RHEONIK RHE46 Coriolis Mass Flow Transmitter
DocumentNo.: 8.2.1.34
Version 1.00
December2022
Foreword
This installation guide contains important information regarding how to
install, start up, and safely operate the RHE46 Coriolis transmitter as
intended together with an RHM Coriolis sensor. It supplements the information
provided in the RHE 40 Desktop Reference. The RHE 40 Desktop Reference and
further information are available to
download from the Rheonik Messtechnik GmbH website: ⮫
https://www.rheonik.com.
Copyright
This installation guide is protected by copyright law. Any use that goes beyond the strict limits set by copy-right law is forbidden and liable for prosecution without the prior consent of Rheonik Messtechnik GmbH. This also applies to any reproduction, translation and microfilming in addition to saving and editing in electronic media. Any infringements will be liable for compensation!
Disclaimer
Rheonik Messtechnik GmbH accepts no liability for personal injury, damage to
property or accidents caused by the following:
- Failure to observe the installation guide
- Failure to observe the safety regulations
- Using the RHE46 Coriolis transmitter in an impermissible manner
Liability
Claims for compensation against Rheonik Messtechnik GmbH arising due to a
breach of pre-contractual or contractual obligations or tortious acts shall
exist only in the case of intent or gross negligence on the part of the
seller, their representatives or vicarious agents.
If Rheonik Messtechnik GmbH breaches essential contractual obligations due to
gross negligence, the extent of the liability shall be restricted to the
typically foreseeable loss. If performance is impossible or in the case of
default, the claims for compensation shall be restricted to reimbursement of
the typically foreseeable loss. If a vicarious agent of Rheonik Messtechnik
GmbH acts in a grossly negligent manner, Rheonik Messtechnik GmbH shall be
liable to the extent of the typically foreseeable loss. The liability in the
case of a breach of essential contractual obligations and in the case of
fraudulent intent on the part of the seller remains unaffected by this. In the
case of physical injury or damage to well-being and health, the applicable
statutory regulations shall apply.
Other applicable documents
The following documents contain additional information regarding how to
operate the RHE46 Coriolis transmitter:
- RHE 40 Desktop Reference (operation manual)
- Tutorial RHEComPro (tutorial)
- RHE20/40 Addendum PID Controller (operation manual)
- RHE40 Addendum Data Logging (operation manual)
- RHE40 Addendum Precision Flow Analysis (operation manual)
- RHE40 Addendum Statistics (operation manual)
- RHE16/20/40 Modbus Map (list of Modbus registers)
- RHM Coriolis sensor (operation manual)
The documents are available to download from the Rheonik Messtechnik GmbH website: ⮫ https://www.rheonik.com.
About this manual
General information
This installation guide contains important information regarding how to
install, start up, and safely operate the RHE46 Coriolis transmitter as
intended. Carefully read the information in this installation guide before
starting any work with or on the RHE46 Coriolis transmitter, and make sure you
have understood everything.
Target group
This installation guide is aimed at operating personnel and at anyone who has
been tasked with the installation and start-up work by the operating
personnel.
Storage
This installation guide is an essential part of the RHE46 Coriolis transmitter
and must be stored such that it always remains accessible to personnel.
Declaration of conformity
The versions of the RHE46 Coriolis transmitter described in this user manual
comply with the applicable standards and guidelines/directives.
The manufacturer’s Declaration of Conformity can be found in section 13 of
this user manual.
Symbols used in this manual
In this manual, sections containing particularly important information are
identified in line with the type of information they contain, e.g.
instructions, lists or references.
Symbol | Meaning |
---|---|
1. → | Instruction; step in a process |
⇒ | Result; outcome of a step |
■ | Lists of entries in no particular order |
⮫ | References to figures, sections etc. |
i | Additional information |
Safety
To ensure that the RHE46 Coriolis transmitter is operated safely and as
intended, it is essential to familiarize yourself with the basic safety
instructions and safety regulations. It is also essential to comply with the
rules and regulations on accident prevention that apply at the installation
location.
This installation guide contains important information regarding how to safely
operate the RHE46 Coriolis transmitter as intended. Carefully read the
information in this installation guide before starting any work with or on the
RHE46 Coriolis transmitter, and make sure you have understood everything.
How warnings are structured and what they mean
When performing different types of work either with the or on the RHE46
Coriolis transmitter, the personnel carry out steps that could entail hazards.
These steps are therefore preceded by a warning.
Warnings are always structured as follows:
- Warning symbol
- Signal word:
- DANGER: high risk of death or serious injury if not prevented
- WARNING: moderate risk of death or serious injury if not prevented
- CAUTION: low risk of death or serious injury if not prevented
- ATTENTION: risk of damage to property if information is not observed
- Description of the type of hazard and how it arises
- Description of the potential consequences of disregarding the hazard
- Description of measures for preventing the hazard
Safety and associated safety measures
DANGER
Risk to life if the information in this installation guide is not observed!
If the information in this installation guide is not observed, this will
result in serious injury or death.
- Read and observe the information in this installation guide before working either with the or on the RHE46 Coriolis transmitter.
- Take the safety instructions and warnings seriously and follow the measures for preventing the hazard.
- Work carefully to prevent accidents resulting in personal injury or damage to property.
- Keep the installation guide safe and make it available to anyone working either with the or on the RHE46 Coriolis transmitter.
Modifications, attachments and conversions
Any modifications to or attachments and conversions on the RHE46 Coriolis
transmitter may make the safety devices less effective or entirely
ineffective, resulting in unforeseeable hazards.
WARNING / ATTENTION
Do not carry out any technical modifications or extensions on the RHE46
Coriolis transmitter.
Risk of injury from slipping, tripping or falling when carrying out
installation work
There is a risk of serious injury from slipping on or tripping over electrical
cables, supply lines, and tools that are lying around.
CAUTION / ATTENTION
Make sure that nobody could trip over or fall because of the electrical cables
when carrying out installation work.
Route electrical cables and supply lines such that nobody could slip on or
trip over them and fall.
Personnel qualifications
WARNING / ATTENTION
Damage to property due to inadequately qualified personnel.
All electrical work must only be performed by people who have in-depth
expertise in electrical engineering (e.g. qualified electricians). These
individuals must provide evidence that they have the requisite expertise for
performing work on electrical systems and the associated components (e.g. by
having passed an examination) and have been trained on the specific product(s)
in question.
Mechanical work must only be performed by appropriately qualified and trained
personnel.
Task | Party responsible | Qualification |
---|---|---|
Installation | Operating personnel | Mechanic |
Work on electrical system | Operating personnel | Electrician |
Start-up, Configuration | Operating personnel | Engineer |
Dismantling work | Operating personnel | Mechanic, Electrician |
Disposal | Waste disposal company | Waste specialist |
Intended use
The RHE46 Coriolis transmitter outputs the measured values for RHM Coriolis
sensors. It can output the mass flow of liquids and gases as well as the
medium density and medium temperature. In addition, the RHE46 Coriolis
transmitter is able to calculate variables such as the volume.
Only install and operate the RHE46 Coriolis transmitter in locations with the
following ambient conditions:
- Ambient temperature -20 to +65 °C
- Relative humidity of 0 to 95% (non-condensing)
- In shade; not exposed to direct sunlight
- In non-hazardous areas – not in Zone 0, 1 or 2, respectively not in class 1, div 1 or 2
Impermissible use
The following are examples of impermissible use of the RHE46 Coriolis transmitter (i.e. use contrary to its intended use):
- Using the RHE46 Coriolis transmitter in locations other than those specified above.
- Using the device in life-support systems in the medical, motor vehicle, aircraft, watercraft or mining industries.
Product description
RHE46-RB (standard version) : RHE46-RJ (version with RJ45 socket)
- RHE46 Coriolis transmitter
- Zero point calibration button
- Hardware lock switch (functionality optional)
- Status LEDs
- Mini-USB port
- QR code (download documentation)
- Terminals for connection to RHM Coriolis sensor
- Terminals for power supply, inputs, outputs and interfaces
- DIN rail lock (top and bottom)
- RJ45 socket for interfaces
- Type label
LED Status Indicator
LED | State | Meaning |
---|
Green
| ****
Off
| RHE transmitter is not ready. The device is either initializing after a
system reset, has no power, or is defective.
On| RHE transmitter in operation.
Flashing| Firmware update running. Do not disconnect power.
Yellow
| Off| Hardware lock switch (Custody Transfer Lock) inactive.
On| Hardware lock switch (Custody Transfer Lock) active.
Flashing
| When the yellow and the red LEDs flash together the Zeroing Process is in progress.
Red
| Off| No errors are present and the instrument is operating as intended.
On
| There is at least one error being reported by the instrument. Details of the exact error(s) can be obtained by reading the error status screens Soft Error Stat and Error Status.
Flashing
| When the yellow and the red LEDs flash together, the Zeroing Process is in progress.
Type label
- Type label
- Manufacturer
- Transmitter type
- Serial number
- Order code
- TAG number
- Permissible power supply
- Permissible ambient temperature
- Protection class
- Sxerial number of corresponding RHM Coriolis sensor
Transportation, storage and check of delivery
Transportation
Always transport the RHE46 Coriolis transmitter in its original packaging.
Storage
Store the RHE46 Coriolis transmitter under the following conditions until it
is installed:
- In the original packaging
- Dust-free, dry and not exposed to direct sunlight
- Storage temperature of -20 to +65°C
Check of delivery
- Check the packaging of the RHE46 Coriolis transmitter for any damage.
- If the packaging is damaged, immediately inform the freight forwarder and your local sales/support representative.
- Check the contents to make sure all the items you ordered are present.
- Check whether the model code on the type label matches the model code on the order.
- If the device comes with protective caps fitted, only remove these just before you start the installation work.
- Keep the packaging safe in case you need to ship the device at a later date.
Installing the RHE46 Coriolis transmitter
Mechanical installation
Mounting on the DIN rail
Open the DIN rail lock at the bottom (1). Place the RHE46 on top of the DIN
rail (2) and press it down (2). Close the DIN rail lock (4) to fix the RHE46
on the DIN
Dismounting from the DIN rail
Open the DIN rail lock (1), tilt the RHE46 forward (2) and lift it off the DIN
rail (3).
Electrical installation
Connecting the RHE46 Coriolis transmitter to the RHM Coriolis sensor
Wire the RHE Coriolis transmitter to the RHM Coriolis sensor according to the
corresponding wiring diagram (QR code on the side of the RHE46), e.g.:
Make sure that you connect the associated RHM to the RHE. See type plate of the RHE for the serial number of the associated RHM at “Set for”.
Connecting the power supply
The RHE46 Coriolis transmitter requires a regulated DC power supply of 12 to
24V ±10%. The power consumption is max. 4W.
- Connect the positive DC power connection of the power supply to the +24V connection (20) and protect it with a fuse (1).
- Connect the negative DC power connection of the power supply to the ground connection (21).
- Connect a protective conductor to a corresponding junction point and to the connection for the protective conductor (22).
Connecting the RS485 interface
The RS485 interface is used for digital communications, for remote control
purposes, and to exchange data. It uses the Modbus protocol to establish a
permanent connection to a monitoring control system.
For the wiring of the RS485 interface, please refer to the wiring diagram (QR
code on the side of the RHE46).
The communication settings of the RS485 interface can be read and changed via
the RHEComPro software, see section 8.
Connecting the Modbus TCP / EtherNet/IP / Profinet interface
The Modbus TCP / EtherNet/IP / Profinet interface is used for digital
communications, for remote control purposes, and to exchange data.
For the wiring of the Modbus TCP / EtherNet/IP / Profinet interface, please
refer to the wiring diagram (QR code on the side of the RHE46).
Once the RHE46 Coriolis transmitter has been connected, it is assigned an IP
address, which is required in order to connect the RHE46 Coriolis transmitter
with your own network. The IP address can be read out using the RHEComPro
software, see section 8.
Connecting the analog output
Active output configuration (A)
- Connect the connection for passive analog output 1 (+) (51) to the +24 V connection (50).
- Connect the analog current input (4 – 20 mA) of the local display / PLC (2) to the connection for passive analog output 1 (-) (52).
- If you are using a 24V DC power supply, we recommend inserting a 510Ω protective resistor upstream of the local display / PLC.
- Connect the ground of your local display / PLC (2) to the ground connection (57).
Passive output configuration (B)
- Connect a suitable power supply (max. 24V DC) to the connection for passive analog output 1 (+) (51).
- Connect the analog current input (4 – 20 mA) of the local display / PLC (2) to the connection for passive analog output 1 (-) (52).
- If you are using a 24V DC power supply, we recommend inserting a 510Ω protective resistor upstream of the local display / PLC.
- Connect the ground of your local display / PLC (2) to the ground connection of the external power source.
Connecting the digital outputs
- A Load to ground (Imax 20mA)
- B Load to +24V (Imax 100mA)
- C Relay to +24V (Imax 100mA)
- 1 Load
- 2 Free-wheeling diode for relay (essential)
- 3 Relay
- 4 Connection terminal
- 5 +24V DC power supply
- 6 Ground
- 31 Connection for active digital output 1
- 32 Connection for active digital output 2
- 37 Ground connection (digital inputs and outputs)
Load to ground (A)
- Connect the input of the connected load (1) to the connection for active digital output 1 (31).
- Connect the output of the load (1) to the ground connection (37).
Load to +24V (B)
- Connect the negative connection of the load (1) to the connection for active digital output 1 (31).
- Connect the +24V power supply (5) to the positive connection of the load
- Connect the ground (6) of the external power source to the ground connection (37) (only if you are using an external +24V power supply (5)). If the power supply is being provided via the RHE46 Coriolis transmitter, do not connect the ground connection.
Relay to +24V (C)
- Connect the negative connection of the relay (3) to the connection for active digital output 1 (31).
- Connect the +24V power supply (5) to the positive connection of the relay (3).
- Connect the load circuit of the relay (3) to the load that needs to be switched.
- Connect the ground (6) of the external power source to the ground connection (37) (only if you are using an external +24V power supply (5)).
- If the power supply is being provided via the RHE46 Coriolis transmitter, do not connect the ground connection.
Connecting the digital input
- A Connection for switch
- B Connection for driver
- 1 Switch
- 2 Driver
- 3 Connection terminal
- 4 +24V power supply
- 20 +24V connection
- 35 Connection for passive digital input 1
- 37 Ground connection (digital inputs and outputs)
Connection for switch (A)
- Connect a suitable +24V power supply (4) to the connection for passive digital input 1 (35).
- Connect the ground of the external +24V power supply (4) to the ground connection (37) (only if you are using an external +24V power supply (4)).
- If the power supply is being provided via the RHE46 Coriolis transmitter, do not connect the ground connection.
Connection for driver (B)
- Connect the output of your driver (2) to the connection for passive digital input 1 (35).
- Connect the ground for the driver (2) to the connection for ground (37) when the ground of the driver (2) and the ground of RHE46 Coriolis transmitter are not connected via the power supply.
- If the power supply is being provided via the RHE46 Coriolis transmitter, do not connect the ground connection.
Zero point calibration
Calibrating the zero point ensures that the measurements are accurate within the permissible tolerance.
- Fill the RHM Coriolis sensor completely with the measuring liquid and make sure that there are no gas bubbles (liquid application) or liquid residues (gas application) in the sensor.
- Flush the sensor until the typical process temperature is reached at the sensor.
- Perform the zero calibration, remind the following points:
- Make sure that the typical process temperature and pressure values are present at the sensor.
- Make sure that there is no flow. In the best case, by closing valves before and after the sensor.
The zero point calibration can be carried out using the button on the front of
the RHE (see figure below) or the RHEComPro communications software.
The zero point calibration is signaled by the flashing of the yellow and red
LED on the RHE and requires a maximum of 30 seconds. As soon as the flashing
stops, the zero point calibration is completed. The flowmeter is now ready for
use.
Configuration
The RHE46 Coriolis transmitter is configured using the RHEComPro software. The
software can be downloaded from the Rheonik website or directly from the
following link:
https://www.rheonik.com/fileadmin/Software/RHEComPro.zip
Please refer to the RHEComPro tutorial for more details on connecting the
transmitter to your PC and using the software.
The RHE46 Coriolis transmitter can be connected to the PC and RHEComPro
software via the mini-USB port on the front of the RHE46 or the RS485
interface at the top terminals.
Accessories
For the RHE46 Coriolis transmitter the following accessories are offered:
Order Code : Description : Picture
-
ARHE-RS Interconnection cable to PC, 5m (Mini-USB to PC USB)
-
ARHE-MO Adapter Modbus RS485 to USB with 2m interface cable
Order code
Returns and disposal
Returns
Do not ship any measuring devices if you are unsure whether you have been able
to fully remove harmful substances, e.g. substances that have penetrated into
cracks/crevices or diffused through plastic.
If the measuring device has been inadequately cleaned, any costs incurred for
disposal or personal injury (e.g. chemical burns) will be charged to the
device operator.
Disposal
The RHE46 Coriolis transmitters are not subject to the WEEE Directive
2012/19/EU on the prevention of waste electrical and electronic equipment and
the reduction of this waste through reuse, recycling and other forms of
recovery.
The RHE46 Coriolis transmitters comply with EU Directive 2011/65/EU on the
restriction of the use of certain hazardous substances in electrical and
electronic equipment. Observe the regulations in force in your country!
Technical data
Housing material | Polycarbonate |
---|---|
Protection class | IP20 / Type 1 |
Ambient temperature | -20…65°C (-4…149°F) |
Relative humidity | 0 – 95% (non-condensing) |
Dimensions | width x height x depth: 72.5 x 110.5 x 47.8 mm (2.86 x 4.35 |
x 1.89 in)
Power supply & power| 12 – 24 V DC ±10%, max. 4W
Analog output (optional)| 1 x 4-20mA output, active and passive, nominal
voltage 24 V DC
Pulse/frequency/status outputs| 2 x configurable pulse/frequency/status
outputs acc. to IEC 60946, max. 10kHz, nominal voltage 24 V DC
Digital input| 1 x configurable control input acc. to IEC 60946, nominal
voltage 24 V DC
Digital data interfaces| Modbus RTU (standard)
Modbus TCP/IPv4, EtherNet/IP, ProfiNet RT (optional)
Declaration of conformity
About Rheonik
Rheonik has but one single purpose: to design and manufacture the very best
Coriolis meters available.
Our research and engineering resources are dedicated to finding new and better
ways to provide cost effective accurate mass flow solutions that provide value
to our customers. Our manufacturing group care for each and every meter we
produce from raw materials all the way to shipping, and our service and
support group are available to help you specify, integrate, start-up and
maintain every Rheonik meter you have in service. Whether you own just one
meter or have hundreds, you will never be just another customer to us. You are
our valued business partner.
Need a specific configuration for your plant? Don’t compromise with a
“standard” product from elsewhere that will add extra cost to your
installation. If we can’t configure it from our extensive and versatile
product range, our exclusive AnyPipeFit Commitment can have your flow sensor
customized with any size/type of process connection and face to face dimension
you need.
No matter what control system you use as the backbone in your enterprise, with
our AnyInterface Commitment, you can be sure that connection and communication
will not be a problem. Alongside a wide variety of discrete analog and digital
signal connections, we can also provide just about any network/bus interface
available (for example: HART, ProfibusDP, ProfiNet, EtherCAT, family of
transmitters. Rheonik RHE 40 Series transmitters can connect to your system –
no headache and no conversion needed.
Rheonik Messtechnik GmbH Rudolf-Diesel-Straße 5
D-85235 Odelzhausen Germany
Tel + 49 (0)8134 9341-0 info@rheonik.com