SEJOY DBP-6673B ECG Blood Pressure Monitor Owner’s Manual
- June 4, 2024
- Sejoy
Table of Contents
- SEJOY DBP-6673B ECG Blood Pressure Monitor
- Safety Notice
- Unit Illustration
- Important Testing Guidelines
- Quick Start
- Battery Installation
- System Unit Operation
- BP Unit Operation
- ECG Unit Operation
- Wireless Unit Operation
- Troubleshooting
- Blood Pressure Information
- Blood Pressure Q&A
- Maintenance
- Specifications
- Warranty
- Electromagnetic Compatibility Information
- Additional Notes
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
SEJOY DBP-6673B ECG Blood Pressure Monitor
Safety Notice
Intended Use
ECG Blood Pressure Monitor is intended for non-invasive measuring an adults
and adolecents over 12 years of ageindividual’s systolic, diastolic blood
pressure and heart rate using the oscillometric method. By switching to ECG
measurement, ECG blood pressure monitor intend to display average heart rate
and real-time heart rate, also can view real-time and historical ECG waveforms
in App.
Precautions to Ensure Safe, Reliable Operation
- Do not drop the unit. Protect it from sudden jars or shocks.
- Do not insert foreign objects into any openings.
- Do not attempt to disassemble the unit.
- Do not crush the pressure cuff.
- If the unit has been stored at temperatures below 0 ℃, leave it in a warm place for about 15 minutes before using it. Otherwise, the cuff may not inflate properly.
- If the unit has been stored at temperatures above 40 ℃, leave it in a cool place for about 15 minutes before using it. Otherwise, the cuff may not inflate properly.
- Do not store the unit in direct sunlight, high humidity or dust.
- To avoid any possibility of accidental strangulation, keep this unit away from children and do not drape tubing around your neck.
- Ensure that children do not use the instrument unsupervised; some parts are small enough to be swallowed.
- Some may get a skin irritation from the cuff taking frequent readings over the course of the day, but this irritation typically goes away on its own after the monitor is removed.
- Longest expected contact time between patient and cuff: Each blood pressure measurement should not exceed 3 minutes.
Important Instructions Before Use
- Do not confuse self-monitoring with self-diagnosis. Blood pressure measurements should only be interpreted by a health professional who is familiar with your medical history.
- Contact your physician if test results regularly indicate abnormal readings.
- If you are taking medication, consult with your physician to determine the most appropriate time to measure your blood pressure. NEVER change a prescribed medication without first consulting with your physician.
- Individuals with serious circulation problems may experience discomfort. Consult your physician prior to use.
- For persons with irregular or unstable circulation resulting from diabetes, liver disease, arteriosclerosis or other medical conditions, there may be variations in blood pressure values measured at the wrist versus at the upper arm. Monitoring the trends in your blood pressure taken at either the arm or the wrist is nevertheless useful and important.
- People suffering from vascular constriction, liver disorders or diabetes, people with cardiac pacemakers or a weak pulse, and women who are pregnant should consult their physician before measuring their blood pressure themselves. Different values may be obtained due to their condition.
- People suffering from arrhythmias such as atrial or ventricular premature beats or atrial fibrillation only use this ECG Blood Pressure Monitor in consultation with your doctor. In certain cases oscillometric measurement method can produce incorrect readings.
- Too frequent measurements can cause injury to the patient due to blood flow interference.
- The cuff should not be applied over a wound as this can cause further injury.
- DO NOT attach the cuff to a limb being used for IV infusions or any other intravascular access, therapy or an arterio-venous (A-V) shunt. The cuff inflation can temporarily block blood flow, potentially causing harm to the patient.
- The cuff should not be placed on the arm on the side of a mastectomy. In the case of a double mastectomy use the side of the least dominant arm.
- Pressurization of the cuff can temporarily cause loss of function of simultaneously used monitoring equipment on the same limb.
- A compressed or kinked connection hose may cause continuous cuff pressure resulting in blood flow interference and potentially harmful injury to the patient.
- Check that operation of the unit does not result in prolonged impairment of the circulation of the patient.
- Product is designed for its intended use only. Do not misuse in any way.
- Product is not intended for infants or individuals who cannot express their intentions.
- Prolonged over-inflation of the bladder may cause ecchymoma of your arm.
- Do not disassemble the unit or arm cuff. Do not attempt to repair.
- Use only the approved arm cuff for this unit. Use of other arm cuffs may result in incorrect measurement results.
- The system might produce incorrect readings if stored or used outside the manufacturer’s specified temperature and humidity ranges. Make sure to store the ECG Blood Pressure Monitor , children, pets and pests are outside of accessible range.
- Do not use the device near strong electrical or electromagnetic fields generated by cell phones or other devices, they may cause incorrect readings and interference or become interference source to the device.
- Do not mix new and old batteries simultaneously.
- Replace batteries when Low Battery Indicator“ ”appears on screen. Replace both batteries at the same time.
- Do not mix battery types. Long-life alkaline batteries are recommended.
- Remove batteries from device when not in operation for more than 3 months.
- Dispose batteries properly; observe local laws and regulations.
- Advising operator that Instruction manual/ Booklet must be consulted.
- Do not use the device during transport vehicles for influencing measurement accuracy, such as patient transport in an ambulance or helicopter.
- Contains small parts that may cause a chocking hazard if swallowed by infants.
- Please align the polarities of each battery with the +ve and -ve signs imprinted on the battery housing when you replace the batteries .
- Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
- Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the unit, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
- Do not touch the electrode of this product to other conductors (including grounding).
- During ECG measurement, if your skin or hands are too dry, please moisten it with a damp towel. Implement measurement.
- If the skin or finger is damaged and bleeding during electrocardiogram measurement, please replace the finger for measurement
- During ECG measurement, please do not use the product in reverse directions with your left and right hands.
- The performance of an automatic ECG sphygmomanometer may be affected by extreme temperature, humidity and altitude.
Discard the used product to the recycling collection point according to local
regulations
The Bluetooth® Smart word mark and logos are registered trademarks owned by
Bluetooth SIG, Inc. and any use of such marks by JOYTECH Healthcare Co.,Ltd.
Federla Commulcation Commission (FCC) Interference Statement
- This device complies with part 15 of the FCC Rules. Operation is subject to the condition that this device does not cause harmful interference.
- This device is verified to comply with part 15 of the FCC Rules for use with cable television service.
- This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
- This device may not cause harmful interference, and
- this device must accept any interference received, including interference that may cause undesired operation. Please note that changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
- This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
- This equipment complies with radio frequency exposure limits set forth by the FCC for an uncontrolled environment.
- This device must not be co-located or operating in conjunction with any other antenna or transmitter.
- Essential performance:
Electrosurgery interference recovery|
Refer 202.6.2.101 IEC 80601-2-30
---|---
Limits of the error of the manometer| Refer 202.12.1.102 IEC 80601-2-30
Reproducibility of the BLOOD PRESSURE DETERMINATION|
Refer 201.12.1.107 IEC 80601-2-30
Reproducibility of the ECG DETERMINATION|
Refer 201.12.1.101.3.1 IEC 60601-2-47
Unit Illustration
Monitor Unit
Arm Cuff Medium size cuff (fits arm circumference: 22.0 cm -36.0 cm).
If air is leaking from the arm cuff, replace the arm cuff with a new one. It is generally recomm ended to have the cuff replaced timely to ensure correct functioning and accuracy. Please consult your local authorized Sejoy distributor or dealer.
Display
Contents
Important Testing Guidelines
- Avoid eating, exercising, and bathing for 30 minutes prior to testing.
- Sit in a calm environment for at least 5 minutes prior to testing.
- Do not stand while testing. Sit in a relaxed position while keeping your arm level with your heart.
- Avoid speaking or moving body parts while testing.
- While testing, avoid strong electromagnetic interference such as microwave ovens and cell phones.
- Wait 3 minutes or longer before re-testing.
- Try to measure your blood pressure at the same time each day for consistency.
- Test comparisons should only be made when monitor is used on the same arm, in the same position, and at the same time of day.
- This ECG Blood Pressure Monitor is not recommended for people with severe arrhythmia.
- Do not use this ECG Blood Pressure Monitor if the device is damaged.
Any blood pressure recording can be affected by the following factors:
- The position of the subject, his or her physiologic condition;
- The performance and accuracy of the device;
- Cuff size: too small cuff (bladder) will produce a higher blood pressure value than usual, too big cuff (bladder) will produce a lower blood pressure value;
- Measuring position does not keep level with your heart;
- Speaking or moving body parts while testing;
- Not relaxing for about 5 minutes before taking the measurement.
Quick Start
BP Quick Start
- Install batteries
- Insert cuff air plug into the behind side of the monitor unit. (See Figure B)
- Remove thick clothing from the arm area.
- Rest for several minutes prior to testing. Sit down in a quiet place comfortably, back and arm support on a desk or table, with your legs uncrossed , your arm resting on a firm and your feet flat on the floor. (See Figure C )
- Apply cuff to your left arm and middle of the cuff at the level of your heart. Bottom of cuff should be placed approximately 1-2cm (1/2”) above elbow joint. (See Figures D&E)
- 6. Press ” START/STOP ” Button to start testing.
ECG Quick Star
- Install batteries. (See Figure A )
- Before measuring the device, rinse your hands with water or wipe them with a wet towel to keep the contact points moist.(See Figure B )
- Rest for several minutes prior to testing. Sit down in a quiet place, preferably at a desk or table, with your arm resting on a firm surface and your feet flat on the floor. (See Figure C )
- Hold down the to bind the measurement device to the smart device, press to power off the device, and then press to start the test. ” MEM” ” START/STOP ” ” START/STOP ” ECG Quick Start
- Press the palm of both hands on the electrode with moderate strength, not too tight (note: keep the palm of both hands on the electrode, sit still, relax, do not move or talk during the measurement)(See Figure D)
Battery Installation
Battery Installation
Slide battery cover off as indicated by arrow. Install 3 new AAA alkaline
batteries according to polarity. Close battery cover.
Battery level indicator
When the device is turned on, the icon of the current battery quantity is
displayed. When the battery quantity is too low, the device indicates low
battery quantity and the screen displays . When the battery is fully charged,
the screen displays
System Unit Operation
System Settings
With power off, press button to activate System Settings. The Memory Group
icon flashes.
-
Select Memory Group
While in the System Setting mode, you may accumulate test results into 2 different groups. This allows multiple users to save individual test results (up to 60 memories per group.) Press ” MEM ” button to choose a group setting. Test results will automatically store in each selected group -
Time/Date setting
Press button again to set the Time/Date mode. Set the year first by adjusting the “MEM” button. Press button again to confirm current month. Continue setting the dat, hour and minute in the same way. Every time the button is pressed, it will lock in your selection and continue in succession ( month,day,hour, minute,12/24 hours) -
Time Format setting
Press button again to set the time fornat setting mode.Set the time format by adjusting the“MEM” button. EU means European Time US means U.S Time -
Voice Setting
Press button to enter voice setting mode. Set voice format ON or OFF by pressing the “MEM” button -
Volume Setting
Press button to enter volume setting mode. Set the voice volume by adjusting the “MEM” button . There are six volume levels. -
Saved Settings
While in any setting mode, press ” START/STOP ” button to turn the unit off. All information will be saved.
BP Unit Operation
Applying the Arm Cuff
- Firmly insert air plug into opening located on behind side of monitor unit.
- With sticky nylon section facing outward, insert end of cuff underneath metal ring of cuff
- Fasten cuff about 1-2cm (0.4-0.8”) above the elbow joint. For best results apply cuff to bare arm and keep level with heart while testing
Note: Do not insert air plug into opening located on right side of monitor unit. This opening is designed for an optional power supply only.
Blood pressure test
-
Power On
Press and hold “START/STOP” button to turn the unit on. Press and hold “START/STOP” button to turn the unit on. The LCD screen will display for one second. A voice tone will indicate when unit is ready for testing. Note: Make sure the switch is in BP mode before the measurement begins Unit will not function if residual air from previous testing is present in cuff. The LCD will flash “ ” until pressure is stabilized. -
Testing for Blood
After cuff inflation, air slowly rise as indicated by the corresponding cuff pressure value. A flashing will appear simultaneously on screen signaling heart beat detection. -
Result Display
The screen will display measurements for systolic and diastolic blood pressure with voice broadcast. An indicator representing the current measurement will appear next to the corresponding WHO Classification. Note: Refer to Page 32~33 for detail WHO Blood Pressure Classification Information.
Irregular Heartbeat Indicator
If the monitor detects an irregular heart rhythm two or more times during the
measuring process, the Irregular Heartbeat Symbol “ “appears on screen along
with measurement results. Irregular heartbeat rhythm is defined as rhythm that
is either 25% slower or faster than the average rhythm detected while
measuring systolic blood pressure and diastolic blood pressure. Consult your
physician if the Irregular Heartbeat Symbol ” ” frequently appears with your
test results.
Safety Precaution: If pressure in arm cuff becomes too extreme while
testing, press the ” START/STOP ” button to turn power off. The cuff pressure
will rapidly dissipate once the unit is off.
Power Off
The ” START/STOP ” button can be pressed to turn off the unit in any mode.
The unit can turn off the power itself about 3 minutes no operation in any
mode.
Arm Shake Indicator
If there is arm movement during the measurement, ” ” may be shown.
Indicates that it may lead to abnormal accurate measurement results.The
measurement result shows “ “. When viewing this memory, the memory result
shows” ”
Cuff loose Indicator
When starting the measurement, “ “will be displayed when the cuff is
properly wound. When the cuff is too loose, ” “will be displayed. At this
time, please wear the cuff correctly and start measuring again.
Last 3 Tests Average
With power off,press the “MEM” button to activate screen display. After the
unit performs a self-diagnosis,the screen will display the average test
results from the last 3 readings of the last group used. The “AVG” symbol will
appear along with the corresponding WHO Blood Pressure Indicator. The Memory
Check mode can be accessed by pressing “MEM” button. To check the average
results from other groups, select the desired group first prior to activating
button in the off position.(See “Select Memory Group” on Page 13)
BP Memory Check
You may check past test results by using the “MEM” button. The most recent
test result and oldest test result in memory can be viewed by pressing and
holding the “MEM” button. Upon activating test results. you can press the
“MEM” button to scroll through all test results stored in memory.
BP Memory Deletion
Memory for a selected group may be deleted while in Memory Check mode.
Press and hold the button for approximately 3 seconds to delete all memory
records from the selected group with voice broadcast “Memory Clear” and then
transfer into testing mode. Press the “START/STOP ” button to turn the unit
off.
Low Battery Indicator
The unit will broadcast “Low Battery” when battery life is depleting and
unable to inflate cuff for testing. The ” ” appears simultaneously for
approximately 5 seconds prior to shutting off. Replace batteries at this time.
No memory loss will occur throughout this process.
Static Pressure Measurement
In the power down state, press and hold the ” START/STOP” button, and
theninstall the batteries. Until the LCD screen is full, release the ”
START/STOP” button. When the LCD screen displays the double zero, the ECG
blood pressure meter is in static state.
Note: Only Service personnel permitted to access to this mode, the mode unavailable in normal use.
ECG Unit Operation
Gestures of ECG measurements
ECG Tes
-
Power On
Press and hold “START/STOP” button to turn the unit on. The LCD screen will display for one second -
Testing for ECG
Wait for the signal to stabilize,and the 5S countdown starts to count.Device side display (See Figure A), APP side page display (See Figure B). -
A flashing ” ” will appear simultaneously on screen signaling heart beat detection and The buzzer buzzes.Displays heart rate values and saves 30 seconds of ECG data to the device. Device side display (See Figure A), APP side page display (See Figure B). Note: Keep your fingers on the device
-
Result Display
The screen will display measurements for heat rate. Suggest this measurement heart rate is normal or too fast or too slow.Device side display (See Figure A), APP give suggested symptoms of atrial fibrillation and abnormal heart rhythms(See Figure B).
Atrial Fibrillation Indicator
If the ECG sphygmomanometer detects an abnormality in the RR interval during
the measurement, it also detects different sizes, shapes, F waves with uneven
orientation and spacing. The atrial fibrillation symbol is displayed on the
screen along with the measurement results. If is a common sign in your test
results, consult your doctor.
Heart Rate Indicator
Ifat the end of the measurement, when the 30s average heart rate is greater
than 120beat/min the tachy symbol and ecg results are displayed on the LCD,
when the 30s average heart rate is less than 60beat/min, When the average
heart rate is between 60 and 120beat/min for 30s, the normal heart rate symbol
and ecg result will be displayed on the LCD screen together”
Power Off
The ” START/STOP ” button can be pressed to turn off the unit in any mode. The
unit can turn off the power itself about 3 minutes no operation in any mode.
ECG Memory Check
You may check past test results by using the “MEM” button. The most recent
test result and oldest test result in memory can be viewed by pressing and
holding the “MEM” button. Upon activating test results. you can press the
“MEM” button to scroll through all test results stored in memory.
ECG Memory Deletion
Memory for a selected group may be deleted while in Memory Check mode.
Press and hold the ” SET ” button for approximately 3 seconds to delete all
memory records from the selected groupwith voice broadcast “Memory Clear” and
then transfer into testing mode. Press the “START/STOP ” button to turn the
unit off.
Use of APP
- Search for “BP+ECG” in the Android App Store or Apple App Store to download, register, and log in.
- If you want to use Bluetooth connection, please turn on the Bluetooth function of your phone. If you want to use wifi connection, please turn it on Your phone’s Bluetooth and wifi capabilities. And your phone needs to be connected to a router
- See the App Operation guide for other operations
Wireless Unit Operation
Bluetooth requirements
- The monitor requires a device with:
- Bluetooth 4.0 or later
- Android 5.0 or later
- IOS 9.0 or later
- And works with:
- iphone , iPod, iPad
- Android Phones and Tablets
Bluetooth connection
- Using for the first time
1. Search for “BP+ECG” in the Android App Store or Apple App Store to download, register, and log in.
2. Open the App on your phone or tablet. If requested, you should enable Bluetooth on your device. You can enable Bluetooth under the Settings menu on your smart phone or table.
3. Create a new user login, or login with your existing user name and password.
4. Selection device “ECG Blood Pressure Monitor
- Pairing your monitor with a Smart Device
1. Open the “ ECG Blood Pressure Monitor ” and follow the pairing instructions shown on your smart phone.The date and time on your monitor will automatically be set when you pair it with your smart device.
2. Confirm that your monitor is connected successfully.When your monitor is connected successfully to your smart phone, it will be display like below.
3. Press the BP [START/STOP] button to turn your monitor off.
Wifi requirements
- The monitor requires a device with:
- Android 5.0 or later
- IOS 9.0 or later
- And works with:
- iphone , iPod, iPad
- Android Phones and Tablets
Wifi connection
- Using for the first time
- Open the App on your phone or tablet. If requested, you should enable Bluetooth and wifi on your device. You can enable Bluetooth and wifi from the Settings menu on your smartphone or desk, and your phone is connected to a router.
- Pairing your monitor with a Smart Device
- Open the “ECG Blood Pressure Monitor” and follow the pairing instructions shown on your smart phone.The date and time on your monitor will automatically be set when you pair it with your smart device.
- Confirm that your monitor is connected successfully.When your monitor is connected successfully to your smart phone, it will be display like below.
- Open the App on your phone or tablet, Click the network configuration button and input the router password currently connected to the intelligent device to connect the ECG sphygmomanometer to the router. When your monitor is connected successfully to your router, it will be display like below.
- Press the BP [START/STOP] button to turn your monitor off.
- Using again
- Open the App on your phone or tablet. If requested, you should enable wifi on your device, which is connected to a router. You can enable wifi on your smartphone or desktop under the Settings menu
- Pairing your monitor with a Smart Device
- Open the “ECG Blood Pressure Monitor” and follow the pairing instructions shown on your smart phone.The date and time on your monitor will automatically be set when you pair it with your smart device
- Confirm that your monitor is connected successfully. When your monitor is connected successfully to your smart phone, it will be display like below.
- Press the BP [START/STOP] button to turn your monitor off.
Troubleshooting
Problem
|
Possible Cause
|
Solution
---|---|---
Blood pressure results are not within typical range
|
Cuff is too tight or not properly positioned on the arm
| Firmly reposition cuff approximately1-2cm (1/2″) above the elbow joint
( See Page 15)
Inaccurate test results due to body movement or monitor movement
|
Sit in a relaxed position with arm placed near heart. Avoid speaking or moving body parts while testing. Make sure the monitor unit is placed in a stationary position throughout the testing period.
( See Page 10)
Ecg waveform drift or clutter is large
|
1. Skin too dry or greasy
2. The electrode is not secure in contact with the human body 3.Muscle tension
|
1.Clean skin with soap or water to make it moist and grease-free 2.Apply pressure to the electrode 3.Relax the hand during measurement
” “displayed
|
Cuff fails to inflate properly
|
Make sure hose is properly fastened to cuff and monitor unit
Improper operation
|
Read user manual carefully and re-test properly.
Pressurization is over cuff rated pressure 300mmHg|
Read user manual carefully and re-test properly.
Blood Pressure Information
Blood Pressure
Blood pressure is the force of blood pushing against the walls of arteries.
It is typically measured in millimeters of mercury (mmHg.) Systolic blood
pressure is the maximum force exerted against blood vessel walls each time the
heart beats. Diastolic blood pressure is the force exerted on blood vessels
when the heart is resting between beats. An individual’s blood pressure
frequently changes throughout the course of a day. Excitement and tension can
cause blood pressure to rise, while drinking alcohol and bathing can lower
blood pressure. Certain hormones like adrenaline (which your body releases
under stress) can cause blood vessels to constrict, leading to a rise in blood
pressure If these measuring numbers become too high, it means the heart is
working harder than it should.
WHO Blood Pressure Classification Indicator
The DBP-6673B is equipped with a classification indicator based on established
guidelines from the World Health Organization. The chart below (color coded on
monitor unit) indicates test results.
Health Reminder
Hypertension is a dangerous disease that can affect the quality of life. It
can lead to a lot of problems including heart failure, kidney failure, and
cerebral hemorrhaging. By maintaining a healthy lifestyle and visiting your
physician on a regular basis, hypertension and relative diseases are much
easier to control when diagnosed in their early stages.
Note: Do not be alarmed if an abnormal reading occurs. A better indication of an individual’s blood pressure occurs after 2-3 readings are taken at the same time each day over an extended period of time. Consult your physician if test results remain abnormal.
Blood Pressure Q&A
Q: What is the difference between measuring blood pressure at home or at a
professional healthcare clinic?
A: Blood pressure readings taken at home are now seen to give a more
accurate account as they better reflect your daily life. Readings can be
elevated when taken in a clinical or medical environment. This is known as
White Coat Hypertension and may be caused by feeling anxious or nervous.
Note: Abnormal test results may be caused by:
-
Improper cuff placement
Make sure cuff is snug-not too tight or too loose. Make sure bottom of the cuff is approximately 1-2cm (1/2”) above the elbow joint. -
Improper body position
Make sure to keep your body in an upright position. -
Feeling anxious or nervous
Take 2-3 deep breaths, wait a few minutes and resume testing.
Q: What causes different readings?
A: Blood pressure varies throughout the course of a day. Many factors
including diet, stress, cuff placement, etc. may affect an individual’s blood
pressure.
Q: Should I apply the cuff to the left or right arm? What is the
difference?
A: Either arm can be used when testing, however, when comparing results,
the same arm should be used. Testing on your left arm may provide more
accurate results as it is located closer to your heart.
Q: What is the best time of day for testing?
A: Morning time or any time you feel relaxed and stress free.
Maintenance
- Avoid dropping, slamming, or throwing the unit.
- Avoid extreme temperatures. Do not expose unit directly under sunshine.
- When cleaning the unit, use a soft fabric and lightly wipe with mild detergent.
Use a damp cloth to remove dirt and excess detergent.When the electrode surface is dirty, please wipe it with a damp cloth or alcohol cotton, at least once a month.
Specifications
Product Description | Arm-type Fully Automatic ECG Blood Pressure Monitor |
---|---|
Model | DBP-6673B |
Display | LCD Digital Display Size:95.5mm×54.5mm (3.76″ x 2.15″) |
BP Measurement Method | Oscillometric Method |
BP Pressurization | Automatic Pressurization |
BP Measurement Range
| Systolic Pressure| 60mmHg~260mmHg
Diastolic Pressure| 40mmHg~200mmHg
Pressure| 0mmHg~299mmHg
Pressure| ±3mmHg
Pulse| 30 ~ 180 Beats/Minute
Pulse| ±5%
ECG Measurement Method| Single-channel ECG
ECG Measurement Range
| bandwidth| 0.67~40HZ
Heart Rate| 30 ~ 199 Beats/Minute
Heart Rate| ±5%
Measuring Time| 30 Seconds
Memory
| 2×150 BP Memories in Two Groups with Date and Time
2×20 ECG Memories in Two Groups with Date and Time
BP Function
| Irregular Heartbeat Detection
WHO Classification Indicator
Last 3 Tests Average
ECG Function
| Atrial Fibrillation Indicator
Heart Rate Indicator
Other Function
| Automatic Power-Off
Low Battery Detection
Voice
Backlight
Power Source| 3A AA batteries
Battery Life| Approximately 2 months at 3 tests per day
Unit Weight| Approx.258g (9.1 oz.) (excluding battery)
---|---
Unit Dimensions| Approx.142.5 x 107.2 x 44mm (5.61″ x 4.22″ x 1.73″ )(L x W x
H)
Cuff Circumference| Approx.135 (W) x 485(L) mm
(Medium cuff: Fits arm circumference 22-36 cm)
Operating Environment
| Temperature| 10℃ ~ 45℃ (50℉~113℉)
Humidity| 10% ~ 95%RH
Pressure| 800hPa~1060hPa
Storage Environment
| Temperature| -20℃~55℃ (-4℉~131℉)
Humidity| 15% ~ 93%RH
Transport Environment
| Temperature| -20℃~55℃ (-4℉~131℉)
Humidity| 15% ~ 93%RH
Bluetooth
| Frequency| 2. 4GHz( 2402- 2480Mhz)
Type Of Antenna| Built-in Onboard Antenna
Version| V5.0
Speed| 1Mbps
Transmission power| About 3dbm
WIFI
| Frequency| 2. 4GHz( 2412- 2462Mhz)
Communication Standard| 802. 11b
Bandwidth| About 20 MHZ
Type Of Antenna| Built-in Onboard Antenna
Antenna gain| About 1.5dBi
Transmission power| ≤15dbm
Classification| Internal Powered Equipment,Type BF ,Cuff is the Applied Part
Ingress Protection Rating| Ip21, Indoor Use Only
Warranty
Specifications are subject to change without notice.
This ECG Blood Pressure Monitor complies with the European regulations and
bears the
CE mark“CE 0123”.This ECG Blood Pressure Monitor also complies with mainly
following standards
(included but not limited):
Safety standard:
EN 6060: 1 Medical electrical equipment part 1: General requirements for
safety
EMC standard:
EN 60601: 1-2 Medical Electrical Equipment — Part 1-2: General
Requirements For Basic
Safety And Essential Performance — Collateral Standard: Electromagnetic
Disturbances – Requirements And Tests.
Performance standards:
IEC80601: 2-30, Medical electrical equipment – Part 2-30: Particular
requirements for the basic safety and essential performance of automated non-
invasive sphygmomanometers.
EN 1060: 3 Non-invasive sphygmomanometers – Supplementary requirements
for electromechanical blood pressure measuring systems. ISO 81060-2, non-
invasive sphygmomanometers – part 2: clinical validation of automated
measurement type.
Please note that changes or modifications not expressly approved by the party
responsible for compliance could void the user’s authority to operate the
equipment.
This equipment has been tested and found to comply with the limits for a Class
B digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and
on, the user is encouraged to try to correct the interference by one or more
of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
This device contains licence-exempt transmitter(s)/receiver(s) that comply with Innovation, Science and Economic Development Canada’s licence-exempt RSS(s). Operation is subject to the following two conditions:
- This device may not cause interference; and
- This device must accept any interference, including interference that may cause undesired operation of the device.
The ECG Blood Pressure Monitor is calculated from the date of purchase, and its service life is 2 years. If the ECG Blood Pressure Monitor does not function properly due to defective components or poor workmanship, we will repair or replace it freely. The warranty does not cover damages to your Blood Pressue Monitor due to improper handling. Please contact local retailer for details.
Electromagnetic Compatibility Information
The device satisfies the EMC requirements of the international standard IEC 60601-1-2. The requirements are satisfied under the conditions described in the table below. The device is an electrical medical product and is subject to special precautionary measures with regard to EMC which must be published in the instructions for use. Portable and mobile HF communications equipment can affect the device. Use of the unit in conjunction with non-approved accessories can affect the device negatively and alter the electromagnetic compatibility. The device should not be used directly adjacent to or between other electrical equipment.
Table 1: Guidance and declaration of manufacturer-electromagnetic
emissions
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in
such an environm-ent.
Guidance and declaration of manufacturer-electromagnetic emissions
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in
such an environm- ent.
Emissions test| Compliance| Electromagnetic environment
– guidance
Radiated emission CISPR 11
| ****
Group 1, class B.
| The device uses RF energy only for its internal function. Therefore, its emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Conducted emission CISPR 11
| ****
Group 1, class B.
| The device is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Table 2
Guidance and declaration of manufacturer-electromagnetic immunity
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in
such an environm- ent.
IMMUNITY test| IEC 60601
test level
| Compliance level| Electromagnetic environment
-guidance
Electrostatic discharge (ESD) IEC 61000-4-2
| ± 8 kV
contact
±2 kV,±4 kV,
±8 kV,
±15 kV air
| ± 8 kV
contact
±2 kV,±4 kV,
±8 kV,
±15 kV air
| Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30 %.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
| ****
30 A/m; 50Hz or 60Hz
| ****
30 A/m; 50Hz or 60Hz
| Power frequency magnetic fields should be at levels charactertic of a typical location in a typical comme- rcial or hospital environment.
Radiated RF EM fields
IEC 61000-4-3
| ****
3V/m or 10 V/m 80MHz-2.7
Ghz 80%AM at 1kHz
| ****
3V/m or 10 V/m 80MHz-2.7
Ghz 80%AM at 1kHz
| ****
Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended seperation distance 80 MHz to 800 MHz 800 MHz to 2.7 Ghz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol:
Table 3
Guidance and declaration of manufacturer-electromagnetic immunity|
---|---
Nowadays, many RF wireless equipments have being used in various healthcare
locations where medical equipment and/or systems are used. When they are used
in close proximity to medical equipment and/or systems, the medical equipment
and/or systems’ basic safety and essential performance may be affected. Arm-
type Fully Automatic Digital ECG Blood Pressure Monitor has been tested with
the immunity test level in the below table and meet the related requirements
of IEC 60601-1-2:2014. The customer and/or user should help keep a minimum
distance between RF wireless communications equipment and this medical
equipment and/or systems as recommended below.|
Test frequency (MHz)| Band (MHz)| ****
Service
| ****
Modulation
| Maximum power (W)| Distance (m)| Immunity test level (V/m)|
385| 380-390| TETRA 400| Pulse
modulation 18Hz
| 1.8| 0.3| 27|
450| 430-470| GMRS 460
FRS 460
| FM
± 5 kHz deviation 1 kHz sine
| 2| 0.3| 28|
710| ****
704-787
| ****
LTE
Band 13, 17
| ****
Pulse modulation
217Hz
| ****
0.2
| ****
0.3
| ****
9
|
745|
780|
810| | GSM 800/900,| ****
Pulse modulation
18Hz
| ****
2
| ****
0.3
| ****
28
|
800-960| TETRA 800,
iDEN 820,
|
870|
| CDMA 850,|
930|
| LTE Band 5|
1720| | ****
GSM 1800;
| ****
Pulse modulation
217Hz
| | | |
| CDMA 1900;| | | |
1845|
1700-1990| GSM 1900; DECT;| 2| 0.3| 28|
| LTE Band 1, 3,| | | |
|
| | 4, 25; UMTS| | | |
1970| | | | | |
2450| 2400-2570| Bluetooth,WLAN,
802.11 b/g/n,RFID 2450,LTE Band 7
| Pulse modulation 217Hz| 2| 0.3| 28|
5240| ****
5100-5800
| ****
WLAN 802.11
a/n
| ****
Pulse modulation
217Hz
| ****
0.2
| ****
0.3
| ****
9
|
5500|
5785|
Table 4
Recommended separation distances between portable and mobile RF
communications equipment and the device
The device is intended for use in an electromagnetic environment in which
radiated therefore disturbances are controlled. The customer or the user of
the device can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment
(transmitters) and the device as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum output power of transmitter
W
| Separation distance according to frequency of transmitter m
80 MHz to 800 MHz
| 800 MHz to 2.7 GHz
0.01| 0.12| 0.23
0.1| 0.38| 0.73
1| 1.2| 2.3
10| 3.8| 7.3
100| 12| 23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Additional Notes
Important Instructions Before Use
-
WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
-
WARNING: PORTABLE RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of Arm-type Fully Automatic Digital ECG Blood Pressure Monitor , including cables specified by the MANUFACTURER. Otherwise, degradation of the performance of this equipment could result.
-
The software identifier refer to the software evaluation report , and the file code is JDBP-7304- 154.
-
verify manometer pressure accuracy: In the power down state, press and hold the ” START/STOP” button, and theninstall the batteries. Until the LCD screen is full, release the ” START/STOP” button. When the LCD screen displays the double zero, the bloodpressure meter is in static state. At this point, 500ml gas capacity, calibrated standard pressure gauge and manual pressure device can be connected to the sphygmomanometer through the sleeve interface of the sphygmomanometer, and manual pressure can be applied to the effective display range of the sphygmomanometer, and then the difference between the reading of the sphygmomanometer and that of the standard pressure gauge can be compared.This mode can be used to verify manometer pressure accuracy.
-
The patient is the operator: the PATIENT is an intended OPERATOR. the PATIENT Do not carry out other maintenance operations except to replace the battery.
-
WARNING: Do not modify this equipment without authorization of the manufacturer.
-
ESSENTIAL PERFORMANCE Maintenance advice:
Pressure calibration will be carried out when this product leaves the factory. Patients can use the method described in the section “Verify Manometer Pressure Accuracy” to verify the accuracy. If the accuracy deviation is large, please contact the manufacturer to recalibration. -
Mechanical strength and resistance to heat: The resistance to heat will be retained by device during the EXPECTED SERVICE LIFE of the ME EQUIPMENT.
-
Do not place the ECG Blood Pressure Monitor and cuff at will. It will cause asphyxiation if the child swallows or twine around his neck.
-
The cuff and the case of the ECG Blood Pressure Monitor have been tested for biocompatibility and do not contain allergenic or harmful materials.Please stop using it if allergy occurs during use.
-
Warning: Non-professionals do not modify the equipment, otherwise it will make the equipment measurement is not accurate.
- Do not expose the equipment for a long time, otherwise it will reduce the performance of the equipment.
- This device is not used for children and pets
-
Clean:
The equipment can be cleaned by lay operator according to rule 3 of maintenance in the instructions -
Warning: Do not use a damaged cuff for blood pressure measurement.
- When measuring with the cuff, if the tester feels seriously uncomfortable, press the button of the ECG Blood Pressure Monitor to deflate the cuff, or remove the cuff directly from the arm.
- If an unexpected reading occurs, the operator can take several more measurements and consult a doctor.
- This equipment is used outside the specified environment, may damage the equipment, and may be inaccurate measurement.
- ME equipment not intended for use in conjunction with flammable agents “ME equipment not intended for use in oxygen rich environment”
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product indicates that it should not be disposed
with other household waste at the end of its life. To prevent potential harm
to the environment or to human health, please separate this product from other
types of wastes and recycle it responsibly. When disposing this type of
product, contact the retailer where product was purchased or contact your
local government office for details regarding how this item can be disposed in
an environmentally safe recycling center.
Business users should contact their supplier and check the terms and
conditions of the purchasing agreement. This product should not be mixed with
other commercial wastes for disposal. This product is free of hazardous
materials.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>