SUN series Quick Steam Sterilizer
General Introduction
This Autoclave is a precision instrument specializing in the operation
occasion in need of simple, fast and efficient sterilization by the doctor or
his/her assistants, and the doctor shall take the responsibility for such
work.
This Autoclave is equipped with the microprocessor for intelligent control and
man-machine interface for convenient operation. With the distinctions in
displaying parameter and condition actively, evaluating malfunction
automatically, shutting off for self-protection automatically in case of
excessive temperature and pressure, this type autoclave is reliable product in
sterilizing.
Application Scope
This product is N-level sterilizer for dental, ophthalmology, operating room, laboratory, small dental equipment, surgical instruments sterilization 135 ℃ for rapid sterilization of items
Tech. Parameter
Inner chamber of the quick steam sterilizer (Diameter X Depth) 125mm X 195mm
Rated Voltage………… 220V±22V,50Hz±1Hz ; 110V±11V,60Hz ± 1Hz
Rated Power………………………………………………………………500VA
Sterilizing:(Pressure/Temp.)……………………………………2.0~2.3bar/134℃
(Both of above are Under Standard Atmospheric pressure)
Fuse-Melting pipe………………………..…….…….…… T15A; T30A
Water Consumption volume per recycling round…………..……………..0.15L
Operating Temp……………………………………………………………0-40ºC
Diagram Specification
Attention and warning
I Sort B model device
**** Protective Grounding
Installation
The clearance between the exterior side surface of the autoclave and the
ambient neighboring articles shall be 10cm as retained, and the top surface
shall be 20cm retained.
The air-cooling window attached to the exterior surface of the autoclave shall
be non-clog by dust or articles, and the autoclave shall be located in the
proper air flowing environment by recommendation of the supplier.
The autoclave shall be installed on the horizontal working base.
PREPARATION BEFORE OPERATION
Before operation, please connect well to the power source. Connect the power
source hole under the back left side of the autoclave and the power socket by
the attached power cable inside the product’s package so as to switch on the
autoclave.
Push down the main power button at the right bottom corner of the front side,
when LCD display bright it means that the autoclave get through with the
electric power, and the process then is in initialization status with “LOAD”
being shown on the screen. The autoclave is not heated while under such
initialization status.
Note: If the boot display “DOOR”, need to open the door knob, so that the work status display “LOAD”.
Control Panel
6.1 Air pressure display
Display on time the air pressure inside the cavity. Unit: Bar
6.2 Temperature display
Display on time the temperature inside the cavity. Unit: ºC
6.3 Sterilization time display. Unit: Minute
6.4 Status/ malfunction code number display
Applied for displaying kinds of status of the autoclave during its working
time.
See more details in “Appendix working status display form”.
When alarm occurs to the autoclave, the digital pipe of the display will show
the alarm code number, the user can find out the malfunction part. (Once the
malfunction occurring to the autoclave, please contact the service center or
distributor immediately).
Operation
Before operation, please connect well to the power source. when LCD display
bright it means that the autoclave get through with the electric power, and
the process then is in initialization status with “LOAD” being shown on the
screen. The autoclave is not heated while under such initialization status.
Otherwise, you need to close the door by handle completely.Then press the
“Start” button to start the sterilizing.
First sterilizing need to fill no less than 150 ml of distilled water, after
that, each sterilizing just need to fill 30 ml of distilled water. (Note the
water level should keep at the position of supporting basket frame)
Note: please use distilled water, in order to prolong the service life of sterilization.
7.2 After modification, on the status of “READY”, press the “START” key to
start working.
Note: If the door is not completely closed, the machine can’t work when
press down the “START” button. If the door is opening during working process,
the working status display will flash “E5”as alarm and the autoclave will stop
working until restart again by switching off the power and dismissing the
alarm.
7.3 Status display “END”, and has a beep after sterilizing process, then you
can open the door and take the sterilized apparatus out.
Note: In case of scald, use the attached hand-support plate to take out
take out the sterilized apparatus.
After opening the door of autoclave, the process is reset at “LOAD” state,
which means the autoclave is under heat preservation status and waiting for
another sterilizing procedure.
7.4 The waiting time for next sterilizing is at least 10 minutes, and ensure
temperature decreasing enough.
7.5 Please shut off the main power supply when you don’t use the autoclave.
(The light ofpower is off).
Emergency Condition
During the running time, the autoclave will automatically raise the alarm,
release pressure, stop heating and display the warning code. (The definition
of the warning code is as the below: malfunction code and resolving method)
for ensuring the safety of the operator if the emergencies occur.
Any emergencies happen, please check the warning code table for resolving the
according malfunctions. If not possible to resolve, please contact the
dealers. We will try our best to provide help to you in short time.
Malfunction code and resolving measures
1| E1| Heating failure| Check the heating heater
Check inlet valve of water is normal
Check water level
Check for gas leaking away
---|---|---|---
2| E2| Failure to maintain the temperature| Check temperature sensor
Check inlet valve of water is normal
Check water level
3| E3| Excessive Temperature| Check temperature sensor
4| E4| Pressure can not be released| Check the exhaust, drain valve
5| E5| Door switch bounce| Check the door knob
Maintenance and Service
9.1 Clean the pure water tank by medical disinfectant once a week.
9.2 Disinfect and clean the interior surface of the autoclave room once a
month..
9.3 Replace the fuse-melter.
(1) Shut off the power
(2) Revolve the fuse base by counter-clockwise by screwdriver to bring out the
melting fuse to be replaced;
(3) Replace the new fuse and reset the fuse base on the original position,
then use the screwdriver to fasten the base by clockwise.
(4) Check the correctness of the parameter of the new fuse replaced.
9.4 Clean the sealing ring periodically
The user shall clean the sealing ring periodically for avoiding the influence
to the seal caused by the dust/dirt left due to the long period use. Use the
smooth cloth with distilled water to wipe the surface of the sealing ring or
sealing cap gently.
The user shall discharge the sealing ring for further cleaning or replacement
if the air-leaking problem can not be resolved after the above process (The
user shall discharge the sealing ring periodically for cleaning and inspection
as recommendation by the supplier). Put back the ring to the original door
groove after cleaning. Most Important: the inset work must be done equably to
the groove During the installation, the laid four equal points of the ring
must be inserted to the groove firstly, later for the same work for the
remaining segments of the ring.
After finishing, press the sealing ring equally by the force of hand.
9.5 Shut off the power before service or components’ replacement, and the
service or replacement work shall be done by the supplier or his designated
technician.
Transportation and Storage
The autoclave shall be transferred and stored with the below conditions:
Ambiance temperature: -40ºC -+55ºC
Relative humidity: < 85%
Atmospheric pressure: 500HPA-1060HPA
APPENDIX
I. Working status display table:
NO. | Display | Explain |
---|---|---|
1 | DOUR. | Note opening the door |
2 | LOAD | Loading status, prepare working |
3 | READY | In this state you can start the sterilization process |
4 | END | That the normal sterilization process is completed |
5 | T | That temperature, unit: °C |
6 | P | That pressure, unit: kgflcm2 |
II. Steam Way Map
III. Wiring Map