DEO DENTAL Multi Purpose Heater
Product Information
Specifications:
- Product Name: Multi-Purpose Heater
- Power Supply: 120-volt circuit
- Grounding: Equipped with grounding wire and plug
- Warranty: 2 years for defects in material or workmanship
- Safety Standards: CE-CB, EMC, LVD, RoHS, REACH, cUL ISO 9001-2015 Certified
Product Usage Instructions
Programme Selection:
Select the desired programme on the heater. For example, P3 for green
stick.
Wax Heating Process:
- Open the cassette and place a 4cm/2 inch wide strip wax inside.Close the cover and wait for approximately ten seconds.
- Do not place the cassette back in the heater with the sheet wax. Open the cassette to find the sheet wax ready to use.
- Remove the 4cm/2 inch sheet wax for application. Return the cassette to the rectangular opening for reheating.
Temperature Adjustment:
Adjust the display to your preferred temperature by pressing the Plus or
Minus buttons on the control panel. The set temperature will flash when
adjusted and maintain the desired heat level.
New Temperature Setting:
If selecting a new temperature, it will flash until the new setting is
reached. After 10 minutes, the display will show the new selected temperature.
FAQ (Frequently Asked Questions):
-
Q: How should I ensure safety while using the heater?
A: Avoid draping the cord where it can be pulled by children or tripped over. Ensure proper grounding as per safety instructions provided in the manual. -
Q: What is covered under the warranty?
A: The warranty covers defects in material or workmanship for two years from the date of purchase. It does not affect your legal rights as a consumer.
P1 – Select this programme for Hot Plate to trim the wax for occlusal registration rims
Instruction for use of the Hot Plate
-
Place the silicone catch dish in the chamber.
-
Place the specially designed circular 100% Aluminium melting plate on the chamber supplied.
The excess molten wax will be collected in the silicone catch dish in the chamber. -
Ensure the unit is mounted on the stand and is sloping away from the user and the hot plate handle is facing rearwards. ( do not use any other chamber )
-
Switch on the heater.
-
Select P1 – FH 15 Minutes – 125°C.
This program will display 125 and within a few seconds will start flashing FH and P1 (This is just a reminder of the programme selected)
FH will continue to flash for 15 minutes. After 15 minutes the display will show 125 and will be solid.
This is the time it takes to increase the temperature of the Aluminium plate to trim the wax for occlusal registration rims The heater will now maintain the temperature for as long as it is switched on.
From time to time, 125 will flash. This indicates that the heater is heating up.
As soon as it reaches temperature it will be solid again. -
Adjust the display to the user’s preferred temperature. The temperature can be changed by pressing the Plus or Minus on the control panel.
From time to time the adjusted set temperature will flash. This indicates that the heater is maintaining the set temperature.
Please go to the website to see video www.deo-dental.com
Cleaning
- Cleaning is best carried out when the heater is warm.
- The hot plate will melt any residual wax quickly and it will flow into the silicone catch dish.
- Any residual wax can be wiped off with a tissue. Take care as the plate will be hot. Use the handle to remove.
- The top plate, silicone catch dish and aluminium chamber can then be passed through a dental washer/disinfector machine at 90 – 95°C for a minimum of 1 minute with a chemical PH of 105 on a cycle for any clinical hand instrument.
- After cleaning the hot plate, aluminium chamber and silicone catch dish, they can be placed for autoclaving following the normal protocol for any dental instruments.
- 134 -137°C. 2.1 -2.25 bar gauge pressure for at least 3 minutes holding time.
- Casing, cover and angle stand are all disinfected using clinical wipes as would be used on all hard surfaces in a clinical environment.
- P2 – Select this program for Water bath
Instructions to use the Water Bath
- Fill the drop in water chamber with warm water from the tap (not exceeding 60°C) to the max level marked.
- Put the stainless steel basket in the chamber.
- Put the water chamber and basket in the main heater.
- Put the cover on and insert the temperature probe.
- Switch on the heater
- Select P2 – FH 10 Minutes – 65°C
This programme will display 65 and within a few seconds will start flashing FH
and P2 (This is just a reminder of the programme selected).
FH will continue to flash for 10 minutes. After 10 minutes the display will
show 65 and will be solid. Adjust the display to the manufacturer’s
recommended temperature for the brand of the impression compound or user
preferences.
The temperature can be changed by pressing the Plus or Minus on the control
panel. The heater will now maintain the temperature for as long as it is
switched on.
From time to time the adjusted set temperature will flash. This indicates that
the heater is maintaining the set temperature. As soon as it reach temperature
it will be solid again.
Please go to the website to see the video www.deo-dental.com
Cleaning
Cleaning is best performed when the heater is warm. Take care as the chamber
will be hot.
The aluminium chamber, basket and silicone ring can then be passed through a
dental washer/disinfector machine. 90-95°C for a minimum of 1 minute with a
chemical PH of 105 on a cycle for any clinical hand instrument.
After the cleaning process, the water chamber, basket and silicone ring can be
placed for autoclaving following the normal protocol for any dental
instruments.
134 -137°C. 2.1 -2.25 bar gauge pressure for at least 3 minutes holding time.
Casing, cover and thermostat are all disinfected using clinical wipes as would
be used on all hard surfaces in a clinical environment.
P3 – Select this programme for green stick and sheet wax
Instructions to use the Green / tracing sticks and sheet modelling wax
(4 Syringes Included)
-
Place the silicone catch dish in the drop in chamber.
-
Place the specially designed circular Green stick and sheet wax plate on the chamber supplied.
-
Any excess wax will be collected in the silicone catch dish in the chamber. (do not use any other chamber or silicone ring)
-
Place the two syringes with the end tip into the two circular holes and put one green stick in each syringe.
-
Place the two empty cassettes in the rectangular opening. Switch on the heater.
-
Select P3 – FH 17 Minutes – 125°C
This programme will display 125 and within a few seconds will start flashing FH and P3 ( This is just a reminder of the programme selected)
FH will continue to flash for 17 minutes. After 17 minutes the display will show 125 and will be solid.
This is the time it takes to increase the temperature of the chamber to heat the sheet wax and melt the green stick in the syringes. -
Adjust the display to the users preferred temperature. The temperature can be changed by pressing the Plus or Minus on the control panel.
From time to time the adjusted set temperature will flash. This indicates that
the heater is maintaining the set temperature. After 17 minutes the Green
stick will be in a molten form and can be extruded. The display will show 125.
Remove one cassette and place on a flat surface.
-
Open the cassette and place a 4cm/ 2 inch wide strip wax. Close the cover and wait for ten seconds approximately.
DO NOT PLACE THE CASSETTE BACK IN HEATER WITH SHEET WAX -
Open the cassette and the sheet wax will be ready to use. Remove the 4cm/2 inch sheet wax and return the cassette to the rectangular opening to reheat.
Please go to the website to see video www.deo-dental.com for a demonstration. -
Adjust the display to the user preferred temperature The temperature can be changed by pressing the Plus or Minus on the control panel.
From time to time the adjusted set temperature will flash. This indicates that the heater is maintaining the set temperature.
Cleaning
Cleaning is best performed when heater is warm. the chamber will be hot. Use
the handle to remove the polycarbonate green stick holder.
The silicon catch dish and aluminium chamber and silicone ring can then be
passed through a dental washer/disinfector machine. 90-95°C for a minimum of 1
minute with a chemical PH of 105 on a cycle for any clinical hand instrument.
After the cleaning process, the chamber and silicone catch dish can be placed
for autoclaving following the normal protocol for any dental instruments. 134
-137°C. 2.1 -2.25 bar gauge pressure for at least 3 minutes holding time.
Casing and cover are all disinfected using clinical wipes as would be used on
all hard surfaces in a clinical environment.
Instructions for Intra/Oral Photographic Mirrors
Instructions to use the Mirror
-
Place the module in the drop in chamber and place in the heater
-
Switch on the heater and select P4 – 100°C
This programme will display 100°C (max set temperature) and within a few seconds it will start flashing P4 and 100°C (This is just a reminder of the programme selected)
Using the minus sign reduce the display temperature to the user preference -
Place the Mirror module to be used in the heater and insert two mirrors into the top openings slot for 10 to 15 minutes before the start of the photography
-
P4 and new selected temperature will continue to flash until the New temperature is reached After 10 minutes the display will show the New selected temperature
Recommended Temperature:
- Intra / Oral – Photographic Mirrors 40°C
- The temperature can be changed by pressing the Plus or Minus on the control panel
- From time to time the adjusted set temperature will flash . This indicates that the heater is maintaining the display temperature.
Cleaning
Cleaning is best performed when the heater is warm. The chamber will be hot.
Use the handle to remove the polycarbonate holder.
The mirror holders, and the aluminium chamber can then be passed through a
dental washer/disinfector machine. 90-95°C for a minimum of 1 minute with a
chemical PH of 105 on a cycle for any clinical hand instrument.
After the cleaning process, the chamber and silicone can be placed for
autoclaving following the normal protocol for any dental instrument..
P4 Multi – purpose heater for composite compule Instructions to use Composites, LA and Endodontic Solution Module
-
Place the module in the drop in chamber and place in the heater
-
Switch on the heater and select P4 – 100°C
This programme will display 100 (max set temperature) and within a few seconds it will start flashing P4 and 100 (This is just a reminder of the programme selected) -
Using the minus sign reduce the display temperature for the material of choice to be used
-
Place material to be used in the appropriate opening. Compules and local anaesthetics on the underside and syringes into the top openings slot for 10 -15 minutes before the start of the treatment
P4 and the new selected temperature will continue to flash until the New temperature is reached After 10 minutes the display will show the New selected temperature
Recommended Temperature:
- 40°C – LA
- 75°C – Composite
- 60°C – Endodontics Solutions
The temperature can be changed by pressing the Plus or Minus on the control
panel
From time to time the adjusted set temperature will flash This indicates that
the heater is maintaining the display temperature.
Please go to the website to see video www.deo-dental.com
Cleaning
- Cleaning is best performed when the heater is warm.
- The chamber will be hot. Use the handle to remove the polycarbonate holder.
- The mirror holders, composite compules. local anaesthetic and the aluminium chamber can then be passed through a dental washer/disinfector machine. 90-95°C for a minimum of 1 minute with a chemical PH of 105 on a cycle for any clinical hand instrument.
- After the cleaning process, the chamber and silicone can be placed for autoclaving following the normal protocol for any dental instrument. 134 -137°C. 2.1 -2.25 bar gauge pressure for at least 3 minutes holding time .
- Casing, cover are all disinfected using clinical wipes as would be used on all hard surfaces in a clinical environment
IMPORTANT SAFETY INSTRUCTIONS
When using electrical appliances, especially when children are present, safety precautions must always be followed, including the following:
READ ALL INSTRUCTIONS BEFORE USING
DANGER – To reduce the risk of electrocution:
-
Always unplug this appliance immediately after using.
-
Do not use while bathing.
-
Do not place or store appliance where it can fall or be pulled into a tub or sink.
-
Do not place in or drop into water or other liquid.
-
Do not reach for an appliance that has fallen into water. Unplug immediately.
WARNING : To reduce the risk of burns, electrocution, fire, or injury to persons: -
An appliance should never be left unattended when plugged in.
-
Close supervision is necessary when this appliance is used by, on, or near children or individuals with certain disabilities.
-
Use this appliance only for its intended use as described in this manual. Do not use attachments not recommended by the manufacturer.
-
Never operate this appliance if it has a damaged cord or plug, if it is not working properly, or if it has been dropped, damaged, or dropped into water. Return the appliance to a service centre for examination and repair.
-
Keep the cord away from heated surfaces. Do not wrap the cord around the appliance.
-
Never use while sleeping.
-
Never drop or insert any object into any opening or hose.
-
Do not use outdoors or operate where aerosol (spray) products are being used or where oxygen is being administered.
-
Connect this appliance to a properly grounded outlet only. See Grounding Instructions.
-
Do not use an extension cord with this appliance..
-
Do not touch hot surfaces. Use handles or knobs
-
Use this appliance on a level surface.
-
Do not let cord hang over edge of table or counter where it can be pulled on by children or tripped over unintentionally.
-
Unplug from outlet when not in use. Allow to cool before handling, cleaning, or storing the appliance.
-
The physiological effects of this appliance have not been investigated by UL.
-
Only use the appliance as described in the operating instructions.
-
Burns can occur from touching hot parts.
-
Always remove the mains supply at night or during long periods of inactivity.
-
Never immerse the appliance, power cord or plug in water or any other liquid.
CAUTION : To reduce the risk of electric shock, do not immerse in water or other liquids.
CAUTION : To reduce the risk of burns or injury – Do not drape the cord over a counter or a tabletop such that it can be pulled on by children or tripped over unintentionally.
SAFETY INSTRUCTIONS FOR HOUSEHOLD
TYPE – GROUNDING INSTRUCTIONS
- This appliance must be grounded. In the event of an electric short-circuit, grounding reduces the risk of electric shock by providing a path of low resistance for the electric current. This appliance is equipped with a cord having a grounding wire with a grounding plug. The plug must be plugged into an outlet that is properly installed and grounded in compliance with all local codes and ordinances.
- DANGER : Improper use of the grounding plug can result in a risk of electric shock. Check with a qualified electrician or service personnel if you are in doubt as to whether the appliance is properly grounded. Do not modify the plug provided with the appliance – if it will not fit the outlet, have a proper outlet installed by a qualified electrician. If repair or replacement of the cord or plug is necessary, do not connect the grounding wire to either flat blade terminal. The wire with insulation having an outer surface that is green with or without yellow stripes is the grounding wire.
- This appliance is for use on a nominal 120-volt circuit and has a grounding plug that looks like the plug illustrated in sketch A. A temporary adapter,
which looks like the adapter illustrated in sketches B and C, may be used to connect this plug to a 2-pole receptacle as shown in sketch B if a properly grounded outlet is not available. The temporary adapter should be used only until a properly grounded outlet (sketch A) can be installed by a qualified electrician. The green-coloured rigid ear, lug, or the like extending from the adapter must be connected to a permanent ground such as a properly grounded outlet box cover. Whenever the adapter is used, it must be held in place by a metal screw.
CLEANING AND MAINTENANCE
- To ensure efficient operation, this appliance should be kept clean.
- Never attempt to clean the appliance before switching off, removing the mains supply and ensuring it has cooled.
- Always dry the appliance thoroughly before attempting to switch on.
- Maintenance should only be carried out by the manufacturer or a qualified person. The delicate microelectronics used within the circuitry can be damaged if removed from the casing without the correct antistatic precautions.
WARRANTY
This warranty is offered as an extra benefit and does not affect your legal
rights. This product has been manufactured and tested to an international
quality standard. The manufacturer undertakes to replace or repair, for a
period of two years, any defect in material or workmanship.
This warranty is limited to the original purchaser.
Limitations to warranty
The warranty may be invalid if:
- The appliance has been incorrectly installed or installed to a supply differing from those described in this manual.
- An attempted repair has been made by a non-qualified person.
- The appliance has not been modified in any way.
- Defects has arisen from misuse or neglect. Register your appliance with the manufacturer at the: www.deo-dental.com
Please provide the date and proof of purchase for all warranty claims.
Tel: +44 (0)1245 354800
Web – www.deo-dental.com | Email – hello
@deo-dental.com
https://www.instagram.com/deodental.uk/
https://www.youtube.com/channel/UCQwonVT8M3tgznYmHeN1EZQ
https://www.facebook.com/profile.php?id=61551248384743 Widford Trading
Estate, 36 / 38, Robjohns Rd, Chelmsford CM1 3AG
References
- Dental.com | Dental Photo Assessments Powered by AI
- Jai Ragbir/Mike Gregory Dentures (@deodental.uk) • Instagram photos and videos
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>