Kactoily B0CD25ZP24 7 In 1 Aquarium Water Quality Monitor User Manual

May 15, 2024
Kactoily

Kactoily 7-in-1 Aquarium
Water Quality Monitor
User ManualKactoily B0CD25ZP24 7 In 1 Aquarium Water Quality
Monitor

Overview

The Kactoily 7-in-1 Aquarium Water Quality Monitor has a modular design, so users can change the probe when needed. Its power supply plug is powered by 5V1A and a USB (type C) connection.
The application example is as follow:Kactoily B0CD25ZP24 7 In 1 Aquarium
Water Quality Monitor - powered

Monitoring Function

Kactoily B0CD25ZP24 7 In 1 Aquarium Water Quality Monitor - Monitoring
FunctionLCD Off/Confirm button: Press this button on the main interface for 3 seconds and release it to turn off the display light. There will be no impact on data collection after the screen is turned off. The display can be reopened by pressing any button.

Measuring Range

Kactoily 7-in-1 Aquarium Water Qualiy Monitor

Parameter| Range| Resolution| Accuracy
PH| 0-14| 0.01| 0.05
Temperature| 0-60°C
32-140°F| 0.1
0.5| ±0.1
±0.1
Total Dissolved Solids| 0-999ppm
1-200ppt| 1 ppm
1ppt| ±5%F . S
Electrical Conductivity| 0-999us/cm
1-400ms/cm| 1 us/cm
1ms/cm| ±5%F . S
Salinity| 0.01 %-25%| 0.01%-5%(±0.1%)
5.1%-25%(±1%)| ±2%F.S
Specific Gravity| 1-1.222| /| /
ORP| 0-±999mv| 1mv| /

Product Installation

Kactoily B0CD25ZP24 7 In 1 Aquarium Water Quality Monitor - Product
InstallationPlease remove the probe protective cap counterclockwise before use, and attach the protective cap if not in use for a long time.
Simply connect the extension line to the original line and attach it to the display if the original line is not long enough.
The total length of the connecting line is about 50 cm after using the extension line.

  1. Stick the two suction cups in the slots
  2. Attach the magnetic sticker to the back of the screen
  3. Attach the metal plate to the outside of the aquarium where you want to fix the display

Note: please connect the display and probe before powering on.

Product Size(in)

Kactoily B0CD25ZP24 7 In 1 Aquarium Water Quality Monitor - oduct
Size

Please remove the probe protective cap before use and attach the protective cap if not in use for a long time.

Calibration and Bias Correction

If the probe is dirty or has a lot of algae on it after a period of use, please soak it overnight with a certain proportion of diluted disinfectant before calibrating.
Note: The product has been calibrated before delivery, and can be used directly upon receipt. It is recommended to calibrate once every 6 months.

  1. Two-point pH calibrationKactoily B0CD25ZP24 7 In 1 Aquarium Water Quality Monitor -calibrationEnter the setting interface by pressing the middle button, and click the left button to access the first option “pH-set”. Continue to click the left button to the first option “pH-CAL-2”, and the device will enter the third interface (4.01 calibration interface) Putting the probe into the calibration solution 4.01, the measured value indicates the current device value.
    Then, shake a few times and allow the solution to fully contact the probe for 180 seconds, then press the left button to confirm.
    Put the probe in neutral water or pure water once it jumps to the fourth interface (6.86
    calibration interface), then take it out and wipe the remaining liquid off.
    Put the probe into ph6.86 calibration solution and let it stand for 180 seconds (solution and probe are fully in contact). Press the left button to confirm, then remove the probe from the solution and place it in neutral water or pure water. Then dry the remaining liquid on the probe.

  2. Three-point pH calibration (recommended)Kactoily B0CD25ZP24 7 In 1 Aquarium Water Quality Monitor - pointEnter the setting interface by pressing the middle button, and click the left button to access the first option “pH-set”. Continue to click the middle button to the second option “pH-CAL-3”, and then click the left button to confirm, the device will enter the third interface (4.01 calibration interface).
    Putting the probe into the calibration solution 4.01, the measured value indicates the current device value.
    Then, shake a few times and allow the solution to fully contact the probe for 180 seconds, then press the left button to confirm.
    Put the probe in neutral water or pure water once it jumps to the fourth interface (6.86 calibration interface), then take it out and wipe the remaining liquid off.
    Put the probe into ph6.86 calibration solution and let it stand for 180 seconds (solution and probe are fully in contact). Press the left button to confirm, then put the probe in neutral water or pure water once it jumps to the fifth interface (9.18 calibration interface). Then take it out and wipe the remaining liquid off.
    Next, put the probe into ph9.18 calibration solution and let it stand for 180 seconds (solution and probe are fully in contact). Press the left button to confirm, then put the probe in neutral water or pure water once the progress is finished. Then take it out and wipe the remaining liquid off.

  3. pH bias correctionKactoily B0CD25ZP24 7 In 1 Aquarium Water Quality Monitor - pH bias correctionBias correction can be done directly on the monitor if it shows a difference from other pH testing products while other pH testing results are accurate during comparative use.
    Enter the setting interface by pressing the middle button, and click the left button to access the first option “pH-set”. Continue to click the middle button to the option “PH-Adj”, and then click the left button to confirm. Thus, the device will enter the third interface (4.01 calibration interface), and the compensation value is 0 by default.
    The measured value represents the value displayed by the device in the test solution. You can directly press the left or right button“-/ +”to adjust it, and then press the middle button“OK” to save it.

  4. ORPbias correctionKactoily B0CD25ZP24 7 In 1 Aquarium Water Quality Monitor - ORP bias correctionEnter the setting interface by pressing the middle button, and clicking the middle button to access the second option “ORP-set”. Continue to click the left button to enter the third interface “Adj Value”. You could now place the probe into the prepared ORP calibration solution, and shake it a few times to let the solution and the probe make full contact for 180 seconds. And it will show the measured value, which is the current monitor value. If it needs to be adjusted, press the left or right button “-/ +” to make adjustments, then press the middle button “OK” to save it.
    Note: It is not recommended to calibrate if there are no special requirements. This value has been strictly calibrated before leaving the factory. It is easy to produce deviations during calibration due to the effect of water quality on this value. Please follow these requirements strictly if calibration is needed.
    1: Utilize pure water
    2: Dissolve the ORP calibration powder thoroughly in pure water, and let it stand for at least three hours before calibration.

  5. Electrical conductivity calibrationKactoily B0CD25ZP24 7 In 1 Aquarium Water Quality Monitor - conductivity calibrationEnter the setting interface by pressing the middle button, and click the middle button to the third option “EC-set” to access the third interface. Continue to click the middle button to the firstly option“EC-CAL”,and then click the left button to confirm, the device will enter the third interface (1413us/cm calibration interface).
    Putting the probe into the calibration solution 1413us/cm, the measured value indicates the current device value.
    Then, shake a few times and allow the solution to fully contact the probe for 180 seconds, then press the left button to confirm.
    Put the probe in neutral water or pure water once it jumps to the fourth interface (12.88ms/cm calibration interface), then take it out and wipe the remaining liquid off.
    Put the probe into 12.88ms/cm calibration solution and let it stand for 180 seconds (solution and probe are fully in contact). Press the left button to confirm, then put the probe in neutral water or pure water once it jumps to the fifth interface (111.3ms/cm calibration interface). Then take it out and wipe the remaining liquid off.
    Next, put the probe into 111.3ms/cm calibration solution and let it stand for 180 seconds (solution and probe are fully in contact). Press the left button to confirm, then put the probe in neutral water or pure water once the progress is finished. Then take it out and wipe the remaining liquid off.

  6. Electrical conductivity, TDS, salinity, specific gravity bias correctionKactoily B0CD25ZP24 7 In 1 Aquarium Water Quality Monitor - specificPress the middle button of the monitor to enter the setting interface, click the middle button to the third option “EC-set”, left click to confirm to enter the second interface. Click the middle button to the second option “EC-Adj”, and then click the left button to confirm.
    The measured value represents the value displayed by the device in the test solution. You can directly press left or right “-/ +” to adjust it, and then press the middle button “OK” to save it.

  7. Temperature unit switchingKactoily B0CD25ZP24 7 In 1 Aquarium Water Quality Monitor - switchingPress the middle button on the monitor to enter the setting interface. Click the middle button to the fourth option “Temp-set” and click the left button to confirm it. Afterward, press the left button to switch temperature units. You’re then able to choose the suitable unit with the middle button and click the left button to confirm your choice.

  8. Temperature bias correctionKactoily B0CD25ZP24 7 In 1 Aquarium Water Quality Monitor - correctionPress the middle button on the monitor to enter the setting interface. Click the middle button to the fourth option “Temp-set” and click the left button to confirm it. Afterward, press the middle button to the “Com” option, and click the middle button to confirm your choice after the adjustment by pressing the left or right button “-/ +”.
    The compensation value is 0 by default, and the measured value represents the current value of the device displayed in the test solution.

  9. : Buzzer SetupKactoily B0CD25ZP24 7 In 1 Aquarium Water Quality Monitor - Buzzer Setup Note: This feature is being improved and is not available yet, so please stay tuned!

  10. Return to factory settingsKactoily B0CD25ZP24 7 In 1 Aquarium Water Quality Monitor - factory settingsPress the middle button to enter the setting interface. And click the middle button to the sixth option “Reset”, and click the left button to confirm your choice.
    Note: The device can be restored to factory defaults if the sensor has been replaced or if the data display is abnormal. In this case, the calibration data will be deleted. This operation should therefore be performed with caution.

Frequently Asked Questions

Q1: Why haven't measurements changed after the product is powered on?

A1: It’s necessary to unscrew the clear protective cap of the probe before use.

Q2: Why are there some white marks on the probe when I received the product? Has this product been used before?

A2: Not at all. The white marks are from the probe protective solution, as we added a special protective solution to the probe part to protect it. The protective solution is saturated, volatile, and easy to crystallize. Once the solution has overflowed and volatilized, it appears as white crystals that can be used after washing.

Q 3: Is the monitor accurate and do I need to calibrate it after receiving the product?

A 3: Any instrument will have a slight error in long-term use, generally within the normal error range. We have carried out strict calibration before leaving the factory. However, it is not excluded that storage in the long-term turnover process leads to deviation of value, and it is recommended to calibrate before use. Our monitors are recommended to be calibrated every six months. There are detailed calibration steps in the instruction manual.

Q4: Does this AquariumMonitor need to be connected to the application?

A4: The 7-in-1 Aquarium Monitor does not need to be connected to the application. It can be used directly after a simple installation according to the user manual. It is ready for use once it has been installed and calibrated according to the instruction manual.

Q5: How should I proceed if the measurements do not change with different pH buffer solutions when calibrating?

A5: Please click “Reset” to restore factory settings and re-calibrate.

After-sales Service

This product is guaranteed for one year. During the warranty period, if the product is damaged due to its own reasons, it will be repaired or replaced free of charge.
The following circumstances are not covered by the warranty:

  1. Damage caused by irresistible natural disasters;
  2. Damage caused by improper handling, use, maintenance and storage of consumers (including consumer disassembly and repair);
  3. It shows that the water in some parts of the host is seriously causing corrosion of the circuit board or the power supply interface.

Note: The display is not waterproof, please pay more attention to it in daily use