VARON VP-1 Pulse Flow Oxygen Concentrator User Manual

June 17, 2024
VARON

VARON VP-1 Pulse Flow Oxygen Concentrator

Foreword

Dear users,
This manual introduces the functional features, operation steps, precautions and maintenance of the product in detail. To ensure that you can use this product more effectively, please read this manual carefully before use; Please keep it carefully for unexpected needs. Please don’t open the package before you use it, otherwise, it will lose the qualification of warranty. Please note that our company reserves the right to update the products and technologies. Some illustrations in the manual may be different from what you see in this product. Therefore, please refer to the actual products you have received.

SAFETY PRECAUTIONS

  • To prevent power failure or equipment failure, the oxygen users who in urgent need please assist with other standby oxygen supply devices (such as Oxygen cylinders. This product can not be used for life support or life extension! Not suitable for newborns or infants!
  • Oxygen can support combustion, Do not smoke when the equipment works!
  • Do not use the equipment when there are high temperatures, sparks and fire in the air!
  • To avoid accidents, children are not allowed to operate the equipment alone. When breathing oxygen, do not wrap the hose around the neck;: The elderly who can’t act easily need to be accompanied by a guardian.
  • It is forbidden to move the equipment by pulling the power cord or hose;
  • When the product works, oxygen is directly separated from the air. Please do not use it in the environment of lampblack and poisonous gas;
  • The product cannot be connected with other equipment from another manufacturer;
  • If any accessories are damaged or lost, please contact the supplier first;
  • It is forbidden to use an adapter that is not equipped with the product or does not meet the safety requirements of GB9706 and 1-2007 standards.
  • It is forbidden to use lithium batteries that are not equipped with the product.
  • When the battery is not in use, please pay attention to protecting the battery interface and the battery interface on the device. Do not touch the conductor or touch it directly by hand.
  • Remove the battery before the equipment is discarded, and contact the relevant departments for proper disposal of the equipment and accessories according to the relevant regulations of the local government.

Product Introduction

Product Principle
This product uses air as raw material, DC12V power supply or battery as a power source, adopts a high-quality French imported molecular sieve and small oil-free air compressor, and uses advanced pressure swing adsorption separation method (PSA method) at room temperature to extract high Purity oxygen; this method is a purely physical method which is stable, safe and reliable.

Scope of product application
It is only used for health and beauty, not for medical treatment.

Structural characteristics

  • Full plastic shell, safe and reliable;
  • With a built-in battery, it can achieve outdoor oxygen absorption;
  • Small mini size, lightweight, can be carried around;
  • It has the function of accumulating time which can understand the total time of machine use in real-time;
  • It has an oxygen concentration alarm function, which can prompt the user about the oxygen production status of the machine in real-time;

Technical Parameters

  1. Model: VP-01

  2. Power supply:

    • AC 100 ~ 240V 50/60HZ
    • DC 12 ~ 16.8V
  3. Power: 75 W

  4. Built-in battery capacity: 92 WH 6.4 AH

  5. Flow gear: 1 ~ 5 gear (5 gears have the highest flow)

  6. Oxygen concentration: 93% $ 3

  7. Noise level: <58dB(A) (the highest grade)

  8. Dimensions: 7.09″4.92″10.63″

  9. Net weight: 2.95 kg

  10. Electrical classification: Class II

  11. Oxygen supply mode: pulse oxygen supply

  12. Machine working mode: continuous operation

  13. Normal working environment:
    Ambient temperature range: 50° F~95° F;
    Relative humidity range: 30% ~ 75%;
    Atmospheric pressure range: 860hPa ~ 1060hPa,
    Note: When the storage and transportation temperature is lower than 5°C, the equipment should be stored at normal operating temperature for more than 4 hours before use.

  14. Storage and transportation environment:
    Ambient temperature range: -20°C ~ 60°C;
    Relative humidity range: 10% ~ 93%, and there is no frost phenomenon in the environment;
    Atmospheric pressure range: 700hPa ~ 1060hPa
    Note: The oxygen generator should be stored in a well-ventilated room without strong sunlight, or corrosive gas Inside, to avoid sudden shock and upside-down during transportation.

Installation and use

  1. Please Check The Package When Receiving The Package
    Before installing and using the product, please first check whether the appearance of the product is intact and whether the variety and quantity of accessories are consistent with the packing list in the user manual. If there is any defect, please contact the seller or manufacturer in time.

  2. Outline DiagramVARON-VP-1-Pulse-Flow-Oxygen-Concentrator- \(2\) VARON-VP-1-Pulse-Flow-Oxygen-Concentrator- \(3\)

  3. Introduction To Operation Panel FunctionVARON-VP-1-Pulse-Flow-Oxygen-Concentrator- \(4\)

  4. Button function introduction

  *  **Power Button:** Start or stop, Press the button for a short time and the machine will start to work, While the machine is working, press and hold the button for 3 seconds, the machine will stop working.Level: Adjust the output oxygen flow; Time: Adjust­able between 0~240 minutes
  *  **Grade:** Adjust output oxygen flow;
  *  **Time:** Setting device using time, adjustable from 0~240 minutes
2.  **Display content introduction**
  *  **Accumulated hours:** display how long the device has been working (unit: hour)
  *  **Timing/Min:** display the device and how many minutes the device are working.  

Remark: One screen shows Timing min & Accumulated hours. The default display is accumulated hours. When pushing the Timing/Min button, it is shifted to Timing minutes. If there is no new instruction, it is automatically switched back to the accumulative hour’s display after 10 seconds.

  *  **Gear:** Output oxygen flow gear, a total of five gears (the five gears have the largest flow) ![VARON-VP-1-Pulse-Flow-Oxygen-Concentrator- \(5\)](https://manuals.plus/wp-content/uploads/2024/02/VARON-VP-1-Pulse-Flow-Oxygen-Concentrator-5.jpg)
  * **Power indicator:** Indicates the current built-in battery power, calculated using voltage value, considering the actual usage, this item is for reference only;  

There are a total of four battery indicator lights on the operation panel:

    *  **100%** indicator light is on—-battery power≥90%;
    *  **75%** indicator light is on—-90% > battery power≥75%;
    *  **50%** indicator light is on—-75% > battery power≥50%;
    *  **25%** indicator light is on—-50% > battery power≥25%;
    * The 25% indicator flashes —- battery power <25%.
  *  **Indicator light and Alarm light :**  

Alarm light: The light is on when the device is operated abnormally, meanwhile will display the alarm code.

    *  **F0001—** Low Battery Warning ;
    *  **F0002-** No Breath Detected ;
    *  **F0011—** High Temperature with Alarm ;
    *  **F0012—** Oxygen Sensor Abnormal Alarm
  *  **Low Concentration:** When the oxygen concentration is lower than 82%,the light is on. When the oxygen concentration is lower than 50%, the light is on with alarm.
  *  **Breath Signal:** When the respiratory signal of the user is detected during the operation of the equipment, the light is on. Meanwhile, the device will automatically pulse oxygen supply.
  1. Use GuidanceVARON-VP-1-Pulse-Flow-Oxygen-Concentrator- \(6\)
  2. Connect power
    Select the appropriate power connection according to the use environment.
  *  **Use the built-in battery only**  

The device has a built-in battery installed by default. If the battery is sufficient, it can be turned on.

  *  **Use AC adapter**  

Connect the AC adapter and the AC power cord firmly, then connect the input end plug of the AC power cord to the indoor power strip, and finally connect the output end of the AC adapter to the charging port of the device. Connected, the power indicator on the AC adapter lights up red at this time, indicating that the power connection is normal and the built-in battery is charging; if the power indicator on the AC adapter lights up green, it indicates that the power supply is connected properly and the built-in battery is fully charged.
Requirement input terminal: AC voltage is 100- 240V, frequency 50/60Hz.

  *  **Use the car power cord in the car**  

Connect the input end of the car power cord to the car cigarette lighter, and then connect the output end of the car power cord to the charging port of the device. At this time, the red light of the car power cord lights up, indicating that the power supply is connected normally.
Requirement input terminal: DC voltage is DC 12~ 16.8V, current is 6A.

2.  **Connect disposable oxygen inhaler**  

Connect the flared end of the disposable oxygen inhaler tube firmly to the oxygen outlet of the device, and wear the other end concerning the figure below.

  1. Put on the disposable oxygen inhaler tube as shown in the figure, adjust the vent to a suitable position, and push the fixing ring upward after putting it on for fixation.
  2. Note that the tube of the disposable oxygen inhaler should not be kinked or blocked, so as not to cause equipment-related alarms and affect normal use.
  3. If you need to buy an oxygen inhaler yourself, please contact the equipment provider
  4. Selection under guidance.
3.  **Get started**  

Press the power button to start the machine, use the gear button to select the gear that suits you, and you can start to inhale oxygen after wearing the disposable oxygen inhaler. This device is a pulsed oxygen supply with a built- in breathing sensor. When the user inhales, the machine will automatically release oxygen and the “breathing signal” light will flash; when the user does not inhale, the machine will not provide oxygen.

4.  **Shut down**  

If you need to stop oxygen supply during use, press and hold the “ON key” for 3 seconds to shut down. After use, first unplug the horn of the disposable oxygen inhaler from the oxygen outlet of the device, then unplug the power cord port at the charging port of the device, and finally unplug the power plug to cut off the power.

USE TIPS

Battery Charging

  1. Due to transportation control and other reasons, the built-in battery of the device only retains power for testing when leaving the factory, or there is no power at all, so the device needs to be charged before first use.
  2. The built-in battery will only be charged automatically when the AC adapter is used; if the vehicle power cord is used, it can only supply power to the device and cannot charge the built-in battery.
  3. This device currently does not support the use of non-built­in batteries.
  4. The AC adapter and the car power cord will share the same power socket, so only one of them can be used at a time.
  5. When the power of the built-in battery is less than 25%, the power indicator will flash to remind you to connect the power adapter or charger to power the device.
  6. When the device is not working, the battery in the device will fully charge for about 4-8 hours. When the device is in operation, the charging time will be extended to 9 hours.
  7. When using the vehicle power cord in a car, turn on the engine. When the engine is off, the device cannot be used to prevent the battery from losing power in the car. And only the dedicated car power cord can be used.

Guidelines for Using Electromagnetic Environment
To prevent this machine from being interfered by a strong electromagnetic environment or other electrical products, please keep this machine and other household appliances at least 30cm apart during use.

Non-Professionals are not Allowed to Disassemble the Oxygen Generator Case
If professional maintenance personnel need the electrical schematic diagram of the product, please contact the manufacturer or after-sales service.

Use and Purchase Of Disposable Oxygen Inhaler

  1. This equipment is a pulse oxygen supply mode. To use this equipment, it must be connected to an oxygen inhaler.
  2. For the oxygen concentrator to correctly detect breathing and deliver pulsed oxygen, please ensure that the disposable oxygen inhaler is correctly installed and that the tube of the oxygen inhaler is not kinked or blocked.
  3. Do not use oxygen inhalers that are not compatible with this device. You need to purchase a suitable oxygen inhaler under the guidance of the equipment provider or professional medical staff to ensure the normal operation of the device and the normal use of the user.

MAINTENANCE

Clean the Shell
Wipe and clean the outside of the machine at least once a month: first cut off the power, and then wipe with a slightly damp cotton cloth or sponge. Be careful not to let liquid penetrate the cracks of the chassis.
After cleaning, place the machine in a ventilated place for 2 hours, and wait until the surface is completely dry before using it.

Cleaning the Disposable Oxygen Inhaler
The oxygen inhaler is cleaned once a week, and the nasal suction tip on the oxygen inhaler should be cleaned after each use.
It can be soaked in white vinegar for about 5 minutes and then rinsed with water, or wiped with medical alcohol. It is recommended that the oxygen inhaler be replaced every 2 months of use. (Note: The inside of the oxygen inhaler tube should be kept dry)

Battery Maintenance
If you do not use the machine for a long time, you need to charge the device once every three months to avoid the battery’s long-term low power and non-use which may cause a decrease in battery life. Each time the battery is charged to 25% to 50%.

PROBLEM & SOLVING

Malfunction Problems Solutions
Unable to Turn on Battery is exhausted Connect the adapter to charge the

battery
Poor power cord connection| Check the power connection
The battery is not installed correctly| Reinstall the battery correctly
Machine malfunction| Contact the seller
No Oxygen Output| The tube of the oxygen inhaler is knotted or blocked| Replace with a new oxygen inhaler
The oxygen inhaler is not tightly connected to the equipment, causing air leakage| Reinstall the oxygen inhaler correctly
Machine malfunction| Contact the seller
Smell in Oxygen| The oxygen  inhaler has not been cleaned for a long time, resulting in dirt in the tube| Clean or replace in time oxygen tube
The new oxygen inhaler will have a slight disinfection smell at the beginning of use, this is normal| The smell will disappear automatically after a few days of use
Machine malfunction| Contact the seller
Regular and Slight Exhaust Sound| The sound of nitrogen exhaust when the machine is producing oxygen| This is the normal action of the machine to produce oxygen and discharge nitrogen, which is a normal phenomenon

PACKING LIST

No. Item Amount
1 Oxygen Concentrator 1
2 AC Adapter 1
3 AC Power Cord 1
4 Car Power Cord 1
5 English Manual 1
6 Disposable Oxygen Inhaler 2
7 Warranty Card/Certificate 1
8 Backpack 1

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)