press confirm
User setting list
Scope | Name | Description | Default |
---|---|---|---|
Communication | Baud Rate | Baud rate setting | 9600 bps |
CRC Alignment | Sending mode of CRC 16-bit check data | Little Endian | |
Modbus Address | Modbus protocol address | Ox01 | |
Function | Startup Mode | The device is in local/remote control mode after | |
powering on | Local | ||
Rise Time Of Voltage | Rise Time Of Voltage Setting | 30 mS | |
Fall Time Of Voltage | Fall Time Of Voltage Setting | 0 mS | |
Rise Time Of Current | Rise Time Of Current Setting | 30 mS | |
Fall Time Of Current | Fall Time Of Current Setting | 0 mS | |
Auto-Reco(Fault) | After the fault occurs, disable the output and check whether | ||
the output will be automatically restored after the specified time | 30 S, |
Close
Auto-output(Hold)| After power-on, whether to automatically start output after
the specified time| 30 S, Close
Timing output| Reference time: Use the clock or
power-on time as the reference time Time range Enable: Enables or disables
this time range
On/Off time: set the time range| Disable
ParalleUSeries Connection| Type of connection: independent,
parallel or series
Master/slave: the master or slave| Independent
Dry Contact output| Control mode: Disable, local or remote control
Relationship: logic related to a fault,
the startup, condition setting, or time setting Signal delay: delay from the
dry contact action after the logic is triggered| Disable
Dry Contact Input| Relationship: Disabled; Fault; Start or
buzzer| Disable
Protection| Over-Volt Value| Over-Voltage Protection Value| 105% VRated
Time of Duration| Trigger over-voltage protection time| 1000 mS
Over-Curr Value| Over-Current Protection Value| 105% IRated
Time of Duration| Trigger over-current protection time| 500 mS
Levell Overload Value| Levell overload protection value| 105% PRated
Time of Duration| Trigger Levell overload protection time| 10000 mS
Level2 Overload Value| Level2 overload protection value| 110% PRated
Time of Duration| Trigger Level2 overload protection time| 5000 mS
Level3 Overload Value| Level3 overload protection value| 120% PRated
Time of Duration| Trigger Level3 overload protection time| 1000 mS
Under-Volt Protection| Under-voltage protection switch| Disable
Protection Value| Under-voltage protection value| 10% VRated
Protection Delay| Under-voltage protection detection delay| 1000 mS
Time Of Duration| Trigger under-voltage protection time| 1500 mS
Under-Curr Protection| Under-current protection switch| Disable
Protection Value| Under- current protection value| 10% I Rated
Protection Delay| Under- current protection detection
delay| 1000 mS
Time Of Duration| Trigger under current protection time| 1500 mS
Short-Circuit Protection| Short-circuit protection switch| Disable
Protection Value| Short-circuit protection voltage value| 5% varied
Protection Delay| Short-circuit protection detection delay| 10 mS
Time Of Duration| Trigger Short-circuit protection time| 20 mS
Protection Switches| Relevant protection switches| —
Password| Password| User Default Settings| —
Reset| Factory Reset| Restoring Factory Settings (except for information
records)| —
Error Log reset| Clears fault Records| —
System Data Reset| Clears UI or all system setting|
User Data Reset| Clears selected data|
Warning list
Name | Attribute | Description | Troubleshooting |
---|---|---|---|
Write EEPROM Err | Unrecoverable | ||
error | Write EEPROM Error | Power off, Restart. |
Read EEPROM
Err| Read EEPROM Error| Power off, Restart.
Write FLASH Err| Write FLASH Error| Power off, Restart.
Read FLASH Err| Read FLASH Error| Power off, Restart.
Dill Speci Err| | Different from Master
specifications| Power off, Restart.
External Error| Recoverable
error| A fault was detected
through dry contact input| Check whether dry contact signal input is normal
and exclude alarm signal.
Driver Protect| Driver circuit error| Power off, Restart.
HW Over-Volt P| The hardware
over-voltage circuit
detects an over-voltage error| Confirm start overshoot or steady
overshoot (overshoot in working process), if it is start overshoot, can set
“priority” to “current priority”, can also set “Rise Time Of Volt” parameter
to a reasonable value(voltage priority); If it is a steady state overshoot and
the voltage is not more than 1.3 times the rated voltage, you can tum off the
hardware overvoltage protection function. If the voltage is more than 1.3
times the rated voltage, install an anti-reverse diode on the output side.
HW Over-Curr P| The hardware
over-current circuit
detects an over-current Error| Confirm start overshoot or steady
overshoot (overshoot in working process), if it is start overshoot, can set
“priority” to “voltage priority”, can also set “Rise Time Of Volt” parameter
to a reasonable
value(voltage priority); If steady-state
overshoot occurs, disable hardware
overcurrent Protection.
Over-Volt P| The software detects an over-voltage error| Confirm start
overshoot or steady
overshoot (overshoot in working process), if it is start overshoot, can set
priority to “voltage priority”, can also set “Rise Time” parameter to a
reasonable value; If the overshoot is steady state, the
“over-current protection value” or
“overcurrent duration” can be
appropriately increased;
Over-Cu rr P| The software detects an over-current error| Confirm start
overshoot or steady
overshoot (overshoot in working process), if it is start overshoot, can set
“priority” to “voltage priority”, can also set “Rise Time Of Curr” parameter
to a reasonable
value(current priority); In the case of
steady overshoot, the “overcurrent protection value” or “overcurrent duration”
can be appropriately raised.
Under-Volt P| | The software detects an under-voltage error| Check whether the
error is reasonable. If not, reset under-voltage protection parameters.
Under-Curr P| The software detects an under-current error| Check whether the
error is reasonable. If not, reset under-current protection parameters.
Short-Circuit P| The software detects an short-circuit error| Check whether
the short-circuit protection
occurs. If the short-circuit error occurs,
rectify the short-circuit error. Otherwise,
reset the short-circuit protection
parameters.
Over-Load P| The software detects an overload error| Eliminate overload error
or adjust overload protection parameters.
Over Temperature| The software detects an over- Temperature error| Check
whether the power supply air duct is blocked.
Error Resume| Automatic error recovery is enabled, recoverable errors are
detected, and recovery attempts fail 10 times| After confirming the cause of
the error and troubleshooting, restart the machine. The error alarm can be
cleared by pressing the “EXIT” key on the main UI.
key is locked| Warming| Key locked| Press the “LOCK” key to unlock it.
Return to HOME| Operation method in the home| Return to the main UI and
operate.
Close Output| Method of operation in the closed output state| Operation after
closing output
RemoteCntr:Com ms| Operate keys in remote mode| Press “UR” to switch back to
local control.
RemoteCntrAnalo g
Switching Prior| Cannot start output
during priority switching| Open output later.
Please Later!| The priority cannot be
switched again during
priority switching| Wait 1 second and switch the priority
again.
Step Mode Is En| Cannot enable other
the mode in step mode| Operation after Turn off Step mode.
Chg Mode Is En| Cannot enable other
the mode in charge mode| Turn off charging mode before operation.
Func Mode Is En| Cannot Enable other
mode In function
generator mode| Tum off function generator before
operation.
Exit Setting| Illegal operation| Operation after exiting the Settings.
Invalid Operate| | The save and call
functions are
unavailable in the
current UI| Perform operations on the correct UI.
Value Exceeds| The input value exceeds the legal range| Input valid value.
Value Too Small
Not Be Set To ‘0’| The input value cannot be ‘0’| Input valid value.
Passward Error| Incorrect password input| Input the correct password, if you
forget the password, call our company.
Unset Volt Ref| The output cannot be open without setting the voltage
reference| Set the voltage reference and start the output.
Unset Curr Ref| The output cannot be
open without setting the current reference| Set the current reference and
start the output.
Unset Power Ref| The output cannot be open without setting the power
reference| Set the power reference and start the output.
Illegal Data| Saving a data group is invalid| Save the data group after
setting it
correctly.
Full Data Space| 128 data groups are full| Delete redundant data groups and
save them.
No Dada| The recall data group is empty| Call data after saving the
corresponding data group.
AddrRange :1-247| Invalid MODBUS
address| Input valid address
Func Code Err| Communication error| Invalid function code| Operate according
to the communication protocol;
RegisterAddrErr| Invalid register address| Operate according to the
communication protocol;
Data Range Err| Illegal data| Operate according to the communication protocol;
Local Mode Err| The device is in local
control mode| Switch to remote mode