General Information
1.1 Intended Use
Rheonik Coriolis mass flow measuring instruments consisting of a sensor (RHM),
a transmitter (RHE), and in some cases, an interconnecting cable, are intended
for measuring liquids and gases. The operator must ensure that the instrument
is only used in a way that complies with limitations and certifications listed
on the instrument type plates. These limitations/certifications include, but
are not limited to:
- Pressure and temperature ranges
- Approval limits (e.g., explosion protection, pressure equipment directive)
- Suitability of the materials of construction in contact with the process
The instrument must be installed, operated, and maintained in accordance with
the manufacturer’s instructions and guidelines. The instrument must not be
modified or altered in any way that has not been approved by the manufacturer
or certification body. Regular inspections and preventive maintenance should
be carried out to ensure ongoing safety and performance. Use not in
accordance.
1.2 Intended Use RHE45
The RHE45 Coriolis transmitter displays the measured values for RHM
Coriolis sensors.
It can display the mass flow of liquids and gases as well as the medium
density and medium temperature. In addition, the RHE45 Coriolis transmitter is
able to calculate variables such as the volume.
Only install and operate the RHE45 Coriolis transmitter in locations with the
following ambient conditions:
- Ambient temperature -40 to +60 °C
- Altitude of max. 3000 m above sea level
- Relative humidity of 10 to 95% (non-condensing)
- In shade; not exposed to direct sunlight
1.3 Impermissible use
The following are examples of impermissible use of the RHE45 Coriolis
transmitter (i.e. use contrary to its intended use):
- Using the RHE45 Coriolis transmitter in locations other than those specified above.
- Using the device in life-support systems in the medical, motor vehicle, aircraft, watercraft, or mining industries.
1.4 Important Safety Instructions for operating Coriolis Flowmeters
The RHE45 transmitter model is integrated in the RHM flow sensor. Maximum
temperature limits for each RHM flow sensor are stated on the permanently
attached serial number plate and must not be exceeded.
The closed transmitter box provides a protection class of at least IP66.
The use of a properly grounded supply is highly recommended.
The use of shielded cables for all I/O signals is recommended to prevent
interference from high level EMI.
For safety instructions regarding RHM Coriolis Flow Sensors, please refer to
the RHM installation and startup guide manual.
All national regulations and standards regarding electrical installation must
be observed!
1.5 Manufacturer’s Liability
Rheonik assumes no liability for loss and/or consequential damages stemming
from the use of this product utilized in life-preserving systems in the
medical, motor vehicle, aircraft, watercraft, or mining industries.
Rheonik accepts no liability for loss or damage resulting from the improper
use of any of its products.
Rheonik assumes no liability for the loss of production and/or consequential
damage from the use of this product unless such liability has been expressly
and contractually agreed.
Rheonik provides a standard one year from shipment warranty on all products
for workmanship and materials. Performance of this warranty will be at the
Rheonik manufacturing facility.
Rheonik assumes no liability for determining the suitability of its products
in any specific application. This is the sole responsibility of the end user.
1.6 Additional Resources
Since this document mainly describes the installation and commissioning of
the RHE45 transmitter it does not contain all information related to the
configuration of the transmitter and its features. These items are described
extensively in the documents listed in Table 1. Manuals and further resources
are available for download from the Rheonik website:
Title | Description | Document Number |
---|---|---|
RHE4X Desktop Reference | Extensive description of the configuration options of | |
the RHE4X transmitter series. | 8.2.1.14 | |
Addendum RHE4X Desktop Reference Data Logging | Description of the data logging | |
feature present in the RHE4X transmitter series. | 8.2.1.15 |
Addendum RHE4X Desktop Reference
Statistics
| Description of the statistic feature present in the RHE4X transmitter
series.| 8.2.1.28
RHEComPro Suite User Manual| Installation and use of the RHEComPro Suite for
PC for the operation and configuration of the RHE transmitters.| 8.2.1.18
Quick Start
Carry out the following steps to prepare the flow meter for operation:
-
If not already done, install the RHM sensor in line (refer to the RHM Coriolis Sensor
Installation Guide for additional information on mechanical installation). -
Connect all signal input, output and communications interface wiring as required (section 3.2.3).
-
Connect power supply (section 3.2.2).
NOTE:
For safety, always connect the protective ground. -
Switch on the power supply. The “READY” LED will light up (only visible when transmitter enclosure is opened).
-
Flush the RHM flow sensor with process fluid and ensure that it is free of air bubbles (in case of a liquid process fluid) or liquid droplets (in case of a gaseous process fluid).
-
Wait until the RHM temperature has stabilized and then carry out a zero-point calibration (section 4.3):
• Ensure a zero-flow condition is present in the flow sensor by closing isolation valves
• Initiate zeroing of the sensor and wait for the zeroing procedure to complete
• Open isolation valves to allow flow through the sensor
The device is ready for operation.
Installation
3.1 Mechanical Installation
The RHE45 is connected to the RHM ex works and cannot be removed.
For mounting the RHM refer to the corresponding manual delivered with the RHM.
When hot fluids are to be measured the housing of the RHE45 must not be
thermally isolated from the ambient for avoiding overheating the electronics.
When planning the mechanical installation please leave at least 12.5 cm/5 inch
space for the M12 plugs and the attached cables at the side where the M12
sockets are located.
Technical data and mechanical drawings can be found in Appendix A.
3.2 Electrical Installation
Caution, the surface of the RHE45 may be hot if a hot fluid is running
through the RHM sensor.
Any cable used for connecting to the RHE45 must be specified for 80°C or
higher. The use of shielded cable is strongly recommended.
3.2.1 Connectors
The RHE45 has one 12-pin M12 connector, “A” coded, for power supply and
standard interfaces and one 8-pin M12 connector, “A” coded for high-speed
interfaces of Ethernet 100base TX type.
With the I/O configurations B1, S1, S2, P1, and P2 the 8-pin socket is
internally not connected.
If the 8-pin M12 connector is not used, it must be closed with the protection
cap delivered with the RHE45 for maintaining the ingress protection.
As accessories Rheonik offers readily configured cables. For available
accessories see chapter 8.
3.2.1.1 12-pin M12 Connector
The 12-pin M12 connector of the RHE45 provides power supply, RS485 interface,
digital outputs, digital input, and the analog 4-20mA interfaces.
Caution: Before connecting the 12-pin M12 socket please check that the
power supply is correctly connected to the pins 1 and 3 of the 12-pin M12 plug
to prevent damage.
The following figures show the pin numbering of the 12-pin M12 socket in the
RHE4X transmitter and a respective plug. Table 2: Pin Configuration of the
12-pin M12 Socket
Pin # | Signal | I/O Option | Remark |
---|---|---|---|
1 | +24V | all | Positive supply terminal |
2 | AO1- | S1, S2, P1, or P2 | Negative terminal for analog output 1 |
3 | 0V/GND | all | Ground for power supply |
4 | RS485- | all | Negative signal of the two-wire RS485 interface |
5 | DO1 | all | Digital output 1 (called channel 0/A in the Modbus documentation) |
6 | RS485+ | all | Positive signal of the two-wire RS485 interface |
7 | DI1 | all | Digital input 1 |
8 | DO2 | all | Digital output 2 (called channel 1/B in the Modbus documentation) |
9 | AO1+ | S1, S2, P1, or P2 | Positive terminal for analog output 1 |
10 | I/O GND | all | Ground for digital in- and outputs |
11 | AO2+ | S2, or P2 | Positive terminal for analog output 2 |
12 | AO2- | S2 or P2 | Negative terminal for analog output 2 |
Shield | Ground / PE | all |
3.2.1.2 8-pin M12 Connector
The 8 pin M12 connector is only connected internally, if one of the following
options (see ordering code section 7) is installed:
EA, EB | Modbus TCP |
---|---|
EN | EtherNet/IP |
EP | ProfiNet |
DP | Profibus DP |
EC | EtherCAT |
The drawing shows the pin numbering of a 8 pin M12 connector. The function of
the pins for the different options is described in chapter 5. 3.2.2 Power
Supply and Grounding
For connections longer than 0.5m shielded cables should be used. For
connections longer than 3m, additional ferrite filter beads close to the RHE
are recommended for avoiding RF EMI.
The power supply input is protected by a fuse. As a protection against fire in
case of a short in the cable, the supply side of the cable should be protected
by a fuse (see figure 5, SI1) with a rating not higher than the current
carrying capacity of the cable.
The ground signal of the RHE45 (Terminal 3 and 10 of the 12 pin M12 connector)
is connected to PE and housing via a 10kΩ resistor.
The cables used for power supply and grounding must comply with the national
requirements. If required, certified cables must be used. The minimum cross
section is 0.35mm² (AWG 22).
The RHE45 does not contain a mains switch. A switch or circuit breaker close
to the RHE45must be included in the supply line. The switch must be marked
correspondingly.
3.2.2.1 Grounding and shielding
An RHM with an RHE45 transmitter must be grounded. Refer to the manual of
the RHM for the position of the ground terminal.
The cross-sectional area of the grounding wire must be equal to or larger than
the conductors used for any supply or I/O connection.
Note: National and Local electrical code requirements may contain
additional grounding requirements. Please ensure that all grounding to carried
out to these [legal] requirements.
For connections longer than 0.5m shielded cables should be used. For best EMC
performance the shield should be connected on both ends of the cable. If the
shield cannot be connected on both ends due to potential differences between
PE on both ends, one end might be connected to PE via a 1nF capacitor on one
end.
3.2.2.2 Power Supply The RHE45 requires a regulated DC supply voltage of
12V or 24V. It will operate with a power supply in the range 10V to 28V.
Connect the positive supply to terminal 1 and 0V to terminal 3 of the 12 pin
M12 connector.
Terminal 3 (0V) and housing (PE) are connected via a 10kΩ resistor. The
resistor can withstand a potential difference of up to 30V between 0V and PE,
but voltage differences above 5V should be avoided. For best EMC performance
connect the supply ground to PE close to the RHE45.
3.2.3 Control Inputs and Outputs
In general, shielded cables should be used for cable connections longer than
0.5m. For cable connections longer than 3m, it is recommended that ferrite
filter beads are installed close to the RHE terminations to help avoid RF EMI.
3.2.3.1 Digital Outputs
The RHE45 has two universal digital outputs which can be used as frequency,
pulse, or control/status outputs.
Digital outputs are push-pull outputs compliant to IEC 60946. They can drive
loads connected to ground and loads connected to +24V. Connect the output of
the RHE to the load. Connect the other side of the load to either GND or to
the positive supply rail. With a grounded load, the output current should be
limited to 20mA (480mW at 24V). With a load connected to the positive supply
rail the output can drive up to 50mA (1.2W at 24V).
The output can drive relays directly as well. Relays should be connected
between the output and the positive supply rail. The maximum output current is
internally limited to 50mA. If the output current exceeds this limit due to a
too low load resistance, the output will be disabled automatically.
3.2.3.2 Analog Outputs
Depending on the I/O configuration ordered, the RHE45 has passive, active or
no analog outputs.
The minimum current of the analog 4-20mA interfaces is limited to 3.2mA. Thus,
fire state settings of 2.0mA or 0.0mA are silently corrected to 3.2mA by the
RHE45 firmware.
Passive analog outputs (I/O option P1 or P2): Connect terminal 9 (11 for
output 2) to the positive terminal of a 24V supply (12V would be possible as
well) and terminal 2 (12 for output 2) to the load. The negative terminal of
the load must be connected to the GND terminal of the supply.
A protective resistor can be connected in series with the transmitter output
and the load. Please note, with a 24Vdc supply, the maximum total load
resistance is 600 Ω.
Active analog outputs (I/O option S1 or S2): Connect load between
terminal 9 (11 for output 2) and GND.
A protective resistor can be connected in series with the transmitter output
and the load. Please note, with a 24Vdc supply, the maximum total load
resistance is 600 Ω.
3.2.3.3 Digital Inputs
RHE45 transmitters have one or two IEC60946 compliant digital inputs.
Digital inputs have an input resistance of 24 kΩ and when operated, will draw
approx. 1mA when connected to a 24Vdc supply. Digital inputs can be operated
with an active DC signal instead of a switch. The maximum input voltage must
not exceed 30V DC. The required input voltage for a low-high transit is
typically 11V, maximum 12.5V, corresponding to an input current of about
0.5mA.
The digital input 2 is available with I/O option EA and EB only. The input is
located on the 8 pin M12 connector, pin 2 (DI2) and pin 3 (I/O GND).
Operation and Configuration
4.1 Operating elements
4.1.1 External If the 8-pin M12 connector is not used, it must be closed
with the protection cap delivered with the RHE45 for maintaining the ingress
protection.
4.1.2 Internal operating Elements
When the cover of the RHE45 enclosure is removed a part of the electronics
becomes accessible which contains some user configurable elements, e.g.:
- Mini USB socket (see chapter 5.1)
- Status LEDs, green (ON), yellow (Hard Lock switch active), red (ERROR)
- Hard Lock switch (option)
- RS485 termination switch
- Push button “ZERO”
- Power supply fuse
- Display with control keys (optionally)
Figure depicts the RHE45 transmitter without display when the cover is open.
When opening the cover of the RHE45 enclose make sure that no fluids enter the
enclosure and observe basic EMC measures such as connecting yourself to the
earth potential of the enclosure.
Figure also shows the switch positions for the activation or deactivation of
the RS485 termination and the Hard Lock switches. The activation of the Hard
Lock switch also will be shown by the yellow LED being lit. In this case no
modification of setup parameters is possible.
Zero Offset Calibration w/o interface
After installation of the RHM with RHE45 a zero calibration is strongly
recommended.
- Flush the RHM with the fluid to be measured and make sure that the RHM is 100% filled.
- Operate the RHE45 for at least 15 minutes for warming up.
- Open the top cover of the RHE by removing the 4 screws in the corners of the top cover.
- Stop the flow and make sure that there is no more flow. For best results a valve in front and behind the RHM would be recommended.
- Press the pushbutton “ZERO” with a tool (screwdriver, pencil) for about 2 seconds. The red and the green LED star flashing.
- When the flashing stops and the green LED lights constantly reclose the top cover.
- Reopen the valves.
- Make sure that the top cover gets properly closed for achieving the IP protection.
- The RHE45 must not be opened in wet or dirty (dusty) ambient or when the housing of the RHE45 is wet.
4.2 Operation and Configuration via Display
Optionally, the RHE45 transmitter has a user operation interface consisting of
a color LCD screen and three pushbuttons. The screen and pushbuttons are used
to navigate through a hierarchical menu structure that logically and
intuitively organizes the various features and functions available within the
programming set supplied with the instrument. This chapter gives a short
introduction into operation of the RHE45 through the front panel user
interface. For more detailed information, please refer to the RHE4X Desktop
Reference Manual.
Function of the Pushbuttons
RHE45 transmitters can be operated through three front panel pushbuttons.
These are labeled “ESC” [X], “NEXT” [>], and “ENTER” [ ]. The function of
these pushbuttons depends upon the currently displayed screen.
The following table details button function in specific situations.
Table 3: Functions of the Pushbuttons
Button | Situation | Function |
---|---|---|
ESC | Menu Navigation | Navigates to the menu one level up |
Number Input | Exits number input field and returns to the associated menu |
window without saving any changes
NEXT| Menu Navigation| Selects the next menu item
During Number Input| Increments the number at the cursor position by one.
Number increment is circular: a ‘9’ increments to ‘0’ for decimal numbers, ‘F’
increments to ‘0’ for hexadecimal numbers (used for service password)
Number Sign Change| Changes ‘+’ to ‘-’ and vice versa. When entering a number
with a sign character, the cursor will start at the first number character to
the right of the sign character in the field. In order to reach the sign
character, it is necessary to traverse all the way to the rightmost character
and then start at the beginning
of the input field again
Decimal Point Position Change| Shifts the decimal point and cursor position
one place to the right
End of input line| The entire number will flash upon pressing ENTER at the
rightmost character in an input field. Pressing NEXT when the field is
flashing will return the cursor to the start of the input line
ENTER| Menu Navigation| Navigates to the menu one level down or enters an
input screen
During Number, Sign or Decimal Point entry/ change| Moves the cursor one
character to the right in an input field. Note that when a number is entered
that is not in the parameters’ acceptable range, the field will change to
inverse colors (light on dark instead of dark on light). When a number is not
within an acceptable range, pressing ENTER at the rightmost character in the
field will return the cursor to the leftmost position in the field for re-
editing.
Number Input at the End| Pressing ENTER at the rightmost character in the
field will cause the entire field to flash. Pressing ENTER again will commit
the modified number and return display to the associated menu window
Status Infos – Bit Status Displays| Increment the index of individual bits in
the status words. Successively pressing ENTER in the Error Status, Soft Error
Status, or Warnings items will allow display/read of each individual bit value
and its status description
For example, and as depicted in Figure 13, when navigating the menu structure,
press the ENTER pushbutton once to get from the main display screen to the
top-level menu selection. Press the NEXT button repeatedly to scroll through
the entire menu. To enter any specific menu, press ENTER. 4.2.2 Passcodes
and Menu Selection
To configure and carry out certain functions in an RHE45 transmitter, it is
necessary to enter a predetermined passcode. Passcodes protect configuration
setup and prevent inadvertent changes to the flow meter operating condition.
There are two different access levels with separate passcodes: “USER” and
“SERVICE”. The USER level accesses those functions that relate to the
operation of the flow meter on a day-to-day basis i.e. zeroing and totalizer
reset. The SERVICE level allows access to configuration and setup functions
such as I/O ranging and calibration.
When needed, the screen will prompt for a passcode. Passcodes are entered
using the NEXT and ENTER buttons. Once entry is complete, the entire passcode
will flash. Press ENTER again to enter the menu. If the passcode supplied is
incorrect, the cursor will return to the leftmost character of the passcode
for re-entry.
Factory default passcodes are shown in Table 4. The USER level passcode can be
changed by accessing the “Service Login” menu.
Table 4: Default Passcodes
Access Level | Passcode |
---|---|
User | 1111 |
Service | 5678 |
Zero Offset Calibration
Before carrying out a zero-offset calibration, make sure that there is a zero
flow condition in the RHM sensor i.e. valves have been closed upstream and
downstream of the RHM meter body.
Navigate to the “Zero Now” menu item as shown in Figure 14 to perform the
zeroing procedure.
It will be necessary to enter the user passcode (see section .2). To perform
the zeroing procedure, press NEXT at the Y/N confirmation screen to change the
“N” to “Y”. Press ENTER to start the procedure or ESC to cancel. Once started,
a countdown timer will begin. When it reaches 0, the zero offset calibration
procedure is complete. Return to the main screen by pressing the EXIT
pushbutton twice.
4.2.1 Change of Positive Flow Direction
Rheonik mass flow meters are bi-directional and can operate with flow
passing through them in any direction. In some cases, after installation, the
transmitter may read negatively because of orientation. In this case, the flow
direction indication can be reversed within the transmitter. To change the
positive flow direction, navigate to the “Phase Measurement” menu item under
Service Login (Figure 15).
Change the value of the “PhsFlwDirConfig” parameter from “0” to “1” (or vice
versa) to change the flow direction indication of the transmitter. 4.2.5
Input/Output Configuration
To configure the analog and digital outputs, navigate to the respective
menu items under User Login (Figure 16). 4.2.5.1 Analog Output
Configuration
-
Select “Analog Output” 1 [C] or 2 [B] and press “ENTER” [] to get to the “Configuration” [C/B01] menu
-
Select one of the options stated in Table 5 under ID [C/B01], e.g. Configuration 3 – Density
-
The default settings are: “Maximum mA” [C/B02] = 20mA; “Minimum mA” [C/B03] = 4mA.
This can be changed if necessary -
Assign the maximum and minimum of the measured variable to respective mA level, e.g. 1200 kg/m³ for “Density Max” [C/B08] and 0 kg/m³ for “Density Min” [C/B09]
-
Select a fire state configuration [C/B14] to determine what shall happen if the measured variable range is exceeded or in case of error condition
-
If necessary, set a damping factor [C/B17] and damping band range [C/B18]
Table 5: Important Analog Output Parameters
ID | Abbreviation | Full Name / Description |
---|---|---|
__ [C/B01] | __ Cur Out Config | **Current Output Configuration: |
Assigns an output channel to the current output:
0 – Analog output is off.
1 – Analog output is configured for Mass Flow (default). → set [C/B04],
[C/B05] 2 – Analog output is configured for Volumetric Flow. → set [C/B06],
[C/B07] 3 – Analog output is configured for Density. → set [C/B08], [C/B09] 4
– Analog output is configured for Tube Temperature. → set [C/B10], [C/B11] 5 –
Analog output is configured for Torsion Bar Temperature. → set [C/B12],
[C/B13] 6 – Drive Gain. → set [C/B15], [C/B16]
ID| Abbreviation| Full Name / Description
---|---|---
[C/B14]| __ CurOutFireState| Current Output Fire State:
Determines the behavior (“fail high” or “fail low”) and value of the analog
output when the measured variable range is exceeded (values 1 to 5), i.e. the
measurement is outside of (CurOutCurMax) and (CurOutCurMin), or an error
condition exists in the SoftError or ErrorStatus fields (values 6 to 10).
State 0 – Output tracks the signal and clamps at the CurOutMaxCur and
CurOutMinCur values when its range is exceeded (default).
State 1 – 22 mA = range exceeded State 4 – 3.2 mA = range exceeded State 5 –
3.6 mA = range exceeded State 6 – 22 mA = error condition State 9 – 3.2 mA =
error condition State 10 – 3.6 mA = error condition
[C/B17]| __ Cur Out Damping Tau| Current Output Damping Tau:
Time constant (Tau) of the current output damping in seconds. An
exponential damping mechanism is used within a defined band, see
CurOutDampingBand. When the output values leave the defined band the damping
is disabled. A value of 0.0 disables the damping.
[C/B18]| Cur Out Damping Band| Current Output Damping Band:
** Defines the band range for the damping of the current output in percent of
the range between “Minimum mA” [C/B02] and “Maximum mA” [C/B03]. A value of
100 makes sure that the values never leave the band and that the damping
always is active.
4.2.5.2 Digital Output Configuration
-
Select “Pulse Output” 1 [R], 2 [S] or “Digital Output” A [K], B [L] and press “ENTER” [ ] to get to the “Configuration” [R/S01, I-L01] menu. The pulse outputs [R,S] can be assigned to a certain pulse rate from a measured value, e.g. the mass flow rate.
The digital outputs [K,L] can be used to indicate status or error.
“Pulse Output” 1 [R] and/or 2 [S] must be disabled if the parameters “Digital Output” A [K] and/or B [L] are assigned to universal digital/pulse outputs -
Select a configuration for the pulse output according to the options stated in Table 10 under ID [R/S01] or for the status output under ID [I-L01]
Examples:
a) Configuration of pulse output [R, S] for mass:
o Select value 1 for parameter ID [R/S01] o Acc. to Table 6, the parameter
[R/S02] has to be selected. This parameter specifies the mass equivalent to
one pulse. E.g. when set to 0.001kg, a mass transfer of 1kg will cause 1000
pulses
b) Configuration of a frequency output [R, S] for mass flow rate:
o Select value 13 for parameter [R/S01] o Acc. to Table 6 the parameters
[R/S04 & 07] have to be selected. The parameter [R/S04] specifies the mass
flow rate equivalent to the frequency set under parameter [R/S07]. E.g. when
[R/S04] is set to 1kg/min and [R/S07] is set to 1000Hz a mass flow rate of
5kg/min will cause a frequency of 5000Hz
c) Configuration of a digital output [I-L] for mass flow alarm:
o Select value 1 for parameter [I-L01] o Acc. to Table 6 the parameters
[I-L02, 03 & 04] apply. Select the needed alarm type [I-L02] and enter the low
[I-L03] and high [I-L04] alarm values. The unit of the alarm is determined by
the digital output configuration [I-L01]. For mass flow alarm the unit is
fixed to kg/min
Table 6: Important Pulse and Digital Output Parameters
ID | Abbreviation | Full Name / Description |
---|---|---|
R/S01] | VersDigOutCh0/1Config | Versatile Digital Output Configuration: |
Used to assign different functions to the pulse output.
In simple pulse mode, Pulsel and Pulse2 operate independently. In phase shift
mode, Pulsel and Pulse2 work together to produce the desired pulse output
relationship. In phase shift mode, both outputs (Pulsel & Pulse2) should be
set to the same mode.
0: Pulse output is off.
1:Mass Flow — Simple Pulse, Forward Flow. 4 set [R/S02] 2:Volume Flow — Simple
Pulse, Forward Flow. 4 set [R/S03] 3:Mass Flow — Pulsel & Pulse2 90 Degrees
phase shift. 4 set [R/S02] 4:Volume Flow — Pulsel & Pulse2 90 Degrees phase
shift. 4 set [R/503] 5:Mass Flow — Simple Pulse, Reverse Flow. 4 set [R/S02]
6:Volume Flow — Simple Pulse, Reverse Flow. 4 set [R/S03] 7:Mass Flow Rate
Fast Frequency Output. 4 set [R/SO4 & 07]
8:Volumetric Flow Rate Fast Frequency Output. 4 set [R/S05 & 07]
9:Density Fast Frequency Output. 4 set [R/506 & 07] 10:Mass Flow Rate Slow
Frequency Output. 4 set [R/SO4 & 07] 11:Volumetric Flow Rate Slow Frequency
Output. 4 set [R/505 & 07] 12:Density Slow Frequency Output. 4 set [R/506 &
07] 13:Mass Flow Rate RHE Compatibility Mode Frequency Output. 4 set [R/SO4 &
07] 14:Volume Flow Rate RHE Compatibility Mode Frequency Output.’
4 set [R/505 & 07] _
[14011| DigOutCh0/1/A/BConfig| Digital Output Channel 0 Configuration:
Alarm Channel configuration:
0: Error indication (default)
1:Mass flow alarm 4 set [I-102, 03 & 04] 2:Volumetric flow alarm 4 set [I-L02,
03 & 04] 3:Density alarm 4 set [I-L02, 03 & 04] 4:Tube temperature alarm 4 set
[I-102, 03 & 04] 5:Torsion bar temperature alarm 4 set [I-L02, 03 & 04]
6:Pressure Alarm 4 set [I-L02, 03 & 04] 7:Forward Mass Totalizer TotalMassFwd
for Batch Mode 4 set [I-L02, 03 & 04] 8:Forward Volume Totalizer TotalVolFwd
for Batch Mode 4 set [I-102, 03 & 04] 9:Fatal errors or Zeroing force output
low, else high.
10:Fatal errors or Zeroing force output high, else low.
For values 1 to 8, the ERR output is pulled high when the alarm condition
specified in DigOutCh0/1/A/BAlmType, DigOutCh0/1/A/B Almlow, and
DigOutCh0/1/A/B AlmHigh becomes true.
[I-L021| DigOutCh0/1/A/BAlmType| Digital Output Alarm Type:
Alarm Channel: Digital Alarm Type:
0: Setpoint alarm (hysteresis, default) — Alarm is indicated when the output
is higher
than DigOutCh0A1mHigh and cleared when it returns below DigOutChOAlmlow.
1:Inband alarm — Alarm is indicated when the output is in the range from
DigOutCh0A1mLow to DigOutCh0A1mHigh.
2:Outband alarm — Alarm is indicated when the output is outside the range
DigOutChOAlmlow and DigOutCh0A1mHigh.
An indicated alarm on channel 0 pulls the digital output 4 labeled “ERR” low.
ID| Abbreviation| Full Name / Description
---|---|---
[I-L03]| DigOutCh0/1/A/BAlmLow| Digital Output Alarm Low:
Alarm low value, see DigOutCh0/1/A/B AlmType. Must be a floating-point
value of 0.0 or higher. There is no unit conversion for this register. The
unit depends on the DigOutCh0/1/A/B Config setting:
1: kg/min
2: m3/min
3: kg/m3
4: Degrees Celsius
5: Degrees Celsius
6: Pascal
7: kg
8: m^3
The same units are used for all digital status outputs 0, 1, A, and B
depending on DigOutCh0/1/A/BConfig respectively.
[I-L04]| DigOutCh0/1/A/BAlmHigh| Digital Output Channel 0 Alarm
High:
Alarm high value, see DigOutCh0/1/A/BAlmType. Must be a floating-point
value of 0.0
or higher. See DigOutCh0/1/A/BAlmLow for the unit specification.
4.2.5.3 Digital Input Configuration
To configure the digital inputs, navigate to the menu item “HMI” under User Login (Figure 17).
- Select “HMI” [H] and navigate to the “DI1 Property” (digital input 1) [H02] or to the “DI2 Property” (digital input 2) [H03] menu.
- Select one of the options stated in Table 11.
Table 7: Digital Input Setting Options
ID | Abbreviation | Full Name / Description |
---|---|---|
__ [H02] [H03] | DI1Property DI2Property | **Property of Digital Input 1/2: |
** Functionality of Digital Input 1/2:
0: DI1/2 is disabled. All inputs are ignored (default).
1: DI1/2 causes the start of a Zeroing Process. Logic is positive – a
transition to a high logic level starts the Zeroing.
2: DI1/2 causes the start of a Zeroing Process. Logic is negative – a
transition to a low logic level starts the Zeroing.
3: DI1/2 causes reset of the Totalizers and the start of Batch Processing
if configured, see RHE4X Desktop Reference, Batch Mode. Logic is positive – a
transition to a high logic level starts the Batch.
4: DI1/2 causes reset of the Totalizers and the start of Batch Processing
if configured, see RHE4X Desktop Reference, Batch Mode. Logic is negative – a
transition to a low logic level starts the Batch.
5: A transition from low to high at DI1/2 stops the totalizers when they
are in the running state. A transition from high to low at DI1/2 restarts the
totalizers when they are in the stopped state.
6: A transition from high to low at DI1/2 stops the totalizers when they
are in the running state. A
transition from low to high at DI1/2 restarts the totalizers when they are in
the stopped state.
4.3 Operation and Configuration via USB or RS485
If the RHE45 is connected via RS485, we recommend making all setup via RS485,
if possible, via the remote-control SW RHEComPro.
If there is no RS485 connected, you can use the internal Mini USB interface.
-
The RHE45 must not be opened in wet or dirty (dusty) ambient or when the housing of the RHE45 is wet.
For setting up the RHE45 via the internal USB interface proceed as follows: -
Open the top cover of the RHE by removing the 4 screws in the corners of the top cover.
-
Connect the USB to a PC, preferable to a laptop.
-
Start RHEComPro on the laptop. For more information about installation and usage of the RHEComPro refer to the manual of the RHEComPro.
-
Connect to the RHE45 by clicking on “Connect”. If the SW does not find the RHE45, try a different COM port.
-
After having finished the setup disconnect the RHE45 in RHEComPro (ikon in left sidebar), remove the USB cable and reclose the RHE45.
-
Make sure that the top cover gets properly closed for achieving the IP protection.
4.3.1 Zero Offset Calibration
After installation of the RHM with RHE45 a zero calibration is strongly
recommended.
- Flush the RHM with the fluid to be measured and make sure that the RHM is 100% filled.
- Operate the RHE45 for at least 15 minutes for warming up.
- Stop the flow and make sure that there is no more flow. For best results a valve in front and behind the RHM would be recommended.
- Click 2 times on the arrow in the right upper corner of the RHEComPro main display.
- Click on “Make Zero” in the right lower corner of the RHEComPro display. The field “Zero” gets yellow and the time counter counts down to zero.
- When the counter stops at zero and the field “zero” returns to green, reopen the valves for normal operation.
4.3.2 Input/Output Configuration
The RHEComPro has a menu for configuring the outputs. The menu is self-
explaining, and no special knowledges are required.
Proceed as follows:
- Click on “Configuration” on top of the RHEComPro main display.
- Select the input or output to be configured and follow the explanations in the menu.
- After “OK” or “Cancel” at the end the RHEComPro returns to normal operation.
4.3.3 I/O test
For a fast check of the I/O connections between the RHE45 and the SPS/PLC, the
RHEComPro provides an I/O test menu.
For testing the I/O connections proceed as follows:
- Click on “Maintenance” on top of the RHEComPro main display.
- Select “Standard Maintenance Activities”
- Select “Fast Tests for I/O”
- Select the output to be tested and follow the explanations in the menu.
- After “Exit the Test” the RHEComPro returns to normal operation.
Internal System Time
The RHE45 has a build in real time clock (RTC) which is useful for data
logging or zero history. For further information please refer to the RHE4X
Desktop Reference manual.
- The RHE45 does not feature a backup battery. The date and time must be updated after every system start or reset.
Remote Operation
RHE45 transmitters have a USB interface for fast setup or trouble shooting and
an RS485 interface for digital communication, remote control, and data
exchange.
Optionally Modbus TCP, EtherNet/IP, ProfiNet, Profibus DP and EtherCAT are
available.
5.1 USB
The USB interface is designed for temporary connection of a PC running
RHEComPRO software for setup, configuration, troubleshooting and data logging.
Connection to the USB interface is by Mini-USB cable to a PC-based computer.
For installing, starting and operating the RHECom software please refer to the
RHEComPro Suite User Manual and the RHE4X Desktop Reference manual.
5.2 RS485
The RS485 port uses Modbus protocol for permanent connection to a supervisory
control system. For using Modbus via the RS485 serial port, refer to the RHE45
Desktop Reference Manual for a detailed description of the interface and
register set.
The RHE45 has an internal 120Ω termination resistor for the RS485. This
resistor can be disconnected via a switch. Ex works the resistor is connected
(switch position “ON”).
In a point-to-point connection of the RS485 the termination resistor is
required for cable length of more than about 5m.
In a chain connection the resistor must be “ON” at the last device in the
chain and “OFF” at all devices in between.
For accessing the switch refer to chapter 4.1.2.
5.3 Modbus TCP (I/O options EA and EB)
The options EA or AB provide one Ethernet 100base TX port and the digital
input DI2 on the 8- pin M12 socket.
Table 8 shows the pinning of the M12 connector, the cable colors (accessory
ARHE45-M2) and the RJ45 pinning.
Table 8: 8-Pin M12 Socket Configuration for Options EA, EB and RJ45 wiring
M12 Pin # | Signal | CAT5 Coloring | RJ45 Pin #, Name |
---|---|---|---|
1 | – | white/blue | |
2 | DI2, digital input 2 | white/brown | |
3 | I/O GND | brown | |
4 | Eth 1 TX- | orange | 2, D1- |
5 | Eth 1 RX+ | white/green | 3, D2+ |
6 | Eth 1 TX+ | white/orange | 1, D1+ |
7 | – | blue | |
8 | Eth 1 RX- | green | 6, D2- |
Shield | Ground / PE | Shield |
Make sure that the digital input DI2 (pin 2 and 3) gets not connected to any
Ethernet signal. Otherwise the RHE45 might not work properly.
For further information please refer to the RHE4X Desktop Reference manual.
5.4 EtherNet/IP, ProfiNet and EtherCAT (I/O options EN, EP and EC)
The options EN, EP, and EC provide two Ethernet 100base TX port on the 8-pin
M12 socket.
For easiest connection to the Ethernet use the connection cable ARHE45-MY.
Table 9 shows the pinning of the M12 connector, the cable colors (ARHE45-MY)
and the RJ45 pinning.
Table 9: 8-pin M12 Socket Configuration for Options EN, EP, EC and RJ45 wiring
M12 Pin #| Signal| CAT5 Coloring| 1 st RJ45 Pin #, Name|
2 nd RJ45 Pin #, Name
---|---|---|---|---
1| Eth 2 TX-| white/blue| | 2, D1-
2| Eth 2 RX+| white/brown| | 3, D2+
3| Eth 2 RX-| brown| | 6, D2-
4| Eth 1 TX-| orange| 2, D1-|
5| Eth 1 RX+| white/green| 3, D2+|
6| Eth 1 TX+| white/orange| 1, D1+|
7| Eth 2 TX+| blue| | 1, D1+
8| Eth 1 RX-| green| 6, D2-|
Shield| Ground / PE| | Shield| Shield
These fieldbus interface offer a limited access mainly to the measurement data maintained by the RHE transmitter. Following measurement data items are available to be read out periodically:
Index | Modbus Name | Modbus Reference | Type |
---|---|---|---|
256 | Error Status | See Generic / 0x401A. | Unsigned32 |
257 | Soft Error | See Generic / 0x401C. | Unsigned32 |
258 | Warnings | See Generic / 0x401E. | Unsigned32 |
259 | InfoStatus | See Generic / 0x4020. | Unsigned32 |
260 | Den Comp | See Density / 0x4806. | Float32 |
261 | Mass Flow Rate | See Mass Flow / 0x4900. | Float32 |
262 | Volumetric Flow Rate | See Volumetric Flow / 0x4A00. | Float32 |
Index | Modbus Name | Modbus Reference | Type |
--- | --- | --- | --- |
263 | Total Mass Fwd | See Totalizer / 0x4B00. | Float32 |
264 | Total Vol Fwd | See Totalizer / 0x4B02. | Float32 |
265 | TotInvenMassNet | See Totalizer / 0x4B04. | Float32 |
266 | TotInvenVolNet | See Totalizer / 0x4B06. | Float32 |
267 | TotalMassRev | See Totalizer / 0x4B08. | Float32 |
268 | TotalVolRev | See Totalizer / 0x4B0A. | Float32 |
269 | AdcTubeMeanTemp | See Temp. Measurement / 0x4500. | Float32 |
270 | AdcTorBarMeanTemp | See Temp. Measurement / 0x4502. | Float32 |
271 | OnBrdTemp | See Temp. Measurement / 0x4504. | Float32 |
272 | PrsMean | See Pressure / 0x4604 (RHE2X). | Float32 |
273 | AssuranceFactor | See Generic / 0x4026 (RHE2X). | Float32 |
274 | StdDensity | See Density / 0x480A. | Float32 |
275 | VolPercentMainSubstance | See Density / 0x480C. | Float32 |
276 | MassFlowRateDisplay | See Mass Flow/0x4904 | Float32 |
277 | VolumetricFlowRateDisplay | See Volume Flow/0x4A02 | Float32 |
The measurement data is expanded by their unit codes and unit strings added for convenience. These are intended to be read out at the startup of a system:
Index | Modbus Name | Modbus Address | Type |
---|---|---|---|
320 | TemperatureUnit | 0x6100 | Unsigned32 |
321 | PressureUnit | 0x6102 | Unsigned32 |
322 | MassUnit | 0x6104 | Unsigned32 |
323 | MassFlowUnit | 0x6106 | Unsigned32 |
324 | DensityUnit | 0x6108 | Unsigned32 |
325 | VolumeFlowUnit | 0x610A | Unsigned32 |
326 | VolumeUnit | 0x610C | Unsigned32 |
327 | TemperatureUnitString | Octet String | |
328 | PressureUnitString | Octet String | |
329 | MassUnitString | Octet String | |
330 | MassFlowUnitString | Octet String | |
331 | DensityUnitString | Octet String | |
332 | VolumeFlowUnitString | Octet String | |
333 | VolumeUnitString | Octet String |
The unit strings are 8 characters long and not 0-terminated. They are encoded
in the ISO 8859-1 character set and contain special characters such as ‘°’ or
‘²’ which must be translated to the target character encoding used for the
display of strings.
Further output data items are specified which can be read or written and can
used to influence the RHE transmitter. These are shown in the following table:
Index | Modbus Name | Modbus Address | Type |
---|---|---|---|
384 | UserPassword | 0x6004 | Octet String (0x0A) (4 ASCII Characters) |
385 | TotInvenReq | 0x6B00 | Unsigned32 (0x07) |
386 | ZeroingRequest | 0x6718 | Unsigned32 (0x07) |
387 | Squawk | 0x6F1C | Unsigned32 (0x07) |
Before a Zeroing or the Squawk function can be initiated the currently valid
user password has to be written to the UserPassword data item. The default
user password is “1111”
For further information on the referenced data items please refer to the RHE4X
Desktop Reference manual.
The ProfiNet and EtherNet/IP interfaces also offer a transparent access to a
Modbus TCP protocol which facilitates the access to all the configuration
parameters of the RHE transmitter, e.g. with the help of the RHEComPro
program. This may be used in the system commission phase via the same 100 Base
TX ports used for the fieldbus accesses. The port number for the Modbus TCP
protocol is defined in the setup register Modbus_TCP_Port (0x60C0). By
default, this register is set to 502, the standard Modbus port number.
5.4.1 ProfiNet Specific Information
The two 100 Base TX interfaces present for ProfiNet can be used to implement
an S2 System Redundancy. The standard GSDML file is already prepared for this
option. Rheonik was assigned the Vendor ID 0x056C and the RHE4X transmitter
series got the Device ID 0x0001. Whilst the employed ProfiNet module supports
isochronous transfers (IRT), this feature is not used since the mass flow
measurements are done independently from any master synchronization. When no
dedicated IP address is specified in the setup register IPv4_Addr (0x60C2)
there are several possibilities to assign an IP address to the ProfiNet
interface:
- DCP (ProfiNet)
- Secure HICP
- SNMP
- DHCP
When an IP address has been assigned the selection of the IP address
assignment method also can be done via the onboard HMTL pages. Thus, a
preliminary IP address may be set via IPv4_Addr, the assignment method
selected via the HMTL pages and thereafter a reset of IPv4_Addr to 0.0.0.0
will activate the assignment method. If security concerns are an issue, please
configure the SNMP interface with an appropriate password.
All cyclic measurement values listed above in the index range of 256 to 277
are accessed via the API=0/Slot=0/Subslot=1 address tuple and the index of the
data item as specified above. These data items are also included in the GSDML
file.
The other registers with indices higher than 319 contain acyclic data which
have to be accessed by the ProfiNet “Get Record” or “Set Record” commands.
They are located in the same API=0/Slot=0/Subslot=1 address tuple.
The data area layout for the periodic data is available as General Station
Description (GSDML) file which can be obtained from the Rheonik Service.
5.5 Profibus DP (I/O option DP)
The option DP provides one Ethernet 100base TX port and the digital input
DI2 on the 8-pin M12 socket.
Table 10 shows the pinning of the M12 connector, the cable colors (accessory
ARHE45-M2) and the recommended standard pinning of a 9-pin D-Sub connector.
Table 10: 8-pin M12 Socket Configuration for Option DP
M12 Pin # | Signal | CAT5 Coloring | Reference 9-pin Sub-D |
---|---|---|---|
1 | – | white/blue | |
2 | DI2, digital input 2 | white/brown | |
3 | I/O GND | brown | |
4 | BUS_A (RS485-) | orange | 3 |
5 | +5V | white/green | 6 |
6 | BUS_B (RS485+) | white/orange | 8 |
7 | – | blue | |
8 | DGND | green | 5 |
Shield | Ground / PE | Shield (1) |
The PROFIBUS station address may be set from the master using the PROFIBUS identification of the RHE. This is currently 0xA56. An initial station address may be set in the RHE parameter set in the following Modbus holding register:
Address | Register Name | Type | Description |
---|---|---|---|
0x60CC | FieldbusNodeId | UINT32 | Values in the range of 0 to 125 set the node |
address, other values assume the use of SSA.
All measurement data items transferred on the PROFIBUS interface are 4-byte
entities and are of unsigned integer or IEEE 754 32-bit floating point type.
These are transferred in big endian byte ordering.
The layout of the measurement related SDO data and the default PDO area is
shown in the following table:
Rd_Inp (SAP56)
offset
| Slot / Index| Modbus Name| Modbus Reference| Type
---|---|---|---|---
0| 1 / 0| ErrorStatus| See Generic / 0x401A.| Unsigned32 (0x07)
4| 1 / 1| SoftError| See Generic / 0x401C.| Unsigned32 (0x07)
8| 1 / 2| Warnings| See Generic / 0x401E.| Unsigned32 (0x07)
12| 1 / 3| InfoStatus| See Generic / 0x4020.| Unsigned32 (0x07)
16| 1 / 4| DenComp| See Density / 0x4806.| Floating Point (0x08)
20| 1 / 5| MassFlowRate| See Mass Flow / 0x4900.| Floating Point (0x08)
24| 1 / 6| VolumetricFlowRate| See Volumetric Flow / 0x4A00.| Floating Point
(0x08)
28| 1 / 7| TotalMassFwd| See Totalizer / 0x4B00.| Floating Point (0x08)
32| 1 / 8| TotalVolFwd| See Totalizer / 0x4B02.| Floating Point (0x08)
36| 1 / 9| TotInvenMassNet| See Totalizer / 0x4B04.| Floating Point (0x08)
40| 1 / 10| TotInvenVolNet| See Totalizer / 0x4B06.| Floating Point (0x08)
44| 1 / 11| TotalMassRev| See Totalizer / 0x4B08.| Floating Point (0x08)
48| 1 / 12| TotalVolRev| See Totalizer / 0x4B0A.| Floating Point (0x08)
52| 1 / 13| AdcTubeMeanTemp| See Temp. Measurement / 0x4500.| Floating Point
(0x08)
56| 1 / 14| AdcTorBarMeanTemp| See Temp. Measurement / 0x4502.| Floating Point
(0x08)
60| 1 / 15| OnBrdTemp| See Temp. Measurement / 0x4504.| Floating Point (0x08)
64| 1 / 16| PrsMean| See Pressure / 0x4604 (RHE2X).| Floating Point (0x08)
Rd_Inp (SAP56) offset| Slot / Index| Modbus Name| Modbus
Reference| Type
0| 1 / 0| ErrorStatus| See Generic / 0x401A.| Unsigned32 (0x07)
4| 1 / 1| SoftError| See Generic / 0x401C.| Unsigned32 (0x07)
8| 1 / 2| Warnings| See Generic / 0x401E.| Unsigned32 (0x07)
12| 1 / 3| InfoStatus| See Generic / 0x4020.| Unsigned32 (0x07)
16| 1 / 4| DenComp| See Density / 0x4806.| Floating Point (0x08)
20| 1 / 5| MassFlowRate| See Mass Flow / 0x4900.| Floating Point (0x08)
24| 1 / 6| VolumetricFlowRate| See Volumetric Flow / 0x4A00.| Floating Point
(0x08)
28| 1 / 7| TotalMassFwd| See Totalizer / 0x4B00.| Floating Point (0x08)
32| 1 / 8| TotalVolFwd| See Totalizer / 0x4B02.| Floating Point (0x08)
36| 1 / 9| TotInvenMassNet| See Totalizer / 0x4B04.| Floating Point (0x08)
40| 1 / 10| TotInvenVolNet| See Totalizer / 0x4B06.| Floating Point (0x08)
44| 1 / 11| TotalMassRev| See Totalizer / 0x4B08.| Floating Point (0x08)
48| 1 / 12| TotalVolRev| See Totalizer / 0x4B0A.| Floating Point (0x08)
52| 1 / 13| AdcTubeMeanTemp| See Temp. Measurement / 0x4500.| Floating Point
(0x08)
56| 1 / 14| AdcTorBarMeanTemp| See Temp. Measurement / 0x4502.| Floating Point
(0x08)
60| 1 / 15| OnBrdTemp| See Temp. Measurement / 0x4504.| Floating Point (0x08)
64| 1 / 16| PrsMean| See Pressure / 0x4604 (RHE2X).| Floating Point (0x08)
Rd_Inp (SAP56) offset| Slot / Index| Modbus Name| Modbus
Reference| Type
---|---|---|---|---
68| 1 / 17| AssuranceFactor| See Generic / 0x4026 (RHE2X).| Floating Point
(0x08)
72| 1 / 18| StdDensity| See Density / 0x480A.| Floating Point (0x08)
76| 1 / 19| VolPercentMainSubstance| See Density / 0x480C.| Floating Point
(0x08)
80| 1 / 20| MassFlowRateDisplay| See Mass Flow/0x4904| Floating Point (0x08)
84| 1 / 21| VolumetricFlowRateDisplay| See Volume Flow/0x4A02| Floating Point
(0x08)
The measurement data is expanded by their unit codes and unit strings added for convenience:
Slot / Index | Modbus Name | Modbus Address | Type |
---|---|---|---|
1 / 64 | TemperatureUnit | 0x6100 | Unsigned32 (0x07) |
1 / 65 | PressureUnit | 0x6102 | Unsigned32 (0x07) |
1 / 66 | MassUnit | 0x6104 | Unsigned32 (0x07) |
1 / 67 | MassFlowUnit | 0x6106 | Unsigned32 (0x07) |
1 / 68 | DensityUnit | 0x6108 | Unsigned32 (0x07) |
1 / 69 | VolumeFlowUnit | 0x610A | Unsigned32 (0x07) |
1 / 70 | VolumeUnit | 0x610C | Unsigned32 (0x07) |
1 / 71 | TemperatureUnitString | Octet String (0x0A) | |
1 / 72 | PressureUnitString | Octet String (0x0A) | |
1 / 73 | MassUnitString | Octet String (0x0A) | |
1 / 74 | MassFlowUnitString | Octet String (0x0A) | |
1 / 75 | DensityUnitString | Octet String (0x0A) | |
1 / 76 | VolumeFlowUnitString | Octet String (0x0A) | |
1 / 77 | VolumeUnitString | Octet String (0x0A) |
The unit strings are 8 characters long and not 0-terminated. They are encoded
in the ISO 8859-1 character set and contain special characters such as ‘°’ or
‘²’ which must be translated to the target character encoding used for the
display of strings.
Further output data items are specified which can be used to influence the RHE
transmitter. These are shown in the following table:
Slot /Index | Modbus Name | Modbus Address | Type |
---|---|---|---|
1 / 128 | UserPassword | 0x6004 | Octet String (0x0A) (4 ASCII Characters) |
1 / 129 | TotInvenReq | 0x6B00 | Unsigned32 (0x07) |
1 / 130 | ZeroingRequest | 0x6718 | Unsigned32 (0x07) |
1 / 131 | Squawk | 0x6F1C | Unsigned32 (0x07) |
Before a Totalizer Reset or a Zeroing can be initiated the currently valid
user password has to be written to the UserPassword data item.
The entire data area layout is available as General Station Description (GSD)
file which can be obtained from the Rheonik Service.
For further information please refer to the RHE4X Desktop Reference manual.
Service and Maintenance
6.1 Maintenance and Calibration
RHE45 and the associated RHM flow sensor do not require any regular
maintenance.
In harsh environments or if highest accuracy is required, we recommend a
calibration every 2 years.
For the calibration procedure refer to the RHE4X Desktop Reference Manual
and/or contact Rheonik or your local dealer.
For best performance the zero-point calibration should be performed from time
to time or when the meter shall be used under drastically changed operating
conditions (temperature, density, pressure…).
6.2 Supervision and Troubleshooting
Remote operation:
On the page “Diagnosis” of the RHEComPro (one click left from the main screen)
the stability data and the error status can be checked.
RHE45 with display: Each status window shows a code e.g. ‘0x00000000’. By
pressing the ENTER button, the cursor walks through this code from right to
left. The respective error or warning bit is explained at the bottom of the
display.
For detailed assistance, please refer to the troubleshooting section of the
RHE4X Desktop Reference. This chapter contains comprehensive information on
understanding and troubleshooting errors, warnings, and connection issues with
the flow meter.
6.3 Service
RHE45 transmitters do not contain any user serviceable parts. Please contact
your local sales/support agent or Rheonik Support for assistance.
6.4 WEEE and RoHS
RHE45 transmitters are not subject to the WEEE directive and are fully
compliant to the RoHS directive.
Ordering Code
Accessories
For tests, test installations, and as a reference Rheonik offers a small
number of accessories fitting to the two M12 sockets of the RHE45.
Table 11: RHE45 Accessories
Order Code | Description | Exemplary Pictures |
---|---|---|
ARHE-SO | RHECom Software on USB flash drive | – |
ARHE-RS | Interconnection Cable RHE45 to PC, 5m (Mini- USB to PC USB) |
__
ARHE-MO| Adapter Modbus RS485 to USB with Interface Cable, 2m
Accessory ARHE45-MS recommended for connection to RHE45|
ARHE45-M1| Shielded 12-core Cable with 12-Pin M12 Plug, 3m| __
ARHE45-M2| Shielded 8-core Cable with 8-Pin M12 Plug, 2m|
__
ARHE45-MS| Shielded Cable with 12-Pin M12 Plug and Test Interface, 3m|
ARHE45-MY| Shielded CAT5 Cable for two Ethernet 100Base TX Interfaces with
RJ45 Sockets, 2m|
Appendix A Technical Information
A.1 Technical Data
General Data
Housing: | Closed Box, directly mounted to the RHM |
---|---|
Material: | Aluminum, powder coated |
Enclosure Rating: | IP65; up to IP67 upon request |
Ambient Temperature: | -40°C to +60°C (-44°F to +140°F) |
Humidity: | 10 to 90% relative humidity, non-condensing |
Altitude: | Up to 3000m, higher maximum altitude upon request |
Dimensions: | See drawing chapter B.2 |
Weight: | 0.5kg |
Display: | Optionally – high contrast backlit LCD, screen changes color to |
indicate warning or error
Operation:| Via remote control
Power Supply:| DC: 12-24 VDC +/- 10%, 6W
Connection:| Connectors M12 / 12 pin and M12 / 8pin
Interfaces:| Modbus RTU (RS485) and USB (standard) Modbus TCP, Ethernet/IP,
ProfiNet, Profibus DP, EtherCAT (option)
Analog Outputs
Type: | 4-20mA compliant to NAMUR NE-43 |
---|---|
Signal type and quantity: | 0, 1 or 2, passive floating and active referenced |
to GND, type and quantity depending on I/O type
Output value:| Programmable
Digital Universal Outputs
Output signal: | Pulse, Frequency, Status |
---|---|
Type: | 24V push-pull, according to IEC60946 |
Maximum current: | 50mA |
Frequency range: | 0 … 10 kHz |
Quantity: | 2 |
Output value: | Programmable |
Digital Inputs
Type: | 24V, according to IEC60946 |
---|---|
Input Level: | >13V for high state, < 6V for low state |
Input Current: | 1mA |
Quantity: | 1 to 2 |
Input value: | Programmable |
A.2 Mechanical Drawing
For detailed drawings and dimensions please refer to the RHM data sheets
About Rheonik
Rheonik has but one single purpose: to design and manufacture the very best
Coriolis meters available.
Our research and engineering resources are dedicated to finding new and better
ways to provide cost effective accurate mass flow solutions that provide value
to our customers. Our manufacturing group care for each and every meter we
produce from raw materials all the way to shipping, and our service and
support group are available to help you specify, integrate, start-up and
maintain every Rheonik meter you have in service. Whether you own just one
meter or have hundreds, you will never be just another customer to us. You are
our valued business partner.
Need a specific configuration for your plant? Don’t compromise with a
“standard” product from elsewhere that will add extra cost to your
installation. If we can’t configure it from our extensive and versatile
product range, our exclusive AnyPipeFit Commitment can have your flow sensor
customized with any size/type of process connection and face to face dimension
you need. No matter what control system you use as the backbone in your
enterprise, with our AnyInterface Commitment, you can be sure that connection
and communication will not be a problem. Alongside a wide variety of discrete
analog and digital signal connections, we can also provide just about any
network/bus interface available (for example: HART, ProfibusDP, ProfiNet,
EtherCAT, PowerLink, EtherNet/IP, CAN, ….) with our RHE 40 Series family of
transmitters. Rheonik RHE 40 Series transmitters can connect to your system –
no headache and no conversion needed.
Rheonik Messtechnik GmbH
Rudolf-Diesel-Straße 5
D-85235 Odelzhausen
Germany
Tel. + 49 (0)8134 9341-0
info@rheonik.com
https://www.rheonik.com
Document No.: 8.2.1.21