GREISINGER EBT-IF2 Capacitive Level Sensor Designed Instruction Manual
- June 17, 2024
- GREISINGER
Table of Contents
EBT-IF2 Capacitive Level Sensor Designed
Product Information
Specifications:
-
Measuring probe: internal Pt1000-sensor
-
Accuracy: (at nominal temperature)
-
Min-/Max-value memory: min- and max measured value are
stored -
Output signal: EASYBUS-protocol
-
Connection: 2-wire EASYBUS, polarity free
-
Busload: 1.5 EASYBUS-devices via interface
-
Adjusting: input of offset and scale value
-
Nominal temperature: 0 to 100 %RH
-
Operating temperature:
-
Relative humidity: 0 to 100 %RH
-
Storage temperature:
-
Housing:
-
Dimensions:
- Sleeve:
- Tube length FL:
- Tube diameter D:
- Collar tube length HL:
- Thread:
- IP rating:
-
Electrical connection: polarity free connection via 2-pol
connection cable- Cable length: 1m or on customer requirement
-
EMC:
-
Disposal instructions:
Safety Instructions:
-
Trouble-free operation and reliability of the device can only
be guaranteed if the device is not subjected to any other climatic
conditions than those stated under Specification. -
General instructions and safety regulations for electric, light
and heavy current plants, including domestic safety regulations
(e.g. VDE), have to be observed. -
If device is to be connected to other devices (e.g. via PC) the
circuitry has to be designed most carefully. Internal connection in
third party devices (e.g. connection GND and earth) may result in
not-permissible voltages impairing or destroying the device or
another device connected. -
If there is a risk whatsoever involved in running it, the
device has to be switched off immediately and to be marked
accordingly to avoid re-starting. -
Warning: Do not use this product as safety or emergency stop
devices, or in any other application where failure of the product
could result in personal injury or material damage.
Available Design Types:
-
Design type 1: standard: FL = 100mm, D = 6 mm
-
Design type 2: standard: FL = 100mm, D = 6 mm, thread =
G1/2 -
Design type 3: standard: FL = 50 mm, HL = 100 mm, D = 6 mm,
thread = G1/2
Frequently Asked Questions:
-
1. Q: How do I dispose of the device?
A: The device must not be disposed of in the regular domestic
waste. Send the device directly to us (sufficiently stamped), if it
should be disposed. We will dispose of the device appropriately and
environmentally sound.
-
1. Q: What should I do if the device is not working as
specified?
A: If the device is not working as specified, please return it
to the manufacturer for repair or maintenance.
B03.0.0X.6C-02
Operating Manual EASYBUS temperature sensor module
EBT IF… as of V1.4
page 1 of 2
Specification:
Measuring range: EBT IF1 (standard): EBT IF2 (standard): EBT IF3
(standard):
Measuring probe: Accuracy: (at nominal temperature)
Min-/Max-value memory:
please refer to type plate -30,0 … +100,0 °C -30,0 … +100,0 °C -70,0 … +400,0
°C
internal Pt1000-sensor
±0,2% of meas. value ±0,2°C (EBT-IF1, EBT-IF2) ±0,3% of meas. value ±0,2°C
(EBT-IF3)
min- and max measured value are stored
Output signal: Connection: Busload:
Adjusting:
EASYBUS-protocol 2-wire EASYBUS, polarity free 1.5 EASYBUS-devices via interface by input of offset and scale value
Ambient conditions for electronic (in sleeve):
Nominal temperature:
25°C
Operating temperature:
-25 to 70°C During operation please take care, that even at higher temperatures at the sensor tube ( >70°C) the allowable temperature range of the electronics, placed in the sleeve, may not exceed!
Relative humidity:
0 to 100 %RH
Storage temperature:
-25 to 70°C
Housing: Dimensions: Sleeve: Tube length FL: Tube diameter D:
Collar tube length HL: Thread:
IP rating:
stainless steel housing depending on sensor construction Ø 15 x 35 mm (without screwing) 100 mm or 50 mm or on customer requirement Ø 6 mm or on customer requirement (available Ø: 4, 5, 6 and 8 mm) 100 mm or on customer requirement G1/2″ or on customer requirement (available threads M8x1, M10x1, M14x1.5, G1/8″, G1/4″, G3/8″, G3/4″) IP67
Electrical connection:
polarity free connection via 2-pol connection cable Cable length: 1m or on customer requirement
EMC:
The device corresponds to the essential protection ratings established in the Regulations of the Council for the Approximation of Legislation for the member countries regarding electromagnetic compatibility (2004/108/EG). In accordance with EN61326 +A1 +A2 (appendix A, class B), additional errors: < 1% FS. The tube have to be protected sufficiently against ESD pulses, if the device is used in areas with risk of ESD. When connecting long leads adequate measures against voltage surges have to be taken.
Disposal instructions:
The device must not be disposed in the regular domestic waste. Send the device
directly to us (sufficiently stamped), if it should be disposed. We will
dispose the device appropriate and environmentally sound.
GREISINGER electronic GmbH
D – 93128 Regenstauf, Hans-Sachs-Straße 26
Tel.: +49 9402/9383-0, Fax: +49 9402/9383-33, eMail: info@greisinger.de
B03.0.0X.6C-02
page 2 of 2
Safety instructions:
This device has been designed and tested in accordance with the safety
regulations for electronic devices. However, its trouble-free operation and
reliability cannot be guaranteed unless the standard safety measures and
special safety advises given in this manual will be adhered to when using the
device.
1. Trouble-free operation and reliability of the device can only be
guaranteed if the device is not subjected to any other climatic conditions
than those stated under “Specification”.
2. General instructions and safety regulations for electric, light and heavy
current plants, including domestic safety regulations (e.g. VDE), have to be
observed.
3. If device is to be connected to other devices (e.g. via PC) the circuitry
has to be designed most carefully. Internal connection in third party devices
(e.g. connection GND and earth) may result in not-permissible voltages
impairing or destroying the device or another device connected.
4. If there is a risk whatsoever involved in running it, the device has to be
switched off immediately and to be marked accordingly to avoid re-starting.
Operator safety may be a risk if:
– there is visible damage to the device
– the device is not working as specified
– the device has been stored under unsuitable conditions for a longer time
In case of doubt, please return device to manufacturer for repair or
maintenance.
5. Warning: Do not use this product as safety or emergency stop devices, or
in any other application where failure of the product could result in personal
injury or material damage.
Failure to comply with these instructions could result in death or serious
injury and material damage.
Available design types:
Design type 1: standard: FL = 100mm, D = 6 mm
FL
sleeve
Design type 2: standard: FL = 100mm, D = 6 mm, thread = G1/2″ FL
Design type 3: standard: FL = 50 mm, HL = 100 mm, D = 6 mm, thread = G1/2″
FL
HL
D
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>