BRAUN IRT 6020 Thermo Scan Ear Thermometer User Manual

June 17, 2024
Braun

BRAUN IRT 6020 Thermo Scan Ear Thermometer

Product description

IRT 6020/6500 Series

Product Description

  1. Lens filter (Box of 20)
  2. Probe tip
  3. Probe
  4. ExacTemp light
  5. Lens filter ejector
  6. Display
  7. Power button
  8. Start button
  9. Battery door
  10. Protective cap (IRT 6020 only)
  11. Protective case (IRT 6500 only)

IRT 6510 Series

Product Description

  1. Lens filter (Box of 20)
  2. Probe tip
  3. Probe
  4. ExacTemp light
  5. Lens filter ejector
  6. Display
  7. Memory button
  8. Power button
  9. Start button
  10. Battery door
  11. Protective case

The Braun ThermoScan thermometer has been carefully developed for accurate, safe and fast temperature measurements in the ear.
The shape of the thermometer probe prevents it from being inserted too far into the ear canal which can hurt the eardrum.
However, as with any thermometer, proper technique is critical to obtaining accurate temperatures.
Therefore, read the instructions carefully and thoroughly.
The Braun ThermoScan thermometer is intended for intermittent measurement and monitoring of human body temperature for people of all ages. It is intended for household use only.
Use of this thermometer is not intended as a substitute for consultation with your physician.

WARNINGS AND PRECAUTIONS

  • Please consult your doctor if you see symptoms such as unexplained irritability, vomiting, diarrhea, dehydration, changes in appetite or activity, seizure, muscle pain, shivering, stiff neck, pain when urinating, etc. in spite of absence of fever.
  • Please consult your doctor if the thermometer shows elevated temperature.

The operating ambient temperature range for this thermometer is 50 – 104 °F (10 – 40 °C).
Do not expose the thermometer to temperature extremes (below –13 °F / –25 °C or over 131 °F / 55 °C) or excessive humidity (> 95 % RH). This thermometer must only be used with genuine Braun ThermoScan lens filters (LF 40).

To avoid inaccurate measurements always use this thermometer with a new, clean lens filter attached.

If the thermometer is accidentally used without a lens filter attached, clean the lens
(see «Care and cleaning» section). Keep lens filters out of reach of children.

This thermometer is intended for household use only. Use of this thermometer is not intended as a substitute for consultation with your physician.

This thermometer is not intended for pre-term babies or small-for-gestational age babies.
It is not intended to interpret hypothermic temperatures. Do not allow children under 12 years to take their temperature unattended.

Do not modify this equipment without authorization of the manufacturer.

Parents/guardians should call the pediatrician upon noticing any unusual sign(s) or symptom(s). For example, a child who exhibits irritability, vomiting, diarrhea, dehydration, seizure, changes in appetite or activity, even in the absence of fever, or who exhibits a low temperature, may still need to receive medical attention.

People who are on antibiotics, analgesics, or antipyretics should not be assessed solely on temperature readings to determine the severity of their illness.

Temperature elevation may signal a serious illness, especially in neonates and infants, or in adults who are old, frail, or have a weakened immune system. Please seek professional advice immediately when there is a temperature elevation and if you are taking temperature on:

  • Neonates and infants under 3 months (Consult your physician immediately if the temperature exceeds 99.4 °F or 37.4 °C).
  • Patients over 60 years of age (Fever may be blunted or absent in older patients).
  • Patients having diabetes mellitus or a weakened immune system (e.g., HIV positive, cancer chemotherapy, chronic steroid treatment, splenectomy)
  • Patients who are bedridden (e.g., nursing home patient, stroke, chronic illness, recovering from surgery)
  • A transplant patient (e.g., liver, heart, lung, kidney).

This thermometer contains small parts that can be swallowed or produce a choking hazard to children.
Keep out of reach of children under 12 years.

Body temperature

Normal body temperature is a range. It varies by site of measurement, and it tends to decrease with age. It also varies from person to person and fluctuates throughout the day.
Therefore, it is important to determine normal temperature ranges. This is easily done using Braun ThermoScan. Practice taking temperatures on yourself and healthy family members to determine the normal temperature range.

Note: When consulting your physician, communicate that the ThermoScan temperature is a temperature measured in the ear and if possible, note the individual’s normal ThermoScan temperature range as additional reference.

How does Braun ThermoScan work?

Braun ThermoScan measures the infrared heat generated by the eardrum and surrounding tissues. To help avoid inaccurate temperature measurements, the probe tip is warmed to a temperature close to that of the human body. When the Braun
ThermoScan is placed in the ear, it continuously monitors the infrared energy. The measurement is finished and the result displayed when the thermometer detects that an accurate temperature measurement has been taken.

Why measure in the ear?

The goal of thermometry is to measure core body temperature1 which is the temperature of the vital organs. Ear temperatures accurately reflect core body temperature2, since the eardrum shares blood supply with the temperature control center in the brain3, the hypothalamus. Therefore, changes in body temperature are reflected sooner in the ear than at other sites. Axillary temperatures measure skin temperature and may not be a reliable indicator of core body temperature.
Oral temperatures are influenced by drinking, eating and mouth breathing. Rectal temperatures often lag behind changes in core body temperature and there is a risk of cross-contamination.

How to use your Braun ThermoScan

  1. Remove the thermometer from the protective case.

Remove protective cap.

  1. Push the Power button .
    During an internal self-check, the display shows all segments. Then the last temperature taken will be displayed for 5 seconds.

  2. The lens filter indicator will blink to signal one is needed.
    To achieve accurate measurements, make sure a new, clean lens filter is in place before each measurement.
    Attach a new lens filter by pushing the thermometer probe straight into the lens filter inside the box and then pulling out.
    How to Use Your Braun ThermoScan
    Note: The Braun ThermoScan will not work unless a lens filter is attached.
    The thermometer is ready to take a temperature when the display looks like the images below.

  3. Guyton A C, Textbook of medical physiology, W.B. Saunders, Philadelphia, 1996, p 919

  4. Guyton A C, Textbook of medical physiology, W.B. Saunders, Philadelphia, 1996, p 754-5

  5. Netter H F, Atlas of Human Anatomy, Novartis Medical Education, East Hanover, NJ, 1997, pp 63, 95.

  6. Fit the probe snuggly into the ear canal, then push and release the Start button .

  7. ExacTemp light will pulse while temperature is in progress.
    The light will remain solid for 3 seconds to indicate that a successful temperature reading has been achieved.
    NOTE: If the probe has been properly inserted into the ear canal during the measurement, a long beep will sound to signal the completed measurement.
    If the probe has NOT been constantly placed in a stable position in the ear canal, a sequence of short beeps will sound, the ExacTemp light will go out and the display will show an error message (POS = position error).
    See «Errors and troubleshooting» section for more information

  8. The confirmation beep indicates that an accurate temperature measurement has been taken. The result is shown on the display.

  9. For the next measurement, press Eject button to remove and discard used lens filter, and put on a new, clean lens filter.
    The Braun ThermoScan ear thermometer turns off automatically after 60 seconds of inactivity.
    The thermometer can also be turned off by pressing the Power button .
    The display will briefly flash OFF and it will go blank.

Temperature taking hints

Always replace disposable lens filters to maintain accuracy and hygiene. The right ear measurement may differ from the measurement taken in the left ear. Therefore, always take the temperature in the same ear.
The ear must be free from obstructions or excess earwax build-up to take an accurate reading.

External factors may influence ear temperatures, including:

Factor Yes affects
Poor probe placement
Used lens filter
Dirty lens

In the cases below, wait 20 minutes prior to taking a temperature.

Factor Yes affects
Extreme hot and cold room temperature
Hearing aid
Lying on pillow

Use the untreated ear if ear drops or other ear medications have been placed in the ear canal.

Memory mode

  1. The last temperature taken is stored in the memory and will be automatically displayed for 5 seconds when it is turned on again.
    Memory Mode
    Available only in
    IRT 6510

  2. This model stores the last 9 temperature measurements.
    To display the stored measurement, the thermometer must be turned on.

  3. Press the Mem button.
    The display shows the temperature, and when releasing the Mem button, the stored temperature for that memory number is displayed, together with “M”.

  4. Each time the Mem button is pushed, the stored reading and an M is displayed to indicate each temperature measurement (eg. M2).
    Memory mode is automatically exited by not pressing the memory button for 5 seconds.
    Memory Mode

Night light feature

Available only in
IRT 6510

This thermometer includes a convenient night light to illuminate the display in a dim environment.

The light will turn on when you press any button. It will stay on until the thermometer is inactive for up to 15 seconds even after temperature is taken.

Changing the temperature scale

If you wish to change the temperature scale of your thermometer:

  1. Make sure the thermometer is turned off.
  2. Press and hold down the Power button . After about 3 seconds the display will show this sequence: ° C / SET / ° F / SET …
  3. Release the Power button when the desired temperature scale is shown.
    There will be a short beep to confirm the new setting, then the thermometer is turned off automatically.

Care and cleaning

 The probe tip is the most delicate part of the thermometer. It must be clean and intact to ensure accurate readings. If the thermometer is ever accidentally used without a lens filter, clean the probe tip as follows:

Very gently wipe the surface with a cotton swab or soft cloth moistened with alcohol. After the alcohol has completely dried out, you can put a new lens filter on and take a temperature measurement.

If the probe tip is damaged, please contact Consumer Relations.

Use a soft, dry cloth to clean the thermometer display and exterior. Do not use abrasive cleaners.
Never submerge this thermometer in water or any other liquid. Store thermometer and lens filters in a dry location free from dust and contamination and away from direct sunlight.

Additional lens filters (LF 40) are available at most stores carrying Braun ThermoScan.

Replacing the batteries

The thermometer is supplied with two 1.5 V type AA (LR 06) batteries.
For best performance, we recommend Duracell® alkaline batteries.

  1. Insert new batteries when the battery symbol appears on the display.
  2. Open the battery compartment. Remove the batteries and replace with new batteries, making sure the poles are in the right direction.
  3. Snap battery door into place.

 To protect the environment, dispose of the product and empty batteries at your retail store or at appropriate collection sites according to national or local regulations.

Calibration

The thermometer is initially calibrated at the time of manufacture. If this thermometer is used according to the use instructions, periodic re-adjustment is not required. If at any time you question the accuracy of temperature measurements, please contact Consumer Relations.

Manufacturing date is given by the LOT number located inside the battery compartment. The first three (3) digits after LOT represent the Julian date that the product was manufactured and the next two (2) digits represent the last two numbers of the calendar year the product was manufactured. The last identifiers are the letters that represent the manufacturer.

An example: LOT 11614KTC, this product has been manufactured on the 116th day of the year 2014.

Errors and troubleshooting

Error Message Situation Solution

| No lens filter is attached.| Attach new, clean lens filter.

| The thermometer probe was not positioned securely in the ear.
An accurate measurement was not possible.
POS = position error| Take care that the positioning of the probe is correct and remains stable.
Replace lens filter and reposition.
Press Start button to begin a new measurement.

| Ambient temperature is not within the allowed operating range (50 – 104 °F or 10 – 40 °C ).| Allow the thermometer to remain for 30 minutes in a room where the temperature is between 50 and 104 °F or 10 and 40 °C.

| Temperature taken is not within typical human temperature range (93.2 – 108 °F or 34 – 42.2 °C).
HI = too high
LO = too low| Make sure the probe tip and lens are clean and a new, clean lens filter is attached.
Make sure the thermometer is properly inserted. Then, take a new temperature.

| System error – self-check display flashes continuously and will not be followed by the ready beep and the ready symbol.
If error persists,
If error still persists,| Wait 1 minute until the thermometer turns off automatically, then turn on again.
… reset the thermometer by removing the batteries and putting them back in.
… please contact Customer Service.

| Battery is low, but thermometer will still operate correctly.| Insert new batteries.

| Battery is too low to take correct temperature measurement.| Insert new batteries.

Product specifications

Displayed temperature range: 93.2 – 108 °F (34 – 42.2 °C)
Operating ambient temperature range: 50 – 104 °F (10 – 40 °C)
Display resolution: 0.1 °F or °C

Accuracy for displayed temperature range / Maximum Laboratory Error
95 °F – 107.6 °F (35 °C – 42 °C): ± 0.4 °F ± 0.2 °C
outside this range: ± 0.5 °F ± 0.3 °C
clinical repeatability: ± 0.26 °F ± 0.14 °C

Long term storage ranges
Storage/transport temperature: -13 °F to 131 °F (-25 °C to 55 °C)
Humidity: 15-95% non condensing
Battery life: 2 years / 600 measurements
Service life: 5 years

This thermometer is specified to operate at 1 atmospheric pressure (700-1060hPA) or at altitudes with an atmospheric pressure up to 1 atmospheric pressure (700-1060hPA).

  • Equipment with type BF applied parts
  • See instruction for use
  • Operating temperature
  • Keep it dry
  • Attention, consult ACCOMPANYING DOCUMENTS

Subject to change without notice.

This appliance conforms to the following standards:

Standard Reference Edition Title:
EN 12470-5: 2003 Clinical thermometers – Part 5: Performance of infra-red ear thermometers (with maximum device).

EN 60601-1: 2006 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance.

EN ISO 14971: 2012 Medical devices – Application of risk management to medical devices.

EN ISO 10993-1: 2009 Biological evaluation of medical devices – Part 1: Evaluation and Testing within a risk management process.

EN 60601-1-2: 2007 Medical electrical equipment – part 1-2: General requirements for basic safety and essential performance

  • Collateral standard: electromagnetic compatibility
  • Requirements and tests

EN 980: 2008 Symbols for use in labeling of medical devices.

EN 1041: 2008 Information supplied by the manufacturer of medical devices.

EN 60601-1-11: 2010 Medical electrical equipment — Part 1-11: General requirements for basic safety and essential performance — Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.

This product conforms to the provisions of the EC directive 93/42/EEC.

MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC. For detailed description of EMC requirements please contact Customer Service.

Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT.

Please do not dispose of the product in the household waste at the end of it useful life.
To protect the environment, dispose of empty batteries at appropriate collection sites according to national or local regulations.

Limited three year warranty

Kaz, Inc. warrants this product against any defects that are due to faulty material or workmanship for a period of three years from the original date of consumer purchase or receipt as a gift. This warranty applies when used for normal household use in accordance with the Use and Care Manual and excludes the battery and damage to the product resulting from accident or misuse. This product is not warranted when used in a professional environment.

In no event shall Kaz, Inc. be liable for any special, incidental, indirect or consequential damages in connection with the purchase or use of this product or costs over the original cost of the product.

If the product should not perform to specifications within the warranty period, please contact Consumer Relations. In Canada, you can contact at 510 Bronte Street S, Milton, ON, L7T 2X6, Canada.
Kaz, Inc. reserves the right to replace a non conforming product with the most comparable product currently available. This product has a five year minimum expected lifetime.

While this warranty gives you specific legal rights, you may also have other rights which vary from state to state/province to province.

Guidance and manufacturer’s declaration electromagnetic emissions

The ME equipment is intended for use in the electromagnetic environment specified below.

The customer or the user of the ME equipment should ensure that it is used in such an environment.

Emissions Test| Compliance| Electromagnetic environment guidance
RF Emissions CISPR 11| Group 1| The ME equipment uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF Emissions CISPR 11| Class B| Complies
Harmonic emissions IEC 61000-3-2| Not Applicable| The ME equipment is solely battery powered.
Voltage fluctuations / flicker emissions| Not Applicable
Non-Life Support Equipment Separation Distance Calculation (3Vrms / 3V/m compliance)

Rated maximum output power of transmitter (W)| Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz in ISM bands
| 80 MHz to 800 MHz
| 800 MHz to 2.5 GHz

0.01| 0.12| 0.12| 0.23
0.1| 0.37| 0.37| 0.74
1| 1.17| 1.17| 2.33
10| 3.69| 3.69| 7.38
100| 11.67| 11.67| 23.33
Guidance and manufacturer’s declaration electromagnetic immunity

The ME equipment is intended for use in the electromagnetic environment specified below.

The customer or the user of the ME equipment should ensure that it is used in such an environment.

Immunity test| IEC 60601 test level| Compliance level| Electromagnetic environment guidance
Electrostatic discharge (ESD) IEC 61000-4-2| ±6kV Contact
±8kV Air| Complies| Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Radiated RF IEC 61000-4-3| 3 V/m 80MHz to 2.5GHz| Complies| Field strengths outside the shielded location from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than 3 V/m.
Interference may occur in the vicinity of equipment marked with the following symbol:

Separation distance calculation provided above. If a known transmitter is present the specific distance can be calculated using the equations.
Conducted RF IEC 61000-4-6| 3Vrms 150kHz to 80MHz| Not Applicable (no electrical cabling)
Electrical fast transient IEC 61000-4-4| ±2kV power line
±1kV I/O lines| Not Applicable| The ME equipment is solely battery powered.
Surge
IEC 61000-4-5| ±1kV differential
±2kV common| Not Applicable
Power frequency magnetic field IEC 61000-4-8| 3 A/m| Complies| Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power supply input lines
IEC 61000-4-11|

95% dip 0.5 cycle
60% dip 5 cycles
70% dip 25 cycles
95% dip 5 sec.

| Not Applicable| The ME equipment is solely battery powered.

Contact US

Questions or Comments?
Call: 1-800-327-7226
eMail: consumerrelations@kaz.com
Visit: www.kaz.com/braun
Write: Consumer Relations, Kaz USA, Inc.,
250 Turnpike Road, Southborough, MA 01772, USA

This product is manufactured by Kaz USA, Inc., under license to the ‘Braun’ trademark.
‘Braun’ is a registered trademark of Braun GmbH, Kronberg, Germany.
ThermoScan is a registered trademark of Kaz USA, Inc.

 Kaz USA, Inc.
250 Turnpike Rd.
Southborough, MA 01772
© 2014 Kaz USA, Inc. All rights reserved.
www.kaz.com/braun

Patents / brevets / patentes:
www.kaz.com/patents/braun

Imported by
Kaz Canada, Inc.
510 Bronte St S.
Milton, Ontario L9T 2X6, Canada

Designed in Germany. Made in Mexico.

IRT6020/6500/6510
31IMIR65190
02.26.14

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals