HIOK RM3542 Resistance Hitester Instruction Manual
- June 1, 2024
- HIOK
Table of Contents
- HIOK RM3542 Resistance Hitester
- Specifications
- Product Information
- Product Usage Instructions
- FAQs
- Introduction
- Measurement Preparations
- Measurement Settings
- Customizing Measurement
- System Settings
- Storing and Exporting Data
- Connecting the printer to the instrument
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
HIOK RM3542 Resistance Hitester
Specifications
- Model: RM3542
- Version: RM3542-01
- Type: Resistance HiTESTER
- Edition: Oct. 2018 Revised edition 6 RM3542A981-06 18-10H
Product Information
The RM3542 Resistance HiTESTER is a versatile device designed for measuring resistance with high precision. It features various measurement settings and functions to cater to different testing needs.
Measurement Flow
The measurement flow involves verifying package contents, following safety information, and taking necessary operating precautions before conducting measurements.
Product Usage Instructions
Chapter 1: Overview
- Product Overview and Features: Familiarize yourself with the functions and features of the Resistance HiTESTER.
- Names and Functions of Parts: Understand the different components and their functions.
- Screen Organization: Learn how information is displayed on the screen.
Chapter 2: Measurement Preparations
- Connecting the Power Cord: Plug in the power cord to supply power to the device.
- Connecting Measurement Cables and Test Fixtures: Connect the necessary cables and fixtures for measurement.
- Turning the Power On and Off: Power on/off the Resistance HiTESTER as needed.
Chapter 3: Measurement Settings (Basic Measurements)
- Pre-Operation Inspection: Perform a pre-operation check before starting measurements.
- Measurement Object Types: Select the type of object being measured.
- Setting the Measurement Speed: Adjust the measurement speed based on requirements.
FAQs
- Q: How do I enable the Comparator Function?
- A: To enable the Comparator Function, navigate to the settings menu and select the option for Comparator Function. Follow the on-screen prompts to set upper/lower thresholds as needed.
- Q: What is the Key-Lock Function?
- A: The Key-Lock Function allows you to disable specific key operations to prevent accidental inputs. You can re-enable key operations using the Key-Lock Cancel option in the system settings.
“`
iii
Contents
6 7 8 9 10 11 12
Appendix Index
iv
Contents
Task-Oriented Reference
1
Task-Oriented Reference
To minimize measurement error
Setting the measurement speed (p. 29) Setting the measurement speed integration time option (p. 44) Zero-adjustment (p. 32)
To judge measurement results
Judge measured values (comparator function) (p. 34)
Compare the measurement settings of two instruments (Settings Monitor
function) (p. 53)
To correct faulty measurements
To automatically store measured values
Confirm faulty measurement (p. 38)
Improve probe contact (Contact Improver function) (p. 47)
Test for short-circuited probe (probe short-circuit detection function) (p.
51) Compare the measurement settings of two instruments (Settings Monitor
function) (p. 53)
Store as soon as measurement is stable (Auto Memory function) (p. 71)
To print measurement results
Printing (p. 79)
To measure by PLC connection (PLC: Programmable Logic Controller)
External control (p. 85) Communications (RS-232C/GP-IB interface) (p. 97)
To connect to a computer
Communications (RS-232C/GP-IB interface) (p. 97)
To automatically send measurement data to a computer (RS-232C interface only)
To check operation
Export measurement values automatically (when finished measuring) (p. 77)
Setting Measurement Start Conditions (Trigger Source) (p. 30) Internal trigger
(INT) Calibration (p. A11)
2
Measurement Flow
Measurement Flow
Be sure to read the “Operating Precautions” (p. 7) before use.
Installing, Connecting and Turning On
Install (p. 7) Connect (p. 21)
Connecting the Power Cord (p. 22) GP-IB RS-232C
EXT. I/O SET MONITOR
Computer communications (p. 97)
Printing (p. 79)
External control (p. 85)
Turn the power on (p. 25)
Compare the settings of two instruments (p. 53) Connect the measurement cables (p. 23)
Settings
Set measurement conditions (as needed)
Confirm the screen configuration (p. 17)
Confirm the initial setup (p. 68)
· Basic Settings (p. 27) · Configure settings for your
specific conditions (p. 41) · System-related settings
(p. 59)
When changing settings
Change basic settings such as measurement speed
Change to detailed settings (measurement conditions and system-related)
Set decision criteria (p. 34)
Calculation, Printing, Communication, and External Control Settings
Statistical calculations (p. 74)
Computer communications (p. 97)
When Finished
Data transmission (p. 77)
Printing (p. 79)
External control (p. 85)
Instrument interface settings must be configured before printing or using communications or remote control.
Turn the power off (p. 25)
3
Introduction
Introduction
Thank you for purchasing the HIOKI Model RM3542/ RM3542-01 Resistance
HiTester. To obtain maximum performance from the instrument, please read this
manual first, and keep it handy for future reference.
Model RM3542-01 is the same as the RM3542, but with GP-IB included.
Trademarks
· Windows and Visual Basic are registered trademark of Microsoft Corporation
in the United States and/or other countries.
· TEFLON is a registered trademark or a trademark of The Chemours Company FC,
LLC.
Verifying Package Contents
Inspection
When you receive the instrument, inspect it carefully to ensure that no damage
occurred during shipping. In particular, check the accessories, panel
switches, and connectors. If damage is evident, or if it fails to operate
according to the specifications, contact your dealer or Hioki representative.
Content confirmation
Confirm that these contents are provided.
Model RM3542 or
Instruction Manual (This document)………… 1
RM3542-01 (with GP-IB included) ……………1
Operation Guide…………………………………… 1
Power Cord (2-line + ground)(p. 22) EXT. I/O Male Connector (p. 96)
4
Verifying Package Contents
Options
Contact your dealer or Hioki representative for details.
Measurement Probes and Fixtures (connect to measurement jacks)
Model 9140 4-terminal Probe
Alligator-clip-type measurement probes. These general-purpose dual-electrode
clips fit a wide range of conductor thicknesses.
Interface Cables
Model 9637 RS-232C Cable (9-pin to 9-pin/ crossover cable)
Model 9638 RS-232C Cable (9-pin to 25-pin/ crossover cable)
Model 9151-02 GP-IB Connector Cable (2m)
Model 9262 Test Fixture
This fixture is for measuring lead components. (less than 10 m residual
resistance after zero adjustment)
Model 9263 SMD Test Fixture
This fixture is for measuring chip components. (less than 10 m residual
resistance after zero adjustment)
Safety Information
5
Safety Information
This instrument is designed to comply with IEC 61010 Safety Standards, and has
been thoroughly tested for safety prior to shipment. However, mishandling
during use could result in injury or death, as well as damage to the
instrument. Using the instrument in a way not described in this manual may
negate the provided safety features. Be certain that you understand the
instructions and precautions in the manual before use. We disclaim any
responsibility for accidents or injuries not resulting directly from
instrument defects.
This manual contains information and warnings essential for safe operation of
the instrument and for maintaining it in safe operating condition. Before
using it, be sure to carefully read the following safety precautions.
Safety Symbols
In the manual, the symbol indicates particularly important information that
the user should read before using the instrument. The symbol printed on the
instrument indicates that the user should refer to a corresponding topic in
the manual (marked with the symbol) before using the relevant function.
Indicates AC (Alternating Current).
The following symbols in this manual indicate the relative importance of
cautions and warnings. Indicates that incorrect operation presents a
significant hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury to the
user or damage to the instrument.
Indicates advisory items related to performance or correct operation of the
instrument.
Symbols for Various Standards
This symbol indicates that the product conforms to regulations set out by the
EU Directive.
WEEE marking: This symbol indicates that the electrical and electronic
appliance is put on the EU market after August 13, 2005, and producers of the
Member States are required to display it on the appliance under Article 11.2
of Directive 2002/96/EC (WEEE).
6
Safety Information
Other Symbols
Indicates the prohibited action.
(p. )
Indicates the location of reference information.
Indicates that descriptive information is provided below.
[ ]
Square brackets indicate instrument display labels (such as setting item names).
SET
Bold characters within the text indicate operating key labels.
(Bold characters)
Unless otherwise specified, “Windows” represents Windows 95, 98, Me, Widows NT4.0, Windows 2000, Windows XP, or Windows Vista.
Click: Press and quickly release the left button of the mouse. Double click: Quickly click the left button of the mouse twice.
Accuracy
We define measurement tolerances in terms of f.s. (full scale), rdg. (reading)
and dgt. (digit) values, with the following meanings.
f.s.
(maximum display value)
The maximum displayable value. This is usually the name of the currently selected range.
rdg.
(reading or displayed value)
The value currently being measured and indicated on the measuring instrument.
dgt.
(resolution)
The smallest displayable unit on a digital measuring instrument, i.e., the input value that causes the
digital display to show a “1” as the least-significant digit.
7
Operating Precautions
Operating Precautions
Follow these precautions to ensure safe operation and to obtain the full
benefits of the various functions.
Preliminary Checks
· Before using the instrument for the first time, verify that it operates
normally to ensure that no damage occurred during storage or shipping. If you
find any damage, contact your dealer or Hioki representative.
· Before using the instrument make sure that the insulation on the power cord
is undamaged and that no bare conductors are improperly exposed. Using the
instrument in such conditions could cause an electric shock, so contact your
dealer or Hioki representative for repair.
· Before using the instrument, make sure that the insulation on the
measurement cables is undamaged and that no bare metal is improperly exposed.
If there is any damage, measured values may be unstable and measurement errors
may occur.
Instrument Installation
Storage temperature and humidity: -10°C to 50°C at 80% RH or less (non- condensating) Operating temperature and humidity: 0 to 40°C at 80% RH or less (non-condensating)
Avoid the following locations that could cause an accident or damage to the instrument.
Exposed to direct sunlight Exposed to high temperature
In the presence of corrosive or explosive gases
Exposed to liquids Exposed to high humidity or condensation
Exposed to strong electromagnetic fields Near electromagnetic radiators
Exposed to high levels of particulate dust
Near induction heating systems (e.g., high-frequency induction heating systems and IH cooking utensils)
Subject to vibration
Installation Precautions
· The instrument should be operated only with the bottom downwards. · Do not place the instrument on an unstable or slanted surface.
50 mm or more 50 mm or more
10 mm or more
The instrument can be used with the stand (p. 16). It can also be rack- mounted. (p. A8).
Rear
Unplugging the power cord kills power to the instrument. Be sure to provide enough unobstructed space to unplug the power cord immediately in an emergency.
8
Operating Precautions
Handling the Instrument
· Do not allow the instrument to get wet, and do not take measurements with
wet hands. This may cause an electric shock.
· Do not attempt to modify, disassemble or repair the instrument; as fire,
electric shock and injury could result.
To avoid damage to the instrument, protect it from physical shock when
transporting and handling. Be especially careful to avoid physical shock from
dropping.
This instrument may cause interference if used in residential areas. Such use
must be avoided unless the user takes special measures to reduce
electromagnetic emissions to prevent interference to the reception of radio
and television broadcasts.
Handling the Cords and Probes
· Avoid stepping on or pinching cables, which could damage the cable
insulation. · To avoid breaking the cables, do not bend or pull them. · To
avoid damaging the power cord, grasp the plug, not the cord, when unplugging
it from
the power outlet. · When disconnecting the BNC connector, be sure to release
the lock before pulling off the
connector. Forcibly pulling the connector without releasing the lock, or
pulling on the cable, can damage the connector (p. 23).
· Use only the specified connection cables. Using a non-specified cable may
result in incorrect measurements due to poor connection or other reasons.
· Before using a test fixture, read the instructions provided with it.
Before Turning Power On
Before turning the instrument on, make sure the supply voltage matches that
indicated on its power connector. Connection to an improper supply voltage may
damage the instrument and present an electrical hazard.
Before Connecting EXT. I/O
9
Operating Precautions
To avoid electric shock or damage to the equipment, always observe the
following precautions when connecting to the EXT. I/O connector. · Always turn
off the power to the instrument and to any devices to be connected
before making connections. · Be careful to avoid exceeding the ratings of
external terminals (p. 91). · During operation, a wire becoming dislocated and
contacting another conductive
object can be serious hazard. Make sure that connections are secure and use
screws to secure the external connectors. · Properly insulate any devices and
mechanisms to be connected to the EXT. I/O connector. · The ISO_5V pin of the
EXT. I/O connector is a 5V power output. Do not apply external power to this
pin. · The ISO_12V pin of the EXT. I/O connector is a 12V power output. Do not
apply externalpower to this pin.
Before Connecting to the RS-232C Connector or SET MONITOR Connector
· Use a common ground for both the instrument and connected device. Using
different ground circuits will result in a potential difference between the
instrument’s ground and the connected device. If the communications cable is
connected while such a potential difference exists, it may result in equipment
malfunction or failure.
· Before connecting or disconnecting any the communications cable, always turn
off the instrument and the connected device. Failure to do so could result in
equipment malfunction or damage.
· After connecting the communications cable, tighten the screws on the
connector securely. Failure to secure the connector could result in equipment
malfunction or damage.
10
Operating Precautions
Before Measuring
· Do not apply any voltage to the measurement jacks. Doing so could damage the
instrument.
· Never attempt to measure at a point where voltage is present. In particular,
do not measure a transformer or motor immediately after a temperature increase
test or withstand-voltage test, as the instrument could be damaged by induced
voltage or residual charge.
· Battery internal resistance cannot be measured with this instrument. It will
sustain damage. To measure battery internal resistance, we recommend the HIOKI
3554, 3555, BT3562, BT3563, and 3561 Battery HiTesters.
· To obtain the guaranteed measurement accuracy, allow at least 30 minutes
warm-up. · The instrument internally retains all settings (but not measured
values), such as measure-
ment range and comparator settings. However, measurement settings made through
the RS-232C or GP-IB interface are not memorized. · In the 100 and higher
ranges (LOW POWER set to OFF), thermal emf can cause measurement errors. · The
DC resistance of a power transformer cannot be measured. When measuring
objects with a large L, such as choke coils and other inductors, measured
values may be unstable. In such cases, contact your dealer or Hioki
representative. · Carefully insulate all HCUR, HPOT, LPOT, and LCUR wiring.
Proper 4-terminal measurements cannot be performed and an error will occur if
core and shield wires touch.
Overview
11
1.1 Product Overview and Features
Chapter 1 1
1.1 Product Overview and Features
The Hioki RM3542 Resistance HiTester employs the 4-terminal method to quickly and accurately measure the DC resistance of components such as resistors and ferrite beads. It includes advanced contact-check, comparator, and data output functions. The intuitive user interface and high noise immunity are ideal for use with taping machines and separators.
Resistance Measurement
The factory defaults (initial settings) are optimized for chip-component
resistance measurements. The RM3542 can also measure devices that are
otherwise difficult to measure with high current, such as ferrite-bead and
small multilayer inductors (low-power resistance measurement, p. 28).
Optional Hioki probes and fixtures are available to connect to the measurement
jacks (BNC jacks, p. 4). Alternatively, commercially available cables such as
1.5D-2V coax can be used (p. 24).
Judge Measured Values Measured values are compared with a pre-specified
reference value or thresholds, and the result is output externally and
indicated by the COMP indicators (comparator function, p. 34).
Upper limit
Lower limit
Save and Output Measured Values
Measured values can be stored in internal memory (p. 69) Statistical
calculations can be performed on the stored data, which can be transferred to
a computer in batch form (however, stored data cannot be confirmed
internally).
Send Measurement Data and Calculation Results to a Printer
Use a commercially available printer with a serial interface to print measured
values and calculation results (p. 79).
Connect a PLC or I/O Board
To control from a PLC, connect to the EXT. I/O connector. In addition to
comparator results, various measurement anomaly signals can be output (p. 85).
Compare Two Instrument’s Setting Conditions
When measuring with two interconnected instruments, settings are compared, and
an alarm is output and measurement is inhibited if the settings differ
(Settings Monitor function, p. 53).
Interface Communications
Connect the instrument to a controller via the RS-232C or GP-IB interface to
control measurement data acquisition (p. 97).
12
1.1 Product Overview and Features
Features
Ultra Fast and Accurate Measurements Increase Productivity The factory default
settings are optimized for chipcomponent resistance measurements. Enhanced
contact-to-measurement and contact-check-to-decision times are only 1 ms. The
offset-voltage compensation (OVC) function minimizes the effects of thermal
emf when using low-power resistance measurement and the 100 m to 10 ranges (p.
57). Measurement results are judged pass/fail with 10 ppm resolution, ideal
for high-speed Class B resistor testing.
High-Speed Data Output and Ample Memory
The Data Output function transfers measured data at 5 ms/sample, even via RS-
232C. Up to 30,000 measurements can be stored, and all data can be exported at
the end of measuring each reel. This function is ideal for system setup,
debugging and process management.
Multiple Interfaces
EXT. I/O is isolated from the measurement and control circuits to provide
noise immunity (p. 85). All data can be acquired in real time using the built-
in 38.4-kbps high-speed RS-232C interface. Connect the commercially available
printer with a serial interface to print measured values and statistical
calculation results (p. 79). The GP-IB interface is available for Model
RM3542-01 (specified when ordering, p. 97).
Low-Power Function (p. 28) For ranges from 1000 m to 1000 , low-power resistance measurement is provided to minimize measurement current. Safely measure devices that are otherwise difficult to measure with high current, such as ferrite-bead and multilayer inductors.
Clearly Visible Display and Intuitive Operation
High-contrast LCD provides clear visibility, helping avoid setting mistakes.
The optimum range is selected automatically when comparator thresholds are
entered.
Auto Memory Function Convenient for Sampling Tests (p. 71)
The auto memory function is convenient for sampling tests after screen-
printing. Measured values are automatically acquired as soon as they
stabilize, and statistical calculations proceed until the specified quantity
is obtained, upon which an alert notification (alarm) occurs. Selecting
[PRINT] (screen display) prints measured values and statistical calculation
results (p. 82).
Fixtures for Component Measurements (p. 4)
The BNC-type measurement jacks exhibit good noise immunity. Ready availability
and easy assembly ensure smooth system setup. Various test fixtures are
available for Hioki LCR HiTesters.
13
1.1 Product Overview and Features
Features
1
Reliable Contact Checking (p. 46)
Contact checking (that was previously performed before and after measuring) is
now performed during measurement, so probe bounce and contact resistance
fluctuations can be detected. Contact checking time can be shortened,
improving tact times.
Contact Condition
Probe Bounce
Model RM3542
Contact Check
Measuring
Previously
Minimize Human Error and Risk the Settings Monitor Function (p. 53)
If the settings of two instruments are different, triggering is inhibited and
an alarm notification is generated to avoid setting mistakes due to human
error.
Reject Faulty Data Voltage Level Monitor Function (p. 49)
When the contact resistance of the HCUR and LCUR leads fluctuates, the
measurement current changes momentarily. Such momentary changes are not
detectable by typical contact checking. The Voltage Level Monitor function
detects a contact error if the detection voltage changes significantly, which
can increase the reliability of measured values.
Poor Contact
Good Contact
Good Contact
Contact Condition
Detection Voltage
Error
Excessive detection voltage fluctuation error
ERROR
Contact Improver Function (p. 47)
The Contact Improver function improves bad contacts between probes and test
samples. Contacts errors are reduced by penetrating oxidation and impurities
between probes and samples. Reducing contact errors can increase productivity
and quality. The intensity of the contact improver function can be adjusted
according to probe type.
Contact Improvement
Contact Condition
Contact Improver
Function
ON
Contact Check
Measurement Status
ON
Checking Measuring
POT CUR
Reliable Four-Terminal Measurement Probe Short-Circuit Detection Function
(p. 51) Four-terminal measurements are inhibited when a conductive foreign
object is present between the POT and CUR probe tips. Short-circuit probe
anomalies are detected by checking the resistance between these tips when not
measuring.
DUT electrode
DUT
Foreign object
Measurement Circuit Strongly Immune to Contact Resistance Fluctuations The
effects of contact resistance fluctuations are reduced even when scattering
occurs near the end of probe life. Such effects are minimized by the fast
response of the measurement circuit.
Strong Electrical Noise Immunity The specified measurement accuracy is
achieved even with ±1.5 kV mixed pulse noise. The floating measurement section
design is highly impervious to electrical noise, minimizing the effect on
measured values even when turning large-induction motors on and off. The free-
range power supply input (90 to 264 V) is practically unaffected by voltage
fluctuations, so stable measurements are possible even in under poor power
conditions.
14
1.1 Product Overview and Features
Block Diagram
I
F
G
C
H
D E
A
B
E D
· Constant current (determined by the measurement range) is applied between
the HCUR and LCUR terminals while voltage is measured between the HPOT and
LPOT terminals. The resistance value is obtained by dividing the measured
voltage (B) by the constant current flow (A).
· The effects of large offset voltage such as from thermal emf are reduced by
current flowing in the positive and negative directions (A).
· The constant current source (A) and voltmeter (B) circuit designs are
largely unaffected by contact resistance.
· Faulty measurement values caused by unstable or chattering contact
conditions can be eliminated by monitoring (C) the detection voltage (B)
waveform (Voltage Level Monitor function).
· Stable measurements are ensured by providing sufficient integration time
(the default setting is 0.3 ms). (The integration time can be reduced to 0.1
ms to support even higher speed (B).)
· Before measuring, the Contact Improver circuit (D) optimizes contact when
the probes touch the DUT.
· By also performing contact checking (E), short circuits between CUR and POT
terminals caused by a clogged probe tip can be detected (probe short-circuit
detection function).
· When measurement starts, the contact check circuit (E) and constant current
monitor (F) are activated to monitor for fault conditions while measuring. The
dual-CPU (C and G) design provides ultra-high-speed measurements and fast
system response.
· Immunity from electrical noise is provided by isolation between the
Measurement and Control blocks (H).
· The auto-ranging 100-to-240 V switching power supply (I) can provide stable
measurements even in poor power quality environments.
15
1.2 Names and Functions of Parts
1.2 Names and Functions of Parts
1
Front Panel
Power On/Off
POWER Button Turns the instrument on and off (p. 25). · Unlit: power off
(when no power supplied) · Red light: power off
(while power is supplied) · Green light: power on
Viewing Measured Values and Settings
Display Screen Monochrome graphical LCD There are three general screen types:
Measurement, Basic Settings and Detailed Settings.
Screen Configurations (p. 17)
Setting
JudginFg 1mteoasFu4reÉdLvÅa[lues (p. 34)
(Compar(aFtÉorLfÅun[c: tëioçnè)Ã) Set a reâfeÊreñnþcâeEëv§aÇlu…e
ïaén¶dÇrSaÇngÍÇeÈfoçÄr judging measureñm/ÇesnëtsIë,swÇhµiÇch<Çc²aÅnB be
confirmed by the COMP indicators.
Select this to judge measured values relative to a reference value and
tolerance (%). The REF% setting display appears.
Press to judge measured values relative to upper/lower thresholds. The ABS setting display appears.
Audible Alarm (beeper)
Viewing Comparator
Results
COMP indicator LEDs Indicate the decision result of the measured value (p.
34). Hi Measured value is
above upper limit IN Pass (meets crite-
ria) Lo Measured value is
below lower limit
Connecting Probes
Measurement Terminals Connect measurement cables or a fixture (p. 23). · HCUR
jack: Current source terminal · HPOT jack: Detected voltage high
terminal · LPOT jack: Detected voltage low ter-
minal · LCUR jack: Measurement current
detected terminal · GUARD jack: Shield (measurement
ground) terminal
Entering numerical values
Enter a numerical value. (we call these the “tenkeys”)
Switches the sign of a numerical value.
Selects the unit of measure.
Sets the tolerance values.
Deletes the value in the selected field.
Accepts the displayed comparator threshold values.
Aborts comparator setting and returns to the previous display.
Selecting the setting contents
(“the F keys”) F1 to F4 keys Selects the corresponding item on the right side
of the display.
Cursor keys Move among the displayed setting items. The cursor location is
indicated by reverse characters.
16
1.2 Names and Functions of Parts
Rear Panel Connecting the Power Cord
Connect the supplied power cord (p. 22).
Manufacturer’s Serial Number
Shows the serial number. Do not remove this label, as it is required for
product support.
External Control
EXT. I/O Connector Connect to a PLC or I/O board to control measurement start,
and to acquire comparator results (p. 85).
Compare Two Instruments’ Settings
SET MONITOR Connector
Connect another RM3542 here to compare instrument settings (p. 53).
RS-232C Communications Printer Output
RS-232C Connector The RS-232C interface can be used to connect to a PLC or
computer (p. 97). It is also used by the commercially available printer with a
serial interface (p. 79).
GP-IB Communications
GP-IB Connector (RM3542-01 only) The GP-IB interface can be used to connect to
a computer (p. 97).
Bottom Panel
This instrument can be rack mounted. See: “Appendix 4 Rack Mounting” (p. A8)
Parts removed from this instrument should be stored in a safe place to enable
future reuse.
When using the stand Extend the stand until it clicks into place. Make sure to
extend both legs of the stand.
Collapsing the stand Fold in the stand until it clicks into place.
Stand
Do not apply heavy downward pressure with the stand extended. The stand could be damaged.
17
1.3 Screen Organization
1.3 Screen Organization
1
The instrument has three general display screen types: Measurement, Basic
Settings and Detailed Settings. Refer to “11.3 Error Displays and Remedies”
(p. 181) for error displays.
The screen examples in this guide appear reversed (black on white) for best
visibility on the printed page. However, the instrument screens can actually
be displayed only as white characters on black background.
Measurement Screen (p. 18)
Basic Settings Screen (p. 18)
Return to Previous Screen
Detailed Settings Screens (p. 19) Measurement Settings Screens [MEAS SETTINGS] Data Settings Screens [DATA SETTINGS] System Screens [SYSTEM]
Return to Previous Screen
Indicates a Continued Screen
18
1.3 Screen Organization
Measurement Screen
This screen normally appears while measuring. View currently measured values
and measurement conditions. Some parts of the display depend on the comparator
mode and other settings.
To display the Basic Settings screen
Settings Menu (corresponding to F keys)
Displayed contents depend on the current function set-
tings.
Measured Value Criteria Setting Values (p. 34)
Parentheses ( ) indicate the corresponding F-key
Displayed contents depend on the MENU (F1)
Displays the Basic Settings screen
selected comparator mode. (Ex.: REF% mode)
Measurement Conditions
PRINT (F2)
Print (p. 81) Appears only when the interface is set for the printer
Shows current setting contents. Displayed contents de-
pend on the current settings.
STAT (F3)
Statistical calculation results (p. 74) Appears only when statistical calcu-
INT/ EXT
Trigger source type (p. 30)
lation is enabled
Measurement range (p. 31)
NUMBER (F4) Set Auto-Memory number (quantity)
FAST/ MED/ SLOW Measurement speed (p. 29)
0ADJ/
Appears only when zero-adjust is
OFF (not shown) enabled (p. 32)
(p. 71)
Displays the number of stored data items and the number of passed and failed
products at the lower left.
OVC/ OFF (not shown)
LP/ OFF (not shown)
(OVC: Offset Voltage Compensation) Appears only when OVC is enabled (p. 57)
Appears only when the Low-Power Resistance function is enabled (p. 28)
UNDO (F3) ALLCLR (F4)
Deletes the previously stored measurement and calculation result (only one can
be deleted) (p. 76) Appears only when calculation results are displayed
Clears all memory and calculation
NUM RMT
Appears only when tenkey input is enabled
Remote control (p. 104)
(p. 76) Appears only when calculation results are displayed
M.LOCK
Disables all operations except LOCAL (F1)
Enables local controll (p. 104)
comparator settings (p. 59)
UNLOCK (F1) Cancels the key-lock state (hold 1s)
F.LOCK
Disables all operations including
(p. 60)
comparator settings (p. 59)
Basic Settings Screen
Make basic measurement condition settings on this
screen. Measurement speed and range can be changed
Return to previous screen while viewing measured values. (Trigger source: INT)
Select measurement condition settings Move with cursor keys.
TRG
Measurement condition setting selections
RANGE SPEED 0ADJ COMP LOCK
Change trigger source (measurement start control method) (p. 30) When EXT
triggering is enabled, [MANU] appears beside the F4 key (press to trigger
measurement manually).
Change range (p. 31)
Change measurement speed (p. 29)
Turn zero-adjust on/off (p. 32)
Turn comparator on/off (p. 34)
Enable/disable key lock (p. 59)
MISC
To display the Detailed Settings screen
To display the [MEAS SETTINGS] screen To display the [DATA SETTINGS] screen To display the [SYSTEM] screen
Detailed Settings Screens Measurement Settings Screen [MEAS SETTINGS] Data Settings Screen (Save, Analysis, and Output) [DATA SETTINGS] System Screen [SYSTEM]
19
1.3 Screen Organization
1
Shows detailed settings for measurements. Adjust the measurement speed,
stability and measurement fault detection functions.
DELAY1
Adjust the delay from probing to trigger input (p. 42)
DELAY2
Adjust measurement object electrical response (p. 42)
INT (FAST/ MED/ Make fine adjustment to integration time
SLOW)
(p. 44)
CONT CHECK
Contact check threshold setting (p. 46)
CONT IMP
Contact Improver function setting (p. 47)
VOLT MONITOR Voltage level monitor function setting (p. 49)
CURRENT MODE Current mode setting (p. 50)
These are settings for memory and statistical calculation functions.
AUTO MEMORY Turn Auto-Memory on/off (p. 71)
STATISTICS
Statistical calculation on/off (p. 74)
DATA OUT
Automatically output measured values (communications) (p. 77)
Set instrument system-related settings on this screen.
SET MONITOR
Turn two-instrument setting comparison on/off (p. 53)
PROBE CHECK
Turn probe short-circuit detection on/off (p. 51)
RETRY
Retry function setting (p. 56)
TRIG EDGE
(External I/O) Set trigger rising/falling edge (p. 94)
EOM
(External I/O) Set EOM (end-of-measurement) signal (p. 93)
INTERFACE
Communications interface settings (p. 101)
LOW POWER
Low-power resistance component measurement on/off (p. 28)
JUDGE BEEP
Comparator beeper settings (p. 62)
KEY BEEP
Key beeper on/off (p. 61)
CLOCK (Y-M-D) Set internal clock (p. 64)
LINE FREQ
Set power source frequency (p. 63)
CONTRAST
Adjust screen contrast (p. 65)
BACKLIGHT
Adjust screen backlight (p. 66)
RESET
Initialize (p. 67)
ADJUST
Instrument Adjustment (p. A13)
20
1.3 Screen Organization
21
Measurement Preparations
Chapter 2 2
Be sure to read the “Operating Precautions” (p.7) before installing and connecting this instrument. Refer to “Appendix 4 Rack Mounting” (p. A8) for rack mounting.
5
2
Front Panel
3
1 Install this instrument (p. 7)
2
Connect the power cord (p. 22)
3 Connect measurement cables (p. 23)
Rear Pane
4
4 Connect the external interface (as needed)
· Using the printer (p. 79) · Using the RS-232C or GP-IB interface
(p. 97) · Connecting to a PLC or I/O board
(p. 85) · Automatically comparing the settings
of two instruments (Settings Monitor function) (p. 53)
5 Turn the power on (p. 25)
6 Make instrument settings (p. 27)
Connect to the test sample
When finished measuring, turn the power off (p. 25).
22
2.1 Connecting the Power Cord
2.1 Connecting the Power Cord
· Before turning the instrument on, make sure the supply voltage matches that
indicated on its power connector. Connection to an improper supply voltage may
damage the instrument and present an electrical hazard.
· To avoid electrical accidents and to maintain the safety specifications of
this instrument, connect the power cord only to a 3-contact (two-conductor +
ground) outlet.
· Before using the instrument, make sure that the insulation on the power cord
is undamaged and that no bare conductors are improperly exposed. Any damage
could cause electric shock, so contact your dealer or Hioki representative.
To avoid damaging the power cord, grasp the plug, not the cord, when
unplugging it from the power outlet.
Power inlet Rear Panel
1 Confirm that the mains supply voltage matches the
instrument, and connect the power cord to the power inlet on the instrument.
2 Plug the power cord into the mains outlet.
The POWER button on the front panel lights red. In event of a power outage,
operation resumes with the same settings when power is restored (breaker
reset, etc.).
23
2.2 Connecting Measurement Cables and Test Fixtures
2.2 Connecting Measurement Cables and Test Fixtures
2 Connect your measurement cables, optional Hioki probes or test fixture to
the measurement jacks.
Refer to “Options” (p.4) for details. See the instructions provided with the
fixture for operating details.
· Do not apply a voltage to the measurement terminals. Doing so may damage the
unit. · When disconnecting the BNC connector, be sure to release the lock
before pulling off the
connector. Forcibly pulling the connector without releasing the lock, or
pulling on the cable, can damage the connector.
· We recommend using optional Hioki fixtures.
· Use the GUARD jack only for Faraday shield, and avoid more than 10 mA
current flow. This jack is not for guarding network resistance measurements.
Connection Methods
Example of defeated guard measurement
Connecting measurement cables
Black plugs
BNC Jack Guide Pins (on the instrument)
Red plugs
1
BNC plug slots
2
Lock
Connecting a fixture
Connect the red plugs to the
HCUR and HPOT jacks, and the black plugs to the LCUR and LPOT jacks.
Align the slots in the BNC plug with the guide pins on the jack on the instrument, then push and twist the plug clockwise until it locks. Disconnecting BNC connectors Push the BNC plug, twist it counterclockwise, and pull it out.
Connect directly to the measurement jacks with the label side up, and affix with the levers on the left and right.
Making your own probes and extenders (p. 24)
24
2.2 Connecting Measurement Cables and Test Fixtures
Making Your Own Measurement Cables
Recommended Measurement Cable Specifications
Conductor resistance
500 m/m or less
Capacitance
150 pF/m or less
Cable dielectric material
Polyethylene (PE), TEFLON (TFE), polyethylene foam (PEF) Insulation resistance at least 10 G
Connector insulating material TEFLON (TFE), polybutylene terephtalate (PBT) Insulation resistance at least 10 G
Length
2 m or less
Example: JIS standard 3C-2V, 1.5D-2V MIL standard RG-58A/U
Wiring Diagram
Before Wiring
· Twist together the HPOT and LPOT wires, and the HCUR and LCUR wires. If not
twisted together, measured values may be unstable and errors occur when
measuring with low-power resistance, or low resistance values.
· Refer to the block diagram (p. 14) for internal circuit details. · Probes
and measurement objects should be shielded at BNC or GUARD jack potential.
Extending Measurement Cables
Observe the following when extending measurement cables:
· Measurement cable length should not exceed 2 m (with conductor resistance
500 m/m or less). Long cables are more susceptible to noise, and measured
values may be unstable.
· Extensions should maintain the four-terminal structure. If converted to a
two-terminal circuit in the wiring, correct measurement may not be possible
due to the effects of wiring and contact resistance.
· Cables and measurement objects should be shielded. · After extending
measurement cables, confirm operation and accuracy (“Measurement Specifica-
tions” (p.176)). · If cutting the ends off of optional measurement cables,
make sure that the shield does not touch
the center conductor of the HCUR, HPOT, LPOT and LCUR cables. Correct
measurement is not possible with a shorted cable.
25
2.3 Turning the Power On and Off
2.3 Turning the Power On and Off
Turning Power On
2
Press the POWER button (it lights green).
After Power-On A self-test (instrument diagnostic routine) is performed.
During the self-test, the following information is displayed while the
hardware is verified.
Self-test The following information is displayed during self-testing: ·
Manufacturer and model name · Firmware versions (main, and measurement) ·
Communication setting · Line frequency setting
No Errors
Error
Normal display (measurement screen)
Indicates an error (p. 181).
When the power is turned on, the same setting as when the power was last
turned off appears (backup function). When powered up for the first time, the
default settings appear.
See: “Default Settings” (p.68)
Before Starting Measurement To obtain precise measurements, provide about 30
minutes warm-up after turning power on. Measurement settings are recalled from
when the power was previously turned off (settings backup). However,
measurement settings made through the RS-232C or GP-IB interface are not
retained, although they can be stored using the :SYSTem:BACKup command (p.
135).
Turning Power Off
Press the POWER button (it lights red when the instrument is off). Disconnect
the power cord from the outlet to extinguish the POWER button light. When
power is turned on again, operation resumes with the same settings as when
last turned off.
If a power outage (e.g., breaker trip) occurs when the instrument is on, it
will automatically turn on again when power is restored (without pressing the
POWER button).
26
2.3 Turning the Power On and Off
27
3.1 Pre-Operation Inspection
Measurement Settings
(Basic Measurements)
Chapter 3
3
See “Measurement Flow” (p. 2) for an outline of the measurement process from
preparation to endof-measurement.
3.1 Pre-Operation Inspection
Before using the instrument for the first time, verify that it operates normally to ensure that no damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representative.
1 Peripheral Device Inspection
Is the power cord insulation torn, or is Metal Exposed any metal exposed?
No Metal Exposed
Do not use the instrument if damage is found, as electric shock or short- circuit accidents could result. Contact your dealer or Hioki representative.
Metal Exposed Is the insulation on a measurement cable torn, or is any metal
exposed?
No Metal Exposed
If there is any damage, measured values may be unstable and measurement errors may occur. Replace the cable with an undamaged one.
2 Instrument Inspection
Yes Is damage to the instrument evident?
If damage is evident, request repairs.
No
When turning power on No
Does the self-test screen appear (model no., version no.)? (p. 25)
Yes
Does the Measurement screen appear after self-test?
An error indication occurs (ERR)
Yes Inspection complete
The power cord may be damaged, or the instrument may be damaged internally.
Request repairs.
The instrument may be damaged internally. Request repairs. See: “11.1
Troubleshooting” (p. 179)
“11.3 Error Displays and Remedies” (p. 181)
Please read the “Operating Precautions” (p. 7) before use.
28
3.2 Measurement Object Types
3.2 Measurement Object Types
The instrument provides two measurement methods: resistance measurement, and
low-power resistance measurement. Select the appropriate measurement method
for the type of component to be measured. For general-purpose resistor
measurements, use the factory defaults. The power applied to the DUT =
Resistance Value × (Measurement Current)2.
See: “(6) DUT Becomes Warm” (p. A6)
(Example)If the resistance to be measured is 100
(Measurement Current)
10 mA
100 × 0.012 = 10 mW
1 mA
100 × 0.0012 = 100 µW
(Measurement Method) Normal Resistance Measurement, 100 Range Low-Power Resistance Measurement, 100 Range
General-purpose resistors
Normal Resistance Measurement
Measurement range: 0.0000 m (100 m range) to 120.0000 M (10 ranges)
Hard-to-measure components such as ferrite bead or layered inductors, or other elements sensitive to measurement current
Low-Power Resistance Measurement
Measurement range: 0.000 m (1000m range) to 1200.000 (4 ranges)
LP LP appears at the top of the screen.
The 1000 and higher ranges (with LOW POWER: OFF) are not usable for inductor
measurements.
1 Open the Basic Settings screen.
The Basic Settings screen appears.
2 Open the System screen.
1 Selection
3 Select the low-power mode, as needed. 1 Selection
4 Return to the Measurement screen. The confirmation screen appears.
2 The System screen appears. [SYSTEM] 2
Normal resistance measurement (default) Low-power resistance measurement
Return to the setting screen. Save setting and return to previous screen.
Discard setting and return to previous screen.
29
3.3 Setting the Measurement Speed
3.3 Setting the Measurement Speed
The measurement speed can be set to FAST, MED (medium), or SLOW. A slower measurement speed provides greater measurement precision, and a faster measurement speed results in greater susceptibility to environmental noise. Ensure that measurement cables and the sample are sufficiently shielded.
1 Open the Basic Settings screen.
3
The Basic Settings screen appears.
2 Select the measurement speed.
1 Selection
3 Return to the Measurement screen.
2
See table below
(default)
Press the up/down cursor keys to change the setting.
Relationship Between Measurement Range and Speed
(factory defaults)
Measurement Range
LOW POWER: OFF FAST MED SLOW
LOW POWER: ON FAST MED SLOW
100m
3.8 ms 13 ms
43 ms 36 ms
–
–
–
1000m
2.0 ms
6.4 ms
41 ms 35 ms
2.5 ms 12 ms
42 ms 35 ms
10 100
1.6 ms 0.9 ms
6.0 ms 3.6 ms
41 ms 34 ms
21 ms 17 ms
2.5 ms 1.7 ms
12 ms 6.1 ms
42 ms 35 ms
41 ms 34 ms
1000
0.9 ms
3.6 ms
21 ms 17 ms
7.2 ms 12 ms
47 ms 40 ms
10k
1.0 ms
3.6 ms
21 ms 17 ms
–
–
–
100k
1.3 ms
3.8 ms
21 ms 18 ms
–
–
–
1000k
2.5 ms
6.0 ms
21 ms 18 ms
–
–
–
10M
5.3 ms
23 ms 20 ms
23 ms 20 ms
–
–
–
100M
26 ms 22 ms
46 ms 39 ms
86 ms 72 ms
–
–
–
Integration time can be optionally set for each range (p. 44).
Upper value: for 50-Hz power line frequency
Lower value: for 60-Hz power line frequency
Tolerance: ±10%±0.2 ms
30
3.4 Setting Measurement Start Conditions (Trigger Source)
3.4 Setting Measurement Start Conditions (Trigger Source)
Measurements can be started in two ways.
To measure automatically
Measure with internal (INT) triggering
Trigger signals are automatically generated internally for continuous
measurement.
To measure at specific times To retain measured values
Measure with external (EXT) triggering
Measurements are triggered by an external signal. Manual measurement
triggering is also available. · Apply a trigger signal at the EXT. I/O
connector (p. 85)
· Send the TRG command by communications interface (p. 126)
· Press F4 [MANU] (only appears when EXT is selected)
· When internal triggering is enabled, the EXT. I/O TRIG signal and the “TRG”
command are ignored
(except for memory storage and statistical calculations). · To measure samples
such as inductors that require time to settle, adjust delay time DELAY2. Start
with
a long delay, and gradually shorten it while watching for the measured value
to settle. See: “4.2 Setting Pre-Measurement Delay” (p. 42)
· When external triggering is enabled, the Auto-Memory function is disabled by
force.
1 Open the Basic Settings screen.
The Basic Settings screen appears.
2 Select internal (INT) or external (EXT) triggering. 2
1 Selection
Internal trigger
External trigger (default)
Press F3 [EXT] to display the F4 [MANU] indicator. Press to trigger a measurement manually.
3 Return to the Measurement screen.
Press the up/down cursor keys to change the setting.
Continuous measurement (:INITIATE:CONTINUOUS ON) is the normal trigger state
when operating from
the front panel. Selecting the internal (INT) trigger source activates
continuous triggering (“free-run”). When external (EXT) triggering is
selected, each external trigger event initiates one measurement. Continuous
mea-
surement can be disabled by sending the :INITIATE:CONTINUOUS OFF command via
RS-232C or GP-
IB. When continuous measurement is disabled, trigger acceptance is controlled
only by the host (computer or PLC).
See: for trigger command: “Triggering” (p. 138), “9.8 Data Exporting Methods”
(p. 148)
31
3.5 Selecting the Measurement Range
3.5 Selecting the Measurement Range
The measurement range can be set as follows. When making comparator settings with the panel keys, the measurement range is selected automatically according to the settings (reference/tolerance or upper/lower thresholds, see the following table). When the comparator settings are made by remote control commands, the measurement range is unaffected.
Changing the Range Measurement error is larger when measuring values nearer the bottom of a measurement range.
3
The 1000 and higher ranges (with LOW POWER: OFF) are not usable for inductor measurements.
Auto-Ranging (when making comparator settings)
Low Power OFF (p. 28)
Low Power ON (p. 28)
Reference (REF%) and
Selected
Upper/Lower Threshold (ABS) Ranges Range
Reference (REF%) and Upper/Lower Threshold (ABS) Ranges
0 to 100.09 m 100.1 to 1000.9 m 1.001 to 10.009 10.01 to 100.09 100.1 to 1000.9 1.001 to 10.009 k 10.01 to 100.09 k 100.1 to 1000.9 k 1.001 to 10.009 M 10.01 to 120.00 M
100 m 1000 m 10 100 1000 10 k 100 k 1000 k 10 M 100 M
– 0 to 1000.9 m 1.001 to 10.009 10.01 to 100.09 100.1 to 1200.0
Selected Range
– 1000 m 10 100 1000
Manual Range Selection
1 Open the Basic Settings screen.
The Basic Settings screen appears.
2 Select the range.
1 Selection
2
Increments the
range
Press the up/down
Decrements the cursor keys to
range
change the setting.
3
Selectable ranges depend on the low-power resistance setting (p. 28). · When
low-power resistance measurement is disabled (OFF)
100m, 1000m, 10, 100, 1000, 10k, 100k, 1000k, 10M, 100M(default) · When low-
power resistance measurement is enabled (ON)
1000m, 10, 100, 1000
Return to the Measurement screen.
32
3.6 Zero Adjustment
3.6 Zero Adjustment
When four-terminal measurement (Kelvin connection) is impractical such as when
measuring very small samples, the additional inherent resistance of the two-
terminal wiring should be canceled out. The zero-adjustment function can
cancel out up to 10 additional resistance.
Before Zero Adjustment
· The guaranteed accuracy of the instrument applies to four-terminal
connections without zero adjustment. When using four-terminal connections, do
not execute zero adjustment. Executing zero adjustment with incorrect wiring
may amplify measurement error. However, zero adjustment may be needed even
with fourterminal measurements if they are affected by a large offset voltage,
such as due to thermal emf (LOW POWER OFF, in 100 to 100 M ranges).
· Execute zero adjustment when the ambient temperature has changed, or when a
probe is replaced.
Execute zero adjustment after the warm-up period following power on.
1 Open the Basic Settings screen.
The Basic Settings screen appears.
2 Select the internal (INT) trigger mode.
1 Selection
3 Short the probes together.
2
Internal trigger
Press the up/down cursor keys to change the setting.
4 Confirm that the measured value does not exceed 10 . If no measured value is displayed, increment the measurement range (p. 31).
5 Select whether to enable or disable zero adjustment. 2
1 Selection
Disable zero adjustment (cancel) Execute zero adjustment
After confirming that the measured value does not exceed 10 , execute zero
adjustment.
6 Return to the Measurement screen.
Zero Adjustment Faults
If zero adjustment fails, the following error message appears.
33
3.6 Zero Adjustment
Before attempting zero adjustment again, confirm the following: · With the 10 range selected, confirm that the displayed value does not exceed 10 .
3
· Confirm that the probe connections are correct.
34
3.7 Judging Measured Values (Comparator Function)
3.7 Judging Measured Values (Comparator Function)
Comparator results are available as external output (at the EXT. I/O
connector) when the comparator reference/tolerance or upper/lower threshold
values have been set.
See: “Chapter 8 External Control” (p. 85)
Comparator results are also indicated by the COMP Hi/IN/Lo panel lamps, and by
audible beeper (disabled by default).
See: “Setting the Comparator Decision (“JUDGE”) Beeper” (p. 62)
The comparator decision mode can be set as one of the following:
Measured value is above upper limit
Pass (meets criteria)
Measured value is below lower limit
Decide whether a measured value is within specified tolerance limits relative
to a specified reference value (p. 35)
Decide whether a measured value is between specified upper and lower threshold
values (absolute values) (p. 37)
Select the REF% (relative values) decision mode
example
12.000 k ….. reference value +0.080%…….. positive tolerance -0.080%………
negative tolerance
Positive
Hi
tolerance [%] Reference value [] Negative
IN IN
tolerance [%]
Lo
Select the ABS (absolute values) decision mode
example
100.00 m …. upper threshold 80.00 m …… lower threshold
Upper
Hi
threshold [] Lower
IN
threshold []
Lo
Before Using the Comparator Function
· When the measured value is out of the selected measurement range, comparator
decision indicators appear as follows. In the event of a measurement fault, no
decision is made. See:”3.8 Confirming Faulty Measurements” (p. 38)
Out-of-Range Display +OvrRng -OvrRng
Comparator Decision Indicator Hi Lo
· If power is turned off during comparator setting, changes to settings are
lost as they revert to their previous values. To accept the settings, press
the ENTER key.
· When setting comparator criteria, the appropriate range is selected
automatically. Refer to “Auto-Ranging (when making comparator settings)” (p.
31) for range settings.
35
3.7 Judging Measured Values (Comparator Function)
Enabling and Disabling the Comparator Function
The comparator function is enabled by default. When the function is disabled, comparator settings are ignored.
1 Open the Basic Settings screen.
The Basic Settings screen appears.
3
2 Enable or disable the comparator function.
1 Selection
3 Return to the Measurement screen.
2
Disable the function Enable the function
Press the up/down cursor keys to change the setting.
(When the function is disabled) Comparator decisions are indicated only when the function is enabled.
Decide According to Reference Value and Tolerance (REF% Mode)
Relative Value = (tolerance)
Measured Value – 1
Reference Value
X 100 [%]
Setting range: -9.999% to +9.999% (When 10% or less) -99.99% to +99.99% (When more than 10%)
Example: Set a reference value of 10.5 with ±4.5% decision tolerance.
To abort the setting process, press screen.
. Settings are abandoned and the display returns to the previous
1 Open the relative tolerance setting screen.
Reference value Positive tolerance (upper decision threshold) Negative tolerance (lower decision threshold)
36
3.7 Judging Measured Values (Comparator Function)
2 Set the reference value. Pressing an inoperative key during setting sounds a low-pitch beep (when the key beeper is enabled).
1
Selection
To Reset Numerical Values Deletes entered digits. This key is enabled only when entering numerical values.
(Example: 10.5)
2
1_
10_
10._
To change the value after selecting the
units, use the cursor
keys to select
the item to change, then enter the new val-
10.5_
10.50 ue with the tenkeys.
Press the units key to accept the setting and move the cursor to the positive tolerance.
3 Set the positive tolerance.
1
Selection
(Example: 4.5%)
2
+4_
+4._
+4.5_
+4.500%
To Reset Numerical Values Deletes entered digits. This key is enabled only when entering numerical values.
To change the value after selecting the
units, use the cursor
keys to select
the item to change, then enter the new val-
ue with the tenkeys.
To Set a Negative Value Press this key to change the sign, as needed.
Press the % key to accept the setting and move the cursor to the negative
tolerance value. The negative tolerance is initially set to the same amplitude
as the positive tolerance (change as needed).
4 Set the negative tolerance in the same way (as needed).
Selection
5 Accept the settings and return to the Measurement screen.
· Internal calculations are performed on floating-point values, and decisions
round up any fraction of the least-significant digit.
· Displayed values of the reference and tolerances are rounded according to
the selected range. Internal calculations use unrounded data, so decisions are
based on the entered (setting) values.
· An error message appears if you press ENTER with the positive tolerance set
lower than the negative tolerance.
See: “11.3 Error Displays and Remedies” (p. 181) (ERR:001)
37
3.7 Judging Measured Values (Comparator Function)
Decide According to Upper/Lower Thresholds (ABS Mode)
Setting example: Upper threshold 150 m, lower threshold 50 m
To abort the setting process, press previous screen.
. Settings are abandoned and the display returns to the
1 Open the absolute value threshold setting screen.
3
Upper threshold Lower threshold
2
Set the positive tolerance.
Pressing an inoperative key during setting sounds a low-pitch beep (when the key beeper is enabled). No error message is displayed.
1
Selection
To Reset Numerical Values Deletes entered digits. This key is enabled only when entering numerical values.
(Example: 150 m)
2
1_
15_
150_
150.0m
To change the value after selecting the
units, use the cursor
keys to select
the item to change, then enter the new val-
ue with the tenkeys.
3
Press the units key to accept the setting and move the cursor to the lower
threshold value.
Set the negative tolerance in the same way.
1
Selection
2
(Example: 50 m)
5_
50_
50.0m
Press the units key to accept the setting and move the cursor to the upper
threshold.
4 Accept the settings and return to the Measurement screen.
· Internal calculations are performed on floating-point values, and decisions
round up any fraction of the least-significant digit.
· Displayed values of the reference and tolerances are rounded according to
the selected range. Internal calculations use unrounded data, so decisions are
based on the entered (setting) values.
· An error message appears if you press ENTER with the positive tolerance set
lower than the negative tolerance.
See: “11.3 Error Displays and Remedies” (p. 181) (ERR:001)
38
3.8 Confirming Faulty Measurements
3.8 Confirming Faulty Measurements
When a measurement is not performed correctly, a measurement fault indicator appears and a measurement fault signal is output at the ERR pin of the EXT. I/O connector (except for out-of-range detection). The instrument detects measurement faults by the following four methods.
Out-of-Range
See: “Out-of-Range Detection Function” (p. 39)
Display
+OvrRng -OvrRng
Appears when the measured value is outside of the measurement or display range. Check for a broken sample component. The comparator result is Hi when +OvrRng is displayed, and Lo when -OvrRng is displayed. No external measurement fault signal (ERR) is output.
Contact Check Fault
See: “4.4 Checking for Poor or Improper Contact (Contact Check Function)” (p. 46)
Display
C.E. Hi C.E. Lo
The resistance between the HPOT and HCUR probe contacts, and between the LPOT and LCUR probe contacts, are measured and compared with specified contact fault values. An error message appears when the measured value reaches or exceeds the specified contact fault values. If this error persists, probe wear or cable failure may be the cause. If the error is not cleared by shorting the tips of a known-good measurement cable, the instrument requires repair.
Voltage Level Monitor Fault
See: “4.6 Detecting Measurement Voltage Faults (Voltage Level Monitor Function)” (p. 49)
Display
C.E. Volt
This method monitors the stability of the voltage between HPOT and LPOT probe
contacts. An error message appears when voltage instability is detected due to
chattering of the probe con-
tacts.
If this error persists, the probes may be degraded due to wear.
C.E. Volt may also be displayed when external noise is strong.
Current Monitor Fault
See: “Current Monitor Function” (p. 39)
This method monitors the regulated measurement current for normal flow through
the DUT.
An error is detected mainly when a measurement fault occurs due to an open-
circuit DUT or between the HCUR and LCUR probe’s poor contacts. The error
display depends on the contact check and voltage level monitor states (see the
table below).
– – – – – – –
This display appears after changing measurement settings and before the next measurement is performed.
Display Examples: Display Measurement State and Appearance with Open-Circuit Probe
Display Measurement State
Current Monitor Results Normal (PASS)
Fault (FAIL)
Contact Check Results
Voltage Level Monitor Results
Normal (PASS)
Display: Measured Value
Display: +OvrRng/ -OvrRng
COMP indicator: According to the COMP indicator: Hi/ Lo
measured value
(when connection to the measure-
ment object is broken)
Fault (FAIL)
Display: C.E. Hi/ C.E. Lo/ C.E. Volt Display: C.E. Hi/ C.E. Lo/ C.E. Volt
COMP indicator: No decision
COMP indicator: No decision
EXT. I/O: ERR signal output
EXT. I/O: ERR signal output
The measurement fault display differs according to detection order and settings.
39
3.8 Confirming Faulty Measurements
Measurement Fault Detection Order
Measurement Fault Detection
Probe Short Circuit
No Hi Wiring Contact Error
No Lo Wiring Contact Error
No Voltage Level Monitor Error
No Constant-Current Error
No Below Lower Limit
No Above Upper Limit
No No Measurement Data
Display
ERR: 021 Yes
C.E. Hi Yes
C.E. Lo Yes
EXT. I/O signal PRB_SHORT output, ERR output
ERR output, CE_HI output
ERR output, CE_LO output
C.E. Volt ERR output Yes
+OvrRng HI output Yes
-OvrRng LO output Yes
+OvrRng HI output Yes
– – – – – – – –
Measurement fault detection proceeds in the order shown at the left, ending
with display of the first detected error.
Corresponding measurement
3 fault signals are also output at
the EXT. I/O connector.
Out-of-Range Detection Function
Examples of Out-of-Range Faults
Out-of-Range Detection
Measurement Example
The measured value is outside of the measurement range.
Attempting to measure 13 k with the 10 k range selected
The relative tolerance (%) display of the measured value exceeds the dis-
Measuring 500 (+2400%) with a reference value of 20
play range (999.999%).
The zero-adjusted value is outside of In the 1 range with 0.5 zero-adjustment in effect, measuring 0.1 provides
the display range.
a zero-adjusted value of -0.4 , which is outside of the display range.
While measuring, input voltage exceed the A/D converter input range.
Measuring a large resistance value in an electrically noisy environment
Current Monitor Function
The instrument supplies constant measurement current through the DUT via the
HCUR and LCUR probes. A current monitor fault occurs if constant current
cannot be attained. If the contact check and voltage level monitor results are
normal, the out-of-range and comparator result displays indicate “Hi”.
Example of Current Monitor Fault · Broken DUT (open work) · HCUR or LCUR probe
contact fault · HCUR or LCUR cable break
40
3.8 Confirming Faulty Measurements
41
4.1 Making Range-Specific Measurement Settings
Customizing Measurement
Settings
Chapter 4
(set as needed)
Change measurement settings as appropriate for your application.
Refer to “Detailed Settings Screens” (p. 19) for the available settings.
4
4.1 Making Range-Specific Measurement Settings
These settings can be made for each range (except for the DELAY1 setting).
1 Open the Basic Settings screen.
The Basic Settings screen appears.
2 Open the Measurement Settings Screen.
1 Selection
3 Select the range to use.
1 Selection
2
4 Set the items as needed.
2
The Measurement Settings Screen appears. [MEAS SETTINGS]
42
4.2 Setting Pre-Measurement Delay
4.2 Setting Pre-Measurement Delay
This setting specifies the delay between trigger signal input and the start of measurement. Adjust this setting to delay measurement until the measured value has time to stabilize, so that even if the sample is connected after triggering, measurement starts only after the specified delay. The delay can be set by two methods, as follows.
Adjust this setting to allow for probe contact mechanical stabilization.
Set DELAY1
The DELAY1 setting is common to all ranges. The default setting is 0 ms
(corresponding to trigger signal input at the same time as probe contacts
become stable). Setting DELAY1 affects measurements in all ranges.
Adjust this setting to allow for stabilization of the measurement sample.
Set DELAY2
Set DELAY2 to the time needed for stabilization after measurement current is
applied, such as may be required for inductive components. The setting affects
only the selected range. The default setting is 0 ms (corresponding to
resistance measurement of non-inductive components).
DELAY1 and DELAY2 Timing Chart
Probe Contact Condition
Start TRIG
Stable Contact
Measurement Current
Acquisition of Measured Value
Internal Delay*
Acquisition
End of Measurement Signal EOM
- Internal delay is provided to suit purely resistive (non-reactive) DUTs, and is different for each measurement range.
43
4.2 Setting Pre-Measurement Delay
Determining the Delay Time
Set the delay so that inductance does not affect measurements. To fine tune
the delay, begin with a longer delay than necessary, then gradually shorten it
while watching the measured value.
1 Open the Basic Settings screen.
The Basic Settings screen appears.
2 Open the Measurement Settings Screen.
1 Selection
4
2
The Measurement Settings Screen appears. [MEAS SETTINGS]
3 Set DELAY1 or DELAY2.
1 Selection 2
Tenkeys
DELAY1 is common to all ranges, while DELAY2 can be set for each range
independently (p. 41).
Setting range: 0.0 ms (default) to 100.0 ms
3
4 Return to the Measurement screen.
The confirmation screen appears.
Return to the setting screen.
Save setting and return to previous screen.
Discard setting and return to previous screen.
44
4.3 Setting the Measurement Integration Time Option
4.3 Setting the Measurement Integration Time Option
The integration time can be optionally set for each range by selecting FAST, MED, or SLOW measurement speed. Integration time can be set in ms or PLC* units.
- PLC = Power Line Cycle, where one PLC is the period of the power line waveform. At 50 Hz, one PLC = 1/50th of a second, and at 60 Hz, one PLC = 1/60th of a second.
PLC setting units are useful where measurements may be affected by power line noise (high- or low-resistance measure-
ments)
Default Settings
Range
100m 1000m 10 100 1000 10k 100k 1000k 10M 100M
LOW POWER: OFF (p. 28) *1
Integration Time [INT]
FAST
MED SLOW
OVC
0.5 ms
5.0 ms
1PLC
ON *2
0.3 ms
2.5 ms
1PLC
ON *2
0.3 ms
2.5 ms
1PLC
ON *2
0.3 ms
3.0 ms
1PLC
OFF
0.3 ms
3.0 ms
1PLC
OFF
0.3 ms
3.0 ms
1PLC
OFF
0.5 ms
3.0 ms
1PLC
OFF
1.5 ms
5.0 ms
1PLC
OFF
2.5 ms
1PLC
1PLC
OFF
1PLC
2PLC
4PLC
OFF
LOW POWER: ON (p. 28) *1
Integration Time [INT]
FAST
MED
SLOW
OVC
–
–
–
–
0.5 ms 0.5 ms 0.3 ms 0.3 ms
–
5.0 ms 5.0 ms 2.5 ms 2.5 ms
–
1PLC 1PLC 1PLC 1PLC
–
ON 2 ON 2 ON 2 ON 2
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
1. Low Power = Low-Power Resistance Measurement (p. 28) 2. Two measurements are made within the above integration times.
1 Open the Basic Settings screen.
The Basic Settings screen appears.
2 Open the Measurement Settings Screen.
1 Selection
2
The Measurement Settings Screen appears. [MEAS SETTINGS] (The settings for the
current measurement range are displayed.)
45
4.3 Setting the Measurement Integration Time Option
3 Select the integration setting units.
1
Selection
4 Select the integration time.
2
Set in units of time Set in units of power line cycles
The setting is specific to the selected range (p. 41)
1
Selection
2 3
4
Setting range: · When setting in ms units: 0.1ms to 100.0ms · When setting
power-line-cycle units: 1 to 6PLC (60 Hz),
1 to 5PLC (50 Hz)
5 Return to the Measurement screen.
The confirmation screen appears.
Return to the setting screen.
Save setting and return to previous screen.
Discard setting and return to previous screen.
· The instruments accuracy specifications are applicable only with the default
integration times. Investigate your measurement requirements carefully before
changing the integration time.
· When the effects of power line noise can be ignored, the integration time
can be set longer than the default to reduce scattering of measured values. On
the other hand, if the integration time is too short, scattering increases.
For high-or low-resistance and low-power resistance measurements that are
easily affected by power line noise, we suggest setting according to the power
line period (PLC units).
46
4.4 Checking for Poor or Improper Contact (Contact Check Function)
4.4 Checking for Poor or Improper Contact (Contact Check Function)
This function detects poor contact between the probes and DUT, and broken
measurement cables. The instrument continually monitors the resistance between
the HCUR and HPOT probes and the LCUR and LPOT probes from the start of
integration (including response time) and while measuring. When the resistance
is outside of the specified value, a contact check fault occurs and the C.E.
Hi or C.E. Lo error message appears. No comparator decision is applied to the
measured value. When these error messages appear, check the probe contacts,
and check for broken measurement cables. If the error is not cleared by
shorting the tips of a known-good measurement cable, the instrument requires
repair.
· During low-resistance measurement, poor contact of the HCUR or LCUR probe
may be detected as an out-ofrange measurement.
· When contact checking is disabled, measured values may be displayed even
when a probe is not contacting the DUT.
1 Open the Basic Settings screen.
The Basic Settings screen appears.
2 Open the Measurement Settings Screen.
1 Selection
2
The Measurement Settings Screen appears. [MEAS SETTINGS]
3 Enable the Contact Check function.
1
2
Disables the function (go to step 5) Enables the function (default)
Selection
The setting is specific to the selected range (p. 41)
4
Select the contact check fault threshold resistance.
1
2
Selection
50, 100, 150, 200 (default), 300, 400, 500 A contact fault occurs when a measured value exceeds the threshold setting.
5 Return to the Measurement screen.
The confirmation screen appears.
Return to the setting screen.
Save setting and return to previous screen.
Discard setting and return to previous screen.
47
4.5 Improving Probe Contact (Contact Improver Function)
4.5 Improving Probe Contact (Contact Improver Function)
Probe contacts can be improved by applying current from the POT to the CUR probes before measuring.
The Contact Improver function applies voltage to the sample. Be careful when measuring samples with characteristics that may be affected.
The current used for the Contact Improver functions can be selected as follows.
17 mA, 25 mA, 35 mA (default), 50 mA
Higher current provides more effective contact improvement, but at the cost of faster probe deterioration. Contact Improver current can be set to be disabled (OFF), enabled (ON), or PULSE.
4
The PULSE setting applies the contact improvement current for about 100 µs immediately before measure-
ment. The PULSE setting is usefull to decrease Joule heating if the DUT is susceptible to its current.
DUT current * DUT voltage
100 m-range to 100 k-range 1 M to 100 M-range
2mA max. 20 V max.
60 mA max. 15 V max.
*1 Steady state value. A rush current of approximately 100 mA flows for 100µs, when a probe came in contact with DUT.
*: It takes several microseconds for the DUT current to reach the steady-state value. Until the steady-state value is reached, a transient current that is approximately equal to the contact improvement current setting (default setting: 35 mA) will flow.
Timing Chart
Probe Contact
(Contact Improver Function) Condition
Stable Contact
- Internal delay is different for each range.
Start TRIG
OFF setting
Measurement
Contact Improver current: Off
Internal DELAY 1 DELAY 2 delay *
Measuring
ON setting
Probe Contact Condition
Start TRIG
Measurement Contact Improver current: On
Stable Contact
Internal DELAY 1 DELAY 2 delay *
Measuring
PULSE setting
Probe Contact Condition
Start TRIG
Measurement
Contact Improver current: Pulse
Stable Contact
DELAY 1
Approx.
100 µs
Internal DELAY 2 delay *
Measuring
48
4.5 Improving Probe Contact (Contact Improver Function)
For ranges between 1000 k and 100 M, the [PULSE] setting is enabled by
default. Before measuring in the ranges from 1000 k to 100 M with the Contact
Improver function set to [ON], verify that measurements are not biased.
1 Open the Basic Settings screen.
The Basic Settings screen appears.
2 Open the Measurement Settings Screen.
1 Selection
2
The Measurement Settings Screen appears. [MEAS SETTINGS]
3 Set the Contact Improver current timing to disabled (OFF), enabled (ON), or PULSE.
1 Selection
2
Disable probe contact improvement
(go to step 4).
Enable probe contact improvement.
Apply contact improvement current for about 100 µs immediately before mea-
surement.
The setting is specific to the selected range(p. 41)
(When selecting ON or PULSE) Set the current limit value.
1
Selection
2
17mA, 25mA, 35mA (default), 50mA
4 Return to the Measurement screen.
The confirmation screen appears.
Return to the setting screen.
Save setting and return to previous screen.
Discard setting and return to previous screen.
49
4.6 Detecting Measurement Voltage Faults (Voltage Level Monitor Function)
4.6 Detecting Measurement Voltage Faults (Voltage Level Monitor Function)
When a measurement voltage fault occurs due to probe chattering, the C.E. Volt
error message appears on the measurement screen and an ERR signal is output.
The C.E. Volt error may also appear when external noise is strong.
Check the following if errors occur frequently: · Check for probe
deterioration. · Provide additional noise suppression. “Appendix 3 Unstable
Measurement Values” (p. A3) · Set the voltage level monitor to Loose, or OFF
(disable).
4
1 Open the Basic Settings screen.
The Basic Settings screen appears.
2 Open the Measurement Settings Screen.
1 Selection
2
The Measurement Settings Screen appears. [MEAS SETTINGS]
3 Enable or disable the function.
1
2
Disables the function (go to step 5) Enables the function (default)
Selection
The setting is specific to the selected range (p. 41)
4 (When enabled (ON selected))
Select the voltage level monitor threshold.
1
Selection
2
Loose Normal Severe
*. Default setting: Loose is the default for the 100 M
range, and Normal for all
other ranges.
5 Return to the Measurement screen.
The confirmation screen appears.
Return to the setting screen.
Save setting and return to previous screen.
Discard setting and return to previous screen.
50
4.7 Applying Current Only When Measuring (Current Mode Setting)
4.7 Applying Current Only When Measuring (Current Mode Setting)
When the Contact Improver function is set to Pulse or disabled (CONT IMP:
PULSE or OFF) and measurement current is set for pulse output, open-circuit
voltage when not measuring does not exceed 20 mV.
See: “4.5 Improving Probe Contact (Contact Improver Function)” (p. 47)
When the Contact Improver function is enabled (CONT IMP: PULSE or ON setting),
the current mode setting is ignored even if set to continuous (CURRENT MODE:
CONT setting). The Contact Improver function forces pulse operation with
measurement current applied only during measurement.
1 Open the Basic Settings screen.
The Basic Settings screen appears.
2 Open the Measurement Settings Screen.
1 Selection
2
The Measurement Settings Screen appears. [MEAS SETTINGS]
3 Select whether to apply current when not measuring.
2
Measurement current is applied while
1
awaiting trigger. Measurement current is applied only
while measuring (default).
Selection
The setting is specific to the selected range (p. 41)
To apply measurement current continuously (CONT setting) even when waiting for a trigger, confirm that the Contact Improver function is disabled (CONT IMP: OFF, (p. 47)).
4 Return to the Measurement screen.
The confirmation screen appears.
Return to the setting screen.
Save setting and return to previous screen.
Discard setting and return to previous screen.
51
4.8 Test for Short-Circuited Probe (Probe Short-Circuit Detection Function)
4.8 Test for Short-Circuited Probe (Probe Short-Circuit Detection Function)
Four-terminal measurements are not possible when a conductive foreign object is present between the POT and CUR probe tips. To detect short-circuited probes, this function measures the resistance between the CUR and POT terminals after a specific time (initially 5 ms) following the end of measurement. Probe short-circuit detection is disabled by default.
DUT Electrode DUT
POT CUR
When a probe short-circuit is detected, an error message appears on Foreign Object the measurement screen, and the PRB_SHORT and ERR signals are
4
output. (ERR:021 Probe short error)
Short-circuit detection can also be controlled by asserting the active-low PRB_CHECK EXT. I/O signal. Asserting the PRB_CHECK signal while measuring causes short-circuit detection to be performed after the end of measurement (p. 85).
About Probe Short-Circuit Detection
· If probes are connected to the DUT during probe short-circuit detection, the
short circuit is detected. Ensure that the probes have sufficient time to
separate from the measurement object.
· Probe short-circuit detection occurs within about 1 ms. · The threshold for
probe short-circuit detection is fixed at 500 , so if the resistance between
CUR and POT
probes is larger, detection is not possible.
Timing Chart (Probe Short-Circuit Detection)
Transport
Measurement stage
Contact
DUT1 contact
Transport
Transport
DUT2 contact
Probe short-circuit
Start TRIG
Measurement
End of Measurement EOM
Comparator Hi, IN, Lo
Probe short-circuit PRB_SHORT
DUT1 measurement
Short-Circuit Detection
DUT2 measurement
Detection timing setting Available
No Decision
Short-Circuit
· Even while the probe short-circuit detection function is set to be disabled,
short-circuit detection is performed when the EXT. I/O PRB_CHECK signal is
asserted.
· When the internal trigger [TRG: INT] source is selected, short-circuit
detection is not performed after the end of measurement. However, short-
circuit detection can still be executed by asserting the PRB_CHECK signal.
52
4.8 Test for Short-Circuited Probe (Probe Short-Circuit Detection Function)
Probe Short-Circuit Detection Enable/Disable
1 Open the Basic Settings screen.
The Basic Settings screen appears.
2 Open the System screen.
1 Selection
3 Enable or disable the function.
1
Selection
2
The System screen appears. [SYSTEM] 2
Disables the function (default) (go to step 5) Enables the function
4 (When enabled (ON selected))
Set the probe detection timing.
1
Selection
Short-circuit detection is delayed for the specified time following the end of measurement.
2
Setting range: 1 to 100 ms, 5 ms (default)
3
5 Return to the Measurement screen.
The confirmation screen appears.
Return to the setting screen.
Save setting and return to previous screen.
Discard setting and return to previous screen.
53
4.9 Comparing the Measurement Settings of Two Instruments (Settings Monitor
Function)
4.9 Comparing the Measurement Settings of Two Instruments (Settings Monitor Function)
This function automatically compares the settings of two
instruments to determine whether they are the same.
2nd Stage
Only those measurement settings affecting the comparator and speed are compared. When the settings differ, an alarm notification appears and subsequent TRIG sig-
Automatic Comparison
1st Stage
nal input is prevented from starting measurement.
4
When the settings of two instruments match, TRIG input is accepted and measurement starts. However, if the range defined by the upper and lower thresholds of the second stage is broader than that of the first stage, measurement still starts despite the different threshold settings.
E DCBA
Transport Direction
1 Connect the two instruments’ SET MONITOR connectors together using a Hioki 9637 RS-232C cable.
2 Open the Basic Settings screen.
The SET MONITOR connectors are identical to RS-232C connectors. Be careful to
avoid connecting the wrong connectors.
The Basic Settings screen appears.
3 Open the System screen.
1 Selection
4 Enable the function on both instruments.
1
Selection
2 The System screen appears. [SYSTEM] 2
Disables the function (default) Enables the function
54
4.9 Comparing the Measurement Settings of Two Instruments (Settings Monitor
Function)
5 Select the instrument to serve as the 1st stage, and set its tolerance range.
1
2
Selection
Selects this instrument as the 1st stage Selects this instrument as the 2nd stage
Example: If the 1st stage is set to measure 12 ±0.800%, and the 2nd is to
measure 12 ±1.000%, enter the difference in toler-
ance of 0.300%.
3
Enter the difference in tolerance (%) to be allowed at the 2nd stage from the tol-
erance range set for the 1st stage.
4
Setting range: 0.000 to 9.999%
6 Set the instrument to serve as the 2nd stage (another RM3542).
1
2
Selection
Selects this instrument as the 1st stage Selects this instrument as the 2nd stage
7 Return to the Measurement screen.
The confirmation screen appears.
Return to the setting screen.
Save setting and return to previous screen.
Discard setting and return to previous screen.
Tolerance Range Setting Conditions
Permissible tolerance is calculated using floating-point values, so the setting must be at least 0.001% larger than the difference between 2nd and 1st stage ranges. Set the upper and lower comparator thresholds according to the following conditions: 1st stage upper threshold < 2nd stage upper threshold 1st stage lower threshold > 2nd stage lower threshold
REF% setting 1st stage
2nd stage
ABS setting 1st stage
2nd stage
(upper limit) UPP [%] REF [] LOW [%] (lower limit)
Tolerance range [%]
(upper limit) UPP [] LOW [] (lower limit)
Tolerance range [%]
Tolerance range [%] > 2nd upper limit [%] – 1st upper limit [%]
Tolerance range[%]
2nd upper limit – 1st upper limit x 100
1st upper limit
Tolerance
1st lower limit – 2nd lower limit
range[%] >
1st lower limit
x 100
55
4.9 Comparing the Measurement Settings of Two Instruments (Settings Monitor
Function)
Practical Example
SET MONITOR: ON 1st 0.300% (on the System screen)
1st Stage
SET MONITOR: ON 2nd (on the System screen)
2nd Stage
TRIG Input Accepted
1st Stage
When changing the reference value 2nd Stage
4
TRIG Input Inhibited
1st Stage
The error message appears when the settings do not match.
2nd Stage
Change the reference value to match the 1st stage
TRIG Input Accepted
When an error is displayed ERR:003 Setting monitor error. (COMP)
ERR:004 Setting monitor error. (SPEED)
Comparator settings do not match. Please check. Measurement speed settings do not match. Please check.
56
4.10 Retrying Measurement After a Fault (Retry Function)
4.10 Retrying Measurement After a Fault (Retry Function)
The Retry function causes measurement to be
automatically retried when a measurement fault
occurs due to probe chatter.
Probe Contact Condition
During Retry, all measurement operations including Con-
tact Improvement and DELAY2 (but excluding DELAY1) Start
are restarted.
TRIG
If a measurement fault persists after the specified contin- Contact
uous retry interval (e.g., if the DUT is not connected), Improver
retrying is aborted and the EOM signal is output. When Measurement Retry is enabled, the maximum time to end-of-measure- Current
ment occurs when recovering from a measurement fault Contact Check immediately before the retry interval expires, which
approaches the sum of the retry interval setting plus nor- End of Measurement
mal measurement time. Decreasing test throughput may EOM
indicate probe maintenance is required.
1 Open the Basic Settings screen.
Chatter Retry
The Basic Settings screen appears.
2 Open the System screen.
1 Selection
3 Select whether to enable or disable Retry.
1
Selection
4 (When enabled (ON selected)) Set the continuous retry interval.
2
The System screen appears. [SYSTEM] 2
Retry disabled (go to step 5) Retry enabled (default)
1
Selection
2
Setting range: 1 to 50 ms (default: 50 ms)
3
5 Return to the Measurement screen. The confirmation screen appears.
Return to the setting screen.
Save setting and return to previous screen.
Discard setting and return to previous screen.
57
4.11 Maintaining Measurement Precision (Self-Calibration)
4.11 Maintaining Measurement Precision (Self-Calibration)
To maintain measurement precision, the instrument self-calibrates every ten
minutes to compensate for internal circuit offset voltage and gain drift. This
function cannot be disabled. During self-calibration, the subsequent
measurement is delayed for about 6PLC + 10 ms (PLC = Power Line Cycles) for
internal circuit compensation.
Self-Calibration Timing Within 130 ms at 50 Hz, or 110 ms at 60 Hz
4 · When the timing of self-calibration overlaps with a measurement, self-
calibration is postponed until the end of measurement. · When a trigger signal
is applied during self-calibration, the start of the triggered measurement is
postponed
until self-calibration is finished. · Self-calibration executes automatically
after changing comparator or measurement speed settings. · During self-
calibration, measurement current and the Contact Improver current are
inhibited.
4.12 Compensating for Thermal EMF Offset (Offset Voltage Compensation – OVC)
This function automatically compensates for offset voltage resulting from
thermal emf or internal instrument bias. (OVC: Offset Voltage Compensation)
See: “Appendix 2 Effect of Thermal emf” (p. A2)
The following value is known to be a true resistance value from RP (>0), the
value measured with current flowing in the positive direction, and RN (<0),
the value measured with current flowing in the negative direction.
R—–P—–—–R—-N–2
Offset voltage compensation is automatically enabled in the following
conditions, and cannot be modified or disabled: · When a range from 100 m to
10 is selected. · When low-power resistance measurement is enabled (LOW POWER:
ON).
When the test object is inductive, some delay (DELAY2) is required (p. 42) to
allow adequate current flow before starting measurement.
58
4.12 Compensating for Thermal EMF Offset (Offset Voltage Compensation – OVC)
59
5.1 Disabling and Enabling Key Operations
System Settings
5.1 Disabling and Enabling Key Operations
Disabling Key Operations (Key-Lock Function)
Activate the key-lock function to disable the instrument’s front panel key operations. Three key-lock levels are available to suit specific purposes.
Only comparator settings are enabled.
Disabling All Except Comparator Settings
5 Key operations other than comparator settings (REF%, ABS, units and tenkeys)
and F1 [UNLOCK] keys are disabled. To disable key operations: select [MENU]
[M.LOCK] is displayed when returning to the measurement screen.
Key operations to change settings are disabled (although keylock can be canceled).
Disabling All Key Operations Including Comparator Settings
All key operations except F1 [UNLOCK] are disabled. To disable key operations:
select [FULL] [F.LOCK] is displayed when returning to the measurement screen.
All key operations are disabled.
Disabling All Panel Keys
Asserting (Low) the EXT. I/O KEY_LOCK signal disables all panel keys,
including F1 [UNLOCK] and F1 [LOCAL] (disables remote control) (p. 85). To
disable the key-lock function and re-enable the keys, de-assert (High) the
KEY_LOCK signal.
1 Open the Basic Settings screen.
The Basic Settings screen appears.
2 Enable or disable key operations.
1 Selection
3 Return to the Measurement screen.
2
Key operations enabled (default) Disable all except key-lock cancel Disable
all except key-lock cancel and comparator setting change
[UNLOCK] is displayed only when key-lock is enabled by front panel key
operations.
60
5.1 Disabling and Enabling Key Operations
Re-Enabling Key Operations (Key-Lock Cancel)
Key-lock can be canceled only when [UNLOCK] is displayed. Press and hold F1
[UNLOCK] for one second.
If key operations are disabled by the KEY_LOCK signal, de-assert (High) the
signal to unlock the keys.
61
5.2 Setting the Comparator Decision and Key Beepers
5.2 Setting the Comparator Decision and Key Beepers
Enabling or Disabling the Key Beeper
The key beeper sound can be enabled and disabled. The key beeper is enabled
(ON) by default.
1 Open the Basic Settings screen.
The Basic Settings screen appears.
5
2 Open the System screen.
1 Selection
2
The System screen appears. [SYSTEM]
3 Select whether to enable or disable the key beeper.
2
Disables the beeper
1
Enables the beeper (default)
Selection
4 Return to the Measurement screen.
The confirmation screen appears.
Return to the setting screen.
Save setting and return to previous screen.
Discard setting and return to previous screen.
62
5.2 Setting the Comparator Decision and Key Beepers
Setting the Comparator Decision (“JUDGE”) Beeper
The comparator decision beeper can be enabled and disabled. The decision
beeper is disabled (OFF) by default.
1 Open the Basic Settings screen.
The Basic Settings screen appears.
2 Open the System screen.
1 Selection
2
The System screen appears. [SYSTEM]
3 Select whether to enable or disable the decision beeper.
2
Disables the beeper (default)
1
Enables the beeper
Selection
4 (When enabled (ON selected))
Select the decision beep conditions.
1
Selection
5 Return to the Measurement screen.
The confirmation screen appears.
2 IN (beep when within range)
HI/LO (beep when out of range) LOW (beep when below lower threshold) HIGH
(beep when above upper threshold)
Return to the setting screen. Save setting and return to previous screen.
Discard setting and return to previous screen.
63
5.3 Power Line Frequency Manual Setting
5.3 Power Line Frequency Manual Setting
For proper electrical noise suppression, the instrument needs to be set to
match the power line frequency. With the default setting (AUTO), the
instrument attempts to automatically detect the line frequency, but manual
setting is also available. Unless the line frequency is set correctly,
measured values may be unstable. An error message appears if line noise is
high enough to prevent correct frequency detection (ERR:041(p. 181)). In that
case, set the instrument’s line frequency manually.
When the AUTO setting is selected, the line frequency is automatically set to 50 or 60 Hz when the instrument is turned on or reset.
However, automatic detection is not available when the line frequency changes after turning power on or
resetting.
If the actual line frequency deviates from 50 or 60 Hz, select the closest frequency. Examples:
5
If the actual line frequency is 50.8 Hz, select the 50 Hz setting.
If the actual line frequency is 59.3 Hz, select the 60 Hz setting.
1 Open the Basic Settings screen.
The Basic Settings screen appears.
2 Open the System screen.
1 Selection
3 Select the line frequency being used.
1
Selection
4 Return to the Measurement screen.
The confirmation screen appears.
2
The System screen appears. [SYSTEM] 2 Automatically detect local line
frequency (default) When the line frequency is 50 Hz When the line frequency
is 60 Hz
Return to the setting screen. Save setting and return to previous screen.
Discard setting and return to previous screen.
64
5.4 Setting the Clock
5.4 Setting the Clock
To record and print the correct time when using statistical calculations (p.
74), the clock needs to be set correctly. The time of printing is also output
when printing statistical calculation results.
1 Open the Basic Settings screen.
The Basic Settings screen appears.
2 Open the System screen.
1 Selection
3 Set the date and time.
2
The System screen appears. [SYSTEM]
1
Selection
2
Enter the last two digits of the year, and the month, day, hour, minutes and seconds in that order (the cursor moves automatically). Enter two digits for all values (e.g., 09).
4 Return to the Measurement screen.
Clock settings cannot be canceled.
65
5.5 Adjusting Screen Contrast
5.5 Adjusting Screen Contrast
The screen may become hard to see when ambient temperature changes. In this
case, adjust the contrast.
1 Open the Basic Settings screen.
The Basic Settings screen appears.
2 Open the System screen.
1 Selection
3 Adjust the contrast.
1
2
5
The System screen appears.
[SYSTEM]
2
0 to 100%, 5% step (default: 50%)
Selection
4 Return to the Measurement screen.
The confirmation screen appears.
Return to the setting screen.
Save setting and return to previous screen.
Discard setting and return to previous screen.
66
5.6 Adjusting the Backlight
5.6 Adjusting the Backlight
Adjust backlight brightness to suit ambient illumination.
· When external (EXT) triggering is selected, backlight brightness is
automatically reduced after non-operation for one minute.
· Be aware that the display may be hard to see when brightness is set too low
(near 0%).
1 Open the Basic Settings screen.
The Basic Settings screen appears.
2 Open the System screen.
1 Selection
3 Adjust the backlight.
1
2
The System screen appears. [SYSTEM] 2
0 to 100%, 5% step, (default: 80%)
Selection
4 Return to the Measurement screen.
The confirmation screen appears.
Return to the setting screen.
Save setting and return to previous screen.
Discard setting and return to previous screen.
67
5.7 Initializing (Reset)
5.7 Initializing (Reset)
The instrument can be reset by three methods.
· System reset from the System screen: Returns all settings (except the clock)
to factory defaults. · Turn the instrument on while simultaneously holding the
REF% and ABS keys: Returns all settings (except
the clock) to factory defaults. · Reset by remote control command: returns all
settings (except communication and clock settings) to their
factory defaults.
RST command (non-backup, (p. 123)) :SYSTem:RESet command (p. 137)
This procedure describes system reset from the System screen.
1 Open the Basic Settings screen.
The Basic Settings screen appears.
5
2 Open the System screen.
1 Selection
3 Select RESET.
1
2
The System screen appears. [SYSTEM] 2
Returns all settings to their factory defaults
Selection
4 Select whether to cancel or proceed to execute system reset.
Cancel the operation
Execute The Measurement screen is displayed when system reset finishes.
68
5.7 Initializing (Reset)
Default Settings
Display
MENU MISC MEAS
DATA
SYSTEM
TRG
RANGE
SPEED 0ADJ COMP LOCK MISC
Setting value
Default Settings
INT/ EXT/ MANU / [Low Power: Off] 100m/ 1000m/ 10/ 100/ 1000/ 10k/ 100k/ 1000k/ 10M/ 100M [Low Power: On] 1000m/ 10/ 100/ 1000 SLOW/ MED/ FAST
EXT 100 M FAST
OFF/ ON
OFF
OFF/ ON
ON
OFF/ FULL / MENU
OFF
MEAS/ DATA/ SYSTEM
Setting Description Trigger source selection (p. 30)
Range selection (p. 31)
Measurement speed (p. 29) Zero adjustment (p. 32) Comparator function(p. 34)
Key-Lock function (p. 59) (Miscellaneous settings)
DELAY1 DELAY2 INT (FAST) INT (MED) INT (SLOW)
CONT CHECK
0 to 100 ms (all ranges) 0 to 100 ms
0.1 ms to 100 ms 1PLC to 6PLC (60 Hz) 1PLC to 5PLC (50 Hz)
OFF/ ON 50 / 100 / 150 / 200 / 300 / 400 / 500
CONT IMP
OFF/ ON/ PULSE 17 mA/ 25 mA/ 35 mA/ 50 mA
VOLT
OFF/ ON
MONITOR
LOOSE/ NORMAL/ SEVERE
CURRENT MODE
CONT/ PULSE
AUTO MEMO-
RY
OFF/ ON
STATISTICS OFF/ ON
DATA OUT
OFF/ ON
SET MONITOR
OFF/ ON, 1st/ 2nd, 0.000% to 9.999%
PROBE CHECK
OFF/ ON, 0 to 100 ms
RETRY
OFF/ ON, 1 to 50 ms
TRIG EDGE OFF EDGE/ ON EDGE
EOM
PULSE/ HOLD 1 to 100 ms
INTERFACE GP-IB/ RS232C/ PRINT
LOW POWER OFF/ ON
JUDGE BEEP
KEY BEEP CLOCK LINE FREQ CONTRAST BACK LIGHT RESET ADJUST
OFF/ ON IN/ HI/LO/ LOW/ HIGH OFF/ ON
AUTO/ 50 Hz/ 60 Hz 0 to 100 0 to 100 –
0 ms 0 ms
Probe delay setting (p. 42) DUT response setting (p. 42)
Depends on measurement range Integration time (p. 44)
ON, 200
Contact-check (p. 46)
ON, 35 mA (range from 100 m to
100 k) PULSE, 35 mA (range from1000
Contact Improvement (p. 47)
k to 100 M)
ON, NORMAL (LOOSE when 100 Range)
Voltage level monitor (p. 49)
PULSE
Current mode setting (p. 50)
OFF
OFF OFF OFF, 1st, 0.000%
Auto-Memory function (p. 71)
Statistical calculation function (p. 74) Data output function (p. 77)
Settings Monitor function (p. 53)
OFF, 5 ms
ON, 50 ms ON EDGE PULSE, 5 ms RS232C, 9600bps GP-IB, ADR01, LF OFF
OFF, HI/LO ON
AUTO 50 80 –
Probe short-circuit detection (p. 51)
Retry function (p. 56) Start Logic Setting (p. 94) End-of-measurement pulse
width (p. 93)
Interface setting (p. 101)
Low-Power Resistance Measurement (p. 28)
Comparator decision beeper (p. 61)
Key beeper (p. 59) Clock setting (p. 64) Line frequency (detection) (p. 63)
Screen contrast adjustment (p. 65) Screen backlight adjustment (p. 66) Reset
(p. 67) Calibration (p. A13)
69
Storing and Exporting Data
Chapter 6
Measured values can be stored or automatically exported, according to application. Stored data can be output to a printer, RS-232C or GP-IB. Also, statistical calculations can be applied to internally stored data.
Stored measurements are lost when the instrument is turned off. Therefore, be sure to print out or export important data to a PC.
Store measured values at specific times.
This is convenient for batch exporting data to a controller while switching
reels.
Data Memory Function (p. 70)
Store up to 30,000 measured values using the EXT. I/O TRIG signal or by
pressing F4 [MANU] on the Basic Settings screen.
6
Store data after measured value has stabilized.
This is convenient for sample inspection after printing (vapor deposition)
resistors on a board.
Auto-Memory Function (p. 71)
Measured values are automatically stored as they become stable. When the
specified number of data points (up to 99) is acquired, the beeper sounds and
auto-storing halts.
Automatically output (export) measurements at the end of measurement.
Data Output Function (p. 77)
Minimizes transfer time by eliminating the need for transmit requests from the
remote controller. (RS-232C interface only)
70
6.1 Storing Data at Specific Times (Data Memory Function)
6.1 Storing Data at Specific Times (Data Memory Function)
Measured values are stored in the instrument’s internal memory according to
the following timings. (up to 30,000 points)
· Every time a measurement is performed by external (EXT) triggering · When a
trigger is applied during internally (INT) triggered measurement
The following three storage methods are available: · Store upon receiving an
EXT. I/O TRIG signal (p. 85)
· Store upon receiving a TRG command (p. 126)
· Store by pressing the F4 [MANU] key on the [MENU] [TRG] setting screen.
· This function can only be enabled by remote control. The data memory
function should be enabled by remote control beforehand. This setting is not
available from the front panel.
· Stored memory data cannot be viewed on the instrument’s screen. Use remote
control commands to export stored data.
Data Memory Function Operating Procedure
1 Enable data memory mode.
Send this remote command to enable the data memory function:
:MEMory:MODE MEMory (p. 142)
2 Store measured values.
Execute external trigger measurement, or apply a trigger during internally
triggered measurement.
3 Export the stored data.
Send this remote command to export the measured values stored in the
instrument:
:MEMory:DATA? (p. 143)
4 Clear measurement data from instrument memory.
Send this remote command to erase the data from instrument memory:
:MEMory:CLEar (p. 142)
Stored data is automatically erased at the following times:
· when the memory function setting (including auto-mem- · when printing the statistical calculations (p. 82)
ory) is changed (p. 142)
· when the DUT is changed (p. 28)
· when the range is changed (p. 31)
· upon system reset (p. 67)
· when changing comparator settings (p. 34)
71
6.2 Store as soon as Measurement is Stable (Auto-Memory Function)
6.2 Store as soon as Measurement is Stable (Auto-Memory Function)
This function automatically stores the value measured each time the probes
contact the sample with internal triggering. When the specified number of
values has been acquired, auto-storage operation stops. Statistical
calculations are applied to the stored data, with results output to the screen
or printer (RS232C).
See: “6.3 Performing Statistical Calculations on Measured Values” (p. 74)
“Chapter 7 Printing” (p. 79)
Data storage and printing can be automatically controlled by the Auto-Memory function.
Prepare the printer (p. 79).
Enable Auto-Memory and set the number of values to store.
Set decision criteria Measure (p. 34).
Printing (p. 82)
Beeper notifies when the specified number of values is stored.
6
The Auto-Memory function is disabled by default. Enable the Auto-Memory
function before setting the number of values to store.
Enabling the Auto-Memory function affects other functions as follows: ·
Statistical calculation is forced on. · The voltage level monitor function is
forced off (although the setting itself is not set to OFF, the function is
actually disabled). · The trigger source setting is forced to internal (INT).
When the trigger source is set to external (EXT), the Auto-Memory function is disabled by force.
Deleting Stored Data Stored data is automatically erased at the following times:
· when the memory function setting (including auto-mem- · when printing the statistical calculations (p. 82)
ory) is changed (p. 142)
· when the DUT is changed (p. 28)
· when the range is changed (p. 31)
· upon system reset (p. 67)
· when changing comparator settings (p. 34)
· upon setting the auto-memory number of values to store
· when the power is turned off
(p. 73)
72
6.2 Store as soon as Measurement is Stable (Auto-Memory Function)
Enabling the Auto-Memory Function
1 Open the Basic Settings screen.
The Basic Settings screen appears.
2 Open the Data Memory Settings screen.
1 Selection
3 Enable the function.
1
Selection
2
The Data Settings screen appears. [DATA SETTINGS] 2
Disable the function (default) Enable the function
4
Return to the Measurement screen.
The confirmation screen appears.
When the function is enabled
Return to the setting screen.
Save setting and return to previous screen.
Discard setting and return to previous screen.
73
6.2 Store as soon as Measurement is Stable (Auto-Memory Function)
Setting the Number of Values to Store
1 Open the Auto-Memory Settings screen.
Total Count Pass Count (IN) Fail Count (Hi/Lo)
2 Enter the number of values to store.
Displays the Auto-Memory setting screen
Setting range: 1 to 99
To Reset Numerical Values
Deletes entered digits. This key is enabled only when entering numerical
values.
To abort the setting process, press
. Settings are abandoned and the display returns to the pre-
vious screen.
6
3 Accept the settings and return to the Measurement screen.
(Example: 20 values set to be stored)
Acquiring Measured Values Automatically
1 Momentarily disconnect (open-circuit) the probes. 2 Connect the probes to
the DUT.
When the measurement is stable, the value is automatically stored and the count is incremented. When the count reaches the specified number of values, a long beep sounds, and subsequent measurements are not stored. The (one) last acquired value can be deleted (Undo function (p. 76)).
74
6.3 Performing Statistical Calculations on Measured Values
6.3 Performing Statistical Calculations on Measured Values
Statistical calculations can be performed on up to 30,000 measured values, with results displayed. Printing is also available (p. 82).
Calculation types: average, maximum and minimum values, population standard deviation, sample standard deviation, process compatibility indices
Maximum value Minimum value
Average
Xmax = MAX (x1, ….., xn) Xmin = MIN (x1, ….., xn)
x = x n
Population standard deviation
Standard deviation of sample
Process capability index (dispersion) *
Cp =
Hi – Lo 6 n-1
Process capability index (bias)*
In these formulas, n represents the number of valid data samples.
Hi and Lo are the upper and lower thresholds of the comparator.
*. The process capability indices represent the quality achievement capability created by a process, which is the breadth of the dispersion and bias of the process’ quality. Generally, depending on the values of Cp and CpK, process capability is evaluated as follows: Cp, CpK>1.33 ……… Process capability is ideal 1.33Cp, CpK>1.00 Process capability is adequate 1.00Cp, CpK ……… Process capability is inadequate
· When only one valid data sample exists, standard deviation of sample and
process capability indices are not displayed.
· When n-1 = 0, Cp and CpK are 99.99. · The upper limit of Cp and CpK is
99.99. If Cp or CpK exceeds 99.99, the value 99.99 is displayed · Negative
values of CpK are handled as CpK = 0. · If statistical calculation is turned
off and then back on without first clearing calculation results, calculation
resumes from the point when it was turned off. · Measurement speed is
restricted when statistical calculation is enabled. · When Auto-Memory is
enabled (ON), statistical calculation is enabled (ON) by force. · When
statistical calculation is disabled (OFF), Auto-Memory is disabled (OFF) by
force.
Deleting Statistical Calculation Results Stored data is automatically erased at the following times:
· when the memory function setting (including data-mem- · when printing the statistical calculations (p. 82)
ory) is changed (p. 142)
· when the DUT is changed (p. 28)
· when the range is changed (p. 31)
· upon system reset (p. 67)
· when changing comparator settings (p. 34)
· upon setting the auto-memory number of values to store
(p. 73)
75
6.3 Performing Statistical Calculations on Measured Values
Using Statistical Calculations
When statistical calculation is enabled and an EXT. I/O trigger signal is
applied, operation is as follows depending on the trigger source setting:
· With external (EXT) triggering: One measurement is performed and subjected
to statistical calculation. · With internal (INT) triggering: The next
measured value after the trigger signal is subjected to statistical cal-
culation.
Operation is the same in the following cases:
(Key Operations)
· when pressing the F4 [MANU] key on the [MENU] [TRG] selection screen
· when pressing the F2 [PRINT] key on the Measurement screen (with internal
triggering and Auto-Memory disabled. Appears only when the interface is set
for the printer.)
· when acquiring measured values by the Auto-Memory function (p. 71)
(Remote Control)
· when a TRG remote control command is received
· when an EXT. I/O print signal is applied on the Measurement screen (with
internal triggering and Auto-Memory disabled)
1 Open the Basic Settings screen.
The Basic Settings screen appears.
6
2 Open the Data Memory Settings screen.
1 Selection
2
The Data Settings screen appears. [DATA SETTINGS]
3 Enable or disable statistical calculation.
1
Selection
4 Return to the Measurement screen.
The confirmation screen appears.
2
Disable statistical calculation (default) Enable statistical calculation
Return to the setting screen. Save setting and return to previous screen.
Discard setting and return to previous screen.
When statistical calculation is enabled, F3 [STAT] appears on the Measurement screen. Confirm calculation results (p. 76)
76
6.3 Performing Statistical Calculations on Measured Values
Confirming, Printing, and Erasing Calculation Results
Statistical calculation results are displayed on the screen. Printing is also
available with the commercially available printer with a serial interface.
Calculation results are automatically erased after printing. Before printing,
select the [PRINT] interface setting.
See: “7.2 Instrument Settings” (p. 81)
The number of valid samples can be confirmed on the Calculation Results
screen. · When the number of valid samples is zero, no calculation results are
displayed. · When only one valid data sample exists, no standard deviation or
process capability indices are displayed.
1 Display the Calculation Results screen.
Displays the Calculation Results screen (if statistical calculation is
enabled).
Num Ave Max Min
Total data count Mean Maximum Minimum
Val Sn Sn1 Cp Cpk
Number of valid measured values (error-free data) Population standard deviation Standard deviation of sample Process capability index (dispersion) Process capability index (bias)
2 To print
To print, select the printer as the interface setting on the System screen (p.
81)
Output to the printer. “Example Printouts” (p. 83)
Statistical calculation results and stored data are erased when printing
finishes.
To erase
Erases the last measurement and calculation result (executes only once).
Erases all measured values and statistical calculation results.
After selecting, a confirmation screen appears.
77
6.4 Auto-Exporting Measured Values (at End of Measurement) (Data Output
Function)
6.4 Auto-Exporting Measured Values (at End of Measurement) (Data Output
Function)
At the end of measurement, the measured value is exported to a computer via
RS-232C.
See: “Chapter 9 Communications (RS-232C/ GP-IB Interface)” (p. 97)
· Set the interface to [RS232C] beforehand. This function is not applicable to
the GP-IB Interface.
See:”9.4 Configuring the Communications Protocol” (p. 101) · When internal
(INT) triggering is selected, data is exported only when a TRIG signal is
applied.
· Executing a :READ? query command exports duplicate measured values.
· For other queries, be careful to avoid overlapping query response timing
with auto-exporting measured values.
· The data format for measured values can be selected as ASCII (default) or
BINARY. Transfer time is minimized when BINARY is selected.
See:”:SYSTem:FORMat <ASCii/ BINary>” (p. 137)
1 Open the Basic Settings screen.
6
The Basic Settings screen appears.
2 Open the Data Memory Settings screen.
1 Selection
2
The Data Settings screen appears. [DATA SETTINGS]
3 Enable or disable auto-exporting (DATA OUT)
2 1
Selection
Disable auto-exporting (default) Enable auto-exporting
4 Return to the Measurement screen.
The confirmation screen appears.
Return to the setting screen.
Save setting and return to previous screen.
Discard setting and return to previous screen.
78
6.4 Auto-Exporting Measured Values (at End of Measurement) (Data Output
Function)
Printing
79
7.1 Connecting the Printer
Chapter 7
Connecting the printer to the instrument
Make instrument settings (p. 81)
Make printer settings
Printing (p. 82) · Measurement values and
comparator decisions · Statistical calculation
results
7.1 Connecting the Printer
Before connecting the printer
Because electric shock and instrument damage hazards are present, always
follow the steps below when connecting the printer. · Always turn off the
instrument and the printer before connecting.
7 · A serious hazard can occur if a wire becomes dislocated and contacts
another conductor during operation. Make certain connections are secure.
· As much as possible, avoid printing in hot and humid environments.
Otherwise, printer life may be severely shortened.
· Use only compatible recording paper in the printer. Using non-specified
paper may not only result in faulty printing, but printing may become
impossible.
· If the recording paper is skewed on the roller, paper jams may result.
Compatible printer
The requirements for a printer to be connected to the instrument are as
follows. Confirm compatibility and make the appropriate settings on the
printer before connecting it to the instrument. See: “7.2 Instrument Settings”
(p. 81)
· Interface ………………………….. RS-232C · Characters per line ……………. At least 45 ·
Communication speed……….. 9600 bps · Data bits ………………………….. 8 · Parity
………………………………. none · Stop bits…………………………… 1 · Flow control ……………………… none
80
7.1 Connecting the Printer
Connecting the Printer to the Instrument
Connection Methods
Printer (Example)
2 3
AC adapter RS-232C Cable
1 Confirm that the instrument and the printer
are turned off.
2 Connect the AC adapter to the printer, and
insert the power plug into an outlet.
3 Connect the RS-232C cable to the RS-232C
connectors on the instrument and printer.
4 Turn the instrument and printer on.
Connector Pinouts
1 2 3 4 5
6 7 8 9
RM3542 (9-pin) Connector
Function
Signal Name
Pin
Receive Data
RxD
2
Transmit Data
TxD
3
Signal or Common Ground GND
5
13 ………………….. 1
25 ………………….. 14
9670 Printer (25-pin) Connector (Example)
Pin
Signal Name
2
TxD
3
RxD
7
GND
4
RTS
5
CTS
Function
Transmit Data Receive Data Signal or Common Ground Request to Send Clear to
Send
7.2 Instrument Settings
1 Open the Basic Settings screen.
81
7.2 Instrument Settings
The Basic Settings screen appears.
2 Open the System screen.
1 Selection
2
The System screen appears. [SYSTEM]
3 Select PRINT as the interface type.
2
1
Selection
4 Return to the Measurement screen.
The confirmation screen appears.
To use the printer
7
Return to the setting screen. Save setting and return to previous screen.
Discard setting and return to previous screen.
82
7.3 Printing
7.3 Printing
Before Printing
Verify that the instrument and printer settings (p. 81) are correct.
Printing Measured Values and Comparator Decisions Printing by key operation
Press the PRINT key to print the measured value currently displayed on the
Measurement screen.
Printing by external control
Measured values and comparator decisions print when the (active-low) PRINT
signal (EXT. I/O connector) is connected to ISO_GND. ISO_GND is a pin in
the instrument’s EXT. I/O connector.
When statistical calculation is enabled [STATISTIC: ON] and internal
triggering [TRG: INT] is selected, statistical calculations are performed and
measured values are printed. When external (EXT) triggering is selected, only
measured values are printed. Use the TRIG signal to perform statistical
calculations with external triggering.
Printing Statistical Calculation Results
Statistical calculation results can be printed when auto-memory or statistical
calculation is enabled (ON). To print, select PRINT on the screen or connect
the (active-low) PRINT signal on the EXT. I/O connector to ISO_GND.
To enable auto-memory: See: “6.2 Store as soon as Measurement is Stable (Auto-
Memory Function)” (p. 71) To enable statistical calculation: See: “6.3
Performing Statistical Calculations on Measured Values” (p. 74) (When
statistical calculation is enabled)
If no valid data exists, only the data count is printed. When only one valid
data sample exists, standard deviation of sample and process capability
indices cannot be printed.
Example Printouts
Resistance measurements
1 0.8725mOhm Lo
2
0.484mOhm Lo
3 10.99998 Ohm IN
4 -10.0026 Ohm Lo
27 9.9986 Ohm Hi
28
9.996 Ohm Hi
29 0.01003kOhm Hi
30 0.00012MOhm Hi
Measurement fault values
1 OvrRng
Hi
2 -OvrRng
Lo
3 C.E.Hi
—
4 C.E.Lo
—
5 C.E.Volt
—
6 ——–
—
83
7.3 Printing
Auto-memory data and statistical calculation results
Date: 09-02-01 Time: 06:18:00
Ref: 1000.000 Ohm Upp: +1.000% Low: -1.500%
1 999.885 Ohm -0.011% IN
2 1001.885 Ohm +0.189% IN
3 1002.394 Ohm +0.239% IN
4 1002.892 Ohm +0.289% IN
5 1012.894 Ohm +1.289% Hi
6 1000.897 Ohm +0.090% IN
7 998.902 Ohm -0.110% IN
8 994.888 Ohm -0.511% IN
9 1000.391 Ohm +0.039% IN
10 979.892 Ohm -2.011% Lo
Hi: 1 IN: 8 Lo: 1 OR: 0
Number: 10 Valid: 10
Max 1012.894 Ohm +1.289% ( 5)
Min 979.892 Ohm -2.011% ( 10)
Avg 999.492 Ohm -0.051%
Sn
7.83568 Ohm
Sn-1 8.25953 Ohm
Cp
0.50
CpK
0.42
· The “Valid” statistical calculation result indicates the number (count) of
data samples not subject to errors such as measurement faults.
· Among the comparator decision result counts (Hi, IN, Lo, and OR), “OR”
indicates the number (count) of out-of-range measurements.
7
84
7.3 Printing
85
8.1 External Input/Output Connector and Signals
External Control Chapter 8
The EXT. I/O connector on the rear of the instrument supports external control
by providing output of the end-of-measurement and comparator decision signals,
and accepting input of measurement trigger and key-lock signals. All signals
are isolated by optocouplers (inputs and outputs share a common signal
ground).
Confirm input and output ratings, understand the safety precautions for
connecting a control system, and use accordingly.
Connect the instrument’s EXT. I/O connector to the signal output or input device.
Make instrument settings (p. 93)
Signal input/output
8.1 External Input/Output Connector and Signals
8 To avoid electric shock or damage to the equipment, always observe the
following
precautions when connecting to the EXT. I/O terminals. · Always turn off the
power to the instrument and to any devices to be connected
before making connections.
· During operation, a wire becoming dislocated and contacting another
conductive object can be serious hazard. Make sure that connections are secure
and use screws to secure the external connectors.
· Ensure that devices and systems to be connected to the EXT. I/O terminals
are properly isolated.
To avoid damage to the instrument, observe the following cautions: · Do not
apply voltage or current to the EXT. I/O terminals that exceeds their ratings.
· When driving relays, be sure to install diodes to absorb counter-
electromotive force.
· Be careful not to short-circuit ISO_5V to ISO_COM. · Be careful not to
short-circuit ISO_12V to ISO_COM.
See: “Connector Type and Signal Pinouts” (p. 86)
86
8.1 External Input/Output Connector and Signals
Connector Type and Signal Pinouts
TRIG (Reserved) KEY_LOCK (Reserved) (Reserved)
0ADJ HOLD ISO_5V ISO_COM
ERR HI LO
CE_HI (Reserved) (Reserved) (Reserved) (Reserved) (Reserved) (Reserved)
Connector: (Instrument Side) 37-pin D-sub female with #4-40 screws
19 18 17 16 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 37 36 35 34 33 32 31 30 29 28 27 26 25 24 23 22 21 20
(Reserved) CAL (Reserved) (Reserved) PRB_CHECK (Reserved) PRINT ISO_COM EOM INDEX IN PRB_SHORT CE_LO (Reserved) (Reserved) (Reserved) (Reserved) ISO_12V
Mating Connectors: DC-37P-ULR (solder type) / DCSP-JB37PR (pressure weld type) Japan Aviation Electronics Industry Ltd.
EXT. I/O Connector (Instrument Side) Pos: positive, Neg: negative, -: not applicable
Pin Signal name I/O
Function
1 TRIG
IN External trigger
2 (Reserved)
—
Logic
Pos/ Neg
Edge
—
3 KEY_LOCK 4 (Reserved) 5 (Reserved)
IN Key-Lock — —
Neg Level
— —
6 0ADJ 7 HOLD 8 ISO_5V 9 ISO_COM
IN
Execute zero-adjust
Neg Edge
IN
Enable external triggering
Neg Level
–
Isolated 5 V power output
–
–
–
Isolated common signal ground
–
–
10 ERR
OUT Measurement fault Neg Level
11 HI
12 LO
13 CE_HI 14 (Reserved) 15 (Reserved) 16 (Reserved) 17 (Reserved) 18 (Reserved)
19 (Reserved)
OUT
HI comparator decision
OUT
LO comparator decision
OUT
Probe (HI side) contact error
—
—
—
—
—
—
Neg Level
Neg Level
Neg Level
— — — — — —
Pin Signal name 20 (Reserved)
I/O —
Function
Logic —
21 CAL 22 (Reserved) 23 (Reserved)
IN
Execute self-calibration
Neg
Edge
—
—
—
—
24 PRB_CHECK
IN
Execute probe shortcircuit detection
Neg
Edge
25 (Reserved)
—
—
26 PRINT 27 ISO_COM
IN
Print measured value
Neg Edge
–
Isolated common signal ground
—
28 EOM
OUT End of measurement Neg Edge
29 INDEX
OUT
Analog measurement finished
Neg Edge
30 IN
OUT
IN comparator decision
Neg Level
31
PRB_SHORT
OUT
Probe short-circuit error
Neg Level
32 CE_LO
33 (Reserved) 34 (Reserved) 35 (Reserved) 36 (Reserved)
OUT
Probe (LO side) contact error
—
—
—
—
Neg Level
— — — —
37 ISO_12V
–
Isolated 12V power output
–
–
Reserved pins are not connected inside the instrument. Do not connect to reserved pins.
· The 0ADJ signal should be asserted (Low) for at least 10 ms. · The connector
shell is conductively connected to the metal instrument chassis and the
protective earth pin
of the power plug. Be aware that it is not isolated from ground.
87
8.1 External Input/Output Connector and Signals
Signal Descriptions
Input Signals
TRIG
When external triggering (EXT) is enabled, one measurement is performed at the falling (ON) or rising (OFF) edge of the TRIG signal. Falling (ON) or rising (OFF) edge triggering can be selected on the Settings screen (default: falling (ON) edge). When internal triggering (INT) is enabled, external triggering is disabled. Also, when the Settings Monitor function is enabled and an error occurs, triggering is disabled (p. 53). The TRIG signal performs the following operations in addition external triggering: · Stores statistical calculation data (when statistical calculation is enabled) · Stores measured data to internal memory (when the data memory function is enabled) (also operates with internal triggering)
(p. 94)
0ADJ
Asserting the 0ADJ signal executes zero adjustment once. To avoid malfunction, this signal should be asserted (Low) for at least 10 ms.
(p. 32)
PRINT Asserting the PRINT signal prints the current measurement value.
(p. 82)
CAL
Asserting the CAL signal executes self calibration. The time required for self calibration is as follows: Approximately 130 ms (with 60-Hz line frequency setting), or 110 ms (with 50-Hz setting) If asserted during measurement, executes after the end of measurement.
(p. 57)
HOLD
Holding the HOLD signal low enables external triggering. When the HOLD signal is high, the settings made on the Settings screen or by commands are re- enabled.
PRB_CHECK
Asserting the PRB_CHECK signal executes probe short-circuit adjustment one time. If asserted during measurement, executes after the specified time from the end of measurement.
(p. 51)
KEY_LOCK
While the KEY_LOCK signal is held low, all front panel keys (except POWER button) are disabled (key unlock and remote control cancellation operations are also disabled).
(p. 59)
Output Signals
ERR
This signal indicates that a measurement fault has occurred (except out-of- range detection). It is updated simultaneously with the EOM signal. At this time, comparator decision outputs are all de-asserted (high).
8
(p. 38)
CE_HI
This signal indicates that a contact check error has occurred between HCUR and HPOT contacts. It is updated simultaneously with the EOM signal. At this time, comparator decision outputs are all de-asserted (high).
(p. 46)
CE_LO
This signal indicates that a contact check error has occurred between LCUR and LPOT contacts. It is updated simultaneously with the EOM signal. At this time, comparator decision outputs are all de-asserted (high).
(p. 46)
PRB_SHORT
This signal indicates that a foreign object is shorting the POT and CUR contacts in a four-terminal probe tip. At this time, comparator decision outputs are all de-asserted (high).
(p. 51)
INDEX
This signal indicates that A/D conversion in the measurement circuit is finished. When the asserted (low) state occurs, the measurement sample can be removed.
EOM
This signal indicates the end of a measurement. At this time, the states of the comparator decision outputs and ERR, CE_HI, CE_LO, and PRB_SHORT are all determined.
(p. 93)
HI, IN, LO These are the comparator decision output signals.
· Input signals are ignored when the following are displayed: Basic, Detailed,
and Comparator Settings screens; Statistical Calculation Results screen
(except for the print signal); and error messages (except Setting Monitor
errors).
· EXT. I/O input and output signals are not usable while changing measurement
settings.
88
8.2 Timing Chart
8.2 Timing Chart
Each signal level indicates a corresponding voltage level.
Contact Improver
Measurement currents
Contact check
Measurement processing
Checking
Available
· The EOM signal is operational when the trigger source setting is EXT and the
EOM output setting is Pulse. · A self-calibration measurement (approximately
130 ms) is automatically performed every 10 minutes
(between measurements). TRIG signal input is accepted during that time, but
the corresponding measurement is delayed until self-calibration is finished. ·
Do not apply a TRIG signal while measuring using external triggering (the
signal is ignored). · When changing settings such as measurement range, allow
about 150 ms processing time before applying a TRIG signal. · Input signals
are ignored when the following are dis
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>