stryker 5100-015-250 MD Series Medium Straight Attachment Instruction Manual
- June 3, 2024
- stryker
Table of Contents
5100-015-250 MD Series Medium Straight Attachment
Instruction Manuam
MD Series Attachments
5100-015-250 5100-015-252 5100-015-270 5100-015-272
Instructions For Use
Introduction
This Instructions For Use manual is the most comprehensive source of
information for the safe and effective use of your product. This manual may be
used by in-service trainers, physicians, nurses, surgical technologists,
biomedical equipment technicians, and central supply/sterile processing
technicians. Keep and consult this reference manual during the life of the
product.
The following conventions are used in this manual:
- A WARNING highlights a safety-related issue. ALWAYS comply with this information to prevent patient and/or healthcare staff injury.
- A CAUTION highlights a product reliability issue. ALWAYS comply with this information to prevent product damage.
- A NOTE supplements and/or clarifies procedural information.
For additional information, especially safety information, or in-service
training, contact your Stryker sales representative or call Stryker customer
service. Outside the US, contact your nearest Stryker subsidiary.
NOTE: The user and/or patient should report any serious product-related
incident to both the manufacturer and the Competent Authority of the European
Member State where the user and/or patient is established.
Indications For Use
See the instructions for the use supplied with the handpiece.
Contraindications
See the instructions for the use supplied with the handpiece.
For Use With
The MD Series Attachments are intended for use with the following Stryker
handpieces:
DESCRIPTION | REF |
---|---|
RemB® Electric Micro Drill | 6400-015-000 |
Consolidated Operating Room Equipment (CORETM) Micro Drill | 5400-015-000 |
Total Performance System (TPSTM) Micro Drill | 5100-015-000 |
User/Patient Safety
WARNINGS:
- Only trained and experienced healthcare professionals should use this equipment.
- Before using any system component or any component compatible with this system, read and understand the instructions. Pay particular attention to WARNING information. Become familiar with system components prior to use.
- Upon initial receipt and before each use, clean and sterilize the equipment as indicated. See the care instructions manual supplied with the handpiece.
- Upon initial receipt and before each use, operate the equipment and inspect each component for damage. DO NOT use any equipment if the damage is apparent or the inspection criteria are not met. See the care instructions manual supplied with the handpiece for inspection criteria.
- When operating the handpiece, DO NOT actuate the attachment collar. Actuation may suddenly stop and/or overheat the handpiece.
- DO NOT attempt to insert or remove any cutting accessory or attachment while the handpiece is operating.
- DO NOT apply excessive pressure. Monitor heavy sideloads and/or long operating periods to prevent overheating of the distal tip and body of the handpiece and/or attachment. Excessive pressure may bend or fracture the cutting accessory.
- ALWAYS operate the equipment within the specified environmental condition values. See the Specifications section.
- ALWAYS follow the recommended duty cycle to prevent the equipment from overheating. See the Specifications section.
- Use adequate irrigation during cutting to prevent heat generation.
- When in use, attachments are applied parts. See the instructions for use supplied with the handpiece for information specific to applied parts.
Cutting Accessories
WARNINGS:
- Use only Stryker-approved system components and accessories unless otherwise specified.
- DO NOT modify any system component or accessory unless otherwise specified.
NOTES:
- A variety of specialized attachments are available for use with the Micro Drill. Each specialized attachment is supplied with specific instructions.
- For a complete list of accessories, contact your Stryker sales representative or call Stryker customer service. Outside the US, contact your nearest Stryker subsidiary.
The following Stryker-approved accessories are sold separately:
DESCRIPTION | REF |
---|---|
E-Type Attachment Bur Guard | 5100-015-301 |
Cutting Accessories
WARNINGS:
- Upon initial receipt and before use, visually inspect the package for damage to confirm the integrity of the sterile barrier. Do not use the product if the damage is apparent, the sterile barrier is compromised, or the package is unintentionally opened.
- All cutting accessories are intended for single use only. Reuse significantly increases wear on the handpiece and attachment.
- DO NOT reuse, reprocess, or repackage a single-use device. A single-use device is intended for single-use only. A single-use device may not withstand chemical, chemical vapor, or high-temperature sterilization reprocessing. Design features may make cleaning difficult. Reuse may create a serious risk of contamination and may compromise the structural integrity of the single-use device resulting in operational failure. Critical product information may be lost if the single-use device is repackaged. Failure to comply may lead to infection or cross-infection and result in patient and/or healthcare staff injury.
- DO NOT use a cutting accessory with an overall length greater than 2.25 inches [5.72 cm] with the Micro Drill attachments described in this instruction booklet.
NOTE: Sterile cutting accessories are sterilized by irradiation.
DESCRIPTION | REF |
---|---|
5.0 mm Egg Bur | 0277-010-050 |
3.0 mm Matchhead Bur, Extended | 0277-010-208 |
Carbide Burs | 0277-010-221 |
0277-010-235
0277-010-421
5300-010-421
5300-010-50X series
5300-010-51X series
5300-010-70X series
5300-010-71X series
5300-010-90X series
6.1 mm Pear Bur| 1608-002-005
Metal Cutting Diamond Disc| 1608-006-099S4
Toe Fusion Sets| 1608-010-100SX series
Resurfacing Tools| 1900-010-040
1900-015-040
1900-015-025
Drill Bits, Medium| 5300-010-041
5300-010-059
5300-010-063
5300-010-071
Swanson Pilot Point Burs, Medium| 5300-010-13X series
Round Diamond Burs, Medium| 5300-010-303
5300-010-307
5300-010-311
Dermabraders| 5300-010-911
5300-010-94X series
If the following symbol is on the device label, the device meets the stated
criteria:
This device contains the following substance(s) defined as CMR 1B in a
concentration above 0.1% weight by weight: Cobalt; CAS No. 7440-48-4; EC No.
231-158-0. Current scientific evidence supports the position that medical
devices manufactured from cobalt alloys or stainless steel alloys containing
cobalt do not cause an increased risk of cancer or adverse reproductive
effects.
Definitions
The symbols located on the equipment and/or labeling are defined in this section or in the Symbol Definition Chart. See the Symbol Definition Chart supplied with the equipment.
SYMBOL | DEFINITION |
---|---|
RUN | Operate Position |
LOAD | Insert/Remove Position |
General warning sign |
Instructions
NOTE: The Micro Drill handpiece may be operated with the Stryker
footswitch or hand switch.
Assembly
W ARNING: Ensure the handpiece is off. DO NOT attach any cutting
accessories or attachments while the handpiece is running. 1. Install the
attachment onto the handpiece with a twisting motion until it securely snaps
into place. See figure 1.
Figure 1. Attachment Installation
NOTE: Ensure the attachment is secure before running the Micro Drill.
Observe the small gap between the attachment and the Micro Drill housing
caused by the anti-rotation pin. This gap will not hinder the function of the
Micro Drill or the attachment. The gap should not be wider than 0.025 inches
[0.65 mm].
2. Insert the cutting accessory as follows:
2. 1. Rotate the lock collar to the LOAD position. See figure 2.
2. 2. Insert the cutting accessory into the attachment with a twisting motion
until it bottoms.
Figure 2. Cutting Accessory Installation
2. 3. Rotate the lock collar to the RUN position. See figure 3.
Figure 3. Cutting Accessory Locked Into Place
2. 4. Gently tug on the cutting accessory to ensure that it is locked into
place.
WARNING: Ensure the cutting accessory is correctly installed. A burn
injury may result if the accessory is incorrectly installed and friction
occurs between the accessory and the attachment.
Disassembly
-
Rotate the lock collar to the LOAD position. See figure 4.
-
Pull the cutting accessory straight out of the handpiece.
Figure 4. Cutting Accessory Removal -
Grasp the attachment and pull it away from the handpiece.
Care Instructions
For processing instructions and disposal/recycling information, see the care instructions manual supplied with the handpiece.
Troubleshooting
WARNING: DO NOT disassemble or service this equipment, unless otherwise
specified.
NOTE: For service, contact your Stryker sales representative or call
Stryker customer service. Outside the US, contact your nearest Stryker
subsidiary.
PROBLEM | CAUSE | SOLUTION |
---|---|---|
The attachment is difficult to install or remove from the handpiece. | The | |
connector requires a break-in period. | Install and remove the attachment |
repeatedly until the connector functions
smoothly.
Attachment will not retain the cutting
accessory.| The lock collar is not fully engaged.| Rotate the collar to the
LOAD position. Reinstall the cutting accessory. Fully
rotate the collar to the RUN position. Gently tug on the cutting accessory to
ensure it is secure.
The handpiece will not run.| The lock collar is not fully engaged.| Fully
rotate the collar to the RUN position.
Cutting accessories will not disengage from the attachment.| The lock collar
is not fully disengaged.| Fully rotate the
collar to the LOAD position.
The cutting accessory seizes in the attachment.| The lock collar is not fully
engaged.| Rotate the collar to the LOAD position. Reinstall the cutting
accessory. Fully rotate the collar clockwise to the RUN position.
Gently tug on the cutting accessory to ensure it is secure.
The attachment is not properly installed into the handpiece.| Reinstall the
attachment. Ensure the connection is flush to the handpiece.
Attachment runs at a gradually higher temperature over its operating life.|
The attachment is beginning to wear.| Return the attachment to Stryker for
evaluation
and possible replacement.
Attachment heats up rapidly during use.| Debris is trapped inside the
attachment.| Clean and sterilize the attachment.
The internal components of the attachment are damaged.| Return the attachment
to Stryker for evaluation
and possible replacement.
The attachment contains moisture and leaks fluid during use.| The attachment
was improperly cleaned and sterilized.| Clean and sterilize the attachment.
Comply with Dry Cycle requirements.
Specifications
WARNING: ALWAYS check any documentation that accompanies attachments and/or cutting accessories for special duty cycle and usage instructions.
Model:| MD Series Straight Attachment
(REF 5100-015-250)
---|---
Dimensions:| 3.4 inch [87.2 mm] length
0.8 inch [19.6 mm] diameter
Mass:| 0.15 lb [69.3 g]
Model:| MD Series Angled Attachment
(REF 5100-015-252)
Dimensions:| 5.2 inch [130.9 mm] length
0.8 inch [19.6 mm] diameter
Mass:| 0.26 lb [117.0 g]
Model:| MD Series Long Straight Attachment
(REF 5100-015-270)
Dimensions:| 4.3 inch [109.0 mm] length
0.8 inch [19.6 mm] diameter
Mass:| 0.18 lb [80.3 g]
Model:| MD Series Long Angled Attachment
(REF 5100-015-272)
Dimensions:| 6.0 inch [153.5 mm] length
0.8 inch [19.7 mm] diameter
Mass:| 0.28 lb [127.7 g]
Equipment Type:| See the instructions for use supplied
with the handpiece.
Mode of Operation:
Duty Cycle:
Rest Between Cycles:| Non-continuous Operation
20 seconds on/20 seconds off, 10 times
30 minutes
Applied Part:| When used with handpieces, the attachments described in this
manual are applied parts. For more information, see the instructions for use
supplied with the handpiece.
Maximum
Temperature of
Applied Part:| Less than 124 °F [51 °C] (Maximum surface temperature as tested
to the standards listed under Product Safety Certification in the instructions
for use supplied with the console.)
Environmental Conditions:| Operation
Storage and Transportation
Temperature Limitation:|
Humidity Limitation:
Atmospheric Pressure Limitation:
ES/DE/FR/IT/NL…………….5100-015-718
JA/ZH/KO……………………..5100-015-722
SV/DA/FI/PT/NO………………………5100-015-731
PL/EL…………………………….5100-015-750
Stryker Instruments
1941 Stryker Way
Portage, Michigan
(USA) 49002
1-269-323-7700
1-800-253-3210
2021-05
5100-015-735 Rev-AB
www.stryker.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>